Generator for thermal tests

ALPP-RTH200

High accuracy

High stability

Ranges 2A – 20A – 200A

Links RS485 and RS232

Control via PLC

Overview :
The RTH200 generator is primarily intended for testing thermal tripping of circuit breakers in
production . The current setpoint is transmitted to the generator via digital RS485 link. The
transmission functions of the current, measurement of the current and continuity check are directly
accessible on the logic inputs fit most PLCs. The source also returns to the PLC the informations of
the source states.

Specifications :

Output electrical specifications


Output type DC
Output voltage 6Vdc max
Output current Adjustable from 0 to 200Adc
Output power 1200W max
Ranges 2A / 20A / 200A
Network regulation 0.05%
Load regulation (0-100%) 0.1%
Generator accuracy 0.15% + 0.01% of range
Time accuracy 1ms
Typical rise time 20ms without overshoot

ELECTRONA – 7 allée de l'Europe – 67960 ENTZHEIM – Tél : (33) 3 88 10 30 40 – Fax : (33) 3 88 10 30 49 – www.electrona.fr
Programm / Read
Résolution courant de sortie 5.10-5 de la gamme
Résolution durée de génération 1ms
Précision de mesure courant 0.25%

Input electrical specifications


Supply type AC single phase
Supply voltage 230Veff +/-10%
Supply frequency 50/60Hz
Input protection Fuse 5x20 250V/16A T
Input current < 10Aeff

Temperature and cooling


Operating temperature 0°C - +40°C
Storage temperature -10°C - +55°C
Cooling Forced air ventilation (suction Front)

Isolation et diélectrique
Insulation imput / mass >100Mohms – 500Vdc
Insulation output / mass >100Mohms – 500Vdc
Dielectric strength input / output 2500Veff – 10s
Dielectric strength open output 2500Veff – 10s

Mechanical specifications
Dimensions without connection and handles 19’’x H134mm x P460mm
Weight 20kgs / 44Lbs