You are on page 1of 9

Pressure Calibration

Application Note

Pressure instrumentation is found These pressure calibrators


in virtually every process plant. accurately measure pressure
Periodic calibration of these by using:
pressure, level, and flow instru- • Internal sensors, or
ments is required to keep plants • External Pressure Modules
operating efficiently and safely.
Fluke provides a broad range of A pressure source may be
pressure calibration tools to help provided by:
you quickly and reliably calibrate • A self-contained internal
your pressure instrumentation. pressure pump, or
• An external source such as an
accessory pump or a
pressure bottle / regulator.
A summary of the pressure
calibration capabilities of Fluke
Process Tools is shown below.

725 and 726 Multifunction Process Calibrators

753 Documenting Process Calibrator

754 Documenting Process Calibrator


719 Electric Pressure Calibrator
Pressure Calibration
with
717 Pressure Calibrator

718 Pressure Calibrator


Fluke Calibrators

5520A Calibrator
Fluke 719 Electric Pressure Calibrator
Function
Measure Pressure to 10,000 psi /
700 bar with internal sensor •

Measure Pressure to 300 psi / 20 bar
with internal sensor •

Measure Pressure to 10,000 psi / 700 bar
with Fluke 700Pxx Pressure Modules • • • • • • •
Pressure switch test • • • • • •
Source pressure with accessory pumps • • • • • • •
Source pressure with built-in pump • •
Measure mA • • • • • •
Source • • • • •
Loop power supply • • • • •
Multifunction source and measure • • •
Electronic data capture • •
Communication to PC • • •
Integrated HART communication •


How to use the 718 to calibrate a pressure switch
1. Depressurize and isolate the 6. Release the pressure slowly
pressure switch from the pro- using the fine adjust vernier
cess. until RCL is displayed.
2. Plumb the 718 and make con- 7. Press the Switch Test but-
nections as per the ton once to read the pressure
illustration. values for switch opening and
3. Turn on the 718 and open again to see the pressure at
the vent valve. Press the Zero switch closing.
button to clear the zero offset. 8. Press and hold the Switch Test
Close the vent. button for 3 seconds to clear
4. Press the Switch Test button the test results and start over.
to enter the switch test mode. 9. Adjust the pressure switch
5. Apply pressure slowly with the setpoint until the switch
hand pump until you contacts open and close at
Pressure switch calibration approach the setpoint. Using the desired pressure.
Verify the setpoint and deadband of pressure the fine adjust vernier adjust
switches using the 718 Pressure Calibrators. the pressure until the switch
opens and OPEN is displayed
on the 718.

Typical Pressure Applications


How to calibrate a P / I Transmitter
Close the bleed valve. Set the
pressure / vacuum valve to +,
for positive pressure. Innovative new pump design
RE R
SSU
PRE IBRATO
CAL
718
HOL
D
Black

DAMP

5. Use the hand pump to apply Red

• Pumps can be easily contaminated with


UNITS

roughly 3 psi to the transmit-


P
LOO ER CLR .
POW ADJ
TRIC
MIN OME
HOLD
BAR

process fluids
O MAX
ZER

ter. Partial pump strokes will


mA

GE Pa
RAN

– Often requires repair


06k
i 0-2
0-30ps 30V X
MA
COM
S I G N A L

apply small increments of


kPa
si 620
90p MAX
+ –

• New design reduces repairs and cost


pressure. Use the fine-adjust
T E S T

of ownership
knob to get reasonably close to
3.00 psi. Has two clean out ports!
6. Press the HOLD key, and record • Remove fluids, clean with a cotton swab
the psi and mA readings. Press • Easy access, can be servied in the field
the HOLD key to resume read-
With a built-in hand pump, ing.
precision measurement of both 7. Calculate and record the error,
pressure and current, and a 24 using: Error = ([ (i-4)/16] –
volt loop supply, the Fluke 718 [(P-3)/12]) * 100 where
Pressure Calibrator is a complete, Error is in % of span, i is your
self-contained tool for the calibra- measured current in mA and
tion of P/I transmitters. To calibrate P is your measured pressure
a 3-15 psi / 4-20 mA transmitter in psi.
using a Fluke 718 30G Pressure 8. Repeat steps 5 through 7 at
Calibrator: mid-range, around 9 psi, to
1. Depressurize the transmitter, check linearity at mid-span.
and then plumb the transmitter 9. Repeat steps 5 through 7, now
to the 1/8 inch NPT pressure at 15 psi, for a check at 100 %
port of the 718. Connect the of span.
test leads per the figure above.
2. Turn the calibrator on. (If you If your calculated errors are
need to power the transmitter, within tolerance, the transmitter
hold down the UNITS key while has passed your As-found test,
turning on the calibrator.) and you are done. If necessary,
3. Press the UNITS key until PSI perform your zero and span
shows in the display. adjustments, then repeat steps
4. With the 718’s bleed valve 5 through 9 for an As-left test.
open to atmosphere, press the Depressurize the line, and
ZERO key. disconnect the 718.

2 Fluke Corporation Pressure Calibration


Intrinsic Safe Version Available

Measuring less than nine inches in length are available. The 718 can also measure
and weighing just over two pounds, the pressure using any of the 29
rugged 718 is easy to carry into the field. Fluke 700Pxx Pressure Modules, to cover
The 718 is offered in 1 psi, 30 psi, 100 psi applications up to 10,000 psi. The 718
and 300 psi models. Media compatibility is comes complete with protective holster, test
dry air and non-corrosive gasses. A built-in leads, test clips, Users Manual, and
pump generates pressure or vacuum. Min, two 9-volt batteries (installed).
Max, Hold and error % calculator functions

Functional Pressure Specifications


Model Range Resolution Over Pressure Functions
717 1G -1 PSI to 1 PSI, (-68.9 mbar to 0.001 psi, 0.001 mbar Over Pressure 5xFS Zero, Min, Max, Hold, Damp
68.9 mbar, -6.89 to 6.89 kPa)
717 15G -12 PSI to 15 PSI, (-850 mbar to 0.001 psi, 0.1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
1.03 bar, -85 to 103 kPa)
717 30G -12 PSI to 30 PSI, (-850 mbar to 2 0.001 psi, 0.1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
bar, -85 to 206.84 kPa)
717 100G -12 PSI to 100 PSI, (-850 mbar to 0.01 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
6.895 bar, -85 to 689.48 kPa)
717 300G -12 PSI to 300 PSI, (-850 mbar to 0.01 psi, 1 mbar Over Pressure 375 PSI, Zero, Min, Max, Hold, Damp
20.68 bar, -85 to 2068.4 kPa) 25 bar
717 500G 0 PSI to 500 PSI, (0 mbar to 0.01 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
34.47 bar, 0 to 3447.4 kPa )
717 1000G 0 PSI to 1000 PSI, (0 mbar to 0.1 psi, 1 mbar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
68.95 bar, 0 to 6894.8 kPa )
717 1500G 0 PSI to 1500 PSI, (0 mbar to 0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
103.42 bar, 0 to 10342 kPa)
717 3000G 0 PSI to 3000 PSI, (0 mbar to 0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
206.84 bar, 0 to 20684 kPa)
717 5000G 0 PSI to 5000 PSI, (0 mbar to 0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
344.74 bar, 0 to 34474 kPa)
717 10000G 0 PSI to 10,000 PSI, (0 mbar to 0.1 psi, 0.01 bar Over Pressure 2xFS Zero, Min, Max, Hold, Damp
700 bar, 0 to 70,000 kPa)

Pressure mA
Max non-
718, 719 destructive
Models Range Resolution Accuracy pressure Range Resolution Accuracy
718-1G -1 psi to 1 psi .001 psi 0.05 % FS 5X full scale 0-24 mA 0.001 mA 0.015 % + 2 counts
(-68.9 mbar to 68.8 mbar) (0.001 mbar) Measure
718-30G -12 psi to 30 psi 0.001 psi 0.05 % FS 2X full scale 0-24 mA 0.001 mA 0.015 % + 2 counts
(-850 mbar to 2 bar) (0.01 mbar) Measure

718-100G -12 psi to 100 psi 0.01 psi 0.05 % FS 2X full scale 0-24 mA 0.001 mA 0.015 % + 2 counts
(-850 mbar to 7 bar) (0.01 mbar) Measure

718-300G -12 to 300 psi 0.01 psi 0.05 % FS 375 psi or 0-24 mA 0.001 mA 0.015 % + 2 counts
(-850 mbar to 20 bar) (1 mbar) 25 bar Measure
719-30G -12 psi to 30 psi 0.001 psi 0.025 % FS 2X full scale 0-24 mA 0.001 mA 0.015 % + 2 counts
(-850 mbar to 2 bar) (0.01 mbar) Measure
Source
719-100G -12 psi to 100 psi 0.01 psi 0.025 % FS 2X full scale 0-24 mA 0.001 mA 0.015 % + 2 counts
(-850 mbar to 7 bar) (0.01 mbar) Measure
Source

Specifications are based on a one year calibration cycle and apply for ambient temperature from +18 °C to +28 °C. “Counts” are the number of increments or decrements
of the least significant digit. 719 accuracy specification is for 6 months.

3 Fluke Corporation Pressure Calibration


754 DOCUMENTING PROCESS CALIBRATOR

Isolation +
Valve
MULTIFUNCTION
725 PROCESS CALIBRATOR

Pressure
Module
Pressure
V mA
LOOP Module
V mA
MEAS
SOURCE
TC RTD C/F

STORE
SETUP 100%

RECALL 25%

25%

Red
0%

Black

Measuring pressure I to P device calibration


To measure pressure, the appropriate pressure module The I to P device is used to convert 4 mA to 20 mA
for the pressure to be tested is attached to the calibrator. electrical analog loop control to pneumatic analog loop
The measured pressure can be displayed in a variety of control, generally 3 psi to 15 psi. Here, a typical configu-
engineering units. A Fluke 725 or 726 multifunction pro- ration for using a pressure module with a 750 Series DPC
cess calibrator could be used here. is demonstrated.

Pressure
Module
754 DOCUMENTING PROCESS CALIBRATOR
Isolation +
754 DOCUMENTING PROCESS CALIBRATOR

Valve

Hand
Pump

Differential
Module
L H

Sourcing pressure Differential measurements


To calibrate an instrument with pressure input, pressure Differential pressure modules are useful in a wide variety
from an external source (such as a hand-held pump) is of applications, e.g., measuring the fluid level in a tank
applied. Prompts on the 750 Series Calibrator display or calibrating a differential pressure transmitter. A Fluke
indicate when to increase or decrease the input pressure, 754 Documenting Process Calibrator is shown.
and when the specified test points are achieved. Here, a
Fluke 754 Documenting Process Calibrator is shown.

Hand Pump
754 DOCUMENTING PROCESS CALIBRATOR

Pressure Module

Red

Black

P to I device calibration Pressure switch calibration


The P to I device is used to convert pneumatic analog Verify and document the setpoint and deadband of pressure
loop control signals of 3 psi to 15 psi to electrical loop switches using the 750 Series Documenting Process Calibrators.
analog control signals of 4 mA to 20 mA. Here, a
Fluke 717 Pressure Calibrator is used.

4 Fluke Corporation Pressure Calibration


Pressure Modules
Pressure module This innovative technique pro-
performance vides several benefits:
Fluke 700 Series pressure mod- 1. Digital communication elimi-
ules are highly accurate, with nates errors due to poor
total specifications that apply connections and electrical
from 0 % to 100 % of full span interference.
and from 0 °C to 50 °C (32 °F to 2. The modules are inherently
122 °F)—a feature that sets them temperature-compensated from
apart from other pressure cali- 0 °C to 50 °C (32 °F to 122 °F).
brators. Many ranges have total 3. The modules are fully inter-
uncertainties of 0.05 % of full changeable because all
scale and reference uncertainties measurements are completed
of 0.025 % of scale (see Table, in the pressure module itself
page 5). and then communicated to
This performance is possible the calibrator in digitized
through the innovative applica- form. Modules are calibrated
A complete family of tion of mathematics and micro- independently of the calibrator,
pressure modules processor power. Fluke pressure and can be used with any 700
A family of 29 pressure modules modules have silicon piezo- Series calibrator. Each module
covers the most common pressure resistor sensors which consist has its own serial number to
calibrations from 0-1˝ H2O of a resistive bridge fabricated facilitate traceability.
(0-.25 kPa) to 0-10,000 psi in a silicon diaphragm. Pressure
(0-70,000 kPa). applied to the diaphragm causes Sensor protection in
Gage pressure modules have a change in the balance of the isolated modules
one pressure fitting and mea- bridge which is proportional to Many of these modules (see
sure the process pressure with the applied pressure. The bridge Table) incorporate a stainless
respect to atmospheric pressure. balance change is not linear and steel diaphragm to isolate the
Differential pressure modules is very sensitive to temperature. sensor. With these modules, any
have two pressure fittings and However, since these effects medium that is compatible with
measure the difference between are quite stable with time and stainless steel can be used on the
the applied pressure on the high through repetitive changes of high side of the module.
fitting versus the low fitting. condition, the sensors can be
Each module is clearly labeled for very accurate in measuring pres- Rugged construction
range, overpressure specification, sure provided they are carefully A urethane overmolding protects
and media compatibility. A metric characterized. against shock if a module is ac-
adapter is included with all but During manufacture, Fluke cidentally dropped and also seals
the P29 through P31 high pres- pressure modules are character- against dirt, dust, and moisture.
sure modules. ized by reading temperature and Pressure connections are 1/4˝
pressure at more than 100 points. NPT. A BSP/ISO adapter is also
Quick and easy A least-squares regression is used provided on all but the P29, P30
measurements to calculate the coefficients of a and P31.
Fluke 700 Series pressure polynomial expression for pres-
modules are easy to operate. To sure. The coefficients, unique to
Convenient setup
measure pressure, the technician A one-meter cable between the
that pressure module, are stored
plumbs the pressure module to pressure module and calibrator
in the module’s memory.
a pressure source, and connects reduces the length of connect-
Each module has its own
the pressure module cable to ing tubing to the pressure source.
microprocessor, allowing it to run
the calibrator. Pressure is ap- The remote pressure head also
the measurement circuitry and
plied, measured by the pressure provides an extra margin of safety
to communicate digitally with a
module, and displayed digitally and convenience by removing the
calibrator. When connected to
on the calibrator. At the touch of calibrator and operator from the
the calibrator, the modules coef-
a button, the pressure may be pressure source.
ficients are uploaded from the
displayed in up to 11 different pressure module to the calibrator.
engineering units. When used Then, as pressure measurements
with the 750 Series Document- are made, raw sensor values for
ing Process Calibrators, pres- pressure and temperature are
sure readings can be date/time digitally loaded to the calibrator,
stamped and stored electronically where the raw sensor values and
for later retrieval. This saves time, coefficients are manipulated to
eliminates errors, and supports derive and display the pressure
compliance with quality stan- reading.
dards and regulations.
5 Fluke Corporation Pressure Calibration
Pressure Performance
Summary calibrator specifications: (one year, 18 °C to 28 °C)
718 / 719
725 / 726

5520A
717

753
754


Function Range Resolution Accuracy Notes


Measure Pressure
(internal sensor)
1
1 psi to 10000 psi
see table
0.0001 psi to
0.01 psi
0.05 % full scale Gasses/liquids 3 (non
corrosive) Zero, Min,
see table Max, Hold, Damp

Measure Pressure1
( internal sensor)
1 psi to 300 psi
see table
0.0001 psi to
0.01 psi
0.05 % full scale (718)
0.025 % full scale (719)
Gasses (non corrosive)
Zero, Min, Max, Hold,
100 psi (719) see table Damp
• • • • • •


Measure Pressure1
(with Pressure Modules)
29 Pressure Modules,
1 in. H2O / 0.25 kPa
To 0.0001 psi, per To 0.025 % of full span,
Pressure Module per Pressure Module
Media compatibility per
Pressure Module specs2
Over pressure per pressure to specs2 specs2 Zero, Min, Max, Hold,
module specs2 10,000 psi / 69 MPa Damp



Source pressure
built-in pump
-12 psig to full scale N/A N/A

Pressure or vacuum,
overpressure protected
• •
Measure mA 0 to 24 mA 0.001 mA 0.05 % reading + 1 count 719 Source mA
• Measure mA 0 to 24 mA 0.001 mA 0.020 % reading + 2 count
• • Measure mA 0 to 24 mA 0.001 mA 0.010 % reading + 0.015 %
of full scale
• • • • • Loop power supply 24 V dc N/A + 10 %

1
Supported Pressure Units on 701 and 702 include: psi, kPa, bar, inches Hg, mm Hg, inches H2O (@ 4 °C), and feet H2O (@4 °C) Supported Pressure Units on 753, and 754 include all of the above,
plus kg/cm2, inches H2O (@ 60 °F), mm H2O (@ 4 °C). Supported Pressure Units on 717, 718, and 725 include psi, kPa, bar, mbar, kg/cm2, inches Hg, mm Hg, inches H2O (@ 4 °C), inches H2O
(@ 20 °C), cm H2O (@ 4 °C), and cm H2O (@ 20 °C).
2
For Pressure Module specifications, see table below.
3
Gasses only for 718 and 719.

Pressure module specifications (all specifications in % of full span. Specifications reflect a confidence interval of 95 %.)
1
Total uncertainty, one
Reference High2 Low2 Max over- year for temperature
Range/ Range (approx)/ uncertainty Stability Temperature Total1 side side Fitting pressure range 0 °C to +50 °C.
Model Resolution Resolution (23 ± 3 ºC) (1 year) (0 to 50 ºC) uncertainty media media material (x nominal) Total uncertainty, 1.0
% of full span for temp-
Differential
erature range -10 °C
FLUKE-700P00 1 in. H2O/0.001 0.25 kPa/0.0002 0.300 0.025 0.025 0.350 Dry Dry 316 SS 30x to 0°C. For P00 module
FLUKE-700P01* 10 in. H2O/0.01 2.5 kPa/0.002 0.200 0.050 0.050 0.300 Dry Dry 316 SS 3x only, compensated
FLUKE-700P02 1 psi/0.0001 6900 Pa/0.7 0.150 0.070 0.080 0.300 Dry Dry 316 SS 3x temperature range is
15 °C to 35 °C.
FLUKE-700P22 1 psi/0.0001 6900 Pa/0.7 0.100 0.020 0.030 0.150 316 SS Dry 316 SS 3x
2
“Dry” indicates dry
FLUKE-700P03 5 psi/0.0001 34 kPa/0.001 0.050 0.020 0.030 0.100 Dry Dry 316 SS 3x
air or non-corrosive
FLUKE-700P23 5 psi/0.0001 34 kPa/0.001 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x gas as compatible
FLUKE-700P04 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 Dry Dry 316 SS 3x media. “316 SS” in-
FLUKE-700P24* 15 psi/0.001 103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x dicates media compat-
ible with Type 316
Gage Stainless Steel.
FLUKE-700P05* 30 psi/0.001 207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x “C276” indicates
FLUKE-700P06* 100 psi/0.01 690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x media compatible
with Hastelloy C276.
FLUKE-700P27* 300 psi / 0.01 2070 kPa / 0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
FLUKE-700P07 500 psi/0.01 3400 kPa/0.1 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x Use of pressure zero is
required prior to mea-
FLUKE-700P08 1000 psi/0.1 6900 kPa/0.7 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x surement or source.
FLUKE-700P09* 1500 psi/0.1 10 MPa/0.001 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 2x Maximum overpressure
Absolute (not compatible with Fluke 701 or 702) specification includes
common mode pres-
FLUKE-700PA3 5 psi/0.0001 34 kPa/0.001 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x
sure. Modules are
FLUKE-700PA4* 15 psi/0.001 103 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x rated. Metric adapter(s):
FLUKE-700PA5 30 psi/0.001 207 kPa/0.01 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x 1/4˝ NPT
FLUKE-700PA6 100 psi/0.01 690 kPa/0.07 0.050 0.010 0.010 0.070 316 SS N/A 316 SS 3x female to male BSP/
ISO 1/4-19, tapered
Vacuum (not compatible with Fluke 701 or 702) thread, included with
FLUKE-700PV3 -5 psi/0.0001 -34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x all modules except
FLUKE-700PV4 -15 psi/0.001 -103 kPa/0.01 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x P29, P30, and P31.
Effective October 1996,
Dual
all modules include a
FLUKE-700PD2 ±1 psi/0.0001 ±6900 Pa/0.7 0.150 0.025 0.025 0.200 316 SS Dry 316 SS 3x NIST traceable certifi-
FLUKE-700PD3 ±5 psi/0.0001 ±34 kPa/0.001 0.040 0.015 0.015 0.070 316 SS Dry 316 SS 3x cate and test data.
FLUKE-700PD4 ±15 psi/0.001 ±103 kPa/0.01 0.025 0.010 0.015 0.050 316 SS Dry 316 SS 3x *Intrinsically safe
FLUKE-700PD5 -15/30 psi/0.001 -100/207 kPa/0.01 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x version available for
use with the 718Ex
FLUKE-700PD6 -15/100 psi/0.01 -100/690 kPa/0.07 0.025 0.010 0.015 0.050 316 SS N/A 316 SS 3x
and 725Ex.
FLUKE-700PD7 -15/200 psi/0.01 -100/1380 kPa/0.1 0.040 0.015 0.015 0.070 316 SS N/A 316 SS 3x
High
FLUKE-700P29* 3000 psi/0.1 20.7 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2x
FLUKE-700P30 5000 psi/0.1 34 M Pa/0.001 0.050 0.010 0.020 0.080 C276 N/A C276 2x
FLUKE-700P31 10000 psi/1 69 M Pa/0.007 0.050 0.010 0.020 0.080 C276 N/A C276 1.5x

6 Fluke Corporation Pressure Calibration


Pressure Accessories
Fluke 700PTP-1 Fluke 700HTP-1 Fluke 700LTP-1
Pneumatic Test Pump Hydraulic Test Pump Low-Pressure Test Pump

For use with: Fluke 700 Series For use with: Fluke 700 Series Description: The Fluke 700LTP-1
Pressure Modules and the Fluke Pressure Modules and the Fluke is a hand operated pressure
710 Series Pressure Calibrators. 710 Series Pressure Calibrators. pump designed to generate either
vacuum to -12 psi/-.85 bar or
Description: The Fluke 700PTP-1 Description: The Fluke 700HTP-
pressures to 100 psi/6.9 bar. The
is a handheld pressure pump 1 is designed to generate pres-
Fluke 700LTP-1 has two
designed to generate either sures up to 10,000 psi/700 bar.
pressure ports with push fit
vacuum to -13 psi/-0.9 bar or The Fluke 700HTP-1 has two
connectors. These push fit con-
pressure to 600 psi/40 bar. pressure ports:
nectors, one for the reference
• 3/8-BSP (ISO228) female
The Fluke 700PTP-1 has two port for connection to a Fluke
parallel thread fitting for the
pressure ports: 700 series pressure module and
reference gauge or pressure
• 3/8-BSP (ISO228) female one to connect to a unit under
module
parallel thread fitting for the test, connect to the supplied
• 1/8-BSP (ISO228) female par-
reference gauge or pressure test hoses. These test hoses are
allel thread fitting for the unit
module terminated with 1/4-BSP (ISO228)
under test
• 1/8-BSP (ISO228) female female parallel thread fittings that
Note: The user must provide a hose with
parallel thread fitting for the appropriate end fittings from this port to can be adapted using the fittings
unit under test the unit under test. included.
Application: The Fluke 700PTP- Application: This pump can Application: The Fluke 700LTP-1
1 features an integral pressure provide up to 10,000 psi using is primarily intended for low
adjustment vernier which varies distilled water or mineral-based pressure applications. It features a
the pressurized volume by 2.0 hydraulic oil. The pump is operat- fine adjust vernier with
cc over approximately eleven ed by pumping several strokes to .00145 / PSI resolution at low
turns of the vernier knob. The prime the system, then switching pressures. The pressure variation
pressure variation achievable to high pressure mode when the achievable with the vernier will
with the vernier will depend on resistance increases. An integral depend on the nominal pressure
the nominal pressure and total pressure adjustment vernier knob and total pressurized volume,
pressurized volume, but with a varies the pressurized volume by but with minimum volume and
minimum volume and maximum 0.6 cc. The pressure variation maximum pressure the vernier
pressure, the vernier provided achievable with the vernier will provides 30 psi ± 6 psi. The
600 ± 20 psi adjustment range. depend on the nominal pressure adjustable pressure relief valve
With a minimum volume and no and total pressurized volume, but features a slow-bleed capability
pressure applied, the vernier can with a minimum volume, the that allows the user to slowly re-
also be used to provide a 0 to vernier provided 150-3,000 psi lease pressure at a controlled rate
70˝ H2O range. Larger volumes (at 150 psi nominal) and 3,000- to achieve a desired pressure.
will provide a smaller range of 10,000 psi (at 3,000 psi nomi-
adjustment, but greater resolu- nal) adjustment ranges. With a
tion. The length of the stroke can minimum volume and no pressure
be adjusted to limit the maximum applied, the vernier can also be
output pressure. Maximum output used to provide a 0 to 1.7 psi
pressure is adjustable from range. Larger volumes will pro-
2.5 psi to 600 psi. vide a smaller range of adjust-
ment, but greater resolution.

7 Fluke Corporation Pressure Calibration


Fluke 700HTH-1 Fluke 700ILF Fluke 700PMP
Hydraulic Test Hose In-line Filter Pressure Pump

Description: The Fluke For use with: Fluke 717, 718, For use with: Fluke 700 Series
700HTH-1 hydraulic test hose is and 719 Pressure Calibrators to Pressure Modules and the Fluke
a 10,000 psi, 700 bar working help isolate the calibrator from 710 Series Pressure Calibrators.
pressure test hose. The hose uses liquid contamination.
self-sealing fittings with easy Description: The Fluke 700PMP
finger tight connections. Description: Use to isolate cali- is a hand-operated pressure
brator from contact with liquids. pump to provide pressures up to
Application: The Fluke Input is 1/8˝ NPT female. Output is 150 psi/1000 kPa. Output fitting
700HTH-1 allows connections 1/8˝ NPT male. is 1/8 FNPT.
to a unit under test from a Fluke
700HTP-1 hydraulic test pump Application: The Fluke 700ILF Application: Linear stroke
in use with the Fluke 700 series can be used to isolate the calibra- of 1.6˝ (4 cm). Multi-turn vernier
pressure modules. The 700HTH-1 tor from incidental contact with for fine adjustment of pressure.
is compatible with water and fluids present in the gas medium Includes controlled pressure
non-corrosive oil. being measured. Particularly bleed valve.
useful with the 718/719 calibra-
Fluke 700PRV-1 Pressure tor to help keep moisture or oils
Relief Valve Kit from contaminating the on-board
pump. The 1/8˝ NPT male thread- Pressure
For use with: Fluke 700HTP-1 ed output threads directly into Calibrator
Hydraulic Test Pump. the 717, 718, and 719 calibra-
Description: The Fluke 700PRV-1 tors. Maximum pressure is 100
consists of two relief valves (1360 psi. Burst pressure is 375 psi.
and 5450 psi) to be used with the Maximum flow is 10 SCFM and
700HTP-1 Hydraulic Test Pump. filtration to 1 micron. Pressure Interface
These relief valves will protect Module Unit
the Fluke pressure modules from Fluke 700 PCK
damage due to over-pressuriza- Calibration Kit
tion. 1/4 BSP male parallel thread The Fluke 700PCK Pressure
to fit Fluke 700HTP-1. Calibration Kit makes it pos-
Application: Repeatability sible to calibrate your pressure
± 10 % of nominal setting. Multi- modules at your facility using
turn adjustment screw to set your own precision pressure Line Power

preload on internal disc springs. standards. The kit consists of


a power supply, an interface PC Serial Port
Fluke 71X Hose Kit adapter, appropriate cables, 8 7 6 5 4 3 2 1

and Fluke 700PC Pressure


For use with: Fluke 718 and 719
Module Calibration software.
Pressure Calibrators.
When installed on your PC,
Description: The Fluke 71X hose the Windows®-based software
kit includes (2) quick disconnect easily steps you through an
fittings to connect to the 718 or as-found verification, a calibra-
719, three 1 meter translucent tion adjustment, and an as-
hoses and one BSP adapter. left verification. Calibration
data is captured for import to
your database. A 386 or better
PC, running Windows 3.1, or
later is required, along with a
precision pressure standard
with an uncertainty of less
than 1/4 that of the pressure
module being verified. Note: With a Fluke 700PCK and any Fluke Pressure Module, a
Fluke 5520A Calibrator becomes a precision pressure standard.

8 Fluke Corporation Pressure Calibration


Pressure Terminology
Absolute pressure — absolute Gage pressure transducer — PSID — pounds per square inch
pressure measurements are a transducer that measures differential.
referenced to zero pressure, pressure relative to atmospheric
(a perfect vacuum.) PSIG — pounds per square inch
pressure.
gage (same as psi).
Absolute pressure Ideal Gas Law — combining
transducer — a transducer that Square root extractor — an
Boyle’s Law and Charles’ Law,
has an internal reference instrument or software program
results in the Ideal Gas Law:
chamber sealed at or close to zero that takes the square root of input
PV=nRT, where nR is constant
pressure (full vacuum) and puts the result on its output.
for a particular gas analogous to
when exposed to atmosphere Square root extraction is needed
the number of molecules and the
a reading of approximately to linearize many flow signals.
relative size of the molecule.
14.7 psi results. Example: orifice plates, ventu-
I/P (I to P) — a current to pres- ris, target flow meters, and pitot
Boyle’s Law — the volume of a sure transmitter. A common tubes all require the transmitter’s
gas is inversely proportional instrument in modern industrial output signal to be linearized.
to the pressure of the gas at plants. A typical large paper mill Mag flow meters, turbine flow
constant temperature: V=1/P. or refinery could have 5,000 I/Ps meters, Doppler flow meters, and
Charles’ Law — essentially in use. vortex shedding flow meters don’t
states for a fixed volume of gas, require square root extraction.
Line pressure — the maximum
if the temperature is raised, the pressure in the pressure vessel Static pressure — the zero-
pressure will increase. or pipe for differential pressure velocity pressure at any arbitrary
P = Constant x T. measurement. point within a system.
Common mode pressure — the Orifice plate — a very low cost Wet/dry differential — a dif-
underlying common pressure and common primary sensing el- ferential pressure transducer
(or static pressure) within a ement (PSE) for measuring flow. It or transmitter that uses a metal
system from which a differential must be used in conjunction with diaphragm at the wet port where
measurement is being made. a d/p cell. It creates a venturi and fluids can be applied, and no dia-
D/P: Differential pressure, a resulting P is developed across phragm at the dry port. The dry
(pronounced DP) — other names the plate whose square root is port exposes the sensor material
used to mean the same thing proportional to flow. to the medium, so only clean dry
are d/p cell, d/p transmitter and gas can be applied to this port.
P/I (P to I) — a pressure to cur-
∆P transmitter (where ∆ is delta rent transducer. Wetted parts — the diaphragm
or differential). This is the most and pressure port material that
common type of transmitter used Pneumatic relay — refers to a
comes in direct contact with the
in most process industries. It can pneumatic instrument that per-
medium (gas, liquid).
be used to measure level, flow, forms a function to its input and
pressure, differential pressure, provides the result on its output
and density or specific gravity. (Example: square root extractor,
With some modifications, it can adder, etc.).
measure such things as tempera- PSI — pounds per square inch
ture and oxygen purity. The d/p (same as psig).
transmitter can be pneumatic, MA
NAGEMENT Fluke. Keeping your world
PSIA — pounds per square inch
SY

up and running.®
TY

electromechanical, or solid state.


ST
I
AL

EM


QU

absolute. ISO 9001


It can also be a smart transmitter.
CE

A typical large process plant can


1

Fluke Corporation
00
RT

FI
9

O
I

ED S
TO MEET I

have hundreds or thousands of PO Box 9090, Everett, WA 98206 U.S.A.


d/p transmitters in service. Fluke Europe B.V.
PO Box 1186, 5602 BD
Gage pressure — the pressure Eindhoven, The Netherlands
relative to atmospheric pressure. For more information call:
In the U.S.A. (800) 443-5853 or
Gage pressure = absolute pres- Fax (425) 446-5116
sure minus one atmosphere. In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2006-2011 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 7/2011 1281620J D-EN-N

Modification of this document is not permitted


without written permission from Fluke Corporation.

9 Fluke Corporation Pressure Calibration

You might also like