Professional Documents
Culture Documents
candidate reservoir must be capable of withstanding an upper portion of the cell. Pressure and temperature in the
average reservoir pressure greater than the CO2 MMP for cell and chambers are displayed continuously on the
obtaining the highest recovery3. It also determines the monitor.
model to be used in predicting or simulating reservoir
performance as a result of CO2 injection.
A prior well-known method used to determine MMP H igh
is the 'slim-tube' method2,3,4,5,10,11 which involves loading oil Temperature
Densitometer T V Monitor &
sample into a column packed with porous media (having Video Recorder
internal diameter (ID) of 4 to 19 mm5) and then injecting
gas/gases to displace/extract the oil. VLE-IT Cell & Enclosure
The problems with this conventional apparatus include
the difficulties associated with the relatively large column Frame Grabber
Board
Computer Control
diameter used and the difficulties in obtaining uniform & Data Processing
packing. The non-uniform porosity and flow rates that
occur as a result give rise to transverse concentration
fingers, making it more difficult to obtain accurate and Micro Video Camera
& Illumination Printer
reproducible data repeatedly in a consistent manner
between laboratories. To minimise this effect, it is often
necessary to operate at extremely low flow rates so that Figure 2: VLE-IT Equipment Major Components
molecular diffusion can dampen the unwanted transverse
compositional variations. As a consequence, it can take Test selections are made interactively, allowing the
several days to run optimal tests with a slim-tube packed user to set the test criteria. The chamber volumetric rates
column. Therefore, there is still a need for a measuring can be varied from 1 to 50 cm3/min and a constant pressure
apparatus that can determine MMP more accurately and mode is used to adjust liquid and/or vapour volumes during
quickly with direct measurement of displacement recovery mixing, and to eventually set the position of the gas-liquid
and visual confirmation to avoid error and ambiguity as interface in the VLE-IT cell.
disclosed in the prior method. The densities of liquid and gas phases are measured
This paper demonstrates the use of the VLE-IT using a PAAR DMA 512P density-measuring cell, which is
equipment to measure IFT of stock tank oil (STO) with CO2 connected to mPDS 2000 evaluation unit. The external cell
gas at reservoir temperature of 77 oC and various pressures. is heated to the desired temperature by a computer-
By making a plot of IFT against pressure, the MMP value is controlled heating block.
obtained by extrapolation to zero IFT. This paper also Drop volume and pendant drop techniques are used in
describes a new approach in determining MMP whereby an the IFT measurement. The drop volume technique
unpacked transparent capillary column with ID of 0.32 mm requires running the variable volume chambers of the
is used rather than the conventional packed column. Due to equipment at slow rates. The volume of a drop (Vd) is given
the standardised nature of the column, repeatability between by:
laboratories can be achieved. This new approach can thus
be adopted as a standard method to experimentally Vd = flow rate x [time/number of drops] ………...[A1]
determine the CO2 MMP for an oil.
The IFT (σ) is then calculated using equation:
Approach I: MMP from Interfacial Tension (IFT) σ = [Vd (ρ1 - ρ2)gc . F / r] ..………………. [A2]
Measurement
where F is a function of r/Vd1/3 based on the work by
Apparatus and procedure. The VLE-IT equipment and its Harkins and Brown7, extrapolated from tables given by
major components are illustrated in Figure 26. It consists of Lando and Oakley8.
a fixed volume cell made of 316 s/s with glass windows on In the alternative pendant drop technique, the image
opposite sides and it can withstand pressures up to 10,000 of the drop hanging from the tip is captured and analysed.
psi and temperatures up to 150 oC. There is a light source to The image analysis applies a computational procedure
illuminate the windows to permit the capture of drop profile derived by Rotenberg etal9 for obtaining IFT and contact
images by using a camera system on the opposite side. The angle from the shape of an axisymmetric fluid interface.
camera system is connected to a monitor and a PC equipped In the procedure, an interface profile from a
with a frame grabber and image analysis software. mathematically derived curve that satisfies the Laplace
There are also two variable volume chambers capillary equation is constructed:
connected to the cell by computer-controlled switching
valves, which determine the mixture flow between the first σ(1/ R1+ 1/ R2 ) = ∆P ………………………. [A3]
variable volume chamber and the capillary tip located in the
57286 New Experimental Approaches in Minimum Miscibility Pressure (MMP) Determination 3
where ∆P = pressure difference across the interface system since miscibility between two phases is achieved
surface of a pendant drop, R1, R2 = two principle radii of when there is no IFT between them. It can be seen in Figure
curvature 3 that the best-fitted line to the IFT data up to 1913 psig,
In this study, the IFT value obtained by the drop intersects the zero IFT axis at a pressure of about 2400 psig.
volume technique was used as the first estimate to match
the shape of the physically observed pendant drop to the Pressure IFT values (mN/m)
(psig)
mathematically derived one. Iterative calculations were then
Drop Volume Pendant drop
made using an in-house program called DROP until the best 1206 7.24 7.00
fit between the two shapes was obtained. The IFT values 1330 5.49 5.40
were then plotted against pressure. The curve was 1435 3.98 4.00
extrapolated to zero IFT in order to get the MMP value. 1515 3.53 3.50
1913 0.64 0.41
Validation with published values. To validate the
performance of the VLE-IT equipment, IFT was determined Table 3: IFT for STO-CO2 @ 77 oC
for a methane-pentane mixture. A good agreement of results
8
was obtained for both drop volume and pendant drop
Drop Volume
methods. The values are also within the range of published 7
Pendant drop
results as shown in Table 1. 6
Calculated
5
IFT (mN/M)
Pressure IFT by drop IFT by pendant IFT reported in literature
4
(bar) volume drop (mN/m)
(mN/m) (mN/m) 3
0
Table 1: IFT for methane-pentane @ 24 oC 0 500 1,000 1,500 2,000 2,500 3,000 3,500 4,000
PRESSURE (PSIG)
Results. The measurements of equilibrium IFT were Figure 3: Interfacial tension for STO with CO2
performed at reservoir temperature and selected pressures
on a STO the properties of which are given in Table 2.
Approach II: MMP from Micro-slim Tube
Compo. Weight % Compo. Weight % Displacement Tests
C7 0.46 C18 2.88
C8 3.41 C19 2.58 Apparatus and procedure. The Gas Extraction &
C9 7.72 C20 2.14 Miscibility Analyser (GEMA), which incorporates a micro-
C10 10.36 C21 1.81 slim tube (MST) apparatus, is shown schematically in
C11 12.03 C22 1.62
Figure 4.
C12 11.60 C23 1.51
C13 10.26 C24 1.34
C14 8.92 C25 1.43
C15 7.89 C26 1.22
C16 4.96 C27 1.03 TV Monitor &
Video Recorder
C17 4.02 C28 0.80 Micro-slim tube
in oven
Density = 0.8267 g/cc
MWt. = 170.4
The Pore Volume of the system is approximately 3.71 cm 3. Validation with published values. The micro-slim tube
A coiled stainless steel column having ID of 0.32 to 0.53 was calibrated against displacement tests on hexadecane-
mm can also be used especially when operating at much CO2 at 50 oC as conducted by Johnson & Pollin10. Table 4
higher pressures. and Figure 5 show the comparison of the two systems.
The supply of the solvent and the sample is by The recovery profile in both cases is the same i.e. the
mercury-free, rapid recharge high-pressure syringe pumps. recovery increases gradually until it levels off at certain
The pumps are computer programmable and pulseless with pressure. Considering the different nature of the 'slim-tube'
option of constant pressure, constant rate and used, the 12.5% difference in the obtained MMP values is
programmable pressure or rate. deemed acceptable.
There are 4-ports and 6-ports switching valves which
determine the flow path of the fluid. The MST is housed in Parameter This study Johnson & Pollin
a constant temperature oven. Elevated temperatures are Slim tube 30m x 0.32 mm ID 15 m x 6.35 mm ID
Packing un-packed 100-140 mesh glass beads
achieved by means of Watlow heating tapes inside the oven
Linear rate, cm/hr 18 650 500
and also around the pumps, controlled by electronic MMP obtained 2100 2400
temperature controller through a computer program.
The ultra-violet (UV) detector at the outlet end of the Table 4: Displacement tests on C16 - CO2 @ 50 oC
MST indicated the CO2 breakthrough. When using the
fused silica capillary column, the interface of the fluids
100
flowing in the tube throughout the test can also be observed MMP~2100 psig
using a scanning micro-video camera mounted near the
80
periphery of the column.
A series of displacement tests were conducted at
60
various pressure levels. For each displacement test, the MMP~2400 PSIG
% recovery
MST was filled with oil at 77oC and test pressure. The small
40
bore capillary column permits a high linear displacement this study
rate of 311 cm/min (0.25 cm3/min) resulting in time needed
20 Johnson & Pollin
per test of only 2 to 3 hours (including cleaning). The
relatively small height equivalent to a theoretical plate 0
(HETP) in a capillary column ensures thousands of multiple 0 500 1,000 1,500 2,000 2,500 3,000
Pressure, psig
contacts between the injected gas and oil, which cannot be
practically achieved in the conventional slim tube. Figure 5: Comparison of MMP from micro-slim tube
The volumes of displaced oil and CO2 injected at and conventional slim tube for hexadecane with CO2
breakthrough were either obtained directly from the pump
readings or by multiplying the displacement rate with the Results. The results of the displacement tests are
displacement time. The displacement time is referred to as summarised in Table 5. Although the displacement tests
the time taken for the oil to first hit the UV detector until data obtained were not indicative of the ultimate recovery to
the onset of CO2 breakthrough. The test was terminated be achieved from a core flood or on a reservoir scale, they
when the response stabilised. were used to distinguish miscible displacements from
The recoveries at breakthrough as calculated using immiscible ones and to show the effect of pressure on
equation [A4] were used in determining the MMP. It has displacement efficiency. Thus, the CO2 MMP determined
been reported2,3 that the ultimate recovery as well as was the minimum pressure at which CO2 could generate
breakthrough recovery result in the same interpretation for miscibility with the test oil through multiple-contact
miscibility pressure. process. Nevertheless, the experiment gives potential
operating pressure/s, which is one of the most critical
% Recovery = elements in reservoir management.
Apart from achieving 80% recovery, the UV response Both techniques are faster than the conventional slim-
and the observation of flowing fluids interface were also tube displacement test or the sand-pack experiment, which
used to determine whether each test was a miscible usually take 3 to 4 weeks to complete.
displacement. For example, the displacement at 1500 psig
was considered immiscible although the percentage Comparison with predicted MMP by EOS & published
recovery exceeded 80% since sharp fluid interface was correlations. A phase-behaviour computer software
observed throughout the test and the UV response was (LABSIM) which consists of several programs utilising the
erratic at breakthrough indicating that two phases existed. Peng-Robinson Equation-of-state (EOS) was used to
Figure 6 shows the difference in the UV response for simulate the multiple contact processes. The EOS was
immiscible displacement at 1500 psig and miscible calibrated by matching the saturation pressure of the
displacement at 3500 psig. reservoir from which the STO was taken.
As shown in Table 6, the MMP values obtained from
both techniques in the present study are in good agreement
P = 1500 psig (immiscible) CO2 breakthrough with each other and fall within the range of predicted MMP
by EOS and published correlations.
absorbance
2182
Cronquist 2400
Conclusions
Figure 6: Comparison of the UV detector response for
immiscible (at 1500 psig) and miscible (at 3500 psig) 1. A new approach using prototype VLE-IT
displacement for STO with CO2 equipment involving the measurement of oil/gas
interfacial tensions was successfully applied in
It is shown in Figure 7 that the two best straight lines determining the CO2 MMP with a STO.
representing the miscible and immiscible breakthrough
intersect at 2300 psig. Thus, the MMP is taken to be 2. A new approach in slim-tube test using unpacked
approximately 2300 psig. capillary column was proven to be capable of
determining the MMP. The tests yielded an MMP
100
value equivalent to the MMP determined from the IFT
90 technique.
80
% recovery
Acknowledgement
The authors would like to thank the management of
PETRONAS and PETRONAS Research and Scientific
Services Sdn. Bhd. for permission to publish this work.
Sincere appreciation is also extended to the technicians and
support staff for their valuable contribution.
References