Acoustic Ouput Tables

for

Transducers

P/N: 4710.00242A03

Note: Some transducer models are not available in certain countries.


Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm
2
Aaprt>1cm
2 scan
Maximum index value 1.364 1.769 # # # #
pr, α (MPa) 2.033
P (mW) 226.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.000 #
deq(zb) (cm) #
fa wf (MHz) 2.219 2.219 # # # #
Dim of Aaprt X (cm) 2.048 # # # #
Y (cm) 1.350 # # # #
td (µsec) 0.684
prr (Hz) 4684
pr at max Ipi (MPa) 3.344
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 237.048
Focal FLx (cm) 8.0 # # #
Length FLy (cm) 8.1 # # #
Mode B # # # # #
Operating
Focus (cm) 8 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
2
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm
2
Aaprt>1cm
2 scan
Maximum index value 1.333 1.436 # # # #
pr, α (MPa) 2.093
P (mW) 225.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) 3.000 #
parameters z at max Ipi.α (cm) 3.017 #
deq(zb) (cm) #
fa wf (MHz) 2.464 2.464 # # # #
Dim of Aaprt X (cm) 2.048 # # # #
Y (cm) 1.350 # # # #
td (µsec) 0.638
prr (Hz) 4684
pr at max Ipi (MPa) 3.680
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 287.984
Focal FLx (cm) 8.0 # # #
Length FLy (cm) 8.1 # # #
Mode THI # # # # #
Operating
Focus (cm) 8 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
3
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm Scan
Maximum index value 1.122 1.734 # 1.207 4.409 #
pr, α (MPa) 1.671
P (mW) 226.400 # 181.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 114.332
zs (cm) 3.000
Associated zbp (cm) 2.810
acoustic zb (cm) . 5.788
parameters z at max Ipi.α (cm) 3.000
deq(zb) (cm) 0.374
fa wf (MHz) 2.219 2.219 # 2.217 2.217 #
Dim of Aaprt X (cm) 2.048 # 2.020 2.020 #
Y (cm) 1.350 # 1.350 1.350 #
td (µsec) 0.687
prr (Hz) 133
pr at max Ipi (MPa) 2.602
Other
deq at max Ipi (cm) 0.383
information 2
Ipi.α at max MI (W/cm ) 178.522
Focal FLx (cm) 8.0 # 8.0 #
Length FLy (cm) 8.1 # 8.1 #
Mode B/M B # M M #
Operating
Focus (cm) 8 8 # 8 8 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
4
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm
2 scan
Maximum index value 0.986 # # 1.960 4.510 #
pr, α (MPa) 1.547
P (mW) # # 122.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 73.431
zs (cm) 3.000
zbp (cm) 2.810
Associated
zb (cm) 5.930
acoustic
z at max Ipi.α (cm) 3.000
parameters
deq(zb) (cm) 0.194
fa wf (MHz) 2.458 # # 2.458 2.458 #
Dim of Aaprt X (cm) # # 2.020 2.020 #
Y (cm) # # 1.350 1.350 #
td (µsec) 1.514
prr (Hz) 5500
pr at max Ipi (MPa) 2.559
Other
deq at max Ipi (cm) 0.194
information
Ipi.α at max MI (W/cm2 ) 134.691
Focal FLx (cm) # # 8.0 #
Length FLy (cm # # 8.1 #
Mode ) PW # # PW PW #
Operating Focus 8 # # 8 8 #
control (cm)
PRF 5.5 # # 5.5 5.5 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
5
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm
2
Aaprt>1cm
2 scan
Maximum index value 1.179 2.209 # # # #
pr, α (MPa) 1.831
P (mW) 372.000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 3.000 #
parameters
deq(zb) (cm)
fa wf (MHz) 2.414 2.414 # # # #
Dim of Aaprt X (cm) 2.048 # # # #
Y (cm) 1.35 # # # #
td (µsec) 1.555
prr (Hz) 5500
pr at max Ipi (MPa) 3.063
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 142.187
Focal FLx (cm) 8.0 # # #
Length FLy (cm) 8.1 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 8 8 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
6
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm 2 scan
2
Maximum index value 1.032 2.933 cm
# # # #
pr, α (MPa) 1.590
P (mW) 372.000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.000 #
deq(zb) (cm) #
fa wf (MHz) 2.206 2.206 # # # #
Dim of Aaprt X (cm) 2.048 # # # #
Y (cm) 1.350 # # # #
td (µsec) 1.733
prr (Hz) 5000
pr at max Ipi (MPa) 2.530
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 118.560
Focal FLx (cm) 8.0 # # #
Length FLy (cm) 8.1 # # #
Mode DPI B-DPI # # # #
Operating
Focus 8 8 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
7
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: CW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm
2 scan
Maximum index value 0.235 # # 0.232 4.263 0.453
pr, α (MPa) 0.332
P (mW) # # 39.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 21.940
zs (cm) 2.817
zbp (cm) 2.810
Associated
zb (cm) 4.570
acoustic
z at max Ipi.α (cm) 2.817
parameters
deq(zb) (cm) 0.068
fa wf (MHz) 3.035 # # 3.035 3.035 #
Dim of Aaprt X (cm) # # 2.020 2.020 #
Y (cm) # # 1.350 1.350 #
td (µsec) 4.996
prr (Hz) #
pr at max Ipi (MPa) 0.535
Other
deq at max Ipi (cm) 0.067
information
Ipi.α at max MI (W/cm2 ) 4.036
Focal FLx (cm) # # 4.0 #
Length FLy (cm) # # 4.0 #
Mode CW # # CW CW #
Operating Focus 4 # # 4 4 #
control (cm)
PRF 8.0 # # 8.0 8.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
8
Acoustic Output Tables
Transducer Model: 2P1 Phased Array Operating Mode: TDI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm 2 scan
2
Maximum index value 1.132 1.621 cm
# # # #
pr, α (MPa) 1.315
P (mW) 406.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.343 #
deq(zb) (cm) #
fa wf (MHz) 2.018 2.018 # # # #
Dim of Aaprt X (cm) 2.048 # # # #
Y (cm) 1.350 # # # #
td (µsec) 1.878
prr (Hz) 5500
pr at max Ipi (MPa) 1.818
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 112.289
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode TDI B-TDI # # # #
Operating
Focus 4 4 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
9
Acoustic Output Tables

Transducer Model: 2P1 Phased Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.571 1.645 # 0.853 3.611 #
pr, α (MPa) 1.241
P (mW) 413.100 # 125.30 #
min of [Pα(zs),Ita.α(zs)] (mW) 63.360 0
zs (cm) 3.493

Associated zbp (cm) 2.810


acoustic zb (cm) 5.797
parameters z at max Ipi.α (cm) 3.341
deq(zb) (cm) 0.253
fa wf (MHz) 2.017 2.017 # 2.826 2.826 #
Dim of Aaprt X (cm) 2.048 # 2.020 2.020 #
Y (cm) 1.350 # 1.350 1.350 #
td (µsec) 1.878
prr (Hz) 5500
pr at max Ipi (MPa) 1.722
Other
deq at max Ipi (cm) # 0.251
information 2
Ipi.α at max MI (W/cm ) 89.075
Focal FLx (cm) 4.0 # 4.0 #
Length FLy (cm) 4.0 # 4.0 #
Mode B B/COLOR # PW PW #
Operating 4 4 # 4 4
Focus (cm) #
control
PRF (kHz) 5.5 5.5 # 5.5 5.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
10
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.031 1.804 # # # #
pr, α (MPa) 2.050
P (mW) 102.100 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.500 #
deq(zb) (cm) #
fa wf (MHz) 4.146 4.146 # # # #
Dim of Aaprt X (cm) 1.020 # # # #
Y (cm) 0.950 # # # #
td (µsec) 0.337
prr (Hz) 7680
pr at max Ipi (MPa) 3.261
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 163.205
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode B # # # # #
Operating
Focus (cm) 4 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
11
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.301 1.221 # # # #
pr, α (MPa) 1.987
P (mW) 69.000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.500 #
deq(zb) (cm) #
fa wf (MHz) 4.083 4.083 # # # #
Dim of Aaprt X (cm) 1.020 # # # #
Y (cm) 0.95 # # # #
td (µsec) 0.349
prr (Hz) 7680
pr at max Ipi (MPa) 3.160
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 158.990
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 4 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
12
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.098 1.812 # 1.946 0.420 #
pr, α (MPa) 2.228
P (mW) 102.100 # 99.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 11.576
zs (cm) 3.361
Associated zbp (cm) 1.664
acoustic zb (cm) . 3.361
parameters z at max Ipi.α (cm) 3.361
deq(zb) (cm) 2.047
fa wf (MHz) 4.146 4.146 # 4.120 4.120 #
Dim of Aaprt X (cm) 1.020 # 1.020 1.020 #
Y (cm) 0.950 # 0.950 0.950 #
td (µsec) 0.368
prr (Hz) 133
pr at max Ipi (MPa) 3.595
Other
deq at max Ipi (cm) 2.016
information 2
Ipi.α at max MI (W/cm ) 209.109
Focal FLx (cm) 4.0 # 4.0 #
Length FLy (cm) 4.0 # 4.0 #
Mode B/M B # M M #
Operating
Focus (cm) 4 4 # 4 4 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
13
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.669 # # 0.509 1.181 #
pr, α (MPa) 1.340
P (mW) # # 26.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 13.290
zs (cm) 2.500
zbp (cm) 1.664
Associated
zb (cm) 3.817
acoustic
z at max Ipi.α (cm) 2.500
parameters
deq(zb) (cm) 0.176
fa wf (MHz) 4.018 # # 4.017 4.017 #
Dim of Aaprt X (cm) # # 1.020 1.020 #
Y (cm) # # 0.950 0.950 #
td (µsec) 0.939
prr (Hz) 4000
pr at max Ipi (MPa) 2.277
Other
deq at max Ipi (cm) 0.173
information
Ipi.α at max MI (W/cm2 ) 57.809
Focal FLx (cm) # # 4.0 #
Length FLy (cm # # 4.0 #
Mode ) PW # # PW PW #
Operating Focus 4 # # 4 4 #
control
(cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
14
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.120 1.267 # # # #
pr, α (MPa) 989
1.693
P (mW) 70.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 2.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 4.304 4.304 # # # #
Dim of Aaprt X (cm) 1.02 # # # #
Y (cm) 0.95 # # # #
td (µsec) 0.870
prr (Hz) 4000
pr at max Ipi (MPa) 2.454
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 105.274
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 4 4 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
15
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.849 1.273 cm
# # # #
pr, α (MPa) 1.766
P (mW) 70.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.500 #
deq(zb) (cm) #
fa wf (MHz) 4.327 4.327 # # # #
Dim of Aaprt X (cm) 1.02 # # # #
Y (cm) 0.95 # # # #
td (µsec) 0.878
prr (Hz) 4000
pr at max Ipi (MPa) 2.698
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 111.091
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode DPI B-DPI # # # #
Operating
Focus 4 4 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
16
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: CW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.168 # # 0.751 3.661 0.811
pr, α (MPa) 0.351
P (mW) # # 36.100 #
min of [Pα(zs),Ita.α(zs)] (mW) 16.982
zs (cm) 2.500
zbp (cm) 1.667
Associated
zb (cm) 2.642
acoustic
z at max Ipi.α (cm) 2.500
parameters
deq(zb) (cm) 0.069
fa wf (MHz) 4.367 # # 4.367 4.367 #
Dim of Aaprt X (cm) # # 1.020 1.020 #
Y (cm) # # 0.950 0.950 #
td (µsec) 1.037
prr (Hz) #
pr at max Ipi (MPa) 0.523
Other
deq at max Ipi (cm) 0.068
information
Ipi.α at max MI (W/cm2 ) 4.533
Focal FLx (cm) # # 4.0 #
Length FLy (cm # # 4.0 #
Mode ) PW # # PW PW #
Operating Focus 4 # # 4 4 #
control
(cm)
PRF 8.0 # # 8.0 8.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
17
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: TDI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 1.158 2.673 cm
# # # #
pr, α (MPa) 2.302
P (mW) 194.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.500 #
deq(zb) (cm) #
fa wf (MHz) 3.951 3.951 # # # #
Dim of Aaprt X (cm) 1.024 # # # #
Y (cm) 0.950 # # # #
td (µsec) 0.971
prr (Hz) 5000
pr at max Ipi (MPa) 3.506
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 183.800
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode TDI B-TDI # # # #
Operating
Focus 4 4 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
18
Acoustic Output Tables

Transducer Model: 5P1 Phased Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.301 1.838 # 0.549 2.227 #
pr, α (MPa) 989
1.706
P (mW) 73.900 # 28.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 14.339
zs (cm) 2.500

Associated zbp (cm) 1.664


acoustic zb (cm) 0.183
parameters z at max Ipi.α (cm) 2.500
deq(zb) (cm) 0.253
fa wf (MHz) 4.304 4.304 # 4.018 4.018 #
Dim of Aaprt X (cm) 1.020 # 1.020 1.020 #
Y (cm) 0.950 # 0.950 0.950 #
td (µsec) 0.870
prr (Hz) 3000
pr at max Ipi (MPa) 2.473
Other
deq at max Ipi (cm) # 0.183
information 2
Ipi.α at max MI (W/cm ) 106.909
Focal FLx (cm) 4.0 # 4.0 #
Length FLy (cm) 4.0 # 4.0 #
Mode B B/COLOR # PW PW #
Operating 4 4 # 4 4
Focus (cm) #
control
PRF (kHz) 3.0 3.0 # 4.0 4.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
19
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.070 0.639 # # # #
pr, α (MPa) 2.503
P (mW) 30.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 5.475 5.475 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.223
prr (Hz) 7040
pr at max Ipi (MPa) 3.499
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 209.932
Focal FLx (cm) 4 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 4.0 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
20
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.986 0.529 # # # #
pr, α (MPa) 2.353
P (mW) 24.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 5.693 5.693 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.210
prr (Hz) 7040
pr at max Ipi (MPa) 3.250
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 208.175
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 2.7 # # #
Mode THI # # # # #
Operating
Focus (cm) 4.0 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
21
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.070 0.639 # 0.717 1.429 #
pr, α (MPa) 2.438
P (mW) 30.6000 # 27.500 #
min of [Pα(zs),Ita.α(zs)] (mW) 5 6.024
zs (cm) 1.361
Associated zbp (cm) 1.352
acoustic zb (cm) . 1.361
parameters z at max Ipi.α (cm) 1.361
deq(zb) (cm) 1.867
fa wf (MHz) 5.475 5.475 # 5.444 5.444 #
Dim of Aaprt X (cm) 1.28 # 2.59 2.59 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.133
prr (Hz) 133
pr at max Ipi (MPa) 3.149
Other
deq at max Ipi (cm) 1.841
information 2
Ipi.α at max MI (W/cm ) 115.362
Focal FLx (cm) 4.0 # 2.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B/M B # M M #
Operating 4 4 # 2 2
Focus (cm) #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
22
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm 2 scan
Maximum index value 0.747 # # 1.649 3.120 #
pr, α (MPa) 1.822
P (mW) # # 58.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 31.423
zs (cm) 1.500
zbp (cm) 1.352
Associated
zb (cm) 1.533
acoustic
z at max Ipi.α (cm) 1.500
parameters
deq(zb) (cm) 0.333
fa wf (MHz) 5.948 # # 5.948 5.948 #
Dim of Aaprt X (cm) # # 2.59 2.59 #
Y (cm) # # 0.50 0.50 #
td (µsec) 0.657
prr (Hz) 5500
pr at max Ipi (MPa) 2.497
Other
deq at max Ipi (cm) 0.333
information 2
Ipi.α at max MI (W/cm ) 101.359
Focal FLx (cm) # # 4.0 #
Length FLy (cm # # 2.7 #
Mode ) PW # # PW PW #
Operating 4.0 # 4.0 4.0
Focus # #
control
(cm)
PRF 5.5 # # 5.5 5.5 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

23
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.907 0.846 # # # #
pr, α (MPa) 989
2.199
P (mW) 32.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 5.881 5.881 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.644
prr (Hz) 5000
pr at max Ipi (MPa) 3.011
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 166.299
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 4.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 4.0 4.0 # # # #
control
(cm)
PRF 5 5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
24
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.808 0.513 cm
# # # #
pr, α (MPa) 1.965
P (mW) 22.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 5.919 5.919 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.638
prr (Hz) 5500
pr at max Ipi (MPa) 2.670
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 151.369
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 4.0 4.0 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
25
Acoustic Output Tables

Transducer Model: 6V1 Micro-Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.092 0.699 # 1.807 2.121 #
pr, α (MPa) 989
2.233
P (mW) 37.300 # 63.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 34.447
zs (cm) 1.500

Associated zbp (cm) 1.352


acoustic zb (cm) 1.533
parameters z at max Ipi.α (cm) 1.500
deq(zb) (cm) 0.361
fa wf (MHz) 5.881 5.881 # 5.948 5.948 #
Dim of Aaprt X (cm) 1.28 # 2.59 2.59 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.644
prr (Hz) 5500
pr at max Ipi (MPa) 3.058
Other
deq at max Ipi (cm) # 0.361
information 2
Ipi.α at max MI (W/cm ) 171.477
Focal FLx (cm) 4.0 # 4.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating # 4.0 4.0
Focus (cm) 4.0 4.0 #
control
PRF (kHz) 5.5 5.5 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
26
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.904 0.302 # # # #
pr, α (MPa) 2.071
P (mW) 17.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.000 #
deq(zb) (cm) #
fa wf (MHz) 5.245 5.245 # # # #
Dim of Aaprt X (cm) 1.17 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.260
prr (Hz) 7104
pr at max Ipi (MPa) 2.983
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 183.763
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 3.0 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
27
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.899 0.198 # # # #
pr, α (MPa) 2.067
P (mW) 11.100 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.000 #
deq(zb) (cm) #
fa wf (MHz) 5.276 5.276 # # # #
Dim of Aaprt X (cm) 1.17 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.260
prr (Hz) 7104
pr at max Ipi (MPa) 2.979
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 173.101
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 2.7 # # #
Mode THI # # # # #
Operating
Focus (cm) 3.0 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
28
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.878 0.313 # 0.422 0.149 #
pr, α (MPa) 2.011
P (mW) 17.300 # 16.900 #
min of [Pα(zs),Ita.α(zs)] (mW) 5 6.773
zs (cm) 2.013
Associated zbp (cm) 1.293
acoustic zb (cm) 2.013
parameters z at max Ipi.α (cm) 2.013
deq(zb) (cm) 1.237
fa wf (MHz) 5.245 5.245 # 5.248 5.248 #
Dim of Aaprt X (cm) 1.17 # 3.49 3.49 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.260
prr (Hz) 200
pr at max Ipi (MPa) 2.897
Other
deq at max Ipi (cm) 1.237
information 2
Ipi.α at max MI (W/cm ) 176.216
Focal FLx (cm) 3.0 # 3.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B/M B # M M #
Operating
Focus (cm) 3 3 # 3 3 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
29
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.731 # # 0.784 1.399 #
pr, α (MPa) 1.780
P (mW) # # 27.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 12.267 0
zs (cm) 2.000
zbp (cm) 1.293
Associated
zb (cm) 2.033
acoustic
z at max Ipi.α (cm) 2.000
parameters
deq(zb) (cm) 0.207
fa wf (MHz) 5.922 # # 5.922 5.922 #
Dim of Aaprt X (cm) # # 3.49 3.49 #
Y (cm) # # 0.50 0.50 #
td (µsec) 0.638
prr (Hz) 4000
pr at max Ipi (MPa) 2.698
Other
deq at max Ipi (cm) 0.207
information
Ipi.α at max MI (W/cm2 ) 138.572
Focal FLx (cm) # # 3.0 #
Length FLy (cm # # 2.7 #
Mode ) PW # # PW PW #
Operating 3.0 # 3.0 3.0
Focus # #
control
(cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
30
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.795 0.579 # # # #
pr, α (MPa) 1.930
P (mW) 39.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 2.000 #
parameters
deq(zb) (cm)
fa wf (MHz) 5.899 5.899 # # # #
Dim of Aaprt X (cm) 1.17 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.641
prr (Hz) 4000
pr at max Ipi (MPa) 2.928
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 146.481
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 2.7 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 3.0 3.0 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
31
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.778 0.580 cm
# # # #
pr, α (MPa) 1.890
P (mW) 39.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.000 #
deq(zb) (cm) #
fa wf (MHz) 5.908 5.908 # # # #
Dim of Aaprt X (cm) 1.17 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.641
prr (Hz) 5500
pr at max Ipi (MPa) 2.842
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 158.589
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 3.0 3.0 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
32
Acoustic Output Tables

Transducer Model: 6V3 Micro-Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.921 0.282 # 0.823 1.386 #
pr, α (MPa) 1.898
P (mW) 40.300 # 29.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 12.885
zs (cm) 2.000

Associated zbp (cm) 1.293


acoustic zb (cm) 2.033
parameters z at max Ipi.α (cm) 2.000
deq(zb) (cm) 0.207
fa wf (MHz) 5.899 5.899 # 5.922 5.922 #
Dim of Aaprt X (cm) 1.17 # 3.49 3.49 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.641
prr (Hz) 5000
pr at max Ipi (MPa) 2.881
Other
deq at max Ipi (cm) # 0.207
information 2
Ipi.α at max MI (W/cm ) 141.754
Focal FLx (cm) 3.0 # 3.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating # 3.0 3.0
Focus (cm) 3.0 3.0 #
control
PRF (kHz) 5.0 5.0 # 4.0 4.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
33
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.923 0.827 # # # #
pr, α (MPa) 1.619
P (mW) 76.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.000 #
deq(zb) (cm) #
fa wf (MHz) 3.075 3.075 # # # #
Dim of Aaprt X (cm) 2.496 # # # #
Y (cm) 1.245 # # # #
td (µsec) 0.320
prr (Hz) 6272
pr at max Ipi (MPa) 2.653
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 109.027
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 8.1 # # #
Mode B # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
34
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.918 1.015 # # # #
pr, α (MPa) 1.642
P (mW) 93.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.000 #
deq(zb) (cm) #
fa wf (MHz) 3.200 3.200 # # # #
Dim of Aaprt X (cm) 2.496 # # # #
Y (cm) 1.245 # # # #
td (µsec) 0.305
prr (Hz) 6272
pr at max Ipi (MPa) 2.782
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 109.799
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 8.1 # # #
Mode THI # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
35
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.170 0.831 # 0.145 0.382 #
pr, α (MPa) 1.843
P (mW) 76.400 # 94.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 10.219
zs (cm) 4.746
Associated zbp (cm) 2.979
acoustic zb (cm) 4.746
parameters z at max Ipi.α (cm) 4.764
deq(zb) (cm) 2.104
fa wf (MHz) 3.075 3.075 # 2.989 2.989 #
Dim of Aaprt X (cm) 2.496 # 4.992 4.992 #
Y (cm) 1.245 # 1.245 1.245 #
td (µsec) 0.321
prr (Hz) 150
pr at max Ipi (MPa) 3.009
Other
deq at max Ipi (cm) 2.069
information 2
Ipi.α at max MI (W/cm ) 154.771
Focal FLx 6.0 # 6.0 #
Length (cm)
FLy 8.1 # 8.1 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 6 6 # 6 6 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
36
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.532 # # 2.219 3.469 #


pr, α (MPa) 0.955
P (mW) # # 274.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 133.185
zs (cm) 2.985
zbp (cm) 2.979
Associated
zb (cm) 3.035
acoustic
z at max Ipi.α (cm) 2.985
parameters
deq(zb) (cm) 0.959
fa wf (MHz) 3.499 # # 3.499 3.499
Dim of Aaprt X (cm) # # 2.496 2.496 #
Y (cm) # # 1.245 1.245 #
td (µsec) 1.002
prr (Hz) 4000
pr at max Ipi (MPa) 1.417
Other
deq at max Ipi (cm) 0.932
information
Ipi.α at max MI (W/cm2 ) 35.767
Focal FLx (cm) # # 6.0 #
Length FLy (cm) # # 8.1 #
Mode PW # # PW PW #
Operating Focus 6 # # 6 6 #
control (cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
37
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.988 2.538 # # # #
pr, α (MPa) 1.769
P (mW) 210.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 3.000 #
parameters
deq(zb) (cm)
fa wf (MHz) 3.036 3.036 # # # #
Dim of Aaprt X (cm) 2.496 # # # #
Y (cm) 1.245 # # # #
td (µsec) 1.248
prr (Hz) 5000
pr at max Ipi (MPa) 2.736
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 118.967
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 8.1 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 6.0 6.0 # # # #
control
(cm)
PRF 5 5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to e nter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
38
Acoustic Output Tables

Transducer Model: C344 Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.814 2.537 cm
# # # #
pr, α (MPa) 1.417
P (mW) 210.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.985 #
deq(zb) (cm) #
fa wf (MHz) 3.033 3.033 # # # #
Dim of Aaprt X (cm) 2.496 # # # #
Y (cm) 1.245 # # # #
td (µsec) 1.254
prr (Hz) 5500
pr at max Ipi (MPa) 2.405
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 95.253
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 8.1 # # #
Mode DPI B-DPI # # # #
Operating
Focus 6 6 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
39
Acoustic Output Tables

Transducer Model: C344 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.969 0.785 # 2.173 3.680 #
pr, α (MPa) 1.803
P (mW) 207.300 # 268.30 #
min of [Pα(zs),Ita.α(zs)] (mW) 130.414 0
zs (cm) 2.985

Associated zbp (cm) 2.979


acoustic zb (cm) 3.035
parameters z at max Ipi.α (cm) 2.836
deq(zb) (cm) 0.885
fa wf (MHz) 3.036 3.036 # 3.499 3.499 #
Dim of Aaprt X (cm) 2.496 # 4.992 4.992 #
Y (cm) 1.245 # 1.245 1.245 #
td (µsec) 1.248
prr (Hz) 5500
pr at max Ipi (MPa) 2.836
Other
deq at max Ipi (cm) # 0.860
information 2
Ipi.α at max MI (W/cm ) 127.877
Focal FLx (cm) 7.0 # 6.0 #
Length FLy (cm) 8.1 # 8.1 #
Mode B B/COLOR # PW PW #
Operating 7 7 # 6 6
Focus (cm) #
control
PRF (kHz) 5.5 5.5 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
40
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.176 1.134 # # # #
pr, α (MPa) 2.012
P (mW) 142.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.214 #
deq(zb) (cm) #
fa wf (MHz) 2.930 2.930 # # # #
Dim of Aaprt X (cm) 3.05 # # # #
Y (cm) 1.185 # # # #
td (µsec) 0.355
prr (Hz) 4352
pr at max Ipi (MPa) 3.205
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 176.101
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode B # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
41
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.160 0.908 # # # #
pr, α (MPa) 1.928
P (mW) 114.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.214 #
deq(zb) (cm) #
fa wf (MHz) 2.924 2.924 # # # #
Dim of Aaprt X (cm) 3.05 # # # #
Y (cm) 1.185 # # # #
td (µsec) 0.332
prr (Hz) 4352
pr at max Ipi (MPa) 3.123
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 178.882
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
42
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.198 1.141 # 0.179 0.554 #
pr, α (MPa) 3.031
P (mW) 142.700 # 139.900 #
min of [Pα(zs),Ita.α(zs)] (mW) 13.076
zs (cm) 4.365
Associated zbp (cm) 3.213
acoustic zb (cm) 4.365
parameters z at max Ipi.α (cm) 4.365
deq(zb) (cm) 2.392
fa wf (MHz) 2.930 2.930 # 2.923 2.923 #
Dim of Aaprt X (cm) 3.05 # 6.1 6.1 #
Y (cm) 1.185 # 1.185 1.185 #
td (µsec) 0.340
prr (Hz) 133
pr at max Ipi (MPa) 3.134
Other
deq at max Ipi (cm) 2.392
information 2
Ipi.α at max MI (W/cm ) 187.035
Focal FLx 6.0 # 6.0 #
Length (cm)
FLy 6.0 # 6.0 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 6 6 # 6 6 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
43
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 1.055 # # 1.537 5.150 #


pr, α (MPa) 1.833
P (mW) # # 209.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 106.892
zs (cm) 3.214
zbp (cm) 3.213
Associated
zb (cm) 4.699
acoustic
z at max Ipi.α (cm) 3.214
parameters
deq(zb) (cm) 0.287
fa wf (MHz) 3.020 # # 3.020 3.020
Dim of Aaprt X (cm) # # 6.1 6.1 #
Y (cm) # # 1.185 1.185 #
td (µsec) 1.257
prr (Hz) 4000
pr at max Ipi (MPa) 2.993
Other
deq at max Ipi (cm) 0.284
information
Ipi.α at max MI (W/cm2 ) 141.991
Focal FLx (cm) # # 6 #
Length FLy (cm) # # 6 #
Mode PW # # PW PW #
Operating Focus 6 # # 6 6 #
control (cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
44
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.197 1.557 # # # #
pr, α (MPa) 2.085
P (mW) 191.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 3.214 #
parameters
deq(zb) (cm)
fa wf (MHz) 3.034 3.034 # # # #
Dim of Aaprt X (cm) 3.05 # # # #
Y (cm) 1.185 # # # #
td (µsec) 1.250
prr (Hz) 3500
pr at max Ipi (MPa) 3.149
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 234.247
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 6 6 # # # #
control
(cm)
PRF 3.5 3.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
45
Acoustic Output Tables

Transducer Model: C354 Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 1.248 1.558 cm
# # # #
pr, α (MPa) 2.174
P (mW) 191.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.214 #
deq(zb) (cm) #
fa wf (MHz) 3.035 3.035 # # # #
Dim of Aaprt X (cm) 3.05 # # # #
Y (cm) 1.185 # # # #
td (µsec) 1.248
prr (Hz) 4000
pr at max Ipi (MPa) 3.316
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 252.168
Focal FLx (cm) 6 # # #
Length FLy (cm) 6 # # #
Mode DPI B-DPI # # # #
Operating
Focus 6 6 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
46
Acoustic Output Tables

Transducer Model: C354 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.194 1.118 # 1.534 5.309 #
pr, α (MPa) 2.143
P (mW) 192.100 # 208.60 #
min of [Pα(zs),Ita.α(zs)] (mW) 106.688 0
zs (cm) 3.214

Associated zbp (cm) 3.213


acoustic zb (cm) 4.699
parameters z at max Ipi.α (cm) 3.214
deq(zb) (cm) 0.245
fa wf (MHz) 3.034 3.034 # 3.151 3.151 #
Dim of Aaprt X (cm) 3.05 # 6.1 6.1 #
Y (cm) 1.185 # 1.185 1.185 #
td (µsec) 1.250
prr (Hz) 4000
pr at max Ipi (MPa) 3.237
Other
deq at max Ipi (cm) # 0.243
information 2
Ipi.α at max MI (W/cm ) 247.527
Focal FLx (cm) 6 # 6 #
Length FLy (cm) 6 # 6 #
Mode B B/COLOR # PW PW #
Operating 6 6 # 6 6
Focus (cm) #
control
PRF (kHz) 4.0 4.0 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
47
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.080 0.800 # # # #
pr, α (MPa) 1.883
P (mW) 131.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.500 #
deq(zb) (cm) #
fa wf (MHz) 3.038 3.038 # # # #
Dim of Aaprt X (cm) 2.50 # # # #
Y (cm) 1.33 # # # #
td (µsec) 0.227
prr (Hz) 4416
pr at max Ipi (MPa) 3.149
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 225.765
Focal FLx (cm) 8.0 # # #
Length FLy (cm) 8.1 # # #
Mode B # # # # #
Operating
Focus (cm) 8 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
48
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.917 0.726 # # # #
pr, α (MPa) 1.708
P (mW) 100.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.500 #
deq(zb) (cm) #
fa wf (MHz) 3.471 3.471 # # # #
Dim of Aaprt X (cm) 2.5 # # # #
Y (cm) 1.33 # # # #
td (µsec) 0.209
prr (Hz) 5184
pr at max Ipi (MPa) 3.082
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 158.986
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
49
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.991 0.794 # 0.130 0.356 #
pr, α (MPa) 1.726
P (mW) 131.800 # 99.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.987
zs (cm) 5.223
Associated zbp (cm) 3.082
acoustic zb (cm) 5.223
parameters z at max Ipi.α (cm) 5.223
deq(zb) (cm) 2.175
fa wf (MHz) 3.038 3.038 # 3.236 3.236 #
Dim of Aaprt X (cm) 2.5 # 7.48 7.48 #
Y (cm) 1. 33 # 1.33 1.33 #
td (µsec) 0.231
prr (Hz) 150
pr at max Ipi (MPa) 2.984
Other
deq at max Ipi (cm) 2.140
information 2
Ipi.α at max MI (W/cm ) 174.116
Focal FLx 8.0 # 6.0 #
Length (cm)
FLy 8.0 # 6.0 #
Mode (cm) B/M B # M M #
Operating 8 8 # 6 6
Focus (cm) #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
50
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.621 # # 1.330 3.041 #


pr, α (MPa) 1.168
P (mW) # # 185.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 78.793
zs (cm) 3.500
zbp (cm) 3.082
Associated
zb (cm) 6.652
acoustic
z at max Ipi.α (cm) 3.500
parameters
deq(zb) (cm) 0.351
fa wf (MHz) 3.651 # # 3.651 3.651
Dim of Aaprt X (cm) # # 7.48 7.48 #
Y (cm) # # 1.33 1.33 #
td (µsec) 1.071
prr (Hz) 4000
pr at max Ipi (MPa) 2.639
Other
deq at max Ipi (cm) 0.351
information
Ipi.α at max MI (W/cm2 ) 72.636
Focal FLx (cm) # # 6 #
Length FLy (cm) # # 6 #
Mode PW # # PW PW #
Operating 6 # 6 6
Focus # #
control (cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
51
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.868 0.872 # # # #
pr, α (MPa) 1.514
P (mW) 125.100 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 3.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 3.042 3.042 # # # #
Dim of Aaprt X (cm) 2.5 # # # #
Y (cm) 1.33 # # # #
td (µsec) 1.246
prr (Hz) 3000
pr at max Ipi (MPa) 2.567
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 106.885
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 6 6 # # # #
control
(cm)
PRF 3.0 3.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
52
Acoustic Output Tables

Transducer Model: C362 Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.927 0.874 cm
# # # #
pr, α (MPa) 1.620
P (mW) 125.100 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.500 #
deq(zb) (cm) #
fa wf (MHz) 3.050 3.050 # # # #
Dim of Aaprt X (cm) 2.5 # # # #
Y (cm) 1.33 # # # #
td (µsec) 1.244
prr (Hz) 4000
pr at max Ipi (MPa) 2.782
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 127.123
Focal FLx (cm) 6 # # #
Length FLy (cm) 6 # # #
Mode DPI B-DPI # # # #
Operating
Focus 6 6 # # # #
control
(cm)
PRF 4 4 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
53
Acoustic Output Tables

Transducer Model: C362 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.110 0.878 # 1.324 2.842 #
pr, α (MPa) 1.562
P (mW) 128.300 # 183.40 #
min of [Pα(zs),Ita.α(zs)] (mW) 77.817 0
zs (cm) 3.500

Associated zbp (cm) 3.082


acoustic zb (cm) 5.023
parameters z at max Ipi.α (cm) 3.500
deq(zb) (cm) 0.371
fa wf (MHz) 3.042 3.042 # 3.512 3.512 #
Dim of Aaprt X (cm) 2.5 # 7.48 7.48 #
Y (cm) 1.33 # 1.33 1.33 #
td (µsec) 1.246
prr (Hz) 4000
pr at max Ipi (MPa) 2.647
Other
deq at max Ipi (cm) # 0.243
information 2
Ipi.α at max MI (W/cm ) 113.653
Focal FLx (cm) 6 # 6 #
Length FLy (cm) 6 # 6 #
Mode B B/COLOR # PW PW #
Operating 6 6 # 6 6
Focus (cm) #
control
PRF (kHz) 4.0 4.0 # 3.5 3.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
54
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.923 0.577 # # # #
pr, α (MPa) 2.266
P (mW) 24.200 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.000 #
deq(zb) (cm) #
fa wf (MHz) 6.030 6.030 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.241
prr (Hz) 6803
pr at max Ipi (MPa) 3.437
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 223.123
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
55
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.953 0.732 # # # #
pr, α (MPa) 2.344
P (mW) 29.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.950 #
deq(zb) (cm) #
fa wf (MHz) 6.056 6.056 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.237
prr (Hz) 6803
pr at max Ipi (MPa) 3.524
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 241.399
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
56
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.999 0.526 # 0.664 0.177 #
pr, α (MPa) 2.376
P (mW) 24.200 # 23.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 7.352
zs (cm) 1.957
Associated zbp (cm) 1.352
acoustic zb (cm) 1.957
parameters z at max Ipi.α (cm) 1.957
deq(zb) (cm) 1.357
fa wf (MHz) 6.030 6.030 # 5.904 5.904 #
Dim of Aaprt X (cm) 1.28 # 2.58 2.58 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.228
prr (Hz) 133
pr at max Ipi (MPa) 3.618
Other
deq at max Ipi (cm) 1.357
information 2
Ipi.α at max MI (W/cm ) 242.581
Focal FLx 2.0 # 2.0 #
Length (cm)
FLy 2.7 # 2.7 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
57
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.778 # # 0.776 1.337 #


pr, α (MPa) 1.887
P (mW) # # 27.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 12.539
zs (cm) 1.950
zbp (cm) 1.352
Associated
zb (cm) 1.957
acoustic
z at max Ipi.α (cm) 1.950
parameters
deq(zb) (cm) 0.211
fa wf (MHz) 5.884 # # 5.884 5.884
Dim of Aaprt X (cm) # # 2.58 2.58 #
Y (cm) # # 0.50 0.50 #
td (µsec) 0.631
prr (Hz) 5000
pr at max Ipi (MPa) 2.808
Other
deq at max Ipi (cm) 0.211
information
Ipi.α at max MI (W/cm2 ) 113.284
Focal FLx (cm) # # 2.0 #
Length FLy (cm) # # 2.7 #
Mode PW # # PW PW #
Operating Focus 2 # # 2 2 #
control (cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
58
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.895 0.895 # # # #
pr, α (MPa) 2.185
P (mW) 32.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.950 #
parameters
deq(zb) (cm)
fa wf (MHz) 5.964 5.964 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.635
prr (Hz) 5000
pr at max Ipi (MPa) 3.269
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 189.480
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2 2 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
59
Acoustic Output Tables

Transducer Model: C611 Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.749 0.891 cm
# # # #
pr, α (MPa) 1.824
P (mW) 32.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.950 #
deq(zb) (cm) #
fa wf (MHz) 5.937 5.937 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.639
prr (Hz) 5000
pr at max Ipi (MPa) 2.721
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 107.114
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
60
Acoustic Output Tables
Transducer Model: C611 Convex Array Operating Mode: CW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm
2 scan
Maximum index value 0.170 # # 0.451 2.201 0.601
pr, α (MPa) 0.356
P (mW) # # 27.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 10.208
zs (cm) 2.500
zbp (cm) 1.352
Associated
zb (cm) 2.642
acoustic
z at max Ipi.α (cm) 2.500
parameters
deq(zb) (cm) 0.053
fa wf (MHz) 4.367 # # 4.367 4.367 #
Dim of Aaprt X (cm) # # 2.58 2.58 #
Y (cm) # # 0.5 0.5 #
td (µsec) 1.037
prr (Hz) #
pr at max Ipi (MPa) 0.531
Other
deq at max Ipi (cm) 0.051
information
Ipi.α at max MI (W/cm2 ) 4.661
Focal FLx (cm) # # 4.0 #
Length FLy (cm) # # 2.7 #
Mode CW # # CW CW #
Operating 4 # 4 4
Focus # #
control (cm)
PRF 8.0 # # 8.0 8.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
61
Acoustic Output Tables
Transducer Model: C611 Convex Array Operating Mode: TDI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm 2 scan
2
Maximum index value 1.164 0.598 cm
# # # #
pr, α (MPa) 2.314
P (mW) 33.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.500 #
deq(zb) (cm) #
fa wf (MHz) 3.951 3.951 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.971
prr (Hz) 5500
pr at max Ipi (MPa) 3.525
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 185.794
Focal FLx (cm) 4.0 # # #
Length FLy (cm) 2.7 # # #
Mode TDI B-TDI # # # #
Operating 4 4
Focus # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
62
Acoustic Output Tables

Transducer Model: C611 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.935 0.551 # 0.790 1.400 #
pr, α (MPa) 2.214
P (mW) 33.100 # 28.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 12.765
zs (cm) 1.950

Associated zbp (cm) 1.352


acoustic zb (cm) 1.957
parameters z at max Ipi.α (cm) 1.950
deq(zb) (cm) 0.205
fa wf (MHz) 5.964 5.964 # 5.884 5.884 #
Dim of Aaprt X (cm) 1.28 # 2.58 2.58 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.635
prr (Hz) 4000
pr at max Ipi (MPa) 3.313
Other
deq at max Ipi (cm) # 0.205
information 2
Ipi.α at max MI (W/cm ) 194.532
Focal FLx (cm) 2.0 # 2.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating # 2 2
Focus (cm) 2 2 #
control
PRF (kHz) 4.0 4.0 # 5.5 5.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
63
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.803 0.745 # # # #
pr, α (MPa) 1.927
P (mW) 35.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 5.755 5.755 # # # #
Dim of Aaprt X (cm) 485 1.02 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.231
prr (Hz) 7042
pr at max Ipi (MPa) 2.610
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 139.410
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
64
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.793 0.740 # # # #
pr, α (MPa) 1.970
P (mW) 32.3000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 6.164 6.164 # # # #
Dim of Aaprt X (cm) 1.02 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.222
prr (Hz) 7042
pr at max Ipi (MPa) 2.7184
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 127.550
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
65
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.799 0.724 # 0.634 0.148 #
pr, α (MPa) 1.915
P (mW) 35.300 # 23.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 4.345
zs (cm) 1.540
Associated zbp (cm) 1.207
acoustic zb (cm) 1.540
parameters z at max Ipi.α (cm) 1.540
deq(zb) (cm) 1.921
fa wf (MHz) 5.755 5.755 # 5.740 5.740 #
Dim of Aaprt X (cm) 1.02 # 2.05 2.05 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.232
prr (Hz) 133
pr at max Ipi (MPa) 2.599
Other
deq at max Ipi (cm) 1.910
information 2
Ipi.α at max MI (W/cm ) 138.198
Focal FLx 2.0 # 2.0 #
Length (cm)
FLy 2.7 # 2.7 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
66
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.567 # # 0.925 1.507 #


pr, α (MPa) 1.279
P (mW) # # 38.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 22.555
zs (cm) 1.767
zbp (cm) 1.207
Associated
zb (cm) 1.767
acoustic
z at max Ipi.α (cm) 1.500 .767
parameters
deq(zb) (cm) 0.314
fa wf (MHz) 5.085 # # 5.085 5.085
Dim of Aaprt X (cm) # # 2.05 2.05 #
Y (cm) # # 0.50 0.50 #
td (µsec) 0.763
prr (Hz) 5000
pr at max Ipi (MPa) 1.745
Other
deq at max Ipi (cm) 0.312
information
Ipi.α at max MI (W/cm2 ) 75.197
Focal FLx (cm) # # 2.0 #
Length FLy (cm) # # 2.7 #
Mode PW # # PW PW #
Operating Focus 2.0 # # 2.0 2.0 #
control (cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
67
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.665 1.742 # # # #
pr, α (MPa) 1.497
P (mW) 85.200 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 5.072 5.072 # # # #
Dim of Aaprt X (cm) 1.02 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.758
prr (Hz) 5500
pr at max Ipi (MPa) 1.956
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 101.022
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2.0 2.0 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
68
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.660 1.821 cm
# # # #
pr, α (MPa) 1.489
P (mW) 94.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 5.086 5.086 # # # #
Dim of Aaprt X (cm) 1.02 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.760
prr (Hz) 5500
pr at max Ipi (MPa) 1.958
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 97.916
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
69
Acoustic Output Tables

Transducer Model: EC9-5 Micro-Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.848 0.869 # 0.983 1.764 #
pr, α (MPa) 1.515
P (mW) 87.300 # 40.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 23.972
zs (cm) 1.500

Associated zbp (cm) 1.207


acoustic zb (cm) 1.767
parameters z at max Ipi.α (cm) 1.500
deq(zb) (cm) 0.285
fa wf (MHz) 5.072 5.072 # 5.085 5.085 #
Dim of Aaprt X (cm) 1.02 # 2.05 2.05 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.758
prr (Hz) 5000
pr at max Ipi (MPa) 1.956
Other
deq at max Ipi (cm) # 0.283
information 2
Ipi.α at max MI (W/cm ) 103.373
Focal FLx (cm) 2.0 # 2.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating # 2.0 2.0
Focus (cm) 2 2 #
control
PRF (kHz) 5.0 5.0 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
70
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.764 0.220 # # # #
pr, α (MPa) 1.963
P (mW) 24.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.817 #
deq(zb) (cm) #
fa wf (MHz) 6.800 6.800 # # # #
Dim of Aaprt X (cm) 2.3 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.123
prr (Hz) 7040
pr at max Ipi (MPa) 3.053
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 228.265
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
71
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.769 0.203 # # # #
pr, α (MPa) 1.997
P (mW) 21.200 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.817 #
deq(zb) (cm) #
fa wf (MHz) 6.754 6.754 # # # #
Dim of Aaprt X (cm) 2.3 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.124
prr (Hz) 7040
pr at max Ipi (MPa) 2.951
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 216.911
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode THI, # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
72
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.781 0.615 # 0.145 0.221 #
pr, α (MPa) 2.014
P (mW) 24.500 # 34.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 4.587
zs (cm) 1.821
Associated zbp (cm) 1.812
acoustic zb (cm) 1.821
parameters z at max Ipi.α (cm) 1.821
deq(zb) (cm) 2.039
fa wf (MHz) 6.800 6.800 # 6.654 6.654 #
Dim of Aaprt X (cm) 2.3 # 4.6 4.6 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.123
prr (Hz) 150
pr at max Ipi (MPa) 3.060
Other
deq at max Ipi (cm) 2.029
information 2
Ipi.α at max MI (W/cm ) 226.090
Focal FLx 2.0 # 2.0 #
Length (cm)
FLy 2.7 # 2.7 #
Mode (cm) B/M B # M M #
Operating 2 2 # 2 2
Focus (cm) #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
73
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.502 # # 0.336 3.121 #


pr, α (MPa) 1.124
P (mW) # # 26.400 #
min of [Pα(zs),Ita.α(zs)] (mW) 14.072
zs (cm) 1.815
zbp (cm) 1.812
Associated
zb (cm) 1.828
acoustic
z at max Ipi.α (cm) 1.815
parameters
deq(zb) (cm) 0.263
fa wf (MHz) 5.019 # # 5.019 5.019
Dim of Aaprt X (cm) # # 4.6 4.6 #
Y (cm) # # 0.5 0.5 #
td (µsec) 0.759
prr (Hz) 5000
pr at max Ipi (MPa) 1.543
Other
deq at max Ipi (cm) 0.263
information
Ipi.α at max MI (W/cm2 ) 66.279
Focal FLx (cm) # # 2.0 #
Length FLy (cm) # # 2.7 #
Mode PW # # PW PW #
Operating 2.0 # 2.0 2.0
Focus # #
control (cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
74
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.740 0.255 # # # #
pr, α (MPa) 1.964
P (mW) 24.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.817 #
parameters
deq(zb) (cm)
fa wf (MHz) 7.050 7.050 # # # #
Dim of Aaprt X (cm) 2.3 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.545
prr (Hz) 5500
pr at max Ipi (MPa) 3.082
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 162.890
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2 2 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
75
Acoustic Output Tables

Transducer Model: L741 Linear Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.816 0.231 cm
# # # #
pr, α (MPa) 2.163
P (mW) 24.500 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.817 #
deq(zb) (cm) #
fa wf (MHz) 7.030 7.030 # # # #
Dim of Aaprt X (cm) 2.3 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.561
prr (Hz) 5000
pr at max Ipi (MPa) 3.384
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 157.010
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 2.7 2.7 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
76
Acoustic Output Tables

Transducer Model: L741 Linear Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.758 0.230 # 0.347 3.400 #
pr, α (MPa) 1.969
P (mW) 25.700 # 23.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 14.499
zs (cm) 1.821

Associated zbp (cm) 1.812


acoustic zb (cm) 1.828
parameters z at max Ipi.α (cm) 1.817
deq(zb) (cm) 0.248
fa wf (MHz) 7.050 7.050 # 5.019 5.019 #
Dim of Aaprt X (cm) 2.3 # 4.6 4.6 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.545
prr (Hz) 5000
pr at max Ipi (MPa) 3.090
Other
deq at max Ipi (cm) # 0.248
information 2
Ipi.α at max MI (W/cm ) 163.767
Focal FLx (cm) 2.0 # 2.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating 2 2 #
Focus (cm) 2. 2 #
control
PRF (kHz) 5.0 5.0 # 5.5 5.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
77
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.978 0.503 # # # #
pr, α (MPa) 2.676
P (mW) 29.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.450 #
deq(zb) (cm) #
fa wf (MHz) 7.511 7.511 # # # #
Dim of Aaprt X (cm) 485 1.28 # # # #
Y (cm) 0.36 # # # #
td (µsec) 0.114
prr (Hz) 4773
pr at max Ipi (MPa) 3.847
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 413.053
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.4 # # #
Mode B # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
78
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.018 0.413 # # # #
pr, α (MPa) 2.749
P (mW) 25.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.450 #
deq(zb) (cm) #
fa wf (MHz) 7.310 7.310 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.36 # # # #
td (µsec) 0.113
prr (Hz) 4773
pr at max Ipi (MPa) 3.960
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 393.469
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.4 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
79
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.978 0.514 # 0.706 0.187 #
pr, α (MPa) 2.676
P (mW) 29.600 # 19.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.291
zs (cm) 1.463
Associated zbp (cm) 1.147
acoustic zb (cm) 1.463 1.463
parameters z at max Ipi.α (cm) 1.816 1.816
deq(zb) (cm) 1.123
fa wf (MHz) 7.511 7.511 # 7.507 7.507 #
Dim of Aaprt X (cm) 485 1.28 # 3.84 3.84 #
Y (cm) 0.36 # 0.36 0.36 #
td (µsec) 0.114
prr (Hz) 200
pr at max Ipi (MPa) 3.907
Other
deq at max Ipi (cm) 1.123
information 2
Ipi.α at max MI (W/cm ) 413.053
Focal FLx 2.0 # 2.0 #
Length (cm)
FLy 2.4 # 2.4 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
80
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 1.360 # # 0.633 1.863 #


pr, α (MPa) 3.804
P (mW) # # 17.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.199
zs (cm) 1.350
zbp (cm) 1.147
Associated
zb (cm) 1.350
acoustic
z at max Ipi.α (cm) 1.350
parameters
deq(zb) (cm) 0.084
fa wf (MHz) 7.832 # # 7.832 7.832
Dim of Aaprt X (cm) # # 3.84 3.84 #
Y (cm) # # 0.36 0.36 #
td (µsec) 0.485
prr (Hz) 5000
pr at max Ipi (MPa) 5.478
Other
deq at max Ipi (cm) 0.084
information
Ipi.α at max MI (W/cm2 ) 611.925
Focal FLx (cm) # # 2.0 #
Length FLy (cm) # # 2.4 #
Mode PW # # PW PW #
Operating # 2.0 2.0
Focus 2.0 # #
control (cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
81
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.842 0.391 # # # #
pr, α (MPa) 2.241
P (mW) 17.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 7.112 7.112 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.36 # # # #
td (µsec) 0.528
prr (Hz) 3500
pr at max Ipi (MPa) 3.287
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 179.775
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.4 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2.0 2.0 # # # #
control
(cm)
PRF 3.5 3.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
82
Acoustic Output Tables

Transducer Model: L742 Linear Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.836 0.392 cm
# # # #
pr, α (MPa) 2.225
P (mW) 17.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 7.095 7.095 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.36 # # # #
td (µsec) 0.530
prr (Hz) 4000
pr at max Ipi (MPa) 3.287
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 149.825
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.4 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2.0 2.0 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
83
Acoustic Output Tables

Transducer Model: L742 Linear Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.963 0.520 # 0.685 2.017 #
pr, α (MPa) 2.268
P (mW) 19.200 # 18.400 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.874
zs (cm) 1.350

Associated zbp (cm) 1.147


acoustic zb (cm) 1.350
parameters z at max Ipi.α (cm) 1.500
deq(zb) (cm) 0.085
fa wf (MHz) 7.091 7.091 # 7.826 7.826 #
Dim of Aaprt X (cm) 1.28 # 3.84 3.84 #
Y (cm) 0.36 # 0.36 0.36 #
td (µsec) 0.528
prr (Hz) 4000
pr at max Ipi (MPa) 3.328
Other
deq at max Ipi (cm) # 0.085
information 2
Ipi.α at max MI (W/cm ) 184.262
Focal FLx (cm) 2.0 # 2.0 #
Length FLy (cm) 2.4 # 2.4 #
Mode B B/COLOR # PW PW #
Operating 2.0 2.0 #
Focus (cm) 2.0 2.0 #
control
PRF (kHz) 4.0 4.0 # 5.5 5.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
84
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.877 0.496 # # # #
pr, α (MPa) 2.304
P (mW) 27.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.501 #
deq(zb) (cm) #
fa wf (MHz) 6.902 6.902 # # # #
Dim of Aaprt X (cm) 485 1.57 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.108
prr (Hz) 7104
pr at max Ipi (MPa) 3.319
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 225.350
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode B # # # # #
Operating
Focus (cm) 2.0 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
85
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.836 0.487 # # # #
pr, α (MPa) 2.221
P (mW) 26.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.501 #
deq(zb) (cm) #
fa wf (MHz) 7.058 7.058 # # # #
Dim of Aaprt X (cm) 1.57 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.109
prr (Hz) 7104
pr at max Ipi (MPa) 3.202
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 290.704
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
86
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.845 0.501 # 1.140 1.209 #
pr, α (MPa) 2.255
P (mW) 27.600 # 33.600 #
min of [Pα(zs),Ita.α(zs)] (mW) 6.284
zs (cm) 1.526
Associated zbp (cm) 1.497
acoustic zb (cm) 1.526
parameters z at max Ipi.α (cm) 1.526
deq(zb) (cm) 1.793
fa wf (MHz) 6.902 6.902 # 7.122 7.122 #
Dim of Aaprt X (cm) 1.57 # 4.7 4.7 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.107
prr (Hz) 200
pr at max Ipi (MPa) 3.283
Other
deq at max Ipi (cm) 1.783
information 2
Ipi.α at max MI (W/cm ) 278.860
Focal FLx 2.0 # # #
Length (cm)
FLy 2.7 # # #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
87
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.754 # # 0.706 1.089 #


pr, α (MPa) 2.094
P (mW) # # 19.200 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.624
zs (cm) 1.501
zbp (cm) 1.497
Associated
zb (cm) 2.317
acoustic
z at max Ipi.α (cm) 1.501
parameters
deq(zb) (cm) 0.120
fa wf (MHz) 7.720 # # 7.720 7.720
Dim of Aaprt X (cm) # # 4.7 4.7 #
Y (cm) # # 0.5 0.5 #
td (µsec) 0.513
prr (Hz) 5000
pr at max Ipi (MPa) 3.886
Other
deq at max Ipi (cm) 0.120
information
Ipi.α at max MI (W/cm2 ) 186.920
Focal FLx (cm) # # 2.0 #
Length FLy (cm) # # 2.7 #
Mode PW # # PW PW #
Operating Focus 2 # # 2 2 #
control (cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
88
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.745 0.618 # # # #
pr, α (MPa) 1.990
P (mW) 29.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.501 #
parameters
deq(zb) (cm)
fa wf (MHz) 7.128 7.128 # # # #
Dim of Aaprt X (cm) 1.57 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.557
prr (Hz) 4000
pr at max Ipi (MPa) 3.109
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 134.660
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2 2 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
89
Acoustic Output Tables

Transducer Model: L743 Linear Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.681 0.618 cm
# # # #
pr, α (MPa) 1.817
P (mW) 29.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.501 #
deq(zb) (cm) #
fa wf (MHz) 7.130 7.130 # # # #
Dim of Aaprt X (cm) 1.57 # # # #
Y (cm) 0.50 # # # #
td (µsec) 0.551
prr (Hz) 5000
pr at max Ipi (MPa) 2.850
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 106.815
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.7 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
90
Acoustic Output Tables

Transducer Model: L743 Linear Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.846 0.519 # 0.750 1.273 #
pr, α (MPa) 2.007
P (mW) 31.300 # 20.400 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.163
zs (cm) 1.501

Associated zbp (cm) 1.497


acoustic zb (cm) 2.317
parameters z at max Ipi.α (cm) 1.501
deq(zb) (cm) 0.109
fa wf (MHz) 7.128 7.128 # 7.720 7.720 #
Dim of Aaprt X (cm) 1.57 # 4.7 4.7 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.557
prr (Hz) 5000
pr at max Ipi (MPa) 3.136
Other
deq at max Ipi (cm) # 0.513
information 2
Ipi.α at max MI (W/cm ) 136.971
Focal FLx (cm) 2.5 # 2.0 #
Length FLy (cm) 2.7 # 2.7 #
Mode B B/COLOR # PW PW #
Operating 2.5 2.5 #
Focus (cm) 2 2 #
control
PRF (kHz) 5.0 5.0 # 5.5 5.5 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
91
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.077 0.307 # # # #
pr, α (MPa) 1.847
P (mW) 47.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.875 #
deq(zb) (cm) #
fa wf (MHz) 2.942 2.942 # # # #
Dim of Aaprt X (cm) 485 2.4 # # # #
Y (cm) 1.2 # # # #
td (µsec) 0.237
prr (Hz) 5888
pr at max Ipi (MPa) 2.867
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 171.033
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode B # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
92
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.965 0.260 # # # #
pr, α (MPa) 1.712
P (mW) 37.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.875 #
deq(zb) (cm) #
fa wf (MHz) 3.144 3.144 # # # #
Dim of Aaprt X (cm) 2.4 # # # #
Y (cm) 1.2 # # # #
td (µsec) 0.224
prr (Hz) 5888
pr at max Ipi (MPa) 2.631
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 207.308
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 6 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
93
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.019 0.312 # 0.165 0.348 #
pr, α (MPa) 1.753
P (mW) 47.400 # 60.500 #
min of [Pα(zs),Ita.α(zs)] (mW) 11.718
zs (cm) 4.172
Associated zbp (cm) 2.868
acoustic zb (cm) 4.172
parameters z at max Ipi.α (cm) 4.172
deq(zb) (cm) 1.674
fa wf (MHz) 2.942 2.942 # 2.958 2.958 #
Dim of Aaprt X (cm) 2.4 # 4.8 4.8 #
Y (cm) 1.2 # 1.2 1.2 #
td (µsec) 0.234
prr (Hz) 200
pr at max Ipi (MPa) 2.685
Other
deq at max Ipi (cm) 1.674
information 2
Ipi.α at max MI (W/cm ) 117.593
Focal FLx 6.0 # 6.0 #
Length (cm)
FLy 6.0 # 6.0 #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 6 6 # 6 6 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
94
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.637 # # 0.525 1.498 #


pr, α (MPa) 1.197
P (mW) # # 63.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 31.260
zs (cm) 2.875
zbp (cm) 2.868
Associated
zb (cm) 4.734
acoustic
z at max Ipi.α (cm) 2.875
parameters
deq(zb) (cm) 0.299
fa wf (MHz) 3.529 # # 3.529 3.529
Dim of Aaprt X (cm) # # 4.8 4.8 #
Y (cm) # # 1.2 1.2 #
td (µsec) 1.064
prr (Hz) 3500
pr at max Ipi (MPa) 2.131
Other
deq at max Ipi (cm) 0.292
information
Ipi.α at max MI (W/cm2 ) 66.001
Focal FLx (cm) # # 6 #
Length FLy (cm) # # 6 #
Mode PW # # PW PW #
Operating Focus 6 # # 6 6 #
control (cm)
PRF 3.5 # # 3.5 3.5 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
95
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.892 0.894 # # # #
pr, α (MPa) 1.680
P (mW) 86.900 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 2.875 #
parameters
deq(zb) (cm)
fa wf (MHz) 3.544 3.544 # # # #
Dim of Aaprt X (cm) 2.4 # # # #
Y (cm) 1.2 # # # #
td (µsec) 1.058
prr (Hz) 3500
pr at max Ipi (MPa) 2.614
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 154.484
Focal FLx (cm) 6.0 # # #
Length FLy (cm) 6.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 6 6 # # # #
control
(cm)
PRF 3.5 3.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
96
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.925 0.894 cm
# # # #
pr, α (MPa) 1.740
P (mW) 86.900 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 2.875 #
deq(zb) (cm) #
fa wf (MHz) 3.541 3.541 # # # #
Dim of Aaprt X (cm) 2.4 # # # #
Y (cm) 1.2 # # # #
td (µsec) 1.062
prr (Hz) 4000
pr at max Ipi (MPa) 2.640
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 143.972
Focal FLx (cm) 6 # # #
Length FLy (cm) 6 # # #
Mode DPI B-DPI # # # #
Operating
Focus 6 6 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
97
Acoustic Output Tables

Transducer Model: VC6-2 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.102 0.312 # 0.556 1.659 #
pr, α (MPa) 1.621
P (mW) 88.100 # 66.700 #
min of [Pα(zs),Ita.α(zs)] (mW) 33.096
zs (cm) 2.875

Associated zbp (cm) 2.868


acoustic zb (cm) 2.317
parameters z at max Ipi.α (cm) 2.875
deq(zb) (cm) 0.286
fa wf (MHz) 3.544 3.544 # 3.529 3.529 #
Dim of Aaprt X (cm) 2.4 # 4.8 4.8 #
Y (cm) 1.2 # 1.2 1.2 #
td (µsec) 1.058
prr (Hz) 4000
pr at max Ipi (MPa) 2.547
Other
deq at max Ipi (cm) # 0.280
information 2
Ipi.α at max MI (W/cm ) 143.854
Focal FLx (cm) 6 # 6 #
Length FLy (cm) 6 # 6 #
Mode B B/COLOR # PW PW #
Operating 6 6 # 6 6
Focus (cm) #
control
PRF (kHz) 4.0 4.0 # 4.0 4.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
98
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.969 0.415 # # # #
pr, α (MPa) 2.394
P (mW) 28.700 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.534 #
deq(zb) (cm) #
fa wf (MHz) 6.360 6.360 # # # #
Dim of Aaprt X (cm) 1.36 # # # #
Y (cm) 0.60 # # # #
td (µsec) 0.114
prr (Hz) 5328
pr at max Ipi (MPa) 3.373
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 372.542
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode B # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
99
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.949 0.231 # # # #
pr, α (MPa) 2.334
P (mW) 17.000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.534 #
deq(zb) (cm) #
fa wf (MHz) 6.305 6.305 # # # #
Dim of Aaprt X (cm) 1.36 # # # #
Y (cm) 0.60 # # # #
td (µsec) 0.124
prr (Hz) 5328
pr at max Ipi (MPa) 3.288
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 277.908
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
100
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.890 0.415 # 1.217 0.200 #
pr, α (MPa) 1.732
P (mW) 28.700 # 38.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 5 14.630
zs (cm) 1.561
Associated zbp (cm) 1.527
acoustic zb (cm) 1.561
parameters z at max Ipi.α (cm) 1.561
deq(zb) (cm) 0.286
fa wf (MHz) 6.360 6.360 # 6.162 6.162 #
Dim of Aaprt X (cm) 2.700 # 1.360 1.360 #
Y (cm) 0.600 # 0.600 0.600 #
td (µsec) 0.114
prr (Hz) 5328
pr at max Ipi (MPa) 3.373
Other
deq at max Ipi (cm) 0.286
information 2
Ipi.α at max MI (W/cm ) 191.907
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
101
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm 2 scan
Maximum index value 1.687 # # 0.517 2.049 #
pr, α (MPa) 4.081
P (mW) # # 17.800 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.372 0
zs (cm) 1.614
zbp (cm) 1.527
Associated
zb (cm) 1.614
acoustic
z at max Ipi.α (cm) 1.534
parameters
deq(zb) (cm) 0.022
fa wf (MHz) 6.100 # # 6.100 6.100
Dim of Aaprt X (cm) # # 2.7 2.7 #
Y (cm) # # 0.6 0.6 #
td (µsec) 0.624
prr (Hz) 5000
pr at max Ipi (MPa) 5.734
Other
deq at max Ipi (cm) 0.022
information 2
Ipi.α at max MI (W/cm ) 90.888
Focal FLx (cm) # # 2.0 #
Length FLy (cm # # 2.0 #
Mode ) PW # # PW PW #
Operating Focus 2 # # 2 2 #
control
(cm)
PRF 5.0 # # 5.0 5.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

102
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.105 0.326 # # # #
pr, α (MPa) 2.677
P (mW) 31.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.534 #
parameters
deq(zb) (cm)
fa wf (MHz) 6.110 6.110 # # # #
Dim of Aaprt X (cm) 1.36 # # # #
Y (cm) 0.6 # # # #
td (µsec) 0.624
prr (Hz) 5000
pr at max Ipi (MPa) 3.710
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 287.985
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2 2 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
103
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Operating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 1.009 0.326 cm
# # # #
pr, α (MPa) 2.445
P (mW) 31.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.641 #
deq(zb) (cm) #
fa wf (MHz) 6.114 6.114 # # # #
Dim of Aaprt X (cm) 1.360 # # # #
Y (cm) 0.600 # # # #
td (µsec) 0.625
prr (Hz) 4000
pr at max Ipi (MPa) 3.457
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 229.206
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
104
Acoustic Output Tables

Transducer Model: 10I2 Linear Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.953 0.413 # 0.519 2.052 #
pr, α (MPa) 2.672
P (mW) 30.413 # 18.100 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.271
zs (cm) 1.614

Associated zbp (cm) 1.527


acoustic zb (cm) 1.614
parameters z at max Ipi.α (cm) 1.527
deq(zb) (cm) 0.024
fa wf (MHz) 6.053 6.053 # 6.093 6.093 #
Dim of Aaprt X (cm) 1.36 # 2.7 2.7 #
Y (cm) 0.6 # 0.6 0.6 #
td (µsec) 0.622
prr (Hz) 3500
pr at max Ipi (MPa) 3.713
Other
deq at max Ipi (cm) # 0.024
information 2
Ipi.α at max MI (W/cm ) 282.652
Focal FLx (cm) 2 # 2.5 #
Length FLy (cm) 2 # 2.5 #
Mode B B/COLOR # PW PW #
Operating 2 2 # 2.5 2.5
Focus (cm) #
control
PRF (kHz) 3.5 3.5 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
105
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.866 0.240 # # # #
pr, α (MPa) 2.398
P (mW) 17.400 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.410 #
deq(zb) (cm) #
fa wf (MHz) 8.001 8.001 # # # #
Dim of Aaprt X (cm) 1.86 # # # #
Y (cm) 0.33 # # # #
td (µsec) 0.113
prr (Hz) 7040
pr at max Ipi (MPa) 3.541
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 171.968
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
106
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.866 0.133 # # # #
pr, α (MPa) 2.334
P (mW) 8.300 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.410 #
deq(zb) (cm) #
fa wf (MHz) 8.001 8.001 # # # #
Dim of Aaprt X (cm) 1.86 # # # #
Y (cm) 0.33 # # # #
td (µsec) 0.113
prr (Hz) 7040
pr at max Ipi (MPa) 3.541
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 172.159
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 2 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
107
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.865 0.240 # 0.426 0.611 #
pr, α (MPa) 2.398
P (mW) 17.400 # 11.300 #
min of [Pα(zs),Ita.α(zs)] (mW) 5 5.198
zs (cm) 1.410
Associated zbp (cm) 1.324
acoustic zb (cm) 1.410
parameters z at max Ipi.α (cm) 1.410
deq(zb) (cm) 1.398
fa wf (MHz) 8.001 8.001 # 8.001 8.001 #
Dim of Aaprt X (cm) 1.860 # 3.680 3.680 #
Y (cm) 0.330 # 0.330 0.330 #
td (µsec) 0.113
prr (Hz) 150
pr at max Ipi (MPa) 3.541
Other
deq at max Ipi (cm) 1.398
information 2
Ipi.α at max MI (W/cm ) 160.496
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode B/M B # M M #
Operating
Focus (cm) 2 2 # 2 2 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
108
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.518 # # 0.244 0.798 #
pr, α (MPa) 1.162
P (mW) # # 12.500 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.406 0
zs (cm) 1.400
zbp (cm) 1.324
Associated
zb (cm) 1.400
acoustic
z at max Ipi.α (cm) 1.400
parameters
deq(zb) (cm) 0.241
fa wf (MHz) 4.103 # # 4.103 4.103
Dim of Aaprt X (cm) # # 3.68 3.68 #
Y (cm) # # 0.33 0.33 #
td (µsec) 0.624
prr (Hz) 4000
pr at max Ipi (MPa) 1.062
Other
deq at max Ipi (cm) 0.241
information
Ipi.α at max MI (W/cm2 ) 27.164
Focal FLx (cm) # # 2.0 #
Length FLy (cm # # 2.0 #
Mode ) PW # # PW PW #
Operating Focus 2 # # 2 2 #
control
(cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
109
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.439 0.169 # # # #
pr, α (MPa) 2.398
P (mW) 18.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.400 #
parameters
deq(zb) (cm)
fa wf (MHz) 8.001 8.001 # # # #
Dim of Aaprt X (cm) 1.86 # # # #
Y (cm) 0.33 # # # #
td (µsec) 0.113
prr (Hz) 3000
pr at max Ipi (MPa) 3.541
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 10.303
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 2 2 # # # #
control
(cm)
PRF 3.0 3.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
110
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.401 0.169 cm
# # # #
pr, α (MPa) 2.246
P (mW) 18.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.400 #
deq(zb) (cm) #
fa wf (MHz) 6.353 6.353 # # # #
Dim of Aaprt X (cm) 1.86 # # # #
Y (cm) 0.33 # # # #
td (µsec) 0.362
prr (Hz) 5000
pr at max Ipi (MPa) 1.484
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 168.952
Focal FLx (cm) 2.0 # # #
Length FLy (cm) 2.0 # # #
Mode DPI B-DPI # # # #
Operating
Focus 2 2 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
111
Acoustic Output Tables

Transducer Model: 10L1 Linear Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.872 0.243 # 0.242 0.796 #
pr, α (MPa) 2.392
P (mW) 19.230 # 13.124 #
min of [Pα(zs),Ita.α(zs)] (mW) 8.403
zs (cm) 1.400

Associated zbp (cm) 1.324


acoustic zb (cm) 1.400
parameters z at max Ipi.α (cm) 1.410
deq(zb) (cm) 0.243
fa wf (MHz) 7.983 7.983 # 8.001 8.001 #
Dim of Aaprt X (cm) 1.86 # 3.68 3.68 #
Y (cm) 0.33 # 0.33 0.33 #
td (µsec) 0.114
prr (Hz) 4000
pr at max Ipi (MPa) 3.544
Other
deq at max Ipi (cm) # 0.243
information 2
Ipi.α at max MI (W/cm ) 129.573
Focal FLx (cm) 2.0 # 2.0 #
Length FLy (cm) 2.0 # 2.0 #
Mode B B/COLOR # PW PW #
Operating 2 2 # 2.0 2.0
Focus (cm) #
control
PRF (kHz) 4.0 4.0 # 5.0 5.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
112
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.942 0.613 # # # #
pr, α (MPa) 1.646
P (mW) 76.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.249 #
deq(zb) (cm) #
fa wf (MHz) 2.329 2.329 # # # #
Dim of Aaprt X (cm) 2.79 # # # #
Y (cm) 1.32 # # # #
td (µsec) 0.446
prr (Hz) 9024
pr at max Ipi (MPa) 1.855
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 99.489
Focal FLx (cm) 5.0 # # #
Length FLy (cm) 5.0 # # #
Mode B # # # # #
Operating
Focus (cm) 5 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
113
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.825 0.548 # # # #
pr, α (MPa) 1.263
P (mW) 69.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.249 #
deq(zb) (cm) #
fa wf (MHz) 2.393 2.393 # # # #
Dim of Aaprt X (cm) 2.79 # # # #
Y (cm) 1.32 # # # #
td (µsec) 0.416
prr (Hz) 9024
pr at max Ipi (MPa) 1.653
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 107.218
Focal FLx (cm) 5.0 # # #
Length FLy (cm) 5.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 5 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
114
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 0.942 0.613 # 0.523 2.886 #
pr, α (MPa) 1.425
P (mW) 76.600 # 79.500 #
min of [Pα(zs),Ita.α(zs)] (mW) 47.131
zs (cm) 3.276
Associated zbp (cm) 3.243
acoustic zb (cm) 3.276
parameters z at max Ipi.α (cm) 3.249
deq(zb) (cm) 0.252
fa wf (MHz) 2.329 2.329 # 2.318 2.318 #
Dim of Aaprt X (cm) 2.79 # 4.14 4.14 #
Y (cm) 1.32 # 1.32 1.32 #
td (µsec) 0.446
prr (Hz) 150
pr at max Ipi (MPa) 1.855
Other
deq at max Ipi (cm) 0.252
information 2
Ipi.α at max MI (W/cm ) 99.489
Focal FLx 5.0 # # #
Length (cm)
FLy 5.0 # # #
Mode (cm) B/M B # M M #
Operating
Focus (cm) 5 5 # 5 5 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
115
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: PW mode

TIS TIB

Index label MI Non-scan Non- TIC


Scan 2 2 scan
Aaprt≤1cm Aaprt>1cm

Maximum index value 0.756 # # 0.908 4.102 #


pr, α (MPa) 1.172
P (mW) # # 134.900 #
min of [Pα(zs),Ita.α(zs)] (mW) 77.874
zs (cm) 4.306
zbp (cm) 3.243
Associated
zb (cm) 4.306
acoustic
z at max Ipi.α (cm) 3.249
parameters
deq(zb) (cm) 0.114
fa wf (MHz) 2.448 # # 2.448 2.448
Dim of Aaprt X (cm) # # 4.14 4.14 #
Y (cm) # # 1.32 1.32 #
td (µsec) 1.531
prr (Hz) 4000
pr at max Ipi (MPa) 1.687
Other
deq at max Ipi (cm) 0.114
information
Ipi.α at max MI (W/cm2 ) 79.359
Focal FLx (cm) # # 5.0 #
Length FLy (cm) # # 5.0 #
Mode PW # # PW PW #
Operating Focus 5 # # 5 5 #
control (cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
116
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.668 1.001 # # # #
pr, α (MPa) 1.425
P (mW) 76.600 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 3.249 #
parameters
deq(zb) (cm)
fa wf (MHz) 2.447 2.447 # # # #
Dim of Aaprt X (cm) 2.79 # # # #
Y (cm) 1.32 # # # #
td (µsec) 1.526
prr (Hz) 5500
pr at max Ipi (MPa) 1.855
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 62.933
Focal FLx (cm) 5.0 # # #
Length FLy (cm) 5.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 5 5 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
117
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 0.864 1.002 cm
# # # #
pr, α (MPa) 1.340
P (mW) 128.000 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 3.249 #
deq(zb) (cm) #
fa wf (MHz) 2.450 2.450 # # # #
Dim of Aaprt X (cm) 2.79 # # # #
Y (cm) 1.32 # # # #
td (µsec) 1.527
prr (Hz) 4000
pr at max Ipi (MPa) 1.855
Other
d at max Ipi (cm) #
information eq
Ipi.α at max MI (W/cm2 ) 73.731
Focal FLx (cm) 5.0 # # #
Length FLy (cm) 5.0 # # #
Mode DPI B-DPI # # # #
Operating
Focus 5 5 # # # #
control
(cm)
PRF 4.0 4.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
118
Acoustic Output Tables

Transducer Model: C311 Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 0.965 0.612 # 0.909 4.106 #
pr, α (MPa) 1.424
P (mW) 75.600 # 133.261 #
min of [Pα(zs),Ita.α(zs)] (mW) 76.694
zs (cm) 4.306

Associated zbp (cm) 3.243


acoustic zb (cm) 4.306
parameters z at max Ipi.α (cm) 3.249
deq(zb) (cm) 0.116
fa wf (MHz) 2.451 2.451 # 2.451 2.451 #
Dim of Aaprt X (cm) 2.79 # 4.14 4.14 #
Y (cm) 1.32 # 1.32 1.32 #
td (µsec) 1.523
prr (Hz) 3500
pr at max Ipi (MPa) 1.857
Other
deq at max Ipi (cm) # 0.116
information 2
Ipi.α at max MI (W/cm ) 80.265
Focal FLx (cm) 5.0 # 5.0 #
Length FLy (cm) 5.0 # 5.0 #
Mode B B/COLOR # PW PW #
Operating 5 5 # 5 5
Focus (cm) #
control
PRF (kHz) 3.5 3.5 # 4.0 4.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
119
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: B mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.011 0.473 # # # #
pr, α (MPa) 2.119
P (mW) 23.100 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 4.399 4.399 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.391
prr (Hz) 7040
pr at max Ipi (MPa) 2.182
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 193.617
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 3.0 # # #
Mode B # # # # #
Operating
Focus (cm) 3 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
120
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: THI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.994 0.355 # # # #
pr, α (MPa) 2.100
P (mW) 17.6 # # #
min of [Pα(zs),Ita.α(zs)] (mW) # #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.500 #
deq(zb) (cm) #
fa wf (MHz) 4.516 4.516 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.377
prr (Hz) 7040
pr at max Ipi (MPa) 2.867
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 157.593
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 3.0 # # #
Mode THI # # # # #
Operating
Focus (cm) 3 # # # # #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 # # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
121
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: B/M mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 Scan
Maximum index value 1.027 0.473 # 0.336 0.262 #
pr, α (MPa) 2.151
P (mW) 23.100 # 16.000 #
min of [Pα(zs),Ita.α(zs)] (mW) 10.128
zs (cm) 1.687
Associated zbp (cm) 1.352
acoustic zb (cm) . 1.687
parameters z at max Ipi.α (cm) 1.500
deq(zb) (cm) 0.252
fa wf (MHz) 4.399 4.399 # 4.413 4.413 #
Dim of Aaprt X (cm) 1.28 # 2.59 2.59 #
Y (cm) 0.5 # 0.5 0.5 #
td (µsec) 0.393
prr (Hz) 150
pr at max Ipi (MPa) 2.782
Other
deq at max Ipi (cm) 0.253
information 2
Ipi.α at max MI (W/cm ) 190.611
Focal FLx (cm) 3.0 # 3.0 #
Length FLy (cm) 3.0 # 3.0 #
Mode B/M B # M M #
Operating
Focus (cm) 3 3 # 3 3 #
control
PRF (kHz) # # # # # #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
122
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 0.849 # # 0.494 2.005 #
pr, α (MPa) 2.012
P (mW) # # 17.900 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.818
zs (cm) 1.768
zbp (cm) 1.352
Associated
zb (cm) 1.768
acoustic
z at max Ipi.α (cm) 1.500
parameters
deq(zb) (cm) 0.127
fa wf (MHz) 5.796 # # 5.796 5.796 #
Dim of Aaprt X (cm) # # 2.59 2.59 #
Y (cm) # # 0.5 0.5 #
td (µsec) 0.653
prr (Hz) 4000
pr at max Ipi (MPa) 2.867
Other
deq at max Ipi (cm) 0.048
information
Ipi.α at max MI (W/cm2 ) 142.082
Focal FLx (cm) # # 3.0 #
Length FLy (cm # # 3.0 #
Mode ) PW # # PW PW #
Operating Focus 3 # # 3 3 #
control
(cm)
PRF 4.0 # # 4.0 4.0 #
conditions
(kHz)
Power 100 # # 100 100 #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
123
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: B/CFM mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm2 Aaprt>1cm2 scan
Maximum index value 1.107 0.851 # # # #
pr, α (MPa) 989
2.317
P (mW) 42.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) #
acoustic
z at max Ipi.α (cm) 1.500 #
parameters
deq(zb) (cm)
fa wf (MHz) 4.390 4.390 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.858
prr (Hz) 5000
pr at max Ipi (MPa) 3.035
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 196.053
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 3.0 # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 3 3 # # # #
control
(cm)
PRF 5.0 5.0 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
124
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: DPI mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1 Aaprt>1cm2 scan
2
Maximum index value 1.100 0.851 cm
# # # #
pr, α (MPa) 2.303
P (mW) 42.800 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) #
parameters z at max Ipi.α (cm) 1.821 #
deq(zb) (cm) #
fa wf (MHz) 4.391 4.391 # # # #
Dim of Aaprt X (cm) 1.28 # # # #
Y (cm) 0.5 # # # #
td (µsec) 0.860
prr (Hz) 5500
pr at max Ipi (MPa) 3.035
Other
deq at max Ipi (cm) #
information
Ipi.α at max MI (W/cm2 ) 198.941
Focal FLx (cm) 3.0 # # #
Length FLy (cm) 3.0 # # #
Mode DPI B-DPI # # # #
Operating
Focus 3 3 # # # #
control
(cm)
PRF 5.5 5.5 # # # #
conditions
(kHz)
Power 100 100 # # # #
Note 1 Data(%)should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
125
Acoustic Output Tables

Transducer Model: 6V1A Micro-Convex Array Ope rating Mode: B/COLOR/PW mode

TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm Aaprt>1cm2 Scan
2
Maximum index value 1.034 0.482 # 0.503 2.015 #
pr, α (MPa) 2.318
P (mW) 43.400 # 18.204 #
min of [Pα(zs),Ita.α(zs)] (mW) 9.824
zs (cm) 1.768

Associated zbp (cm) 1.352


acoustic zb (cm) 1.768
parameters z at max Ipi.α (cm) 1.500
deq(zb) (cm) 0.126
fa wf (MHz) 4.413 4.413 # 5.805 5.805 #
Dim of Aaprt X (cm) 1.28 # 2.59 2.59 #
Y (cm) 0.50 # 0.50 0.50 #
td (µsec) 0.859
prr (Hz) 5000
pr at max Ipi (MPa) 3.033
Other
deq at max Ipi (cm) # 0.127
information 2
Ipi.α at max MI (W/cm ) 194.053
Focal FLx (cm) 3.0 # 2.5 #
Length FLy (cm) 3.0 # 3.0 #
Mode B B/COLOR # PW PW #
Operating # 2.5 2.5
Focus (cm) 3 3 #
control
PRF (kHz) 5.0 5.0 # 4.0 4.0 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
126
Acoustic Output Tables

Transducer Model: 7-4MHz C542 Curved Array Operating Mode: THI, B/M
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm 2
Aaprt>1cm 2 scan
Maximum index value 0.95 0.91 # 0.35 0.52 #
pr, α (MPa) 1.71
P (mW) 55.4 # 36.8 #
min of [Pα(zs),Ita.α(zs)] (mW) 18.2
zs (cm) 3.6

Associated zbp (cm) 3.6


acoustic zb (cm) 3.6 3.6
parameters z at max Ipi.α (cm) 3.6 3.6
deq(zb) (cm) 0.41
fa wf (MHz) 3.4 3.4 # 3.4 3.4 #
Dim of Aaprt X (cm) 2.6 # 2.6 2.6 #
Y (cm) 1.3 # 1.3 1.3 #
td (µsec) 0.464
prr (Hz) 8015
pr at max Ipi (MPa) 2.53
Other
deq at max Ipi (cm) 0.41
information 2
Ipi.α at max MI (W/cm ) 306.3
Focal FLx (cm) 4 # 4 #
Length FLy (cm) # # # #
Mode THI, B/M THI, B # M M #
Operating
Focus (cm) 3 5 # 5 5 #
control
PRF (kHz) 8 6.4 # 1.6 1.6 #
conditions
Power (%) 100 100 # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

127
Acoustic Output Tables

Transducer Model: 7-4MHz C542 Curved Array Ope rating Mode: PW


TIS TIB
Index label MI Non-scan Non- TIC
Scan 2
Aaprt≤1cm Aaprt>1cm 2 scan
Maximum index value 0.95 # # 1.15 1.53 #
pr, α (MPa) 1.62
P (mW) # # 118 #
min of [Pα(zs),Ita.α(zs)] (mW) 60.0
zs (cm) 3.8
zbp (cm) 3.8
Associated
zb (cm) 3.6 3.6
acoustic
z at max Ipi.α (cm) 3.6 3.6
parameters
deq(zb) (cm) 0.38
fa wf (MHz) 3.2 # # 3.2 3.2 #
Dim of Aaprt X (cm) # # 2.3 2.3 #
Y (cm) # # 1.5 1.5 #
td (µsec) 1.43
prr (Hz) 1008
pr at max Ipi (MPa) 2031
Other
deq at max Ipi (cm) 0.34
information 2
Ipi.α at max MI (W/cm ) 544.2
Focal FLx (cm) # # 4 #
Length FLy (cm # # # #
Mode ) PW # # PW PW #
Operating
Focus (cm) 5 # # 5 5 #
control
PRF (kHz) 1 # # 4 4 #
conditions
Power (%) 100 # # 100 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

128
Acoustic Output Tables

Transducer Model: 7-4MHz C542 Curved Array Ope rating Mode: CFM-B
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm 2
Aaprt>1cm 2 scan
Maximum index value 0.96 2.23 # # # #
pr, α (MPa) 1.76
P (mW) 144 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) 3.5 #
acoustic
z at max Ipi.α (cm) 3.5 #
parameters
deq(zb) (cm) #
fa wf (MHz) 3.2 3.2 # # # #
Dim of Aaprt X (cm) 3.4 # # # #
Y (cm) 1.3 # # # #
td (µsec) 1.5
prr (Hz) 6472
pr at max Ipi (MPa) 2.46
Other
d at max Ipi (cm) #
information eq
2
Ipi.α at max MI (W/cm ) 582.8
Focal FLx (cm) 4 # # #
Length FLy (cm) # # # #
Mode CFM B-CFM # # # #
Operating
Focus (cm) 5 5 # # # #
control
(cm)
PRF (kHz) 2 6.48 # # # #
conditions
Power (%) 100 100 # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

129
Acoustic Output Tables

Transducer Model: 7-4MHz C542 Curved Array Ope rating Mode: CFM-THI, B
TIS TIB
Index label MI Non-scan Non- TIC
Scan
Aaprt≤1cm 2
Aaprt>1cm 2 scan
Maximum index value 1.01 2.37 # # # #
pr, α (MPa) 1.73
P (mW) 155 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) 3.4 #
parameters z at max Ipi.α (cm) 3.4 #
deq(zb) (cm) #
fa wf (MHz) 3.3 3.3 # # # #
Dim of Aaprt X (cm) 3.5 # # # #
Y (cm) 1.8 # # # #
td (µsec) 1.5
prr (Hz) 6473
pr at max Ipi (MPa) 1.731
Other
deq at max Ipi (cm) #
information
2
Ipi.α at max MI (W/cm ) 615.6
Focal FLx (cm) 4 # # #
Length FLy (cm) # # # #
Mode CFM THI,B- # # # #
Operating CFM
Focus (cm) 5 5 # # # #
control
PRF (kHz) 2 6.48 # # # #
conditions
Power (%) 100 100 # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

130
Acoustic Output Tables

Transducer Model: 6-2MHz C322 Curved Array Ope rating Mode: B/M
TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm scan
Maximum index value 1.350 1.520 # 0.435 1.655 #
pr, α (MPa) 2.350
P (mW) 98.6 # 52.6 #
min of [Pα(zs),Ita.α(zs)] (mW) 28.5
zs (cm) 4.35

Associated zbp (cm) 2.85


acoustic zb (cm) # 4.35
parameters
z at max Ipi.α (cm) 4.35 #
deq(zb) (cm) 0.012
fa wf (MHz) 3.125 3.125 3.125 # 3.125 #
Dim of Aaprt X (cm) 2.200 2.200 # 4.850 #
Y (cm) 1.105 1.105 # 1.235 #
td (µsec) 0.350
prr (Hz) 200
pr at max Ipi (MPa) 3.755
Other
deq at max Ipi (cm) 0.013
information
2
Ipi.α at max MI (W/cm ) 300.65
Focal FLx (cm) 5 5 # #
Length FLy (cm) # # # #
Mode B/M B M # M #
Operating (cm) 5 3 5 # 5 #
Focus
control
conditions PRF (kHz) 0.2 9.1 0.2 # 0.2 #
Power (%) 100 100 100 # 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
131
Acoustic Output Tables

Transducer Model: 6-2MHz C322 Curved Array Ope rating Mode: THI, B/M
TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm scan
Maximum index value 1.025 2.856 # 0.356 1.563 #
pr, α (MPa) 1.562
P (mW) 215.6 # 52.6 #
min of [Pα(zs),Ita.α(zs)] (mW) 29.1
zs (cm) 4.35
zbp (cm) 2.85
Associated
zb (cm) # 4.35
acoustic
parameters z at max Ipi.α (cm) 4.35 #
deq(zb) (cm) 0.012
fa wf (MHz) 2.510 2.510 2.510 # 3.123 #
Dim of Aaprt X (cm) 2.200 2.200 # 4.850 #
Y (cm) 1.105 1.105 # 1.235 #
td (µsec) 0.750
prr (Hz) 9108
pr at max Ipi (MPa) 2.45
Other
d at max Ipi (cm) 0.013
information eq
2
Ipi.α at max MI (W/cm ) 168.5
Focal FLx (cm) 5 5 # #
Length FLy (cm) # # # #
Mode THI, THI, B THI, M # THI, M #
Operating (cm) B/M
5 5 5 # 5 #
Focus
control
conditions PRF (kHz) 9.1 9.1 9.1 # 0.2 #
Power (%) 100 100 100 # 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

132
Acoustic Output Tables

Transducer Model: 6-2MHz C322 Curved Array Ope rating Mode: PW


TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm scan
Maximum index value 0.425 # # 0.315 2.200 #
pr, α (MPa) 0.560
P (mW) # # 42.50 #
min of [Pα(zs),Ita.α(zs)] (mW) 25.6
zs (cm) 2.850
zbp (cm) 2.900
Associated
zb (cm) # 4.580
acoustic
parameters z at max Ipi.α (cm) 2.580 #
deq(zb) (cm) 0.16
fa wf (MHz) 2.350 # 2.350 # 2.350 #
Dim of Aaprt X (cm) # 4.850 # 4.850 #
Y (cm) # 1.235 # 1.235 #
td (µsec) 2.082
prr (Hz) 1000
pr at max Ipi (MPa) 0.850
Other
d at max Ipi (cm) 0.209
information eq
Ipi.α at max MI 2
(W/cm ) 19.8 0.
1.
Focal FLx (cm) # 5 # #
Length FLy (cm) # # # #
Mode PW # PW # PW #
Operating Sample Volume Depth (cm) 5 # 5 # 5 #
control
conditions PRF (KH 1 # 4 # 4 #
Power z) 100
(%) # 100 # 100 #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

133
Acoustic Output Tables

Transducer Model: 6-2MHz C322 Curved Array Ope rating Mode: CFM-B

TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm scan
Maximum index value 0.720 1.795 # # # #
pr, α (MPa) 1.385
P (mW) 136.5 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #
zbp (cm) #
Associated
zb (cm) # #
acoustic
parameters z at max Ipi.α (cm) 4.258 #
deq(zb) (cm) #
fa wf (MHz) 2.560 2.560 # # # #
Dim of Aaprt X (cm) 2.200 # # # #
Y (cm) 1.105 # # # #
td (µsec) 1.580
prr (Hz) 6000
pr at max Ipi (MPa) 1.655
Other
deq at max Ipi (cm) #
information
2
Ipi.α at max MI (W/cm ) 65.925
Focal FLx (cm) 5 # # #
Length FLy (cm) # # # #
Mode B B-Color # # # #
Operating Focus (cm) 5 5 # # # #
control
conditions PRF (KHz) 9.1 6.0 # # # #
Power (%) 100 100 # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.
134
Acoustic Output Tables

Transducer Model: 6-2MHz C322 Curved Array Operating Mode: CFM-THI, B


TIS TIB
Index label MI Non-scan Non- TIC
Scan 2 2
Aaprt≤1cm Aaprt>1cm scan
Maximum index value 0.74 1.693 # # # #
pr, α (MPa) 1.420
P (mW) 131.4 # # #
min of [Pα(zs),Ita.α(zs)] (mW) #
zs (cm) #

Associated zbp (cm) #


acoustic zb (cm) # #
parameters
z at max Ipi.α (cm) 4.261 #
deq(zb) (cm) #
fa wf (MHz) 2.541 2.541 # # # #
Dim of Aaprt X (cm) 2.200 # # # #
Y (cm) 1.105 # # # #
td (µsec) 1.564
prr (Hz) 5000
pr at max Ipi (MPa) 1.579
Other
d at max Ipi (cm) #
information eq
2
Ipi.α at max MI (W/cm ) 69.125
Focal FLx (cm) 5 # # #
Length FLy (cm) # # # #
Mode THI,B THI,B-Colo # # # #
Operating Focus (cm) 5 5r # # # #
control
conditions PRF (KHz) 9.1 5.0 # # # #
Power (%) 100 100 # # # #
Note 1 Data should only be entered in one of the columns related to TIS.
Note 2 Information need not be provide regarding TIC for any TRANSDUCER ASSEMBLY not
intended for transcranial or neonatal cephalic uses.
Note 3 If the requirements of 201.12.4.2a) are met, it is not required to enter any data in the columns
related to TIS, TIB or TIC.
Note 4 If the requirements of 201.12.4.2b) are met, it is not required to enter any data in the columns
related to MI.
Note 5 Focal Length is a NOMINAL value.

135