How to Input the Configure Parameter on Hematology Analyzer

Note: Each Hematology Analyzer has different Configure Parameter, This instruction is used to guide on
how to input the Parameter.

Figure 1 RT-7200 Parameter label Example

Explanation on the Label meaning:

1. On the instrument’s Right side cover, there is the parameter label, like the above photo.
2. Turn on the instrument. When the instrument is at the beginning of the self-Initializing stage,
please click “ALT+F4” to jump self check. And then the instrument will go to the test interface.
3. Logo on with Engineer mode, The password is 1001
And then click “Logon” and select Engneer mode and then input the password 1001.Don’t forgot
to click ok to save.

Note: If you don’t log on with Engineer mode, you can see the below photo in Service button.
After you logo on with Engineer mode, you can see 2 more buttons.

4. Enter Service interface and input the parameters.


4-1 Needle Positions:
Click “Debug” button, and then input the probe parameter NP1~NP5.

Please input the Needle position value and then click “Send” to save. The sequence is Needle
Position 1, 2, 4, 5, 3.
Note: For RT-7200, it doesn’t have NP 5, so the sequence is Needle Position 1, 2, 4, 3.

4-2 Pulse Parameter and Pressure Parameter.

Enter the Engineering Adjusting button, and then you can see the following interface:
This is The value of
R.O.P

This is The value of


W.O.P

The 2 Pressure Value


are the same with T.S.P
value.

4-3 Calibration Parameter

Enter Calibration Interface and then input the Calibration parameters.


Please input the original calibration parameter and then check the instrument Performance. If the
instrument is working well before replace the software, you can directly calibrate the instrument with the
calibrator.

Note: The parameter is not mentioned on the parameter label, please don’t change the parameter.