Note: Each Hematology Analyzer has different Configure Parameter, This instruction is used to guide on
how to input the Parameter.
1. On the instrument’s Right side cover, there is the parameter label, like the above photo.
2. Turn on the instrument. When the instrument is at the beginning of the self-Initializing stage,
please click “ALT+F4” to jump self check. And then the instrument will go to the test interface.
3. Logo on with Engineer mode, The password is 1001
And then click “Logon” and select Engneer mode and then input the password 1001.Don’t forgot
to click ok to save.
Note: If you don’t log on with Engineer mode, you can see the below photo in Service button.
After you logo on with Engineer mode, you can see 2 more buttons.
Please input the Needle position value and then click “Send” to save. The sequence is Needle
Position 1, 2, 4, 5, 3.
Note: For RT-7200, it doesn’t have NP 5, so the sequence is Needle Position 1, 2, 4, 3.
Enter the Engineering Adjusting button, and then you can see the following interface:
This is The value of
R.O.P
Note: The parameter is not mentioned on the parameter label, please don’t change the parameter.