Tridema engineering
User Manual
rel. 2.0
Index
1 Legend of Simbol
2 Chek List
3 Principle
4 Instrument Description
5 Technical Features
6 Operative function
7 Calibration menu
8 Parameters Menu
9 PT
10 APTT
11 Fibrinogen – FIB
12 Thrombine Time – TT
13 PT Calibration
14 Fibrinogen calibration – FIB
15 PT ISI
16 Fibrinogen Calibration Concentration
17 Parameters Editor
18 Parameters Description
19 Appendix A – Bluethoot Printer (optional)
20 Instrument Disposal
21 Troubleshooting
22 Customer Service
23 Contacts
1 Symbol Description
On the packaging and on the label on the instrument appear the symbols hown below. The
symbol have the following meaning:
Relevant news
Manufacturer
2 Check List
• Diamond Instrument
• Power Supply 12v
• CD-ROM User Manual
• CE Certificate
• Certificate of analysis
3 Principle
• 450 nm
• 510 nm
• 630 nm
4 Instrument Description
1 2
4 5
2 24 keys keyboard
4 Power supply in
5 Power switch
5.
5 Technical Features
• 10 MIPS microprocessor
• 300 bytes ram
• 64k flash eprom
• 1024 bytes eprom
6 Operative Function
In the meanwhile you can do whatever else with the Diamond, for example you can correct the
reagents parameters or change the PT calibration curve.
The arrov “Up” ▲ and “Down ▼on keyboard, use for move the indicator on display for set
your choice
PT
APTT
FIB
THROMBIN TIME
You will see the arrow on the right side of the display moving up and down.
When the arrows point the analysis you wich perform, press the enter key and it will start.
6 Funciones Operativas
Pressing the “Left” ◄ and ►”Right” arrows you can move between the other Diamond menu:
to reach the “parameters menu” from main menu you have to pass through the calibrations
menu and vice-versa.
7 Calibration Menu
PT
FIB
ISI-PT
CONCENTRATION
You will see the arrows on the right side of the display moving up and down.
When the arrows will point to the function you want to perform, press the enter key and it will
start.
8 Parameters Menu
PT
APTT
FIBRINOGEN
TT
You will see the arrows on the right side of the display moving up and down.
When the arrows will point to the function you want to perform, press the enter key and it will
start.
9 PT
F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm
9 PT
Now the instrument waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see
9 PT
10 APTT
F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm
The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.
10 APTT
Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see
10 APTT
Now if press exit key, the instrument come back at main menu, for new choice, while if press
“#”, perform a new test Aptt.El abertura de esta panorámica:
The following screen
11 Fibrinogeno – FIB
F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm
The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.
11 Fibrinogeno – FIB
Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see
11 Fibrinogeno – FIB
Now if press exit key, the instrument come back at main menu, for new choice, while if press
“#”, perform a new test Fib.
12 Tiempo de Trombina – TT
F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm
The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.A la presión del interruptor de inicio sobre el display se
12 Tiempo de Trombina – TT
Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see
12 Tiempo de Trombina – TT
13 Calibración PT
When you select the PT calibration phase you will get following screen
Pressing the number 1, 2, 3, or 4 you will be able to change any time value in order to create
a new calibration curve.
With keyboard will be possible to set a new time value, equal to 100% of activity.
For decimal number, use this key “.” for separate the numbers
With radial keys “#” it’s possible to delete the last character
Press “enter” for to confirm the new values, after setting all values press “exit” for to exit.
When you select the Fibrinogen calibration phase you will get following screen
Pressing the number 1, 2, 3, or 4 you will be able to change any time value in order to create
a good calibration curve.
x2 ?
Using the keyboard can input the new time value obtained for a CALIBRATOR X2 concentration
plasma, you can use the 1,2,3,4,5,6,7,8,9,0 keys plus the “.” for decimals
With radial keys “#” it’s possible to delete the last character
Press “enter” for to confirm the new values, after setting all values press “exit” for to exit.
WARNING The concentration values are referred to the calibrator in use, for
example if your calibrator is 320 mg/dl, the four values will be 640 mg/dl, 320
mg/dl, 160 mg/dl and 115 mg/dl, mean CONC x 2, x CONC 5, CONC x 33
15 PT ISI
When you select the PT ISI you will get following screen
Now pressing the 1 key the system will ask for the new isi value
VALUE?
Use the keys 1,2,3,4,5,6,7,8,9,0, “.”To input the value and “#” to delete the last input
When you select the FIBRINOGEN CALIBRATION CONCENTRATION you will get the
following screen
Now pressing the 1 key the system will ask for the new value
VALUE?
Use the keys 1,2,3,4,5,6,7,8,9,0, “.” To input the value and “#” to delete the last input
17 Editor Parameters
Moving with ▲ and ▼ arrows up and down you can choose the parameter you want to change:
• WAVELENGTH
• CUTLINE A
• CUTLINE B
• DEAD TIME
• NO CLOT SEC
• NORM MIN
• NORM MAX
• RATIO SECS
• INCUBATION
• DELAY
18 Parameter Description
WAVELENGTH – This is the wavelength preferred to perform this test, 1 for the blue(450nm),
2 for the red(630nm) and 3 for green(510nm).
CUTLINE A** – Please, not change this parameter, it will be changed only from technical
responsible.
CUTLINE B – Please, not change this parameter, it will be changed only from technical
responsible.
DEAD TIME – This value in seconds during that the clot detector doesn’t work, is a free zone
just to give the time reaction to be more stable optically.
NO CLOT SEC – This is the value in seconds for the NO CLOT message.
SENSPERC - Please, not change this parameter, it will be changed only from technical
responsible.
RATIO SECS – This is value in seconds that will be used in RATIO calculations.
* viene aggiornato in automatico per la PT ad ogni registrazione delle calibrazioni.
INCUBATION – This is value in seconds that will be used from the countdown function in
analysis.
19 Appendix A
Optional Bluetooth printer
• In the version Diamond with printer
results:
TT
WARNING: Before switch on the instrument and printer, switch off all Bluetooth device
20 Waste disposal
Intrument, Used cuvettes, pipette tips, reagent and plasma containers must be disposed in
compliance with biohazard and legal provisions.
The instrument does not fall under Directive 2002/96 / EC
(Directive on waste electrical and electronic di apparecchiature (WEEE).
21 Troubleshooting
Instrument does not switch on Check power line, power supply and
connecting cables.
With duplicate testing Check the quality of the pipet tips (or
reproducibility is not good pipettor) used.
22 Customer Service
For all questions relating to the use of the diamond instruments, repeatability of the results and
possible damage of the instrument, see TROUBLESHOOTING menu or contact your local
distributor.
Our staff will help you solve any problem relating to the instrument.