Tridema engineering

User Manual
rel. 2.0

User Manual - rel. 2.0


Index

Index

1 Legend of Simbol
2 Chek List
3 Principle
4 Instrument Description
5 Technical Features
6 Operative function
7 Calibration menu
8 Parameters Menu
9 PT
10 APTT
11 Fibrinogen – FIB
12 Thrombine Time – TT
13 PT Calibration
14 Fibrinogen calibration – FIB
15 PT ISI
16 Fibrinogen Calibration Concentration
17 Parameters Editor
18 Parameters Description
19 Appendix A – Bluethoot Printer (optional)
20 Instrument Disposal
21 Troubleshooting
22 Customer Service
23 Contacts

User Manual - rel. 2.0


Chapter 1 - Symbol Description

1 Symbol Description

On the packaging and on the label on the instrument appear the symbols hown below. The
symbol have the following meaning:

Consult the user manual

Caution (refer to attached documents) Observe the notes

Relevant news

Manufacturer

User Manual - rel. 2.0


Chapter 2 - Check List

2 Check List

• Diamond Instrument
• Power Supply 12v
• CD-ROM User Manual
• CE Certificate
• Certificate of analysis

User Manual - rel. 2.0


Chapter 3 - Principle

3 Principle

The Diamond is a semi-automatic coagulation analyzer, the reading principle is photometric,


it is based on a solid state source emitting at 3 wavelenghs:

• 450 nm
• 510 nm
• 630 nm

Each of thes wavelengths can be used each different test.


The Beam emitted by the solid state led is driven through a hole to a plastic cup used to hold
the plasma.
The light beam passed through the plasma-reagent solution is read by a solid state photodiode,
amplified and converted to a digita signal.
The microcomputer analyzes the data and after some mathematical calculation gives the result
to user.

User Manual - rel. 2.0


Chapter 4 - Instrument Description

4 Instrument Description

1 2

4 5

1 128x64 monochromatics display

2 24 keys keyboard

3 Sample holder and reading unit

4 Power supply in

5 Power switch
5.

User Manual - rel. 2.0


Chapter 5 - Technical Features

5 Technical Features

• Ready to operate aftre 5 minutes warm up


• 3 wavelength 450nm, 510nm e 630nm
• Solid state photodiode-amplifier adc converter
• Graphic display 128 x 64
• 24 Keys keyboard
• 16 position sample holder 37° C
• 4 position warmed reagents 37° C
• 2 position for reagents at room temperature

• 10 MIPS microprocessor
• 300 bytes ram
• 64k flash eprom
• 1024 bytes eprom

User Manual - rel. 2.0


Chapter 6 - Operative Function

6 Operative Function

Connect the power supply to the instrument and switch on.

When the power on, in the lower side of the display,


a number will indicate the temperature expressed in
degrees Celsius.

The temperature will stabilize around the 37° degrees.

Waiting 5 minutes before perform the tests

In the meanwhile you can do whatever else with the Diamond, for example you can correct the
reagents parameters or change the PT calibration curve.

The arrov “Up” ▲ and “Down ▼on keyboard, use for move the indicator on display for set
your choice

PT
APTT
FIB 
THROMBIN TIME

You will see the arrow on the right side of the display moving up and down.

When the arrows point the analysis you wich perform, press the enter key and it will start.

User Manual - rel. 2.0


Chapter 6 - Funciones Operativas

6 Funciones Operativas

Pressing the “Left” ◄ and ►”Right” arrows you can move between the other Diamond menu:

MAIN MENU’ ◄► CALIBRATION MENU’ ◄► PARAMETERS MENU’

to reach the “parameters menu” from main menu you have to pass through the calibrations
menu and vice-versa.

The “PARAMETERS” menu is protected by a password.


The password can be required to company specialists that sell the instrument.

User Manual - rel. 2.0


Chapter 7 - Calibration Menu

7 Calibration Menu

You can move “Up” ▲ and “Down” ▼arrows on the


keyboard.

PT
FIB
ISI-PT
CONCENTRATION

You will see the arrows on the right side of the display moving up and down.
When the arrows will point to the function you want to perform, press the enter key and it will
start.

User Manual - rel. 2.0


Chapter 8 - Parameters Menu

8 Parameters Menu

You can move “Up” ▲ and “Down” ▼arrows on the


keyboard.

PT
APTT
FIBRINOGEN
TT

You will see the arrows on the right side of the display moving up and down.

When the arrows will point to the function you want to perform, press the enter key and it will
start.

User Manual - rel. 2.0


Chapter 9 - PT

9 PT

When you see these on the display

The instrument is in stand-by mode


Put a reading cup with plasma in channel reading
Now it’s possible to perform the test
If you want change the parameters use F1, F2, F3 and F4 keys

F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm

The Countdown must be set in parameters menu, setting the incubation


parameters, for example, 120 seconds.

User Manual - rel. 2.0


Chapter 9 - PT

9 PT

When press start, on display you see

Now the instrument waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see

It will be possible the absorbance variation in real time.

When the instrument detect the clot, on display you see:

User Manual - rel. 2.0


Chapter 9 - PT

9 PT

Now if press exit key, the instrument come back at main


menu, for new choice, while if press “#”, perform a new test PT.

The following screen:

Mean that was not possible to detect the clot.

User Manual - rel. 2.0


Chapter 9 - APTT

10 APTT

When press start, on display you see

The instrument is in stand-by mode


Put a reading cup with plasma in channel reading
Now it’s possible to perform the test
If you want change the parameters use F1, F2, F3 and F4 keys

F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm

The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.

User Manual - rel. 2.0


Chapter 9 - APTT

10 APTT

When press start, on display you see

Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see

Apenas el instrumento habrá notado un coágulo, enseñará este resultado:

It will be possible the absorbance variation in real time.


When the instrument detect the clot, on display you seeanálisis.

User Manual - rel. 2.0


Chapter 9 - APTT

10 APTT

Now if press exit key, the instrument come back at main menu, for new choice, while if press
“#”, perform a new test Aptt.El abertura de esta panorámica:
The following screen

Mean that was not possible to detect the clot.

User Manual - rel. 2.0


Chapter 11 - Fibrinogeno - FIB

11 Fibrinogeno – FIB

When you see these on the display:

The instrument is in stand-by mode


Put a reading cup with plasma in channel reading
Now it’s possible to perform the test
If you want change the parameters use F1, F2, F3 and F4 keys

F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm

The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.

User Manual - rel. 2.0


Chapter 11 - Fibrinogeno - FIB

11 Fibrinogeno – FIB

When press start, on display you see:

Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see

It will be possible the absorbance variation in real time.


When the instrument detect the clot, on display you see:

User Manual - rel. 2.0


Chapter 11 - Fibrinogeno - FIB

11 Fibrinogeno – FIB

Now if press exit key, the instrument come back at main menu, for new choice, while if press
“#”, perform a new test Fib.

The following screen

Mean that was not possible to detect the clot.

User Manual - rel. 2.0


Chapter 12 - Tiempo de Trombina - TT

12 Tiempo de Trombina – TT

When you see this on the display:

The instrument is in stand-by mode


Put a reading cup with plasma in channel reading
Now it’s possible to perform the test
If you want change the parameters use F1, F2, F3 and F4 keys

F1 countdown start
F2 wavelength 450nm
F3 wavelength 630nm
F4 wavelength 510nm

The Countdown must be set in parameters menu, setting the incubation parameters,
for example, 180 seconds.A la presión del interruptor de inicio sobre el display se

User Manual - rel. 2.0


Chapter 12 - Tiempo de Trombina - TT

12 Tiempo de Trombina – TT

When you see this on the display

Now the instruments waits the operator that pipette the reagents in the reading cup. The
absorbance variation will do start automatically the reaction, and on display you see

It will be possible the absorbance variation in real time.

When the instrument detect the clot, on display you see

User Manual - rel. 2.0


Chapter 12 - Tiempo de Trombina - TT

12 Tiempo de Trombina – TT

While if press “#”, perform a new test TT.

The following screen

Mean that was not possible to detect the clot.

User Manual - rel. 2.0


Chapter 13 - Calibración PT

13 Calibración PT

When you select the PT calibration phase you will get following screen

Pressing the number 1, 2, 3, or 4 you will be able to change any time value in order to create
a new calibration curve.

For example pressing the button 1 the system will ask:


100%?

With keyboard will be possible to set a new time value, equal to 100% of activity.

For decimal number, use this key “.” for separate the numbers

With radial keys “#” it’s possible to delete the last character

Press “enter” for to confirm the new values, after setting all values press “exit” for to exit.

User Manual - rel. 2.0


Chapter 14 - Calibración Fibrinogeno – FIB

14 Calibración Fibrinogeno – FIB

When you select the Fibrinogen calibration phase you will get following screen

Pressing the number 1, 2, 3, or 4 you will be able to change any time value in order to create
a good calibration curve.

For example pressing the button 1 the system will ask:

x2 ?

Using the keyboard can input the new time value obtained for a CALIBRATOR X2 concentration
plasma, you can use the 1,2,3,4,5,6,7,8,9,0 keys plus the “.” for decimals

With radial keys “#” it’s possible to delete the last character

Press “enter” for to confirm the new values, after setting all values press “exit” for to exit.

WARNING The concentration values are referred to the calibrator in use, for
example if your calibrator is 320 mg/dl, the four values will be 640 mg/dl, 320
mg/dl, 160 mg/dl and 115 mg/dl, mean CONC x 2, x CONC 5, CONC x 33

User Manual - rel. 2.0


Chapter 15 - PT ISI

15 PT ISI

When you select the PT ISI you will get following screen

Now pressing the 1 key the system will ask for the new isi value

VALUE?

Use the keys 1,2,3,4,5,6,7,8,9,0, “.”To input the value and “#” to delete the last input

When done press the exit key to exit.

User Manual - rel. 2.0


Chapter 16 - Fibrinogen Calibration Concentration

16 Fibrinogen Calibration Concentration

When you select the FIBRINOGEN CALIBRATION CONCENTRATION you will get the
following screen

Now pressing the 1 key the system will ask for the new value

VALUE?

Use the keys 1,2,3,4,5,6,7,8,9,0, “.” To input the value and “#” to delete the last input

When done press the exit key to exit.

WARNING: The calibrator value HAS TO BE expressed in mg/dl. No other unit


allowed.

User Manual - rel. 2.0


Chapter 17 - Editor Parameters

17 Editor Parameters

Changing these parameters will take you to dramatically changes in the


instrument’s performance.
If you are not a system specialist please don’t change anything and refer to your
service engineers.

In the parameters menu you will see the following screen

Moving with ▲ and ▼ arrows up and down you can choose the parameter you want to change:

• WAVELENGTH
• CUTLINE A
• CUTLINE B
• DEAD TIME
• NO CLOT SEC
• NORM MIN
• NORM MAX
• RATIO SECS
• INCUBATION
• DELAY

Confirmar comprimiendo “ENTER”.

User Manual - rel. 2.0


Chapter 18 - Parameter Description

18 Parameter Description

WAVELENGTH – This is the wavelength preferred to perform this test, 1 for the blue(450nm),
2 for the red(630nm) and 3 for green(510nm).

CUTLINE A** – Please, not change this parameter, it will be changed only from technical
responsible.

CUTLINE B – Please, not change this parameter, it will be changed only from technical
responsible.

DEAD TIME – This value in seconds during that the clot detector doesn’t work, is a free zone
just to give the time reaction to be more stable optically.

NO CLOT SEC – This is the value in seconds for the NO CLOT message.

SENSPERC - Please, not change this parameter, it will be changed only from technical
responsible.

RATIO SECS – This is value in seconds that will be used in RATIO calculations.
* viene aggiornato in automatico per la PT ad ogni registrazione delle calibrazioni.

INCUBATION – This is value in seconds that will be used from the countdown function in
analysis.

DELAY – This is value in seconds that be added to the final result.


* No negative values allowed
** N.B.: This value only for Fibrinogen tests is set 99.0

User Manual - rel. 2.0


Chapter 19 - Appendix A - Optional Bluetooth printer

19 Appendix A
Optional Bluetooth printer
• In the version Diamond with printer

system follow these steps for print the

results:

• Switch on the printer

• Switch on the diamond and wait 10

seconds for link

• Perform any test Pt, Aptt, Fibrinogen or

TT

• When on display appear the result, press

EXIT for print.

WARNING: Before switch on the instrument and printer, switch off all Bluetooth device

near the instrument(for example the smartphone)


User Manual - rel. 2.0
Chapter 20 - Waste Disposal

20 Waste disposal

Intrument, Used cuvettes, pipette tips, reagent and plasma containers must be disposed in
compliance with biohazard and legal provisions.
The instrument does not fall under Directive 2002/96 / EC
(Directive on waste electrical and electronic di apparecchiature (WEEE).

User Manual - rel. 2.0


Chapter 21 - Troubleshooting

21 Troubleshooting

Instrument does not switch on Check power line, power supply and
connecting cables.

Counter starts prior to start Possible interference from sun light.


pipetting Move away from a window

Air bubbles trigger Check quality of pipet tips

NO clot Check if a clot has been formed or


not.

With duplicate testing Check the quality of the pipet tips (or
reproducibility is not good pipettor) used.

Pressing EXIT - bluetooth printer Check battery status. If necessary,


does not print recharge the battery.
Switch-off other bluetooth devices
(cellphone etc) to avoid
possible interferences.

User Manual - rel. 2.0


Chapter 22 - Customer Service

22 Customer Service

For all questions relating to the use of the diamond instruments, repeatability of the results and
possible damage of the instrument, see TROUBLESHOOTING menu or contact your local
distributor.

Not groped to repair or make modifications to the instrument.

Our staff will help you solve any problem relating to the instrument.

User Manual - rel. 2.0


Chapter 23 - Contact

23 Contact
Enquiries:
Antonio Scordari
info@quadralab.it

Sales:
sales@quadralab.it

Customer Service:
Gennaro Murolo
customer@quadralab.it

TRIDEMA ENGINEERING srl

Via Circumvallazione Esterna, 91


80017 - Melito di Napoli

Tel. +39 081 7101661


Fax +39 081 7100658

Web: www.tridema.it
E-mail: info@tridema.it

Via Circumvallazione Esterna, 91 - 80017 Melito di Napoli - Italy