Warning:
Please read the following instructions carefully. Follow the indications
before you use this device.
Instruction Manual Every effort has been made to avoid errors in the text and diagrams. However,
BioSciTec GmbH assumes no responsibility for any errors which may appear in
this publication.
Improvio C1 L
BioSciTec GmbH
Westerbachstr. 47
60489 Frankfurt/Main
Deutschland
Tel:+49 (0) 69.420.828.84
Fax:+49 (0) 69.420.828.86
Copyright Information
The contents of this report are property of BioSciTec GmbH. The contents are
not to be copied, reproduced or transferred to a third person, without a prior
written permission.
Copyright BioSciTec GmbH, All rights reserved.
• General information
• Maintenance
• Pay attention to all WARNING, CAUTION and STOP statements in this 2.2. Functions
publication.
• Generates pictures, depending on resolution, within a few seconds
• Never open the electrical compartment while the equipment is plugged into a
power source. • Picture handover in form of Bitmaps
• Always work under laboratory regulations. Use a lab coat and pay attention to • Control through Imaging-Software with Twain-Interface
the safety standard procedures in you laboratory.
• Easy to use
• Please be careful while working with infectious material and pay attention to
the safety instructions thoroughly. Always wear gloves for your own safety.
Proceed according to the safety instructions of your laboratory when Caution:
infectious substances get in contact with your skin. If the directions given in this manual are not followed, the system may
not work correctly and can it can be damaged
Warning:
If the Improvio C1 L, the software or corresponding pieces are
damaged, the system cannot work correctly and therefore the warranty
will no longer be valid
2.3. Application
2.4. Validation
Every Improvio C1 L passes a test phase of 100 scans. The Reader is calibrated
using a calibration card. It is included in the device delivery box.
Warning:
If you alter the equipment in any way, you put the warranty in danger
and it will no longer be valid.
The following table shows Improvio C1 L specifications The picture shows the Improvio C1 L scanner with an opened drawer and positioned strips
on the pad.
Parameter Specifications
3. Start-up
Plastic contents: 2 parts of body housing
3.1. Introduction
Drawer: mechanical drawer
In this chapter is described how to install and start this device.
Power Supply: 1 x external power supply, CE certified
This includes unpacking, accessories, the software installation and PC
connections.
Input: 100V – 240V 50-60 Hz, 0.6 A
Warning:
3.2. Device and Accessories
Please, pay attention to the temperature. During winter season, the
• Improvio C1 L Reader instrument can be delivered very cold because of shipment. After
positioning, wait until the instrument has reached the environmental
• Improvio C1 L Drawer with foam mat
temperature. Not till then, start running the instrument.
• Calibration card
• VA 70
Warning:
Please pay attention that the moved out drawer does not reach out over
3.3. Unpacking and Positioning of the Device the working place. This would lead to injuries of the employees and can
damage the instrument.
Place the instrument on a flat, horizontal surface. Remove all indicated
safeguarding and adhesive tape. Put no heavy objects onto the Reader.
Avoid vibrations, strong electromagnetic waves (not computer) and the exposure
Warning: to direct sunlight.
Instruments sent with shipping companies will have transportation
Keep the workplace surrounding the instrument free of any aggressive liquids.
safeguarding
Storage
The drawer has levers on the back that can vary in their position.
4.2. Preparation
4.4. Connecting Computer with Improvio C1 L and calibration
During the installation you will be informed, that you have to connect the scanner
with the PC, before you can go on. Please pay attention to this information and
connect the scanner according to the following steps:
Put the black power plug (only use the original power adapter of the Improvio C1 L -
Mod. 3A-183WP24) into the socket and the rear of the scanner.
Drawer with levers swung out Drawer with levers in hinged position
Caution:
Only use the original power adapter of the Improvio C1 L.
First insert the drawer into the guideways with the levers swung out. Before they
crash into the Reader’s side rests, please swing the levers in and close the drawer
Afterwards connect Improvio C1 L to the PC with the grey USB-cable. Type A ending
completely.
has to be put in the USB-port of the PC, Type B ending in the USB-port of the
scanner. The plugs cannot be mixed up due to their different fit. The device has a
power switch at the rear to turn it on. If the Reader is not in use it will automatically
enter the standby mode after 15 minutes. An input signal of the B4C software can
reactivate the scanner.
Turn the power switch on the rear of the instrument “ON”. At the end proceed
with the software installation.
Note:
Pay attention that the allergen lines are positioned upwards. The
membranes should be placed more or less in the midlle of the pad, not
too close to the border. Check that the troughs are not touching each
other.
Close the drawer as far as it will go. You can start the scan-process now.
Open the drawer after scanning and remove the strips.
Warning:
It may be, that strips will not be measurable in case of heavy colour
The picture shows the Improvio C1 L scanner with an opened drawer and the calibration reactions after incubation. In such cases the membranes can not be
card positioned on the pad. evaluated inadequately.
Note:
Perform a minimum of 5 test scans before starting the first analysis. The
warmer the lamp of the device, the better the scan results. After a longer 5.2. Erroneous image acquisition
break between scans, perform a few empty scans as well. If any image does not reflect the visual perception, please check the correct
Never break up the prewarming phase. The results of the following scan proceeding upon the instructions and restart the image acquisition. If you are not
could be wrong. able to recieve a correct image please contact the manufacturer.
At the rear of the device you can find a switch to turn the device on and off. Warning:
Warning: Use rubber gloves before cleaning the scanner.
If you turn off the device it is not disconnected from the power grid.
Slight contaminations
The device is only disconnected from the power network completely if it is After pulling out the drawer completely, wipe the pad (it is glued in the drawer)
unplugged.
and the drawer with a damp cloth carefully.
Note:
Heavy contaminations
Please position the scanner in such a way that you can easily access
the rear to disconnect the device from the power grid. In case of heavy contaminations of the drawer, pull it out completely (see chapter
4.2). Use rubber gloves for your own safety. Use only pure water to clean the
pad.
Stopp:
7. Cleaning and maintenance Never use Acetone, concentrated Alcohol or acids to clean the device.
Note: Note:
Optimal results will be obtained when the device is operated virtually For the cleaning of the drawer (not for the pad!) you can use mild
dust-free. Clean the drawer and the insert from time to time, depending alkaline detergents. For desinfection you can use mild desinfection
on the level of contamination. agent.
7.1. Outside
Warning:
Clean the device only with water and a mild detergent. For external cleaning, use
For cleaning the pad it is only allowed to use water, do not use
a moist cloth and wipe the device evenly.
detergents.
7.2. Inside
7.3. Cleaning of the glass sheet Any required compensatory accessories will need to be charged.
Warning: Relock the device for transportation (please also refer to 4.1) and use the original
Only qualified personnel is allowed to clean the glass plate. Before packaging. Only in doing so, transportation damages can be prevented.
cleaning, take out the pad and the drawer completely. The instrument can only be shipped in a locked state using the original packing.
Warning:
8. Disposal
Always ensure the disconnection of the device from the power grid by
unplugging the reader. Please pay attention to the respective laws and provisions concerning the
disposal of the device and its accessories. Electrical components need to be
disposed separately.
To clean the Improvio C1 L intensively, please unplug the device and pull the
drawer out.
Warning:
Position the device upside-down and wipe the glass plate with a slightly wet lint-
Always pay attention to your laboratory safety standards when working
free cloth carefully. If it is necessary repeat this procedure until the glass sheet is
with infectious substances before disposing the device and its
clean, dry and striation-free.
accessories.
Put the device back to its initial position carefully. Please place the drawer back
into its guideways and close it completely (see chapter4.2). Finally reconnect the
scanner to the power supply.
No further duties are required for the maintenance of the device and its
accessories.
In case the device needs to be sent in for repair, please exclude any calibration
cards and manuals from the return package.