You are on page 1of 6

Article in press - uncorrected proof

Clin Chem Lab Med 2006;44(3):311–316  2006 by Walter de Gruyter • Berlin • New York. DOI 10.1515/CCLM.2006.054 2006/399

Influence of hemolysis on routine clinical chemistry testing

Giuseppe Lippi*, Gian Luca Salvagno, Martina venient corrective solution might be quantification of
Montagnana, Giorgio Brocco and Gian Cesare free hemoglobin, alerting the clinicians and sample
Guidi recollection.
Istituto di Chimica e Microscopia Clinica,
Keywords: hemolysis; laboratory testing; preanalyti-
Dipartimento di Scienze Morfologico-Biomediche,
cal variability.
Università degli Studi di Verona, Verona, Italy

Abstract
Introduction
Background: Preanalytical factors are the main source
of variation in clinical chemistry testing and among
Over the past decade, the increasing availability of
the major determinants of preanalytical variability,
automation and improvements in laboratory technol-
sample hemolysis can exert a strong influence on
ogy have contributed to diminishing the incidence of
result reliability. Hemolytic samples are a rather com-
errors throughout the analytical laboratory process,
mon and unfavorable occurrence in laboratory prac-
increasing in parallel the reliability of laboratory data
tice, as they are often considered unsuitable for
and allowing a major sense of security in the value of
routine testing due to biological and analytical inter-
laboratory results. A major emphasis is currently
ference. However, definitive indications on the ana-
being placed on preanalytical variables, believed to
lytical and clinical management of hemolyzed
exert a significant influence on results of laboratory
specimens are currently lacking. Therefore, the pres-
tests. As a consequence, the accurate standardization
ent investigation evaluated the influence of in vitro
of each aspect of the preanalytical phase seems piv-
blood cell lysis on routine clinical chemistry testing.
otal to achieve reliable and accurate results and to
Methods: Nine aliquots, prepared by serial dilutions
provide clinically relevant information (1).
of homologous hemolyzed samples collected from 12
Besides procedures immediately related to patient
different subjects and containing a final concentration
preparation and blood collection, such as physical
of serum hemoglobin ranging from 0 to 20.6 g/L, were
activity (2), the fasting state, the blood collection tech-
tested for the most common clinical chemistry ana-
nique (3–5) and the tourniquet time (6, 7), there are
lytes. Lysis was achieved by subjecting whole blood
additional preanalytical circumstances that might
to an overnight freeze-thaw cycle.
influence the reliability of laboratory testing. In partic-
Results: Hemolysis interference appeared to be
ular, problems arising from troublesome specimen
approximately linearly dependent on the final concen-
collection or handling, such as wet alcohol transfer
tration of blood-cell lysate in the specimen. This gen-
from the skin to the blood specimen, small-gauge
erated a consistent trend towards overestimation of
needles, difficulty in locating easy venous access,
alanine aminotransferase (ALT), aspartate amino-
small or fragile veins, unsatisfactory attempts, vein
transferase (AST), creatinine, creatine kinase (CK),
missing, partial obstruction of catheters and other col-
iron, lactate dehydrogenase (LDH), lipase, magnesi-
lection devices, application of a negative pressure to
um, phosphorus, potassium and urea, whereas mean
the blood in the syringe, excessive shaking or mixing
values of albumin, alkaline phosphatase (ALP), chlo-
of the blood after collection, exposure to excessively
ride, g-glutamyltransferase (GGT), glucose and sodi-
hot or cold temperature, centrifugation at a too high
um were substantially decreased. Clinically
speed for a prolonged period of time, frequently com-
meaningful variations of AST, chloride, LDH, potas-
promise the integrity of blood and vascular cells,
sium and sodium were observed in specimens dis-
causing leakage of intracellular components and pro-
playing mild or almost undetectable hemolysis by
ducing significant biological and analytical interfer-
visual inspection (serum hemoglobin -0.6 g/L). The
ence (8, 9).
rather heterogeneous and unpredictable response to
Hemolysis is classically understood as the release
hemolysis observed for several parameters prevented
of hemoglobin and other intracellular components
the adoption of reliable statistic corrective measures
from erythrocytes to the surrounding plasma, follow-
for results on the basis of the degree of hemolysis.
ing damage or disruption of the cell membrane.
Conclusion: If hemolysis and blood cell lysis result
Hemolysis may occur either in vivo or in vitro, and is
from an in vitro cause, we suggest that the most con-
a most undesirable condition that influences the accu-
*Corresponding author: Prof. Giuseppe Lippi, MD, Istituto racy and reliability of laboratory testing (9). Along
di Chimica e Microscopia Clinica, Dipartimento di Scienze with preanalytical causes, in vivo blood cell lysis can
Morfologico-Biomediche, Università degli Studi di Verona, originate from hereditary, acquired, and iatrogenic
Ospedale Policlinico G.B. Rossi, Piazzale Scuro, 10, 37121
conditions, such as autoimmune hemolytic anemia,
Verona, Italy
Phone: q39-045-8074517, Fax: q39-045-8201889, severe infections, intravascular disseminated coagu-
E-mail: ulippi@tin.it lation and transfusion reactions; it does not depend
Article in press - uncorrected proof
312 Lippi et al.: Hemolysis and laboratory testing

on the technique of the healthcare provider and is Laboratory testing


thus virtually unavoidable and cannot be resolved
(10). Visible hemolysis, as a hallmark of a more gen- Hemolysis was assayed by measuring the concentration of
free serum hemoglobin by the reference cyanmethemoglo-
eralized process of blood cell damage, is usually not
bin method on the UV-1700 spectrophotometer (Shimadzu
apparent until the separation of serum or plasma has
Italia S.l.r., Milano, Italy) (13). Hemoglobin determination,
occurred. It is commonly defined as an extracellular white blood cell and platelet counts were performed on an
hemoglobin concentration of )0.3 g/L (4.65 mol/L), ADVIA 120 system (Bayer Diagnostics, Newbury, Berkshire,
resulting in a detectable pink-to-red hue of serum or UK). Serum albumin (bromocresol green colorimetric meth-
plasma with a visible appearance in specimens con- od), alkaline phosphatase (ALP) (IFCC method with amino-
taining as low as 0.5% hemolysate (11). Hemolyzed methyl-propanol buffer), alanine aminotransferase (ALT)
specimens are a rather frequent occurrence in labo- (IFCC method with pyridoxal phosphate activation), aspar-
ratory practice, with prevalence described as being as tate aminotransferase (AST) (IFCC method with pyridoxal
phosphate activation), total bilirubin (diazo method, employ-
high as 3.3% of all of routine samples presented to a
ing 2,5-dichlorophenyl diazonium tetrafluoroborate), calcium
clinical laboratory and accounting for nearly 60% of
(colorimetric method, o-cresolphthalein complexone), crea-
rejected specimens, five-fold more than the second tine kinase (CK) (N-acetylcysteine-activated IFCC method),
most common cause (8, 12). Although laboratory test- creatinine (Jaffe kinetic method), g-glutamyltransferase
ing of hemolyzed specimens has traditionally been (GGT) (Szasz-Persijn method, using L-g-glutamyl-3-carboxy-
discouraged to avoid unreliable results of some bio- 4-nitroanilide), glucose (enzymatic, hexokinase method), iron
chemical parameters, there are no definitive guide- (FerroZine method), lactate dehydrogenase (LDH) (DGKC
lines or recommendations on this topic and clinical method), lipase (colorimetric method, using 1,2-O-dilauryl-
laboratories usually follow arbitrary or individual pro- rac-glycero-3-glutaric acid-6-methylresorufin ester), magne-
cedures to handle results for unsuitable samples due sium (chlorophosphonazo method), inorganic phosphorus
(ammonium molybdate method), urea nitrogen (urease/glu-
to blood cell lysis. Therefore, the aim of the present
tamate dehydrogenase kinetic method), and uric acid (uri-
investigation was to evaluate interference caused by
case/peroxidase enzymatic method) were assayed on a
in vitro hemolysis and blood cell lysis on several rou- Roche/Hitachi Modular System P (Roche Diagnostics GmbH,
tine biochemical parameters and to provide tentative Mannheim, Germany), according to the manufacturer’s
guidelines for the clinical management of hemolyzed specifications and using proprietary reagents. Sodium, chlo-
specimens. ride and potassium were measured on a Roche/Hitachi Mod-
ular System by an indirect method using ion-selective
electrodes. Total imprecision, as expressed by the coefficient
of variation (CV), was less than 2.5% for all analytes tested.
All measurements were performed in duplicate within a sin-
Materials and methods
gle analytical session and results were finally reported as the
mean of paired measurements. The significance of differ-
Experimental design and blood sampling ences between samples was assessed by a paired Student’s
t-test; the level of statistical significance was set at p-0.05.
In the morning of the first day of the evaluation, 3.5 mL Bland-Altman and limits-of-agreement analyses were used
of blood from 12 healthy volunteers, who gave explicit to compare the results of the independent measurements
informed consent for the investigation, was separately col- and differences were finally reported as the percentage of
lected into two 6.0-mL siliconized vacuum tubes containing averages (14).
no additives (Becton-Dickinson, Oxford, UK), using a 20G,
0.80=19 mm Venoject multi-sample straight needle (Teru-
mo Europe NV, Leuven, Belgium). Volunteers were select-
Results
ed on the basis of a homogeneous range of values for
white blood cell count (3.25–4.98=103/mL), platelet count
(153–281=103/mL) and hemoglobin concentration (137– As a result of the overnight freeze-thaw cycle, all the
146 g/L). The first specimen (sample 1) was immediately blood cell counts fell beyond the analytical sensitivity
stored at –708C, whereas the second (sample 2) was centri- of the instrument (white blood cell counts -0.2=
fuged at 1500=g for 10 min at room temperature. Serum 103/mL, platelet count -5=103/mL and erythrocytes
was separated and stored at –708C. -0.1=106/mL). The results of the present investiga-
In the morning of the second day of the evaluation, blood tion on hemolyzed specimens are presented in Tables
from each of the same 12 volunteers was collected into four 1 and 2. Hemolysis interference, expressed as mean
additional 6.0-mL tubes of the same type using the same absolute and percentage relative bias, appeared to be
type of needle. Samples were pooled (sample 3) and finally approximately linearly dependent on the final concen-
divided into aliquots of 2 mL. Samples 1 and 2 were thawed; tration of blood cells lysate in the specimen. As
nine serial dilutions of free serum hemoglobin were pre-
expected, the addition of blood cell lysate generated
pared by mixing scalar aliquots of samples 1 and 2. Then
a consistent and dose-dependent trend towards over-
200 mL of each of these dilutions was added to 2-mL aliquots
estimation of ALT, AST, creatinine, CK, iron, LDH, lip-
of blood collected on day 2 (sample 3) to achieve free serum
hemoglobin concentrations in the specimens ranging from ase, magnesium, phosphorus, potassium and urea,
0 to 20.6 g/L, thus being almost representative of the hemol- whereas mean values of albumin, ALP, chloride, GGT,
ysis degree we randomly observe in samples sent to our glucose and sodium were substantially decreased
laboratory. Blood was then centrifuged at 1500=g for 10 min when compared with the baseline specimens (no
at room temperature and serum was separated and imme- lysate). Statistically significant differences in samples
diately analyzed. containing 0.6 g/L of serum hemoglobin or less could
Table 1 Influence of hemolysis on routine clinical chemistry testing. Differences are given as absolute values and percentage relative bias (mean"SD) from the baseline sample (no lysis).
Values are compared with current analytical quality specifications for desirable bias (15). Statistically significant differences were evaluated by Student’s paired t-test (ap-0.05; bp-0.01).
Results with hemolysis-induced variations exceeding the relative desirable bias are marked in bold.

Analyte Desirable Free serum hemoglobin


bias, %
No lysis 0.16 g/L 0.3 g/L 0.6 g/L 1.3 g/L 2.6 g/L 5.1 g/L 10.3 g/L 20.6 g/L

Albumin, g/L "1.3 44.9"1.1 44.8"1.0 44.6"1.3 44.5"1.1 44.3"1.0 44.3"1.2 44.2"1.1 43.7"1.1 42.2"1.1
(y0.1"0.4%) (y0.5"0.9%) (y0.8"0.4%) (y1.1"0.7%) (I1.3"0.8%) (I1.4"0.7%) (I2.5"0.4%) (I5.9"1.0%)
Alkaline phosphatase, U/L "6.4 52"8 52"8 52"7 51"8 50"7b 47"7b 43"7b 36"7b 24"6b
(0.0"1.0%) (I0.4"1.6%) (I1.4"2.1%) (I3.0"1.4%) (I9.2"2.0%) (I18.0"2.3%) (I31.4"3.6%) (I53.7"5.3%)
Alanine aminotransferase, "12.0 20"13 20"13 20"13 20"13 20"13 22"13a 23"13b 26"13b 32"13b
U/L (H1.6"2.8%) (H2.2"8.0%) (H2.3"7.2%) (H2.9"7.8%) (H10.2"11.9%) (H18.2"11.6%) (H37.1"16.3%) (H73.8"28.2%)
Aspartate aminotransferase, "5.4 22"4 23"4b 25"4b 27"4b 31"4b 40"5b 59"7b 95"11b 169"22b
U/L (H4"3%) (H12"5%) (H24"7%) (H42"9%) (H84"14%) (H169"25%) (H336"48%) (H677"87%)
Bilirubin (total), mmol/L "10.0 20.8"10.3 20.8"10.4 20.6"10.4 20.6"10.0 20.4"10.8 20.4"10.9 20.3"10.2 20.3"9.6 20.2"9.3
(I0.1"0.8%) (I1.1"2.9%) (I0.3"2.2%) (I4.0"5.9%) (I4.2"7.4%) (I2.8"4.9%) (I0.6"7.7%) (I1.3"9.2%)
Calcium, mmol/L "0.8 2.40"0.09 2.40"0.10 2.40"0.09 2.40"0.09 2.41"0.09 2.41"0.09 2.42"0.09 2.42"0.10 2.40"0.09
(q0.1"0.9%) (q0.1"0.5%) (q0.1"0.7%) (q0.3"0.5%) (q0.4"0.8%) (q0.6"0.7%) (q0.6"0.7%) (I0.1"0.7%)
Chloride, mmol/L "0.5 105.5"1.9 105.5"1.9 105.0"1.7 104.8"1.6 104.6"1.5 104.3"1.3 104.0"1.5 103.6"1.9 102.5"1.6
(0.0"0.0%) (I0.5"0.5%) (I0.7"0.6%) (I0.8"0.6%) (I1.2"0.6%) (I1.4"0.8%) (I1.8"0.7%) (I2.8"0.9%)
Creatine kinase, U/L "11.5 118"46 120"47a 121"47b 125"47b 131"48b 145"47b 175"47b 240"50b 380"55b
(H2"2%) (H3"2%) (H7"4%) (H13"4%) (H27"10%) (H56"22%) (H120"46%) (H259"102%)
Creatinine, mmol/L "3.4 81"10 82"11 82"11 82"10 83"10a 85"11b 86"10b 86"10b 87"9b
(H0.2"1.8%) (H0.3"2.3%) (H0.4"2.0%) (H1.5"1.4%) (H4.4"2.4%) (H5.8"2.4%) (H6.3"2.3%) (H7.5"3.6%)
g-Glutamyltransferase, U/L "10.8 17"8 16"8a 16"8b 16"8b 16"8b 15"8b 15"8b 14"7b 12"7b
(I4.7"4.6%) (I6.0"3.9%) (I6.4"3.7%) (I7.8"3.1%) (I11.1"5.3%) (I14.5"5.7%) (I18.3"5.4%) (I35.5"9.7%)
Glucose, mmol/L "2.2 5.02"0.66 5.02"0.68 5.02"0.66 5.02"0.64 5.02"0.64 5.02"0.64 5.01"0.64 4.97"0.65b 4.95"0.65b
(I0.2"0.6%) (I0.1"0.4%) (0.0"0.8%) (0.0"0.5%) (I0.1"0.3%) (I0.2"0.9%) (I1.0"0.4%) (I1.4"1.0%)
Iron, mmol/L "8.8 18.4"5.8 18.4"5.8 18.6"5.6 18.6"5.7b 18.8"5.8b 19.2"5.7b 20.1"5.8b 22.2"5.9b 27.3"6.1b
(H0.1"0.9%) (H1.5"2.1%) (H1.5"1.3%) (H2.1"0.6%) (H4.7"2.0%) (H10.6"3.9%) (H23.1"8.3%) (H53.1"16.0%)
Article in press - uncorrected proof

Lactate dehydrogenase, U/L "4.3 338"37 369"36b 397"38b 451"39b 550"42b 746"52b 1203"69b 1906"116b 3804"140b
(H9"2%) (H17"2%) (H34"4%) (H63"7%) (H122"13%) (H259"34%) (H470"69%) (H1039"132%)
Lipase, U/L "10.1 28"6 28"6 29"6 29"5 31"5a 35"6b 39"5b 43"5b 45"5b
(0.0"0.0%) (H0.5"1.3%) (H2.2"4.0%) (H8.2"7.8%) (H23.1"7.7%) (H38.0"9.9%) (H53.1"12.7%) (H60.7"15.9%)
Magnesium, mmol/L "1.8 0.92"0.04 0.92"0.04 0.92"0.04 0.92"0.03 0.93"0.03b 0.94"0.04b 0.95"0.03b 0.98"0.04b 1.02"0.03b
(0.0"0.4%) (0.0"0.8%) (H0.4"2.1%) (H1.5"2.1%) (H2.3"0.7%) (H3.5"1.0%) (H6.8"1.6%) (H11.6"2.3%)
Phosphorus, mmol/L "3.2 1.17"0.12 1.17"0.11 1.17"0.11 1.18"0.12 1.18"0.11 1.20"0.11b 1.23"0.11b 1.29"0.11b 1.42"0.11b
(H0.1"1.3%) (H0.6"1.3%) (H0.8"1.0%) (H1.4"1.6%) (H2.9"1.7%) (H5.5"2.0%) (H10.9"3.4%) (H22.1"7.4%)
Potassium, mmol/L "1.8 4.12"0.22 4.15"0.22a 4.18"0.23b 4.25"0.21b 4.37"0.22b 4.65"0.24b 5.22"0.25b 6.36"0.30b 8.78"0.42b
(H0.7"0.7%) (H1.5"0.8%) (H3.2"1.2%) (H6.1"1.1%) (H12.9"1.4%) (H26.7"1.6%) (H54.4"3.1%) (H113.2"5.6%)
Sodium, mmol/L "0.3 140.1"1.5 139.6"1.1a 139.3"1.0b 138.9"0.8b 138.6"0.9b 138.3"1.2b 137.9"1.4b 136.5"0.9b 133.5"1.0b
(I0.4"0.4%) (I0.6"0.4%) (I0.9"0.6%) (I1.1"0.5%) (I1.3"0.5%) (I1.6"0.5%) (I2.6"0.5%) (I4.7"0.8%)
Urea nitrogen, mmol/L "5.5 6.3"1.2 6.3"1.1 6.3"1.0 6.3"1.1 6.3"1.1 6.4"1.1a 6.5"1.1a 6.6"1.2b 6.8"1.1b
(I0.4"2.3%) (I0.4"2.6%) (I0.6"2.3%) (I0.5"1.5%) (q1.4"1.6%) (q3.0"2.4%) (q3.9"2.2%) (H8.9"2.6%)
Uric acid, mmol/L "4.8 305"89 303"89 303"89 303"88 302"89 301"89 296"89 290"87 275"86
Lippi et al.: Hemolysis and laboratory testing 313

(I0.6"1.0%) (I0.8"1.1%) (I0.9"1.1%) (I1.0"1.1%) (I1.3"1.1%) (I3.2"1.4%) (I5.1"2.0%) (I10.3"2.4%)


Article in press - uncorrected proof
314 Lippi et al.: Hemolysis and laboratory testing

Table 2 Influence of hemolysis on routine clinical chemistry testing: mean coefficient of variation (CV) (ns12) determined
from the reference specimen (no lysis) and each one of the homologous eight samples containing scalar dilutions of free
serum hemoglobin.

Analyte Desirable Free serum hemoglobin


bias, %
0.16 g/L 0.3 g/L 0.6 g/L 1.3 g/L 2.6 g/L 5.1 g/L 10.3 g/L 20.6 g/L

Albumin, g/L "1.3 0.2% 0.4% 0.4% 0.6% 0.7% 0.7% 1.3% 3.1%
Alkaline phosphatase, U/L "6.4 0.2% 0.5% 1.0% 1.6% 4.8% 9.9% 18.7% 36.9%
Alanine aminotransferase, U/L "12.0 0.8% 3.2% 2.6% 3.0% 5.9% 8.1% 15.2% 26.1%
Aspartate aminotransferase, U/L "5.4 2.1% 5.6% 10.5% 17.3% 29.4% 45.5% 62.3% 77.0%
Bilirubin (total), mmol/L "10.0 0.3% 1.2% 1.0% 2.9% 3.0% 1.8% 3.3% 3.9%
Calcium, mmol/L "0.8 0.3% 0.2% 0.3% 0.3% 0.4% 0.3% 0.3% 0.3%
Chloride, mmol/L "0.5 0.0% 0.2% 0.4% 0.4% 0.6% 0.7% 0.9% 1.4%
Creatine kinase, U/L "11.5 1.1% 1.6% 3.4% 6.0% 11.6% 21.3% 36.2% 54.3%
Creatinine, mmol/L "3.4 0.8% 0.8% 0.9% 0.7% 2.1% 2.8% 3.0% 3.6%
g-Glutamyltransferase, U/L "10.8 2.5% 3.2% 3.4% 4.1% 6.0% 7.9% 10.2% 22.0%
Glucose, mmol/L "2.2 0.2% 0.1% 0.3% 0.1% 0.1% 0.3% 0.5% 0.7%
Iron, mmol/L "8.8 0.3% 0.7% 0.7% 1.0% 2.3% 5.0% 10.2% 20.7%
Lactate dehydrogenase, U/L "4.3 4.4% 8.0% 14.4% 24.0% 37.7% 56.2% 69.8% 83.7%
Lipase, U/L "10.1 0.0% 0.3% 1.0% 3.8% 10.2% 15.8% 20.8% 23.0%
Magnesium, mmol/L "1.8 0.2% 0.3% 0.7% 0.9% 1.1% 1.7% 3.3% 5.5%
Phosphorus, mmol/L "3.2 0.4% 0.6% 0.5% 0.9% 1.4% 2.7% 5.1% 9.9%
Potassium, mmol/L "1.8 0.4% 0.7% 1.6% 3.0% 6.1% 11.8% 21.4% 36.1%
Sodium, mmol/L "0.3 0.2% 0.3% 0.4% 0.5% 0.7% 0.8% 1.3% 2.4%
Urea nitrogen, mmol/L "5.5 1.0% 1.1% 0.9% 0.7% 0.7% 1.6% 1.9% 4.3%
Uric acid, mmol/L "4.8 0.3% 0.4% 0.4% 0.5% 0.7% 1.6% 2.6% 5.4%
Values are compared with current analytical quality specifications for desirable bias (15).

already be observed for AST, LDH, potassium and Specimen Acceptance Q-Probes study, as it influenc-
sodium. Albumin, total bilirubin, calcium, chloride es the reliability of laboratory tests for both biological
and uric acid measurements were not significantly and analytical reasons (12). Although the interference
different, even in specimens containing up to 20.6 seems to be related to the degree of lysis and the
g/L of serum hemoglobin. Clinically meaningful vari- specificity of the method being used, it has been
ations, as reflected by deviation from the current ana- reported that several laboratory results can be seri-
lytical quality specifications for desirable bias (15), ously influenced, especially those for potassium, sodi-
were observed for AST, chloride, LDH, potassium and um, calcium, magnesium, bilirubin, haptoglobin, total
sodium in samples with mild or almost undetectable protein, aldolase, amylase, LDH, AST, ALT, phospho-
hemolysis by visual inspection (0.6 g/L of serum rus, alkaline phosphatase, acid phosphatase, GGT,
hemoglobin) (Table 1). Of the analytes evaluated, the folate, and iron (17–19). However, there are no defin-
bias values recorded for total bilirubin, calcium and itive indications on the degree of lysis responsible for
glucose were the only ones that did not achieve a clin- such interference, nor are definitive guidelines avail-
ically significant variation, even at the highest con- able on the most suitable procedure for reporting and
centration of serum hemoglobin (20.6 g/L). Despite handling results of biochemical testing on hemolyzed
the approximately linear relationship of variations specimens. Results of the present investigation raise
to free hemoglobin concentration in serum, an un- two critical issues: the influence of mild or almost
predictable, heterogeneous and sample-specific re- undetectable hemolysis by visual inspection (-0.6
sponse was observed for several parameters, as g/L of serum hemoglobin) on some analytes, such as
shown by the amplitude of the mean CVs calculated AST, LDH, potassium and sodium; and the manage-
from each of the 12 independent patient samples test- ment of laboratory results for hemolytic specimens.
ed (Table 2). The interference of in vitro hemolysis on laboratory
testing might be caused by: (i) leakage of hemoglobin
and other intracellular components into the surround-
Discussion ing fluid, which induces false elevations of some ana-
lytes or dilution effects; (ii) chemical interference by
There are many opportunities for errors in the pre- free hemoglobin in the analytical reaction; and
analytical phase. Phlebotomists must be aware of the (iii) method- and analyte concentration-dependent
possibility of complications and interference in results spectrophotometric interference due to an increase in
of laboratory testing, should they fail to follow an the optical absorbance or a change in the blank value,
accurate and standardized blood collection technique especially for laboratory measurements at 415, 540
(16). Hemolysis, reflecting more generalized blood and 570 nm, where hemoglobin shows strong absor-
cell damage, is a rather common and unfavorable bance (8, 20, 21). Clearly, at high levels of serum
occurrence in laboratory practice. It is the most fre- hemoglobin, such interferences might coincide, caus-
quent reason for specimen rejection, as indicated by ing variations that are not necessarily in the same
the College of American Pathologists (CAP) Chemistry direction and making the situation rather difficult to
Article in press - uncorrected proof
Lippi et al.: Hemolysis and laboratory testing 315

understand. As for ALT, AST, LDH, magnesium, phos- values for hemolysis, as tentatively suggested for
phorus and potassium, the interference mechanism potassium (22, 23).
has previously been characterized and is attributed to Instead of reporting the post-hemolysis corrected
large differences between intracellular and extracel- potassium result, Dimeski et al. recently suggested
lular concentrations for these analytes (21). Lower val- the provision of a qualitative comment indicating the
ues for glucose, sodium and chloride are likely caused likely range of the potassium concentration and rec-
by dilutional effects, whereas hemolysis interference ommended that if results lie in a critical range, they
in the CK assay has been attributed to intracellular should be communicated promptly to the requesting
adenylate kinase, which is not wholly inhibited under clinician (24). A similar option is to report results of
operating conditions (20, 21). For ALP, iron, lipase and hemolyzed specimens that include messages or flags.
GGT, hemolysis interference is likely caused by However, these solutions appear rather questionable
spectral overlap and by a chemical reaction between from both the analytical and clinical perspective, irre-
hemolysate and reaction components (21). spective of the correction formulas calculated from
At variance with biological causes, in vitro blood linear regression of absolute error vs. hemoglobin
cell lysis is preventable, as it is usually caused by concentration. In fact, we believe that whenever
inappropriate collection, handling and processing of results of laboratory testing are reasonably thought
the specimen. Hemolysis and blood cell lysis are to be analytically or clinically unreliable, they should
almost undetectable by visual inspection until whole not be reported to prevent misleading information
blood has been centrifuged, exposing the serum or that could finally compromise the whole clinical rea-
plasma to scrutiny. Although it is unlikely that hemo- soning. Although the elimination of some types of
lyzed and unsuitable samples can be observed in interference, such as that caused by cell lysis, is prob-
problem-free phlebotomy activities, clinical laborato- lematic and often impossible, it is nevertheless
ries should accurately identify and document each conceivable to influence spectral interference by mul-
procedure that appears to be more sensitive to in tiwavelength analysis and blank measurement or
vitro hemolysis and possibly to what extent it is preparation (21). Obviously, this solution would not
affected. Three potential approaches can be identified work for interferences due to leakage of intracellular
to deal with hemolyzed specimens: hemolysis correc- analytes or dilution effects (i.e., AST, ALT and LDH).
tion; result reporting with a clear indication of the Therefore, if hemolysis and blood cell lysis result
potential interference arising from blood cell lysis; from an in vitro cause, we support previous indica-
and sample recollection. Correction factors for facti- tions that the most suitable corrective measure might
tious hyperkalemia in a clinically relevant range were be free hemoglobin quantification, warning of the cli-
identified and result correction for hemolysis was fur- nicians so that in vivo hemolysis can be ruled out and,
eventually, sample recollection (8).
ther recommended as a reliable alternative to provide
A further issue is the clinical management of labo-
clinically useful information for unsuitable specimens.
ratory testing in patients with hemolytic disorders. In
This was achieved by multiplying the hemoglobin
vivo hemolysis because of either a hematological dis-
concentration by the slope obtained from a linear
ease or a medical procedure, including extracorporeal
regression analysis between the bias observed for
circulation, mechanical devices, irradiative treatments
each analyte at the relative free serum hemoglobin
and drugs, generally becomes apparent when the red
concentration. Hypothetically, when the lower bound
blood cell lifespan is shortened by more than 50%.
of the predicted change in potassium results in a cor-
Plasma haptoglobin is capable of binding free hemo-
rected value within the reference range, a second
globin, thus allowing its effective clearance by the
blood draw might be unnecessary (20, 22). Accord-
reticulo-endothelial system (10). Less is known about
ingly, some instruments already report an ability to
the in vivo clearance of other elements and sub-
identify hemolytic specimens and eventually correct
stances released during hemolysis. When an in vivo
results for the degree of hemolysis. The results of the hemolytic state is suspected, should the sample or the
present investigation testify that such an approach results of the biochemical assays, known to be influ-
might be inaccurate and potentially misleading. In enced by hemolysis, be analyzed and/or taken into
fact, given the wide and heterogeneous interference account and reported, or discarded? A recent article,
observed in hemolytic specimens obtained from 12 reporting a critical case with hyperkalemia induces
different patients (Table 2), we do not recommend caution, and suggests that the correct laboratory pro-
hemolysis correction in clinical practice. Although the cedure and reporting should be guided by clinical
main error is approximately linearly dependent on rather than by analytical considerations (25). Further-
hemoglobin concentration and independent of the more, the article raises concern about the non-critical
initial analyte concentration, the mean variations use of automatic potassium suppression when a sus-
observed for AST and LDH exceeded the desirable pected hemolyzed sample is encountered. A different
bias in even mild hemolyzed specimens. Moreover, case is the occurrence of elevated enzyme activities,
for specimens containing 1.3–10.3 g/L free hemoglo- such as AST and CK, in samples from patients with
bin, which includes the majority of visibly hemolyzed suspected in vivo hemolysis. Correct use of the ana-
specimens sent to our laboratory, the mean CV for 13 lytical resources currently available should allow the
out of the 20 analytes tested in 12 separate patient correct assessment of every uncertain condition.
samples exceeded the desirable bias (Table 2), thus We are aware that our investigation has two limi-
preventing the identification of a ratio for correcting tations. First, we have reproduced in vitro blood cell
Article in press - uncorrected proof
316 Lippi et al.: Hemolysis and laboratory testing

lysis by freezing blood specimens at y708C for up to 6. Lippi G, Salvagno GL, Montagnana M, Brocco G, Guidi
24 h. Besides lysate erythrocytes, hemolyzed speci- GC. Influence of short-term venous stasis on clinical
chemistry testing. Clin Chem Lab Med 2005;43:869–75.
mens encountered in laboratory practice might also
7. Lippi G, Salvagno GL, Montagnana M, Guidi GC. Short-
contain destroyed or fragmented platelets, leukocytes term venous stasis influences routine coagulation test-
and vascular cells. Therefore, our study design ing. Blood Coagul Fibrinolysis 2005;16:453–8.
appears to be a suitable surrogate, although it is not 8. Kroll MH, Elin RJ. Interference with clinical laboratory
representative of all the possible events that induce analyses. Clin Chem 1994;40:1996–2005.
whole blood lysis in laboratory practice, especially 9. Carraro P, Servidio G, Plebani M. Hemolyzed specimens:
those represented by troublesome blood collection. a reason for rejection or a clinical challenge? Clin Chem
2000;46:306–7.
We evaluated such interference on single reagents
10. Rother RP, Bell L, Hillmen P, Gladwin MT. The clinical
and a defined analyzer, and thus our results might not sequelae of intravascular hemolysis and extracellular
be universally reproducible with other testing sys- plasma hemoglobin: a novel mechanism of human dis-
tems, especially those capable of identifying hemo- ease. J Am Med Assoc 2005;293:1653–62.
lytic specimens. 11. Burns ER, Yoshikawa N. Hemolysis in serum samples
Remarkable improvements in analytical quality drawn by emergency department personnel versus lab-
oratory phlebotomists. Lab Med 2002;33:378–80.
guarantee that the actual performance of clinical lab-
12. Jones BA, Calam RR, Howanitz PJ. Chemistry specimen
oratories is suitable for improving the patient’s acceptability. A College of American Pathologists Q-
health. However, the large percentage of laboratory Probes study of 453 laboratories. Arch Pathol Lab Med
errors still occurring in the pre- and postanalytical 1997;121:19–26.
phases are a source of major concern (1). The issue 13. National Committee for Clinical Laboratory Standards.
of hemolysis and blood cell lysis has traditionally Reference and selected procedures for the quantitative
plagued clinical laboratories, pointing to a need for determination of hemoglobin in blood: approved stan-
dard, 3rd ed. NCCLS document H15-A3. Wayne, PA:
operative guidelines and recommendations. Collec-
NCCLS, 2000.
tion of specimens of consistent quality, based on 14. Bland JM, Altman DG. Measuring agreement in method
proper training and knowledge of the factors that can comparison studies. Stat Methods Med Res 1999;8:
influence laboratory results, is essential to minimize 135–60.
errors at any step of the laboratory work-up, to opti- 15. Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez
mize resource utilization and quality, and to improve A, Jimenez CV, et al. Current databases on biologic var-
the whole patient management procedure. The fre- iation: pros, cons and progress. Scand J Clin Lab Invest
1999;59:491–500.
quency of in vitro hemolysis and additional pre-
16. Ogden-Grable H, Gill GW. Phlebotomy puncture junc-
analytical interference might be lowered by imple- ture: preventing phlebotomy errors – potential for harm-
menting standardized collection procedures, such as ing your patients. Lab Med 2005;36:430–3.
conventional straight needles and evacuated tube 17. Laessig RH, Hassemer DJ, Paskey TA, Schwartz TH. The
systems, instead of butterfly devices or syringes (5). effects of 0.1% and 1.0% erythrocytes and hemolysis on
Laboratory guidelines on this topic should then serum chemistry values. Am J Clin Pathol 1976;66:
639–44.
encompass procedures for identification of speci-
18. Sonntag O. Haemolysis as an interference factor in clin-
mens with clinically significant hemolysis, such as ical chemistry. J Clin Chem Clin Biochem 1986;24:
automated hemolysis/turbidity detection-quantifica- 127–39.
tion, measurement and immediate transmission of 19. Yucel D, Dalva K. Effect of in vitro hemolysis on 25 com-
results of laboratory testing, which can provide the mon biochemical tests. Clin Chem 1992;38:575–7.
clinician with essential information for clinical situa- 20. Jay D, Provasek D. Characterization and mathematical
tions requiring immediate intervention (9). correction of hemolysis interference in selected Hitachi
717 assays. Clin Chem 1993;39:1804–10.
21. Grafmeyer D, Bondon M, Manchon M, Levillain P. The
influence of bilirubin, haemolysis and turbidity on 20
analytical tests performed on automatic analysers.
References
Results of an interlaboratory study. Eur J Clin Chem Clin
Biochem 1995;33:31–52.
1. Bonini P, Plebani M, Ceriotti F, Rubboli F. Errors in labo- 22. Owens H, Siparsky G, Bajaj L, Hampers LC. Correction
ratory medicine. Clin Chem 2002;48:691–8. of factitious hyperkalemia in hemolyzed specimens. Am
2. Lippi G, Brocco G, Franchini M, Schena F, Guidi G. Com- J Emerg Med 2005;23:872–5.
parison of serum creatinine, uric acid, albumin and glu- 23. Hawkins R. Variability in potassium/hemoglobin ratios
cose in male professional endurance athletes compared for hemolysis correction. Clin Chem 2002;48:796.
with healthy controls. Clin Chem Lab Med 2004;42:644–7. 24. Dimeski G, Clague AE, Hickman PE. Correction and
3. Lippi G, Guidi GC. Effect of specimen collection on routine reporting of potassium results in haemolysed samples.
coagulation assays and D-dimer measurement. Clin Chem Ann Clin Biochem 2005;42:119–23.
2004;50:2150–2. 25. Ismail A, Shingler W, Seneviratne J, Burrows G. In vitro
4. Lippi G, Salvagno GL, Guidi GC. No influence of a butter- and in vivo hemolysis and potassium measurement. Br
fly device on routine coagulation assays and D-dimer Med J 2005;330:949.
measurement. J Thromb Haemost 2005;3:389–91.
5. Lippi G, Salvagno GL, Brocco G, Guidi GC. Preanalytical
variability in laboratory testing: influence of the blood
drawing technique. Clin Chem Lab Med 2005;43:319–25. Received October 31, 2005, accepted December 5, 2005

You might also like