Overview:
4D imaging enables the display of the real time 3D anatomical structure and adds the dimension of
‘movement’ to a 3D image by providing continuous, real-time displays of internal organs or fetus.
3D imaging provides single frame images, while 4D imaging provides continuous high volume
acquisition of 3D images. The following part is about the operating steps of 4D imaging in the
system with touch screen.
01
Adjust the ROI and sample line:
The 2D imaging with ROI and sample line is
displayed in the inactivated 4D mode, only the data
in ROI is acquired for 4D imaging.
Tap【1】,
【2】,
【4】or , , to enter the
full, dual or quad display.
A:X-axis B:Y-axis
C:Z-axis Dynamic
3D
02
Tap【Clip Plane】and move the trackball to
observe reviews by discretionary plane.
② View
Tap the 6 directions or the left/right part of
【Rotate X】,【Rotate Y】,【Rotate Z】,【Move
When you adjust the size of the ROI or the U/D】,【Move L/R】or rotate the button to
position of sample line, you can tap【Volume】, adjust the direction of the images.
【Curve】or【All】to reset.
④ C-Plane
C-Plane is a function to observe the shape of
different planes of the 4D volume.
Tap【A, B, C】to select a desired image , then
the desire image on the corresponding axis is
Tap【A, B, C】to choose a ref. image, then tap
displayed on the screen.
【AB, AC, BC, ABC】to choose the display
Tap【3*3, 4*4, 5*5】to set the slice number.
mode, and you can clearly analyze the structure
Tap the left/right part of【Slice Spacing】to set
of the planes.
the spacing(0.5-2).
Step Five:Save
⑤ M-Slice Press 【 Save 】 to save the 4D Image/Cine
You can observe reviews by using 2 or more /Volume, and the volume data can be modified
slices. For example, each layer of an irregular after being saved.
tumor can be observed by applying this feature. Tap【Exit】to return to the real-time B mode.
04