You are on page 1of 5

Main feature

Leak rate measurement for pharmaceutical isolators:


Practical guidance for operators and test engineers
Tim Coles

Abstract This then means that the leaktightness It would therefore make sense for
This paper starts by defining leaktightness of the isolator needs to be stated at the isolator manufacturers to specify which
of isolators and explaining the difference design stage and then quantified during of these specified classes their products
between leak rate measurement and leak validation at FAT, SAT, OQ and also comply with. They may choose to meet
detection. It then sets out the classes of during subsequent PPM. a slightly different standard. For example;
isolator leakage rates from 14644-7:2004 Isolator operators and test engineers one manufacturer works to 0.50% volume
and ISO 10648-2:1994 and gives will often refer to leak testing. The loss per hour. This is quite acceptable
recommendations as to which class of correct description of the procedure is, since no absolute rules exist but, on the
leak rate is applicable for which operation. in fact, leak rate measurement. other hand, standards are developed to
Leak rate measurement is described in encourage consistency, so why deviate
some detail starting with the different Leak rate and leak detection from accepted norms?
methods available including pressure It is important to understand at an early Leak rate can be expressed in a variety
decay, pressure hold and nitrogen stage the difference between leak rate of terms and units other than percentage
dilution. Theoretical considerations on measurement and leak detection. The leak volume loss per hour. Pharmaceutical
the pressure decay test are followed by rate may be defined as the amount of air Isolators mentions reciprocal hours. Some
practical guidance and examples of the lost from a positive isolator or gained by a manufacturers refer to specific pressure
different expressions for pressure decay. negative isolator per unit time. The most decay rates per unit time (e.g. 16 Pascals
The paper then goes on to describe the convenient way of expressing leak rate is per minute), others express the leak rate
relative merits of various leak detection probably as percentage volume loss per the other way round, namely time to give
methods including the use of helium, hour. By contrast, leak detection is applied a fixed maximum pressure decay (e.g. 6
DOP (dispersed oil particulates), when the isolator has failed a leak rate test, minutes giving a maximum of 60 Pa decay).
ammonia with proprietary bromophenol and detection is then used to find out where Tables comparing these alternative leak
cloth, soap bubbles and ultrasonics. The leakage is taking place in order to fix it. rate expressions are given later in the paper.
paper concludes with sections on testing A further important point to note is the Once again, a more logical approach
gloves and half-suits, when to test, the fact that all isolators leak. The question is surely uses the established standards
distributed leak test and induction leaks. then by how much? and their defined classes. So which class
is appropriate for which operation? As
previously noted, there are no rules,
It is important to understand at an early stage the difference but the following broad advice is offered:
Class 1. Class III microbiological
between leak rate measurement and leak detection. The leak safety cabinets and very high
rate may be defined as the amount of air lost from a positive containment isolators.
Class 2. Negative pressure
isolator or gained by a negative isolator per unit time. The aseptic isolators.
most convenient way of expressing leak rate is probably as Class 3. Positive pressure
aseptic isolators.
percentage volume loss per hour. By contrast, leak detection Class 4. Not appropriate for
is applied when the isolator has failed a leak rate test, and pharmaceutical isolators.

detection is then used to find out where leakage is taking It can also be argued that the class of
place in order to fix it. leak rate chosen should depend on the
grade of cleanroom in which the isolator
is housed; a better cleanroom permits a
Introduction Standards, guidelines and units higher leak rate, especially in the case
Since isolators are designed to maintain a There are no rules on leak rate. of a negative isolator used for aseptic
specialised environment, it makes sense However, we do have some guidelines. processing, but this has to be decided on
to specify the nature, or performance, Both ISO 14644-7:2004 and ISO a case to case basis.
of the containment. This is done partly 10648-2:1994 specify four classes of
by defining the filters on the ventilation leaktightness as follows: Leak rate measurement methods
system and partly by defining the quality Class 1: 0.05% volume loss per hour How then should the leak rate of an
of the barrier. The leaktightness is a Class 2: 0.25% volume loss per hour isolator be measured in a practical and
fundamental measure of the quality Class 3: 1.0% volume loss per hour rational way? There are a number of
of the barrier. Class 4: 10.0% volume loss per hour methods, as follows:

8 Clean Air and Containment Review | Issue 11 | July 2012 www.cleanairandcontainment.com


Main feature

D 3UHVVXUHGHFD\ to the human body of excessive tells you that the allowable decay is 42 Pa.
The majority of users will opt for concentrations of nitrogen as there In many ways, Equation 2 is more useful,
pressure decay testing. This is easy to is with carbon dioxide). but the two equations are the same thing,
perform and does not need specialised just with a different prime subject.
G 3DUMRDQG)RVFR0HWKRGV
equipment or highly trained personnel; Having mastered the leak rate
The nuclear industry is believed to have
indeed the isolator may be equipped to equation, the next decision is the starting
used the Parjo method for gloveboxes.
perform the test itself. It is essentially just test pressure. The four classes of leak rate
The test equipment consists of a
a question of closing off all the ports and may be assumed to be at working
reference vessel inside the glovebox,
valves on the isolator, raising the isolator pressure and we could indeed measure
with a glass tube connected into the top.
to a test pressure (or lowering to a test the leak rate at working pressure;
The tube has a horizontal section into
depression), and noting the change in however, some added safety margin
which a soap bubble is introduced. The
pressure differential to ambient over would be the norm and a quick test is
soap bubble is effectively a frictionless
time. However, there are a number of generally preferred. Thus some multiple of
piston. Any leakage of the sealed
practical considerations and these are working pressure would be rational. Some
glovebox will cause the bubble to move
detailed later. testers use pressures as high as five times
along the tube. The speed at which it
working pressure but three times working
E 3UHVVXUHKROG moves can be used to calculate the leak
pressure is probably adequate. Two times
In many ways, the pressure hold method rate quite easily. The Fosco method
working pressure may also be considered
is an ideal method to measure leak rate replaces the bubble tube with a
adequate as long as the time of the test is
since it actually quantifies the leak rate micromanometer. In practice, these
such that the isolator does not actually
directly. In this test the isolator is sealed methods only work where very low leak
drop below working pressure.
off as if for the pressure decay test and is rates are involved and where conditions
The next consideration is the
then supplied with air from a suitable for testing are well controlled. Not ideal
arrangement of the isolator during
pump, via an air flow meter. By careful then for busy hospital pharmacies!
pressure decay tests, more specifically,
adjustment of the air pump, it should be
what do you do with the sleeves,
possible to raise the isolator to a suitable Theoretical considerations
gauntlets or half-suits? If these item
test pressure and then hold that pressure for pressure decay
move around in the test, the volume
steady. The air flow required to hold that Perhaps the first thing you need to
changes and the pressure changes
steady test pressure can then be read off know here is how to convert raw
accordingly. Perhaps they should be
the flow meter and that reading is, of pressure decay data into percentage
removed during the test and blanked off,
course, the leak rate. volume loss figures. The calculation is
but then the sleeves and suits are the
This method is used successfully by derived from the universal gas law and,
parts most prone to develop leaks. The
at least one isolator manufacturer but in in the easy-to-do form, the leak rate
practical solution for sleeves is to evert
practice it is really only suitable for large equation is as follows:
them, i.e. allow the internal pressure of
and / or relatively leaky isolators, such
Equation 1 L = PD X 6,000 the isolator to push them completely out
as industrial scale filling lines. With
SP X M of the isolator in the case of positive
smaller isolators, it becomes hard to set
pressure isolators or to pull them into the
an air flow rate which holds a steady
Where: isolator in the case of negative isolators.
pressure. The isolator tends to either rise
L = Leak rate in % volume loss per hour In this condition they act as pressure
or fall in pressure with small air pump
PD = Pressure decay in Pascals compensators and hold up the isolator
adjustments. However, if you have a
SP = 101,325 + starting pressure in Pascals pressure, but at least they present a
fairly large isolator, say larger than 3
M = Test time in minutes known and constant error; which is not
cubic metres, give it a go. At least the
ideal, but is at least practical.
maths is easy!
For half-suits, the best plan is to
Or if you want to know the decay
F 1LWURJHQ'LOXWLRQ agitate the suit gently while the isolator
allowed for a given leak rate, the leak
This is a sophisticated test and, whilst is above test pressure until the suit
rate equation looks like this:
the results may be excellent, the settles in a stable position.
equipment required makes the method Equation 2 PD = L X SP X M The pressure decay test procedure
rarely used in the pharmaceutical 6,000 overall then consists of the following
industry. The test involves filling the stages:
isolator with nitrogen and holding it at a As an example for Equation 1, suppose a. Decide what class of leak rate your
negative pressure. A sensitive oxygen that an isolator is tested starting at isolator is required to achieve, on the
meter inside the isolator then follows 200 Pa for 5 minutes and the resulting basis of the design of the isolator, the
the increase in oxygen concentration as pressure decay is 25 Pa. The equation process in the isolator, the transfer
the isolator leaks air inwards. A plot of tells us that the leak rate was 0.30% methods used and the room housing
the oxygen concentration change gives volume loss per hour, just short of the isolator
the leak rate. The test is sensitive but not achieving Class 2 (0.25%).
b. Decide what test pressure to apply.
very practical and there may be some As an example for Equation 2, if you
safety issues with handling nitrogen in wish to test for Class 2 leak rate, you might c. Decide how long the test should be
this way as there is a risk of asphyxiation. choose a starting pressure of 150 Pa and a and specify the maximum allowable
(This is because there is no pre-indication test time of 10 minutes. Equation 2 then pressure decay in that time.

www.cleanairandcontainment.com Clean Air and Containment Review | Issue 11 | July 2012 9


Main feature

d. Alternatively decide a specific pressure has already been suggested and a. The testing is only to meet Class 2
pressure decay magnitude and if an aseptic isolator is running at 50 Pa, or Class 3 leak rates.
specify the minimum time for that then we have a comfortable test pressure
b. The test is short, not more than, say,
decay to occur. It should be noted of 150 Pa. As regards test time, we
10 minutes.
that in the case of an isolator with probably want a quick test, but we
good leaktightness, this method also want enough time to generate a c. The isolator is in thermal equilibrium
could take a very long time! repeatable result. 1 minute is a short test, with the surroundings which
20 minutes is perhaps a long test, so 10 themselves are stable.
e. Ensure that the isolator is in stable
minutes may be a rational compromise.
thermal conditions, which means it d. The outside weather is not changing
What decay is allowed under this
should be turned off and equilibrated quickly – no violent storms are passing
test? Using Equation 2, for Class 2 we
at room temperature. by – and there are no changes within
get an allowable drop of up to 42 Pa.
the building due, for example, to doors
f. Seal up all the ports, doors and valves. However if we apply the same equation
opening or HVAC systems readjusting.
for Class 3, we get a drop of 169 Pa which
g. Attach a suitable fan and valve, and a
is clearly unusable. We therefore need to
calibrated micromanometer.
shorten the test to 5 minutes, giving an Testing to Class 1 is not really
h. Raise the internal pressure to around allowable drop of 84 Pa. relevant to pharmaceutical isolators but,
the test pressure and evert the sleeves Apart from problems with the for installations where it is has to be
and gloves, or shake down the half-suit. movement of sleeves and suits during applied, more information on the
the decay test, we still have the necessary corrections is given in the
i. Raise the internal pressure to above
universal gas law to contend with. If the Appendix at the end of this paper.
the test pressure and allow the
air temperature inside the isolator The more pedantic of critics will
pressure to decay back down to the
changes during the test, then the point out that the pressure decay test
test pressure. Note the time.
pressure inside changes. Likewise if the does not measure leak rate at a constant
j. Run the time duration of the test and ambient atmospheric pressure changes pressure, and that actual leakage will be
note the final pressure (or, if alternative during the test, then the differential more at high pressure, reducing steadily
d). is being used, allow the prescribed pressure changes. Is this significant? as the pressure decays. As mentioned,
pressure decay to occur and note the The practical answer is a qualified the pressure hold test overcomes this
final time). “no”, provided that the following are true: issue, however the decay test is still

k. Calculate the leak rate using


Equation 1 above.
7DEOH/HDN5DWHV&RPSDUHG
l. If the test fails to meet the chosen
,62&ODVV YROORVV 'HFD\UDWH 'HFD\UDWH
leak rate, repeat two or three more
SHUKRXU 3DSHUKRXU 3DSHUPLQXWH
times to obtain consistent results.
- 0.03 30 0.51
m. If the test still fails, proceed to leak
1 0.05 51 0.85
detection – see later.
- 0.1 101 1.7
2 0.25 254 4.2
A negative pressure test may be
applied to negative pressure isolators, - 0.5 507 8.5
but a major concern with negative 3 1.0 1,015 17
isolators, which are normally specified 4 10.0 10,150 169
when containment is important, is that
containment should be maintained if 7DEOH0D[LPXP7LPH$OORZDEOHIRUD3D'HFD\7HVW
there is a pressure reversal and the
isolator goes positive. This may occur ,62&ODVV 'HFD\ 3D 0D[LPXP7LPH 0LQV
due to due to exhaust fan failure, room 1 42 50
pressure faults or even excessively rapid
2 42 25
in and out glove movements. For this
reason, some users will test negative 3 42 2.5
isolators at positive pressure, arguing 4 42 0.25
that this represents the worst case
scenario for leakage. 7DEOH0D[LPXP'HFD\$OORZDEOHIRUDPLQXWH7HVW

,62&ODVV 7LPH 0LQV 'HFD\ 3D


Practical values and further
considerations for leak rate 1 10 8.4
measurement 2 10 42
So what test pressures and test times 3 10 169
might be chosen by the isolator operator
4 10 1,690
or test engineer? Three times working

10 Clean Air and Containment Review | Issue 11 | July 2012 www.cleanairandcontainment.com


Main feature

valid if the test pressure is above raised to perhaps 100 Pa pressure and a. FAT (Factory Acceptance Test)
working pressure. A successful result proprietary bromophenol cloth is
b. SAT (Site Acceptance Test)
will ensure that the leak rate at working applied to the seals and joints of the
pressure is better than the specified isolator. Where ammonia escapes, the c. OQ (Operational Qualification)
class. Most operators will simply accept cloth turns from yellow to blue. At the
d. PPM
this caveat for the decay test. end of the test the wipers are placed in
(Planned Preventative Maintenance)
Tables 1, 2 and 3 may be useful in the plastic bag and the isolator is
comparing leak rate expressions. ventilated, ideally to atmosphere. e. Following any maintenance or
Ammonia is of course pungent, but the repair which has breached the
Leak detection methods quantities used are small and the real sealed structure
If the leak rate test shows that the hazards are small. This method is
isolator does not meet the specified highly recommended to isolator users.
class, then detection is going to be The distributed leak test
d. Soap solution and ultrasonic
needed to track down the source, or Pharmaceutical Isolators mentions a
detection. The soap solution test is
sources of the problem. All leak ‘distributed leak test’. This is really
exactly the same as is applied to
detection methods are time-consuming. only a leak detection and repair
pneumatic tyres. The method is
a. Helium is often the detection method exercise carried out immediately prior
cheap, sensitive but messy and
of choice. In this case the sealed to an overall leak rate measurement
laborious. Ultrasonic detection may
isolator may be carefully brought to test. The objective is to set a baseline
detect large leaks from isolators under
100 Pa or 200 Pa pressure with helium for subsequent leak rate measurement
high test pressure but is generally not
from a cylinder. A helium detector or tests. Whilst there might be benefits
a useful test for isolator leaks.
“sniffer” is then used to search for in carrying out this exercise, it is
points of leakage. The more expensive e. Progressive elimination. When all recognised that leak detection tests
types of helium detector, not else fails, progressively sealing off the may not detect all leaks due to
surprisingly, are more sensitive. The various sections of the isolator, where inaccessibility. Therefore the objective
advantage of helium detection is that practical, may help to pin-point leaks. of setting a baseline might be unrealistic.
the gas is inert; however, the results
are patchy and tend to be not easily Induction leakage
repeatable. Even so, most isolator Component leak testing Pharmaceutical Isolators also mentions
installations will have access to Gloves, sleeves and half-suits are the induction leakage. This may occur
helium leak detection. most likely source of leaks in a well- where local air velocities within the
built isolator since they are subject to isolator are such that the pressure
b. DOP (Dispersed Oil Particulates or
much wear-and-tear. They are within these flows is reduced by the
“smoke”) can be used for leak
essentially consumable items. They can Bernoulli effect. Thus a positive
detection. A hot or cold smoke
be leak tested as part of the complete pressure isolator could locally generate
generator is used to put smoke into
isolator, or they can be tested separately. a negative pressure. Such an effect
the isolator which is then sealed and
Many isolator manufacturers offer could be significant when operators are
taken to a test pressure. Either a
test devices which close off the sleeves leaving the sleeves, gauntlets or
particle counter or a photometer can
or gauntlets of the isolator and then half-suit. Generally-speaking, this
be used then to scan the structure and
apply a manual or an automatic leak test theoretical leakage is not an issue.
locate leaks. In the case of negative
by pressure decay. Similar devices may
pressure isolators the detectors may
be available for half-suits. Training
be inside the isolator which is at test
One manufacturer offers a device Leak rate measurement is a relatively
depression, and smoke is applied to
which leak tests sleeves and gloves, or simple process but even so, some
likely leak points around the outside
gauntlets, in situ, using nitrogen training for test engineers will give
of the isolator. Two disadvantages of
dilution, but this is an expensive and quicker and more reliable results.
this test are the potential deposition of
time-consuming method. Training can be obtained from the
oil and the likely activation of smoke
isolator manufacturers, from
alarms if they are not switched off.
Leak test rationales organisations such as the PHSS, and
c. Ammonia provides an excellent When should leak rate measurement from a variety of specialist training
method for leak detection. It is take place? The following occasions companies and consultants.
sensitive, inexpensive and will trace would seem to be logical:
small leaks with absolute repeatability.
To carry out this test, a bottle of around
50 ml of ammonia solution is placed
inside the isolator together with a pad
Ammonia provides an excellent method for leak detection.
of lint-free wipers and a polyethylene It is sensitive, inexpensive and will trace small leaks with
bag. The isolator is sealed up and using
the gloves, the ammonia is poured out
absolute repeatability.
onto the wipers to disperse through
the isolator volume. The isolator is then

www.cleanairandcontainment.com Clean Air and Containment Review | Issue 11 | July 2012 11


Main feature/Vital vocabulary

Appendix A two-place or even a three-place Measurement of very low leak rates


thermometer will be needed inside the by pressure decay, or indeed any other
&RUUHFWLQJWKHUHVXOWVRISUHVVXUH
isolator. A barometer resolving to 1 Pa method, is difficult and the results must be
GHFD\WHVWVIRU&ODVVOHDNUDWH
will be needed outside the isolator. regarded with some degree of scepticism.
It would seem that the only way
In practice, the theoretical correction
to make a pressure decay test of this
for temperature and pressure is difficult. i. BS EN ISO 14644-7:2004, Cleanrooms
sensitivity supportable is to apply the and associated controlled environments,
Measurement of temperature at a single
appropriate corrections to the decay Part 7: Separative devices (clean air hoods,
point in the isolator is by no means gloveboxes, isolators and mini-environments)
figures. The corrections are made to
representative of the entire volume. Local ii. ISO 10648:1994, Containment enclosures,
the pressure reading at the end of the
heating may not be registered by the Part 2: Classification according to leak
test thus: tightness and associated checking methods
thermometer but will increase pressure.
฀ If atmospheric pressure rises, iii. Pharmaceutical Isolators, B.Midcalf,
Measurement of atmospheric pressure to
add 10 Pa to the final pressure W.Mitchell Phillips, J.Neiger, T.Coles,
within a Pascal is not easy since rapid Pharmaceutical Press, 2004
reading for every 0.1 mb rise
fluctuations take place at this level. iv. Full derivation from the universal gas
during the test.
law can be obtained from the author.
฀ If atmospheric pressure falls,
subtract 10 Pa from the final pressure
reading, for every 0.1 mb fall during 7LP&ROHV, BSc, MPhil, has worked in the field of isolator
the test. technology for over twenty years. He was a founding member
of the UK Pharmaceutical Isolator Working Party that
฀ If the internal temperature rises,
produced Pharmaceutical Isolators, Pharmaceutical Press,
subtract 3.5 Pa from the final
2004, and more recently of the PDA committee that has just
pressure reading, for every 0.01°C
produced Technical Report No 51: “Biological Indicators for
rise during the test.
Gas and Vapour Phase Decontamination Processes” [for the
฀ If the internal temperature falls, add validation of isolator sanitisation]. His book “Isolation Technology - a Practical
3.5 Pa to the final pressure reading, Guide”, CRC Press Inc. 2004, is now in its second edition.
for every 0.01°C fall during the test.

Q and R are for…


Quality including the verification of the pre-test identification of hazards and the analysis
(ICH Q9*) The degree to which a set of conditions. and evaluation of risks associated with
inherent properties of a product, system ISO 14644-6:2007: Vocabulary exposure to those hazards.
or process fulfils requirements. Editor’s note: This is the only definition ISPE Glossary of Pharmaceutical and
ISPE Glossary of Pharmaceutical and that I could find and is specific to ISO Biotechnology Terminology (GAMP 5,
Biotechnology Terminology (GAMP 5, 14644-1. My own definition would be A Risk-Based Approach to Compliant GxP
A Risk-Based Approach to Compliant GxP “The qualification process repeated Computerized Systems)
Computerized Systems) at prescribed intervals throughout the
lifetime of an installation”. Rapid transfer port (RTP)
Qualification A transfer device in the form of a
(ICH Q7*) Action of proving and Risk double-door transfer port system used
documenting that equipment or ancillary Combination of the probability of to move items from one isolator to
systems are properly installed, work occurrence of harm and the severity another without contamination entering
correctly, and actually lead to the expected of that harm. or escaping from the system.
results. Qualification is part of validation, ISO 14644-6:2007: Vocabulary Pharmaceutical Isolators,
but the individual qualification steps Pharmaceutical Press, 2004
alone do not constitute process validation.
ISPE Glossary of Pharmaceutical Hazard *ICH is the International Conference on
and Biotechnology Terminology (ISPE Potential source of harm Harmonisation of Technical Requirements
Baseline® Guide: Vol. 6, Biopharmaceutical ISO 14644-6:2007: Vocabulary for Registration of Pharmaceuticals for
Manufacturing Facilities) Human Use. ICH publishes a series of
Quality Guides. The subject of ICH Q7 is
Requalification Risk assessment Good Manufacturing Practice and of ICH
Execution of the test sequence specified (ICH Q9*) A systematic process of Q9 is Quality Risk Management. These are
for the installation to demonstrate organizing information to support a risk available as free PDF downloads on the
compliance with ISO 14644-1 according decision to be made within a risk ICH website: http://www.ich.org
to the classification of the installation, management process. It consists of the

12 Clean Air and Containment Review | Issue 11 | July 2012 www.cleanairandcontainment.com

You might also like