Professional Documents
Culture Documents
Em Technique
Em Technique
Introduction to
Electron Microscopy
Instrumentation
Human eye
100 m
Cells
Infrared 10 m
Red blood cells
copy
1 m Bacteria i cro s
Visible n m
c tr o
Ele
Light microscope 100 nm Mycoplasma
Ultraviolet
Viruses
10 nm
Proteins
x,-rays
1 nm
Amino acids
Rat intestine
Light microscopy Electron microscopy
Microvilli
1 µm
10 µm
Green…F-actin
Yellow…ß-catenin
Orange…Nuclei
Schwarz & Humbel 2007: Methods in Molecular Biology, vol. 369, Electron Microscopy: Methods and Protocols, Second Edition
Edited by: J. Kuo © Humana Press Inc., Totowa, NJ
Why electron microscopy?
Mouse intestine
Actin filaments
Glycocalix
Junction
Microvilli
1 µm
Mouse intestine
Membrane
(lipid bilayer)
Actin filaments
500 nm
Up to macromolecular level
Cytosceleton after
extraction
Light microscopes
Photons
Electron microscope
Electrons
e-
From photons to electrons
e- Similarities to photons:
Optical properties
(Diffraction, chromatic abberation, spherical abberation,
astigmatism etc.)
The higher the energy of the electrons, the lower the wavelength,
the higher the resolution
h
DeBroglie relation: λ = wavelength
mv h = Planck's constant (6.6 X 10-27)
m = mass of the particle
Electron pathes through potential field v = velocity of the particle
1.23
nm V = accelerating voltage
V
0.61
Abbe’s equation d NA n sin
n sin
d = resolution in nm
0.753
Resolution EM: d nm α = half opening angle of objective (in radians)
V V = accelerating voltage
However:
Electromagnetic
lens Objective lens Glass lens
Electromagnetic
lens Projector lens Glass lens
Phosphorescent
screen Eye
Final image
X-ray,
photomultiplier
Specimen
The types of electron microscopes
High vacuum
Filament is heated
Electrons are emitted from the tip
F…Filament
W…Wehnelt electrode
C…Ceramic high voltage insulator
Rb…Autobias resistor
Ie…Electron emission current
Electron source (Electron gun)
Thermionic
Cold field emission
Tungsten LaB6 Schottky
Normalized brightness
High voltage
Electromagnetic lenses
v…speed of electron
B…magnetic field
F…resulting force
Image rotation:
Reasons:
• Contamination of lenses and apertures
• Inhomegenities of the lens
• Charging of specimen
e- (98 kV)
e- (100 kV)
e- (102 kV)
Filament chamber
Ultra high vacuum: < 10-9 mbar
Specimen chamber
High vacuum: ~ 10-7 mbar
Viewing chamber
High vacuum: ~ 10-5 mbar
Vacuum systems
Rotary pump
Rotary pump
Goniometer: x, y, z, r
Specimen size:
• 3 mm in diameter!
• Ca. 100 nm in thickness
(electron transparent)
Specimen holders and stages
Specimen holder
3 mm
Specimen holders and stages
Gun
Objective lens
Specimen stub
Stub holder
Stage
Specimen size:
• 100 mm in diameter
• 2 cm in z-direction (not electron transparent)
Specimen holders and stages
NOTE:
Elastic Inelastic
(higher angle, E=E0) (low angle, E=E0-∆E)
Unscattered
(E=E0)
Electron – specimen interactions
Inelastic scattering:
K 1
L
Emission of electrons and radiation M
N
Electron – specimen interactions
Primary electrons
X-rays
SEM analysis
Cathode luminescense
Heat
Auger electrons
TEM analysis
Inelastically scattered electrons
Elastically scattered electrons
Unscattered electrons
Electron – specimen interactions
TEM REM
Imaging in the transmission electron microscope
Illumination
Condenser lens
Projector lens
Absorption of electrons
Scattering of electrons
Signal
Intensity
Specimen profile
Imaging in the transmission electron microscope
Objective aperture
Signal
Intensity
Specimen profile
Imaging in the transmission electron microscope
Diffracted ray
Specimen
Objective Objective
lens lens
Projective Projective
lenses lenses
Image plain
Signal
Intensity
Specimen profile
Imaging in the transmission electron microscope
Absorption weak
Objective aperture
Signal
Intensity
Specimen profile
Imaging in the transmission electron microscope
1 µm
Imaging in the scanning electron microscope
Imaging in the scanning electron microscope
Illumination
Photomultiplier
Detector • Photomultiplier
• No CCD camera
Lens system
“condenser”
Beam scanner
Objective
Specimen
Specimen: Bulk specimen
Imaging in the scanning electron microscope
…Secondary electrons
The focused electron beam is moved from one pixel to another. At every
pixel, the beam stays for a defined time and generates a signal (e.g.
secondary electrons) which are detected, amplified and displayed on a
computer screen.
Imaging in the scanning electron microscope
Object
768 px 768 px
1024 px 1024 px
Imaging in the scanning electron microscope
R…interaction volume
Imaging in the scanning electron microscope
Contrast based on SE
R dependent on density of material (Z) and acceleration voltage of PE (0.1 - 30 kV)
Energy of SE independent of acceleration voltage of PE
R λ C: 10 – 100 nm
R λ Cr: 2 – 3 nm
λ Pt: 1 – 2 nm
Imaging in the scanning electron microscope
SE detector
(inlens)
SE detector
Φ SE I BSE
BSE
SE II Small excitation volume
High BSE coefficient
SE Escape depth 1-3 nm
500 nm
Freeze-fractured yeast
Imaging in the scanning electron microscope
BUT:
• Useful if specimen is coated with heavy
metals
R BSE vs. SE
• Less sensitive to charging (higher energy)
• Less topographic contrast
• More material contrast
Imaging in the scanning electron microscope
Contrast SE
SE signal at 20 kV SE signal at 1.7 kV
Little topography Good topography
(Signal based on SE II induced by BSE!) (Signal based on SE I from surface layer)