Introduction Into IEC 62304 Software Life Cycle For
Introduction Into IEC 62304 Software Life Cycle For
Christoph Gerber
4. September 2008
SPIQ
9/5/2008 1
Navigation
Agenda
Current Picture
Summary
References
Q&A
Navigation
Current Picture
Navigation
No choice be compliant!
Navigation
Regulatory
Requirements
Navigation
in general
21 CFR PART 11 Electronic Records & Signatures
Navigation
FDA compliance
EN ISO 13485
Medical devices - Quality management systems
IEC 62304
Overview
Navigation
tion
re
E d i
p
rd
3
la
ce
-4
s
1
01-
1-
06
6
IE C6
IEC 62366
affects
Usability
Navigation
RISK MANAGEMENT
QUALITY MANAGEMENT
AL
DIC
SOFTWARE ENGINEERING E R E
M A
O R T W
F OF
Y
T ES
E
F IC
A
S EV
D
Navigation
Purpose This standard defines the life cycle requirements for medical
device software. The set of
– processes,
– activities, and
– tasks
described in this standard establishes a common framework for
medical device software life cycle processes.
Field of This standard applies to:
Application – the development and maintenance of medical device
software,
– to the development and maintenance of medical device software
when software is itself a medical device or when software is an
embedded or integral part of the final medical device,
– does not cover validation and final release of the medical device,
even when the medical device consists entirely of software.
Navigation
IEC 62304
Key Concepts
Navigation
Software Safety
Classification
Navigation
Software safety class for Software System and Software Items according
to the possible effects on the patient, operator, or other people resulting
from a HAZARD to which the SOFTWARE SYSTEM can contribute.
Class A: No injury or damage to health is possible
Class B: Non-SERIOUS INJURY is possible
Class C: Death or SERIOUS INJURY is possible
Navigation
Segregation of critical SW
Class A clear communication interfaces helps
SW
for rationale
Processor 2 cluster critical SW
Class C
SW Middle grained approach for item
Processor 1 to fine gets impractical
to coarse gets in trouble with SOUP
and SW risk management
M1
Software
Risk Management
Navigation
Functional Analysis
Risk = f L, S
(2) Use risk graph as risk evaluation measure
(3) Move risks from red to green area
Frequent
Intolerable
L
I
Region
Probable
Risk = combination of the K
E ALARP
likelihood (L) of occurrence of K Remote
L
harm and the severity (S) of that I Improbable Broadly
harm [ISO 14971] H
O acceptable region
O Incredible
D
Negligible Light Modest Severe
Severity
Electronic
Severity
identify software items
contributing to hazardous
situation
requires knowledge of
architecture Likelihood
User Needs
Software Requirements
Device Risk Software documented in Specification
Management Requirements Analysis (+ Risk Control Measures)
Risk Analysis
1) on Functions
2) on Components
Architectural
Software Risk documented in Design
Architectural Design
Management Description
C
C B
Ongoing Detailed Design
Activity C A
Implementation
Navigation
SOUP
Navigation
Configuration management
add SOUP software to your SCCS
Navigation
IEC 62304
QM System Mapping
Navigation
SW configuration management
SOP SCCS
process
During development SOP Issues Management
SW problem resolution process
After release SOP Change Management / CAPA
Summary
Navigation
Summary
References
IEC 62304:2006 Medical device software -- Software life cycle processes
http://www.iso.org/iso/iso_catalogue/catalogue_tc/catalogue_detail.htm?csnumber=38421
IEC 60601-1-4- Ed. 1.1 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral Standard: Programmable
electrical medical systems; http://www.iec-normen.de/shop/product.php?artikelnr=209369
Richtlinie 93/42/EWG des Rates vom 14. Juni 1993 über Medizinprodukte
http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=CELEX:31993L0042:DE:HTML
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
(May 11, 2005), http://www.fda.gov/cdrh/ode/guidance/337.html