Professional Documents
Culture Documents
IABP
IABP
EQUIPMENT:
Intra-arterial cannula:
Fluid-Filled Tubing:
- Column of non-compressible fluid w no bubbles between arterial line and transducer for
Hydraulic Coupling
- Short, wide and stiff = reduce damping
- Short; <1.2m = reduce damping
o The fluid filled system has a ‘natural frequency’ of resonance
o NF is proportional to the length of tubing
o So longer tubing = lower natural frequency
o Patient’s pulse oscillation is low frequency so with longer tubing the NF approaches
the patient’s pulse frequency
o This will cause resonance and amplify the signal and the system will be under-
damped
- Stiff;
o Soft IV tubing will absorb the wave and damp the system
- Wide;
o Damping increases by the third power of any decrease in the diameter of tubing
o Ie narrow tubing increases damping
3 way tap:
Pressure Transducer:
Flush System:
- Electrical signal derived from pressure transducer is transmitted via cable to microprocessor
for filtering, amplification, analysis and display on a screen as Pressure v Time wave
MoA
PHYSICS
SOURCES OF ERROR
ADV
- Continuous data
- Waveform analysis (ie contractility)
- Direct measurement of all blood pressures (SBP, MAP, DBP), HR
- Samples blood for ABG analysis and reduces venipuncture requirements
DISADV
- Clotting and backflow through cannula and line set if we lose pressurization
- Arterial damage, or in fact any tissue/nerve damage around the site of insertion
- Infection
- Ischaemia
- Inadvertant drug injection into artery
- Requires periodic check and zeroing to confirm accuracy
- Bleeding from inadvertent disconnection
- Waveform needs to be displayed to detect resonance and damping
- Requires calibration
CALIBRATION
Static
Calibrating (‘zeroing’)
Zeroing – the use of atmospheric pressure as a reference standard against which all other
pressures are measured
ensure the transducer pressure tubing and flush solution are correctly assembled and free of
air bubbles
place transducer at level of the right atrium
‘off to patient, open to air (atmosphere)’
press ‘zero’ -> sets atmospheric pressure as zero reference point
whenever patient position is altered the transducer height should be altered
even though atmospheric pressure is constant
the system is closed the zero reference point & cannula tip need to be exactly the same
level to measure the pressure in the artery
The Zero Reference Point is the level of the top of the fluid column (air fluid interface)
above/below the diaphragm when the tap is open to air
The transducer is measuring the mean pressure in the artery (the height of the fluid column
above the zero point (which is 0mmHg when they were zeroed at the same level)
If patient’s table is elecated above this level 20cm, the pressure recorded will be arterial
pressure + 20cm H2O due to the vertical height of the fluid column (divide by 1.36 to convert
cmH2O to mmHg)
Dynamic
P = Force/Area