Professional Documents
Culture Documents
Bioinstrumentation I
Chapter 7 –Blood Pressure and sound
نعمان النجار.إعداد د
Dr.Eng. Noman AL Najjar- PhD in biomedical engineering
Department of Biomedical Engineering UST-Sana’a
noman_qaed@yahoo.com
Dr.Noman
Dr.Noman 2
in
John G. Webster (Editor)
Medical Instrumentation: Application and Design
John Wiley & Sons, 1998
ISBN 0-471-15368-0
Blood Pressure and sound
3 Outline
Concept of pressure
Measurement of Pressure
Blood Pressure around the Heart
Classification of blood pressure
Direct blood pressure measurement
Blood Pressure Transducers
Strain gage w/ Wheatstone Bridge
Intravascular Sensors
Fiber O ptics based Pressure Sensor
Dynamic Properties of the Measurement System
Bandwidth Requirements
Heart Sounds
Indirect blood pressure measurement
AuscultatoryMethod
Ultrasonic determination of blood pressure
Oscillometric Automated BP Monitors
Practical part Dr.Noman
Blood Pressure and sound
4
What is pressure?
P=F/A
-Measured in Pascal
-1 Pa = 1 Newton/ 1 m2
Dr.Noman
Blood Pressure and sound
5
Measurement of Pressure
1 Atmosphere = force needed to raise the mercury
in column about 760 mm high
-Blood pressure is measured in terms of mm Hg
Why we measure pressure?
• Pressure is the driving force of the dynamics of
the human body
• Pressure is generated by the heart (constant
volume or a constant pressure pump?)
• Changes of the pressure inside heart chambers
cause acceleration and deceleration of the
blood that generate sounds.
Dr.Noman
Blood Pressure and sound
6
Dr.Noman
Blood Pressure and sound-Blood Pressure around the Heart
7
Classification of blood pressure levels for adults 18 years and older. Systolic
pressure, the higher number of a blood pressure reading, is the pressure as the
heart pumps; diastolic pressure is the pressure when the heart relaxes between
beats
CATEGORY SYSTOLIC (mm Hg) DIASTOLIC (mm Hg)
Normal* <130 < 85
High normal 130-139 85-89
Hypertension
STAGE 1 (Mild) 140-159 90-99
STAGE 2 (Moderate) 160-179 100-109
STAGE 3 (Severe) 180-209 110-119
STAGE 4 (Very Severe) >209 >119
Optimal blood pressure is less than 120/80 mm Hg. Unusually low readings should
be evaluated by a physician.
Dr.Noman
Blood Pressure and sound- Classification of blood pressure
9
Sensing
port
Sample and transducer Roller clamp
zero stopcock
Electrical connector
Disposable pressure transducer with an integral flush device
Dr.Noman
Blood Pressure and sound- Blood Pressure Transducers
12
Dr.Noman
Blood Pressure and sound- Blood Pressure Transducers
13
Direct measurements
• Intravascular sensors:
– Sensors a tip of catheter
– Has higher frequency response and less time delay
– More expensive, may break after a few uses
• Sensors used can be:
– Strain gage, piezo-electric crystals, LVDT,
variable inductance, variable capacitance, opteoelectronics,
and semiconductor devices
Dr.Noman
Blood Pressure and sound- Strain gage w/ Wheatstone Bridge
14
Disadvantage of the previous system: The frequency response of the system is limited
by the hydraulic properties, in particular the low pass filter effect of the tubing system
Using an intravascular system eliminates the entire plumming system, by making the
measurement at the site!
•Eliminates the time delay introduced by the tubing system
•Allows high fidelity measurement of the high frequency components of the BP
signal
Typical sensors used:
•Strain gages bonded onto a flexible diaphragm at the catheter tip
•Fiber optic systems where the displacement measurement of the diaphragm is
made optically
Dr.Noman
Blood Pressure and sound- Fiber Optics based Pressure Sensor
16
Figure 7.4 (a) Schematic diagram of an intravascular fiber-optic pressure sensor. Pressure causes deflection in a thin
metal membrane that modulates the coupling between the source and detector fibers. (b) Characteristic curve for the
fiber-optic pressure sensor. Dr.Noman
Blood Pressure and sound- Fiber Optics based Pressure Sensor
17
Figure 7.5 Fiber-optic pressure sensor for intracranial pressure measurements in the
newborn. The sensor membrane is placed in contact with the anterior fontanel of the
newborn.
Dr.Noman
Blood Pressure and sound- Dynamic Properties of the Measurement System
18
Sensor
Errors in measurement of
dynamic pressure can lead to
(a) P Diaphragm
significant consequences, Liquid Catheter
Incremental
underdamped system - D
length
V
Overestimated BP.
The liquid filled catheter-sensor
Rc Lc Rc Lc Rc Lc Rs Ls
is a hydraulic system that can be (b)
Distortion will be seen on all signals acquired by a system that does not possess
the appropriate frequency response.
Dr.Noman
Blood Pressure and sound- Bandwidth Requirements
23
Dr.Noman
Blood Pressure and sound- Bandwidth Requirements
25
Dr.Noman
Blood Pressure and sound- Bandwidth Requirements
26
Dr.Noman
Blood Pressure and sound- Heart Sounds
28
Dr.Noman
Blood Pressure and sound- Heart Sounds
29
Figure 7.15 Correlation of the four heart sounds with electric and
mechanical events of the cardiac cycle.
Dr.Noman
Blood Pressure and sound- Heart Sounds
30
S2: Closure of semilunar valves causing the deceleration and reversal of flow
in the aorta and pulmonary artery
Dr.Noman
Blood Pressure and sound- Heart Sounds
31
(From A. C Burton, Physiology and Biophysics of the Circulation, 2nd ed. Copyright © 1972 by
Year Book Medical Publishers, Inc., Chicago. Used by permission.)
Dr.Noman
Blood Pressure and sound- Indirect Measurements AuscultatoryMethod
32
Figure 7.20 Typical indirect blood-pressure measurement system The sphygmomanometer cuff is inflated by a hand
bulb to pressures above the systolic level. Pressure is then slowly released, and blood flow under the cuff is monitored
by a microphone or stethoscope placed over a downstream artery. The first Korotkoff sound detected indicates systolic
pressure, whereas the transition from muffling to silence brackets diastolic pressure .
Dr.Noman
(From R. F. Rushmer, Cardiovascular Dynamics, 3rd ed., 1970. Philadelphia: W. B. Saunders Co. Used with permission.)
Blood Pressure and sound-Indirect Measurements Ultrasonic determination of
33
BP
Figure 7.21 Ultrasonic determination of
blood pressure A compression cuff is
placed over the transmitting (8 MHz) and
receiving (8 MHz ±D ƒ) crystals. The
opening and closing of the blood vessel are
detected as the applied cuff pressure is
varied.
Cuff pressure mm Hg
34
1
Oscillometric 200
MAP = DP + (1/3)(SP – 160 2
DP)
120
• Start and end of
oscillation indicates 80
systolic and diastolic 40
pressures 0
Dr.Noman
IBP ( Invasive BP ) measurement
37
diaphragm
Blood catheter
Pi cable
vessel Po
pill up with
some liquid strain gages
(ex.saline)
Pi
t Dr.Noman
IBP ( Invasive BP ) measurement
38
V L
Electrical resistance : V Ri , R
A
resistivity
P1 F ( P1 P2)
R
P2 Liquid resistance : op R F
L
R
A
Dr.Noman viscosity
IBP ( Invasive BP ) measurement
39
Dr.Noman
IBP ( Invasive BP ) measurement
40
multi
physics Electric Circuit Fluid Mechanics
voltage pressure
current flow
charge volume
R
V L
[ ] ( )
Pa s 8L
R [ 3 ] ( )
I A F m r 4
L
V P L
dI L ( 2 )
dt dF r
dt
I C Young' s modulus
C
dV
dt
Dr.Noman
Equivalent Circuit Model of IBP
41
Rc Lc
Po diaphragm
catheter cable +
Pi Vi i Vo Cd
-
liquid strain gages Compliance of
diaphragm
di dV0
Vi Rc i Lc V0 i Cd
dt dt
dV0 d 2V0
Vi V0 RcCd Lc Cd 2
: 2nd order ODE
dt dt
D d
Rc Cd
D : operator K 1,
dt 2 Lc
D 2 2D 1
damping
[ 2 1] Vo (t ) KVi (t ) Wn
ratio
Wn Wn Lc Cd
natural
Dr.Noman
frequency
Equivalent Circuit Model of IBP
42
Wn Wn Wn
1 2
tan 1 ( )
w w w Wn
[1 ( ) 2 ]2 4 2 ( ) 2
Wn Wn Wn w
Dr.Noman
Equivalent Circuit Model of IBP
43
1 (critically damped)
2 (overdamped)
Wn w
∠H w
0.5 (underdamped)
-π
Dr.Noman
Equivalent Circuit Model of IBP
44
|H| ∠H f1 f2
K -4/π
0.5
-1.8π
f1 f2 f
Vi (t ) A sin(2f1t ) Vi (t ) A sin(2f 2t )
Vo (t ) 0.5 A sin(2f 2t 1.8 )
Vo (t ) KA sin(2f1t )
4
Dr.Noman
Unit Step Response
46
underdamping
input signal
overdamping
critical damping
Dr.Noman
Transient Step Response
47
Po
balloon
P
bulb
saline
overdamping
critical damping
underdamping
Dr.Noman
Example
48
r 1 DP 12
fn ( ) 91Hz
2 L DL
4 L DV 12
3( ) 0.033
r DP
f n,bub 22 Hz
bub 0.137
1.34 1.95 log f
Dr.Noman
Example
49
(7.2) By changing only the radius of the catheter, redesign the (no-bubble)
catheter of Figure 7.9 to achieve the damping ratio ζ=1.
Calculate the resulting natiral frequency fn.
r 3 r0 0
3
r 3 0.0032
r 0.147
f n f n 0 r r0
f n 29 Hz 1.46 log f
Dr.Noman
50
Thank you.
Dr.Noman
Dr.Noman 51
References
• main text book “)Medical Instrumentation( Webster”
• Dr Baha and Dr Haitham’s class notes at KAU BME
•Lectures of Dr Noman AL Najjar at UST –BME
•http//engineering.roman.edu
YOUNHO HONG