Professional Documents
Culture Documents
Although the correction of abnormal arterial blood oxygen ventilation.2 The increasing range of measurements
and carbon dioxide (CO2) tensions is the main goal of available during mechanical ventilation should assist the
mechanical ventilation, it is of note that an intervention clinician in the optimal setting of the ventilator.
which improves gas exchange does not necessarily improve
patient outcome. Nitric oxide inhalation has been shown
to improve oxygenation but not outcome.11 Similarly, Static measurements
prone ventilation is associated with a significant improve-
ment in gas exchange but in a large multicentre randomized, Compliance and resistance
controlled trial there was no effect upon patient mortality.9 The basic mechanical properties of compliance and
Interventions that improve gas exchange have even been resistance are readily measured during volume control ven-
associated with an adverse effect upon outcome. In the tilation with constant flow inspiration (Fig. 1).10 There must
ARDSnet trial which compared high and low tidal volumes be an adequate end inspiratory pause in order that the
in acute lung injury, gas exchange as reflected by the inspiratory plateau pressure can be accurately measured
PaO2 :F IO2 ratio was significantly higher in the high tidal and there should be no leak in the respiratory circuit. If
volume group up to day 7 of the trial.3 However, despite intrinsic PEEP is present, the compliance value will be
causing improved gas exchange, high tidal volume ventila- underestimated if the end-expiratory pressure measured in
tion was associated with an increased absolute mortality of the airway is used. Total PEEP should be measured after
8.8% compared with using low tidal volumes. This land- an end-expiratory pause (see below) as this reflects the true
mark study confirmed that how the ventilator is set has a end-expiratory pressure within the lungs.
significant influence upon mortality and that an improve- A static compliance (Crs.st) less than 50 ml per cm H2O is
ment in blood gas tensions does not imply that the patient is commonly encountered in ICU and patients with severe
receiving optimal ventilatory support. acute respiratory distress syndrome (ARDS) may have
The ARDSnet tidal volume trial confirmed the large body values less than 20 ml per cm H2O.10 Monitoring compli-
of evidence from laboratory studies that positive pressure ance may be useful for assessing the effectiveness of lung
ventilation can itself cause and exacerbate lung injury.38 The recruitment or the occurrence of lung over-distension after
application of excessive tidal volumes that result in adjustments to ventilator settings. The change in compliance
hyperinflation (‘volutrauma’) appears to be the most impor- that occurs after the application of PEEP may differentiate
tant cause of ventilator-induced lung injury (VILI). Cyclical between lung recruitment or hyperinflation.32
opening and closing of atelectatic lung regions during tidal When interpreting the measured values of Crs.st, the
breathing (‘atelectrauma’) appears to be the other major influence of the chest wall compliance (CW) should be
contributory factor. Lung protective ventilation has been considered. Chest wall abnormalities, increased muscle
described as ensuring maximal recruitment of collapsed tone and abdominal distension may all reduce chest wall
areas of the lung, avoiding regional hyperinflation and compliance. In order to measure the two components of
preventing cyclical recruitment–derecruitment during tidal compliance (CL and CW), an oesophageal balloon is used
The Board of Management and Trustees of the British Journal of Anaesthesia 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org
Macnaughton
60
. Table 1 Airway resistance (cm H2O litre1 s) in a range of different clinical
Flow, V conditions [mean (SD)]. *From Ref. 43
–1
litre s
I time Normal spontaneous breathing 2.5
0
Normal ventilated 4.2 (1.6)
Cardiogenic pulmonary oedema* 12.1 (5.5)
ARDS* 15.5 (4.6)
Chronic airflow limitation* 26.4 (13.4)
40
Volume (ml)
Fig 1 Pressure and flow curves during constant flow ventilation: tidal 600
volume (Vt)=V_ ·I time (constant flow inflation); total respiratory system
static compliance (Crs,st)=Vt/(P2Pexp); total respiratory system
400
dynamic compliance (Crs,dyn)=Vt/(P1Pexp); airway resistance min
(Raw)=(PpeakP1)/V_ ; airway resistance max (Raw)=(PpeakP2)/V_ . Lower inflection point, LIP
200
58
Ventilators for lung performance reporting
whether the application of PEEP will be effective in pro- frequency of 20 bpm. Using the least squares method the
ducing recruitment.23 Furthermore, the process of under- values of best fit are computed for airway resistance, respira-
taking a static PV curve may induce changes that are not tory system compliance and total PEEP.28 The advantage
a true reflection of what happens to the lungs during tidal of this method is that the lung mechanical parameters can be
breathing. Reducing the PEEP level to zero and allowing a measured during any ventilator mode with any inspiratory
prolonged expiration may cause significant de-recruitment flow pattern provided that the patient is relaxed.13 There is
that would not be present during ventilation with a normal no requirement for an end inspiratory or expiratory pause
ventilatory frequency.35 The PV curve may also be influ- although it will be inaccurate if the patient is making
enced by changes in chest wall compliance caused by spontaneous respiratory efforts.33 Intrinsic PEEP can be
abdominal distension in extra pulmonary ARDS.24 It estimated from the difference between the calculated total
appears that ventilation during tidal breathing follows the PEEP and the applied PEEP. However, this measurement
deflation limb of the PV curve and a point of maximum does not appear particularly accurate in patients with
curvature has been described which equates to the onset of significant airflow limitation42 when static measurements
de-recruitment which may be a more appropriate pressure of intrinsic PEEP are most useful clinically. If data
to guide the setting of PEEP.12 The ability to undertake the collected from the whole respiratory cycle are used in the
measurement of PV curves in proprietary ICU ventilators calculation, the assumption is that resistance and compli-
will allow the role of this measurement in clinical practice ance remain constant. This may be a significant source of
to be established. error as resistance tends to decrease with increasing lung
59
Macnaughton
Tracheal PV loops and dynostatic algorithm during dynamic ventilation. With this analysis, an estimated
Some ventilators offer the facility to measure tracheal pres- ‘static’ alveolar PV plot can be produced during any mode
sure directly through a tracheal catheter introduced through of ventilation and without the need for an end inspiratory
the tracheal tube. Measuring tracheal pressure removes the pause (Fig. 3). The dynostatic algorithm PV curve typically
influence of the resistance of the tracheal tube and signifi- shows much greater changes in volume-dependent com-
cantly changes the appearance of the dynamic PV loop.34 A pliance than if compliance is calculated conventionally
more accurate estimate of end inspiratory and expiratory and can be used to indicate over-distension during tidal
pressures is obtained and assuming that the airway resis- ventilation.
tance is low, tracheal pressure will be a reasonable reflection
of alveolar pressure. This results in a valid estimate of
compliance measured under dynamic conditions. Intrinsic PEEP and gas trapping
The difference between tracheal and alveolar pressure
which arises from airway resistance can be estimated Minimizing intrinsic (PEEPi) and the associated hyperinfla-
from analysis of tracheal pressure and flow data collected tion are central to the ventilatory strategy for patients with
at matching lung volumes during inspiration and expira- severe airflow limitation attributable to COPD or asthma.27
tion.16 If inspiratory and expiratory resistance are assumed The adverse effects of PEEPi include increased inspiratory
to be identical at the same lung volumes, alveolar pressure work of breathing during spontaneous ventilation, reduced
can be predicted from the equation of motion and has been ability to trigger the ventilator during assisted modes of
600
exp insp
d
te
200 LIP
la
cu
al
C
100 t
d par
polate
Extra
0
0 5 Pexp 10 15 20 Pinsp 25 30
Pressure (cm H2O)
ƒ ƒ ƒ ƒ
Fig 3 Dynostatic alveolar PV curve. Alveolar pressure is calculated according to the equation: Pdynostatic=(Pexp·V inspPinsp·V exp)/(V inspV exp).
Every point on the dynostatic curve (Pxdyn) is calculated using pressure and flow values at isovolume levels (indicated by broken lines) during
inspiration and expiration in the tracheal P/V loop. Reproduced from Karason and colleagues, with permission from Blackwell Publishing.44
60
Ventilators for lung performance reporting
obtained by recording the pressure change that occurs before Measurements of dead space have been shown to have an
inspiratory flow commences during inspiration. If the increasing number of uses in the ICU. In ARDS an increased
patient is not making any spontaneous efforts this can be physiological dead space has been reported to be an inde-
measured from the change in airway pressure. However, pendent predictor of mortality.26 The mechanism for this
when the patient is spontaneously breathing, oesophageal association is unclear and may reflect that a high dead space
pressure needs to be recorded (as an estimate of pleural is a marker of more extensive endothelial and vascular
pressure) and dynamic PEEPi is extrapolated from the injury. However, increased intra-pulmonary shunt influ-
change in oesophageal pressure which occurs before inspira- ences the measurement of the Bohr–Enghoff dead space
tory flow commences. Dynamic measurements reflect the as the shunted pulmonary venous blood which carries
lowest regional value of PEEPi and may be considerably more CO2 is mixed with the arterial blood causing an
less than static measurements in patients with airflow lim- increased arterial Pco2. The association of dead space
itation who typically have significant inhomogeneity.22 and mortality in ARDS may be explained by an increased
Measurements of PEEPi should be made without any apparent physiological dead space because of a more severe
external PEEP applied. The value of static PEEPi observed shunt.8
in patients with chronic airflow limitation undergoing Physiological dead space measurements may be helpful
mechanical ventilation depends upon the airway resistance in the optimal application of PEEP. Lung recruitment has
and on the applied ventilatory parameters (tidal volume, been associated with a reduction in dead space,39 which
expiratory time and ventilatory frequency). Values up to may reflect a reduction in intra-pulmonary shunt. The appli-
61
Macnaughton
monitor lung performance. Adaptive support ventilation 2 Amato MBP, Barbas CSV, Medeiros DM, et al. Effect of a protec-
(Hamilton Medical, Bonaduz, Switzerland) is a fully tive ventilation strategy on mortality in the acute respiratory
distress syndrome. N Engl J Med 1998; 338: 347–54
automated mode which ensures that the desired minute
3 ARDS Network. Ventilation with lower tidal volumes as com-
volume set by the clinician is delivered either as manda- pared to traditional tidal volumes in acute lung injury and ARDS-
tory pressure-controlled breaths or spontaneous pressure- New Engl J Med 2000; 3421301–8
supported breaths according to the patient’s underlying 4 Arnold JH, Thompson JE, Arnold L. Single breath CO2 analysis:
respiratory effort. From continuous monitoring of dynamic description and validation of a method. Crit Care Med 1996; 24:
lung mechanics by the least squares method, the ventilator 96–102
adjusts the inspiratory pressures automatically in order to 5 Blanch L, Bernabe F, Lucangelo U. Measurement of air trapping,
intrinsic positive end-expiratory pressure, and dynamic hyper-
obtain the optimal combination of ventilatory frequency and
inflation in mechanically ventilated patients. Respir Care 2005;
tidal volume which achieves the target minute ventilation. 50: 110–23
The algorithm includes a lung protective strategy, which 6 Bouadma L, Lellouche F, Cabello B, et al. Computer-driven man-
prevents high tidal volumes and airway pressures above agement of prolonged mechanical ventilation and weaning: a pilot
35 cm H2O. The expiratory time constant (RCe) is moni- study. Intensive Care Med 2005; 31: 1446–50
tored and used to adjust the time for expiration in order to 7 Bugedo G, Bruhn A, Hernández G, et al. Lung computed tomo-
minimize the risk of PEEPi. The only parameters set by the graphy during a lung recruitment maneuver in patients with
acute lung injury. Intensive Care Med 2003; 29: 218–25
clinician during ASV are the minute ventilation, F IO2 and
8 Drummond GB, Fletcher R. Deadspace: invasive or not? Br J
PEEP. Minute ventilation is entered as the percentage
62
Ventilators for lung performance reporting
21 Marini JJ, Pepe PE. Occult positive end-expiratory pressure in 33 Ruiz-Ferrón F, Rucabado AL, Ruiz NS, et al. Results of respiratory
mechanically ventilated patients with airflow obstruction—the mechanics analysis in the critically ill depend on the method
auto-peep effect. Am Rev Respir Dis 1982; 126: 166–170 employed. Intensive Care Med 2001; 27: 1487–95
22 Maltais F, Reissmann H, Navalesi P, et al. Comparison of static and 34 Servillo G, Svantesson C, Beydon L, et al. Pressure volume curves
dynamic measurements of intrinsic PEEP in mechanically in acute respiratory failure: automated low flow inflation versus
ventilated patients. Am J Respir Crit Care Med 1994; 150: occlusion. Am J Resp Crit Care Med 1997; 155: 1629–36
1318–24 35 Stenqvist O. Practical assessment of respiratory mechanics.
23 Mergoni M, Volpi A, Bricchi C, Rossi A. Lower inflection point Br J Anaesth 2003; 91: 92–105
and recruitment with PEEP in ventilated patients with acute 36 Sulzer CF, Chioléro R, Chassot PG, et al. Adaptive support
respiratory failure. J Appl Physiol 2001; 91: 441–50 ventilation for fast tracheal extubation after cardiac surgery.
24 Mergoni M, Martelli A, Volpi A, Primavera S, Zuccoli P, Rossi A. Anesthesiology 2001; 95: 1339–45
Impact of positive end-expiratory pressure on chest wall and 37 Tassaux D, Dalmas E, Gratadour P, Jolliet P. Patient–ventilator
lung pressure–volume curve in acute respiratory failure. interactions during partial ventilatory support: a preliminary
Am J Respir Crit Care Med 1997; 156: 846–54 study comparing the effects of adaptive support ventilation
25 Mols G, Brandes I, Kessler V, et al. Volume-dependent compliance with synchronized intermittent mandatory ventilation plus
in ARDS: proposal of a new diagnostic concept. Intensive Care inspiratory pressure support. Crit Care Med 2002; 30: 801–7
Med; 251084–91 38 Tremblay LN, Slutsky AS. Ventilator-induced lung injury: from
26 Nuckton TJ, Alonso JA, Kallet RH, et al. Pulmonary dead-space the bench to the bedside. Intensive Care Med 2006; 32: 24–33
fraction as a risk factor for death in the acute respiratory distress 39 Tusman G, Böhm S, Suarez-Sipmann F, Turchetto E. Alveolar
syndrome. N Engl J Med 2002; 346: 1281–6 recruitment improves ventilatory efficiency of the lungs during
63