Professional Documents
Culture Documents
net/publication/15700672
CITATIONS READS
2 761
2 authors:
Some of the authors of this publication are also working on these related projects:
All content following this page was uploaded by Peter H Pan on 13 January 2015.
Positive end-expiratory
pressure and lung
co .mpliance: effect on
Peter H. Pan MSEEMD~Jan J. van der Aa PhD delivered tidal volume
The effects o f positive end-expiratory pressure (PEEP) and lung Les effets de la pression expiratoire positive (PEEP) et de la
compliance (CO on delivered tidal volume (Vraet) and ventilator compliance pulmonaire (CI..) sur le volume courant gdndrd
output were evaluated in the following anaesthesia machine/ ('VTdel) et le ddbit du ventilateur m~canique sont dvaluds sur
ventilator systems: Narkomed III with a Model AV-E ventilator les appareils d'anesth~sie dquipds d'un ventilateur: Narkomed
(III/AV-E system) and an Ohmeda Modulus II with either a III avec un ventilateur AV-E (systdme III/ AV-E) et un appareil
7810 anaesthesia ventilator (II[7810 system) or a Model 7000 d'anesth~sie Ohmeda Modulus H ~quipd d'un ventilateur
anaesthesia ventilator (II/7000 system). With a standard circle d'anesthlsie 7810 (syst~me 11[7810) ou d'un ventilateur d'anes-
anaesthesia breathing circuit connected to a test lung simulating th~sie 7000 (syst~me 11/ 7000~ Avec une circuit standard avec
Ct~ gas flow was measured and integrated over time at each absorption branch,, sur un poumon artificiel simulant la CL,
combination o f V T settings (Vrset), 500 ml or 1000 mk CL le d~bit gazeux et mesur~ et int~grd par rapport au temps pour
settings, 0.15 to 0.01 L" cm 1-120-1 decreased incrementally; chacune des combinaisons de r~glage du V r (VTset), 500 ml
and PEEP settings, 0 to 30 cm 1120 increased in 5-cm 1-120 OU 1000 m k de riglage de la CL de 0,15 ?~0,01 L" cm 1120 -1
increments. The integral o f gasflow at the Y-piece o f the breath- diminu~e par plateau; r~glage du PEEP augmenti par plateau
ing circuit was recorded as VTdel a n d at the output o f the ven- de 0 it 30 cm HzO Lorsque que la CL it 0,01 L" cm 1120 -1
tilator bellows as ventilator output. As CL decreased to 0.01 et que le PEEP augmente it 30 cm H20 aux V~e t de 500
L" cm 1-120-1 and PEEP increased to 30 cm 1-120, at Vrset m l e t 1000 ml, le Vrdel diminue lindairement ~ 251 4- 6 ml
o f 500 ml and 1000 ml, respective Vrael decreased linearly to et 542 4- 7 ml avec le III/AV-E~ ~ 201 4- 5 et 439 4- 5 avec
251 4- 6 ml and 542 4- 7 with the III/AV-E, 201 + 5 and le 11/7810, et ~ 181 4- 4 et 433 4- 7 ml avec le 11/7000 (P
439 + 5, with the II/7810, and 181 • 4 and 433 4- 7 ml < 0,05 entre les trois syst~mes.) Laperte de VTdel due au PEEP
with the 11/7000 (P < 0.05 among the three systems). Loss seul, qui n'augmente que l~gkrement quand le Vrset est aug-
in VTdetdue to PEEP alone, which increased only slightly when ment~, explique un plus grand pourcentage de V~e t quand
Vrset was increased, accounted for an increasingly greater per- celui-ci est diminu~, ce qui est moins prononc~ quand la CL
centage o f VT~t as it Was decreased, which was less pronounced est basse. Les effets du PEEP et de la Ct, sur le d~bit du
with low CL. Effects o f PEEP and CL on ventilator output ventiiateur sont lea m~mes que ceux notes sur le VT.ae~ mais
were similar to those on VTdel but o f lesser magnitude. During de rnoindre importance. Pendant le PEEP, Vrset doit ~tre aug-
PEEP, Vrset must be increased to compensate for loss in Vraet menti pour compenser pour la perte de Vrae~ Le V T expir$
and expired V r must be monitored to prevent hypoventilation. doit dtre monitor$ pour pr~venir l~ypoventilation.
Key words
For controlled mechanical ventilation of anaesthetized pa-
ANAESTHESIA: equipment;
tients during surgery, an anaesthesia machine is com-
EQUIPMENT: ventilators;
monly connected in sequence with an anaesthesia ven-
Ltm6: tidal volume.
tilator. Patients with respiratory failure who undergo
From the Department of Anesthesiology, University of Florida surgery may require high levels of minute ventilation,
College of Medicine, Gainesville, Florida. Dr. Pan is Currently peak inspiratory airway pressure, and positive end-
Associate Professor of Anesthesiology at Medical College of expiratory pressure (PEEP). The tidal volume delivered
Virginia, Virginia Commonwealth University, Richmond, to the patient (VTdcOis usually determined not only by
Virginia. the tidal volume setting on the ventilator (VT~t), but also
Address correspondence to: Attn: Editorial Office, Dr. Pan, by the fresh gas flow rate, the compliance of breathing
Department of Anesthesiology, University of Florida College of circuit tubing, and the compressibility of gas. I-4 The
Medicine, P.O. Box 100254, Gainesville, FL 32610-0254. greater the peak inspiratory pressure, the greater is the
Accepted for publication 19th May, 1995. compression of delivered gas volume and the more the
FIGURE 1 Diagram of experimental arrangement. The application of PEEP requires additional gas compression within the volume enclosed by
the ventilator bellows, the CO2 absorber, the inspiratory valve, and the PEEP valve to increase pressure from ambient (shaded area).
circuit tubing is distended; these effects decrease Va'd~. For each system, a series of spring-loaded PEEP valves
Further, while it improves oxygenation in some patients,5 (Vital Signs, Totowa, NJ), in sequence from 5 to 30 cm
PEEP may also affect ventilator performance and, thus, H20, were placed at the junction of the expiratory valve
decrease Va'dct. This study was undertaken to quantify and the expiratory limb of the circle breathing circuit
the effect of PEEP on VTd~lat different levels of lung (Figure 1). A pneurnotachograph (model #2, Gleisch,
compliance (CO during mechanical ventilation with three Blue Bell, PA) and a differential pressure transducer
commonly used anaesthesia machine/ventilator systems. (Statham-Gonld, Anaheim, CA) were placed at the out-
put of the ventilator bellows to measure ventilator output
Methods and at the connection of the Y-piece to an adult test
The following three commonly used anaesthesia ma- lung (Michigan Instruments, Grand Rapids, MI) to
chine/ventilator systems were evaluated: measure VTd~(Figure 1)..The adult test lung was used
- I I I [ A V - E : A Narkomed III anaesthesia machine to simulate CL. Flow signals were integrated (Grass Med-
(North American D~ger) with the vaporizer turned ical Instruments, Quincy, MA) to measure volume. The
off, an AVE ventilator (North American Dr~iger, Tel- pneumotaehograph was calibrated under ambient tem-
ford, PA) and carbon dioxide absorber, 90 cm of 22- perature and pressure with a calibration syringe (model
mm corrugated tubing connecting ventilator and ab- #021156, WE Collins, Braintree, MA). The differential
sorber, and a disposable polyethylene circle breathing pressure transducer was calibrated against a water ma-
circuit (Intertech/Ohio, Bannoekburn, IL). nometer (Wescor). The anaesthesia machine/ventilator
- II/7810: A Modulus II anaesthesia machine (Ohmeda, systerns were all tested (including for leaks) and calibrated
Madison, WI) with the vaporizer turned off, a model according to manufacturers' recommendations before the
7810 ventilator (Ohmeda), a carbon dioxide absorber study.
(GMS, Ohmeda), and a disposable polyethylene circle For the experiment, the anaesthesia machine and ven-
breathing circuit (Intertech/Ohio). tilator were set as follows: fresh gas flow of 0 L- min -l,
- 11/7000: The same configuration as the 11/7810 but inspiratory-to-expiratory ratio of 1:2, and respiratory rate
with a model 7000 ventilator (Ohmeda). of 10 breaths-min -~. Before data collection, ventilators
Pan and van der AA: PEEP, CL A N D DELIVERED VT 833
V-rset [ ~ vr~t
350
750 ~ I~) 1'5 2~ 2'5 3~
PEEP(cmH20) PEEP (cnl H20)
FIGURE 4 Ventilator output (mean) at lung compliances of 0.05 FIGURE 5 Ventilator output (mean) at lung compliances of 0.05
L" cm H2 - j (open symbols) and 0.01 L "cm H2 - I (soild symbols) L ' c m H2 -1 (open symbols) and 0.01 L ' em H2 -1 (solid symbols)
during positive end-expiratory pressure (PEEP) and ventilation with a during positive end-expiratory pressure (PEEP) and ventilation with a
tidal volume setting (VTseO of 500 ml on the following three tidal volume setting (VTset) of 1000 ml on the following three
anaesthesia machine/ventilator systems: Narkomed III and an AV-E anaesthesia machine/ventilator systems: Narkomed III and an AV-E
ventilator (squares) and a Modulus II with either a 7810 ventilator ventilator (squares) and a Modulus II with either a 7810 ventilator
(circles) or a 7000 ventilator (triangles). The SD for all measurements (circles) or a 7000 ventilator (triangles). The SD for all measurements
never exceeded 11. never exceeded 11.
4 with the II/7000 and from VT~ of 1000 ml to, re- Discussion
spectively, 542 5: 7, 439 5: 5, and 433 5 : 7 ml (P < We measured ventilator output at the ventilator bellows
0.05 among the three systems) (Figures 4 and 5). The in order to evaluate the portion of change in VTde~due
VrPEEP~= increased only slightly but significantly when solely to ventilator performance, that is, the portion of
VT~ was increased from 500 to I000 ml (Table), thus, VT~I independent of the breathing circuit and anaesthesia
the decrease (VTPEEPIoss) as a percentage of VTset Was machine. The effect of PEEP on Vrdet results from gas
greater at VTm of 500 than of 1000 ml. compression of the space within the ventilator bellows,
The effect of PEEP and C L on ventilator output was the inspiratory valve, and the PEEP valve (Figure 1,
similar to that on VTd~l,but of lesser magnitude (Figures shaded area), which is required to increase ambient pres-
4 and 5). Compared with the ventilator output with the sure. The VT~l can be derived from the following equa-
11/7810 and 11/7000, that with the III/AV-E was affected tion. 1,5,6
least (P < 0.05). When CL was decreased from 0.15 to
VTdel : VTset "~-fresh gas flow rate. inspiratory time
0.01 L. cm H20 -1, ventilator output with the III/AV-
-- VTPEEPIoss - - VT(C+CL)Ioss
E decreased less than 8% at VTm of 500 ml and less
t h a n 6 % a t a VTset o f I 0 0 0 m l . w h e r e VT(CL+C)1oss is t h e loss d u e solely to CL a n d c o r n -
TABLE Loss of delivered tidal volume caused by positive end.expiratory pressure alone (PEEP) at different simulated lung compliances and
different tidal volume settings (VTsv) on three anaesthesia machine/ventilator systems
Lung compliance of O.05 L" cm 1-120"-1 Lung compliance of O.Ol L" cm 1t20"1
Anaesthesia/ raachine
ventilator system Vrset of 1000 ml Frsetof 500 ml Vrset of 1000 ml VTm of 500 ml
Values are means :t: SD. PEEP of 30 era" H20 represents the worst case.
*P < 0.05 compared with VT~ of 1000 ral at the same lung compliance.
Pan and van tier AA: PEEP, CL AND DELIVERED VT 835
Acknowledgements
This material was presented in part at the 1990 Annual
Meeting of the American Society of Anesthesiologists,
Las Vegas, Nevada, October 19-23, 1990. The authors
thank Lynn Dirk of the Department of Anesthesiology,
University of Florida for editing the manuscript.