Professional Documents
Culture Documents
and a valve
YueYang Yuan, JianGuo Sun, Baicun Wang, Pei Feng, and ChongChang Yang
FIG. 1. Pneumatic and electric circuit block diagram. Dashed arrow lines denote gas flow. Solid arrow lines denote electrical signal.
of patient airway.12 The pressure is monitored by the pressure The corresponding messages for confirming the settings
sensor. A leakage hole is deliberately placed in the mask to are displayed on LCD. After the ventilator is started, the
eliminate CO2 (carbon dioxide). During the inspiration time, blower is driven to carry a constant gas flow into the valve. By
part of flow is inhaled by the patient into airway and the rest of regulating the electric current of solenoid, the valve transports
the flow leaks out from the mask. During the expiration time, a gas flow to the patient end in proportion. These electrical
gases from both the ventilator and from patient are taken out analogy signals from two flow sensors and one pressure sensor
from the mask and dissipate into the atmosphere. are processed by their signal processes circuits and fed back
The mechanism of valve for regular gas flow is explained to the microcontroller.
in Fig. 2. The valve output flow (QVOUT ) is in proportion to the Based on the hardware, the embedded software for
valve input flow (QVIN ). It is regulated by the valve leakage this pressure-controlled ventilator is programmed into the
flow (QVCON ). The leakage flow rate is in positive proportion to microcontroller. In this software, three basic ventilation modes
the gap size (gap). And the gap size is in positive proportion are prepared: the spontaneous controlled mode combining
to the electric current in the valve solenoid (IV ). A smaller HFOV (S-HFOV mode), time-cycled mode combining HFOV
current in the valve solenoid means a smaller gap size, which (T-HFOV mode), and the continuous positive airway pressure
means a smaller leakage flow and a larger output flow. Their ventilation mode13 combining HFOV (CPAP-HFOV mode).
relationship can be expressed in Equation (1). K1 and K2 stand Fig. 3 describes the embedded software.
two proportional coefficients, When the ventilator is powered on, its system restores
the parameters back to the last operation condition. Those
QVOUT = QVIN − QVCON parameters can also be set by buttons in the panel and then
QVLEAK = K1 × gap
. (1) will be displayed on LCD. After the ventilation is started, the
corresponding output can be carried out. The PID14 (Propor-
Gap = K2 × IV
tion Integration Differentiation) control algorithm is designed
The electric circuit system is mainly composed of a micro- to control the ventilator to work in preset requirements. The
computer, sensor signal process circuits, the blower driver blower outputs a constant gas flow to the valve, and then the
circuit, the valve driver circuit, operation panel, and LCD valve transports the flow to produce a positive pressure at
(Liquid Circuit Display). Through the operation panel, the user the patient end. Finally the system will calculate the volumes
may set the ventilation pressures, IE ratio (the time ratio of based on the pressure and flow rate with Equation (2) and
inspiration and expiration), BPM (breathing rate per min), etc. display them on LCD,
FIG. 2. Valve and its flow control method. Q VIN , Q VOUT , and Q VCON stand input flow, output flow, and leakage flow.
025113-3 Yuan et al. Rev. Sci. Instrum. 87, 025113 (2016)
TABLE I. The main parameters set in ventilator. The unit of pressures (IPAP, PAP, CPAP, ∆PI, ∆PE, ∆P) is cm
H2O. NA denotes not applied.
S-HFOV 18 6 12 8 12 1 5
T-HFOV 18 6 12 8 12 1 5
18 12 NA NA 5
CPAP-HFOV 18 10 NA NA 2–10
18 5 NA NA 2–10
025113-4 Yuan et al. Rev. Sci. Instrum. 87, 025113 (2016)
FIG. 5. Data obtained under operating nHFOV ventilator with different ventilation mode: “In” and “Ex” express inspiration and expiration. (a) S-HFOV mode
(5 Hz of HFV), (b) T-HFOV mode (5 Hz of HFV), and (c) CPAP-HFOV mode (5 Hz of HFV).
TABLE II. Measured peak-peak pressures and flows at each frequency as preset oscillation pressure 10 cm H2O and 5 cm H2O. Data were expressed as
mean ± SD of 10 measurements randomly obtained during the entire experiment.
Measured data
Preset Measured 2 Hz 3 Hz 4 Hz 5 Hz 6 Hz 7 Hz 8 Hz 9 Hz 10 Hz
∆P 9.76 ± 0.22 9.16 ± 0.04 8.43 ± 0.08 7.68 ± 0.25 7.21 ± 0.42 6.59 ± 0.27 6.32 ± 0.4 6.14 ± 0.29 5.85 ± 0.48
∆P = 10
∆F 72.64 ± 2.39 71.26 ± 2.72 64.84 ± 2.78 63.52 ± 3.61 65.24 ± 4.82 59.35 ± 3.9 60.29 ± 3.28 53.61 ± 1.63 56.28 ± 4.21
∆P 5.08 ± 0.03 4.82 ± 0.04 4.13 ± 0.03 3.61 ± 0.06 3.23 ± 0.09 3.10 ± 0.06 2.87 ± 0.21 2.80 ± 0.08 2.47 ± 0.1
∆P = 5
∆F 44.10 ± 4.36 42.97 ± 2.42 41.28 ± 2.05 38.18 ± 2.47 37.63 ± 1.7 35.69 ± 2.15 36.87 ± 2.17 33.15 ± 2.09 30.67 ± 1.79
H2O. And measured ∆PE is 5.08 ± 0.06 cm H2O (n = 24, tube is necessary for Sensormedics 3100A. The ET-tube also
p < 0.01) when it is preset to 8 cm H2O. The measured losses22 about 34%–90% pressure when the diameter of the
IPAP (or CPAP) is 17.77 ± 0.78 cm H2O (n = 36, p = 0.086) ET tube ranges between 6.5 mm–2.5 mm. However, despite
when it is preset to 18 cm H2O. And the measured EPAP is leakages from the valve and from the mask in this designed
6.88 ± 0.36 cm H2O (n = 24, p < 0.01) when it is preset to nHFOV ventilator, according to the existing results under our
6 cm H2O. experimental conditions, the therapy oscillatory pressures are
Fig. 6 describes the peak-to-peak oscillatory pressures in directly delivered to the airway though the nasal mask with
the mask and the oscillatory flow rates in the main airway when low-loss pressure. According to experimental results given in
the frequency is changed from 2 Hz to 10 Hz. The spectrums Fig. 6 and Table II, the max oscillatory pressure amplitude
(pressure spectrums and flow rate spectrums) disclose that has a loss ≈50% at 10 Hz and nearly zero loss at 2 Hz. The
the oscillatory pressures and the oscillatory flow rates are measured mean pressure at each frequency is a steady-going
inversely proportional to the oscillatory frequency of HFV value which is closes to the preset pressure of CPAP.
under given experimental conditions. The measured peak-to-
peak oscillatory pressures (∆P) and oscillatory flow rates (∆F)
are read from Figs. 6(a) and 6(b) and listed in Table II. At preset IV. CONCLUSIONS
∆P = 10 cm H2O, the ∆P and ∆F are ranged from 9.76 cm H2O In this work, an experimental nHFOV ventilator was
(n = 10, p < 0.009) at 2 Hz down to 5.85 cm H2O (n = 10, achieved by utilizing a high-speed blower and an invented
p < 0.001) at 10 Hz and from 72.64 l/min at 2 Hz down to valve. With this ventilator, three ventilation modes S-HFOV,
56.28 l/min at 10 Hz, respectively. Similarly, at preset ∆P = 5 T-HFOV, and CPAP-HFOV are operated to ventilate the flash
cm H2O, ∆P and ∆F are ranged from 5.08 cm H2O (n = 10, sheep lungs in a typical experimental setup.
p < 0.006) at 2 Hz down to 2.47 cm H2O (n = 10, p < 0.001) The experimental results demonstrate that this nHFOV
at 10 Hz and from 44.1 l/min at 2 Hz down to 30.67 l/min at ventilator can provide a low-loss oscillatory pressure into nasal
10 Hz. The measured mean pressure of 18.7 cm H2O ± 0.33 mask and large valuable minute ventilation volume as well as
(n = 18, p < 0.001) read from Fig. 6(a) is close to the preset small tidal volume into the patient’s lungs though the common
of 18 cm H2O. breath tube and nasal mask.
According to the experimental results shown in Figs. 5(b) The nHFOV ventilator designed according to specifica-
and 5(c), the minute ventilation volume (MV) is large but the tions of this analysis is able to provide proper ventilation
tidal volume is small (in general, MV ≈ 6000 − 9000 ml and and avoid the practice of inserting the ET-tube into patient’s
VT ≈ 500 ml for a normal healthy adult17). The maximum airway.
tidal volume VT(max) is even smaller than the anatomical
dead space (≈150 ml for normal healthy adult17). In classic
ACKNOWLEDGMENTS
concept of pulmonary ventilation, the amount of gas reaching
the alveoli shall be equal to the required life-sustaining The technical assistance from Curative Medical is grate-
volume (≈500 ml for normal healthy adult). Though, in our fully acknowledged. This project is supported by Chinese
experiments, the low tidal volume (or oscillation volume) is Universities Scientific Fund (Item No.: 14D310302).
still not a satisfactory explanation for whether the HFOV
is effective in life support.18,19 The large minute ventilation 1S. Mehta, “High frequency ventilation,” Curr. Opin. Crit. Care 6(1), 38–45
volume provided in our experiments with the test nHFOV (2000).
2C. C. Dos Santos and A. S. Alutsky, “Overview of high-frequency ventila-
ventilator is believed to be one of the key factors for sustaining tion modes, clinical rationale, and gas transport mechanisms,” Respir. Care
life. Clin. North Am. 7(4), 549–575 (2001).
3M. Amit, S. Balpreet, H. Salhab El et al., “Use of noninvasive high-
So far, we have not found other technical papers that
discuss about how to realize the nHFOV with the mask in a frequency ventilation in the neonatal intensive care unit: A retrospective
review,” Am. J. Perinatol. 30(2), 171–176 (2015).
non-invasive ventilator. The only available literature offered 4M. A. Borrello, “Closed loop control system for a high frequency oscillation
by De Luca et al.20,21 describes how the nHFOV works through ventilator,” U.S. patent 8434482 (05 July 2013).
5B. A. Simon and W. Mitzner, “Design and calibration of a high-frequency
nasal prongs with an invasive ventilator Sensormedics 3100A
oscillatory ventilator,” IEEE Trans. Biomed. Eng. 38(2), 214–218 (1991).
and a neonatal lung model. According to their records, about 6K. Suwa and M. Taqami, “High-frequency animal ventilator using a loud-
22%–38% of pressures were reduced by the controlling circuit speaker and its gas exchange characteristics,” Jpn. J. Physiol. 31(6), 859–868
and prongs. Furthermore, in generally application, an ET- (1981).
025113-6 Yuan et al. Rev. Sci. Instrum. 87, 025113 (2016)
7B. Ibsen, “The anesthetist’s viewpoint on the treatment of respiratory lung model with Babylog®; 8000 plus,” Pediatr. Pulmonol. 50(2), 173–178
complications in poliomyelitis during the epidemic in Copenhagen, 1952,” (2013).
Proc. R. Soc. Med. 47(1), 72–74 (1954). 16R. W. Robinson, D. P. White, and C. W. Zwillich, “Moderate alcohol
8J. Downar and S. Mehta, “Bench-to-bedside review: High-frequency oscil- ingestion increases upper airway resistance in normal subjects,” Am. Rev.
latory ventilation in adults with acute respiratory distress syndrome,” Crit. Respir. Dis. 132(6), 1238–1241 (1985).
Care 10(6), 1–8 (2006). 17B. Maury, “The respiratory system in equations,” in Ms&A—Modeling
9R. Dembinski, M. Max, R. Bensberg et al., “High-frequency oscillatory Simulation & Applications (Springer, Verlag, Italia, 2013), p. 217.
ventilation in experimental lung injury: Effects on gas exchange,” Intensive 18H. K. Chang, “Mechanisms of gas transport during ventilation by high-
Care Med. 28(6), 768–774 (2002). frequency oscillation,” J. Appl. Physiol.: Respir., Environ. Exercise Physiol.
10L. M. Hannan, G. S. Dominelli, Y. W. Chen et al., “Systematic review of 56(5), 553–563 (1984).
non-invasive positive pressure ventilation for chronic respiratory failure,” 19J. Jane pillow, “High-frequency oscillatory ventilation: Mechanisms of gas
Respir. Med. 108(2), 229–243 (2014). exchange and lung mechanics,” Crit. Care Med. 33(Suppl. 3), S135–S141
11J. G. Sun and X. M. Zhang, “Three-way pressure balance valve and medical (2005).
respirator,” Chinese invention patent CN202010105871 (03 August 2013). 20D. De Luca, M. Piastram, D. Pietrini, and G. Conti, “Effect of amplitude
12B. Diong, H. Nazeran, P. Nava et al., “Modeling human respiratory imped- and inspiratory time in a bench model of non-invasive HFOV through nasal
ance,” IEEE Eng. Med. Biol. Mag. 26(1), 48–55 (2007). prongs,” Pediatr. Pulmonol. 47(10), 1012–1018 (2012).
13A. Antonescu-Turcu and S. Parthasarathy, “CPAP and bi-level PAP therapy: 21D. De Luca, V. P. Carnielli, G. Conti, and M. Piastra, “Noninvasive high
New and established roles,” Respir. Care 55(9), 1216–1229 (2010). frequency oscillatory ventilation through nasal prongs: Bench evaluation
14D. Son and H. Choi, “A proportional integral differential control of flow over of efficacy and mechanics,” Eur. J. Intensive Care Med. 36(12), 2094–2100
a circular cylinder,” Philos. Trans. R. Soc., A 369(1940), 1540–1555 (2011). (2010).
15N. Ken, T. Etsushi, N. Fumikatsu et al., “The temperature change in an endo- 22Operator’s Manual 3100A High Frequency Oscillatory Ventilator, C 2005
tracheal tube during high frequency ventilation using an artificial neonatal VIASYS Respiratory Care, Inc.