SMART PIPETTE - USER GUIDE
n S = Standard Deviation
∑ ( Vi - V)2
S= i=1 V = Mean Value
n-1 n = number of readings
CV% = 100 · S
_
V
| 10
SMART PIPETTE - USER GUIDE
0.25 12 6
1 2.5 1.8
10 1 0.6
5 2 2
50 0.6 0.3
10 3 1
100 3 0.6
2 3 2
20 0.9 0.4
20 2.5 0.7
1000 3 0.6
11 |
SMART PIPETTE - USER GUIDE
SFA-0020 20 µl 20 1 0.5
SFA-0025 25 µl 25 1 0.3
1 8 5
SMA8-010
SMA12-010 0.5-10 µl 0.02 5 4 2
10 2 1
2 7 3
SMA8-020
2-20 µl 0.02 10 3 2
SMA12-020
20 2 1.6
5 3 2
SMA8-050
5-50 µl 0.1 25 1.5 1
SMA12-050
50 1 0.7
10 3 2
SMA8-100 10 – 100
0.1 50 1 0.8
SMA12-100 µl
100 0.8 0.3
| 12
SMART PIPETTE - USER GUIDE
20 5 1.4
SMA8-200 20-200
0.2 100 1 0.4
SMA12-200 µl
200 0.7 0.25
30 3 1
SMA8-300 30-300
0.2 150 1 0.5
SMA12-300 µl
300 0.6 0.3
Calibration Adjustment
13 |
SMART PIPETTE - USER GUIDE
MAINTENANCE
When the SMART pipette is not in use, please make sure that it is
stored in an upright position. Accumax recommends the SMART
stand for this purpose.
You should check the pipette at the beginning of each day for accu-
mulation of dust and dirt on the outside surfaces. Attention should
be paid to the tip cone. No other solvents except Iso-propanol should
be used to clean the pipette. If the pipette is used daily, it should be
inspected every three months.
1 2 3
1. For opening press 2. Rotate the tip ejector 3. Turn out the tip ejector,
the tip ejector closer closer in the direction slowly
upwards indicated until the
snap
• Pull out the piston and other parts by unscrewing the tip cone.
Push out with piston the rest of the piston assembly including the
Seal and O-ring. Remember to keep all parts in order for reassem-
bly.
• Clean the piston, the piston spring, Seal and the O-ring with a iso-
propanol and lint free tissue. Allow them to dry.
• Check the tip cone for foreign particles and remove, if any.
• Grease the cleaned parts with the approved lubricant provided with
each pipette.
| 14
SMART PIPETTE - USER GUIDE
1
Align the notch on the tip ejector with the slots on tip
ejector closer as indicated
2
Ensure the notch on the main handle matches the
slots on the tip ejector closer. Push the tip ejector
closer upwards as indicated.
3
Insert both parts in the aligned manner as indicated.
For correct re-assembly, it is crucial that the slots on the tip ejector closer match
the notches on the inside of the main handle. Please follow the above procedure.
15 |
SMART PIPETTE - USER GUIDE
SP 377-1
SP 302-1
SP 70-1
SP 23-1
SP 317-1
SP 392-1
SP 259-1
SP 44-1 SP 314-1
SP 14-1
| 16
SMART PIPETTE - USER GUIDE
Group 2 Group 3
SP 41-1 SP 41-2
SP 301-1
SP 301-1
SP 48-2 SP 48-2
SP 258-1
SP 48-1 SP 259-1
SP 48-1
SP 38-1
SP 38-1
SP 25-1
SP 25-1
SP 314-1
SP 314-1
SP 14-1
SP 14-1
17 |
SMART PIPETTE - USER GUIDE
Group 4 Group 5
Variable Volume 2-20µl Variable Volume 5-50µl
SP 39-3
SP 40-1
SP 301-1
SP 306-1
SP 48-2 SP 45-1
SP 391-1
SP 265-1
SP 260-1
SP 47-1
SP 48-1
SP 38-1
SP 38-1
SP 25-1
SP 24-1
SP 314-1
SP 314-1
SP 14-1
SP 14-1
| 18
SMART PIPETTE - USER GUIDE
Group 6 Group 7
Variable Volume 10-100µl Variable Volume 20-200µl
SP 39-1 SP 366-1
SP 303-1
SP 305-1
SP 45-1
SP 45-1
SP 386-1
SP 261-1 SP 388-1
SP 264-1
SP 46-1
SP 45-1
SP 38-1 SP 38-1
SP 24-1 SP 24-1
SP 314-1 SP 314-1
SP 14-1
SP 14-1
19 |
SMART PIPETTE - USER GUIDE
Group 8 Group 9
SP 31-2 SP 30-2
SP 51-1 SP 50-1
SP 53-1
SP 52-1
SP 353-1 SP 353-1
SP 304-1 SP 304-2
SP 57-2 SP 57-1
SP 263-1 SP 262-1
SP 42-1 SP 43-1
SP 28-1
SP 29-1
SP 26-1
SP 27-1
SP 312-1
SP 313-1
SP 55-1
SP 54-1
| 20
SMART PIPETTE - USER GUIDE
autoclaving
Accumax SMART pipette is completely autoclavable at 2 bar pres-
sure, 121°c temperature for 20 minutes exposure time according to
DIN.
21 |