International Journal of Trend in Scientific Research and Development (IJTSRD)

International Open Access Journal | www.ijtsrd.com

ISSN No: 2456 - 6470 | Volume - 3 | Issue – 1 | Nov – Dec 2018

Real Time Signal Quali


Quality
ty Aware Internet of Things (IOT)
Framework for FPGA Based ECG Telemetry
lemetry System and
Development oof Android Application
Dhanashri P. Yamagekar, Dr. P. C. Bhaskar
Department off Technology, Shivaji University Kolhapur, India

ABSTRACT
Day by day the scope & use of the electronics as well as the size and position of the chambers, the
concepts in bio-medical
medical field is increasing gradually. presence of any damage to the heart, and the effects of
A novel approach to the design of real time ECG the drugs or devices used to regulate the heart, such as
signal acquisition system for patient mo monitoring in a pacemaker.
medical application, FPGA (Field Programmable Gate
Array) is the core heart of proposed system which is A typical ECG tracing of normal heartbeat consists of
configured and programmed to acquire using ECG a P wave, a QRS complex and a T wave. Data
(Electrocardiogram) sensor. In this paper a new acquisition products server as a focal point in a
concept of ECG telemetry system is discussed along system, tying together a wide variety of products such
with signal quality aware IoT framework for energy as sensor that indicate temperature, flow, levels, pulse
efficient ECG monitoring system. Tele monitoring is rate, electrocardiograph etc.
a medical practice that involves monitoring patients
who are not at the same location as the healthcare The core heart of the proposedoposed system is field
provider. The purpose of the present study is use to programmable gate array (FPGA) which is configured
identify heart condition and give the information to and programmed to acquire a real time data. Real time
the doctor. The objective of the study is to improve data from the process is acquired using ECG sensor.
the doctor-patient
patient ratio and evaluation of cardiac Signal conditioners are designed for each sensor.
diseases in the rural population. The proposed system Signal conditioners are interfaced
terfaced with FPGA through
for the electrocardiogram (ECG) mon monitoring ADC and MAX 232 for PC interfacing. Output of the
controlled by FPGA and implemented in the form of system is displayed waveform on PC.
android application.
An electrocardiogram is a test that checks for problem
Keyword- FPGA (Field programmable gate array), with the electrical activity of our heart. An ECG
Electrocardiogram (ECG) signals, IoT (Internet of translates the heart’s electrical activity
act into line
Things) tracing on paper. The spikes and dips in the line
tracing are called waves. The ECG signal consist of P,
I. INTRODUCTION Q, R, S, T waves. The function of each wave is-
is
The word electrocardiography is evolved from Greek  P Wave- represents the atrial contraction
word Kardia which means Heart. ECG that is
 QRS complex represents the ventricular
electrocardiography is a process of interpretation of
contraction.
heart activity over the period of time and is detected
by electrodes attached to the surface of body. An ECG  R peaks- represents a heartbeat.
is used to measure the heart’s electrical conduction  T wave- represent the last common wave in an
system. It picks up electrical impulses generated by ECG. The electrical signal is produced when the
the polarization and depolarization of cardiac tissue ventricles are repolarizing.
and translates into a waveform. The waveform is then
used to measure the rate and regularity of heartbeats,

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 216
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
MATLAB environment to de-noisede ECG signal and
test the designed algorithm on modelsim these signal
transfer digital filter algorithm on FPGA plateform
and to analyze the complexity and accuracy of digital
filter on MATLAB.[15]

P. Girish, B .V .Ramana, [2014] A real time ECG


monitoring system featuring HEV analysis and
wireless data transmission through radio frequency
based on a SoC development platform. In this system
modelsim Xillinx Edition and Xillinx ISE will be used
Fig. 1- ECG signal Characteristics simulation and synthesis respectively. The Xilinx
chip-scope
scope tool will be used to test the FPGA inside
II. RELATED WOEK result while the logic running on FPGA. [11]
Several implementations of finding response
sponse time of
Real-Time
Time System exist. Also there are several Yongming Yang, [2007] A real time ECG monitoring
implementations of performance analysis of IoT based system is proposed based on FPGA. The FPGA chip
Health Care Monitoring for Real-Time
Time System. Also is taken as the centrol microprocessor and applies
there are novel implementations of Signal Quality structured design on VHDL to collect and transmit
Aware ECG Telemetry System for continuous cardiac real-time
time ECG signals. It contain low power
health monitoring applications. A brief survey’s consumption to carry around and long usage time.
mentioned as below- The digital filtering and data compression arithmetic
are also integrated in the FPGA chip and can be
Udit Satija, [2016] proposes a novel signal quality improved or added with some other modules with
aware IoT-enabled
enabled ECG telemetry system for software modification. It is more intelligent compared
continuous cardiac health monitoring applications. with existing portable ECG monitoring system. [20]
The main purpose of these system is design and
development of light weight ECG signal quality III. BLOCK DIAGRAM
assessment method for automatically classifying the The system is used to monitor ECG. Data is digitized
acquired ECG signal into acceptable or unacceptable with an FPGA microcontroller and send to a
class and real time implementation of proposed IoT computer. The system s portable and runs on less
enabled ECG monitoring framework.[1]
[1] power. The system improves the quality of patient
monitoring and eliminates the need for nurses to
R.Harini1, [2015] ECG is identify heart condition in repeatedly manually
ally perform these measurement.
the rural underserved population where the doctor
patient ratio is low and access to health care is
difficult. The purpose of these system was clinical
validation of handheld ECG as a screening tool for
evolution of cardiac diseases in the rural population.
[9]

B. Khaleelu Rehman1, [2017] The methodology of


ECG FIR filter based on FPGA. The ECG signal is Fig.2- Block diagram of FPGA based ECG patient
processed and implemented on FPGA platform. The monitoring system.
filtered signal is subject to short time Fourier
Transform by medical experts. A implication can be A] FPGA-
made by medical experts. A recorded ECG signal is A field-programmable
programmable gate array (FPGA) is an
used as test input and to test the implemented modules integrated circuit designed to be configured by the
on FPGA. [5] customer or designer after manufacturing, hence
"field-programmable".
programmable". The FPGA configuration is
Sande Seema, [2010] propsed the de-noising
noising of ECG generally specified
fied using a hardware description
signal also design algorithm for IIR
R and FIR filter in language (HDL), similar to that used for an

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 217
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
application-specific integrated circuit (ASIC) FPGAs serves the function to transfer the data in between PC
can be used to implement any logical function that an and the FPGA, for the further operation of the system
ASIC could perform.
E] COMPUTER /PC-
The ability to update the functionality after ship
shipping, Computer / PC is used in a project to visualize the
partial of the portion of the design and the low non
non- output of the application. Computer / PC is interfaced
recurring engineering costs relative to an ASIC design for the Electrocardiography. A waveform of ECG can
(notwithstanding the generally higher unit cost), offer be visualize on Computer /PC.
advantages for many applications. FPGAs contain
programmable logic components called ""logic IV. PROPOSED METHODOLOGY
blocks", and a hierarchy of reconfigurable A] For IOT Framework –
interconnects that allow the blocks to be "wired The main modules of signal quality-aware
quality (SQA)-IoT
together" somewhat like many (changeable) logic framework are illustrated in Fig. It consists of three
gates that can be inter-wired
wired in (many) different modules:
ules: (i) ECG signal sensing module, (ii)
configurations. Logic blocks can be configured to automated signal quality assessment module, and (iii)
perform complex
mplex combinational functions, or merely signal-quality
quality aware ECG analysis and transmission
simple logic gates like AND and XOR. The Spartan
Spartan-6 module.
family of Field Programmable Gate Array specifically
designed to meet the needs of high volume cost cost-
sensitive consumer electronic.

B] ECG Sensor-
The AD8232 is an integrated signal conditioning
block for ECG and other bio potential measurement Fig.- 3 Signal Quality--aware framework
applications. It is designed to extract, amplify, and
filter small bio potential signals in the presence of The main focus on design and real-time real
noisy conditions, such as those created by motion or implementation of automated ECG signal quality
remote electrode
ctrode placement. This design allows for an assessment method and validation of the effectiveness
ultra-low power analog-to-digital
digital converter (ADC) or of the proposed SQA-IoT IoT framework under resting,
an embedded microcontroller to acquire the output ambulatory and nd physical activity conditions, The
signal easily. proposed automated ECG signal quality assessment
(ECG-SQA)
SQA) method consists of three steps such as
C] ADC- flat-line
line (or ECG signal absence) detection, abrupt
An analog-to-digital converter (ADC, A/D, or A-to- baseline wander extraction, and high-frequency
high noise
D) is a system that converts an analog signal
signal, such as detection and extraction to compute the signal quality
a sound picked up by a microphone or light entering index (SQI) for assessing the clinical acceptability of
a digital camera, into a digital signal.. An ADC may ECG signals. In this his work, the ECG-SQA
ECG is
also provide an isolated measurement such as implemented based on the-basedbased filtering, turning
an electronic device that converts an input points and decision rules.
analog voltage or current to a digital number
representing the magnitude of the voltage or current. I] HF Noise Detection –
Typically the digital output is a two's The high frequency (HF) noises such as muscle
complement binary number that is proportional to the artifacts, power line interference, motion artifacts,
input, but there are other possibilities. pause and instrument noise are ar introduced in the
acquired ECG signal. Apply decision rules for
D] RS232- detecting the ECG segment with HF noises-
noises
RS 232 is a serial communication cable used in the
system. Here, the RS 232 provides the serial
communication between the microcontroller and the
outside world such as display, PC or Mobile etc. So
1it is a media used to communicate between
microcontroller and the PC. In our system the RS232 where R1 is true if a total number of turning points in
any block (tpk)) exceeds 5% of the number of samples

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 218
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
(P ∗ F s, where, F s is the sampling frequency of the The detection decision rule is given by-
by
signal) in that block and R2 is true if the distance
between two turning points rk is less than two.

If SQIHF = 1 then the segment is classified as a HF Where SQIABW denotes the signal quality index
noise segment. Otherwise the segment is classified as (SQI) for the presence of abrupt baseline drift (ABW)
a high frequency noise-free
free ECG segment. event. Here, the γ is chosen as 0:2
0 mV for detecting
abrupt baseline wander which can distort the ST
II] ECG signal absence detection- segment and other low-frequency
frequency components of the
Due to the disconnection of electrodes with skin and ECG signal.. The ECG signal with abrupt baseline
the electronic component saturation, sensing device wander event is classified as an unacceptable. The
exhibits the absence of ECG signal information in the ECG signal with slowly varying baseline wander is
acquired signal. In practice, observation
rvation to the classified as an acceptable since it can be removed
recording shows the presence of zero amplitude flat from the signal without significantly distorting the
line (ZFL), only baseline wander (OBW), and the PQRST complexes.
long pause with physiological and external noises.
Existing approaches were developed for detection of IV] ECG Signal Quality Grading-
Grading
ZFL event. It is based on turning poin
points (TP) which The grading of acquired ECG signal based on the
can be computed as mentioned in Algorithm. decision scores obtained for the detection of abrupt
baseline wander, ECG signal absence, and high
frequency noises. The ECG signal is graded with
three classes such as Good,ood, Intermediate and bad
based on the HF noise score. The presence of flat line
If SQIFL = 1 then the segment is classified as a flat and abrupt baseline wander may result in noisy
line or ECG signal absence segment. Otherwise the clinical features. Based upon assessment results, it is
segment is further processed for detecting the noted that the some morphological features and RR
presence of high frequency noises. intervals can be measuredred from the ECG signal with
some level of HF noises. Thus, we grade the noisy
III] Abrupt Change Detection- ECG signal into intermediate and bad. The value of
The ECG signals are corrupted by baseline wanders γH is chosen as 0.1% of the maxima or minima in the
that are mainly caused by respiratory activity, body HF noise detection stage for detecting severe noise or
movements, skin-electrode
electrode interface, varying bad quality signal. While the noise level threshold γH
impedance between electrodes and skin due to poor is set to 0.05% of the maxima or minima for grading
electrode contact. the noisy signal into intermediate class.

By using FIR Based –The The basic structure of FIR filter


is like a sub-section
section of the delay lines. Each section of
the output of the weighted cumulative, the FIR
coefficient is-

y(nT). = . ℎ(𝑘) 𝑥(𝑛 − 𝑘


𝑘) Where overall signal quality index (SQI) is computed
as

The baseline wander signal is obtained as


as- SQI = SQIABW + SQIFL + SQIHF
b[n] = x[n] - x~ [n]
Good quality indicates the ECG signal with
Where b (n)) is the baseline wander signal and x~ (n) acceptable level of noise which does not distort the
is the baseline wander removed signal which is morphological features while intermediate and bad
computed as indicate the medium and high level of noise or poor
quality, respectively. The intermediate class of ECG
signals
ls may be considered for application-specific
application
ECG signal analysis systems.

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 219
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
B] For Androd Application- significant amounts of muscle noise and baseline
 ALGORITHM- wander. Recorded ECG ‘ma’ contains primarily prima
1. Initialize central processing unit muscle noise (EMG) with spectrum, which overlaps
2. Initialize Ports that of the ECG, which extends to higher frequency.
3. Initialize memory
4. Initialize ADC 1.2 MATLAB
5. Connect 3-leads of electrode in correct place MATLAB is a powerful, comprehensive, and easy to
6. Read the signals from the sensor and transmit use environment
ment for technical computations, one of its
signals to the amplifier greatest strengths is that MATLAB allows building its
7. Convert analog signal to digital tal signal uusing own reusable tools. Mat lab is being used for ECG
ADC3204 database extraction which is stored at
8. Display the ECG signal on PC www.physiobank.org.. Stored ECG signal can be
9. Also display on Android phone called in Simulink model from workspace of mat lab.
Real time EMG noise is accessed from MIT-BIH MIT
V. SYSTEM IMPIMENTATON noise stress database in the form of .mat file then it is
A] Software Implementation imported into workspace which is added with pure
1.1 Mit-Bih Database ECG in Simulink model. For designing of FIR low
The dataset used in this study is obtained from pass filter FDA Tool from MATLAB has been used.
Physiobank entitled MIT-BIH BIH Arrhythmia Database Different types of FIR digital filter designed to
available on-line. MIT-BIH BIH Arrhythmia Database remove high frequency EMG noise.
which having sources of ECGs includes set of over
4000 long-term
term Holter recorders, which was obtain 1.3 FIR FILTER
from Beth Israel Hospital Arrhythmia Laboratory. FIR filters are widely used because of inherent
Each signal is recorded with 360 sampling frequency stability and due to the powerful design algorithms
which having 11-bit bit resolution with 10 mv range. which are exist from them. When FIR filters
MIT-BIH
BIH Arrhythmia Database where annotated ECG implement in non-recursive
recursive form,
f linear phase one
signals are described by a text header file (.hea), a can obtain with ease. FIR low pass filter is designed
binary file
ile (.dat) and a binary annotated file (.atr). by equiripple filtering, least square method,
Header file consists of detailed information such as windowing method for designing have been used. In
number of samples, sampling frequency, format of windowing method such as Kaiser, Rectangular,
ECG signal, type and number of ECG leads, patient’s Hamming, Hanning and Blackman functionsfunction are
history and the detailed clinical information. In binar
binary designed.
data signal file, the signal is stored in 212 format
which means each sample requires number of lead 1.4 Flow
ow chart for MATLAB programming
times 12 bits to be stored and the binary annotation
file consists of beat annotation from which half of this
database has been freely available .ForFor EMG dataset
is obtained from MIT-BIHBIH Noise stress database from
PhysioBank. Noise stress database includes 12 half half-
hour ECG recordings and 3 half-hourhour recordings. This
is having a noise typical in ambulatory ECG
recordings. PhysioBank noise recordings were taken
by physical activities of volunteers. For ECG
recording 3 noise records ds were assembled. These
recordings were contain predominantly baseline
wander (as 'bw'), muscle artifact (as 'ma') which also
known as EMG noise and third one is electrode
motion artifact (as 'em'). From noise stress database
records of ‘bw’ contains primarilyarily baseline wander
with low frequency signals, which was caused by
motion of the volunteer or the leads. Recorded ECG
‘em’ contains electrode motion artifacts with 1.5 Simulink model for FIR
F filter

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 220
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
FLOW CHART

Fig4- Simulink
imulink Model of ECG De
De-noising

In above fig-44 it show that the design of FIR filter.


The filters are designed using MATLAB FDA Tool
by specifying the filter order, cut off frequency and
sampling frequency. In the present section, the design
considerations of low pass filter, high pass filter,
notch filter are elaborated separately for removal of
noise signals from ECG using FIR R filter. For ECG de
de-
noisingg the digital filter is designed with various The experimental setup of the system which includes
orders. Output of this various designs are simulated FPGA board, ECG sensor, Arduino&
Arduino Bluetooth, the
and analyzed with the help of performance parameters ECG waveform display on monitormon and Android
like SNR and MSE. applicatonis
is given in below figure:

B] HARDWARE IMPLEMENTATION
The implementation of the design is fundamental to
move from software to a hardware realization. The
implementation technique exploits the parallel
structure and the versatility of Field Programmable
Gate Arrays (FPGAs) in order to yield a high
performance and yet low cost implementation.

The step by step procedures that are used to complete


this project are highlighted. They are explained in Fig 6-Experimental
Experimental setup of system
below. The explanations are important in complete the
objectives of the project. As shown in below flow VI. RESULT AND DISCUSSION
chart, the implementation procedure will be followed. I] Matlab Simulation Result
The board at transmitter side will be FPGA board and Clean ECG is taken from MIT-BIH
MIT which is added
at the receiver side the board will be Bluetooth model with low frequency noise. Simulation was done on
and Arduino in this digital hardware system. SIMULINK available with MATLAB. Fig.7shows
pure ECG from MIT-BIH BIH Arrhythmia database of
patient shows noisy ECG. Fig.8shows
Fig. the application
of different FIR filter methods on low frequency
EMG noise. Equation (1) is used to calculate SNR of
filtered signal and mean square error (MSE) from
equation (2) is calculated.

N∑ (ECG raw) 2SNR= 10 log10 i=0

N∑ (ECG raw – ECG filtered) 2 i= ----- (1)

N∑ (ECG raw – ECG filtered)2 MSE = i=0


Fig5-Experimental setup of telemetry
try system N -----(2)

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 221
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
B] Hardware Implementation Results
Result
I] Waveform Analysis-
The output waveforms generated by FIR filter give
better filtered output. The output waveforms are
generated with normal ECG signal,
signal filtered high pass
signal, filtered low pass signal and peak detection as
shown in fig.9 below

Fig 7- Noisy ECG Signal

Fig.9- De-noising
noising ECG Signal by FIR Filter by using
Spartan Hardware kit.
Fig 8-Filtered
Filtered ECG Signal
When the real time ECG signal display on the PC then
at time it display on the android application. The ECG
Types and ECG SNR Mean Square
output waveforms display on the Android phone as
order of the Sample in dB Error (MSE)
shown in fig. 10 –
FIR Filter
Blackman low
Ecg24 19.21 25.67
pass filter (100)
Hamming low
Ecg24 19.20 25.67
pass filter (100)
Hanning low
Ecg24 19.21 25.67
pass filter (100)
Kaiser low pass
Ecg24 18.67 25.42
filter (100)
Rectangular low Fig. 10- ECG waveform display on
o Android
Ecg24 19.11 25.62 applicaton.
pass filter (100)
Table 1 - Performances and evaluation of FIR low
pass filters on MIT-BIH
BIH database. II] DEVICE UTILIZATION SUMMARY
The value shows SNR in dB. AND POWER ANALYSIS-
ANALYSIS
a) Device Utilization-
In Table 1 comparison of different FIR method for After implementation on Spartan FPGA kit,
SNR and MSE is given. Here it can be concluded that utilization of resources i.e. no of LUTs, Slices,
FIR method effectively remove the EMG nois noise Bonded I/O has been calculated for different types of
without distortion of the ECG signal. In case of FIR filter. Fig. 11- shows that Equiripple requires
Equiripple, Hamming, Hanning order of the filter was highest no. of slices, flip-flop,4
flop,4 input LUTs, bonded
increased which increases the complexity of filter IOB with 1GCLK. In window method Kaiser has
designing. So that SNR of FIR filter with Kaiser is minimum utilization of resources. As it require 87
(18.67) less than other type of filter. In case oof other slices, 156 slices, Flip flop, 111 LUTs, 17 bonded
method the order of the filter easily grow very much IOB, with 1 GCLK.
.It increases the number of filter coefficient which
leads to the large memory requirement and problems
in hardware implementation.

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 222
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
VIII. FUTURE SCOPE
The implementation of FIR High Pass filter
techniques gives better results on higher end FPGA
devices like Virtex, which are having more resources
and facilities than Spartan devices. This method can
be successfully applied with the cascading of IIR and
FIR filters for real time ECG systems. Finally, it is
seen that an optimum combination can be
Fig.11- Device Utilization Report of FIR Filter implemented using FPGA technique.

b) Power analysis- IX.REFERANCES


Power analysis will generate the timing, ISE report 1. Yongming Yang I, Xiaobo Huang 1, Xinghuo Yu,
and power utilization report for the particular design. “Real-Time
Time ECG Monitoring System Based on
Snapshots of Power Analyzer for FIR filter shown in FPGA ”, The 33rd Annual Conference of the
Fig. 12- IEEE Industrial Electronics Society (IECON)
Nov. 5-8, 2007.
2. Borromeo S, Rodriguez-Sanchez
Sanchez C, Machado F,
Hernandez-Tamames JA, A, de la Prieta R. “A
Reconfigurable, Wearable, Wireless ECG System
”, the 29th Annual International Conference of the
IEEE EMBS Cité Internationale, Lyon, France
August 23-26, 2007.
Fig.12- Device Utilization
tilization Report of FIR Filter 3. Brian A. Walker#, Ahsan H. Khandoker#, and Jim
Black, “Low Cost ECG Monitor
Monit for Developing
VII. CONCLUSON Countries” , 2009
In this research paper, the implementation of FIR 4. ndrej Krejcar, Dalibor Janckulik, Leona Motalova,
filter on FPGA has been introduced. It has been Karel Musi, “Real
Real Time Processing of ECG
investigated that, the improvement of noisy ECG Signal on Mobile Embedded Monitoring Stations
signals can be achieved by combination of FIR filters. ”, Second International Conference on Computer
In order to measure the performance of de de-noising, Engineering and Applications, 2010.
SNR of processed ECG is calculated and MSE was
determined to find the degree of mismatch between 5. Ankit J. Gordhandas and Thomas Heldt and
noisy ECG and filtered. The designed FIR filter with George C. Verghese, “Real
Real-Time Extraction and
Kaiser Window works excellent in removing High Analysis of Key Morphological Features in the
frequency interference from noisy ECG conditions Electrocardiogram, for Data Compression and
utilizing low order of filter. Utilization of resources Clinical Decision Support”, AAAI 2011.
using FIR filter has been minimizedi. e. less power 6. Abhilasha M. Patel, Pankaj K. Gakare, A. N.
consumption (74mw)and and minted the quality of ECG Cheeran, “Real
Real Time ECG Feature Extraction and
signal. Arrhythmia Detection on a Mobile Platform”,
Platform
International Journal of Computer Applications
This method is more applicable in variou various (0975 – 8887) Volume 44– – No.23, April 2012
applications of ECG monitoring. Use of optimization
technique using FPGA platform will improve the 7. Sonal K. Jagtap, Research Student,
Student “A Real Time
quality and processing time. FPGA are finding wide Approach: ECG Noise Reduction
Red in Chebyshev
acceptance in medical system. For their ability for Type II Digital Filter”, International Journal of
rapid prototyping of a concept that requires Computer Applications (0975 – 8887) Volume
hardware/software co-design,
design, for performing custom 49– No.9, July 2012.
processing in parallel at high data rates. 8. Sayanti Chattopadhyay1, Susmita Das2, Avishek
Nag3, Jayanta Kumar Ray4, Soumyendu
Bhattacharjee5, Dr. Biswarup Neogi, “Design
“Des and
Simulation Approach Introduced to ECG Peak

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 223
International Journal of Trend in Scientific Research and Development (IJTSRD) ISSN: 2456-6470
2456
Detection with Study on Different Cardiovascular Stefanizzi, and Luciano Tarricone, “An IoT- IoT
Diseases”, December 2012. Aware Architecture for Smart Healthcare Systems
”, IEEE Internet of Things Journal, DOI
9. Masudul Haider Imtiaz & Md. Adnan Kiber,
10.1109/JIOT.2015.
“Design and Implementation of a Real
Real-Time
Automated ECG 13. Sudip Deb, Sheikh Md. Rabiul Islam, Jannatul
Diagnosis (AED) System”, 2013 RobaiatMou, Md. Tariqul Islam,” Design and
Implementation of Low Cost ECG Monitoring
10. Bhushan N. Patil, “A
A Review of ECG Monitoring
System for the Patient using
u Smart Device”, ,
System Using Wavelet Transform” International
February 16-18,
18, 2017Feb 2017.
Journal of Research in Advent Technology, Vol.2,
No.4, April 2014. 14. Udit Satija, Student Member, IEEE, Barathram.
Ramkumar, and M. Sabarimalai Manikandan,
11. Prerana N Gawale1, A N Cheeran2, Nidhi G
Member, IEEE “Real-Time
Time Signal Quality-Aware
Quality
Sharma, “Android
Android Application for Ambulant ECG
ECG Telemetry System for IoT-Based
IoT Health
Monitoring”, International Journal of Advanced
Care Monitoring” IEEE Internet of Things
Research in Computer and Communication
Journal, DOI 10.1109/JIOT.2017.
Engineering
Vol. 3, Issue 5, May 2014. 15. R.Harini1, B. Rama Murthy2 , K. Tanveer Alam,
“DEVELOPMENT
OPMENT OF ECG MONITORING
12. Luca Catarinucci, Danilo De Donno, Luca
SYSTEM USING ANDROID APP ”, March
Mainetti, Luca Palano, Luigi Patrono, Maria
2017.
Laura

@ IJTSRD | Available Online @ www.ijtsrd.com | Volume – 3 | Issue – 1 | Nov-Dec


Dec 2018 Page: 224