Edit User Factors

Setting Mining or Mining Plus User Factors

Example: A group of samples was homogenized and split. A portion of each sample was sent to an outside lab for
analysis. The other portion was analyzed with the Innov-X analyzer.

The results of analysis prior to calibration are as follows:

Analyser Cu LAB Cu
Std 1 4624 6370
Std 1 4368 6370
Std 1 4576 6370
Std 1 4375 6370
Std 1 4451 6370
Std 1 4420 6370
Std 1 4302 6370
Std 1 4641 6370
Std 2 9624 13806
Std 2 9248 13806
Std 2 9365 13806
Std 2 9913 13806
Std 2 9509 13806
Std 2 9203 13806
Std 2 9517 13806
Std 2 10063 13806
Std 2 9410 13806

Innov-X Africa, PostNet Suite 126, Private Bag X1 Edenglen 1613 RSA CK 1996/050610/23 VAT: 4110236298
Tel.: +27 (11) 453 0023/ +27 (0) 82 881 3152; Fax +27 (86) 556 2124; E-mail: stuart@innovxafrica.co.za
Tel: +27 (0) 83 676 1020; Fax: +27 (86) 684 7819; E-mail: ingrid@innovxafrica.co.za
DELTA
Edit User Factors

PAGE 2 of 5

All the measurements on the Innov-X analyzer are reading too low for Cu. In this case, it is necessary to calculate a
User factor that could be applied to this type of sample.

Procedure:

Plot the data on an XY scatter plot in excel. The Innov-X results must be on the X-axis, the Lab values on the Y-axis.
(*this order is important). Add a trend line and insert the equations

The slope and intercepts for these graphs can be entered directly into the analyzer. In some cases it is sufficient to
enter just a correction for the slope. In others the slope and intercept should be entered.

The software allows a user to enter multiple sets of User factors for different applications, or different ore bodies. A
group of factors is given a name, and then the factors are entered. The factors can then be selected by name.

To enter the User Factors on the Delta:


Home
Mode
Mining
Analysis
Tools
More
User Factor

Innov-X Africa, PostNet Suite 126, Private Bag X1 Edenglen 1613 RSA CK 1996/050610/23 VAT: 4110236298
Tel.: +27 (11) 453 0023/ +27 (0) 82 881 3152; Fax +27 (86) 556 2124; E-mail: stuart@innovxafrica.co.za
Tel: +27 (0) 83 676 1020; Fax: +27 (86) 684 7819; E-mail: ingrid@innovxafrica.co.za
DELTA
Edit User Factors

PAGE 3 of 5

Innov-X Africa, PostNet Suite 126, Private Bag X1 Edenglen 1613 RSA CK 1996/050610/23 VAT: 4110236298
Tel.: +27 (11) 453 0023/ +27 (0) 82 881 3152; Fax +27 (86) 556 2124; E-mail: stuart@innovxafrica.co.za
Tel: +27 (0) 83 676 1020; Fax: +27 (86) 684 7819; E-mail: ingrid@innovxafrica.co.za
DELTA
Edit User Factors

PAGE 4 of 5

Innov-X Africa, PostNet Suite 126, Private Bag X1 Edenglen 1613 RSA CK 1996/050610/23 VAT: 4110236298
Tel.: +27 (11) 453 0023/ +27 (0) 82 881 3152; Fax +27 (86) 556 2124; E-mail: stuart@innovxafrica.co.za
Tel: +27 (0) 83 676 1020; Fax: +27 (86) 684 7819; E-mail: ingrid@innovxafrica.co.za
DELTA
Edit User Factors

PAGE 5 of 5

To select the correct User Factor, follow the procedure above and choose “Select Model” OK
The Selected model name will now appear on the top of the screen. Go back to the measurement screen and
remeasure the samples.

Analyser Cu Lab Cu
Std 1 6670 6370
Std 1 6301 6370
Std 1 6601 6370
Std 1 6311 6370
Std 1 6421 6370
Std 1 6376 6370
Std 1 6206 6370
Std 1 6695 6370
Std 2 13883 13806
Std 2 13340 13806
Std 2 13509 13806
Std 2 14300 13806
Std 2 13717 13806
Std 2 13275 13806
Std 2 13728 13806
Std 2 14516 13806
Std 2 13574 13806

Your analyser is now ready for analysis with the User factor applied.

Should you require any assistance, please do not hesitate to contact our Service department on +27 11 453 0023 or
e-mail ingrid@innovxafrica.co.za

Innov-X Africa, PostNet Suite 126, Private Bag X1 Edenglen 1613 RSA CK 1996/050610/23 VAT: 4110236298
Tel.: +27 (11) 453 0023/ +27 (0) 82 881 3152; Fax +27 (86) 556 2124; E-mail: stuart@innovxafrica.co.za
Tel: +27 (0) 83 676 1020; Fax: +27 (86) 684 7819; E-mail: ingrid@innovxafrica.co.za

You might also like