Professional Documents
Culture Documents
Intro
The calculation algorithms in modern treatment planning system are capable of highly
accurate radiation dose calculations provided they have an accurate dataset to work with. We can
currently reliably achieve dose calculation accuracy that is within 1-2% if a model-based
algorithm is used. The addition of heterogeneity correction in the most popular treatment
planning systems used today help to achieve a more realistic dose calculation even in the
presence of tissue heterogeneities. The most relevant tissues and cavities from a radiation
dosimetry perspective are the ones that radiologically different from water, including lungs, oral
Methods
For this assignment I set out to generate two treatment plans for a solitary lung tumor,
one with heterogeneity correction and one without. The plans are simple, utilizing the lowest
energy available in my clinic, 6MV, and a typical two beam arrangement with one anterior and
one posterior beam. Margin for both beams was set to what the physician would typically select
in my clinic, 5mm in all directions around the PTV. The plans were calculated to isocenter which
is in the PTV with a prescription of 250cGy per fraction for 24 fractions for a total dose of
6000cGy. The primary treatment planning system used in my clinic is RayStation by RaySearch
and not only does it have heterogeneity correction built-in, but it is always on. In a
found to be satisfactory with central axis calculations agreeing with measurements within 2% for
Treatment Planning Project 2
open fields and 3% for wedged fields.2 Since there is no option to turn off heterogeneity
corrections, it was recommended by the staff at my clinic to set the overall density of the external
contour of the patient dataset to the density of water. This will somewhat simulate but not
directly compare the difference between a treatment plan that uses heterogeneity correction to
one that does not. However, it demonstrates how lung tissue affects the dose calculation in a
modern treatment planning system which utilizes an accurate calculation algorithm with
heterogeneity correction.
Both plans are demonstrated in axial, sagittal, and coronal planes in cuts through the
tumor level at isocenter. A plan document was generated for both which demonstrates the beam
sets, a DVH including a pre-defined set of contours. The plan labeled Non-Hetero located in the
top left of each CT figure indicates the plan which has the density of the external contour set to
water. The figures will alternate between each plan for comparisons sake.
As we can see from the beam sets, the monitor units for the plan with heterogeneity corrections
is lower than the plan with the density of the external contour set to water. This is due the
heterogeneity correction accounting for the lower density of the lung tissue. Lung tissue has a
average density of .31 g/cm3, compared to water which has an average density of 1 g/cm3. While
the 95% line of the water density plan at a glance looks like it provides more coverage, we can
Treatment Planning Project 3
see the 100% isodose line curving inwards at the isocenter at midline. The DVH and region of
interest tables later provide more information regarding coverage. The patient’s external contour
has a thickness which is narrower laterally, and the resulting sloping of the contour promotes
Hetero Plan
We can tell from the dose volume histograms here that the coverage for the water density plan is
superior. Average dose, D95, and D99 of the PTV are all higher in the in the water density plan.
This is not realistic since it is not taking into account the lower density of the lung tissue.
Because there is less build-up in the lung for the plan with hetero corrections, only some dose
builds up in the actual tumor when normalized to 100%. If this plan was prescribed as 100% of
the prescription dose to 95% of the PTV, the resulting monitor units would be significantly
higher than they are now. If the axial slice of each plan is compared, note the increased dose
medially and just beyond the lung tissue. This is because there is an increase in dose in soft
As mentioned previously, two other common sites of significant heterogeneity are air
cavities such as the sinuses and bone. The air-tissue interfaces between the cavities and soft
tissue are complex to measure or calculate due to lack of electron equilibrium. Underdosing
effects occur at both the distal and proximal air cavity interfaces.1 However, we know that using
lower energy photon energies will help to improve coverage over higher energies when air or
low-density tissue is present. As for bone, it both reduces the dose to tissue beyond it, but it also
has unique effects for the tissue immediately surrounding it. Beyond 1cm of hard bone we can
particularly applicable to dental implants or prostheses such as hip implants. For example, one
study examined the backscatter dose from dental implants and found it increases up to a
maximum of 53% and is primarily dependent on the physical density and electron density of the
metal crown allow.4 This has been a problem at my clinical site, as the team is still searching for
the most ideal way to deal with the backscatter of metal dental work. It can have detrimental
Treatment Planning Project 8
reactions to the oral mucosa. The effect of high Z hip prostheses were studied in the AAPM Task
Group 63. The figure below indicates the dosimetric effect of various beams on a Co-Cr-Mo slab
at various depths. 1 Note the dose build up at the interface of the slab at beam entry and the
immediate reduction in dose just after the slab. One way to account for a hip prosthesis in the
treatment planning system utilizing heterogeneity corrections is to contour the implant and the
density can be set by the planner. The problem is that the CT number to electron density
Treatment Planning Project 9
conversion factor may not account for the high-Z inhomogeneity and there are also image
artifacts.1
Treatment Planning Project 10
References
2. Mzenda B, Mugabe KV, Sims R, et al. Modeling and dosimetric performance evaluation
of the RayStation treatment planning system. J Appl Clin Med Phys. 2014;15(5):4787.
https://10.1120/jacmp.v15i5.4787
3. Khan FM, Gibbons JP. Khan's The Physics of Radiation Therapy. Lippincott Williams &
Wilkins; 2014.
4. Chang KP, Lin WT, Shiau AC, et al. Dosimetric distribution of the surroundings of
different dental crowns and implants during LINAC photon irradiation. Rad Phys Chem.
2014;104. https://doi.org/10.1016/j.radphyschem.2013.11.026