Sonel SA
Test and Calibration Laboratory
“14 Wokulskiego St.
58-100 Swidnica
Tol: #48 74 86 83 864
Fax: +48 74 85 83 809
CALIBRATION CERTIFICATE
Issue Date: 18.02.2019, Certificate No:119836/14
Device Tested _—_Isolation Resistance Tester MIC-2510 s sero1t
‘echnical Data According to operating manual MIC.2510.
Calibration ‘Sonel SA, 58-100 $widniea, Wokulskiego 11 St.
quested by
Measurement errors According to operating manual MIC-2510,
requirements
Calibration carted Sone! SA, Test and Calibration Laberatory
uti
Methodology snd Comparison using cirect measurement method. Testing methods are outined in Intemal Standard
extent of testing ZN-03/SONEL046, fst eeu publishes Apel 2041
Calibration Date 18.02.2019
References to official Reasings ofthe calbrated device were refered to:
‘metrological “calibrator CALMET C101 sin: 10004,
‘Standards voltage probe KEITHLEY 40KV sin: G17AT,
“multimeter AGILENT model No 34401” In: USG6O72148,
“resistance calibrator SONEL SRP-S0K0 -STO sin: 470023,
+ standard decade resistor ZELAP 001-0108 ev 94/2006,
Ambient conditions Tests were carried out in ambient temperature of (2382)"C and relative humidity of (40210)%.
Measurement Expanded uncertainty ofthe measurements performed, atthe 95% level of confidence and a coverage factor
uncertainty ‘established in accordance with recommendations oulined in EA-402 EU Expression of the Uncertainty
of Measurement in Calibration’ Paish edition published in 2001
Results of the ‘The instrument meets the manufacturer's declared functional and metrological
calibration: ‘specifications atthe tested points. The results have been presented on pages
from 2/3 t0 3/3.
‘Approved by R. Pankjewicz
er o.2510/1905