You are on page 1of 8

The YELLOW JACKET Anememoter

/ Thermometer 68915 is a versatile


instrument that measures air velocity

YELLOW JACKET ® (FPM - feet per minute; or MPS -


meters per second), air flow volume
(CFM - cubic feet per minute; or CMS
- cubic meters per second) and
Anemometer/Thermometer temperature. Temperature is read at
the same time as velocity or flow
FPM, CFM and temperature readings with one hand-held instrument measurements in your choice of °F or
°C. The unit eliminates analog charts
that are hard to read and calculate.

Features include:
• Microprocessor controller
• Minimum and maximum
recordings
• Averaging of FPM, MPS or CMS
multipoint readings
• “Hold” button to freeze LCD
reading
• Switchable imperial to metric
• Automatic power shut-off after 20
minutes unless set for continuous
operation in non-sleep mode
• RS-232 output

Specific applications in the HVAC


industry include environmental paint
booths, air balancing, and many others
where measurement of air velocity,
Operation UPC No. flow and temperature are essential.

and Maintenance
686800 68915 Check for damage immediately.

Carton contents include:


1 pc. hand-held instrument
1 pc. vane type probe head with cable
1 pc. 9-volt alkaline battery
1 pc. maintenance & operation
instructions
1 pc. carrying case
Before starting When you turn on the
unit, an “all functions” This button provides
1. Remove battery compartment display (fig. 1) will several functions:
ON/OFF appear on the LCD, HOLD
screw and lid. Insert 9 volt RESET ADVANCE
alkaline battery. and then automatically • Freezes a reading in
change to the air velocity/ tempera- the LCD.
2. Make sure the unit is off before ture display (fig. 2). Turning off the
attaching the probe. Align the • Stores readings for minimum,
unit will erase any air velocity
probe connector key to the maximum, and minimum/
(FPM) or flow volume (CFM)
maximum averages for FPM.
keyway in the unit and insert readings. The unit will reset to 0000
the connector. velocity or flow volume. The “Free • Increases the numerical value of
Area” number (see below) will not each individual digit in the “Free
NOTE: The unit must be off before change from the four digit number Area” number. For example,
replacing the battery. you’ve entered and saved. pressing the button will change the
vel FLOW AREA Hold Ph Knots vel
number from 1.111 to 2.111;
Button functions
8.8.8.8 0000
ft/m ft²
m/s m²
ft/m pressing again, 3.111. Each digit
mil/h clm
km/h cms can be increased up to 9 before
-1.8.8.8. 70.2
MIN REC °C
returning to “0.” To increase the
MAX AVG °F °F
value of the second or other digit,
YELLOW JACKET Fig. 1 Fig. 2 use the Record Single Point button
Pressing this button changes the (see below) to select the digit you
display from air velocity (FPM) to want to change.
“Free Area” • Changes Baud Rate for RS-232
number to flow output.
volume (CFM),
and back to
DIGITAL ANEMOMETER MODE(SELECT) velocity (FPM).
Air velocity is the
speed of air MIN/MAX- This button provides
• VEL (FPM) RECORD
movement in SINGLE POINT several functions:
• FREE AREA
• FLOW (CFM) FPM (ft/m) or
HOLD
ON/OFF
RESET MODE(SELECT) ADVANCE
MPS (m/s). Air • Displays continuous moving
flow is the volume of air in CFM average in FPM.
(ft³) or CMS (cm³). “Free Area” • Displays average, minimum, and
number is the open area in the duct maximum FPM readings for singe
• VEL (FPM)
MIN/MAX- • FREE AREA NEXT DIGIT→ or grill in square feet. When measur-
RECORD MULTI-POINT and multipoint measurements.
• FLOW (CFM) ing through a diffuser you must
SINGLE POINT AVERAGE
subtract the amount of diffuser • Saves changes made to “Free Area”
blockage. The “Free Area” default is number, imperial to metric switch-
68915 1.111 (ft²) which is equivalent to a ing, and Baud Rate.
10” x 16” duct. You can change this
number (see page 5) to different
dimensions if necessary.

2
Specifications
Air flow ranges: 125 - 4900 ft/min
This button provides
NEXT DIGIT→ (0.7 - 25 m/sec)
MULTI-POINT several functions:
Flow resolution: 1 (0.01)
AVERAGE • Selects individual
digits in the “Free Area” Flow accuracy: ±2%
number for which you want to Temperature: 14 to 122°F
change the value. For example, in (-10 to 50°C)
1.111, with this button you can Temp. resolution: .01 (0.1)
select the second “1.” Using the Temp. accuracy: ±1.0°F (±0.6°C)
HOLD / ADVANCE button, you
can then change the second digit to Power: 9 volt/100 hour
1.211 or higher (see HOLD / Fan diameter: 2-7/8” (70mm)
ADVANCE above).
Measuring Air Velocity (FPM)
• Displays multipoint averages for
FPM, and the number of points (See changing to metric on page 6)
measured.
SINGLE POINT FPM
Auto Power Off
1. Turn unit on by pressing ON/OFF
If powered and unused, the unit will button.
turn off after 20 minutes to save the
battery. Auto shut-off wull be 2. Unit is ready when LCD
proceeded by three beeps. To restart, shows “vel” in upper left and
simply press the ON button. temperature vel
at lower right
0000
ft/m

Non-Sleep Mode (fig. 2).


To bypass “Auto Power Off” for 70.2 °F

continuous operation follow these


steps whn turning the unit on:
3. Hold vane probe head in
1. Press the ON and HOLD buttons front of the air source for a
at the same time. measurement vel

0080
2. Release only the ON button. in FPM ft/m

3. After an “n” appears on the LCD, (fig. 3).


release the HOLD button. 72.3 °F

The 68915 will remain on until the Fig. 3


OFF button is pressed.

UPC# Description
68915 Anemometer
68916 Replacement 2.75 (70mm)
probe assembly
3
Continuous Moving Average Multipoint FPM
1. Turn unit on. 1. Turn unit on.
2. Place vane probe head in front of 2. Place vane probe head at first point
air flow source. to be measured.
3. Press MIN/MAX RECORD button. 3. After you see the reading in the
The unit makes a reading once a LCD, press HOLD/ADVANCE.
second and displays a moving Release the button when you hear
average for up to two hours. a single beep. The LCD will show
HOLD above vel Hold
MIN/MAX/AVG the reading
0075
ft/m

1. Turn unit on. (fig. 4).


2. Place vane probe head in front of Fig. 4
72.3 °F

air flow source.


3. Press MIN/MAX button and the 4. Press MIN/MAX RECORD.
unit will begin recording. Release the button when you hear
4. Press HOLD button to store a single beep. The LCD will show
readings before moving the vane a number from 1-8 representing the
probe head away from the air reading point that was saved
flow source. (fig. 5). The unit will return
5. Press MIN/MAX once to see AVG automatically to the “vel” display
velocity readings. ready for the vel Hold

0075
next reading. ft/m
6. Press MIN/MAX again to see MIN

1
velocity readings.
°F
7. Press MIN/MAX again to see MAX Fig. 5
velocity readings.
8. Press MIN/MAX again to return to
the velocity reading you stored with 5. Repeat steps 2-4 until you’ve
the HOLD button. measured and recorded all desired
9. Press ON/OFF RESET to clear the points up to 8.
MIN/MAX average readings. 6. Press NEXT DIGIT/MULTIPOINT
button to view the average air
NOTE: At any time you can use velocity and the number of points
MODE (SELECT) to switch from measured vel Hold
FPM velocity to CFM flow readings.
0068
(fig. 6). ft/m

Before switching, be sure the “Free


Area” number matches the open square
footage of the grill or duct you are Fig. 6 AVG 8
measuring (see page 5 for “Setting
Free Area Number”).

4
Measuring air flow (CFM) 7. When the flashing stops, press the
SETTING “FREE AREA” HOLD button to save the changed
“Free Area” number. The unit will
NUMBER
go to the CFM measurement mode
(fig. 9).
To measure CFM, the “Free Area” FLOW

0000
number in the unit is equal to the
amount of open area in the duct or clm

grill you are taking readings from.


To calculate this number, go to Fig. 9
68.9 °F

Appendix A.

If you do not need to change the APPENDIX A


number, press MODE/(SELECT) to VOLUME AIRFLOW
go into the CFM measurement mode CALCULATIONS
(fig. 7). FLOW

0000 clm
The “Free Area” is equal to the cross
sectional area of the duct or diffuser,
Fig. 7
68.9 °F
minus the amount of blockage. In
rectangular duct work, this cross
sectional area is the length times the
To change the number, take the follow- width. In a round duct, the area is the
ing steps: radius squared times (3.14).
1. Turn the unit on.
2. Press MODE SELECT once. You i.e. (LxW=Area), (RxRx3.14=Area)
will hear one beep and see AREA
and 1.111 with ↑
a flashing digit ↓ AREA ↓ ↓
1.111
ft²

(fig. 8).
↑ To convert square inches to square
feet, divide by 144 inches / ft².
Fig. 8

3. To increase the value of the flashing EXAMPLE:


digit from 1 to 2 or higher (up to 9), If an air duct is rectangular and the
press HOLD/ADVANCE. Press height is 24 inches with a length of 12
once for each increase. inches, the air velocity reading in the
duct was 450 FPM.
4. To make the next digit to the right
flash, press the NEXT DIGIT H x L =A
button. When the next digit is A (ft²) x FPM = CFM
flashing, repeat step 3. 24 inches x 12 inches = 288 sq. in.
5. Repeat steps 3 and 4 until all digits
288 sq. in. ÷144 in. / ft² =
match the desired number.
2 square feet (ft²)
6. Press MIN/MAX RECORD button.
5
If grill blockage is 20%, then multiply SINGLE AND
2 ft² times .9 to get a “FREE AREA” MULTIPOINT CFM
of 1.8 (ft²).
With the unit in CFM mode, operation
450 FPM x 1.8 “FREE AREA” = for single and multipoint readings and
810 CFM averages is the same as for FPM (see
page 4).
EXAMPLE: An air duct is round
and the diameter is 12 inches. The Changing from Imperial to
air velocity reading in the duct is Metric
450 FPM:
To change unit readings from feet/min
R x R x 3.14 = A and °F to meters/sec and °C, follow
A (ft²) x FPM = CFM these steps:
1. Turn the unit on by pressing the
6 inches x 6 inches x 3.14 = 113 sq. in. ON/OFF and NEXT DIGIT /
AVERAGE buttons at the same
113 sq. in. ÷ 144 inches / ft² = time.
0.785 sq. ft. (ft²) 2. Release the ON button first.
3. When the LCD shows a small
If grill blockage is 35%, then multiply ft/m, m/s, °F and °C, release the
0.785 ft² times .65 to get a “FREE AVERAGE button. A small ft/m
AREA” of .510 (ft²). and °F will remain on the LCD.
4. Press HOLD to change to m/s
450 FPM x .510 “FREE AREA” =
and °C.
229.5 CFM
5. Press RECORD button and a “5”
will appear on the LCD.
6. Press HOLD/ADVANCE and a
Efficiency Tip Baud Rate of 1200 or 2400 will
You do not need the correct “Free appear (see section for
Area” number in the 68915 to changing Baud Rate).
measure air velocity (FPM). But 7. Press RECORD and a “5” will
by setting the “Free Area” appear on the LCD.
number before measuring (FPM), 8. Press HOLD to save setting and
you can increase efficiency. With hold until the “vel” display appears
the number set as the first step in and the unit is ready for m/s and
the measurent process, you can °C reading.
change back and forth from FPM
to CFM by simply press the
MODE (SELECT) button during
any function.

6
Changing from
Metric
3. When the LCD shows a small
to Imperial
ft/m, m/s, ˚F and ˚C, release the
To change unit readings from NEXT DIGIT/AVERAGE button.
meters/sec and ˚C to feet/min and ˚F, A small fm/s and ˚C or ft/m and
follow these steps: ˚F will remain on the LCD.
1. Turn the unit on by pressing the 4. Press RECORD button and a “5”
ON/OFF and NEXT DIGIT/ will appear on the LCD.
AVERAGE buttons at the 5. Press HOLD/ADVANCE and a
same time. Baud Rate of 1200 or 2400 will
2. Release the ON button first. appear. To change from 1200 to
3. When the LCD shows a small 2400, press NEXT DIGIT. To
ft/m, m/s, ˚F and ˚C, release the change from 2400 to 1200 press
NEXT DIGIT/ AVERAGE button. HOLD/ADVANCE.
A small m/s and ˚C will remain on 6. Press RECORD and a “5” will
the LCD. appear on the LCD.
4. Press the NEXT DIGIT to change 7. Press HOLD to save settings and
to ft/m and ˚F. hold until the “vel” display
5. Press RECORD button and a “5” appears and the unit is ready for
will appear on the LCD. a reading.
6. Press HOLD/ADVANCE and a 8. Press NEXT DIGIT to return to
Baud Rate of 1200 or 2400 will the “vel” display ready for
appear (see section for a reading.
changing Baud Rate.)
7. Press RECORD and a “5” will
Troubleshooting
appear on the LCD. LOW BATTERY
8. Press HOLD to save settings and When you see the low battery indicator,
hold until the “vel” display turn off the unit and replace the 9 volt
appears and the unit is ready for alkaline battery. Do not replace the
ft/m and ˚F reading. batter when the unit is on.
E6
Changing Baud Rate for When you see E6 in the LCD, the probe
RS-232 Output cable is improperly connected or discon-
nected from the meter body.
1. Turn the unit on by pressing the
ON/OFF and NEXT DIGIT/
AVERAGE buttons at the
B
same time.
2. Release the ON button first.

7
Warranty
YELLOW JACKET® products from YELLOW JACKET
Ritchie Engineering Company, Inc. Products Division
are guaranteed to be free from mate- Ritchie Engineering Company, Inc.
rial and workmanship defects for a 10950 Hampshire Avenue South
period of one year. This guarantee Bloomington, MN 55438-2623
does not cover products that have Phone: 952-943-1333
been altered, abused, misused or or 800-769-8370
improperly maintained. Fax: 952-943-1605
or 800-322-8684
All returns MUST have an RGA
number and proof of purchase/copy of Some specialty products are not
invoice. Please obtain an RGA stocked and thus not eligible for
number from Ritchie Engineering return. If the product is found to be
Customer Service by calling defective, Ritchie Engineering will
(800)769-8370. All electronic prod- repair, replace or issue a credit at our
ucts must be returned to Ritchie discretion.
Engineering prepaid.

P/N 405926_A

You might also like