5 BADAN TENAGA NUKLIR NASIONAL V KAN
PUSAT TEKNOLOGI KESELAMATAN DAN METROLOGI RADIASI ante esas Nasal
rf Jalan Lebak Bulus Raya No. 49, Jakarta 12440 Laboratorium Kabat
Kotak Pos 7043 JKSKL, Jakarta 12070 LK-142-10N
batan — Telp : +62-21-7513906 (Hunting), Fax : +62-21-7657950, E-mail : ptkmr@batan.go.id tsonec 17025; 2008
SERTIFIKAT KALIBRASI
CALIBRATION CERTIFICATE
No. 2460 / KN 04 02/ KMR 5.1/ 05 / 2018
IDENTITAS PEMILIK / OWNER IDENTITY:
Nama / Name : PT. Terra Samudra Eng
Jl, Cibereum Ruko Pekopen No.3 RT 02 RW 02 Lambang Jaya
eres * Tambun Selatan Bekasi
No. Permohonan / Request No. : 0857/KN 04 02/ KMR 5.1/05/ 2018
Diterima / Accepted : 07 Mei 2018 / May 7, 2018
ALAT UKUR RADIASI YANG DIKALIBRASI / INSTRUMENT CALIBRATED:
Nama / Jenis Alat Dosimeter Saku Gamma
Name / Type of instrument Gamma Pocket Dosimeter
Tipe & No.Seri Elektrometer
Electrometer Type & Serial No. Soe a ea
Tipe & No. Seri Detektor a .
Detector Type & Serial No.
Fabrikan / Manufacturer Arrow-Tech
ALAT UKUR RADIASI STANDAR / STANDARD RADIATION INSTRUMENT :
Tipe & No. Seri Elektrometer F
Ekchamotor Type & Sarat No : Keithley 6517B # 1331921
Tipe & No. Seri Detektor :
Detector Type & Serial No : Exradin A6 # XQ100534
Ketertelusuran / Traceability + Slmelalui LDSS / Si through SDL
No. Sertifikat / Certificate No. : LDSS/KAL-INT/II-17/02
METODE KALIBRASI / CALIBRATION METHOD :
Kalibrasi telah dilakukan pada tanggal 23 Mei 2018 dengan metode substitusi (AEA Safely Reports
Series No. 16, 2000). Detektor disinari di medan radiasi gamma '*"Cs
The calibration has been performed on May 23, 2018 by the substitution method (IAEA Safety Reports Series No. 16,
2000). Detector was exposed in a gamma radiation field of "Cs.
Kondisi Kalibrasi / Calibration Condition : Terlampir / Attached
Hasil Kalibrasi / Calibration Result Terlampir / Attached
‘Tanggal Dikeluarkan / Date of Issue 25 Mei 2018 / May 25, 2018
Sub Bidang Kalibrasi Alat Ukur Radiasi
Dra. Nur Rohmah
NIP. 19620421 198901 2 001
FM 001 SOP 016 003/KN 04 027KMR 5.1 (September 2017) Hal.t dari Page 1 of 2. BADAN TENAGA NUKLIR NASIONAL V KAN
PUSAT TEKNOLOGI KESELAMATAN DAN METROLOGI RADIASI omit aiuotas nasonal
Nf Jalan Lebak Bulus Raya No. 49, Jakarta 12440 Lebornerian Kabéren
Kotak Pos 7043 JKSKL, Jakarta 12070 LK. 142-10 N
batan ~ Telp : +62-21-7513906 (Hunting), Fax : +62-21-7657950, E-mail : ptkmr@batan.go.id isonec 17025 : 2005
LAMPIRAN / ATTACHMENT
No. 2460/ KN 04 02/ KMR 5.1 / 05/2018
Tanggal / Date: 25 Mei 2018 / May 25, 2018
KONDISI KALIBRASI to
CALIBRATION CONDITION :
Tekanan / Suhu / Kelembaban :
Pressure/Temperature/Humidity Fe eee
Posisidetektor ._ Tegak lurus pada sumbu berkas radiasi
Detector Positioning * Perpendicular to the radiation beam axis
Medan Radiasi . Terkolimasi
Radiation Field * Collimated:
Jarak sumber — detektor : -
Source to detector distance (SDD) pelegen
Laju dosis ekiv.ambien/kerma udara/paparan standar—. 494.5 mR
Standard ambient dose equivalent/air kerma/exposure rate
HASIL KALIBRASI :
CALIBRATION RESULT .
EEE =
9 | Ketidakpastian
Rentang / Skala | Eaktor Kalibra
Range / Scale | Calibration Factor Coa
|
(0-200) mR 1,05 65
+ Faktor kalibrasi harus dikalikan dengan bacaan alat ukur radiasi.
The Calibration factor must be multiplied by the radiation measuring instrument readings
+ Ketidakpastian bentangan pada faktor kalibrasi dalam sertifikat ini diperoleh dari ketidakpastian standar
gabungan dikalikan dengan faktor cakupan k=2, dengan tingkat kepercayaan 95 %.
The expanded uncertainty of calibration factor in this certificate is based on the combined standard uncertainty
‘muttiplied by a coverage factor of k=2, which provides a confidence level approximately 95%.
CATATAN / REMARKS :
Berdasarkan Pasal 11 Peraturan Kepala BAPETEN No.1/2006, kalibrasi ini dilakukan secara berkala
sekurang-kurangnya sekali dalam 1 (satu) tahun.
Based on article of the BAPETEN Chairman Decree No. 1/2006, this calibration should pe performed periodically at
least once in a year.
Sub Bidang | Kalibrasi Alat Ukur Radiasi
g r]
ra. Nur Rohmah
NIP. 19620421 198901 2.001
FM 001 SOP 016.003/KN 04 02/KMR 5.1 (September 2017) Hal.2 dari 2/Page 2 of 2