MULTISIZE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Modo1No, 46482714P12 and P13
REV 0 DIRECTION 46017547
TABLE OF CONTENTS
CHAPTER TITLE
INTRODUCTION
OPERATING MANUAL
PREINSTALLATION
SERVICE MANUAL
SCHEMATICS
DIAGNOSTICS
RENEWAL PARTS
REV 0 DIRECTION 46-017547
%
1 ,
CHAPTER 1 = INTRODUCTION
cj
P
,' ' -pr
FEATURES
0 The Cassette.Holder plpvides ,counte$alanced support for metnc cassettes, ~
The Cassette Holder provides four choices of operabon either a nght or left
hand model, both of which will accept either metnc or inch cassettes
A Cassette Size Control Lever adjusts the Cassette Holder to accept one of
four different size cassettes, either metnc or inch
1-1
I
s
e Scale
& 1
MULTWEE VERTlCAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 46482714P12 and P13
REV 0 DIRECTION 48017547
TO DETERMINE SID
Partially insert cassette into the Cassette Holder Extend measunng tape
from cohmator to cassette to measure SID (Refer to Automatx Collimator
Operaung Manual for addittonal information )
Insen a gnd into 1/4" (64 mm) wide slot behmd chest plate from nght side
of the Cassette Holder for nght hand model and from left side of the Cassette
Holder for left hand model Shde it 111 slot untd it stops When fully rnsened
gnd edge wdl be recessed approxlmately 1/8" (3 2 mm) in slot
For removal, move the tab on the bottom of the cassette holder front section,
then shde out the gnd
PHOTOTIMING
The Cassette! holder can be used for phototmng wth the rnstallation of the
Ion Chamber Detector The detector works in conjunctton with the x-ray
control The three field pattern on the front panel of the Cassette holder
corresponds to the three sensing areas of the Ion Chamber Detector
RATING PLATE
The Raung Plate for the Model 46-182714P12 Right Hand Cassette Holder
is located in the upper, nght hand comer of the column (see Illustrauon 2-1).
and, for the Model 46-1827 14P13 Left Hand Cassette Holder IS ~ theI I upper,
left hand comer of the column
Before ordenng replacement parts, be sure to make a note of the Model Number
and Send Number on the Rating Plate, as both numbers are essential for proper
idenucatron of a part
2-3
YULMQE VERTICAL CASSETE HOLDER
GE MEDICAL SYSTEMS Mod01 b 6182714P12 .ndP13
REV 0 DIRECTION -01 7547
ANNUAL MAINTENANCE
The followmg checks should be performed every twelve months to insure complete and
proper operatlon of the cassette holder
Release the vemcal lock The toothed lock plunger should completely disengage from
the lock stnp, the vemcal carnage should move freely and smoothly over the entue range
of travel with m m a l side play The camage should r em m posiuon or dnft slowly
upward when released If necessary, add or subtract tnmmer weights untd counterbalance
1s smsfactory
Reengage the vemcal lock The lock plunger teeth should fully engage the lock stnp,
prevenmg further vertlcal motlon If the lock does not hold or bmds, check lock hardware
for rmsahgnment, wear, etc
NOTE The procedun autllned In thb section mqulrss nmoval of tho front covers from
tho cassotto holdw
Visually mspect the counterweight cables for wear, fraying. lunluag, or rust To detect
broken strands, move the cassette holder through its en- range of vemcal travel while
r u m g a soft cloth or cotton along the cables (The enwe length of the cables must be
checked ) If any broken wires are found, irnmednte replacement of both cables 1s
recommended Knked cables and cables with any corrosion or rust at or near an anchor
point should also be replaced Also check for s h y spots whxh indate cable wear
In cases of premature cable wear. check for and ehmmate the cause if possible M i a g n e d
pulleys, for example, may be responsible In such cases the =des of the grooves may show
evidence of rubbing Should cable replacement be made, all pulleys showing excessive
wear should be replaced as well
Check the conditlon of cable anchor points The anchor hardware should be structurally
sound, replace If any structural weaknesses are detected
Finally. lubncate the cables by wiping them wth a cloth saturated wth STP (or slrmlar
lubncant) Apply the lubncant hghtly so that an oil film sjust wible, wipe off any excess
Make sure the poruon of each cable which bears agunst the pulley groove 1s carefully
lubncated SOLVENTS SHOULD NEVER BE USED TO CLEAN THE CABLES
The pulleys on which the counterweight cables nde have a permanently sealed beanng
which should provide lubncation for the life of the equipment Lubncatron 1s not required
as pan of the mamtenance schedule unless it should be mdicated that lubncauon is
inadequate (excessive fnction, squealung, etc ) Should hoccur, a few drops of oil
may be injected as a temporary measure untd the pulleys can be replaced
24
MULTlSQE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 46-182714P12 end P13
REV 0 DIRECTION 46-017547
DATE OF INSPECTION
Use this form in conjunction with PMS work and functional checks Under RECORDED
INFORMATION enter descnptive word such as 'OK' 'ADJUSTED' 'LUBED', or "/A'
if not applicable
a Hardware Tightness
b Extenor Cleanliness
c Paint Touchup
2 Counterweight Cables
a Condition
b Condition of Anchors
c Lubncation
d Pulley Condition
3 Vertical Carnage
B FUNCTIONAL CHECK
2 Vertical Motion
PMS PERFORMED BY
FORM NO 83515
25
U U L r r S m VERTICAL CASSElTE HOLDER
QE MEDlCAL SYSTEMS I l 0 d . l Nor 46182714P12md P13
REV 0 DIRECTION 46017547
2 25'
(5 72 cm
1
I Centerfine of 2' (5 08 an)-
4 0 5'4
26 67 an
k 1 3 12
a*5'
(211 5 cm)
1
-
L-350'
(8 89 cm)
Tradc
i Max. 74'
4 125'
(1OUcm)
i
(188 0 an)
Mn t
i+- (101 6cm)
(5842cm)
I I
I
mters of grawty
carnage in tughest
-4-L-J
I &lowestpostbars
pc
I !.....I-- ........
ILLUSTRATION 3-1
The Cassette Holder can be installed either at the head or foot end of the
table where space permits and where the tube unit can be positioned at the
correct focal distance from the the film plane See IUustmon 3-1 for
dimensions of the cassette holder
MULTCSRE VERTICAL CASSETTE HOLDER
QE MEDICAL SYSTEMS Model N o r 46-182714P12 and P13
REV 0 DIRECTION 6 0 1 7 5 4 7
Where wall structure is not suitable for dmct mountmg, install back plates
dunng wall constructton or mount suitable plates after wall construction to
support the cassette holder For mountmg detads, see Ulustratron 3-2
2 312' --m
(5 9 cm)
B Verttcal centerline
of cassette holder
and cassette in place
4 375'-
(11 lcm
ILLUSTRATION 3-2
3.3
UuLTCsm VERTICAL CASSElTE HOLDER
QE MEDICAL SYSTEMS Model Ne8 46482714P12 .ndP13
REV 0 DIRECTION46017547
For wrmg mformatron, refer to the MIS Maps and MIS Charts for the system
being mtalled Also refer to the site p h n g drawrngs
NOTE
If the Ion Chamber Detector IS to be ~nstalled,allow about 10 (3 0 m)of
slack detector cable between the Cassette Holder and the conduit enuy point
to pemt full vemcal travel
M U L W R E VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 46-182714P12 and P13
REV 0 DIRECTION 46-017547
-
CHAPTER 4 SERVICE MANUAL
TABLE OF CONTENTS
1 INSTALLATION ITEMS 4- 1
2 UNPACKING AND MOUNTING THE COLUMN ASSEMBLY 4-2
4-1
MULTCSDE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 46482714P12 and P13
REV 0 DIRECTION 48017547
-
CHAPTER 4 SERVICE MANUAL
Thls chapter gives the necessary mformmon for instalhng the
MULTI-SIZE Vemcal Cassette Holder
SEC 1 INSTALLATIONITEMS
4-1
UULTWRE VERllCAL CASS€IlE HOLDER
QE MEDtCAL SYSTEMS Model Noa 46-182714P12 and P13
REV 0 DIRECTION -017547
1
I
I
I
I
I
I
I
I
I
I
I
I
ILLUSTRATION 4-1
to.--
2-1 Open long crate
2-2 L~ftthe Column Assembly from its contamer, lay on floor as per
Illustrauon 4-1
2-3 Remove the (8) Column Front Cover screws, carefully remove both
Front Covers Retam screws for later use
2-4 Situate the Column Assembly in the designated posltlon agamst the wall
Note The vemcal centerline of the column coincldes
with the vertical centerhe of the cassette
2-5 Vemcally level the column left to nght, mark on wall the locations
of all eight obround mounting holes
4-2
M U L M U E VERTICAL CASSElTE HOLDER
GE MEDICAL SYSTEMS Model Nor 46-182714P12 and P13
REV 0 DIRECTION 46-017547
-”.
r .” ’ t I
24’
(61 cm)
24
t 24
(61 cm)
Verbal centerlines
of d u m n assembly
and cassette in place
are ccnncidental
K (10 2 cm )
ILLUSTRATION 4-2
Because of the many types of wall constnrdlon, wall fasteners are not Included
Use eight sultable, threaded steel fasteners, such as lag screws or toggle bolts
not less than 1/4“ ( 63 cm) dlameter Each fastener should be secured to a rigid
structure of the wall and be capable of indlviduaiiy supporting at leest 25 Ibs
(11 4 kg ) pull
4-3
YULTlSEE VERTICAL C A S S € f E HOLDER
GE MEDICAL SYSTEMS Model Nor 4&182714P12 and P13
REV 0 DIRECTION 46-017547
SEC 3 INSTALLINGCOUNTERWEIGHTS
I-' Tnmmer
b
Weight
ILLUSTRATION 4-3
4-4
MULTCSQE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model No8 46-182714P12and P13
REV 0 DIRECTION 46017547
' , tig 1' I p4.$
Bottom Bracket 2
ILLUSTRATION 4-4
4-2 Remove all paciung secunng the Front Section dunng shipment
4-5
MULTCSPE VERTICAL CASS€llE HOLDER
GE MEDICAL SYSTEMS Model No8 6182714P12 and P13
REV 0 DIRECTION 46017547
Carnage
Front Section
Column h m b l y
Counterweight Contatner
ILLUSTRATION 4-5
4-6
MULTCSRE VERTlCAL CASSEllE HOLDER
GE MEDICAL SYSTEMS Model 4&182714P12 and P13
REV 0 DIRECTION 46-017547
I1 w
Ion Chamber . Column Top Cover
P2 Female Connector
P2 Male Connector
Ground Stud
wth Nuts
CSS Cable
Ill HI I I /
Ion Chamber Detector
(Before Installatron)
Adhesve Covenng
ILLUSTRATION 4-6
5-1 If an ion chamber detector is not included, remove screws and height
reference scale from package inside chamber cover Rewrap mylar and
replace inside chamber cover Omt steps 2,3,4 and 5
4-7
MULrrSQE VERTICAL CASSElTE HOLDER
GE MEDICAL SYSTEMS Model Nor 4W82714P12 and P13
REV 0 DIRECTION 46-017547
5-3 Shde the detector and mylar into the 112" (1 27 cm) wide opening on
the side of the front s e a o n Do not force the detector Shght in and
out pressure wdl seat detector and mylar in slot When fully inserted,
the detector and mylar wlll project about 1/16" ( 16 cm) from slot
5-4 Route the ion chamber cable through cable clamps no 1 and 2 on the
front sectlon Remove the unused spht snap bushmg from the connector
housing, and fasten the ion chamber cable to unused clamp no 3
5-5 Place the spht bushmg and the ion chamber cable, and resnap into the
connector housing Dress the ion chamber cable to run parallel with
the CSS cable
5-6 Remove the column top cover, set the connector housing in back of
the column top Remove the column cable channel, place over the
cable or cables (if ion chamber is included ) Refasten channel to column
5-7 Insert the P2 female connector of the automatic cohmator cable and the
ground wire through the column top cover Connect the €2connector,
and fasten the ground wire to the ground stud
5-8 Run the ion chamber cable (if included) through the column top cover,
fasten the top cover to the connector housmg and column top
4-8
MULTCSEE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 46-182714P12 and P13
REV 0 DIRECTION 48017547
SEC 6 VERTICALLOCKTEST
Height Reference Scale;-
2% & f l y ,
-
/ -\ 14' x 1 7 (35cm x 43 cm)
Cassette
q/ -Grid
-
/- Back
!4
r (12) Lead Pieces
Front
E%- w
- Hex Nuts
I% Counterweight Container
6-1 Replace the ion chamber lead hned cover on front secuon
(See Illustration 4-7 )
6-2 Insert Gnd and 14" x 17" (35 cm x 43 cm) Cassette
4-9
YULTl-SRE VERTICAL CASSETIE HOLDER
GE MEDICAL SYSTEMS Model No8 46182714P12 and P13
REV 0 DIRECTION 46-017547
6-3 Test the vemcal lock The lock plunger should completely disengage
from the lock stnp, the vemcal carnage suppomng the front secbon
should move freely and smoothly over the entm range of travel
6-4 Reengage the vemcal lock The lock plunger should fully engage the
lock stnp, prevenmg further vemcal mouon
If the lock does not hold or buds, check lock hardware for rmsahgnment
or damage
7-1 Test the counterbalancing with a 14" x 17" (35 cm x 43 cm) Cassette.
a 17" x 17" (43 cm x 43 cm) Gnd and an Ion Chamber Detector
(if included) in place (See Illustration 4-6 ) When properly counter-
balanced the effort to move v e m d y m either dmction should not
exceed 9 lbs (4 0 kg )
-
7-2 IMPORTANT BEFORE ADJUSTING COUNTERWEIGHTS
a Check that vertical lock is fully engaged
b Do not attempt to disengage the vemcal lock and move carnage
while adjusting counterweights
7-3 Adjust counterweights by removing one lead block at a ume from the
contamer Replace tnmmer weights on top of remarning blocks
7-4 When smaller increments of weight adjustment are needed, remove lead
pieces from tnmmer weights (See Illustration 4-7 )
410
MULll-SRE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS Model Nos 6182714P12 and P13
REV 0 DIRECTION 46-017547
NOTE (Refer to Section 8 and 9 when using metnc size cassettes in this cassette
holder and sections 11 through 13 when using inch size cassettes )
8-1 Attach metnc height reference scale (See Illustratron 4-7 ) Set the
Cassette Size Control Lever to 35 cm x 43 cm for metnc cassettes
Adjust the chestplate honzontal centerlme to correspond with the tube
honzontal centerhe Without removmg the baclung from the height
reference scale, temporanly position it so that the top of the front section
corresponds to the height mdicated on the height reference scale of the
tube hanger or tube stand Mark posiuon with a pencil Remove baclung
of the scale and permanently attach to column
9-2 Test the cohmation for each of the 4 film sizes uslng the Colhmator Light
Field, adjust the corresponding potentiometers if required, per Table 4-1
and Illustration 4-8 (Refer to Automatic Colhmator Instruchons to
complete installahon )
TABLE 4-1
CSS CALIBRATION DATA (BASED ON +10 00 VDC APPLIED TO POTENTIOMETERS)
Calibrated Output
Cassettes Size Voltages +VDC (+/- 50 mv)
Centimeters Measured at Cobmator Electronics
24 3 860V
30 5 585V
35 7 164v
43 9 357v
4-1 1
YULMQE VERTICAL CASS€llE HOLDER
GE MEDICAL SYSTEMS Model Nor 46482714P12 and P13
REV 0 DIRECTION 46-017547
Operatmg Instructions
r
b-,Cassette
Vertrcal Lock
size
Control Plate
Cassette Size
Control Lever
"\I
Gnd Ejector Tab
\
ILLUSTRATION 4-8 \
Cassette Holder is supphed with the Chest Plate for Metnc Cassettes in place
4-12
MULTCSIZE VERTICAL CASSETIE HOLDER
GE MEDICAL SYSTEMS Model N o s 46-182714P12 and P13
REV 0 DIRECTION 46017547
. " " J f 1.
10-1 To convert cassette holder from the metnc to the inch cassette system,
remove Ion Chamber lead h e d cover and Ion Chamber if installed
See Illustrations 4-4 and 4-6
10-2 Remove two screws from gnd ejector tab Remove gnd ejector from
its slot
10-3 Remove four slotted screws (two on top, two on bottom) from stamless
front cover Shde off cover with chest plate Reverse Metnc Chest Plate
(The Inch Chest Plate is on the other side of the Metnc Chest Plate )
10-4 Reverse procedure, and replace all the items removed in steps 10-1, 10-2
and 10-3
10-6 Locate the Inch Control Plate attached to the back of the Connector
Houslng (See Illustration 4-6 ) Remove the Metnc Cassette Size Control
Plate by unscrewmg the four socket screws Replace Metnc with Inch
Control Plate with the lever in the 14" x 17" positlon The Inch Plate is
in position on a nght hand unit when the 10" x 12" cassette size is in the
upper nght hand comer of the plate It is in position on a left hand unit
when the 10" x 12" cassette size is in the upper left hand comer of the
plate Replace the Ball Knob
Set the Cassette Size Control Lever to the 14" x 17" position
Adjust the chest plate honzontal centerline to correspond with the tube's
honzontal centerhne Without removing the backmg from the height
reference scale, temporanly position it so that the top of the front section
corresponds to the height indicated on the height reference scale of the
tube hanger or tube stand Mark position with a pencil Remove
backmg of the scale, permanently attach to column
4-13
YULTlslzE VERTICAL CASSEllE HOLDER
GE MEDICAL SYSTEMS Model Noa 46182714P12 and P13
REV 0 DIRECTION 46017547
12-1 The field srnng potenuometers, which control the colhmated size of
the X-Ray field have been factory tested, but need to be re-cahbrated
for the rnch cassettes
Test the cohmation for each of the 4 film sizes using the Colhmator
hght Field, adjust the corresponlng potentiometers d requued, per
Table 4-2 and Illustration 4-8 (Refer to Automauc Colhmator
Ins~uctlonsto complete rnstJlatlon )
TABLE 4-2
CSS CALIBRATION DATA
(BASED ON +10 00 VDC APPLIED TO POTENTIOMETERS)
Calibrated Output
Cassettes Size Voltages +VDC (+/- 50 mv)
Inches Measured at Collmtor Electromcs
10 in 4 240v
11 ln 4 970V
12 111 5 700v
14 in 7 164v
17 in 9 357v
4-14
MULTCSlZE VERTICAL CASSEllE HOLDER
GE MEDICAL SYSTEMS Model Noa 46-182714P12and P13
REV0 - DIRECTION 46-017547
r--------
I
I
I
I
10" x 12"
or
24cmx30cm
Adjustment pots
10 in.
24cm
TRANSVERSE
0 LONGITUDINAL
L -- ---- - -
r-------- .------------
I
I
I
11" x 14"
or
30cmx35cm
30cm.
11 in.
0 1 0 I------------
r----------1 I
35cm iI
14in I
1
I
I
I
I
Adjustment pots I
L--------
I
I
I
I
I I
r-------- .- - - - - - - - - - - - I- - - - - - - - - - -
I
I 14" x 17" I
I
or I
I I
I 35cmx43cm I
I Adjustment pots I
I I
I
I
I
I
I I
I 17" x 14" I
I or I
I
II 43crnx~cm I
I Adjustment pots I
ILLUSTRATION 4-8
4-15
M U L M U E VERTICAL CASSETE HOLDER
GE MEDICAL SYSTEMS Model Nos 46-182714P12 and P13
REV 0 DIRECTION 46-017547
-
CHAPTER 5 SCHEMATICS
51
M U L W Q E VERTICAL CASSEITE HOLDER
GE MEDICAL SYSTEMS Model Nos 46182714P12 and P13
REV 0 DIRECTION 46-017547
CHAPTER 6 DIAGNOSTICS -
DIAGNOSTICSAPPLICABLE TO MODEL NOS 46-182714P12 and 46-182714P13
The potentiometers are actrvated by custom made, light force, highly reliable gold
'crosspoint' contact swdches
a) If the cassette holder Is used infrequently work the control lever several times
through each settmg to selfclean the contacts
d) ll all but one held size are incorrect look for a faulty switch
e) The #7584K83A Quad Pot Switch Plate Assembly has four (4) assemblies of
two (2) 17584K82 pot swtches mounted to the pot switch plate
1) The #7584K84A Electnc Master Plate and PC Board Assembly consists of six
(6) #7584K84AOl pots soldered to one (1) rr7584K84A-02 PC Bead Assembly
which snaps onto the #75841(84AO3 Master Plate
YUL- VERTICAL CASS€ITE HOLDER
OE MEDICAL SYSTEMS Y0d.l Nos 46-182714P12 and P13
REV 0 DIRECTION 48017547
rTEM PART
NUMBER NUMBER DESCRIpnON
1 PM7584K01-C ColumnandC.mageAmwnbly
2 PM7584K02 Cdumn Assembly
3 PM7584K03-C Cdumn Tracks and S w
4 PM7584K04 Column Top Assembly
5 PM7584KW-02 Boanng CaM.Pulley
6 PM7584K05 Cdumn Bartom h s m b l y
7 PM7584K06-C countermlght CeMes (mstdwd rwt)
8 PM7584K07-C Countemecght Roller A8sembly
9 PM7584K08 Carnage Assembly (leascablos and vetl lock)
10 PM7584KOE-01 Carnage Angles hsemMy
11 PM7584K08-02 Ball Beanng Carnage
12 PM7584K09 Verbcal Lock Assembly
13 PM7584K10-C Weight Conmrwr Assembly
14 PM7584K11-C Tnmmer Weight Assembly
15 PM7584K12-C Lead Blocks
16 PM7584K13 Cdunn Front Cover
17 PM7584K14 Column Top Cwer
18 PM7584K15 Verbcal Reference Scale INCH
19 PM7584K16 V O W Reference ScaleMETRlC
20 PM7584K17 Cdumn W e chenml
7-1
MULTCSRE VERTICAL CASSETTE HOLDER
GE MEDICAL SYSTEMS -1 NO8 6182714P12 8nd P13
REV 0 DIRECTION 4&017547
72