Professional Documents
Culture Documents
Manual v1.2
Updated Jan 2008
F. File List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
A. Trolley Panel
Tuning mode
Detector use
B. Ion optics
Trolley
positions
C. Submenus
Figure 2-1: View of Main Tuning Menu.
A. Trolley Panel
Count rate
Radius settings
Selected ion
Deflector plate
tuning
Detector tuning
B. Additional Controls
Tuning mode
Detector use
Ion optics
Trolley positions
???
???
Detector Use
Multi Collection. Uses multiplier tubes after the mass
spectrometer stage (Shown is the selected state for this
button).
Faraday Cups. Use of Faraday cups to detect total ion
current. The options are FCp to detect the primary ion
current at the end of the primary column, FCo to detect
the primary ion current at the sample stage and TIC to
measure the total secondary ion current prior to analysis
in the mass spectrometer stage.
???
Ion Optics
Magnetic Field. Sets magnetic field strength in the mag-
netic sector of the mass spectrometer stage and thus
determine the radii of the secondary ion trajectories.
Enter field strengths numerically in Gauss
Magnetic Field. Alternative input for magnetic field
strength in Gauss. Minimum and maximum for the field
strength are given.
Center Beam. Centers secondary ion beam maximum in
detector if the tuning sub-modes are used.
C. Submenus
Figure 2-4: View of Bottom of Tuning Panel, Containing Buttons for Submenus.
1. Reset Menu
2. Detector Calibration
a) Setup window
b) Display Window
Line used to
select
deflection
voltage.
The line in Figure 2-7 indicates the deflection voltage. You move the line to change the
deflection voltage. When you put your mouse over the window, a new line appears,
which you can move back and forth and place in any location. Currently the line is set
to select for just 14N. If you moved it to the right in the middle of the other plateau,
you d be choosing just 12C.
Figure 2-9: View of Main Mass Table Menu with Three Panels Shown for Edit Mass (a), Assign
Trolleys (b), and Place Trolleys (c).
Mass Table/Real.
6. Energy Menu
a) Energy Tuning
b) Energy Display
8. PHD Menu
a) PHD Conditions
b) PHD Display
In Figure 2-17, move the line to select the threshold voltage. You change the voltage of the
actual detector to shut out noise. The line should always be at the vertex of the curve, as shown in this
example.
9. Tools Menu
Lens to be tuned
Progress bar
b) Tools Display
b) HMR Display
Figure 2-21:View of HMR Display Window. Moving the bar sets the voltage of the deflector or lens
that you re checking.
1. Mapping of
points of mea- Alternate holders:
surement
2. Positioning of
sample holder
BIOLOGY
3. Map of the sam-
ple holder in the
sample chamber
GEOLOGY
4. Position of the primary
beam
5. Sample stage
movement
Figure 3-1: Display Window of Holder Menu. To bring up this window, click on the top edge of the
frame.
???
?????
???
B. Edit Menu
????
????
C. Points Menu
Show
D. Crater Menu
Coordinates of
selected crater
File management
Print data for current crater
Coordinates
Editable comment, file name of the original acquisition
experiment, and the date of the experiment
Figure: 4-1:View of Source Menu. This is an example of a window that requires that the top bar be
blue before you can type in it.
File List. Opens the Sources File List with sets of param-
eter files for the primary ion sources (see Figure 4-2).
Figure: 4-3: Window that appears when you stop/start machine. Enter the delay time in hours and
minutes; if you want to stop/start right away, enter 0.
A. Main Panel
Figure 5-1:The Main Panel of the Preset Menu. In this window the top bar must be blue in order to
type in it.
Same as “save”
B. File Management
Edit Submenu
E. Use of the Preset L1=7000V Toggle Switch for High Spatial Resolution Analysis
To obtain the highest spatial resolution analysis (or smallest Cs Beam diameter), the voltage
on lens L1 needs to be brought from the normal operating value of 0 volts up to a value of
7000 V.
It is not advisable to simply use the thumbwheel to ramp up the L1 voltage as the Cs beam
current will go through a maximum around 3000V. Even a brief exposure to this high Cs
current can damage a sample, in particular if it is an ultrathin/EM section.
The toggle switch will change the L1 voltage from 0 to 7000V. It will also change the values
of Cox, C0y, C1x, C1y to the optimum values corresponding to L1=7000V.
To use the toggle switch, click on the “Preset” button on the NanoSIMS toolbar on the right
of the screen.
When you open the preset menu, a window will open with six rectangular boxes (below).
The top box will be titled “Large Current”. If you click on the rectangular box entitled
“G0_080103”, it will change the L1 voltage to 7000V and also the Cox, C0y, C1x, C1y values.
To switch back to the condition where L1 = 0V, simply click on the rectangular box again.
67
F. File List
69
A. Selection
71
B. File Manager
72
C. Depth Profile
73
74
D. Line Scan
E. Isotopes
Ratios button
b.
a.
c.
Figure 6-6: Menus for Isotopes Menu within Analysis Method Menu. Clicking on the ‘Ratios’ button
in panel ‘b’ brings up panel ‘c’.
76
77
F. Image Acquisition
Species Symbol
78
80
G. Grain Mode
H. Batch Analysis
82
I. Chain Analysis
83
84
(Nb = 1)
(Nb = 2)
85
86
A. Main Menu
87
88
89
B. Results Menu
Figure 7-2: View of Results Menu. You can scroll down this page by using the scroll bar on the right.
90
C. Display Menu
91
92
D. Analytical Parameters
93
94
2D Visualization tool
2D image processing
95
97
C. 2D Processing
98
99
100
101
102
103
A. Holder
104
105
B. Tuning
106
107
C. Keyboard
108
SE FC
Raster
B1
B2X
B2Y
B3
Field (um)
Field DAC
Coef
LF4
LF4 (DAC)
TIC
Polarity
LF2
LF3
SS100 Int
SS100 Ext
CV
109
D. Hardware
110
FCs Plates Deflector plates for Faraday cup after magnetic sector
C7
Options
Source Type of ion sources in the primary column - Cesium car-
bide and oxygen duoplasmatron, or only a cesium carbide
source
Secondary ???
Slits
Selects exit slits for detectors after the mass spectrometer
stage.
Selects size for exit slits
111
112
Slits
Entrance Entrance slit. Shapes the beam between the electrostatic
and magnetic sectors.
Aperture Aperture slit. Shapes beam prior to entry into the mass
spectrometer.
Exit Exit slit. Limits mass range that can enter the mass detec-
tors.
113
F. Detector
114
115
G. B Field
(Truncated)
Figure 10-7: View of ‘B Field’ Tab on Setup Menu
116
H. Directories
(Truncated)
117
I. Sources
(Truncated)
Figure 10-9: View of ‘Sources’ Tab on Setup Menu.
118
119
120
????
121
122
123
Sets status for the turbo pumps TP1 (air lock), TP3 (multi
collection chamber) and TP5 (source column) as well as
the primary pump PM. Toggles between “START” and
“STOP?”
124
????
Initializes lines????
SP authorization?
125
C. Vacuum recorder
126
127
???
???
128
B. Tools
129
C. Periodic Table
Submenus
130
131
132