Calibrating Process Polymat Plus S
ATTENTION: THE REQUIRED INSTRUMENTS FOR A PERFECT OPERATION OF
THE BELOW PROCEDURES ARE:
1. DIGITAL OSCILLATION METER 60MHz;
2. DIGITAL MULTIMETER 4 % ;
3. mAs METER;
ATTENTION: NEVER START THE NEXT STEP WITHOUT CONCLUDED THE
FOREGOING STEP!
1" STEP
1.1- Disconnect three-phase supply of the inverter and insulate the terminals.
1.2. Remove the power supply fuses F16 and F17 of the plate D5-S filament.
1.3. Remove the fuses F7 and F8 of the anode starter supply in the D5-S
plate.
2° STEP
CALIBRATING KV REF. IN THE D2-S BOARD
3.1- Measure with 4 digit multi-meter in the R26 resistor of the D3-S plate and
Select 40 KV on the panel. Adjust the value of 0,992V in R62 (TRIMPOT)
of the D2-S. Grounding point of D2-S is the dissipator of the tension regulator.
3.2 Select 125KV on the panel and adjust 3, 10V in R69 (TRIMPOT) of D2-S.
Tension reference frame for the calibration of the D2-S plate trimpots
kV V. Ref Kv V. Ref,
40 0,992 66 1,637
41 1,017, 70 1,736
42 1,042 [ 73 1,810
44 4,091 77 1,910
46 4,141 81 | 2,009
48 1,190 85 2.108
50. 1.240 90) 2235
52 1,290 96 2.381
55, 1,364 102 2,530
57 1.414 709
60, 1,488 117
63 1,562 [Eeni25)
3™ STEP
INITIAL CALIBRATING OF THE D3-S PLATE2.2. Equip the Cathotest and connect once more the equipment Ifthe
filaments become lighted, the problem will be in the con
2.3. ATTENTION: Equipment disconnected! Verify fuses F16 ae ie in
of the D5-S plate. If they are burned verify if there is some short-ci
the K9 connector of the D3-S plate. ;
24- Repeat 2” step of the service manual
25- ATTENTION: Equipment disconnected! Disconnect the K8 cabling of
the D3-S plate and consider the value of 4,2 ohm in the cable between
K6-1 and KB-2, 4,2ohm between K8-1 and K8-3, 8,4 ohm between KB.
and K8-3.
If you do not find these values, check the filament cable, the connection to
the high tension transformer and the generator terminals.
3. ERROR 003 : BLOCKING OF OVER-CURRENT
3.1- Repeat the 2” step of the service manual.
3.2- Recalibrate the currents.
4. ERROR 004 : FAULT OF COMMUNICATION WITH THE PANEL
4.- Check the paneligenerator interconnection cable.
5, ERROR 005 : BLOCKING OF KV MAXIMUM/OPEN DOOR
5.1- Repeat the 3” step and then the 10 step of the service manual.
5.2- Check the K19 connector of the D6-S plate.
5.3- Check the X1 connector of the D6-S.
5.4- Check the closing of the room door and the sensor.
6. ERROR 006 : BLOCKING OF IGBT
6.1- Check the resistance of the room network. Usually this blocking occurs
when the network does not succeed providing power to the equipment.
IMPORTANT: If this will be the case, the power limit of the equipment must be
changed so that the inverter module will not be damaged!
6.2- Repeat the 10” step of the service manual. If the calibratin,
R19 ofthe DB-S pat is very close o 3V and the network is mot proper "
is jing can occur. ly decrease this value at i
‘equipment behavior. eee
6.3- If this blocking occurs in discharges of low power, th:
the nines le ot ar Wie HV teh ortior aie ema rece oeaes
7. THE PANEL DOES NOT TURN ON.
7.1- ATTENTION: Equipment disconnected!
eee! Check the Diazed fuses of the
7.2- Check the F13 fuse of the D5-S plate.
18