Professional Documents
Culture Documents
I3031-5.0 en
d
3
Content
1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.1 About the QuantumX documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.2 The QuantumX family . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3 Channel overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4 Sychronization of QuantumX modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.1 QuantumX Assistant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.2 catmanAP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.3 Lab VIEW library . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.4 Programming interface (API) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5 Firmware update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.6 TEDS Editor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.7 DIAdem driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.8 DASYLab driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4 Housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.1 Mounting housing clips on modules with degree of protection IP20 . . . . . . . . . . . . . . . . . . . . . . . 28
4.2 Mounting housing clips on modules with degree of protection IP65 . . . . . . . . . . . . . . . . . . . . . . . 31
4.3 Connecting housings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.4 CASE-FIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.5 BPX001 backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.5.1 Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.5.2 Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.5.3 Mounting the modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
QuantumX HBM
4
6 Transducer connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.1 Shielding design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.2 Adjustable sensor supply for active transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
6.3 TEDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.3.1 Retrofitting the transducer plug with TEDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
6.4 MX840 universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
6.5 MX840A universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
6.6 MX440A universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
6.7 MX410 highly dynamic universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
6.8 MX460 frequency measuring amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
6.9 MX1609 / 1609Pmodule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
6.9.2 MX1609/MX1609-P status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
6.10 MX471 CAN module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
6.10.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
6.10.2 MX471 pin assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
6.10.3 Receiving CAN messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
6.10.4 LED status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
6.11 MX1601 universal amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
6.11.1 MX1601pin assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
6.11.2 MX1601status display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
6.12 Transducder technologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
6.12.1 Strain gages full bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
6.12.2 Strain gage full bridges, inductive full bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
6.12.3 Piezoresistive full bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
6.12.4 Strain gage half bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
6.12.5 Inductive half bridges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
6.12.6 Quarter bridge, strain gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
6.12.7 Connection of transducers with double shield technique . . . . . . . . . . . . . . . . . . . . . . . . 103
6.12.8 Potentiometric transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
6.12.9 LVDT transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
6.12.10 Piezoelectric sensors IEPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
6.12.11 DC voltage sources 100 mV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
6.12.12 DC voltage sources 10 V and 60 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
6.12.13 DC power sources 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6.12.14 Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
6.12.15 Resistance thermometers Pt100, Pt1000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
6.12.16 Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
6.12.17 Frequency measurement without directional signal (RS 485: Differential signals) . . . 116
6.12.18 Frequency measurement with directional signal (RS 485: Differential signals) . . . . . . 117
6.12.19 Frequency measurement without directional signal (single−pole mode) . . . . . . . . . . . 118
6.12.20 Frequency measurement with directional signal (single−pole mode) . . . . . . . . . . . . . . 119
6.12.21 Pulse counting, (RS 485: Differential signals) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
HBM QuantumX
5
8 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136
8.1 System accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
8.1.1 BPX001 backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
8.1.2 Housing connection elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
8.2 Voltage supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
8.2.1 Power pack NTX001 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140
8.2.2 Supply cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8.3 FireWire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8.3.1 FireWire cable (module to module; IP20) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
8.3.2 FireWire cable (module to module; IP65) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
8.3.3 Connection cable (PC to module; IP20) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
8.4 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
8.4.1 Plug kit with TEDS chip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
8.4.2 Port saver SubHD 15pol. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
8.4.3 Adapter D−Sub−HD 15-pin to D−Sub 15-pin. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144
8.5 MX840 accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
8.5.1 Cold junction for thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
8.5.2 Adapter cable (CAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
8.6 Accessories MX410 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
8.7 MX1609 and 1609P accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
8.7.1 Thermo-connector with integrated RFID chip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
8.7.2 RFID chip for sticking onto thermo-connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
9 Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
QuantumX HBM
6 Safety instructions
1 Safety instructions
Appropriate use
The module and the connected transducers may be used for measurement and directly
related control tasks only. Any other use is not appropriate.
To ensure safe operation, the module may only be used as specified in the operating
manual. It is also essential to follow the respective legal and safety regulations for the
application concerned during use. The same applies to the use of accessories.
Each time, before starting up the modules, you must first run a project planning and risk
analysis that takes into account all the safety aspects of automation technology. This
particularly concerns personal and machine protection.
Additional safety precautions must be taken in plants where malfunctions could cause major
damage, loss of data or even personal injury. In the event of a fault, these precautions
establish safe operating conditions.
This can be done, for example, by mechanical interlocking, error signaling, limit value
switches, etc.
Safety rules
A module must not be connected directly to a power supply network. The maximum
permissible supply voltage is 10...30 V (DC).
The supply connection, as well as the signal and sense leads, must be installed in such a
way that electromagnetic interference does not adversely affect device functionality (HBM
recommendation: ”Greenline shielding design”, downloadable from the Internet at
http://www.hbm.com/Greenline).
Automation equipment and devices must be covered over in such a way that adequate
protection or locking against unintentional actuation is provided (such as access checks,
password protection, etc.).
When devices are working in a network, these networks must be designed in such a way that
malfunctions in individual nodes can be detected and shut down.
Safety precautions must be taken both in terms of hardware and software, so that a line
break or other interruptions to signal transmission, such as via the bus interfaces, do not
cause undefined states or loss of data in the automation device.
HBM QuantumX
Safety instructions 7
Conditions on site
For modules in IP20 protected housing:
− Protect the modules from humidity or effects of the weather such as rain, snow, etc.
− The permissible relative humidity at 31°C is 80% (non condensing); linear reduction down
to 50% at 40°C.
− Ensure that the ventilation openings on the sides are not covered.
For all modules:
− Do not expose the instrument to direct sunlight.
− Observe the maximum permissible ambient temperatures given in the specifications.
Remaining dangers
The scope of supply and performance of the module covers only a small area of
measurement technology. In addition, equipment planners, installers and operators should
plan, implement and respond to the safety engineering considerations of measurement
technology in such a way as to minimize remaining dangers. Prevailing regulations must be
complied with at all times. There must be reference to the remaining dangers connected with
measurement technology. After making settings and carrying out activities that are
password−protected, you must make sure that any controls that may be connected remain in
safe condition until the switching performance of the module has been tested.
Working safely
Error messages should only be acknowledged once the cause of the error is removed and
no further danger exists.
QuantumX HBM
8 Safety instructions
Outputs
It is essential to take safety precautions when using a module’s digital, analog or CANbus
outputs. Make sure that status or control signals do not perform any actions that result in
damage to human health or the environment.
Qualified personnel
Qualified persons means persons entrusted with the installation, fitting, commissioning and
operation of the product who possess the appropriate qualifications for their function. This
module is only to be installed and used by qualified personnel, strictly in accordance with the
specifications and the safety rules and regulations.
This includes people who meet at least one of the three following requirements:
• Knowledge of the safety concepts of automation technology is a requirement and as
project personnel, you must be familiar with these concepts.
• As automation plant operating personnel, you have been instructed how to handle the
machinery and are familiar with the operation of the modules and technologies described
in this documentation.
• As commissioning engineers or service engineers, you have successfully completed the
training to qualify you to repair the automation systems. You are also authorized to
activate, to ground and label circuits and equipment in accordance with safety
engineering standards.
It is also essential to comply with the legal and safety requirements for the application
concerned during use. The same applies to the use of accessories.
NOTE
The safety instructions in this document also apply to the NTX001 power supply and
the BPX001 active back plane.
HBM QuantumX
Safety instructions 9
In this manual, the following symbols are used to point out residual dangers:
Symbol: DANGER
Meaning: Maximum danger level
Warns of an imminently dangerous situation in which failure to comply with safety
requirements will result in death or serious bodily injury.
Symbol: WARNING
Meaning: Dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety
requirements can result in death or serious bodily injury.
Symbol: CAUTION
Meaning: Potentially dangerous situation
Warns of a potentially dangerous situation in which failure to comply with safety
requirements could result in damage to property or some form of bodily injury.
Symbol:
Meaning: Electrostatic Sensitive Devices
Devices marked with this symbol can be destroyed by electrostatic discharge. Please
observe the precautions for handling electrostatic sensitive devices.
QuantumX HBM
10 Safety instructions
Symbols pointing out notes on use and waste disposal as well as useful information:
Symbol: NOTE
Points out that important information about the product or its handling is being given.
Symbol:
Meaning: CE mark
The CE mark enables the manufacturer to guarantee that the product complies with the
requirements of the relevant EC directives (the declaration of conformity is available at
http://www.hbm.com/hbmdoc).
Symbol:
Meaning: Statutory marking requirements for waste disposal
National and local regulations regarding the protection of the environment and recycling of
raw materials require old equipment to be separated from regular domestic waste for
disposal.
For more detailed information on disposal, please contact the local authorities or the dealer
from whom you purchased the product.
HBM QuantumX
Safety instructions 11
Conditions on site
For modules in housings with IP20 degree of protection:
− Protect the modules from moisture and dampness or weather such as rain, snow, etc.
− The permissible relative humidity at 31oC is 80% (non condensing); linear reduction to
50% at 40 oC.
− Make sure that you do not cover the ventilation openings at the side.
For all modules:
− Do not expose the modules to direct sunlight
− Comply with the maximum permissible ambient temperatures as stated in the
specifications.
− Ensure sufficient ventilation during installation into the BPX001 backplane.
QuantumX HBM
12 Introduction
2 Introduction
• A comprehensive online help with index and easy search options which is available after
the installation of a software packet (e.g. QuantumX Assistant, catmanEASY)
HBM QuantumX
Introduction 13
The QuantumX family is a modular and universally applicable measurement system. The
modules can be individually combined and intelligently connected according to the
measurement task. The distributed operation makes it possible to position individual modules
close to the measuring points, resulting in short sensor lines.
QuantumX HBM
14 Introduction
S 2 FireWire interfaces
− For optional voltage supply (see data sheet)
− For optional data communication via PC
− For medule synchronization
− For internal transmission of measurement data
S Connector (VG strip) for installation on an active module carrier (not applicable for IP65
variants)
S Status LEDs for displaying general system and channel states
S A working standard calibration certificate has been stored on every amplifier. It can be
read out via QuantumX Assistant.
S AutoBoot (module configuration is retained)
HBM QuantumX
Introduction 15
QuantumX Modules
MX1609
MX840A MX440A MX410 MX460 MX1601 MX878 MX471 CX27 CX22W
MX1609-P
Number of channels (total) 8 4 4 4 16 16 - - - -
Data rate (Samples/s) 19200 19200 96000 96000 300 19200 - - - -
Bandwidth (Hz) 3200 3200 38000 38000 14 3000 - - - -
Analog outputs • •
Digital IN (static) • •
Digital OUT (static) • •
•
Mathematics • • • •
Local recording of measured data •
¹) A Smart module (1-EICP-B-2) is required for connecting current-fed piezoelectric transducers.
See data sheets for precise technical specifications. The pin assignments can be found in the following chapters.
QuantumX HBM
16 Introduction
To enable measurement signals to be synchronized for processing and analysis, they need
to be synchronously measured.
All QuantumX modules can be synchronized among themselves. This ensures simultaneous
measurement on all channels. All the analog-digital converter rates, measuring rates and
bridge excitation voltages are also synchronized.
Synchronization methods:
Synchronization via FireWire
All the modules are synchronized automatically when they are connected via the FireWire
cable. This is the recommended method.
No CX27 module present in the system:
The module with the highest serial number takes over the Master function.
HBM QuantumX
Introduction 17
QuantumX HBM
18 Introduction
EtherCAT
Ethernet
NTP
Synchronizing via NTP (without FireWire)
NTP
NTP
NTP
NTP
Time base :
NTP server
Ethernet
switch
Successful synchronization:
For accurate synchronization, the corresponding channels should be parameterized using
identical filter settings. The delay is not automatically corrected. Filter delays are given in the
datasheet. After booting and successful synchronization, the system LED shows green. If
synchronization is disturbed, or not yet established, the system LED shows orange.
Example: MX840
System LED
There are several synchronization methods to choose from (also see Fig. 2.1 page 18):
S synchronizing via FireWire
S synchronizing via EtherCAT (CX27)
S synchronizing via NTP (Network Time Protocol) with FireWire
S synchronizing via NTP without FireWire
QuantumX HBM
20 Software
3 Software
The supplied QuantumX System CD contains a powerful software package consisting of the
QuantumX Assistant, Lab View library, programming library for .NET/COM, TEDS-Editor,
FireWire driver as well as a program for module firmware updating. catmanEASY and
DIAdem driver are software products that can be obtained as separate product packages.
Modules:
• Search and configuration (e.g. TCP/IP communication), naming
• Reset to factory settings
• Read factory calibration certificate
• Analysis (information, status, log file)
• Save configuration to operating PC
Channels/sensors:
• Configuration (name, connection type, TEDS, semi-automatic assignment)
• Measurement (alphanumeric display)
• Open the TEDS editor and read/write to TEDS
• Activate/deactivate isochronous operation via FireWire
Individual signals:
• Set sampling rate and filter (type, cut-off frequency)
Sensor database
• Output signals (scaled, filteresd)
• Modify and expand existing sensor databases (e.g. own sensors, dbc data files)
HBM QuantumX
Software 21
3.2 catmanAP
S EasyScript is based on current VBA standard (Visual Basic for Applications) and enables
writing of own scripts for individual measurement tasks
QuantumX HBM
22 Software
Lab VIEW is a graphic programming system from the company ”National Instruments”. The
acronym stands for ”Laboratory Virtual Instrumentation Engineering Workbench”.
The main application areas for Lab VIEW are in measurement, control and automation
technology.
A Lab VIEW device driver is a VI (Virtual Instrument) or sub-program that is used in Lab
VIEW programs for easy operation of devices. Device drivers are used to initialize, open and
close interfaces when required, to initialize and configure devices, to make settings, trigger
and query measurements, etc.
A setup program is available on the QuantumX System CD with which you can install the
Lab VIEW driver on your computer.
The abbreviation API stands for ”Application Programming Interface” and designates
so-called programming interfaces. Programmers can directly access functions of other
programs via APIs and use these in their own programs.
With the API, you have full access to all QuantumX functions through an individually
programmed application, e.g. your own operator interface.
The API can be used in the form of programming libraries in .NET or COM technologies. The
libraries enable the creation of own applications in programming languages such as e.g.
Visual Basic, C++, C# or Delphi. Functions such as communication connection, configuration
of measurement channels, implementation of measurements and troubleshooting are
components of the library.
The API can be easily installed from the QuantumX System CD. Application-based examples
and practice-oriented documentation enable rapid a quick start.
You can easily check the firmware status of the modules and update them with the
”QuantumX firmware update” software.
Before updating your firmware, please check whether your PC software needs to be
updated.
We recommend that you check and, if necessary, update the firmware
• if you want to use a new PC software package
• if you expand your system with new modules
QuantumX Assistant also enables you to check the firmware status of your modules:
• Right−click on a module −> Details −> System parameters
HBM QuantumX
Software 23
• Compare your version with the actual firmware version on the Internet under:
www.hbm.com\quantumX
QuantumX HBM
24 Software
If the firmware version number of your module is lower than the actual number on the
Internet, an update is necessary:
• Download the actual firmware from the HBM website and save it in the download
directory of the firmware updater (in most cases: C:\Programs\HBM\QuantumX Firmware
Update\Download)
• Download the actual software package from the HBM website
• Close all open HBM software, install the new software and start the QuantumX Firmware
Update
• Click on the ”Find modules” symbol or press the function key F4.
• Press the ”Start” button and wait until the update is complete (do not interrupt the
process/do not switch off the modules/do not interrupt the connection)
HBM QuantumX
Software 25
HBM’s TEDS Editor enables TEDS data to be read, edited and written directly via a
measurement channel or the HBM TEDS dongle. When retrofitting TEDS, the editor provides
corresponding templates for the different transducer types. Some templates can be saved
and loaded.
HBM’s DIAdem driver enables measurement data acquisition using the following QuantumX
amplifiers: MX840, MX840A, MX440A, MX410, MX460 and MX1609.
Latest information can be found in the corresponding release notes at www.hbm.com.
The partner company IMP sells a DASYLab driver for measurement data acquisition using
the follwing QuantumX modules: MX840, MX840A, MX440A and MX1609.
Latest information can be found at www.impkoeln.de.
QuantumX HBM
26 Housings
4 Housings
The degree of protection given in the technical data indicates the suitability of the housings
for various ambient conditions and also the protection of persons against potential risks
when used. The letters IP (International Protection), which are always present in the
designation, are followed by two digits. These indicate which degree of protection a housing
offers against contact or foreign bodies (first digit) and moisture (second digit).
The QuantumX modules are available in housings with degree of protection IP20 and, in
some cases, with IP65 (as per DIN EN 60529).
IP 2 0
IP 6 5
Code index Degree of protection against Code index Degree of protection against water
contact and foreign bodies
2 Protection against contact with 0 No water protection
fingers, protection against foreign
bodies with ∅ > 12 mm
6 Complete protection against contact, 5 Protection against water jet(nozzle)
protection against penetration of dust from any angle
Both housing types can be connected together with the aid of two lateral housing clips
(1-CASECLIP, not included in scope of delivery). To do this, the existing lateral covers must
be removed and the housing clips screwed on.
Housing clip
HBM QuantumX
Housings 27
QuantumX HBM
28 Housings
The module electronics are integrated in a metal housing that is surrounded by a case
protection (CASEPROT). This also serves for centering when several devices are stacked on
top of each other and offers a certain degree of protection against mechanical damage.
Case protection
MX840 housing
Cover
The mounting of the housing clips shown in the following pictures must be implemented on
both sides of the housing.
a.f. 2.5
HBM QuantumX
Housings 29
Cover
Housing clip
a.f. 2.5
QuantumX HBM
30 Housings
a.f. 2.5
HBM QuantumX
Housings 31
The mounting shown in the following pictures must be implemented on both sides of the
housing.
Cover
Housing clip
a.f. 2.5
QuantumX HBM
32 Housings
HBM QuantumX
Housings 33
The following pictures show the connection of two IP20 housings. The procedure is identical
for IP65 housings.
Press
Lever
Catch
QuantumX HBM
34 Housings
HBM QuantumX
Housings 35
4.4 CASE-FIT
A fitting panel (CASE−FIT) allows flexible mounting of IP20 protected modules of the
QuantumX series. A lashing strap or CASE−CLIP can be used for fastening the modules.
132
169,5
Ø 5,6
Laschen für die
zusätzliche fixierung
mit Spanngurten
approx. 30
11,4
22
QuantumX HBM
36 BPX001 Backplane
You can connect up to 9 modules with each other using the BPX001 backplane without the
need for complicated cabling and you can link these via the two FireWire interfaces of the
backplane with other modules or module groups. You can also set up a direct connection to
a PC via one of the FireWire interfaces of the backplane. The FireWire interfaces of the
individual modules are actively connected to each other.
The supply voltage (18 V ... 30 V DC) for the modules must be supplied externally. The
circuits of the FireWire connections and the modules are protected with a total of four fuses
with control displays (arrangement see Tab. 4.1).
The modules can be positioned anywhere in the backplane. The backplane is designed for
wall or control cabinet installation and has boreholes for attachment.
Slot 9
Slot 1
HBM QuantumX
BPX001 Backplane 37
4.5.1 Connection
VG strip module
connection
FireWire X1 / X2
Supply voltage
18 V ... 30 V DC
5 A max. + −
Ground
No. Protection
1 FireWire X1 connection
2 FireWire X2 connection
3 Slots 1 to 4
4 Slots 5 to 9
QuantumX HBM
38 BPX001 Backplane
A total of 10 boreholes are provided in the backplane for wall mounting (∅ 6.5 mm). We
recommend using the outer 4 boreholes for wall mounting.
NOTE
Only use countersunk head bolts for fastening, otherwise the modules cannot be
mounted correctly.
Note the following information when installing one or more backplanes in a control cabinet:
• When installing in a control cabinet, the temperature limits given in the technical data of
the backplanes must be complied with
• Depending on the installation situation, sufficient ventilation (vertical air flow) or cooling
must be provided (the maximum total output on a backplane is approx. 150 Watt)
• The ventilation slits of the modules must not be covered (e.g. by cable ducts)
HBM QuantumX
BPX001 Backplane 39
TOOLS
We recommend a T-handle socket spanner 4x150 (across flats 4 mm, length 150 mm).
NOTE
The modules can only be fastened in backplanes in housings with IP20 protection
without case protection, case clips or lateral covers If these are present, remove as
shown in Chapter 4.
Mounting sequence:
1. Remove the cover of the connecting plug (rear of module).
Cover
2. Unscrew the upper and lower screwed clamping glands of the backplane up to the stop
(the screws are secured against falling out!).
QuantumX HBM
40 BPX001 Backplane
3. Position the module vertically on the backplane and push it in carefully on the lower guide
rail back up to the stop.
a.f. 4.0
Guide rails
2.
1.
QuantumX HBM
42 Connection
Connect the modules to a DC voltage of 10 V ... 30 V (recommended 24 V), see the table for
the power consumption per device.
CAUTION
A rule of thumb for voltage distribution via FireWire says:
”An external supply voltage of equal potential is required at every third module”.
Defects in the device cannot be excluded if a supply voltage > 30 V is used. If the supply
voltage drops below 10 V, the modules switch off.
In the event of battery operation in a vehicle, we recommend incorporating an uninterrupted
voltage supply between battery and module to compensate for voltage drops during starting
processes.
Module Typical power consumption, including transducer excitation (Watt)
MX840 13
MX840A 12
MX440A 10
MX1601 13
MX410 15
MX460 9
MX1609 6
MX1609-P 6
CX22 12
CX27 7
MX878 7
MX471 6
If several modules are connected to each other via FireWire for time−synchronous data
acquisition (see Fig.5.4), the supply voltage can be looped through. The power pack used
must be able to provide the appropriate output.
The maximum permissible current on the FireWire connection cable is 1.5 A.If the chain is
longer repeating the supply connection is mandatory.
If several amplifiers are operated non−synchronously (see Fig.5.3), they must be supplied
separately.
NTX001
Oder X104
1−Kab271−3
X101/X102
1−Kab269
FireWire
HBM QuantumX
Connection 43
10 V ... 30 V DC
1-NTX001 or X104
1-KAB271-3
NOTE
You must use an Ethernet crossover cable with older computers. Newer PCs/laptops
have Ethernet interfaces with autocrossing function. You can also use Ethernet patch
cables here.
QuantumX HBM
44 Connection
Patch cable
Patch cable
Modules can be connected to the PC with standard Ethernet switches. We recommend patch
cables.
With the star structure displayed here, measurement data from other modules is not lost in
case of a cable break in the ethernet cable.
HBM QuantumX
Connection 45
10 V ... 30 V DC
Patch cable
Patch cable
FireWire connection
1-Kab269-x: Connection cable with various lengths (x m)
The supply voltage for the modules is looped through FireWire in the configuration shown
above (max. 1.1 A through FireWire; power consumption for 1 module, see specifications in
data sheet).
Advantage of this connection structure: The other modules remain active in case of a cable
break in the ethernet cable.
QuantumX HBM
46 Connection
NOTE
Ensure that your PC has a valid IP address.
Install and start up the latest QuantumX Assistant version on your PC.
(All screenshots shown in this quick start guide display the menus in the operating system
WindowsXP).
• Click on the icon (Find modules) or press the function key F4.
HBM QuantumX
Connection 47
The next dialog window offers you the several options for network searching. For the first
startup we recommend:
• Search the whole network and select from the result (recommended setting)
Recommended
button.
NOTE
The network connection can be influenced by:
w An activated WiFi connection on your PC: Switch this connection off and start the
network search again
w The use of a standard patch cable in a direct connection (peer-to-peer)
If the module you have selected appears in black in the list, immediate operation is possible.
QuantumX HBM
48 Connection
If the module appears gray in the list, mark it and click on the ”Edit network settings” button.
HBM QuantumX
Connection 49
Automatic Configuration
Module settings
PC settings
Manual Configuration
PC settings
QuantumX HBM
50 Connection
• Click on ”OK”
• Confirm the settings with the ”Yes” button, the module is then restarted with the current
settings.
The system LED of the module should now light up green, if not please check your network
settings again! If the network settings are ok, the module names are shown in black.
• Mark the relevant modules by checking the relevant checkbox
• Confirm with ”OK” and you are now ready for your first measurement job
HBM QuantumX
Connection 51
• In the tab ”Alternate Configuration”, select the option ”User-defined” and enter your data
in the ”IP address line” and ”subnet mask”.
QuantumX HBM
52 Connection
• Confirm the settings with the ”Yes” button, the module is then restarted with the current
settings.
HBM QuantumX
Connection 53
The system LED of the module should now light up green, if not please check your network
settings again! If the network settings are ok, the module names are shown in black.
QuantumX HBM
54 Connection
General information
S Baud rate of 400 MBaud (approx. 50 MByte/s)
S Asynchronous (all nodes) or isochronous (in real time to a specific node) data
transmission
S Data synchronization
S Supply voltage via FireWire connection cable (max. 1.5 A)
10 V ... 30 V DC
1-NTX001 or 1-KAB271−3
X104
X101
1-KAB270-3
Adapter
PC: PCI or PCIe card
Notebook: PC card (previously PCMCIA)
NOTE
Please check in advance whether a firmware or software update is required.
Software/firmware downloads can be found on the HBM website:
www.hbm.com\downloads
HBM QuantumX
Connection 55
10 V ... 30 V DC
(e.g. NTX001)
10 V ... 30 V DC
(e.g. NTX001)
X102
X101
FireWire connection
1-Kab269-2 1-Kab270-3
1−Kab269-0.2
2 m connection cable 3 m connection cable
0.2 m connection cable
Data are transferred, modules are synchronized in timing and voltage is supplied via the
FireWire connections. You can connect a maximum of 12 modules in series with each other.
NOTE
Different voltage sources need to supply the same voltage level.
QuantumX HBM
56 Connection
NOTE
For trouble shooting purposes you can switch to the original FireWire driver with the
tool “t1394bus_installwizard.exe”. You can find the tool after having installed the
driver on your hard disc.
• After installation and configuration, connect the FireWire cable first to the PC adapter and
then to the first module. Activation is confirmed acoustically by Windows.
Whenever you connect a new module via FireWire to your computer, the operating system
will prompt you to register the module. Always refer here to the driver “hbm1394.sys”.
• Install and then start the latest QuantumX Assistant on your PC.
• Click on the icon (Find modules) or press the function key F4.
The field ”Modules found” displays all modules connected via FireWire.
• If your modules are not yet displayed, mark ”Search the complete network” and click
HBM QuantumX
Connection 57
NOTE
If no modules are found via FireWire this may be caused by the following:
− The modules have not been properly registered. Click on the FireWire driver in
Systray, check the driver behind the modules and reinstall it, if necessary
(hbm1394.sys).
− Check all connections between the modules.
QuantumX HBM
58 Connection
• Compare your version with the latest firmware version on the Internet under:
www.hbm.com\quantumX
HBM QuantumX
Connection 59
If the firmware version number of your module is lower than the current number on the
Internet, an update is necessary:
• Download the latest software package from the HBM website (QuantumX firmware
downloader, QuantumX Assistant, etc.)
• Close all running HBM software, install the new software and start QuantumX Firmware
Updater
• Download the latest firmware from the HBM website and save it in the download directory
of the firmware updater (in most cases: C:\Programs\HBM\QuantumX Firmware
Update\Download)
• Click on the ”Find modules” symbol or press function key F4 (see chapter 2.1).
• Click on the ”Start” button and wait until the update is complete (do not interrupt the
process/do not switch off the modules/do not interrupt the connection)
NOTE
You can update the firmware of the modules via FireWire, directly via Ethernet or via
the Ethernet connection of the Gateway CX27.
QuantumX HBM
60 Connection
1 2 3
P4
Hub
6 7 8 9 10 11
P3
5
12
1 2 3 4
Displayed:
Total number of modules: 15
Total number of Hops: 12
Data sink
QuantumX module
P3 = Port 3
Fig.5.8: Example of a star topology with two chains and one hub
The number of modules connected in series (daisy chain) is limited to 12. If you want to
connect more modules (maximum 24), you must use hubs. Hubs are devices that connect
network chains together in star configurations. This connection mode is again limited to 14
hops.
A hop is the transition from one module to another (this means n−1 hops for n QuantumX
modules in a chain).
Depending on the connection situation, 1 to 2 hops are counted in one hub (see Fig.5.9).
To count the total number of hops, the longest chain to the data sink must be counted (worst
case).
In FireWire systems that consist of several hubs with several module series, only the hops of
the longest series are counted, the others are not taken into account.
HBM QuantumX
Connection 61
Phy 1 Phy 2
Port
4
Port 4
Note
Always connect the chain with the most modules to Port 3 or Port 4.
QuantumX HBM
62 Connection
Greater distances (>5m) in FireWire networks can be bridged with opto-hubs that enable
distances of up to 500 m with the use of a glass fiber cable.
1 Hop
1 2 Max. 500 m
Opto- Opto-
P3
Hub Hub
P4 P4 7 8
P3 6
1 2 3 4 5
Displayed:
Total number of modules: 11
Total number of hops: 8
QuantumX module
HBM QuantumX
Connection 63
10 V ... 30 V DC
Opto-Hub
Opto-Hub
QuantumX HBM
64 Connection
6 Transducer connection
Sources of interference may cause electromagnetic fields which couple interference voltages
into the respective measuring circuits inductively or capacitively through connecting cables
and equipment housings and thus affect the way the equipment works. It is essential to
ensure that the equipment used in the system itself does not emit electromagnetic
disturbances. Electromagnetic compatibility (EMC) which combines the concepts of required
electromagnetic immunity or susceptibility (EMS) and permitted electromagnetic interference
(EMI) has become increasingly more important over the past years.
The HBM Greenline shielding design
The complete measuring chain is fully enclosed in a Faraday cage through the unique design
of the cable shield. The cable shield is flattened, connected to the transducer housing, and
routed to the amplifier housing through conductive connectors. These measures substantially
reduce the effects of electromagnetic disturbances.
Conductive housing
ensuring connection
to plug and equipment
housing Cable shield connected
Signal-carrying to conductive housing
contacts through strain relief
HBM QuantumX
Connection 65
Note
All components of the measuring chain (including all cable joints such as connectors
and couplings) have to be surrounded by a closed, EMC-tested shield. Shield
transitions must be realized in the form of flattened, closed and low-impedance joints.
This is ensured when original HBM connectors are used.
QuantumX HBM
66 Connection
Some modules can supply active transducers with an excitation voltage of 5…24 V.
When using the adjustable transducer excitation, electrical isolation from the supply voltage
of the amplifier is not required.
The maximum permissible power consumption is 700 mW per channel, in total not more than
2 W. If the power consumption is more than 700 mW on one channel, the transducer
excitation of this channel will switch off. If the power consumption exceeds a total of 2 W, the
device may switch off.
MX840
1 ... 8
4
9
CAUTION
The output voltage during idling is up to 20% higher than set when the voltage <8 V. If
a current of 2 mA is drawn, the voltage falls back to the set value.
The condition is met in sensors with a power consumption of min. 20 mW.
This can also be met with an additional resistance of 3 kΩ if the power consumption is
lower or if the sensors are sensitive.
Check the correct voltage setting when connecting a sensor. Too high a voltage can
destroy the sensor. The voltage value is a part of the MX840 parameterization and can
only be changed with a new parameterization. The sensor supply is switched off in the
delivery condition.
HBM QuantumX
Connection 67
6.3 TEDS
The acronym TEDS stands for ”Transducer Electronic Data Sheet” and denotes a transducer
or sensor’s electronic data sheet which is saved to a small electronic chip or a corresponding
module and is intrinsically linked to the transducer.
In addition, valuable meta data is supplied, for example, calibration data providing essential
information for the traceability of measurements or tests. The electronic data sheet can be
located in the transducer housing, in the inseparable cable or connector plug.
The function and principle of operation are defined in Standard IEEE1451.4.
Data mode
QuantumX MX840A
Measurement mode
QuantumX HBM
68 Connection
S One TEDS module can be accessed via two separate cable wires (”one-wire circuit”) or
the transducer plug can be retrofitted with TEDS.
S Amplifiers enabling IEPE transducers to be directly connected support TEDS version 1.0.
S Some HBM transducers feature a special integrated TEDS module which enables TEDS
data to be transmitted via the transducer’s feedback line (patented ”zero-wire circuit”).
After digital communication (data mode), the amplifier switches to the measurement
mode. The U93 force transducer is an example for such a transducer.
S Thermocoulpe amplifiers with an RFID chip integrated in the transducer plug use TEDS
technology, for example, to electronically connect the measuring point with the
transducer.
The data sheet to the respective amplifier provides more specifications relating to TEDS, for
example the maximum permitted cable length to the transducer. If no TEDS is used, the
permitted cable length can be substantially longer.
Note
Further information about TEDS topics can be found at http://www.hbm.com/teds
HBM QuantumX
Connection 69
The IEEE standard 1451.4 defines a generally accepted process which enables sensors to
be identified. The sensor is identified via the respective data sheet which is stored in
electronic format in the sensor, cable or plug on a 1−wire EEPROM (TEDS − Transducer
Electronic Data Sheet). The amplifier uses a serial 1−wire interface for communication with
this EEPROM, reads out the data sheet and configures the amplifier accordingly.
Retrofilling TEDS in the plug is shown in the figure below.
1 11
5 15
10
4
Bridge
1 1
2 Data
3 No function 1 2 3
1-wire EEPROM
(Maxim DS2433+) View from below
QuantumX HBM
70 Connection
You can connect up to eight transducers to the universal amplifier MX840. The transducers
are connected via 15-pin D-SUB-15HD device connectors. All measuring channels are
electrically isolated von one another and from the mains. When using the adjustable
transducer excitation, electrical isolation from the supply voltage of the amplifier is not
required.
MX840 connectable transducers
Transducer type Connector sockets see page
LVDT 1 ... 8 85
Current 1 ... 8 89
Potentiometers 1 ... 8 84
Thermocouples 1 ... 8 92
HBM QuantumX
Connection 71
CANbus 1 103
QuantumX HBM
72 Connection
6
Bridge
1 11
9
4
5 15
Fig.6.2: Pin arrangement of connection plug, view from the solder side
Pin Connection
1 TEDS (+)
2 Bridge Excitation voltage (−), 0° reference pulse (zeroing pulse) (−)
3 Bridge Excitation voltage (+), 0° reference pulse (zeroing pulse) (+)
4 Always connect with Pin 9! (Plug identification)
5 Measurement signal (+), potentiometer measurement signal(+),
voltage input 100 mV (+), f1(−) signal differential, SSI data (−)
6 TEDS (−), ground frequency measurement
7 Sense lead (−), f2(−) signal differential, CAN-High, SSI clock (−)
8 Sense lead (+), f2(+) signal differential, CAN-Low, SSI clock (+)
9 Signal ground
10 Measurement signal (−), f1(+) signal differential, SSI data (+)
11 Active sensor supply 5 ... 24 V (0 V)
12 Active sensor supply 5 ... 24 V (+)
13 Current input "30 mA (+)
14 Voltage input 10 V (+), 60 V (+)
15 Vacant
NOTE
Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter
cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device
connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable (see
chapter 8.4.3).
HBM QuantumX
Connection 73
The front panel of the universal amplifier has a system LED and 8 connection LEDs. The
system LED indicates the status of the device, the connection LEDS the states of the
individual connections.
Connection LED
System LED
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
CAN LEDs
Green CAN bus activated, CAN data can be received
Orange CAN bus in ”WARNING” state, CAN data received but bus is occasionally
disturbed; buffer overflow, individual data lost
Red CAN bus in ”ERROR” or ”BUS-OFF” state, CAN data cannot be received or
processed
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX HBM
74 Connection
The MX840A universal amplifier corresponds to the MX840 amplifier, with the additional
option to connect strain gage half bridges and variable ohmic resistors.
MX840A connectable transducers
Transducer type Connector sockets see page
LVDT 1 ... 8 85
Current 1 ... 8 89
Resistance 1 ... 8 90
Potentiometers 1 ... 8 84
Thermocouples 1 ... 8 92
HBM QuantumX
Connection 75
CANbus 1 103
QuantumX HBM
76 Connection
6
Bridge
1 11
9
4
5 15
Fig.6.4: Pin arrangement of connection plug, view from the solder side
Pin Connection
1 TEDS (+)
2 Bridge Excitation voltage (−), 0° reference pulse (zeroing pulse) (−)
3 Bridge Excitation voltage (+), 0° reference pulse (zeroing pulse) (+)
4 Always connect with Pin 9! (Plug identification)
5 Measurement signal (+), potentiometer measurement signal(+),
voltage input 100 mV (+), f1(−) signal differential, SSI data (−)
6 TEDS (−), ground frequency measurement
7 Sense lead (−), f2(−) signal differential, CAN-High, SSI clock (−)
8 Sense lead (+), f2(+) signal differential, CAN-Low, SSI clock (+)
9 Signal ground
10 Measurement signal (−), f1(+) signal differential, SSI data (+)
11 Active sensor supply 5 ... 24 V (0 V)
12 Active sensor supply 5 ... 24 V (+)
13 Current input "30 mA (+)
14 Voltage input 10 V (+), 60 V (+)
15 Calibration signal T10F(S) and T40, 5 V/max. 10 mA
NOTE
Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter
cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device
connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable (see
chapter 8.4.3).
HBM QuantumX
Connection 77
The front panel of the universal amplifier has a system LED and 8 connection LEDs. The
system LED indicates the status of the device, the connection LEDS the states of the
individual connections.
Connection LED
System LED
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
CAN LEDs
Green CAN bus activated, CAN data can be received
Orange CAN bus in ”WARNING” state, CAN data received but bus is occasionally
disturbed; buffer overflow, individual data lost
Red CAN bus in ”ERROR” or ”BUS-OFF” state, CAN data cannot be received or
processed
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX HBM
78 Connection
You can connect up to eight transducers to the universal amplifier MX440A. The transducers
are connected via 15-pin D-SUB-15HD device connectors. All measuring channels are
electrically isolated von one another and from the mains.
The transducer types that can be connected and the status display are identical to those of
the MX840A universal amplifier (see page 74).
System LED
Connection LED
HBM QuantumX
Connection 79
You can connect up to four transducers to the highly dynamic universal amplifier MX410. The
transducers are connected via 15-pin D-SUB-15HD device connectors. You will need BNC
adapters (accessories 1-IEPE-MX410) for the connection of IEPE transducers.
All measuring channels are electrically isolated von one another and from the mains. When
using the adjustable transducer excitation, electrical isolation from the supply voltage of the
amplifier is not required.
Current 1 ... 4 89
QuantumX HBM
80 Connection
6
Bridge
1 11
9
4
5 15
Fig.6.7: Pin arrangement of connection plug, view from the solder side
Pin Connection
1 TEDS (+)
2 Bridge Excitation voltagen (−)
3 Bridge Excitation voltage (+)
4 Always connect with Pin 9! (Plug identification)
5 Measurement signal (+)
6 TEDS (−)
7 Sense lead (−)
8 Sense lead (+)
9 Signal ground
10 Measurement signal (−)
11 Active sensor supply (−)
12 Active sensor supply (+)
13 Current input "30 mA (+)
14 Voltage input 10 V, IEPE (+)
15 Reset external charge amplifier
The analog output can be tapped via BNC. Please see chapter 7 ”Functions and outputs” for
information about the configuration.
NOTE
Many HBM transducers are fitted with 15-pin D-SUB connectors (2 rows). The adapter
cable 1-KAB416 can be used for connection to the 3 row D-SUB-15HD device
connectors of the MX840. Pins 4 and 9 are already bridged in this adapter cable.
HBM QuantumX
Connection 81
Connection LED
System LED
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red Sensor supply overload
Analog output LEDs
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Orange System is not ready, boot procedure running
Rot Over current at analog output
Orange Input signal overload
Rot Overload resulting from invalid scaling of analog output
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX HBM
82 Connection
You can connect up to four transducers to the frequency measuring amplifier MX460. The
transducers are connected via 15-pin D-SUB-15HD device connectors. All measuring
channels are electrically isolated von one another and from the mains. When using the
adjustable transducer excitation, electrical isolation from the supply voltage of the amplifier is
not required.
HBM QuantumX
Connection 83
6
Bridge
1 11
9
4
5 15
Fig.6.9: Pin arrangement of connection plug, view from the solder side
Pin Connection
1 TEDS (+)
2 Reference pulse 0° (zeroing pulse) (−)
3 Reference pulse 0° (zeroing pulse) (+)
4 Always connect with Pin 9! (Plug identification)
5 Frequency input f1 (−)
6 TEDS (−), signal ground
7 Frequency input f2 (−)
8 Frequency input f2 (+)
9 Reference voltage Vref (2.5 V)
10 Frequency input f1 (+)
11 Active sensor supply 5 ... 24 V (−)
12 Active sensor supply 5 ... 24 V (+)
13 No function
14 f1 AC+ (for passive inductive transducers)
15 Calibration signal T10F(S) and T40, 5 V/max. 10 mA
QuantumX HBM
84 Connection
Connection LED
System LED
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
HBM QuantumX
Connection 85
MX1609
−
+
Thermocouple
white
− −
+ +
green
Post−scaling
The MX1609 has a post−scaling function. A table that converts degree oC to degree oC
enables errors resulting from thermocouples and mounting conditions to be minimized.
The MX1609 processes a maximum of 64 pairs of values. The ”Calibration Table” TEDS
template enables 14 pairs of values to be saved, if no additional optional templates are being
used.
This function provides the best results when the ambient temperature of the MX1609 and
thus the temperature of the cold junction is kept constant.
QuantumX HBM
86 Connection
HBM QuantumX
Connection 87
1-THERMO-RFID
(integrated by HBM)
The chip for measuring point identification is already integrated in the HBM THERMO-MINI.
1) RFID = Radio Frequency Identification: Method for communication between transponder and read/write device
with magnetic fields or electromagnetic waves.
QuantumX HBM
88 Connection
Top view
Glue the RFID chip as shown above with an adhesive (recommended: Two-component
adhesive).
HBM QuantumX
Connection 89
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red Sensor supply overload
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
QuantumX HBM
90 Connection
The MX471 module provides four independent CANbus nodes that are all electrically
isolated from each other and the mains.
Connected devices are not directly addressed during data transmission on a CANbus. A
unique identifier denotes the contents of a message (e.g. rotational speed or engine
temperature).
The identifier also signals the priority of the message.
Note
To ensure normal operation the CANbus needs to be terminated at both ends (and
only there) using appropriate termination resistors.
A 120−ohm termination resistor can be individually connected in the module by
software. Termination is also required when short cables with low bit rates are used.
Please refer to the data sheet for the relation between bit rate and maximum bus line length.
The configuration of a node is retained after switching the modules off and on.
For decoding signals at a rate greater than 2000/s, please set up signal inputs 1 to 8 on the
MX471. The signal buffers of these signal inputs have been expanded accordingly.
HBM QuantumX
Connection 91
1
6
9
5
Fig.6.12: Pin arrangement of connection plug, view from the solder side
Pin Connection
1 no function
2 CAN−Low
3 GND
4 no function
5 CAN Shield
6 GND
7 CAN−High
8 no function
9 no function
QuantumX HBM
92 Connection
To enable CAN messages to be received, it is essential that the relevant messages can be
identified by the node. This may be done directly at the node or, in a reproducible way, by
previously generated messages in the sensor database. Individual messages can be linked
to the node by drag & drop from the sensor database.
Type *.dbc CAN databases, too, can be loaded into the sensor database. Should a CAN
database not be available, it can be generated by the user. Editors for this purpose are
provided by different companies.
Received CAN messages are instantly ”time−stamped” in the measurement mode. This
enables directly acquired measured quantities and CAN messages to be acquired and
analyzed in parallel and synchronously in the entire system.
Note
MX471 is not a classical data logger that logs the complete CAN data stream on the bit
level. The parameterized node ”listens” on the CANbus and extracts the signals from
the relevant CAN messages to transmit them as measured values.
HBM QuantumX
Connection 93
System LED
CAN LEDs “BUS”
System LED:
Green Error−free operation
Yellow System is not ready, boot procedure running
Flashing yellow Download active, system is not ready
Red Error, faulty synchronization
Ethernet LED:
Green on Ethernet link status is ok
Flashing yellow Ethernet data transmission ongoing
QuantumX HBM
94 Connection
Up to 16 freely configurable inputs for voltage (10 V, 100 mV) or current (20 mA) or
current−fed piezoelectric sensors (IEPE) can be connected to the MX1601 universal
amplifier.
The transducers are connected via 8−pin plug terminal connectors (Phoenix Contact FMC
1,5/8−ST−3,5−RF (order no. 1952089)).
All measuring channels are electrically isolated von one another and from the mains. When
using the adjustable transducer excitation, electrical isolation from the supply voltage of the
amplifier is not required.
HBM QuantumX
Connection 95
2
Bridge
3
1 5
2 Data
3 no function 1 2 3 6
1-Wire-EEPROM 7
(Maxim DS2433+)
Bottom view
8
Fig.6.13: Pin assignment of the connector plug, view from the connection side
Pin Connection
1 Voltage input 10 V (+), 100 mV (+), IEPE (+)
2 Signal ground, TEDS (−)
3 Current input 20 mA (+)
4 TEDS (+)
5 Always connect with Pin 2! (Plug identification)
6 Active sensor supply (+)
7 Active sensor supply (−)
8 Housing (shield connection)
NOTE
The transducer excitation voltage can be set in the range of 5 ... 24 V and is only
available at channels 1 ... 8. At channels 9 ... 16, the supply voltage (10 V ... 30 V) is
output less approx. 1 V. A current drain of max. 30 mA is possible; the current limiter
cuts the transducer excitation voltage at a higher current drain.
QuantumX HBM
96 Connection
The front panel of the universal amplifier has a system LED and 16 connection LEDs. The
system LED indicates the status of the device, the connection LED the status of the
individual connections.
Connection LED
System LED
Connection LED
System LED
Green Error-free operation
Orange System is not ready, boot procedure running
Flashing orange Download active, system is not ready
Red Error
Connection LEDs
All LEDs are orange Boot procedure running (system is not ready)
All LEDs are flashing orange Firmware download active (system is not ready)
Orange Connection newly assigned, transducer identification running (calibration)
Green Error-free operation (”Ignore TEDS” or ”if available” set,
channel is manually configured)
Flashing green (5s), then Reading TEDS data
green
Flashing orange (5s), then Manual configuration running (ignore TEDS)
green
Red Amplifier overload,
No sensor connected
Channel error (incorrectly parameterized, connection error, invalid TEDS data)
Red Sensor supply overload
General rule: Brief flashing → TEDS identified (green: is used, orange: is not used).
HBM QuantumX
Connection 97
1 11
5 15
10
wh
Measurement signal (+) 5
bk Excitation (−) 2
2 bu Excitation (+) 3
rd
Measurement signal (−) 10
1 4 ye
Cable shield Hsg.
3 gn
Sense lead (+) 8
gy
Sense lead (−) 7
4
1 1
2 Data
3 No function 1 2 3
1-wire EEPROM
(view from below) Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX HBM
98 Connection
1 11
5 15
10
Measurement
wh
signal (+) 5
bk Excitation (−) 2
2 bu Excitation (+) 3
rd Measurement
signal (−) 10
1 4 ye
Cable shield Hsg.
3 gn
Sense lead (+) 8
gy
Sense lead (−) 7
4
1 1
2 Data
3 No function 1 2 3
1-wire EEPROM
(view from below) Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM QuantumX
Connection 99
1 ... 8
1 11
5 15
10
Measurement
5
signal (+)
Excitation (−) 2
2 Excitation (+) 3
Measurement
signal (−) 10
1 4
Cable shield Hsg.
3
Sense lead (+) 8
Sense lead (−) 7
4
9
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(view from below) Hsg. = housing
QuantumX HBM
100 Connection
1 11
5 15
10
wh Measurement
signal (+) 5
bk
Excitation (−) 2
bu Excitation (+) 3
ye
Cable shield Hsg.
gn
Sense lead (+) 8
gy
Sense lead (−) 7
1 1
2 Data
3 No function 1 2 3
1-wire EEPROM
(view from below)
Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM QuantumX
Connection 101
1 11
5 15
10
wh
Measurement
signal (+) 5
bk
Excitation (−) 2
bu Excitation (+) 3
ye
Cable shield Hsg.
gn
Sense lead (+) 8
gy
Sense lead (−) 7
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(optional)
view from below Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX HBM
102 Connection
Available adapters:
Strain gage with 120 ohm: order number: SCM−SG120
Strain gage with 350 ohm: order number: SCM−SG350
2 1 1’
HBM QuantumX
Connection 103
1 11
5 15
10
Measurement
signal (−) 10
Measurement
5
signal (+)
Excitation (−) 2
2
Excitation (+) 3
1 4
Cable shield Hsg.
3
Sense lead (+) 8
Sense lead (−) 7
4
RB / 2 9
(on transducer)
Hsg. = housing
We recommend this connection technique for very small measurement ranges, in particularly
disturbed environments and when using long cables.
This applies for all bridge connections.
With cable lengths >50 m, a resistor with half the value of the bridge resistance (RB/2) must
be connected in each sense lead of the transducer.
QuantumX HBM
104 Connection
1 11
5 15
10
1
Cable shield 1) Case
3
Sense lead (+) 8
Sense lead (−) 7
4
9
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM (optional)
(view from below)
HBM QuantumX
Connection 105
1 11
5 15
10
LVDT transducer
Measurement
signal (+) 5
Excitation (−) 2
Excitation (+) 3
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(view from below)
Hsg. = housing
QuantumX HBM
106 Connection
Current fed piezoelectric transducers are supplied with a constant current of e.g. 5.5 mA and
feed a voltage signal to the amplifier. This transducer type is also called IEPE or ICP
transducer.
IEPE means ”Integrated Electronics Piezo−Electric”.
ICP is a registered trademark of PCB Piezotronics
Is supported by the following modules:
MX410 (direct to SubHD or via BNC adapter), MX1601 (directly to the connector)
MX840, MX840A, MX440A with 10 V analog input and 24 V supply via a Smart module
1 11
5 15
10
IEPE 4
(−) 9
(+) 14
Hsg. = housing
HINWEIS
IEPE transducers with TEDS version 1.0 are supported.
HBM QuantumX
Connection 107
Connection diagram for MX840, MX840A, MX440A with external Smart module:
1 11
5 15
10
Maximum input voltage to housing and signal
ground : "60 V
PIN:
white
2 (+) 14
6 Supply voltage zero 4
BNC green
(−) 9
IEPE 3 Cable shield Hsg.
5 red
24 V 12
4 black
0V 11
Smart module
(1−EICP−B−2) 6
1
1
2 Data
1 2 3
3 No function
Hsg. = housing
1-wire EEPROM (optional)
(view from below)
QuantumX HBM
108 Connection
Retrofilling the TEDS chip in the transducer plug when using the Smart module:
The Smart module cannot read TEDS data directly saved to the IEPE transducer. The
QuantumX plug can be retrofitted with TEDS to enable the Smart module to be read and to
automate channel configuration according to the IEPE transducer.
TEDS Editor enables settings to be made.
S TEDS−specific settings:
f ”High Level Voltage Output Sensor”
f Physical Measurand: Acceleration (m2/s or g)
f Electrical range: standard +/− 10 V
f Power requirements: required
CAUTION
Replacing the IEPE transducer on the Smart module may result in incorrect device
settings.
HBM QuantumX
Connection 109
1 11
5 15
ground : "60 V
(+) 5
U
(−) 9
10
4
Cable shield Hsg.
1
1
2 Data
1 2 3
3 No function
QuantumX HBM
110 Connection
1 11
5 15
10
(+) 14
U
Supply voltage zero 4
(−) 9
Cable shield Hsg.
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM (optional)
(view from below) Hsg.=Housing
HBM QuantumX
Connection 111
1 11
5 15
10
Maximum input voltage to housing and signal ground:
"60 V
(−) 9
4
I
(+) 13
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM Hsg. = housing
(view from below)
QuantumX HBM
112 Connection
6.12.14 Resistance
1 11
5 15
10
Four-wire circuit
Excitation (−) 2
Measurement 10
signal (−) 1)
ϑ
Cable shield Hsg.
Measurement 5
signal (+)
Excitation (+) 3
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(optional)
Hsg. = housing
view from below
HBM QuantumX
Connection 113
1 11
5 15
10
Four-wire circuit
Excitation (−) 2
Measurement 10
signal (−)
ϑ
Cable shield Hsg.
Measurement 5
signal (+)
Excitation (+) 3
4
9
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(optional)
Hsg. = housing
view from below
QuantumX HBM
114 Connection
6.12.16 Thermocouples
6
Connection with MX840, MX840A, MX440A
1 11
5 15
10
Thermocouple
1−THERMO−MX BOARD
(soldered into connector plug and contains a cold spot
compensation element and TEDS)
When recording temperatures with thermocouples using MX840, MX840A, MX440A, you
must use a connector with integrated cold spot compensation element (1−THERMO−MX
BOARD).
1) Nicrosil
HBM QuantumX
Connection 115
When recording temperatures with thermocouples, you must solder the ”1-THERMO-MX
BOARD” board into the connector plug.
6 10
PT1000
1 5
PINs to be soldered
• Insert the 1-THERMO-MX BOARD in the correct position between the plug pins
Note
Check the position with the plug shape (see picture above).
In this position, the PT1000 of the cold spot compensation element is on top.
PT1000
• PIN 1 TEDS
PIN 6 TEDS
PIN 5 Thermocouple (+)
PIN 10 Thermocouple (−)
PIN 9 Signal ground
PIN 7 PT1000 cold junction
PIN 8 PT1000 cold junction
PIN 2 Supply (−)
PIN 3 Supply (+)
QuantumX HBM
116 Connection
Schematic diagram
5V
6
f1 (+)
200 mV
f1 (−) 1 11
0V
5 15
10
gy
Ground 6
4
6
1
1
2 Data
1 2 3
3 No function
Hsg. = housing
1-wire EEPROM
(view from below)
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM QuantumX
Connection 117
Schematic diagram
5V
f1 (+) 6
200 mV
f1 (−) 1 11
0V
5 15
10
gy
f2 (+) 8
9
f1 (−) = Measurement signal speed, 0° (−)
f1 (+) = Measurement signal speed, 0° (+) 4
f2 (−) = Measurement signal speed, 90° (−)
f2 (+) = Measurement signal speed, 90° (+)
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(view from below)
Hsg. = housing
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
QuantumX HBM
118 Connection
f1
5V
3.5 V Thresholds 6
1 11
1.5 V Thresholds
0V
5 15
10
gy
Ground 6
9
4
5
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(optional)
Hsg. = housing
view from below
Adjustable sensor supply: Pin12: 5 V ... 24 V
Pin 11: 0V
Cable color code: wh= white; bk= black; bu= blue; rd= red; ye = yellow; gn= green; gy= gray
HBM QuantumX
Connection 119
f1 6
5V
3.5 V Thresholds
1 11
1.5 V Thresholds
5 15
0V
10
f1 (+) 10
f2 (+) 8
Industrial
Cable shield Hsg.
pulse
generator 6
4
5
7
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(optional)
view from below Hsg. = housing
QuantumX HBM
120 Connection
Schematic diagram
6
5V
1 11
f1 (+)
200 mV
f1 (−) 5 15
0V 10
Zeroing pulse + 3
Zeroing pulse − 2
f1 (+) 10
Industrial f1 (−) 5
pulse Cable shield Hsg.
generator f2 (−) 7
f2 (+) 8
6
9
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
Hsg. = housing
(optional)
view from below
HBM QuantumX
Connection 121
f1 6
5V
3.5 V Thresholds 1 11
1.5 V Thresholds
5 15
0V
10
f2 (+) 8
Industrial
pulse
generator f1 (+) 10
Cable shield Hsg.
Zeroing pulse 3
9
4
7
5
2
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM Hsg. = housing
(optional)
view from below
QuantumX HBM
122 Connection
1) Schematic diagram
1 11
5 15
10
Ground 6
f1 (−) 5
f1 (+) 10
SSI f2 (−) 7
f2 (+) 8
9
f1 (−) = Data (−)
f1 (+) = Data (+) 4
f2 (−) = Shift clock (−)
f2 (+) = Shift clock (+)
Hsg. = housing
HBM QuantumX
Connection 123
1) Schematic diagram
U
1 11
5 15
10
Maximum input voltage to housing and signal
ground : "60 V
f1 AC+ 14
Cable shield Hsg.
Ground 6
9
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM
(view from below)
Hsg. = housing
QuantumX HBM
124 Connection
1 11
5 15
10
Transducer
f1 (+) 10
Cable shield Hsg.
4
7
5
2
6
1
1
2 Data
1 2 3
3 No function
1-wire EEPROM Hsg. = Housing
(view from below)
HBM QuantumX
Connection 125
6.12.26 CANbus
Chanel 1
SubHD-15pol. SubD-9pol.
6
1
6
1 11
9
5 15 5
10
CAN-High 7 7
CAN-Low 8 2
CAN-GND 6 6
Note
Ensure correct termination with termination resistors is made, as shown in Fig.6.1.
The MX840 does not have any termination, the MX471 features internal termination
that can be activated by software.
Termination Termination
resistance resistance
120Ω 120Ω
Node 1 Node n
... ...
The adapter cable 1-KAB418 is used to connect the D-SUB-15HD device connectors of the
MX840 to standard CAN plugs (D-SUB-9).
QuantumX HBM
126 Functions and outputs
The MX878, MX410 and MX460 modules enable mathematics functions to be performed
in real time and their results to be provided as standard system signals. These system
signals can be used like actual measurement signals for subsequent tasks (analog output,
EtherCAT signal, source signal for mathematics function, data visualization and storage).
The MX878 and MX410 modules feature analog outputs that can be connected to a
system or source signal, for example an actual measurement signal (additionally scaled,
filtered) or the result of a mathematics function.
Measurement channels that are used in mathematics functions or directly for analog outputs
need to be activated for ”isochronous data transfer” (e.g. in the QuantumX−Assistant
software, ”Signals” tab).
Note
The module configuration is immediately active after system reboot (Auto Startup). No
operating PC is required for configured signal outputs to run stand−alone.
max
S Peak values (Peak)
+ −
x =
S Add & Multiply
min min
max max
HBM QuantumX
Functions and outputs 127
7.1 MX410
The MX410 module is a module with eight analog outputs that can be accessed on the front
panel via BNC sockets.
The outputs are directly assigned to the above located inputs.
Note
After configuring an analog output its function (configuration, scaling) continues to
be available, even if the computer is disconnected. No connection to a PC is required.
Keep in mind the following notes when working with peak monitor channels:
S The maximum output (sample rate) is restricted to 4800 Hz
S Resetting of PEAK values only via PC software (system input signals will follow)
S The output rate of the peak monitor channel must not be higher than the sample rate of
the input channel
S Filters set for the MX410 do not apply to peak monitor channels
S These channels are always unfiltered. However, the input signal is filtered.
S The peak units do not accept other peak units or RMS as input − only the 4 analog inputs
are allowed
QuantumX HBM
128 Functions and outputs
max Each peak detection unit can monitor either the Min or the Max of one of the module’s 4
analog input channels. A peak unit may operate in different modes:
S RUN: peak value will be continuously updated
S HOLD: last peak value will be ”frozen”
S PEAK: peak detection enabled
S FOLLOW: peak detection disabled, i.e. the channel yields the original signal of the input
channel
Measuring
Output signal
Vi, Vo
t
Funktion Run Hold Run
Betriebsart Peak Follow
HBM QuantumX
Functions and outputs 129
Where f(x) denotes the input channel signal and T the time window (in ms).
Keep in mind the following notes when working with RMS channels:
S The maximum sample rate is restricted to 4800 (2400) Hz
S The output (sample) rate of the RMS channel must not be higher than the sample rate of
the input channel
S Filters set for the MX410 do not apply to RMS channels. These channels are always
unfiltered. However, the input signal is filtered
QuantumX HBM
130 Functions and outputs
7.2 MX460
The MX460 supports 4 special math channels for analysis of rotational parts: Rotational
vibration and angle difference.
HBM QuantumX
Functions and outputs 131
7.3 MX878
The MX878 module is a module with eight analog outputs that can be accessed on the front
panel via BNC sockets.
The MX878 can receive all signals that are isochronously available on the Firewire.
The settings for this are implemented with the QuantumX Assistant. Before output at the
analog output, the signal passes through an output characteristic curve parameterized by the
user (2 point scaling) and a filter also parameterized by the user. In addition, the rate for DAC
is reduced to 96 kS/s through interpolation.
QuantumX HBM
132 Functions and outputs
HBM QuantumX
Functions and outputs 133
max Keep in mind the following notes when working with peak monitor channels:
S The maximum sample rate is restricted to 4800 Hz
S The sample rate of the peak monitor channel must not be higher than the sample rate of
the input signal
S The peak units do not accept other peak units or RMS as input
Each peak detection unit can monitor either the Min or the Max of one of 4 signals ”visible as
isochronous” within the system.
Measuring
Output signal
Vi, Vo
t
Funktion Run Hold Run
Betriebsart Peak Follow
QuantumX HBM
134 Functions and outputs
You need to connect all modules via FireWire (or use a backplane) for the MX878 to become
operational! Refer to the ”ISO” column in the catman channel configuration window. This
column indicates if a channel provides its data via the isochronous link (indicated by a
symbol). Click the column or use the column’s context menu to enable or disable
isochronous transfer for a channel.
Providing the data via isochronous transfer may use up significant computing power on the
module (in particular on the MX410 and MX460 high−speed modules). Do not activate the
isochronous transfer needlessly!
HBM QuantumX
Functions and outputs 135
7.4 MX471
The modular QuantumX data acquisition system can be individually assembled. FireWire
enables quantities measured using amplifiers or computed values to be transmitted to
specific modules in real time (isochronous signals).
For example: To send a measured temperature as a so−called ”Process Data Object” (PDO)
over the CANbus via a specific node of the MX471, the temperature signal simply has to be
dragged onto the node using drag & drop. Parameterization is then performed automatically.
The CAN message needs to be completed in the corresponding dialog.
A parameterized MX471 node enables any signals or measured quantities to be cyclically
sent as CAN messages.
The CAN messages are sent to the CANbus at the data rate set for the source signal. With a
data rate > 4.800 Hz, the transmission rate is limited. A configurable divider enables the
transmission rate on the CANbus to be generally reduced.
It is essential to specify via software the baud rate and, if necessary, activation of a
termination resistor (bus termination) to ensure correct bus parameterization.
Note
In general, a CAN network enables both message types (standard/extended) to be
used.
No analysis of Sync messages or Remote frames is made.
The four CAN nodes are handled with equal priority.
A node parameterized as sender, works as an autonomous gateway and does no
longer require connection to a PC.
SDOs are not supported; the MX471 can thus not be considered as a CANopen slave.
The individual nodes’ statuses are directly displayed at the respective connection
point.
CAUTION
If a CAN node uses an incorrect baud rate for transmission over the bus, the entire
data transmission on the bus may fail.
QuantumX HBM
136 Accessories
8 Accessories
System accessories
Article Description Order No.
Module carrier QuantumX (standard) Module carrier for maximum 9 modules of the QuantumX 1-BPX001
family;
General information:
− Wall or control cabinet installation (19”);
− Connection of external modules via FireWire possible;
− Power supply: 24 V DC;
− Power consumption: Max 5 A (150 W);
Note: Only modules with degree of protection IP20 can be
inserted.
Housing connection element for QuantumX Mechanical connection elements for QuantumX modules 1-CASECLIP
modules (IP20 / IP65); Set consisting of 2 housing clips, including
assembly material for fast connection of 2 modules.
Module accessories
Voltage supply
IP20 IP65
AC−DC plug−in power supply unit; Input: 100-240 V Connector − Voltage supply QuantumX
AC ("10%); (module in protection class IP65)
1.5 m cable with international plug set
Output: 24 V DC, Max. 1.25 A; 2 m cable with plug for
IP20 modules
Order No.: 1-NTX001 Order No.: 1-CON−P1001
3 m cable for voltage supply of a QuantumX module;
Plug for IP20 modules on one side and exposed
wires at the other end.
Note when using multiple modules: The supply
voltage can be looped through FireWire connections
(max. 1.5 A)
Order No.: 1-KAB271-3
FireWire
IP20 IP65
FireWire cable (module to FireWire connection cable between QuantumX FireWire connection cable between QuantumX
module) modules in IP20 design (length: 0.2m/2m/5m); Fitted modules in IP65 design (length: 0.2m/2m/5m); Fitted
both ends with appropriate plugs. both ends with appropriate plugs.
Note: The cable can be used to optionally supply Note: The cable can be used to optionally supply
connected QuantumX modules with voltage (max. 1.5 connected QuantumX modules with voltage (max. 1.5
A, from source to last acceptor). A, from source to last acceptor).
Order No.: 1-KAB269-0.2 Order No.: 1-KAB272-0.2
Order No.: 1−KAB269−2 Order No.:: 1−KAB272−2
Order No.: 1−KAB269−5 Order No.: 1−KAB272−5
1−CASEFIT Fitting panel for mounting of QuantumX modules Order No.: 1-CASEFIT
using case clips (1−CASECLIP), lashing strap or
cable tie. Basic fastening by 4 screws
HBM QuantumX
Accessories 137
FireWire
IP20 IP65
Hub to module connection FireWire connection cable between hub and FireWire connection cable between hub and
cable 3 m QuantumX modules in IP20 design (length: 3 m); QuantumX modules in IP65 design (length: 3 m);
Fitted both ends with appropriate plugs. Fitted both ends with appropriate plugs.
Note: The cable can be used to optionally supply Note: The cable can be used to optionally supply
connected QuantumX modules via the hub with connected QuantumX modules or hub with voltage
voltage (max. 1.5 A, from source to last acceptor). (max. 1.5 A, from source to last acceptor).
Order No.: 1-KAB275-3 Order No.: 1-KAB276-3
PC to module connection FireWire connection cable between PC and
cable 3 m QuantumX module (length: 3 m); Fitted both ends
with appropriate plugs.
Note: The cable cannot be used to supply QuantumX
modules with voltage.
Order No.: 1−KAB270-3
FireWire PC-Card FireWire PC−Card with FireWire 1394b interface for
connection of QuantumX amplifiers to notebook or
PC (via PCCARD adapter)
Order No.: 1-IF−001
Ethernet
Ethernet crossover cable Ethernet patch cable between PC and QuantumX
module in IP65 design (length: 5 m); Fitted both ends
with appropriate plugs.
Order No.: 1-KAB239-2 Order No.: 1-KAB273-5
Transducer side
General information
Article Description Order No.
D−Sub−HD 15-pin plug set with TEDS chip Plug kit D−Sub−HD 15-pin (male) with TEDS chip for storing a 1-SUBHD15-MALE
sensor data sheet; housing Metallized plastic with knurled
screws.
Note: The TEDS chip is blank.
Port Saver D−Sub−HD 15 pin Four fully-wired male to female port savers protecting the 1−SUBHD15−SAVE
wear and tear of D−Sub−HD 15 pin ports for frequent
plugging and unplugging of transducers. Extends contact
durability by min. 500.
This adaptor attaches securely with screws.
SCM−SG120/350 SCM−SG350 Strain Bridge Module (120 Ohm or 350 Ohm) 1−SCM−SG120
Signal conditioning module for QuantumX full bridge input 1−SCM−SG350
(D−Sub−HD15 pin).
Integrated 350 Ohm bridge completion resistor, shunt
calibration, TEDS, D−Sub−HD-device connection.
Soldering point for transducer cable.
Adapter D−Sub−HD 15-pin to D−Sub 15-pin. Adapter D−Sub HD 15-pin to D−Sub 15-pin for connecting 1-KAB416
transducers with pre−assembled D−Sub plugs on MX840
(length approx. 0.3 m)
Note: Pre-assembled for full bridge (6-wire).
QuantumX HBM
138 Accessories
MX840/MX840A accessories
Article Description Order No.
Cold junction for thermocouple on MX840, Electronics for temperature compensation for measurements 1-THERMO-MXBOARD
MX840A, MX440 with thermocouples on MX840, MX840A, MX460, comprising:
− PT1000 cold junction
− Including 1-wire TEDS chip for transducer identification
Note: Mounting in D−Sub−HD 15-pin transducer plug.
Adapter D−Sub−HD15 to D−Sub9 (CAN) on Adapter for connecting CAN devices to MX840, MX840A, 1-KAB418
MX840, MX840A MX460. D−Sub−HD 15-pin (plug) to D−Sub 9-pin (socket);
(length: approx. 30 cm).
MX410 accessories
Article Description Order No.
BNC adapter for IEPE sensors with BNC Adapter for connecting IEPE sensors with BNC connection 1-IEPE-MX410
cable to MX410 (BNC socket to 15-pin D−Sub−HD; length:
approx. 7 cm).
HBM QuantumX
Accessories 139
Housing clip
QuantumX HBM
140 Accessories
Mains Modules
UK mains cable
HBM QuantumX
Accessories 141
3m 30
Approx. 10−15 mm
black 0V
brown + Power supply
8.3 FireWire
0.2 m
2.0 m
5.0m
Plug ODU
SX1LOC−P08MFG0−0001
3m
FireWire B-plug
PC adapter socket
HBM QuantumX
Accessories 143
Plug kit D−Sub−HD 15-pin (male) with TEDS chip for storing a sensor data sheet.
Order No.: 1-SUBHD15-MALE
Front Back
When frequently connecting and disconnecting transducers we recommend that you use port
savers for protecting the transducer sockets of the QuantumX module. The port saver is
easy to screw into place and can be replaced after several hundred mating cycles. No
expensive module repair is necessary any more.
QuantumX HBM
144 Accessories
bk
bu
gy
gn
D-SUB HD plug wh D-SUB socket
rd
Wire bridge
CAUTION
This cable is only for transducers with full bridge and 6-wire circuits! If other transducers are connected,
the universal amplifier can be damaged or even destroyed.
HBM QuantumX
Accessories 145
Electronics for temperature compensation for measurements with thermocouples. Board for installation in a 15-pin
D−Sub−HD plug.
View from above View from below
12
Position of board in
plug
6 10
12,5
1 5
PINs to be soldered
QuantumX HBM
146 Accessories
bk
bu
gy
gn D-SUB socket
D-SUB HD plug
wh
rd
Hsg. Hsg.
HBM QuantumX
Accessories 147
RFID
21
4,9
12,1
All dimensions in mm
QuantumX HBM
148 Accessories
9 Support
Headquarters world-wide
Europe
Hottinger Baldwin Messtechnik GmbH:
Im Tiefen See 45, 64293 Darmstadt, Germany
Tel. +49 6151 8030, Fax +49 6151 8039100
E−mail: info@hbm.com
www.hbm.com
Asia
Hottinger Baldwin Measurement (Suzhou) Co., Ltd.
106 Heng Shan Road, Suzhou 215009, Jiangsu, VR China
Tel. (+86) 512 68247776, Fax (+86) 512 68259343
E−mail: hbmchina@hbm.com.cn
HBM QuantumX
Support 149
QuantumX HBM
E Hottinger Baldwin Messtechnik GmbH. All rights reserved.
All details describe our products in general form only.
They are not to be understood as express warranty and do
not constitute liability whatsoever.
7−2002.3031