You are on page 1of 3

ORIGINAL STUDY

Confirmation of High Specificity of an Automated


Enzyme Immunoassay Test for Serological Diagnosis
of Syphilis: Retrospective Evaluation Versus Results
After Implementation
Laura van Dommelen, MD, PhD,* Christian J.P.A. Hoebe, MD, PhD,*† Frank H. van Tiel, MD, PhD,*
Carel Thijs, MD, PhD,‡ Valère J. Goossens, MD, PhD,* Cathrien A. Bruggeman, PhD,*
and Inge H.M. van Loo, MD, PhD*
We have previously evaluated the performance of the
Background: The optimal algorithm for serological syphilis screening Bioelisa Syphilis 3.0 EIA as a primary screening assay, in com-
is still a matter of debate. We have previously evaluated the perfor- parison with the performance of the Treponema pallidum particle
mance of the Bioelisa Syphilis 3.0, using a selection of archived sera, agglutination (TPPA) in a selected collection of serum samples.1
and in this study compare these results with the Bioelisa results after This EIA detects IgM and IgG antibodies against T. pallidum
clinical implementation. antibodies. The sample collection included 145 sera from syphilis
Methods: All Bioelisa Syphilis 3.0 results obtained since clinical imple- patients with active or latent disease, 41 sera from healthy con-
mentation were analyzed. Bioelisa-positive or borderline samples were trols, and 144 sera from patients with underlying conditions which
retested using Treponema pallidum particle agglutination, rapid plasma may influence T. pallidum antibody testing. The sensitivity and
reagin test, fluorescent treponemal antibody-absorption test, and/or im- specificity were both 100% (95% confidence interval [95% CI]
munoblot. On sera sent in together with cerebrospinal fluid, occasionally sensitivity 97.5%–100%, specificity 98.0%–100%). Because the
both the T. pallidum particle agglutination and Bioelisa were performed. Bioelisa Syphilis 3.0 can be used on an automated EIA reading
Results: The Bioelisa was performed on 14,622 sera. Bioelisa-positive system, this EIA was implemented in our laboratory to replace
samples, which were not retested by the previously described assays, the TPPA as a screening method.
were withdrawn from the database (n = 36). In 1.3% of the samples Because this high specificity was based on the results of
(187/14,586), the Bioelisa was positive or borderline and, ultimately, tests performed on a selected sample collection, with an artifi-
115 sera were considered true positive (prevalence 0.8%). The specificity cially high syphilis prevalence (44%), it was not surprising that
of the Bioelisa was 99.5%. we noticed a relative high number of false-positive results in the
Conclusions: Based on the results of all performed diagnostic assays, first year after implementation of the Bioelisa as a screening-assay
the specificity of the Bioelisa of 99.5% is very consistent with that found for syphilis. A false-positive Bioelisa was defined as a positive re-
in the initial study (100%; 95% confidence interval was 98.0%–100%). sult in the Bioelisa with all confirmatory testing with TPPA, immuno-
Interpreting (positive) test results is difficult in the absence of a gold stan- fluorescense, rapid plasma regain, and/or additional testing with an
dard, especially when the disease prevalence is low. Results should be immunoblot at a reference laboratory, being negative. Therefore,
viewed in the light of the patients' characteristics. we decided to perform a retrospective analysis of the Bioelisa Syphilis
3.0, to assess whether the high specificity found in the initial evalua-

I n the absence of a gold standard, the optimal serological syphilis


screening algorithm is still a matter of debate. Generally, there
are 2 screening methods: (1) using a treponemal assay (e.g.,
tion would stand up in clinical practice.

enzyme immunoassay [EIA]), which is called the “reverse algo- MATERIALS AND METHODS
rithm” in the literature, or (2) a nontreponemal assay (rapid The Bioelisa Syphilis 3.0 (Biokit SA, Barcelona, Spain)
plasma reagin test [RPR] or Venereal Disease Research Labora- was introduced in our laboratory on October 1, 2007. All Bioelisa
tory). Also, there is variation in the confirmatory assays used. results (n = 14,622) obtained between October 2007 and February
2010 were analyzed. The Bioelisa detects IgM and IgG antibodies.
From the *Department of Medical Microbiology, CAPHRI School of Pub- If the Bioelisa was positive, the TPPA (MHA-TP; Fujirebio,
lic Health and Primary Care, Maastricht University Medical Centre, Tokyo, Japan), RPR (Syfacard-R*; Abbott Murex, Dartford,
AZ, Maastricht, the Netherlands; †Department of Sexual Health, Infec-
UK), and fluorescent treponemal antibody-absorption test (FTA-
tious Diseases and Environmental Health, Public Health Service South
Limburg, HA, Geleen, the Netherlands; and ‡Department of Epidemi-
Abs; Trepo Spot IF, BioMerieux SA, Marcy l'Etoile, France) were
ology, CAPHRI School of Public Health and Primary Care, Maastricht additionally performed on this sample for confirmation. The TPPA
University Medical Centre, AZ, Maastricht, the Netherlands and FTA-Abs both detect IgM and IgG antibodies against T.
Conflict of interest: None declared. pallidum. The RPR uses tissue lipid cardiolipin (antigen) to detect
Correspondence: Laura van Dommelen, MD, PhD, Department of Medical “reagin,” that is, antibodies directed against tissue components,
Microbiology, CAPHRI School of Public Health and Primary Care, which appear in blood of a patient with syphilis as a reaction to
Maastricht University Medical Centre, PO Box 5800, 6202 AZ, Maastricht, tissue damage. If the previously mentioned serological tests yielded
the Netherlands. E-mail: lauravandommelen@yahoo.com. an inconclusive serological profile (e.g., a discrepancy between the
Financial support: None declared. Bioelisa and TPPA and/or FTA Abs results), the serum was sent
This study is approved by the Medical Ethics Committee of the Maastricht
University Medical Centre (no. 12-4-118) to the national reference laboratory (National Institute of Public
Received for publication July 30, 2014, and accepted December 12, 2014. Health and the Environment, Bilthoven, the Netherlands) for addi-
DOI: 10.1097/OLQ.0000000000000240 tional testing with an immunoblot. The TPPA was occasionally
Copyright © 2015 American Sexually Transmitted Diseases Association performed on Bioelisa-negative sera, which were sent to our labo-
All rights reserved. ratory together with cerebrospinal fluid, to exclude neurosyphilis. All

120 Sexually Transmitted Diseases • Volume 42, Number 3, March 2015

Copyright © 2015 by the American Sexually Transmitted Diseases Association. Unauthorized reproduction of this article is prohibited.
Automated EIA Test for Diagnosis of Syphilis

TABLE 1. Bioelisa Results After Clinical Implementation2,3


Final Conclusion
Positive Negative Total 95% CI
Total Bioelisa Positive 115 72 187 Se 100% 96.8%–100%
Negative 0 14399 14399 Sp 99.5% 94.6%–99.6%
Total 115 14471 14586 PPV 61.5% 54.1%–68.5%
NPV 100% 100%–100%
Syphilis prevalence 0.79%

No clinical data Bioelisa Positive 67 26 93 Se 100% 94.6%–100%


Negative 0 4684 4684 Sp 99.5% 99.2%–99.6%
Total 67 4710 4777 PPV 72.0% 61.8%–80.9%
NPV 100% 100%–100%
Syphilis prevalence 1.40%

STD screening Bioelisa Positive 38 24 62 Se 61.3% 90.1%–73.4%


Negative 0 4468 4468 Sp 99.5% 99.4%–99.7%
Total 38 4492 4530 PPV 61.3% 48.1%–73.4%
NPV 100% 99.9%–100%
Syphilis prevalence 0.55%

Pregnancy Bioelisa Positive 3 11 14 Se 100% 29.2%–50.8%


Negative 0 3154 3154 Sp 99.7% 99.4%–99.8%
Total 3 3165 3168 PPV 21.4% 4.6%–50.8%
NPV 100% 99.9%–100%
Syphilis prevalence 0.09%

Se indicates sensitivity; Sp, specificity; NPV, negative predictive value; STD, sexual transmitted diseases; 95% CI, exact binomial 95% CI, confidence
interval; PPV, positive predictive value.

diagnostic assays were performed according to the manufacturers' Bioelisa was positive or borderline. Because only 6 of these 187
instructions, as previously described.1 Data were unlinked from pos- samples were tested borderline and confirmatory testing was similar
sible patient identifiers, and each patient was included only once in to the Bioelisa-positive samples, these samples were considered
the database. Results were analyzed using the 2-way contingency Bioelisa positive in further analysis. Sixty-two Bioelisa-positive sam-
table analysis (ctab2x2) and exact binominal CIs calculation ples (33%) showed discrepant results in the confirmatory assays and
(confint) programs from www.statpages.org.2,3 A test result were sent to the national reference laboratory. Immunoblot testing re-
was considered as true positive if the Bioelisa was positive and sulted in an additional 6 true positive samples. All Bioelisa border-
was confirmed by a positive TPPA and FTA-Abs, or by a positive line samples were TPPA negative and considered as true negatives.
immunoblot result. Exceptions are displayed in Table 2. All Because TPPA is routinely performed on serum and cerebrospinal
Bioelisa-negative samples were considered true negative results, fluid to exclude neurosyphilis, confirmatory assays, including TPPA,
based on the results of our previous study. were also performed on 41 such serum samples that had been
tested Bioelisa negative. One of these samples showed reaction
RESULTS in the TPPA (1:80) and FTA-Abs, whereas RPR result was nega-
All Bioelisa-positive samples that were not tested according tive and reference laboratory conclusion stated “borderline” in
to the algorithm described in the previous section were withdrawn the immunoblot. In total, 115 (0.8%) of 14,586 samples were con-
from the database (n = 36). In 1.3% of samples (187/14,586), the sidered true positives (Tables 1 and 2).

TABLE 2. Results of Confirmatory Tests in Bioelisa-positive samples*


TPPA RPR FTA Reference Laboratory Final Conclusion Positive (n) Final Conclusion Negative (n) Total (n)
Positive Positive Positive NA 56 0 56
Positive Negative Positive NA 51 0 51
Positive Negative Positive Positive 4 0 4
Positive Negative Positive Negative 0 1 1
Negative Positive Negative Negative 0 1 1
Negative Negative Positive Positive 2 1† 3
Negative Negative Positive Negative 0 18 18
Negative Negative Positive NA 2 1 3
Negative Negative Positive Negative 0 5 5
Negative Negative Negative NA 0 15 15
Negative Negative Negative Positive 0 3† 3
Negative Negative Negative Negative 0 27 27
115 72 187

*Postive and bordeline positive samples are grouped in this table to preserve readability.

Result from reference laboratory bordeline positive.
NA indicates not applicable.

Sexually Transmitted Diseases • Volume 42, Number 3, March 2015 121


Copyright © 2015 by the American Sexually Transmitted Diseases Association. Unauthorized reproduction of this article is prohibited.
van Dommelen et al.

The main reasons for syphilis screening were sexually trans- above-mentioned results, it is unlikely that a significant number
mitted disease screening (n = 4530; 31% of all samples), screening of patients were missed using the Bioelisa. An overview of the
during pregnancy (n = 3168; 22%), and workup for infertility performance of several other syphilis EIAs can be found in our pre-
(n = 1167; 8%). In another 6% (n = 944) of the cases, syphilis vious article.1 In general, EIAs are considered equally sensitive or
screening was requested for other reasons (e.g., suspicion of more sensitive for the detection of T. pallidum antibodies compared
neurosyphilis), leaving 33% (n = 4777) of the patients for whom with TPPA or T. pallidum hemaglutinine agglutination.9–13
no clinical data were available. Considering all true positive samples, Another interesting point in the study by Castro et al.,9
clinical data were unfortunately only available for 1 patient: an HIV- mentioned previously, is that they question the use of the FTA-Abs
positive man with lymphadenopathy, malaise, and weight loss (EIA as a confirmatory assay based on their results because it was the least
positive, TPPA negative, RPR negative, FTA-Abs strongly positive). sensitive. Also, the TPPA was found less sensitive compared with the
Based on the results of all performed diagnostic assays, the Bioelisa in their study. Several underlying conditions, such as preg-
specificity of the Bioelisa was 99.5% (95% CI, 99.4%–99.6%) and nancy or autoimmune disease, can interfere with all syphilis assays,
the positive predictive value (PPV) was 62% (95% CI, 54.1%– resulting in false-positive results.1 In case of discrepancy between as-
68.5%; Table 1). The overall specificity of 99.5% is very similar says and/or weakly positive results, it is therefore difficult to assess if
to that found in the initial evaluation (100% with the 95% CI being it concerns false-positive results or (weak) true positive results. Un-
98.0%–100% in the initial study), with an even narrower CI of spec- fortunately, the sensitivity of syphilis nucleic acid amplification tests
ificity in this study (95% CI, 99.4%–99.6%). When analyzing the in the absence of skin lesions is poor.13–15 As shown in Table 2, even
results of the largest patient categories according to available clinical with the same test results, it is not always easy to come to a final con-
data, the values of specificity were very similar (Table 1). clusion. In the absence of a gold standard, results should always be
viewed in light of the patients' medical history, symptoms, and sex-
DISCUSSION ual history before treatment decisions are made.
Performing a prospective evaluation is time consuming and REFERENCES
costly. It took us several years to gather 115 serologically confirmed
1. van Dommelen L, Smismans A, Goossens VJ, et al. Evaluation of a
syphilis cases. In our previous evaluation study, we have therefore rapid one-step immunochromatographic test and two
used a selection of archived sera. In the present study, we have com- immunoenzymatic assays for the detection of anti–Treponema
pared the results of the previous study with the performance after pallidum antibodies. Sex Transm Infect 2008;84:292–296.
implementation of this assay in our laboratory. In the initial evaluation, 2. Clopper CS, Pearson ES. The use of confidence or fiducial limits
all syphilis serologically positive samples comprehended 44% of the illustrated in the case of the binomial. Biometrika 1934;26:404–413
total number of sera. The prevalence of syphilis in the Netherlands, (http://statpages.org). Accessed April 13, 2014.
however, is only 0.2% among pregnant women and 2.3% in men hav- 3. Rosner B. Fundamentals of Biostatistics, 2006. Available at:
ing sex with men.4 Our current analysis shows that the high specificity http://statpages.org). Accessed April 13, 2014.
found in the initial study stands up after implementation in a popula- 4. Vriend HJ, Koedijk FDH, van den Broek IVF, et al. Sexually
transmitted infections, including HIV, in the Netherlands in 2010.
tion with a low syphilis prevalence (0.8%). The specificity was sim- Bilthoven, the Netherlands: National Institute for Public Health and
ilar in all subgroups of patients, and hence, differences in PPV are the Environment, 2011.
solely due to differences in the prevalence of syphilis, as can be seen 5. Billings PR, Bernstein MS. Physicians poor at prevalence and positive
in Table 1, with PPVs ranging from 21.4% (in the low-risk group of predictive value. JAMA 1985;254:1173–1174.
pregnancy screening) to 72.0% (in the group without clinical data). 6. Lijmer JG, Mol BW, Heisterkamp S, et al. Empirical evidence of
Even with a very high specificity, the high number of nondiseased design-related bias in studies of diagnostic tests. JAMA 1999;282:
can result in a substantial number of false-positive results. Under- 1061–1066.
standing this phenomenon is important when interpreting a test re- 7. Ebel A, Bachelart L, Alonso JM. Evaluation of a new competitive
sult.5 Confirmatory assays are therefore essential in the serological immunoassay (BioElisa Syphilis) for screening for Treponema pallidum
antibodies at various stages of syphilis. J Clin Microbiol 1998;36:358–361.
diagnosis of syphilis, especially in low-prevalence patient groups. 8. Donkers A, Levy HR, Letens-van Vliet A. Syphilis detection using
A comparison between both our evaluations is hampered the Siemens ADVIA Centaur Syphilistreponemal assay.
by the fact that, in our initial study, the TPPA was performed on Clin Chim Acta 2014;433C:84–87.
all samples, whereas in our current analysis—performed after 9. Castro A, Jost H, Cox D, et al. A comparison of the analytical level
implementation—the TPPA was almost strictly limited to Bioelisa- of agreement of nine treponemal assays for syphilis and possible
positive samples. This causes a verification bias compared with implications for screening algorithms. BMJ Open 2013;3:e003347.
the initial study: not all samples were subjected to the reference 10. Binnicker MJ, Jespersen JD, Rollins LO. Treponema-specific tests
test.6 False-negative Bioelisa results could have been missed. for serodiagnosis of syphilis: comparative evaluation of seven assays.
As was seen in our evaluation, one Bioelisa-negative sample was J Clin Microbiol 2011;49:1313–1317.
11. Maple PA, Ratcliffe D, Smit E. Characterization of Treponema
reactive in the TPPA and FTA-Abs and borderline reactive in the pallidum particle agglutination assay–negative sera following
immunoblot. In our previous evaluation, however, the Bioelisa screening by treponemal total antibody enzyme immunoassays.
did not miss any positive sample. The first version of the Bioelisa Clin Vaccine Immunol 2010;17:1718–1722.
Syphilis was evaluated by Ebel et al.7 in 1998 using 824 sera. 12. Sena AC, White BL, Sparling PF. Novel Treponema pallidum
The T. pallidum hemagglutination assay and FTA-Abs were both serologic tests: A paradigm shift in syphilis screening for the 21st
positive in 53% of these samples. The sensitivity and specificity century. Clin Infect Dis 2010;51:700–708.
found in their study were 99.5% and 99.4%, respectively. The 13. Vulcano F, Milazzo L, Volpi S, et al. Italian national survey of blood
current version of the assay (3.0) was recently evaluated by donors: external quality assessment (EQA) of syphilis testing.
Donkers et al.8 in 1251 samples (syphilis prevalence 8.0%), and J Clin Microbiol 2010;48:753–757.
14. Grange PA, Gressier L, Dion PL, et al. Evaluation of a PCR test for
they found an agreement of 100% with the other 2 EIAs used detection of treponema pallidum in swabs and blood. J Clin Microbiol
in their study. Castro et al.9 analyzed the analytical level of agree- 2012;50:546–552.
ment between several EIAs in comparison with the TPPA and 15. Heymans RJJ, van der Helm JJ, de Vries HJ, et al. Clinical value of
found that the Bioelisa (version not stated) was relatively sen- Treponema pallidum real-time PCR for diagnosis of syphilis.
sitive compared with the other EIAs used. Taking together all the J Clin Microbiol 2010;48:497–502.

122 Sexually Transmitted Diseases • Volume 42, Number 3, March 2015

Copyright © 2015 by the American Sexually Transmitted Diseases Association. Unauthorized reproduction of this article is prohibited.

You might also like