You are on page 1of 20

FCI ST100 Series Thermal Mass Flow Meters

Advanced Solutions for Process and Plant


Gas Flow Metering
Mass Flow, Temperature and Pressure

FCI ST100 SERIES


Chemical
Wastewater Treatment, Landfills
Refineries
Oil & Gas
Mining
Metals
Manufacturing
Cement, Stone, Brick, Glass
Power Utilities
Pulp & Paper
Food & Beverage
And more...
ST100 Series Features
g Direct gas mass flow measuring
g Flow rate, total flow, temperature,
pressure measurement
g No moving parts, lowest maintenance
g Most extensive and comprehensive
global approvals for Ex installations
Model ST100 g Temperature service to 850 °F [454 °C]
with -FP style
flow element g Rugged, industrial transmitter enclosure
Model ST100 readout
g Graphic, multivariable LCD readout
g Integral and remote electronics versions
g Multi-gas calibrations
g Multiple analog outputs ­– 4-20 mA, pulse
g HART®, Foundation™ Fieldbus, PROFIBUS PA,
Modbus RS-485
g Dual-element systems
Model ST100L in-line
flow meter
g Patented VeriCal™ In-Situ Calibration
Verification System
g SIL compliant
g On-board data logger
g Sensor configurations to optimize application
performance
g Extensive selection of process connections

ST100 Series – The Future of Gas Flow Metering


With FCI’s new ST100 Series, the science of process gas flow or upgrade, ST100 can be converted to any of these outputs with
metering has never been so advanced or so future-ready! a card change, in the field. That takes “never obsolete” to a whole
ST100 is an all-new thermal dispersion technology gas flow meter new level.
that combines feature- and function-rich electronics with the industry’s Then there’s ST100’s display/readout which brings new meaning
most advanced flow sensors to achieve a truly state-of-the-science to the term “process information.” Simply, the ST100’s graphical
flow metering solution for industrial processes and plant applications. display provides the most comprehensive information on process
The ST100 Series transmitter is unsurpassed in meeting measurements and conditions of any thermal flow meter available
your current and future need for outputs, process information and (see page 7).
communications. Whether your output needs are traditional 4-20 mA In the gas stream, ST100 flow elements and process connections
analog, frequency/pulse or advanced digital bus communications ensure the best possible measurements and effective installation.
such as HART, Foundation Fieldbus, PROFIBUS or Modbus, ST100 FCI flow element designs have been perfected and optimized through
has your solution. What’s more, should you ever need to change six decades of R&D and real world experience in thermal dispersion

2
technology and flow metering applications. With thermal dispersion Find your gas here?
there are no moving parts to foul or clog. This results in the greatest
reliability and valuable savings on maintenance. FCI has provided thermal mass flow meter solutions
Fast response, accuracy, repeatability and long-life are for all of these and more…
hallmarks of FCI flow elements available with ST100. And, all new,
are the STP models that include pressure measurement. Mass flow, Acetaldehyde Ethyl Acrylate Ketene Phenol
Acetic Acid Ethyl Alcohol Krypton Phosgene
temperature and pressure in a single tap point thermal dispersion
Acetone Ethyl Amine Landfill Gas Propadiene
meter — that’s an industry first and FCI exclusive! Acetonitrile Ethyl Benzene M-Cresol Propane
Then, consider other ST100 Series exclusives like dual- Acetyl Chloride Ethyl Bromide Mercury Propanol
element averaging systems (see page 8) and FCI’s patented Air Ethyl Chloride Methane Propyl Chloride
VeriCal option that provides significant cost and time savings by Allyl Chloride Ethyl Fluoride Methanol Propylene
Ammonia Ethyl Mercaptan Methyl Acetate Propylene Oxide
performing dynamic wet calibration verification in-situ, without
Aniline Ethylene Methyl Alcohol Propyne
ever removing the instrument from the process (see page 10). To
Argon Ethylene Methyl Amine P-Xylene
ensure both ease of installation and a match with your process Benzene Dichloride Methyl Butane R-11
conditions, the ST100 Series offers a comprehensive selection of Bio-Gas Ethylene Oxide Methyl Fluoride R-12
process connections and installation options. Boron Trifluoride Flare Gas Methyl Formate R-13
Your smart investment in an ST100 Series flow meter is Bromine Fluorine Methyl Hexane R-13B1
Bromobenzene Fluorobenzene Methyl Hydrazine R-14
complete with FCI precision calibration. FCI’s NIST traceable flow
Butadiene Fluoroform Methyl R-21
calibration facility is unmatched by any other thermal dispersion Freon-11 Mercaptan
Butene R-22
flow meter manufacturer. Your ST100 Series flow meter is custom Butylene Oxide Freon-12 Methyl Octane R-23
calibrated to exacting process and fluid conditions, essential to Butyne Freon-13 Methyl Pentane R-112
providing the highest precision, longest lasting solution that meets Carbon Dioxide Freon-14 Methylal R-113
its specifications and your specific application. Carbon Disulfide Freon-21 Methylene R-114
Freon-22 Chloride R-114B2
Carbon Monoxide
Freon-23 Morpholine R-115
Carbon
Tetrachloride Furan M-Xylene R-116
ST100 Series Model Summary Carbonyl Sulfide Halon Naphthalene R-134A
Chlorine Helium Natural Gas R-142B
ST Series N-Butane
Chlorobenzene Heptene R-152A
Model ST100 ST100L ST102 ST110 ST112 Chloroethane N-Butane
Hexanol R-216
Basic Type Insertion In-Line Insertion Insertion Insertion Chloroform Hexene N-Butanol R-500
Flow a a a a a Chloromethane Hydrazine N-Butyl Alcohol R-502
N-Decane
Temperature a a a a a Chloroprene Hydrogen
N-Dodecane
R-503
Cis-2-Butene Hydrogen R-504
Dual-Element a a Cis-2-Hexene Bromide Neon R-C318
VeriCal In-Situ
Calibration Verification a a Compressed Air Hydrogen Neopentane Radon
Chloride N-Heptane
STP Series Cumene Silane
Hydrogen N-Hexane
Cyanogen Silicon
Model STP100 STP102 STP110 STP112 Cyanide
Cyclobutane Nitric Oxide Tetrachloride
Basic Type Insertion Insertion Insertion Insertion Hydrogen
Cyclohexane Nitrogen Styrene
Deuteride
Flow a a a a Cyclooctane Hydrogen
Nitrogen Dioxide Sulfur Dioxide
Temperature a a a a Cyclopentane Fluoride Nitromethane Sulfur
Hexafluoride
Nitrous Oxide
Pressure a a a a Cyclopropane Hydrogen Iodide
N-Nonane Sulfur Trioxide
Hydrogen
Dual-Element a a Decene
Deuterium Peroxide N-Octane Superheated
VeriCal In-Situ Thiophene
Calibration Verification a a Deuterium Oxide Hydrogen Sulfide Nonene
Titanium
Diethyl Amine Iodine N-Pentane Tetrachloride
Diethyl Ether Isobutane N-Propanol Toluene
The ST100 Series is comprised of two core families – the “ST” and the Diethyl Ketone Isobutene N-Propyl Alcohol Trans-2-Butene
“STP,” with nine total models. The ST family features instruments that Digester Gas Isobutyl Alcohol N-Propyl Amine Trimethyl Amine
measure both mass flow and temperature. The STP family is an FCI Dimethyl Ether Isoheptane N-Undecane Triptane
exclusive thermal dispersion flow meter that additionally measures a Dimethyl Isohexane Octene Uranium
Propane Isooctane Oxygen Hexafluoride
third parameter – pressure. Within both families are basic single-point
Dimethyl Sulfide Isopentane O-Xylene Vinyl Acetate
models, dual-element averaging systems and models with VeriCal, an Ethane Isoprene Ozone Vinyl Chloride
FCI patented in-situ calibration verification system. Ethanol Isopropyl Alcohol Pentanol Vinyl Fluoride
Ethyl Acetate Isopropyl Amine Pentene Vinyl Formate

3
Inputs, Outputs and Bus Communications Digital Bus Communications Options
for Today and Tomorrow The ST100 Series features the most comprehensive selection of
ST100 Series’ extensive and comprehensive offering of field- digital bus communications available in a thermal dispersion flow
interchangeable outputs is your assurance of compatibility with meter. All bus communications provide actual measurements,
your process control system for the long term. Whether your including flow rate, total flow, temperature and, with STP models,
network is analog based or bus comm based, there is a compatible pressure. Instrument diagnostics and service information are
solution with ST100. And, if your plant has ever considered continuously accessible via the bus. All ST100 Series user set-up
changing or upgrading system communications in the future, the and configuration parameters may also be performed over the bus.
ST100 Series is ready to change with you. Each of the ST100 output Included with the analog output cards is a two-way HART bus
types is modular and completely interchangeable in the field, at any over the #1 4-20 mA output. The HART bus complies with revision
time. This assures adaptability and compatibility with your process level 7 protocol and is fully compatible with all versions of HART field
measurement and control strategy – today and tomorrow. communicators and HART control systems.
Instead of analog outputs / HART, you can select a digital bus of:
Foundation Fieldbus H1, PROFIBUS PA or Modbus RS-485. These bus
options are full two-way I/O communications.
HART, Foundation Fieldbus and PROFIBUS communication options
are provided with device interface software, either EDD or DTM
type, for easy, seamless integration with process control systems
and PC-based frame applications such as PACTware. These software
packages and updates can be downloaded from FCI’s website
and from the respective bus organization’s websites. FCI obtains
registration and certification on its products from the respective bus
organizations. *

www.hartcomm.org www.profibus.com www.fieldbus.org www.fdtgroup.org

* ST100 Series with Foundation™ Fieldbus communications is a certified and


Analog Outputs with HART registered device with FieldComm Group™. Registered features include alarms
and events, function block linking, trending, multi-bit alert reporting and field
Comprehensive, scalable and flexible describe the ST100 Series’ diagnostics. The ST100 was one of the first two devices to be compliant and
analog outputs bundle. The analog output card features triple certified with advanced field diagnostics. Manufacturer ID (HEX) 01FC49.
4-20 mA outputs that are fully assignable to flow rate and/or ST100 Series with HART® communications is a certified and registered device
temperature, and with STP models, pressure. All 4-20 mA outputs with FieldComm Group, registration number L2-06-1000-168. Manufacturer
are user scalable to represent the full range or any subset and ID (HEX) 00A6.
are fully assignable to any of the measured variables. Outputs are
isolated and have fault indication per NAMUR NE43 guidelines,
user selectable for high (> 21.0 mA) or low (< 3.6 mA). Digital I/O Ports
Additionally, full two-way HART communication is included and
available over the #1 4-20 mA output. The analog card also includes All ST100 Series models include a USB port for connection to a laptop
a user selectable 0-1 kHz or 0-10 kHz frequency or pulse output of or other computer for all user set-up programming, field enactable
flow. The frequency output is proportional to flow rate while the device changes and servicing diagnostics. Also provided standard
pulse output can be used for external flow totalizers. is an Ethernet 10/100 network port designed to facilitate remote
access by FCI service personnel to perform service and diagnostics.
Temporary connection to both USB and Ethernet connections can be
made easily by removing the rear cover of the instrument enclosure,
or the unit can be permanently attached through the conduit ports.

4
Digital Display / Readout
The ST100 Series’ optional digital readout provides a best-in-class troubleshooting. (In the blind unit, these functions must be performed
information suite, backlighted LCD and four (4) optical touch buttons. through the USB port or bus communications.) These buttons are
The display backlight has a unique proximity detector that illuminates activated through the front panel glass. There is no need to open
it only when a person approaches, or can be set to “always on.” The the enclosure or declassify a hazardous area to perform functions.
display and button functions can be rotated electronically, via the An atypical button sequence is required for button activation which
buttons, in 90° increments to optimize display viewing and button prevents accidental or unauthorized access to the menus.
activations. The display provides readouts of the process variables and
The display buttons provide convenient, ready access to all service diagnostics. The normal process measuring mode screen is the
instrument set-up and programming, as well as diagnostics and most comprehensive of any thermal dispersion flow meter available.

Backlighted LCD

Instrument fault/service icon  101 segment flow rate bargraph


Only appears if fault detected
Flow rate Engineering units
Flow totalizer
Temperature
Alarm status indicators
Only appear if alarm is tripped Pressure (STP models)
Current selected group
With multiple calibration groups User programmable 17 character field
Show tag, process name or gas type, etc. on display

Four (4) optical touch buttons for


programming and display selections
— Activated through front panel window; Scroll up Easy to follow sequence
no need to open enclosure menu for set-up and
selections
— Protected from accidental or
unwanted activation

Go back/
escape Enter / select

Scroll down

5
Sensors Optimized to Meet the Application Extensive Selection of Process Connections
High quality thermal mass flow measurement An unmatched selection of process connection choices assures a
begins with the sensor element in the flow stream. match for any installation requirements.
Only FCI combines six decades of field experience Insertion flow meters – For most applications an adjustable “U”
with extensive R&D and performance validation length (insertion depth) configured with compression fittings offers
“FPC” Style
Element utilizing an on-premise flow laboratory with more the most installation flexibility. Adjustable elements are available in
than 18 flow stands. FCI is the leader in thermal 6 inch [152 mm], 12 inch [305 mm], 21 inch [533 mm], 36 inch [914
mass flow sensor innovation, reducing the size mm], and 60 inch [1524 mm] lengths. All adjustable “U” elements
of instruments, accelerating response time and include permanent, laser-etched depth gauge markings, in both
improving overall operation in applications with inches and millimeters, to ensure accurate insertion depth within the
“FP” Style dramatic temperature swings. Further, ST100 installation. Male NPT compression fittings are available in either
Element
Series flow elements are all-welded to ensure 3/4 inch or 1 inch, with Teflon or metal ferrules, or with a tapped and
maximum strength, durability and leak prevention, threaded ANSI or DIN flange. Fixed length elements made to your
and they are offered standard in either 316L specifications with a fixed threaded fitting or a welded ANSI or DIN
stainless steel or Hastelloy-C276. flange are also available for all ST100 Series models.
To maximize performance and long life for For hot tap installations, retractable packing glands with a
“S” Style
Element your applications, ST100 Series is engineered and choice of pressure rating and packing material are available for
offered with a choice of flow element styles. The all ST100 insertion flow meter models. Pressure service ratings of
FPC style is an exclusive FCI design that features a patent-pending either 50 psi [3,4 bar] or 500 psi [34 bar] are standard with higher
flow conditioner, fast response time and a protective shroud. For ratings available on special request. To ensure compatibility with
many applications with clean gases and for compressed air using an the process fluid and integrity of the installation, packing material is
insertion style instrument, the FPC is the recommended solution. For selectable as either graphite or Teflon 1.
the in-line Model ST100L, and instruments with the VeriCal option,
the FP is the preferred solution. In other applications such as wet, In-line flow meter (Model ST100L) – The spool-piece flow body
dirty, erosive, or corrosive gases or with rapidly changing flow rates, length is nine times its nominal diameter, and can be fabricated as
where response time is less important, the S is regularly specified. required with male NPT, female NPT, ANSI flange or DIN flange and
Table 1 provides a summary of conditions and characteristics to butt weld prepared process connections.
consider in selecting a flow element style.
1 Teflon packing material may be required for certain process media such as
ozone, chlorine or bromine; consult FCI’s AVAL program for recommendation
ST100 Series Flow Element Selection 2 Chromium carbide coating also available ; contact FCI

Flow Element –FPC –FP –S


Conditions or Characteristics
All clean gases   
Air / compressed air   
Slightly dirty gases   
Damp / moist gases   
Very dirty gases   
Wet gases   
Erosive 2   
Corrosive   
Particulates in flow stream 2   
Pulsing    Adjustable “U” Length Tapped and threaded
Fast response time    with Compression Fitting ANSI or DIN Flange
Dynamic temperature swings   
Rapid or erratic changes in flow requir-
ing smoothed response and output    Low Pressure Packing Gland
Less-than-ideal straight run with Ball Valve (top) and
(without Vortab® flow conditioning)    Medium Pressure Packing
Frequent cleaning required
Gland with Ball Valve (right)
  
With VeriCal option and STP1XX models 
 = Excellent performance, preferred solution
 = Good performance, acceptable solution
 = Not recommended

6
ST100 Series Insertion Flow Meters
Model ST100 ST102 ST110 ST112 STP100 STP102 STP110 STP112
Flow Element
­FPC Sensor Style a a
­FP Sensor Style a a a a a a a a
S Sensor Style a a
316L SS Sensor a a a a a a a a
Hastelloy-C276 Sensor a a
All Welded Element a a a a a a a a
Temperature Service ≤ 350 °F [177 °C] ≤ 350 °F [177 °C] ≤ 350 °F [177 °C] ≤ 350 °F [177 °C] Standard: 32 °F to 176 °F [0 °C to 80 °C]
and Measuring Range ≤ 500 °F [260 °C] ≤ 500 °F [260 °C] ≤ 500 °F [260 °C] ≤ 500 °F [260 °C] Ex Rated: -22 °F to 212 °F [-30 °C to 100 °C]
All ST models -40 °F [-40 °C] to: ≤ 850 °F [454 °C] ≤ 850 °F [454 °C] Limited by pressure transducer; contact FCI for special order, higher temp. service

Pressure Measurement  0-50 psig [3,4 bar (g)]  0-500 psig [34 bar (g)]
Ranges  0-160 psig [11 bar (g)]  0-1000 psig [70 bar (g)]
Process Connections
Compression Fittings with variable “U” length up to 6″, 12″, 21″, 36″ or 60″ [152 mm, 305 mm, 533 mm, 914 mm or 1524 mm]
3/4″ MNPT a a
1″ MNPT a a
Flange, Tapped and Threaded for
3/4″ NPT Compression Fitting a a
Retractable Packing Glands * with variable “U” length up to 6″, 12″, 21″, 36″ or 60″ [152 mm, 305 mm, 533 mm, 914 mm or 1524 mm] * Contact FCI for pressures > 500 psi [34 bar (g)]
Low Pressure
≤ 50 psi [3,4 bar (g)] a a a a a a a a
Medium Pressure
≤ 500 psi [34 bar (g)] a a a a a a a a
Fixed specify “U” length up to 60″ [1524 mm] maximum
1″ MNPT a a a a a a a a
Flanged a a a a a a a a

ST100L In-Line For Line Sizes 1″, 1 1/2″, 2″ [25 mm, 40 mm, 51 mm]
Single-point insertion flow meter for pipe
sizes 2 1/2″ [64 mm] and larger. Calibration
ranges from 0.25 SFPS to 1000 SFPS [0,07
Model ST100L
NMPS to 305 NMPS].
Flow Element
– Models ST100, STP100
­FP Sensor Style a
S Sensor Style a
316L SS Sensor a In-line spool-piece flow meter for use in line sizes
1″, 1 1/2″, and 2″ [DIN25, DIN40, and DIN50].
Hastelloy-C276 Sensor a Typical calibration range is from 0.006 SCFM to
All Welded Element a 838 SCFM [0,01 NCMH to 1425 NCMH].

Temperature Service ≤ 257 °F [125 °C] – Model ST100L

Process Connections
1″ Tubing a Dual-element insertion flow meters with
1″ Pipe a averaging mode for improved accuracy and
repeatability in line sizes ≥ 12″ [305 mm].
1 1/2 ″ Pipe a
– Models ST102 A, ST112 A, STP102 A, STP112 A
2 ″ Pipe a
NPT (Male) a
NPT (Female) a
Flanged a VeriCal™ In-Situ Calibration models are
insertion flow meters with the FCI exclusive
Butt Weld a VeriCal in-situ calibration verification system.

– Models ST110, STP110, ST112, STP112

7
Two Sensor Averaging System
Applications involving line sizes 12 inches [305 mm] or greater With any two-element system, one element is always remote
can realize improved installation accuracy and repeatability by mounted while the other may be integrally mounted or remote
averaging the flow rates of two elements. Distorted, swirling mounted from the transmitter.
and non-repeatable flow profiles can result in decreased
accuracy of single point meters. In many cases, it is impractical
or impossible to provide the required straight-run for a fully
With Integral Configuration
developed flow profile in large lines.
Models ST102  A, ST112  A, STP102  A and STP112  A
overcome these flow profile concerns with a simple, economical
“dual-element averaging system.” The transmitter electronics
will average the input from two independent flow elements
into a single output. Each flow element can be independently
configured for insertion length and process connection to
allow installation flexibility. One flow element can be integral
with the flow transmitter, or both can be configured as remote
to accommodate easy access to terminal blocks and the
optional digital display/optical four button interface. The flow
transmitter also provides independent information for each
flow element, saving time when performing service checks.
To help determine if your application will benefit
from using a dual-element averaging system, contact your With Remote Transmitter
local representative or visit FCI online to utilize AVAL, FCI’s
proprietary evaluation software.

8
Add Pressure Measurement with “STP” Models On-Board Data Logger
The “STP” models are the only thermal dispersion gas mass flow ST100 Series electronics include a user selectable and programmable
meters available featuring pressure measurement to create a triple data logging feature. Readings are stored to a removable, internal,
function solution (flow + temperature + pressure) in a single tap point board-mounted, micro-SD card (micro secure digital memory card).
insertion instrument – an industry first and FCI exclusive feature. The supplied micro-SD card has a 2 GB capacity capable of storing
A precision strain gauge pressure transducer is built into the approximately 21 million readings. Recording time base is user
STP probe assembly and integrated into the transmitter inputs selectable with maximum rate of 1 reading per second. The logging
and outputs. The transducer, connections and wiring assembly feature is selectable via the front panel menu or via the serial port
are enclosed within a compact, protective aluminum housing and and configuration software tools. Readings are stored in ASCII, comma
incorporated into the flow element assembly. Four pressure ranges separated format for easy uploading to spreadsheets such as MS-Excel.
are available: 50, 160, 500 and 1000 psig [3,4; 11; 34 and 70 bar (g)]
to maximize resolution and accuracy for your application. Pressure
transducers are available in both Ex and non-Ex rated types. Full Instrument Agency Approvals
In addition to the labor savings over installing and running power
ST100 Series meters are agency approved for hazardous environment
to multiple sensors, bus communication installation reduces bus
installations. FCI products undergo rigorous agency testing and obtain
wiring, nodes and interfacing because all three measurements are
their approvals on the entire instrument, not just the enclosure. This
provided over a single bus connection. Analog outputs and alarm values
highest level of agency certification is your evidence of FCI’s integrity.
are also user assignable to pressure measurement as required by your
This ensures that every FCI instrument meets the intended safety
application. In STP models with the optional graphical/digital display,
requirements of your installation.
the pressure measurement value is added and continuously displayed.
FM, FMc: Class I, Division 1, Explosion proof, Groups B, C, D
Model STP100: Single-point insertion flow element with flow, Class II/III, Division 1, Dust ignition proof, Groups E, F, G
temperature and pressure Class I, Division 2, Nonincendive, Groups A, B, C, D
Model STP102: Dual-element insertion flow element; both Class II, Division 2, Nonincendive, Groups E, F, G
elements with flow, temperature and pressure Class III, Division 1, 2, Dust ignition proof, indoor/outdoor
Hazardous (classified) locations
Model STP110: Single-point insertion flow element with flow,
Type 4X/ IP67; NEC 500
temperature and pressure and VeriCal
ATEX: EX d IIC Gb T6/T1 Ta = - 40°C to 65°C;
Model STP112: Dual-element insertion flow element; both Ex tb IIIC T85°C/T450°C Db Ta = - 40°C to 65°C; IP67
elements with flow, temperature and pressure, IECEx : II 2 G Ex d IIC Gb T6/T1 Ta = - 40°C to 65°C; IP67
both elements with VeriCal CPA, NEPSI, Inmetro, EAC (TRCU) Russia
CE approved, CRN
Probe complies with Canadian Electrical code requirements of
ANSI/ISA 12.27.01-2011 as a single seal device
SIL 1 compliant
FCI continually obtains additional agency approvals; contact FCI for
availability of other agency approvals

Models STP100, Shown with optional Models STP110,


STP102 low pressure packing gland STP112

9
VeriCal ™ In-Situ Calibration Verification g VeriCal kit with fixtures and fittings to meter and control a
precise flow of nitrogen 4 across the ST flow element. The VeriCal
Models ST110, ST112, STP110 and STP112
kit provides a specially designed 100 psig [6,9 bar (g)] pressure
regulator and LCD digital pressure gauge packaged in a NEMA 4
g Validate performance on-site in minutes (IP66) rated enclosure, and a 25 foot [7,6 m] air hose with quick
g Save costs, no need to remove flow meter from process disconnect fittings to connect the kit to the VeriCal ready flow
element. The kit is fully portable or can be permanently mounted.
g Comply with ISO and local regulations for periodic The VeriCal kit, part number 020849-03, is ordered separately as
calibration verification a second line item. A single VeriCal kit can be used with and
g Provides in-situ flow element cleaning support any number of VeriCal-ready flow elements.
g Nitrogen source (user supplied) with a hose, regulated 125 psig
Models ST110, ST112, STP110 and STP112 include FCI’s exclusive to 150 psig [8,6 bar (g) to 10,3 bar (g)], either from an installed plant
and patented VeriCal system. VeriCal provides you with the ability to line or a portable supply tank. Typical test requires approximately
perform periodic field validation and verification of the flow meter’s 40 ft3 [1,1 m3] of nitrogen.
measuring performance and calibration without extracting the flow
meter from the pipe or process.
In gas flow processes with procedures or regulations requiring How It Works
periodic calibration verification, an FCI flow meter configured with
VeriCal provides the most convenient and lowest cost solution. With the VeriCal kit attached to the ST flow meter and the nitrogen
Performing a verification with VeriCal also provides a secondary source, a positive-pressure nitrogen injection is introduced 4. With
benefit of cleaning the flow element which helps ensure performance the ST flow element fully retracted from the actual flow stream and
and reduce the need for routine maintenance. the ST set in the VeriCal Calibration group mode, the user adjusts the
pressure at the VeriCal kit to inject small, controlled doses of nitrogen
The VeriCal option is comprised of three components: at specific pressures (prescribed on the FCI VeriCal calibration
g VeriCal ready flow element – This specially-constructed flow document). At each pressure point, flow readings are taken from the
element includes a flow port at the sensors, internal tubing, sonic ST flow meter and compared to the respective baseline readings in
nozzle and precision orifice, a ball valve to select “normal” or the documentation provided by FCI. This procedure is performed at
“VeriCal” operation and a welded and sealed inlet check valve to initial installation, then as often as desired.
prevent hazardous process gases from escaping. After the flow Model ST110: Single-point insertion flow meter configured
meter system has been precision calibrated in FCI’s NIST traceable with VeriCal
flow laboratory, the lab also performs an additional nitrogen
Model ST112: Dual-element flow meter; both elements
calibration and obtains five (5) baseline measurement points across configured with VeriCal
the flow range to which all field checks using the VeriCal system are
Model STP110: Single-point flow meter with pressure
compared. With each ST110, ST112, STP110 or STP112, FCI provides measurement configured with VeriCal
a second calibration curve and printed document showing the five
Model STP112: Dual-element flow meter with pressure measurement
baseline VeriCal flow readings for use by your field technicians. This option; both elements configured with VeriCal
VeriCal calibration curve is stored within one of the ST100 Series’
P/N 020849-03: VeriCal kit
five calibration groups. The VeriCal calibration is quickly and easily
4 Other gases may be used; contact FCI for specific information
accessed and recalled from the first menu screen on the display.

10
FCI’s World Class Calibration
Ensures Installed Accuracy
ST100 Series models are tested and calibrated to rigorous
standards so that you get the instrument that does the job
specified. To design and produce the highest quality flow
instrumentation, FCI operates a world-class flow calibration
laboratory with calibrations performed on more than 19 different
flow stands, using equipment traceable to NIST (US National
Institute of Standards and Technology), and ISO/IEC 17025
(International Standards for test lab quality systems).
To achieve the highest possible accuracy in ST100 Series,
Up to Five Unique Calibrations FCI utilizes these precision flow stands to flow actual gases
and reference fluids matched to the temperature and process
While a single calibration is sufficient for many operations, the conditions of your application. Further utilizing the extensive
ST100 Series can provide up to five (5) unique calibration groups. flow laboratory capabilities, FCI has developed an all new, patent
Depending on your application need, this feature can provide pending and scientifically engineered SpectraCal gas equivalency
significant cost and time savings. Some examples include: algorithm as an economical alternative to actual gas calibration.
Broaden measuring range – Extend the turndown ratio up to Other suppliers are often limited only to air calibrations
1000:1 or have different ranges for the same gas to maximize and rely on non-field tested or un-validated theoretical
accuracy and resolution under changing conditions. This is equivalencies for other gases and gas mixtures. This procedure
particularly beneficial with flares, which can require both a low can be inadequate and create measurement and output
flow range for normal operation and leak detection, and also a errors well outside published specifications. FCI calibration
very high flow range for upset conditions. Multiple calibration capabilities are un-matched in the industry, providing you with
group settings saves on the cost of using multiple flow meters. total confidence that your installation meets its published
specifications and your application needs.
Different mixtures of same gases – Embed calibrations to
optimize flow measurement accuracy in dynamic or seasonally
affected processes (e.g. Digester Gas Group 1 is 65% CH4, 35% More than 19 precision flow
stands to match NIST traceable
CO2; Group 2 is 62% CH4, 38% CO2). fluids, process conditions, flow
rates and line sizes specified in
Different gases – Portable or temporary installations for
your application.
multiple applications, or to reduce spare parts inventory in
plants with multiple installations and applications. Multiple
gases, such as dual fuel sources (e.g. natural gas and propane).

11
ST100 Series Features
Four conduit ports provide greatest signal integrity and separation for power input, analog Extensive analog and digital communications
output lines, digital I/O, relays and/or auxiliary input signals; choice of NPT or M20 threads output choices
— Triple 4-20 mA with HART
SIL 1 compliant — Foundation Fieldbus H1
AC or DC power supply — PROFIBUS PA
— Modbus RS-485
Weather-proof, ruggedized, — 0-1 kHz or 0-10 kHz frequency or pulse
Ex rated enclosures — USB port
— Choices for local — Ethernet
or remote mounting
— NEMA 4X, IP67 On-board data logger
— Aluminum standard, stainless
steel optional Four (4) optical touch buttons
— Proximity activation, no need to
Global agency approvals of entire open enclosure
instrument system for hazardous — Full instrument programmability
location installations: — Protected against unwanted activation
FM, FMc, ATEX, IECEx, NEPSI, CPA,
Inmetro, EAC (TRCU) Russia Comprehensive informational display
— Digital readout of all measured
Multiple calibrations parameters; flow rate, total flow,
— Up to five independent, separate calibrations temperature and pressure with
— Multiple gases or mixed gas compositions engineering units
— Same gas, different flow range to optimize — Analog flow rate bar graph
accuracy and extend turndown up to 1000:1 — Alarm status indication
Precision calibration and calibration choices — Instrument fault indication
— Specific gas and application matched — User programmable 17 character field
calibration using equipment traceable (example: display gas type, tag number or
to NIST and ISO/IEC 17025 application/location)
— Display orientation rotates in 90°
Extensive selection of process connections increments electronically
— Simple, adjustable installation with threaded
— Backlighted: auto-on activation via
NPT connector
proximity sensor or set for always on
— Teflon or metal ferrule seals
— Fixed connections Multi-function: measures mass flow rate and
— ANSI or DIN flanges temperature; STP Series adds
— Retractable assemblies pressure measurement

Stainless steel or Hastelloy-C276 wetted parts Permanent laser-etched depth gauge


markings; ensures accurate centering
of adjustable-length elements
All welded sensor elements for maximum
service life and leak-proofing
Precision, wide-ranging platinum RTD sensors
Exclusive equal mass sensors provide
Choice of three flow element styles to optimize optimum performance in processes with wide
application performance temperature swings
— FPC style
— FP style (shown)
— S style

Optional stainless steel Optional


enclosure sun shield

Remote up to 1000 ″ [300 m] ST100L in-line style ST100 with sun shield

12
ST100 Series Mass Flow Meter Specifications
Instrument Other:

NEPSI, Inmetro, EAC (TRCU) Russia, CRN, CPA, CE marking
Probe complies with Canadian Electrical code requirements
g Measuring Capability
of ANSI/ISA 12.27.01-2011 as a single seal device
ST1XX Models: Flow rate, total flow and temperature
STP1XX Models: Flow rate, total flow, temperature and pressure g SIL: SIL 1 compliant
g Basic Style SFF % AC Powered DC Powered
ST100: Insertion, single-point Single element types 79.8 % 79.9 %
Dual element types 88.9 % 88.9 %
ST100L: In-line (spool piece), single-point
ST102: Insertion, dual-element system g Calibration: Performed on flow stands with equipment traceable
ST110: Insertion, single-point with VeriCal™ capability to NIST and ISO/IEC 17025
ST112: Insertion, dual-element system with VeriCal capability
STP100: Insertion, single-point with pressure measurement
STP102: Insertion, dual-element system with pressure measurement Flow Element
STP110: Insertion, single-point with pressure measurement g Material of Construction
and VeriCal capability All-welded 316L stainless steel; Hastelloy-C optional
STP112: Insertion, dual-element system with pressure measurement g Operating Pressure
and VeriCal capability ST100, ST102, ST110, ST112 Insertion Style
g Flow Measurement Range Metal ferrule: 1000 psig [70 bar (g)]
Insertion Style: 0.25 SFPS to 1000 SFPS [0,07 NMPS to 305 NMPS] Teflon ferrule: 150 psig [10 bar (g)] (200 °F [93 °C] maximum)
ST100L In-line: 0.0062 SCFM to 838 SCFM Fixed connection NPT: 1000 psig [70 bar (g)]
[0,01 NCMH to 1425 NCMH] Fixed connection flanged: per flange rating
– Air at standard conditions; 70 °F and 14.7 psia [0 °C and 1,01325 bar (a)] STP Models See pressure sensor specifications
g Temperature Measurement Range Same as ST type above or per pressure sensor specification, whichever is lower
Up to 850 °F [454 °C] commensurate with element; see Operating Model ST100L In-line style
Temperature in Flow Element specification Tubing Sch 40 Pipe Sch 80 Pipe
g Pressure Measurement Range (STP Models) Psig Bar(g) Psig Bar(g) Psig Bar(g)
Available Ranges: F Style Sensor
0 psig to 50 psig [0 bar (g) to 3,4 bar (g)] 1 ″ 2400 165 2500 172 3000 207
1 1/2 ″ 1750 121 2500 * 172 *
0 psig to 160 psig [0 bar (g) to 11 bar (g)]
2 ″ 1500 103 2250 * 155 *
0 psig to 500 psig [0 bar (g) to 34 bar (g) S Style Sensor
0 psig to 1000 psig [0 bar (g) to 70 bar (g)] 1 ″ 2400 165 2500 172 2500 172
g Media: All gases that are compatible with the flow element material 1 1/2 ″ 1750 121 2500 * 172 *
Accuracy 2 ″ 1500 103 2250 * 155 *
g

Flow: Gas specific calibration: ± 0.75% reading, ± 0.5% full scale * 1 1/2 ″ and 2 ″ Sch 80 available by special order only; contact FCI
3/4 ″ pipe also available by special order
Temperature: ± 2 °F [± 1,1 °C] (display only, flow rate must be greater
g Operating Temperature (Process)
than 5 AFPS [1,5 m/sec])
Pressure (STP Models): ± 1.0% of span ST100, ST102 Insertion Style
g Temperature Coefficient All Flow Elements (FPC, FP and S):
With optional temperature compensation; valid from 10% to 100% of -40 °F to 350 °F [-40 °C to 177 °C]
full scale calibration -40 °F to 500 °F [-40 °C to 260 °C]
-40 °F to 850 °F [-40 °C to 454 °C]
Flow: Maximum ± 0.015% of reading / °F up to 850 °F
[± 0.03% of reading / °C up to 454 °C] ST110, ST112 Insertion Style
g Repeatability FP Style Flow Element:
-40 °F to 350 °F [-40 °C to 177 °C]
Flow: ± 0.5% reading -40 °F to 500 °F [-40 °C to 260 °C]
Temperature: ± 1 °F [± 1 °C] (flow rate must be greater than 5 AFPS
[1,5 NMPS]) STP Series Insertion Style
FP Style Flow Element:
Pressure (STP Models): ≤ 0.1% of span
Standard: 32 °F to 176 °F [0 °C to 80 °C]
g Turndown Ratio
Ex Rated: -22 °F to 212 °F [-30 °C to 100 °C]
Normally factory set and field adjustable from 2:1 to 100:1 within Limited by pressure transducer; contact FCI for special order, higher
calibrated flow range; up to 1000:1 possible with factory evaluation of temperature service
application
g Temperature Compensation ST100L In-line Style
FP and S Style Flow Element:
Standard: ± 30 °F [± 16 °C] -40 °F to 257 °F [-40 °C to 125 °C]
Optional: ± 100 °F [± 55 °C] g Process Connection
g Agency Approvals
FM, FMc: Class I, Division 1, Explosion proof, Groups B, C, D Compression Fittings Models ST100 and ST102 only
Class II/III, Division 1, Dust ignition proof, Groups E, F, G 3/4″ or 1″ male NPT, stainless steel with adjustable Teflon ferrule or metal
Class I, Division 2, Nonincendive, Groups A, B, C, D ferrule; or flanged tapped and threaded for 3/4″ fitting, ANSI or DIN flanges 
Class II, Division 2, Nonincendive, Groups E, F, G Compression fittings not available with 850 °F [454 °C] temperature
Class III, Division 1, 2, Dust ignition proof, indoor/outdoor versions of ST100 or ST102
Hazardous (Classified) locations Retractable Packing Glands
Type 4X/ IP67 (per NEC 500) Low pressure 50 psig [3,5 bar (g)] or medium pressure 500 psig [34 bar (g)]
ATEX: EX d IIC Gb T6/T1 Ta= - 40°C to 65°C; with graphite or Teflon packing material; 1 1/4″ male NPT or ANSI or DIN
Ex tb IIIC T85°C/T450°C Db Ta = - 40°C to 65°C; IP67 flange 
IECEx : II 2 G Ex d IIC Gb T6/T1 Ta = - 40°C to 65°C; IP67 Teflon packing required when process media is ozone, chlorine or bromine; remote
mount required when medium pressure packing gland is required

ST100 Series specifications continued on next page 13


ST100 Specifications Cont’d
Fixed Fittings / All Welded g Enclosures
1″ male NPT or ANSI or DIN flange  Main Transmitter / Electronics
Insertion Length NEMA 4X, IP67; polyester powder coated aluminum or optional in
Field adjustable lengths: stainless steel; 4 conduit ports threaded as 1/2 ″ NPT or M20x1.5; 7.74 ″ x
1″ to 6″ [25 mm to 152 mm] 5.40 ″ x 5.00 ″ [196,6 mm x 137,2 mm x 127 mm] 
1″ to 12″ [25 mm to 305 mm] Local Enclosure (Remote Configuration):
1″ to 21″ [25 mm to 533 mm] Model ST100L, Models ST100 and ST102 without packing gland option:
1″ to 36″ [25 mm to 914 mm]
NEMA 4X, IP67; polyester powder coated aluminum; 2 conduit ports threaded
1″ to 60″ [25 mm to 1524 mm]
as 1/2 ″ NPT or M20x1.5; 3.75 ″ x 4.00 ″ x 3.24 ″ [95 mm x 102 mm x 82 mm]
Fixed lengths from 2.6″ to 60″ [66 mm to 1524 mm]
Models ST100 and ST102 with packing gland option; ST110, ST112
g ST100L In-line Flow Tube and all STP Models:
Flow element is welded to an in-line flow tube, calibrated and supplied as a spool- NEMA 4X, IP67; polyester powder coated aluminum; 1 conduit port threaded
piece; options include low flow injection tubes and built-in Vortab flow conditioners as 1 ″ NPT or M20x1.5; 5.40 ″ x 4.82 ″ [137,2 mm x 122 mm]
for optimum low flow rangeability and performance
All models with stainless steel enclosure option:
Size: 1″ diameter tubing; 1″, 1 1/2″ or 2″ schedule 40 pipe
Length: 9 nominal diameters NEMA 4X, IP67; stainless steel; 1 conduit port threaded as 1 ″ NPT or
Process Connections: Female NPT, male NPT, ANSI or DIN flanges, or M20x1.5; 5.40 ″ x 4.82 ″ [137,2 mm x 122 mm]
g Data Logger
butt weld prepared
Option: Flanges sized for flow tube User programmable for readings per time increment to a maximum of
g Remote Transmitter Configurations: Transmitter may be mounted remotely 1 reading/second; removable, circuit board-mountable 2 GB micro-SD
from flow element using interconnecting cable (up to 1000′ [300 m]); remote (secure digital) memory card supplied; stores approximately 21 M
configuration required with medium pressure packing gland process connection readings in ASCII comma-separated format
g STP Models: Specifications on Pressure Sensor g Readout/Display and Optical Touch Buttons Optional

Standard Sensor PSI Bar PSI Bar PSI Bar PSI Bar
 Large 2″ x 2″ [50 mm x 50 mm] LCD; digital plus bar graph and
Pressure range 50 3,4 160 11 500 34 1000 70
engineering units
Over pressure safety 100 7 290 20 1000 70 1740 120  Digital displays of flow rate, total flow, temperature and pressure
Burst pressure 250 17,24 500 34 2500 172 7975 550 (with STP models); user selectable for engineering units
Wetted Materials:  Analog bar graph of flow rate
Connection 316L 316L 316L 316L
Pressure Sensor PH 13-8 SS 316L 316L 316L  Alarm status indication
 User programmable 17 alphanumeric character field associated with
Ex Rated Sensor PSI Bar PSI Bar PSI Bar PSI Bar
Pressure range 50 3,4 160 11 500 34 1000 70
each calibration group
Over pressure safety 240 16,55 1160 80 1160 80 1740 120  Set-Up & Service mode displays text and service codes
Burst pressure 290 20 1390 95,84 5800 400 7970 549,5  Backlighted – backlight activated by proximity motion detection, or
Wetted Materials: user may set for always on
Connection Stainless steel Stainless steel Stainless steel Stainless steel
Pressure Sensor Stainless steel Stainless steel SS and Elgiloy SS and Elgiloy  Four (4) optical touch buttons for user programming of instrument set-
up and service interrogation
 Optical touch button activation through front window – no need to
Flow Transmitter/Electronics open enclosure to access or activate
g Operating Temperature:  Display is electronically rotatable in 90° increments to optimize
Single element models: -40 °F to 150 °F [-40 °C to 65 °C] viewing angle
Dual element models: ST102, ST112, STP102, STP112 Note: If readout/display not ordered, all user set-up and service inter-
-40 °F to 120 °F [-40 °C to 49 °C] rogation must be done via computer link to bus comm and/or USB port.
g Input Power

AC: 85 Vac to 265 Vac, 50 Hz to 60 Hz;


Single element: 10 watt; dual element: 13.1 watt maximum
Other Options
DC: 24 Vdc ± 20%;
g Vortab Flow Conditioners
Single element: 9.6 watt; dual element: 13.2 watt maximum Available for all line size applications; standard choice with
g Outputs
Model ST100L (in-line)
g Sun Shield
Analog Shades main transmitter, electronics and display from direct sunlight;
Standard: Three (3) 4-20 mA*, 0-1kHz, or 0-10 kHz pulse/frequency 316L stainless steel
4-20 mA outputs are user assignable to flow rate, temperature and/or if FCI PN 023241-01 Integral transmitter
so equipped, pressure; outputs are user programmable to full flow range FCI PN 023237-01 Remote transmitter
or subsets of full flow range; pulse/frequency output is user selectable g
as pulse for external counter/flow totalizer, or as 0-1 kHz or 0-10 kHz Ball Valves / Cable Glands
g Element Coatings and Materials
frequency representing flow rate
* Outputs are isolated and have fault indication per NAMUR NE43 For service in highly corrosive gases or with erosive particulates, FCI can
guidelines, user selectable for high (> 21.0 mA) or low (< 3.6 mA) provide special coatings and wetted materials to protect the element
and provide longer service life; examples include Kynar, Tantalum, and
Digital
chromium carbide
Standard: USB, Ethernet (for use in configuration and as service port)
g Certification and Testing Documentation
Optional: HART (comes standard with analog outputs, V7 compliant)
Foundation Fieldbus H1, PROFIBUS PA or Modbus RS-485
g Auxiliary Inputs Specifications at reference operating conditions of 70 °F, 14.7 psia [21,1 °C, 1013
4-20 mA input channel used for FCI administered special configurations bar (a)] and straight pipe run 20 d upstream, 10 d downstream
to allow ST100 Series to accept inputs from external devices such as FCI is a continuous improvement company; specifications subject to change without
gas analyzers, gas composition or pressure sensors notice

14
ModelST100L
Model ST100L In-Line
In-Line Flow
FlowMeterMeter
Meter
Model ST100L In-Line Flow
In-LineConfiguration
In-Line Configuration
Integral Configuration

Enclosure meets explosion


Enclosure meets
proof, water explosion
and dust tight
5.00 proof, water and
approvals (seedust tightfor
manual
5.00127 approvals
127 specific (see manual
approval for
types)
specific approval types)
Optional display 7.75
Optional display 7.75197
1/2 inch NPT
5.40 1/2 inch NPT 197
5.40137 or M20X1.5
Mounting orientation or M20X1.5
Mounting orientation 137 9.98
of flats to be parallel to flow 9.98253
of flats to be parallel to flow
253
Flow direction External ground screw
Flow direction External ground screw

Sensor element types:


Sensor element
"S" style types:
(shown) 1/2 X˚
Optional Vortab "S""F"
style (shown) 1/2 X˚
Optional Vortab style (optional)
"F" style (optional) X˚

7 Dia 2 Dia 2x flanged
7 Dia 2 Dia 2x process
flanged connections
"A" Length process connections
"A" Length

3.63 4.13
3.6392 4.13105
92 105

RemoteTransmitter
Remote Transmitter
Remote Transmitter
Enclosure meets explosion
Enclosure meets
proof, water explosion
and dust tight
proof, water and 3.27 3.75
approvals (seedust tightfor
manual 3.2783 2X 1/2 NPT 3.7595
approvals (see manual for
specific approval types) 83 2Xor1/2M20X1.5
NPT 95
specific approval types) or M20X1.5
3.56
Mounting orientation 3.5691 External
Mounting orientation
of flats to be parallel to flow External
ground screw
91 7.34
of flats to be parallel to flow 7.34186 ground screw
Flow direction 186
Flow direction
Sensor element types: 1/2 X°
Sensor element
"S" style types:
(shown) 1/2 X°
"S""F"
style (shown)
style (optional) X°
"F" style (optional) X°
Optional Vortab 2x flanged
Optional Vortab 2x process
flanged connections
7 Dia 2 Dia process connections
7 Dia 2 Dia
"A" length
"A" length

7.75 5.00
7.75197 5.00127
197 1/2 inch NPT Optional display
1/2orinch NPT Optional display 127
M20X1.5
or M20X1.5
External
External
ground screw
ground screw
10.00
10.00
254
ø.27 TYP 254
3.50
3.5089
2.75
2.7570
ø.27 TYP
7
7
89 70 Optional,
Optional,
aluminum "L" bracket
aluminum "L" bracket
1.75 4.50
1.7544 4.50114
2.25 44 114
2.2557
57 2.80 7.00
2.8071 7.00178
71 178

15
Model ST100 Single
ModelPoint
ST100&Single
ST102 Point
Dual Point
& ST102 Dual Point
Insertion FlowInsertion
Meter Flow Meter
Model ST100 Single-Point,
IntegralST102 Dual-Element
Configuration Insertion Flow Meter
Integral Configuration

Integral Configuration
Compression fitting, Compression fitting,
optional teflon or metal ferrule
optional teflon or metal ferrule
Process connection: Process connection:
Display, Optional Display, Optional Compression fitting, optional.
Compression fitting, optional.
Teflon or Metal ferrule; Teflon or Metal ferrule;
3/4 inch or 1 inch NPT 3/4 inch or 1 inch NPT
.80 .80
5.00 5.00 20 20
127 127

Horizontal flow orientation


Horizontal flow orientation
10.00 Max 10.00 Max shown shown
"U" length "U" length
254 254

Enclosure meets explosionEnclosure meets explosion


proof, water and dust tight
proof, water and dust tight
approvals (see manual forapprovals (see manual for
specific approval types)specific approval types)
5.40 5.40
External grounding External
137grounding Mounting
137 orientation of Mounting
flats Vertical flow orientationVertical flow orientation
orientation of flats
screw screw to be parallel to flow to be parallel to flow shown shown
"S" Style head "S" Style head

7.75 7.75 .88 .88


197 197 22 22

"FP" or "FPC" Style head "FP" or "FPC" Style head


4X 1/2 NPT or M20X1.5 4X 1/2 NPT or M20X1.5

Remote Transmitters
Remote Transmitters

With Ferrule typeRemote Transmitter


With Ferrule type With Packing gland
With Packing gland
Compression fitting
Compression fitting
With Ferrule Type Compression Fitting With Packing Gland
Enclosure meets explosionEnclosure meets explosion
proof, water and dust tight
proof, water and dust tight 4.68 5.03 4.68 5.03
approvals (see manual forapprovals (see manual for 119 128119 128
specific approval types)specific approval types) Enclosure meets
2X 1/2 NPT or M20X1.5 2X 1/2 NPT or M20X1.5 Enclosure meet
explosion proof, water explosion proof
3.27 and dust tight approvalsand dust tight a
3.27
3.75 3.75 (see manual for specific(see manual for
83 4.69 4.69
9583 95 119 approval types) approval types)
119
1 inch NPT 1 inch NPT
3.56 3.56 14.61 3.61 3.61
14.61
90 90 371 92 92 Mounting orientation Mounting orientatio
371
8.00 Max 8.00 Max MAX MAX of flats to be parallel of flats to be paralle
203 203 to flow to flow
6.31 6.31
2.19 2.19
56
Mounting orientation of Mounting
flats orientation of flats160 160
56 to be parallel to flow to be parallel to flow
1 1/4 inch NPT packing gland.
1 1/4 inch NPT packin
3/4 inch tube Teflon or graphite packing
Teflon or graphite pa
3/4 inch tube Process connection: Process connection: material (optional flanged
"U" length "U" length material (optional flan
Compression fitting, optional.
Compression fitting, optional. type provided) type provided)
Teflon or metal ferrule; Teflon or metal ferrule; "U" length "U" length
3/4 inch or 1 inch NPT 3/4 inch or 1 inch NPT
3/4 inch pipe 3/4 inch pipe

.80 "FP" or "FPC" Style head "FP" or "FPC" Style head "S" Style head "S" Style head
.80 1.13 1.13
20 20 29 29 "FP" or "FPC" Style head "FP" or "FPC" Style head
"S" Style head "S" Style head

7.75 7.75
197 197 5.00 5.00
1/2 inch NPT 1/2 inch NPT 127 127
or or
Optional Display Optional Display
M20X1.5 M20X1.5
External ground screw External ground screw
5.40 5.40
137 137
10.00 10.00
254 254 10.00 10.00
3.50 2.75 .27 254 .27 254
3.50 2.75
89 70 7 7
89 70 TYP TYP
Aluminum "L" bracket Aluminum "L" bracket
1.75 1.75
2.25 442.25 44 4.50 4.50
57 57 114 114
2.80 2.80 7.00 7.00
71 71 178 178

16
Models ST110 Single Point With Verical
& ST112 Dual Point With Verical
Models Insertion
ST110 Single
FlowPoint
MeterWith Verical
& ST112 Dual
Integral
Model ST110 Single-Point, Point With
Configuration
ST112 Verical
Dual-Element Insertion Flow Meter
Insertion Flow Meter
4X 1/2 inch NPT Integral
IntegralConfiguration
Configuration
or M20x1.5 1 1/4 inch NPT packing gland;
VeriCal box assembly Teflon or graphite packing material
4X 1/2 inch NPT (optional flanged type provided)
or M20x1.5 1 1/4 inch NPT packing gland;
VeriCal box assembly Teflon or graphite packing material
1.13
(optional flanged type provided)
7.75 29
197
1.13
7.75 29
1/4 turn plug valve
197 "FP" Style head
External grounding
1/4 turn plug valve
"FP" Style head
External grounding
Enclosure meets explosion
proof, water and dust tight Mounting orientation
5.40 of flats to be parallel to flow
approvals (see manual for 137
specific approval
Enclosure meetstypes)
explosion 1/8 inch hose plug
proof, water and dust tight Mounting orientation
5.40 of flats to be parallel to flow 3/4 inch pipe
approvals (see manual for 137
specific approval types) N2 in 1/8 inch hose plug
5.00
127 3/4 inch pipe
N2 in "FP" Style head
5.00
127
8.00 8.75 2.00 "FP" Style head
203 222 51

8.00 8.75
25.06 2.00
"U" length
203 222
637 51

25.06
"U" length
637
Remote Transmitters
Remote Transmitters
Remote Transmitter
1 inch NPT 1 1/4 inch NPT packing gland;
Teflon or graphite packing material
(optional flanged type provided)
1 inch NPT 1/4 turn plug valve 1 1/4 inch NPT packing gland;
Teflon or graphite packing material
(optional flanged type provided)
1/4 turn plug valve

Enclosure meets explosion 1.13


proof, water and dust tight 1/8 inch hose plug 29
approvals (see manual for VeriCal box assembly
specific approval
Enclosure types)
meets explosion 1.13
proof, water and dust tight 1/8 inch hose plug 29
approvals (see manual for 24.50 "U"
VeriCal box assembly
specific approval types) 622 length
4.69 8.75 6.31
119 22224.50 160 "U"
4.68 622 length
119 4.69 8.75 6.31
119 222 N2 160
in
4.68 5.74
119 146
N2 in "FP" Style head
5.74
146
2.31
59 "FP" Style head
5.23
1332.31
59
7.75 5.00
5.23
197 127
133
4X 1/2 inch NPT External ground screw
7.75 or Optional Display 5.00
197 M20X1.5 127
4X 1/2 inch NPT External ground screw
or 5.40
Optional Display
M20X1.5 137
10.00
254
7.50 5.40
3.50 2.75 191 137
10.00 89 .27 TYP
70 7
254
7.50 Aluminum "L" bracket
3.50 2.75 191
.27 TYP
89 70 1.75 7
4.50
44 114
2.25 Aluminum "L" bracket
1.75 57 7.00
4.50
2.8044 178
114
71
2.25
57 7.00
2.80 178
71 17
Models STP100 Single Point With Pressure
& STP102 Dual Point With Pressure
Insertion
Models STP100 FlowPoint
Single Meter
With Pressure
Integral
Model STP100 Single-Point,
& STP102 DualConfiguration
STP102 Dual-Element
Point Insertion Flow Meter
With Pressure
4x 1/2 inch NPT
Insertion Flow Meter
or M20x1.5 Integral Configuration
Integral Configuration
Pressure transducer 1 1/4 inch NPT packing gland;
assembly housing Teflon or graphite packing material
4x 1/2 inch NPT (optional flanged type provided)
or M20x1.5
Pressure transducer 1 1/4 inch NPT packing gland;
1.13
assembly housing Teflon or graphite
29 packing material
7.75 (optional flanged type provided)
197
1.13
7.75 "FP"29Style head
197
External grounding

"FP" Style head

Enclosure meets explosion External grounding


proof, water and dust tight Mounting orientation
approvals (see manual for 5.40 of flats to be parallel to flow
specific approval types) 137
Enclosure meets explosion
proof, water and dust tight Mounting orientation 3/4 inch pipe
approvals (see manual for 5.40 of flats to be parallel to flow
specific approval types) 137
5.00
127 3/4 inch pipe
"FP" Style head
5.00
127 8.00 8.75 2.00
203 222 51 "FP" Style head

25.06
8.00 8.75 2.00 "U" length
637
203 222 51

25.06
"U" length
637

Remote Transmitters
Remote Transmitter
1 inch NPT Remote Transmitters 1 1/4 inch NPT packing gland;
Teflon or graphite packing material
(optional flanged type provided)
1 inch NPT 1 1/4 inch NPT packing gland;
Teflon or graphite packing material
(optional flanged type provided)

Enclosure meets explosion 1.13


proof, water and dust tight 29
approvals (see manual for
specific approval types) Pressure transducer
Enclosure meets explosion assembly housing 1.13
proof, water and dust tight 24.41 "U" 29
approvals (see manual for 620 length
specific approval types) Pressure transducer
4.69 8.75assembly housing6.31
119 222
24.41 160 "U"
4.68 620 length
119
4.69 8.75 6.31
5.74 119 222 160
4.68
146
119
"FP" Style head
5.74
2.31
146
59 "FP" Style head
5.23
1332.31
59
5.23 7.75 7.00
133 197 1/2 inch NPT 178
External ground screw
or Optional Display
M20X1.5
7.75 7.00
197 178
1/2 inch NPT 5.40 External ground screw
or Optional Display 137
10.00 M20X1.5
10.00
254 254 5.40
3.50 2.75 .27 TYP 137
10.00 89 70 7 10.00
254 254
Aluminum "L" bracket

3.50 2.75 1.75 4.50


44 .27 TYP
89 70 7 114
2.25 Aluminum "L" bracket
57 7.00
1.75
2.80 4.50
178
71 44 114
2.25
57 7.00
18 2.80
178
71
Models STP110 Single Point With Pressure and Verical
& STP112 Dual Point With Pressure and Verical
Models STP110Insertion FlowWith
Single Point Meter Pressure and Verical
Model STP110 Single-Point,
& STP112 Dual
STP112
Integral
Dual-Element
Configuration
Point
Insertion Flow Meter
With Pressure and Verical
Insertion Flow Meter
4x 1/2 inch NPT
or M20x1.5 Integral Configuration
Integral Configuration
VeriCal and pressure transducer 1 1/4 inch NPTpacking gland;
assembly housing Teflon or graphite packing material
4x 1/2 inch NPT (optional flanged type provided)
or M20x1.5
VeriCal and pressure transducer 1 1/4 inch NPTpacking gland;
assembly housing 1.13
Teflon or graphite packing material
7.75 29
(optional flanged type provided)
197
1.13
7.75 1/4 turn plug valve 29
"FP" Style head
197
External grounding
1/4 turn plug valve "FP" Style head
Enclosure meets explosion External grounding
proof, water and dust tight Mounting orientation
approvals (see manual for 5.40 of flats to be parallel to flow
specific approval types) 137
Enclosure meets explosion 1/8 inch hose plug
proof, water and dust tight Mounting orientation
5.40 3/4 inch pipe
approvals (see manual for of flats to be parallel to flow
specific approval types) 137 N2 in
1/8 inch hose plug
5.00
127 3/4 inch pipe
N2 in "FP" Style head
5.00
127 8.00 8.75 2.00
203 222 51 "FP" Style head

8.00 25.06
8.75 2.00 "U" length
203 637
222 51

25.06
"U" length
637
Remote Transmitters

RemoteTransmitters
Remote Transmitter
1 inch NPT 1 1/4 inch NPTpacking gland;
Teflon or graphite packing material
(optional flanged type provided)
1/4 turn plug valve 1 1/4 inch NPTpacking gland;
1 inch NPT
Teflon or graphite packing material
(optional flanged type provided)
1/4 turn plug valve
Enclosure meets explosion 1.13
proof, water and dust tight 29
approvals (see manual for 1/8 inch hose plug
specific approval types) VeriCal and pressure
Enclosure meets explosion transducer assembly housing 1.13
proof, water and dust tight 29
approvals (see manual for 1/8 inch hose plug
specific approval types) 24.50 VeriCal and pressure "U"
622 transducer assembly housing length
4.69 8.75 6.31
119 22224.50 160 "U"
4.68 622 length
119
4.69 8.75 6.31
119 222 N2 in160
4.68 5.74
119 146
"FP" Style head
N2 in
5.74
2.31
146
59 "FP" Style head
5.23
133 2.31
59
5.23 7.75
133 197 5.00
1/2 inch NPT 127 External ground screw
or Optional Display
7.75 M20X1.5
197 5.00
1/2 inch NPT 127 External ground screw
Optional Display 5.40
or 137
10.00 M20X1.5 10.00
254 254
3.50 2.75 .27 TYP 5.40
70 7 137
10.00 89 10.00
254 254 Aluminum "L" bracket
3.50 2.75 .27 TYP
70 1.75 7 4.50
89 114
44 Aluminum "L" bracket
2.25 7.00
57 4.50
178
2.80 1.75 114
71 44
2.25 7.00
57
2.80 178 19
71
Locally Represented By: Visit FCI online at www.FluidComponents.com | FCI is ISO 9001 and AS 9100 Certified
FCI World Headquarters
1755 La Costa Meadows Drive | San Marcos, California 92078 USA
Phone: 760-744-6950 Toll Free (US): 800-854-1993 Fax: 760-736-6250

FCI Europe
Persephonestraat 3-01 | 5047 TT Tilburg, The Netherlands | Phone: 31-13-5159989 Fax: 31-13-5799036

FCI Measurement and Control Technology (Beijing) Co., LTD | www.fluidcomponents.cn


Room 107, Xianfeng Building II, No.7 Kaituo Road, Shangdi IT Industry Base, Haidian District | Beijing 100085, P. R. China
Phone: 86-10-82782381 Fax: 86-10-58851152

0816 500 Doc. No. 02MK011474 F 


© Copyright 2010 by Fluid Components International LLC. All rights reserved. Manufactured in accordance with one or more of the following patents: US Patent Numbers: 4,981,368, 5,111,692, 5,117,216, 5,134,772, 5,152,049, 5,167,153, 5,780,737, 5,600,528, 5,913,250,
6,208,254, 6,340,243, 6,628,202, 6,843,110, 7,191,645; China Patent Numbers: ZL00815586.0, ZL03817053.1. FCI is a registered trademark of Fluid Components International LLC. Information subject to change without notice.