Professional Documents
Culture Documents
GT-70A Series
User's Manual
Read this manual before use.
Keep this manual in a safe place for future reference.
Introduction
This manual describes the basic operations and hardware functions of the GT-70A
Series. Read the manual carefully to ensure performance and function of the GT-70A
Series for safe use.
Keep this manual in a safe place for future reference.
■ Symbols
The following symbols alert you to matters concerning the prevention of human injury
and product damage.
Note
Provides additional information on proper operations.
General Cautions
• At startup and during operation, be sure to monitor the functions and performance
of the GT-70A Series.
• Do not attempt to open or modify the GT-70A Series or use it in any way other than
as described in the specifications. If the GT-70A Series is modified or used other
than as described, the warranty will be voided.
• When the GT-70A Series is used in combination with other devices, its function and
performance may be degraded, depending on the operating conditions and
surrounding environment.
• Do not use the GT-70A Series for the purpose of protecting the human body.
• To aid in preventing malfunction, do not expose the GT-70 to extreme temperature
fluctuations.
Handling Abnormalities
GT-M-E 96029E 1
Installation environment
To use the GT-70A Series correctly and safely, avoid installing it in the following
locations; doing so may lead to breakdown of the unit.
• Location that is humid, dusty or poorly ventilated
• Location with a high temperature such as a place exposed to direct sunlight
• Location where there are flammable or corrosive gases
• Location where the unit may be directly subjected to vibration or shock
• Location where water, oil or chemicals may splash onto the unit
• Location where static electricity is present
Do not install the GT-70A Series near a power source or high-voltage cable,
otherwise noise may cause the GT-70A Series to malfunction. Take corrective action
for noise by using noise filters, laying cables separately, and/or installing insulation
on the amplifier and the sensor head.
A change in the ambient temperature may cause the sensor's readings to fluctuate.
Be sure to keep the ambient temperature constant.
When the ambient temperature changes by 10°C, it takes 40 minutes for the GT's
internal temperature to equalize.
Warming up
Before using the GT-70A Series, wait approximately 30 minutes after the power is
turned on. Otherwise, the displayed value may gradually fluctuate.
• As this product is a precision measuring instrument, do not drop the unit or expose
it to any other forms of shock.
• Be sure not to apply force that exceeds the loads shown below.
• Although the sensor head is water proof, do not use it in water or where liquids
such as oil splashes directly.
30N 100N 1N
2 GT-M-E
■ Air cylinder
• The GT-A10/A22 uses a single-acting cylinder. The built-in spring returns the
cylinder to its home position when the air is supplied. Do not supply air to the
exhaust valve.
• The air cylinder cannot be removed.
■ Air
• The pressure of the air supplied to GT-A10/A22 should be kept constant by using a
regulator.
• Use an air filter, a mist separator, or other equipment to supply dry air.
• Coupling socket/Exhaust valve. The coupling socket and exhaust valve attached
to the air cylinder are for exclusive use. They cannot be replaced.
Do not cut or extend the cable between the amplifier and the sensor head. This may
change the characteristics of the unit not to meet the product's specifications.
Other Precautions
Power supply
• Noise superimposed on the power supply may cause malfunction. Be sure to use
a DC stabilized power supply provided with an insulation transformer.
• If you use a commercially available switching regulator, be sure to ground the
frame GND terminal or the GND terminal.
Effects of vibration
When the detection target is vibrating, the displayed value may fluctuate. In this
case, increase the response time to achieve a more stable reading.
Exposing the sensor head to magnetic fields will affect its output.
Avoid installing the GT-70A Series near motors or other equipment and devices that
generate strong magnetic fields.
GT-M-E 3
Dust boot
In case cracks occur on the dust boot of the sensor head, optional dust boots (for
GT-H10/GT-A10 : OP-78041 , for GT-H22/GT-A22: OP-78042)* are available.
Please note that we cannot guarantee the IP67 protection rating when an optional
dust boot is used. To ensure an IP67 enclosure rating, replace the entire sensor head.
For GT-A10/GT-A22 to ensure the IP67 protection rating, the following must be
satisfied.
• An air tube with inner diameter of φ 2.5 mm must be connected to the exhaust
valve. The air tube must be securely inserted into the base of the connection port.
The unit must be installed in such a manner that no water will splash onto the end
of the air tube and no dust will adhere to it.
Threaded portion
Exhaust valve
Air tube
• Tighten the threaded portion of the exhaust valve by applying the specified torque
(0.5N•m). If the threaded portion is loose, IP67 cannot be guaranteed.
CE Mark
The GT-70A Series conforms to the CE Marking. The applicable standards (EMC
Directives) are explained as follows:
EMI :EN61326, class A
EMS :EN61326
4 GT-M-E
Contents
Introduction
Safety Precautions ....................................................................................1
General Cautions .................................................................................... 1
Handling Abnormalities .......................................................................... 1
Precautions for Usage ............................................................................ 1
Other Precautions ................................................................................... 3
CE Mark .................................................................................................. 4
List of Functions .......................................................................................9
GT-M-E 5
Chapter 3 Basic Usage of the GT-70A Series
6 GT-M-E
13. Displayed Number of Digits......................................................... 4-42
14. Hysteresis .................................................................................... 4-43
15. Input Line Function 1 ............................................................... 4-45
16. Input Line Function 2 ................................................................... 4-48
4-4 Additional Function Setting Mode .............................................. 4-49
Parameters in Additional Function Setting Mode .............................. 4-49
20. Store Preset Value ..................................................................... 4-51
21. Power Save Function ................................................................... 4-52
22. Core Alarm .................................................................................. 4-53
Core Alarm Teaching......................................................................... 4-58
23. Core Alarm Checkpoint Setting ................................................... 4-59
24. Display Filter Function ................................................................ 4-61
4-5 Calibration Setting Mode ........................................................... 4-62
Parameters in Calibration Setting Mode ............................................ 4-62
30. Calibration Function Setting....................................................... 4-64
1st Detection/Reading ....................................................................... 4-65
31. 2nd Target Setting ....................................................................... 4-66
2nd Detection/Reading.................................................................... 4-67
4-6 Calculation Mode ....................................................................... 4-69
Applications That Can Be Used in Calculation Mode ....................... 4-69
Precautions When Using Calculation Mode ...................................... 4-72
C0. Calculation Function OFF............................................................ 4-75
C1. Maximum .................................................................................... 4-76
C2. Minimum...................................................................................... 4-77
C3. Degree of Flatness ..................................................................... 4-78
C4. Average....................................................................................... 4-79
C5. Reference Difference .................................................................. 4-80
C6. Twist ............................................................................................ 4-82
C7. Warpage...................................................................................... 4-84
C8. Thickness .................................................................................... 4-86
Chapter 5 Specifications
GT-M-E 7
External Input Circuit Diagram ............................................................. 5-8
5-3 Timing Chart ................................................................................. 5-9
Standard Detection .............................................................................. 5-9
Peak Hold, Bottom Hold, Peak-to-Peak ............................................... 5-9
NG Hold ............................................................................................... 5-9
Preset Input ........................................................................................ 5-10
Bank Input .......................................................................................... 5-10
Reset Input ......................................................................................... 5-11
Minimum Input Time .......................................................................... 5-11
5-4 Dimensions ................................................................................. 5-12
Sensor Amplifier ................................................................................. 5-12
Sensor Head ...................................................................................... 5-14
Panel Cutout Dimensions ................................................................... 5-17
Head Mounting Bracket A (OP-76874) Mounting Hole Diagram ....... 5-17
Appendix
8 GT-M-E
List of Functions
GT-M-E 9
MEMO
10 GT-M-E
Before Use 1
This chapter describes the overview of the GT-70A Series and the
name and function of each part.
1
1-1 Overview of the GT-70A Series................................ 1-2
1-2 Checking the Package Contents.............................. 1-7
1-3 Identifying Part Names and Functions ................... 1-12
GT-M-E 1-1
1-1 Overview of the GT-70A Series
Determines whether the height of the detected workpiece is within the tolerance
range against the reference workpiece (master workpiece).
Before Use
When the workpiece cannot be directly detected, status of the workpiece can be
determined by detecting the fixture.
1-2 GT-M-E
1-1 Overview of the GT-70A Series
By detecting the assembled workpiece, you can determine whether the parts are
properly installed or whether too many parts have been inserted.
1
Example Packing insertion check
Before Use
Confirms that targets are properly seated.
"Tolerance calibration" (page 3-17)
GT-M-E 1-3
1-1 Overview of the GT-70A Series
The GT-70A Series include main units (GT-71A/71AP, GT-75A/75AP) and expansion
1 units (GT-72A/72AP, GT-76A/76AP). By connecting the expansion units to the main
unit, you will be able to detect various kinds of workpieces.
Before Use
Detecting warpage
Calculates the degree of warpage of the workpiece by using three detection points
and determines whether it is within the specified range.
Note
"C7. Warpage Display" can be set only when two expansion units are
connected to a main unit (total of 3-unit configuration).
Detecting flatness
Determines whether the degree of flatness is within the specified range by extracting
the maximum and minimum values from the amplifiers (main and expansion units)
and calculating the delta (maximum value - minimum value).
1-4 GT-M-E
1-1 Overview of the GT-70A Series
By using the main unit as a reference, the expansion unit calculates and determines
whether the delta falls within the specified range.
1
Example Determining pin height
Before Use
Checks if pin height is within the specified
range of a reference surface.
"C5. Reference Difference" (page 4-80)
Detecting twist
Calculates the degree of twist of the workpiece using four detection points and
determines whether it is within the specified range.
Note
"C6. Twist Display" can be set only when three expansion units are
connected to a main unit (total of 4-unit configuration).
GT-M-E 1-5
1-1 Overview of the GT-70A Series
DL-RS1A DL-RB1A
Programmable PC Programmable
controller controller
For details on the communication unit, refer to "DL-RS1A User's Manual" and
"DL-RB1A User's Manual".
1-6 GT-M-E
1-2 Checking the Package Contents
Before using the unit, make sure that the following equipment and accessories are
included in the package.
Sensor Amplifier 1
DIN rail mounting type
Before Use
GT-71A/GT-71AP (main unit) GT-72A/GT-72AP (expansion unit)
P-H HI P-H
HI B-H B-H
P-P GO P-P
GO CALC CALC
LO LO
SE S ET
T 1
1
2 2
3 3
TIM T IM
SET HI SET
HI PV
PV
LO PRE LO PRES
SET ET
MO MOD
E
DE SER IES
SERIE
S
Sensor amplifier x 1
Instruction Manual x 1 Sensor amplifier x 1
P-H
HI B-H
P-P
GO CALC
LO
1
2
3
TIM
HI
RV
LO PRESE
T
GT-M-E 1-7
1-2 Checking the Package Contents
1 HI
GO
LO
1
2
P-H
B-H
P-P
CALC
3
TIM
HI
RV
LO PRESET
Before Use
Sensor Head
GT-H10/GT-H22
1-8 GT-M-E
1-2 Checking the Package Contents
GT-A10/GT-A22
Before Use
Supplementary manual x 1
Sensor head x 1 Precautions for usage x 1
GT-M-E 1-9
1-2 Checking the Package Contents
1
Before Use
Standard contact
(Key wrench supplied)
Panel amplifier fixture Model: OP-77678
Model: OP-76876 Included with
(Included with panel-mountable End unit (two pieces) (mounted on) the sensor head.
amplifier units) Model: OP-26751
Ceramic contact Dust boot for GT-H10/A10 Dust boot for GT-H22/A22
(Key wrench supplied) (Key wrench supplied) (Key wrench supplied)
Model: OP-81970 Model: OP-78041 Model: OP-78042
A ceramic ball is used Included with (mounted on) Included with (mounted on)
at the tip. the sensor head. the sensor head.
1-10 GT-M-E
1-2 Checking the Package Contents
Spacer
1
(Key wrench supplied) Speed controller for
Model: OP-77684 GT-A10/GT-A22
Before Use
Used to extend the spindle Model: OP-82133
for 12.2 mm.
GT-M-E 1-11
1-3 Identifying Part Names and Functions
*1 When shipped from the factory, the protective cover for expansion slots is installed.
*2 It is not installed on the main unit (GT-71A/GT-71AP).
Amplifier control
Expansion unit
unit
connector (lower)
I/O cable connector
* The main unit (GT-75A/GT-75AP) has no expansion connector for expansion unit (upper side).
1-12 GT-M-E
1-3 Identifying Part Names and Functions
1
(2)
Before Use
(3)
Item Description
(1) Digital LED display Displays the detected values and setting.
(2) Position indicator Indicates HI/GO/LO positions.
(3) [SET] button Used to automatically adjust the set point.
(4) Bank indicator Indicates which program is currently being selected.
Indicates when the GT series starts detection; either
(5) Timing input indicator automatically when it exceeds a pre-set level or manually
when externally selected.
Normal
display Lights the LEDs according to the increase/decrease of the
measured values.
mode *1
(6) Bar LED
OK/NG • If the measured value is "GO", all the LEDs are lit in green.
display • If the measured value is "HIGH (or LOW)", all the LEDs are
mode *1 lit in red.
(7) Preset indicator Indicates that the preset value is being adjusted.
(8) PV indicator Peak Value is currently being displayed.
(9) [MODE] button Used for adjusting the settings and display modes.
Menu selection
• Left/right arrow button: Used to select the menu.
(10) Arrow buttons
• Up/down arrow button: Used to change the contents of the
menu.
Indicates peak hold, bottom hold, peak-to-peak hold, and
(11) Status indicator calculation.
"3-1 Main Screens" (page 3-2)
GT-M-E 1-13
1-3 Identifying Part Names and Functions
*1 Briefly pushing the [MODE] button on the peak value display screen will switch the bar
LED display mode.
1
Press the
Normal display mode [MODE] button OK/NG display mode
Before Use
GT-H10/GT-H22/GT-H10L/GT-H22L
(1)
(3)
(5) (3)
(6) (2)
(4)
Item Description
When the measured value is HIGH/LOW: Lights in red
During normal
When the measured value is GO: Lights in green
operation
Measurement wait state*1: Turns off
1-14 GT-M-E
1-3 Identifying Part Names and Functions
Before Use
• When the P.V. value display appears as "-------"
*2 The blinking color is the same as the lighting color of the normal operation.
*3 The light turns off only when "All (Full eco)" is selected in "21 Eco (Power save
function)".
"21. Power Save Function" (page 4-52)
*4 When ErC (overcurrent error) is displayed, the indicator is the same as in the normal
operation.
"Error Messages" (page A-5)
*5 The dust boot can be replaced, but the IP67 protection rating cannot be guaranteed.
To ensure the IP67 protection rating, replace the entire sensor head.
"Replacing the Dust Boot" (page A-9)
Note
GT-H10L/H22L has no dust boot. The protection rating is IP40.
Do not install the dust boot in GT-H10L/H22L. Doing so would cause
the spindle not to return smoothly.
GT-M-E 1-15
1-3 Identifying Part Names and Functions
GT-A10/GT-A22
(8)
1
(7)
Before Use
(2)
(1)
(3) (9)
(5)
(6)
(3)
(4)
(4)
Item Description
When the measured value is HIGH/LOW: Lights in red
During normal
When the measured value is GO: Lights in green
operation
Measurement wait state*1: Turns off
1-16 GT-M-E
1-3 Identifying Part Names and Functions
Before Use
location where there is no risk of water splashing onto the
end of the air tube or dust adhering to it.
*1 The measurement wait status indicates the following cases:
• When the internal timing is used, the period from when the value exceeds (or drops
below) the timing level until the data is determined
• When the P.V. value display appears as "-------"
*2 The blinking color is the same as the lighting color of the normal operation.
*3 The light turns off only when "All (Full eco)" is selected in "21 Eco (Power save
function)".
"21. Power Save Function" (page 4-52)
*4 When ErC (overcurrent error) is displayed, the indicator is the same as in the normal
operation.
"Error Messages" (page A-5)
*5 The dust boot can be replaced, but the IP67 protection rating cannot be guaranteed.
To ensure the IP67 protection rating, replace the entire sensor head.
"Replacing the Dust Boot" (page A-9)
*6 Discharge the air into the atmosphere.
*7 The built-in spring returns the air cylinder to its home position. Do not supply the air to
the exhaust valve.
GT-M-E 1-17
1-3 Identifying Part Names and Functions
MEMO
1
Before Use
1-18 GT-M-E
Installation and Connection 2
GT-M-E 2-1
2-1 Mounting and Wiring the Sensor Amplifier
2 1 Fit the tab of the lower part of the main unit to the DIN rail. While pushing
the main unit in the direction of Arrow (1), push the body down in the
direction of Arrow (2).
Installation and Connection
(3)
(2)
(1)
2 To detach the amplifier, while pushing the main unit in the direction of
Arrow (1), pull the body up in the direction of Arrow (3).
2-2 GT-M-E
2-1 Mounting and Wiring the Sensor Amplifier
The expansion unit can only be used in addition to the main unit.
Up to nine expansion units can be added to one main unit.
Note
• When adding an expansion unit(s), be sure to use a 24 VDC power
supply.
• An expansion unit of different output type cannot be added (for
example, an expansion unit of NPN output cannot be added to a
main unit of PNP output).
• An expansion unit of the DIN rail mounting type cannot be added to a
main unit of the panel mounting type.
Main unit
Expansion cover
3 Push and fix the expansion unit to the connector of the main unit until it
clicks.
Main unit
Expansion unit
Connector
GT-M-E 2-3
2-1 Mounting and Wiring the Sensor Amplifier
4 Mount the end units (OP-26751: a set of two pieces) on both sides of the
amplifiers (the main unit and the expansion unit), then fix the end units
with screws on the upper part of each end unit (2 points x 2 units).
The mounting method of the end unit is the same as that of the amplifier.
2
End unit
Installation and Connection
End unit
45
(Unit: mm)
Min. 63
+0.4
-0
21
45 (Unit: mm)
Note
When mounting the units horizontally, ensure the center pitch
between the sensor amplifiers is 63 mm or more.
2-4 GT-M-E
2-1 Mounting and Wiring the Sensor Amplifier
2 Insert the panel mounting bracket in the direction shown below into the main
unit from the back, then fit the front protective cover to the main unit face.
Front protection cover
Amplifier
Panel
Tab
The expansion unit can only be used in addition to the main unit. Up to nine
expansion units can be added to one main unit.
Note
• When adding an expansion unit(s), be sure to use a 24 VDC power supply.
• An expansion unit of different output type cannot be added (for
example, an expansion unit of NPN output cannot be added to a
main unit of PNP output).
• An expansion unit of the panel mounting type cannot be added to a
main unit of the DIN rail mounting type.
GT-M-E 2-5
2-1 Mounting and Wiring the Sensor Amplifier
2 X = 24 x (n) + 21
X
Installation and Connection
(Unit: mm)
Note
The expansion units cannot be mounted in the horizontal orientation
side by side.
Mount the units in such a manner that the main unit is placed on the top.
Main unit
Expansion unit 1
Expansion unit 2
.
.
.
2-6 GT-M-E
2-1 Mounting and Wiring the Sensor Amplifier
Expansion
Expansion cable 2
unit 1
Expansion unit connector
Expansion cable
CAUTION
Note
Do not connect between "expansion unit expansion connector
(upper) - expansion unit expansion connector (upper)" or
"expansion unit expansion connector (lower) - expansion unit
expansion connecter (lower)".
GT-M-E 2-7
2-1 Mounting and Wiring the Sensor Amplifier
Black
HIGH output
White
LOW output
Gray
GO output
Pink
PRESET input
Purple
Bank A or Reset input *2
Pink/Purple
Bank B or Timing input *3
*1 Supply 24 VDC when connecting expansion units.
*2 Bank A input is the factory default setting.
*3 Timing input is the factory default setting.
Black
HIGH output
White
LOW output
Gray
GO output
Pink
PRESET input
Purple
Bank A or Reset input *1
Pink/Purple
Bank B or Timing input *2
*1 Bank A input is the factory default setting.
*2 Timing input is the factory default setting.
2-8 GT-M-E
2-1 Mounting and Wiring the Sensor Amplifier
Power cable
Connecting the power cable Disconnecting the power cable
Note
• The number of power cable wires is different between the main unit
and the expansion unit (Main unit: 8 wires, Expansion unit: 6 wires).
• The power of the expansion unit is supplied through the expansion
cable being connected to the main unit. If the I/O line of the
expansion unit is not used, cut the power cable (I/O cable) at the
edge of the connector.
Brown
Blue 12-24 VDC *1
Black
HIGH output
White
LOW output
Gray
GO output
Pink PRESET input
Purple
Bank A or Reset input *2
Pink/Purple
Bank B or Timing input *3
*1 Supply 24 VDC when connecting expansion units.
*2 Bank A input is the factory default setting.
*3 Timing input is the factory default setting.
GT-M-E 2-9
2-1 Mounting and Wiring the Sensor Amplifier
2-10 GT-M-E
2-2 Connecting and Mounting the Sensor Head
Wrench
±1°
45°
A
2 Insert the collet into the hole (from the chamfered side), and loosely
tighten with the nut.
Collet
Chamfered side
Nut
GT-M-E 2-11
2-2 Connecting and Mounting the Sensor Head
3 Insert the sensor head into the collet. While securing the tightening
sleeve with the included wrench, tighten the nut with a separate wrench.
Dust boot
10 N·m max.
Note
GT-H10L/H22L has no dust boot.
2-12 GT-M-E
2-2 Connecting and Mounting the Sensor Head
To mount the sensor head on the side face of a table or something similar, use the
included Head mounting bracket B (OP-76875).
Mounting diagram Package contents of the standard
bracket B (OP-76875)
2
Collet Nut Mounting bracket
Mounting bracket
Nut
Proper collet alignment
Note
If the collet is inserted from the wrong side, the sensor head
cannot be secured.
2 Insert the sensor head into the collet, and tighten the nut with a wrench.
Nut
Wrench
(10 N·m max)
Dust boot
GT-M-E 2-13
2-2 Connecting and Mounting the Sensor Head
M4 screw
2
Installation and Connection
Note
GT-H10L/H22L has no dust boot.
2-14 GT-M-E
2-2 Connecting and Mounting the Sensor Head
1 Unlock the sensor head connector, then insert it into the connector on the
side face of the amplifier. 2
2 Turn the round end of the connector clockwise until it clicks to lock it.
Click
Locked state
To detach the connector of the sensor head, follow the above procedures in the opposite order.
1 Unlock the sensor head connector, then insert it into the connector on the
back face of the amplifier.
Unlocked state
2 Turn the round end of the connector clockwise until it clicks to lock it.
To detach the connector of the sensor head, follow the above procedures in the
opposite order.
GT-M-E 2-15
2-2 Connecting and Mounting the Sensor Head
MEMO
2
Installation and Connection
2-16 GT-M-E
Basic Usage of the GT-70A Series 3
This chapter describes how to install and wire the GT-70A Series
as well as the basic setting method.
3
3-1 Main Screens ........................................................... 3-2
3-2 Origin Alignment .................................................... 3-12
3-3 Setting the Tolerances ........................................... 3-15
3-4 Displaying the Detection Result ............................. 3-21
3-5 Useful Functions .................................................... 3-22
GT-M-E 3-1
3-1 Main Screens
The main screen is a screen displayed during the normal operation (during
detection), and the display can be switched with the [Right]/[Left] arrow buttons.
The display pattern of the main screen differs depending on the setting of "00.
Calculation mode".
"4-6 Calculation Mode" (page 4-69)
3 When the GT-70A Series is used only with the main unit or when "C0. Calculation
function OFF" is set for "00. Calculation mode" with the expansion unit added, the
following main screen appears. (Common for the main and expansion units)
Basic Usage of the GT-70A Series
The main screen is switched as shown below with the [Right]/[Left] arrow buttons.
Reference "00. Calculation mode" is set to "C0. Calculation function OFF" when
shipped from the factory.
(5) Preset value setting screen (2) R.V. value display screens
(4) LOW setting value screen (3) HIGH setting value screen
3-2 GT-M-E
3-1 Main Screens
The P.V. (= Present Value) display shows the current measurement of the displaced
contact.
When the hold function is used (when the external input is performed while the
setting other than "Standard" is set for "01. Detection mode" or "External timing input"
is set for "03. Timing type"), the value being held is displayed.
When the hold function is not used, the same value as the R.V. value is displayed.
"01. Detection Mode" (Page 4-5)
3
"03. Timing Type" (Page 4-31)
Reference · Approximately two seconds after the title screen appears, it changes to
the digital LED display.
· From the "P.V. value setting screen", the tolerances can be set.
"Tolerance calibration" (page 3-17)
· When the calculation function is used, the status indicator "CALC" is lit
in red.
"4-6 Calculation Mode" (page 4-69)
· The preset, measurement increase/decrease direction, multiplier,
calibration function setting, and hold function are applied to the P.V.
value and displayed.
"Preset Function" (page 3-22)
"10. Measurement Increase/decrease Direction" (page 4-39)
"11. Multiplier" (page 4-40)
"30. Calibration Function Setting" (page 4-64)
GT-M-E 3-3
3-1 Main Screens
The R.V. (= Raw Value) value display shows the actual measured displacement of the
target, via the position of the spindle.
Reference · Approximately two seconds after the title screen appears, it changes to
the digital LED display.
· On the R.V. value display, the hold value (the value displayed using the
hold function) is not displayed.
· The preset, measurement increase/decrease direction, multiplier, and
calibration function setting are applied to the R.V. value and displayed.
"Preset Function" (page 3-22)
"10. Measurement Increase/decrease Direction" (page 4-39)
"11. Multiplier" (page 4-40)
"30. Calibration Function Setting" (page 4-64)
3-4 GT-M-E
3-1 Main Screens
Reference · The display switches to the value display screen in two seconds after it
switched to the title screen.
· The two-point calibration can be set in "(3) HIGH setting value screen"
and "(4) LOW setting value screen".
"Two-point calibration" (page 3-19)
GT-M-E 3-5
3-1 Main Screens
Reference · The display switches to the value display screen in two seconds after it
switched to the title screen.
· The two-point calibration can be set in "(4) LOW setting value screen"
and "(3) HIGH setting value screen".
"Two-point calibration" (page 3-19)
3-6 GT-M-E
3-1 Main Screens
For details on the preset function, refer to "Preset Function" (page 3-22).
The Preset setting The PRESET position
value blinks in red. indicator blinks in red.
Reference The display switches to the value display screen in two seconds after it
switched to the title screen.
GT-M-E 3-7
3-1 Main Screens
Main screens when using the calculation function differ depending on the calculation
mode set for "00. Calculation mode".
This section describes the main screens displayed when a setting other than "C5.
3 Reference difference" is selected for "00. Calculation mode".
"4-6 Calculation Mode" (page 4-69)
■ Main screens on the main unit
Basic Usage of the GT-70A Series
"(5) Preset value setting screen" (page 3-7) "(6) Calculated value display screen" (page 3-9)
"(4) LOW setting value screen" (page 3-6) "(2) R.V. value display screen" (page 3-4)
3-8 GT-M-E
3-1 Main Screens
Reference The display switches to the value display screen in two seconds after it
switched to the title screen.
"(2) R.V. value display screen" (page 3-4) "(5) Preset value setting screen" (page 3-7)
Note
When a setting other than "C5. Reference difference" is selected, the
function setting mode (set by pressing and holding the [MODE]
button) cannot be set on an expansion unit.
When pressing and holding the [MODE] button,
"CAL.Loc" appears.
(Basic Setting Mode, Additional Function Setting Mode, and
Calibration Setting Mode can be set)
GT-M-E 3-9
3-1 Main Screens
This section describes the main screens displayed when "C5. Reference difference"
is selected for "00. Calculation mode".
"4-6 Calculation Mode" (page 4-69)
■ Main screens on main unit
When "C5. Reference difference" is selected for "00. Calculation mode", the main
3 screens that can be displayed on the main unit are only "(2) R.V. value display" and
"(5) Preset value display".
Basic Usage of the GT-70A Series
"(2) R.V. value display screen" (page 3-4) "(5) Preset value setting screen" (page 3-7)
Note
When "C5. Reference difference" is selected, only "00. Calculation
mode" and "02. Response time" can be set for the function setting
mode (set by pressing and holding the [MODE] button) on the main
unit.
"02. Response Time" (page 4-29)
3-10 GT-M-E
3-1 Main Screens
"(4) LOW setting value screen" (page 3-6) "(2) R.V. value display screen" (page 3-4)
GT-M-E 3-11
3-2 Origin Alignment
When you use the GT-70A Series for the first time or when you replace the sensor
head, you must reset the origin.
The tolerance range will be now based on the new origin.
"3-3 Setting the Tolerances" (page 3-15)
The origin can be adjusted either by the sensor directly or from a remote external
input.
3 Note
• You cannot change the origin when "------", "-FFFF", or "FFFF" is displayed.
• Aligned origin is written to the internal memory every time, so the
Basic Usage of the GT-70A Series
changed origin is saved even after the power is turned off. The
origin alignment can be performed up to about a million times.
• If you frequently perform origin alignment through an external
input, select "OFF" for "20. Preset Memory". If "OFF" is selected,
origin alignment can be set, however it is not written to the internal
memory. (When the power is turned off, it returns to the previously
stored origin.)
"20. Store Preset Value" (page 4-51)
Reference We recommend that you should periodically reset the origin for more
accurate judgment.
This section describes the procedures of the origin alignment using the operation
buttons of the sensor amplifier.
1 Place the sensor's contact on the desired target to serve as the origin.
Target
3-12 GT-M-E
3-2 Origin Alignment
2 While in the main display, press the [Up] and [Down] arrow buttons at the
same time.
After "PrESEt" blinks several times on the amplifier display screen, "0.000" * is
displayed on the P.V. value display, and the origin has been set.
Press the [Up] and [Down] arrow buttons at the same time
* When using the preset function, the specified preset value is displayed.
Reference To reset the origin to factory default, press and hold the [Up] and [Down]
arrow buttons.
After "PrESEt" is displayed, "rESEt" blinks, indicating that the origin
alignment is reset to the factory default setting.
1 Set the display screen of the sensor amplifier to one of the main screens,
and perform detection of the master workpiece.
Master workpiece
<P.V. value display screen>
GT-M-E 3-13
3-2 Origin Alignment
2 In the preset value setting screen, set a value that you want to display
when detecting the master workpiece.
Example If you want to make the detected value for the master workpiece 10.000,
set the preset value to 10.000.
3 While the target is being detected, press the [Up] and [Down] arrow
buttons at the same time.
After "PrESEt" blinks several times on the amplifier display screen, "10,000" * is
displayed on the P.V. value display screen.
Master Master
workpiece workpiece
Detecting the master workpiece "PrESEt" blinks Detecting the master workpiece
(Before origin alignment is performed) (Performing the origin alignment) (After origin alignment is performed)
Press the [Up] and [Down] arrow buttons at the same time
By short-circuiting "Preset input" (pink line) from the external device, origin alignment
is performed.
"5-2I/O Diagram" (Page 5-7)
"Minimum Input Time" (page 5-11)
Reference When origin alignment is performed by the external input, "PrESEt" is not
displayed.
3-14 GT-M-E
3-3 Setting the Tolerances
The tolerances of the GT series are comprised of upper and lower limits. Any
measured value that goes beyond the upper or lower limits will activate the HIGH or
LOW outputs respectively. Any measured value that lies between the tolerances will
activate a GO output.
These outputs are displayed on both the sensor amplifier as well as the head. A HIGH
or LOW output is displayed as a red indicator light whereas a GO output is a green
light. The output indicator on the head is easy visible so that a user can simply glance
at the sensor and know whether the target is good or defective.
3
The head indicator blinks while the sensor head is being set, so even if multiple devices
Operation indicator
HI/GO/LO indications
Note
When setting the tolerances manually or by two-point calibration, be
sure to set them as "HIGH setting value > LOW setting value".
Reference The criterion output when setting as "HIGH setting value < LOW setting value" is as
follows.
· Regardless of the detected value, the GO criterion is not output.
· When the value that is over the HIGH setting value and under the LOW
setting value is detected, the HIGH criterion and the LOW criterion are
output at the same time.
GT-M-E 3-15
3-3 Setting the Tolerances
This is a method for directly setting the tolerances (HIGH setting value, LOW setting
value).
1 On the main screen of "HIGH setting value", set a value using the [Up]/
[Down] arrow button.
" (3) HIGH setting value screen" (page 3-5)
3
Pushing the [Up] arrow button increases
the setting value.
Basic Usage of the GT-70A Series
2 On the main screen of "LOW setting value", set a value using the [Up]/
[Down] arrow button.
" (4) LOW setting value screen" (page 3-6)
3 To return to the P.V. value display, press the [Right]/[Left] arrow buttons.
Reference When the HIGH setting value or LOW setting value is set, the new setting
values immediately take effect.
3-16 GT-M-E
3-3 Setting the Tolerances
The upper limit value (HIGH setting value) and lower limit value (LOW setting value)
are automatically set.
Tolerances can be set with "Tolerance calibration" when a master workpiece is
available and with "Two-point calibration" when actual good and defective targets are
available.
Tolerance calibration 3
This is a method for setting a range based on the detection value of a master
Note
• Tolerance calibration can be set only on the P.V. value display.
• Tolerance calibration cannot be performed
when "------", "-FFFF", or "FFFF" is
displayed.
Attempting to adjust the tolerances under
these conditions will flash a "no vAL" message.
1 While on the "P.V. value display", position the sensor head to detect the
master workpiece (you want to make as a reference for the acceptable value).
GT-M-E 3-17
3-3 Setting the Tolerances
Displays
3 <Tolerance setting screen> alternately. <Setting range for tolerance>
"SEt" blinks, and the HIGH setting value and LOW setting value are determined.
After it blinks for about a second, it returns to the P.V. value display.
Reference The HIGH setting value and LOW setting value can be adjusted after the
tolerances are set.
" (3) HIGH setting value screen" (page 3-5)
" (4) LOW setting value screen" (page 3-6)
3-18 GT-M-E
3-3 Setting the Tolerances
Two-point calibration
This is a method to set the intermediate value between a good target and a defective
target as the tolerance when a good target, a HIGH defective target and a LOW
defective target are available.
Note
The two-point calibration cannot be performed when "------", "-FFFF", or "FFFF" is
shown on the R.V. value display.
3
1 On the main screen of "HIGH setting value
GO
<HIGH setting value screen>
2 Press the [SET] button while the good target is being detected.
Displays
alternately. Detected value for the good target
3 Press the [SET] button while the HIGH defective target is being detected.
The value for the HIGH defective target is captured.
GT-M-E 3-19
3-3 Setting the Tolerances
GO
5 Press the [SET] button while the good target is being detected.
Basic Usage of the GT-70A Series
Displays
alternately.
Detected value for the good target
6 Press the [SET] button while the LOW defective target is being detected.
The value for the LOW defective target is captured.
Reference The HIGH setting value and LOW setting value can be adjusted after
the two-point calibration is set.
" (3) HIGH setting value screen" (page 3-5)
" (4) LOW setting value screen" (page 3-6)
3-20 GT-M-E
3-4 Displaying the Detection Result
After setup of the GT-70A Series is completed the following results are displayed on
the sensor amplifier and sensor head.
In this example, the tolerances "HIGH setting value = 6.000" and "LOW setting value
= 3.000" are set.
3
Lights in red Lights in green Lights in red
GO range
HIGH setting
value
LOW setting
value
LOW range
HIGH GO LOW
GT-M-E 3-21
3-5 Useful Functions
Preset Function
When the preset value is set, the displayed value can be changed by adding the
desired value (preset value).
Example For displaying the value by adding "10.000"
• Adjust the preset on the main screen of "Preset value setting screen".
" (5) Preset value setting screen" (page 3-7)
• The criterion outputs are judged by setting the value to which the preset value is
added as the detected value. After the preset values is set, set the criterion values
(HIGH setting value/LOW setting value) again.
• By combining with the origin alignment, a desired value can be displayed as the
detected value of the workpiece (master workpiece).
"Origin Alignment" (page 3-12)
3-22 GT-M-E
3-5 Useful Functions
Bank Function
With GT-70A Series, you can register up to four pre-registered HI/GO/LO settings.
• HIGH setting value
• LOW setting value
• Preset value
By using the bank function, registered setting items can be switched easily
according to the changes of the type of the detection target. 3
How to switch the bank
<Bank 0 is selected>
<Bank 1 is selected>
<Bank 3 is selected>
GT-M-E 3-23
3-5 Useful Functions
Note
The input line function 2 does not default to "Bank B" when shipped
from the factory.
3-24 GT-M-E
3-5 Useful Functions
Input line 1
Bank Input line 2 (pink/purple)*
(purple)*
Bank 0 OFF OFF
Bank 1 ON OFF
Bank 2 OFF ON ON: Short-circuited
Bank 3 ON ON OFF: Open
Brown* Brown*
+V +V
Purple/
Blue* pink stripe*
0V
Blue*
0V
Purple/pink
stripe*
* The expansion units (GT-72A/GT-76A, GT-72AP/GT/76AP) have no
brown-blue line. (The power wires of the main unit and those of the
expansion unit are bussed together.)
"I/O Diagram" (page 5-7)
GT-M-E 3-25
3-5 Useful Functions
Hysteresis
If the P.V. value is fluctuating around the tolerances, the outputs may repeatedly turn
on and off. By increasing the hysteresis value, the delta between the SET and
RESET distance widens to reduce output chattering.
You can set the hysteresis in the basic setting mode of "14. Hysteresis".
Refer to "14. Hysteresis" (page 4-43) for details on hysteresis.
3
Displayed value
Basic Usage of the GT-70A Series
Hysteresis
Hysteresis
LOW setting value
time
ON
HIGH output
OFF
ON
GO output OFF
ON
LOW output
OFF
Reference The hysteresis is set to "0.003" when shipped from the factory.
3-26 GT-M-E
3-5 Useful Functions
Initialization
1 On the main screen, while holding down the [MODE] button, press the
[SET] button five times. 3
The initializing screen appears.
<Initialization screen>
"YES" is blinking.
GT-M-E 3-27
3-5 Useful Functions
The display unit can be switched from the initial reset screen.
This section explains how to switch from "mm" display to "inch" display.
3-28 GT-M-E
3-5 Useful Functions
Key Lock
Under the key locked state, while holding down the [MODE] button, press the [Up]
arrow button for two seconds or more.
The key lock cancel screen appears and then changes to the main screen.
GT-M-E 3-29
3-5 Useful Functions
MEMO
3
Basic Usage of the GT-70A Series
3-30 GT-M-E
Setting Various Functions 4
GT-M-E 4-1
4-1 Detection Functions
With the GT-70A Series, you can set various functions to support many kinds of
detection methods.
Additional Power save and core alarm control. [MODE] button and the
function
"4-4 Additional Function Setting Mode" left arrow key at the same
setting
(page 4-49) time for at least two
mode
seconds.
Press and hold the
Calibration Sets the calibration function. [MODE] button and the
setting right arrow key at the
mode "4-5 Calibration Setting Mode" (page 4-62) same time for at least two
seconds.
Calculation based on the multiple detected Press and hold the
targets to check the degree of flatness, twist, or [MODE] button for at least
Calculation
warpage. two seconds. (This mode
mode
can be set only when the
"4-6 Calculation Mode" (page 4-69) expansion unit is added.)
Reference Each setting mode can be entered only from the main screen.
"3-1 Main Screens" (page 3-2)
4-2 GT-M-E
4-2 Function Setting Mode
Reference
Setting item Setting range Description
page
Detection mode Standard/
NG hold/
Types of detection that are
peak hold/ 4-5
application specific.
bottom hold/
Peak-to-peak
Response time
hSP (1.5)/5/10/100/ Period of time prior to sampling
4-29
500/1000/5000 (ms) data.
Self-timing delay
type Specifies how to display the value
Static hold/delay timer when detection exceeds (or drops 4-33
below) the self-timing level.
User-specified
delay time
0 to 9999 (ms) Sets the delay time. 4-34
Maximum value/
minimum value/degree
Calculation mode of flatness/average
value/reference "4-6 Calculation Mode" 4-69
difference/twist/
warpage
/thickness
GT-M-E 4-3
4-2 Function Setting Mode
· On the main screen, press and hold the [MODE] button for at least two
seconds.
[MODE] button
4
Main screen (P.V. value display)
Setting Various Functions
· Press and hold the [MODE] button for at least two seconds.
· Or press the right arrow key several times until the setting completion display
appears, and press the right arrow key once more.
4-4 GT-M-E
4-2 Function Setting Mode
MODE
Standard
NG hold
4
Peak hold
Peak-to-peak
Initial
Detection mode Description
value
Standard detection On the P.V. value display, the displayed values and
tolerances are always updated and can be held during the
timing input. X
"Standard detection" (page 4-6)
NG hold detection Holds the output when the detected value exceeds (drops
below) the HIGH (LOW) setting during external timing.
"NG hold detection" (page 4-17)
Peak hold
Holds the peak value during external (or internal) timing.
"Hold detection (peak hold) "(page 4-21)
Hold detection
Bottom hold
Holds the bottom value during external (or internal) timing.
"Hold detection (bottom hold)" (page 4-26)
Peak-to-peak Holds the peak-to-peak value by within the "Sampling
period" defined external timing.
"Hold detection (peak-to-peak) "(Page 4-23)
GT-M-E 4-5
4-2 Function Setting Mode
Standard detection
In the standard detection, the displayed value and tolerances are always updated on
the P.V. value display.
In addition, these same values can be held via external or internal timing.
■ When the external timing is used
Timing chart
This section describes changes of the displayed value and tolerances by the
external timing input.
4 In the following example, "N.O" is set for "12. Output mode".
"12. Output Mode" (page 4-41)
Setting Various Functions
Display value
The displayed value is held while
the External timing input is ON.
R.V. value
HIGH setting value
Time
ON
HIGH output
OFF
ON
GO output OFF
ON
LOW output OFF
ON
External timing input OFF
(1) While the external timing input is off, the displayed value and tolerances are updated.
(2) The P.V. value is held when the external timing is active.
(3) When the external timing input is off, the hold state of the P.V. value is cancelled.
4-6 GT-M-E
4-2 Function Setting Mode
Setting procedure
MODE
Reference To use the external timing, "External timing Press for
input" must be set for "Input 2" in the basic 2 sec
setting mode (default setting when shipped
Return to the main display
from the factory).
"16. Input Line Function 2" (page 4-48)
GT-M-E 4-7
4-2 Function Setting Mode
Time
(Old) Output criterion Delay time (New) Output criterion result hold period Delay time
result hold period
HIGH output ON
OFF
GO output ON
OFF
LOW output ON
OFF
4-8 GT-M-E
4-2 Function Setting Mode
Note
If the detected value drops below the self-timing
level before the delay time has elapsed, the data
cannot be held. This results in a display error as
shown on the right.
"Error Messages" (page A-5)
In this situation, the Hi and Lo outputs will be on
while the Go is off.
Press the SET button to release the error state.
GT-M-E 4-9
4-2 Function Setting Mode
Setting procedure
4-10 GT-M-E
4-2 Function Setting Mode
MODE
Press for
2 sec
Return to the main display
GT-M-E 4-11
4-2 Function Setting Mode
Timing chart
This section describes changes of the displayed value and outputs according to the
internal timing.
In the following example, "N.O" is set for "12. Output mode".
"12. Output Mode" (page 4-41)
Time
(Old) Output criterion Sampling period (New) Output criterion result hold period Sampling period
result hold period
HIGH output ON
OFF
GO output ON
OFF
LOW output ON
OFF
4-12 GT-M-E
4-2 Function Setting Mode
Note
If the detected value drops below the self-timing
level before the detected value is stabilized, an
error display appears as shown in the right.
"Error Messages" (page A-5)
In this situation, the Hi and Lo outputs will be on
while the Go is off.
Press the SET button to release the error state.
When the error display is reset, the criterion
standby display as shown in the right is
displayed. Criterion standby
display
4
GT-M-E 4-13
4-2 Function Setting Mode
Setting procedure
4-14 GT-M-E
4-2 Function Setting Mode
GT-M-E 4-15
4-2 Function Setting Mode
MODE
4-16 GT-M-E
4-2 Function Setting Mode
NG hold detection
In the NG hold detection mode, the output is held if the HIGH or LOW output turns on
during the period specified by the external timing (while the external timing input is
off).
Time
(Old) Output criterion result hold period Output criterion period (New) Output criterion result hold period
HIGH output ON
OFF
ON
GO output OFF
LOW output ON
OFF
(1) (2) (3) (4)
(1) The last output result is held while the external timing input is on.
(2) The HIGH/LOW outputs are monitored while the external timing input is off.
(3) If neither the HIGH or LOW outputs are on then the GO output will be active so long as external timing
input is off.
(4) The hold status of the outputs and displayed value that have been held will be released when the
next external timing input turns off.
GT-M-E 4-17
4-2 Function Setting Mode
Display value Holds the HIGH criterion output Holds the P.V. value
4 Time
(Old) Output criterion result hold period Output criterion period (New) Output criterion result hold period
OFF
HIGH output ON
OFF
ON
GO output OFF
LOW output ON
OFF
(1) (2) (3) (4) (5)
(1) The last output result is held while the external timing input is on.
(2) The HIGH/LOW outputs are monitored while the external timing input is off.
(3) If neither the HIGH or LOW outputs are on then the GO output will be active so long as external timing
input is off.
(4) The hold status of the outputs and displayed value that have been held will be released when the
next external timing input turns off.
4-18 GT-M-E
4-2 Function Setting Mode
Time
4
(Old) Output criterion result hold period Output criterion period (New) Output criterion result hold period
(1) The last output result is held while the external timing input is on.
(2) The HIGH/LOW outputs are monitored while the external timing input is off.
(3) If the LOW is detected while the external timing input is off, the LOW output is held. Once the output
is held, it will never be changed by further measurements. At this time, the displayed value is not held
and the detected value is displayed.
(4) Because the LOW has been detected while the external timing input is off, the LOW output is held
even after further measurements. The P.V. value at the rise time of the external timing input is held
as a displayed value.
The displayed value is held within the HIGH range until the next external timing input turns off, upon
which the LOW is held.
(5) The hold status of the outputs and displayed value that have been held will be released when the
next external timing input turns off.
GT-M-E 4-19
4-2 Function Setting Mode
Setting procedure
MODE
Press and hold the button for two seconds or Return to the main display
more to end the setting.
Hold detection
The hold detection detects peak value, bottom value, and the peak-to-peak value (only
when the external timing is used).
4-20 GT-M-E
4-2 Function Setting Mode
R.V. value
Sampling period C
peak value
4
LOW setting value
External timing ON
OFF
input
HIGH output OFFON
ON
GO output OFF
ON
LOW output OFF
(1) The maximum value (peak value) is held while the external timing input is off (sampling period A ).
(2) At the rise time of the external timing input (end of the sampling period A ), the peak value held during
the sampling period A is displayed. This new peak value is now considered the measured value and
the GO output activates.
(3) The maximum value (peak value) is held while the external timing input is off (sampling period B ).
(4) At the rise time of the external timing input (end of the sampling period B ), the peak value held
during the sampling period B is displayed. This new peak value is now considered the measured
value and the HIGH output activates.
(5) The maximum value (peak value) values is held while the external timing input is off (sampling period
C ).
(6) At the rise time of the external timing input (end of the sampling period C ), the peak value held during
the sampling period C is displayed. This new peak value is now considered the measured value and
the GO output activates.
GT-M-E 4-21
4-2 Function Setting Mode
input
ON
HIGH output OFF
ON
GO output OFF
ON
LOW output OFF
(1) The minimum value (bottom value) is held while the external timing input is off (sampling period A ).
(2) At the rise time of the external timing input (end of the sampling period A ), the bottom value held
during the sampling period A is displayed. This new bottom value is now considered the measured
value and the GO output activates.
(3) The minimum value (bottom value) is held while the external timing input is off (sampling period B ).
(4) At the rise time of the external timing input (end of the sampling period B ), the bottom value held
during the sampling period B is displayed. This new bottom value is now considered the measured
value and the GO output activates.
(5) The minimum value (bottom value) is held while the external timing input is off (sampling period C ).
(6) At the rise time of the external timing input (end of the sampling period C ), the bottom value held
during the sampling period C is displayed. This new bottom value is now considered the measured
value and the LOW output activates.
4-22 GT-M-E
4-2 Function Setting Mode
(1) The maximum value (peak value) and the minimum value (bottom value) are held while the external
timing input is off (sampling period A ).
(2) At the rising edge of the external timing input (end of the sampling period A ), the bottom value is
subtracted from the peak value held during the sampling period A . This new delta value is now
considered the measured value and the HIGH output activates.
(3) The maximum value (peak value) and the minimum value (bottom value) are held while the external
timing input is off (sampling period B ).
(4) At the rising edge of the external timing input (end of the sampling period B ), the bottom value is
subtracted from the peak value held during the sampling period B . This new delta value is now
considered the measured value and the GO output activates.
GT-M-E 4-23
4-2 Function Setting Mode
Setting procedure
Note
If "Peak-to-peak" is selected in step 1,
"03. Timing Type" cannot be specified.
(Because it is fixed to "t-in".)
Reference To use the external timing, "External timing input" must be set for "Input
line function 2" in the basic setting mode.
"16. Input Line Function 2" (page 4-48)
4-24 GT-M-E
4-2 Function Setting Mode
Note
When the power is turned on or the error
display is reset, the criterion standby display
occurs until the first P.V. value is fixed and all Criterion standby display
the outputs are turned off.
Time
(Old) Output criterion result hold period Sampling period A (New) Output criterion result hold period Sampling period B
ON
HIGH output OFF
GO output ON
OFF
ON
LOW output
OFF
GT-M-E 4-25
4-2 Function Setting Mode
Self-timing level
P.V. value previously held Holds the P.V. value and
the output criterion result
HIGH setting value
Sampling period
A bottom value P.V. value currently held
Time
(Old) Output criterion result hold period Sampling period A (New) Output criterion result hold period Sampling period B
Setting Various Functions
ON
HIGH output
OFF
ON
GO output OFF
ON
LOW output OFF
4-26 GT-M-E
4-2 Function Setting Mode
Setting procedure
cannot be performed.
GT-M-E 4-27
4-2 Function Setting Mode
MODE
Press for
2 sec
4-28 GT-M-E
4-2 Function Setting Mode
The response time is the time from when the sensor head starts detection until the
value at the detection position is determined as a comparator value (P.V. value).
MODE
Setting value
input screen
Press and hold
MODE
for at least 2 sec
4
"Entering function setting mode" (page 4-4)
GT-M-E 4-29
4-2 Function Setting Mode
Example If you set a response time of 5 ms (average time: 4 ms, number of times
for averaging: 8 times), data will be obtained and output as follows:
Sampling period (0.5 ms)
1 2 3 4 5 6 7 8 9 10 11
Averages data 1 to 8
Average measurement time
and outputs data A.
Measured data A B C
4 Refresh cycle Refresh cycle
Setting Various Functions
Reference • If you want to detect instantaneous changes, set a short response time.
• If you want to ignore instantaneous changes or fluctuations of values,
set a long response time.
4-30 GT-M-E
4-2 Function Setting Mode
4
MODE
Default
Item Description
Setting
External timing
input Select this when inputting a timing using the input*1 (pink/purple
stripe line) from an external device. X
"5-2 I/O Diagram" (page 5-7)
When the detected value exceeds the self-timing level (a
change from the value below the self-timing level to the one
Rising edge above the self-timing level), the timing input occurs.
self-timing
Self timing level
Timing input occurrence
Detected value
"04. Self-timing Level" (page 4-32)
When the detected value drops below the self-timing level (a
change from the value above the self-timing level to the one
Falling edge below the self-timing level), the timing input occurs.
self-timing Detected value Timing input occurrence
Self timing level
*1 To use the external timing, "External timing input" must be set for "Input line function 2"
in the basic setting mode (default setting when shipped from the factory).
"16. Input Line Function 2" (page 4-48)
GT-M-E 4-31
4-2 Function Setting Mode
The self-timing level is the set point to trigger the timing input automatically when it
detects a change in the values.
Reference This item can be set only when either (Rising edge self-timing) or
SELF.~L (Falling edge self-timing) is selected for " (Timing
type)".
"03. Timing Type" (page 4-31)
4
MODE
Setting Various Functions
Setting value
input screen MODE
Press and hold for
at least 2 sec
Default
Item Setting range
Setting
Self timing level -99.999 to 999.999 (unit: mm) 0.500
4-32 GT-M-E
4-2 Function Setting Mode
With the self timing delay, select the method to determine the criterion value after the
timing input is generated by the self-timing level.
Reference This item can be set only under the following conditions.
• Skd (Standard) is selected for "01. fn < (Detection mode)".
• Either (Rising edge self-timing) or SELF.~L (Falling edge self-
timing) is selected for " (Timing type)".
"01. Detection Mode" (page 4-5)
"03. Timing Type" (page 4-31) 4
Static hold
MODE
Delay timer Press and hold for
at least 2 sec
Timing input
Self-timing level
Response time
R.V. value
Timing input
Self-timing level
Delay time
GT-M-E 4-33
4-2 Function Setting Mode
Set the delay time to fix the criterion value after the timing input occurs using the self-
timing input.
Reference This item can be set only when "k.d (Delay timer) is selected for "05.dLY
(Self-timing delay type)".
"05. Self Timing Delay" (page 4-33)
4
MODE
Setting Various Functions
Setting value
input screen MODE
Press and hold for at least 2 sec
Default
Item Setting range
Setting
User-specified delay time 0 to 9999 (unit: ms) 1000
4-34 GT-M-E
4-2 Function Setting Mode
Select the setting method for the value of the stability width in the static hold delay
stability.
Reference This item can be set only when "Skb.d (Static hold) is selected for 05.dLY
(Self-timing delay type)".
"05. Self Timing Delay" (page 4-33)
4
MODE
MODE
User Press and hold for
at least 2 sec
Default
Item Description
Setting
The stability width is set to the default value (0.010 mm).
R.V. value
X
Timing input
Self-timing level
R.V. value
Timing input
Self-timing level
GT-M-E 4-35
4-2 Function Setting Mode
Set the value of the stability width when the static hold delay is set.
Reference This item can be set only when "USEr (User) is selected for "07.Skb
(Static hold delay stability criterion)".
"07. Static Hold Delay Stability" (page 4-35)
4 MODE
Setting Various Functions
Setting value
input screen MODE
Press and hold for
at least 2 sec
Default
Item Setting range
Setting
Static hold
0.000 to 999.999 (unit: mm) 0.010
delay stability width
4-36 GT-M-E
4-3 Basic Setting Mode
Reference
Setting item Setting range Description
page
Measurement
increase/decrease Displays the value detected when the
direction sensor head spindle is pressed as a
Normal/Reverse 4-39
positive (normal) or negative (reverse)
Multiplier
value.
4
Sets the multiplier for the displayed
0.1 to 100.0 4-40
detection value.
Displayed number
of digits
0.001/0.01/0.1/1 Sets the number of displayed digits. 4-42
Hysteresis
0.000 to 999.999 Sets the hysteresis. 4-43
GT-M-E 4-37
4-3 Basic Setting Mode
From the main screen, press and hold the [MODE] and [SET] buttons at the
same time for at least two seconds.
4
Setting Various Functions
Use one of the following methods to exit basic setting mode and display the main
screen.
• After the settings are complete, press and hold the [MODE] button for at least
two seconds.
• After the settings are complete, press the right arrow key several times until
the setting completion display appears, and press the right arrow key once
more.
4-38 GT-M-E
4-3 Basic Setting Mode
Display the value detected when the sensor head spindle is pressed as a positive
(normal) or negative (reverse) value.
MODE
Normal 4
MODE
Reverse
Measurement
Default
increase/decrease Description
Setting
direction
Value increases based on the Display value
Normal amount that the spindle is
FFFF
pushed in (amount of
movement). X
-FFFF
0 Spindle movement
-FFFF
0 Spindle movement
Note
When changing the measurement increase/decrease direction, reset
the following parameters:
• HIGH setting value ( "(3) HIGH setting value screen" (page 3-5))
• LOW setting value ( "(4) LOW setting value screen" (page 3-6))
• Timing type ( "03. Timing Type" (page 4-31))
• Self timing level ( "05. Self Timing Delay" (page 4-33))
GT-M-E 4-39
4-3 Basic Setting Mode
11. Multiplier
When the detection point is located beyond a lever or in other similar cases, the
display value can be displayed by multiplying the actual detected value.
MODE
Setting value
4 input screen
Press and hold
MODE
for at least 2 sec
Setting Various Functions
Default
Item Setting range
Setting
Multiplier 0.1 to 100.0 1.0
4-40 GT-M-E
4-3 Basic Setting Mode
MODE
N.O.
Default
Output mode Description
Setting
N.O.
(Normal open)
Outputs ON for criterion output. X
N.C.
(Normal close)
Outputs OFF for criterion output.
Example
Judgment Output mode Criterion output
HIGH output: On
N.O. (Normal open) GO output: Off
LOW output: Off
HIGH criterion
HIGH output: Off
N.C. (Normal close) GO output: On
LOW output: On
GT-M-E 4-41
4-3 Basic Setting Mode
Set the number of digits of the display value in the digital display.
Change this setting when you want to display only necessary digits.
MODE
4 Displayed
digits selection MODE
Press and hold for
Setting Various Functions
at least 2 sec
Default
Item Setting range
Setting
Displayed number of digits 0.001/0.01/0.1/1 0.001
Reference • Changes to the displayed number of digits are applied only to the P.V.
value, R.V. value, and calculation value displays.
• The displayable range is not affected by the displayed number of digits
setting. Therefore, the displayable range is -99 to 999 even if the
displayed number of digits is set to "1."
4-42 GT-M-E
4-3 Basic Setting Mode
14. Hysteresis
Reference • Hysteresis refers to the characteristic where the criterion value that 4
turns on the criterion output and the one that turns off the criterion
output are different.
MODE
Setting value
input screen
MODE
Press and hold for
at least 2 sec
Default
Item Setting range
Setting
Hysteresis 0.000 to 999.999 0.003
GT-M-E 4-43
4-3 Basic Setting Mode
Setting the hysteresis to 0 causes fluctuations in the P.V. value to directly affect the
criterion output, so it is useful for applications that require strict tolerance measurements.
However, when the detected values fluctuate close to the tolerances, the criterion
output becomes unstable.
GO area
R.V. value
Time
ON
HIGH output OFF
ON
GO output OFF
ON
LOW output OFF The criterion output is unstable when both HIGH and
GO criterion outputs are on or off
When the hysteresis is set, the output does not react excessively to fine variations in
the P.V. value close to the tolerances and becomes stable.
A large hysteresis tends to produce a more stable output, but it also becomes more
difficult to output for GO when the P.V. value varies close to the tolerances.
GO area
R.V. value
HIGH criterion
value Hysteresis
Hysteresis width
Hysteresis LOW
criterion value
Time
ON
HIGH output OFF
ON
GO output OFF
ON
LOW output OFF Output is stable even when
the P.V. value fluctuates
4-44 GT-M-E
4-3 Basic Setting Mode
4
MODE
Default
Item Description
Setting
Bank A
Sets the input line 1 (purple) as the bank input line.
The bank switches as the input line 1 is turned on and off. X
Reset input Sets the input line 1 (purple) as the reset input line.
The P.V. value, R.V. value, and criterion output are reset
when the input line 1 is turned on or off.
Reference When performing bank switching with external input, input line 1 can be
turned on and off in combination with input line 2 (pink/purple) to set bank
Nos. 0 to 3.
"Bank Function" (page 3-23)
GT-M-E 4-45
4-3 Basic Setting Mode
Reset input
When a reset signal is input from an external device, all of the outputs are turned off
(when N.O. is set) and the P.V. and R.V. values become "- - - - - -".
Reference The resetting time is the same as the response time set in "02. Response
Time" (page 4-29).
Display value
Start reset
End reset
Time
Max. 20 ms
Reset input ON
OFF
Min. 20 ms
HIGH output ON
OFF
GO output ON
OFF
LOW output ON
OFF
20 ms pass after previous Resetting time
reset input falling edge (depends on response time)
Note
Wait until 20 ms have passed after the previous reset input has fallen
before entering a reset input. If fewer than 20 ms pass, the reset input
may not be recognized.
4-46 GT-M-E
4-3 Basic Setting Mode
Reference The resetting time is the same as the response time set in "02. Response
Time" (page 4-29).
Sampling period A
Peak value Sampling period B peak value
Displayed Resetting time
value R.V. value (depends on response time)
currently held
R.V. value
currently held
Time
Sampling period A Sampling period B Sampling period C
Max. 20 ms
ON ⑤
Reset input OFF
Min. 20 ms
ON
External timing input OFF
HIGH output ON
OFF
GO output ON
OFF
LOW output ON
OFF
(1) The peak value is stored while the external timing input is off (sampling period A ).
(2) At the rise time of the external timing input (end of the sampling period A ), the peak value stored during the
sampling period A is displayed. This peak value is taken as a measured value and the GO output turns on.
(3) Peak value is stored while the external timing input is off (sampling period B ).
(4) At the rise time of the external timing input (end of the sampling period B ), the peak value stored
during the sampling period B is displayed. The peak value is taken as the measured value, In this
case, The HIGH output turns on.
(5) When a reset signal currently held is activated, all of the criterion outputs are turned off 20 ms after
the reset input started at the latest. At this time, "- - - - - -" is displayed on the sensor amplifier.
(6) After the resetting time ends, sampling starts when the external timing input turns off. Sampling continues
until the next external timing input turns on and the peak value is saved.
(7) At the rise time of the external timing input (end of the sampling period C ), the peak value stored
during the sampling period C is displayed. This peak value is taken as a measured value and the
GO output turns on.
Note
Wait until 20 ms have passed after the previous reset input has fallen before entering
a reset input. If fewer than 20 ms pass, the reset input may not be recognized.
GT-M-E 4-47
4-3 Basic Setting Mode
4 MODE
MODE
Bank B Press and hold
for at least 2 sec
Default
Item Description
Setting
External timing input
Sets the input line 2 (pink/purple) as the external timing
input line. X
Bank B input
The bank switches as the input line 2 (pink/purple) is
turned on and off.
Reference When performing bank switching with external input, input line 2 can be
turned on and off in combination with input line 1 (purple) to set bank
Nos. 0 to 3.
"Bank Function" (page 3-23)
4-48 GT-M-E
4-4 Additional Function Setting Mode
Reference
Setting item Setting range Description
page
Store preset value
Sets whether to store the position set
YES/NO 4-51
for origin alignment.
Display filter
function Set this parameter to prevent
ON/OFF 4-61
fluctuations on the display value.
GT-M-E 4-49
4-4 Additional Function Setting Mode
The following method can be used to enter additional function setting mode.
On the main screen, press and hold the [MODE] button and the left arrow key at
the same time for at least two seconds.
[MODE] button
4
Main screen (P.V. value display)
Setting Various Functions
Use one of the following methods to exit additional function setting mode and display
the main screen.
• After the settings are complete, press and hold the [MODE] button for at least
two seconds.
• After the settings are complete, press the right arrow key several times until
the setting completion display appears, and press the right arrow key once
more.
4-50 GT-M-E
4-4 Additional Function Setting Mode
The origin information is stored into memory. The information is saved even after the
power is turned off.
When the sensor amplifier turns on, it loads the latest saved origin information.
"3-2 Origin Alignment" (Page 3-12)
Note
Information can be saved in the storage area even when the power is
turned off about one million times.
Set "NO" in situations such as when the origin is frequently aligned 4
with an external signal.
YES
MODE
NO Press and hold for at least 2 sec
Default
Item Description
Setting
YES (save) The origin information is saved in memory and is not cleared
when the power is turned off.
Set this parameter when the origin for the detected
X
workpieces does not change frequently.
The origin information is saved temporary in memory but is
NO (do not save) cleared when the power is turned off.
The origin information is cleared when the power is turned
off, so origin alignment must be performed every time that
the power is turned on.
GT-M-E 4-51
4-4 Additional Function Setting Mode
Set this parameter when you want to reduce the consumption power or when you do
not want to display the specific values.
When operating in the power save state, pressing any button will return to the normal
display. If buttons are not pressed in the normal display for one minute, the GT-70A
Series returns to power save mode.
Note
When operating in power save mode, only "Origin alignment" is active. To perform
4 other operations, press any button to return to the normal display before operating.
Setting Various Functions
MODE
Off
Half-eco
MODE
Press and hold for at least 2 sec
Full eco
Default
Item Description
Setting
Off
Power save is turned off.
Sensor amplifier: Normal display X
Sensor operation indicator: Normal display
Half-eco Only the sensor amplifier enters power save mode.
Sensor amplifier: Animation display*
Sensor operation indicator: Normal display
Full eco The sensor amplifier and sensor head both enter power save
mode.
Sensor amplifier: All indicators off (nothing is displayed)
Sensor operation indicator: Indicators off
* Only one segment flashes sequentially from right to left on the digital
display. If an error occurs, the error is displayed.
4-52 GT-M-E
4-4 Additional Function Setting Mode
If a contact sensor is used for a long period of time, the spindle may lose its ability to
fully expand, which reduces its ability to perform detection correctly. The core alarm
function checks for this malfunction and outputs error.
Note
Core alarm can only be set if "t-in" (External timing) is selected in
"03. Timing type".
OFF
ON
MODE
User Press and hold for at least 2 sec
Default
Item Description
Setting
OFF
The core alarm function is not used. X
GT-M-E 4-53
4-4 Additional Function Setting Mode
Core alarm
The core alarm checks whether the spindle returns to its original position (if it is fully
extended or fully compressed) during the non-detection state*1 that occurs after
detection ends and before the next detection begins. An error outputs if the spindle
4 fails to returns to its original position.
The non-detection state*1 is determined by the external timing input. The non-
detection state occurs when the external timing input is on. While the external timing
Setting Various Functions
input is on, if the spindle reaches a fully extended (or fully compressed) state even
once*2, this is recognized as normal operation. If the fully extended (or fully
compressed) state is not reached*2, an error is output.
Normal Error
Fully extended Fully extended Not fully extended
Detecting Detecting
②
①
Start detection End detection Start detection End detection Start detection
External timing input ON
OFF Non-detection Non-detection
ON
HIGH output OFF
The core alarm is indicated
by both HIGH and LOW
outputs activating at the
GO output ON same time
OFF
LOW output ON
OFF
4-54 GT-M-E
4-4 Additional Function Setting Mode
In the diagram shown above, the detection starts (external timing input is off) with the
spindle fully extended and ends (external timing input is on: output hold) when it
reaches the detection point.
No error is output in (1) since the spindle is fully extended during the non-detection
state (external timing input is on), while an error (the core alarm error " ") is
output (both the HIGH output and the LOW output are on) in (2) as soon as entering
the detection state (external timing input is off).
Fixture
Workpiece
(1) During non-detection, the sensor head is already touching the fixture, so the
spindle is not fully extended (does not reach the -FFFF position).
(2) During detection, the fixture is used to detect the workpiece.
(3) If the spindle sticks, detection cannot be performed correctly.
GT-M-E 4-55
4-4 Additional Function Setting Mode
Setting procedure
4-56 GT-M-E
4-4 Additional Function Setting Mode
Core alarm
checkpoint
Set
the checkpoint
Teaching
position 4
MODE
Press for
2 sec
Return to the main screen
GT-M-E 4-57
4-4 Additional Function Setting Mode
Sets the reference point when setting a desired position for the core alarm checkpoint.
Note Core alarm teaching can only be set if "22. Core Alarm" is set to
"USEr (User)".
4
MODE
Press and hold for
Set reference
Setting Various Functions
at least 2 sec
position
4-58 GT-M-E
4-4 Additional Function Setting Mode
Note Core alarm teaching can only be set if "22. Core Alarm" is set to
"USEr (User)".
MODE 4
Default
Item Setting range
Setting
Core alarm checkpoint setting -65.535 to 65.535 (Unit: mm) 0.500
Note
Do not set values higher or lower than the setting range (-99.999 to
-65.536 and 65.536 to 99.999). Errors may occur during normal
operations.
GT-M-E 4-59
4-4 Additional Function Setting Mode
Setting range:
Core alarm checkpoint
0 to 65.535*
Core alarm teaching
4-60 GT-M-E
4-4 Additional Function Setting Mode
The display filter function makes the display response speed slower and prevents
fluctuations in display values. The display becomes clearer and easier to read.
MODE
ON 4
MODE
OFF Press and hold for
Default
Item Description
Settings
ON
Prevents fluctuations in the display value. X
GT-M-E 4-61
4-5 Calibration Setting Mode
Reference
Setting item Setting range Description
page
Calibration
function setting Set when performing "Origin alignment"
Default/Adjustment or "Span adjustment" after situations 4-64
such as replacing the sensor head.
4 1st detection/
reading
(Origin alignment) ———— Sets the origin position. 4-65
Setting Various Functions
2nd detection/
reading
(Span adjustment) Sets the detection position for span
———— 4-67
adjustment.
4-62 GT-M-E
4-5 Calibration Setting Mode
On the main screen, press and hold the [MODE] button and the right arrow key
at the same time for at least two seconds.
[MODE] button
4
<Main screen (P.V. value display)>
Use one of the following methods to exit calibration setting mode and display the
main screen.
• After the settings are complete, press and hold the [MODE] button for at least
two seconds.
• After the settings are complete, press the right arrow key several times until
the setting completion display appears, and press the right arrow key once
more.
GT-M-E 4-63
4-5 Calibration Setting Mode
After replacing a sensor head, this parameter can be used to perform origin
alignment or span adjustment. This reduces the effects of misalignment that occurs
during installation and produces detection values that are closer to the values
achieved before replacement.
4 MODE
Default
Setting Various Functions
MODE
Adjustment Press and hold for 2 sec.
Adjustment
Performs origin alignment and span adjustment.
4-64 GT-M-E
4-5 Calibration Setting Mode
1st Detection/Reading
Sets the origin, which is used as the reference for the calibration settings.
Note
The parameter can be set only when "30. Calibration function setting" is
set to "AdJ (Adjustment)".
4
1st detection
GT-M-E 4-65
4-5 Calibration Setting Mode
Note
The parameter can be set only when "30. Calibration function setting" is
set to "AdJ (Adjustment)".
4
MODE
Setting Various Functions
Setting value
input screen MODE
Press and hold for 2 sec
Initial
Item Setting range
value
2nd target setting -99.999 to 99.999 (Unit: mm) 5.000
4-66 GT-M-E
4-5 Calibration Setting Mode
2nd Detection/Reading
Reads the detection value for span adjustment and corrects for the amount of
difference between the 2nd target value.
Note
The parameter can be set only when "30. Calibration function setting" is
set to "AdJ (Adjustment)".
Detection
reading MODE
Press and hold for 2 sec
GT-M-E 4-67
4-5 Calibration Setting Mode
Note
• The error screen appears as the calibration
error in the following cases. When the error
screen appears, the calibration is not
performed. Calibration error screen
・ When the sensor head position is not within
the detection range
・ When the 1st position is not determined and the 2nd position is
determined
・ When the values between the target and detection values does
not match
・ When the results of the span adjustment is more than twice the
factory default value
(Increase the Multiplier before performing calibration.)
• When performing calibration, reset the following parameters:
・ HIGH setting value ( "(3) HIGH setting value screen" (page 3-5))
・ LOW setting value ( "(4) LOW setting value screen" (page 3-6))
・ Self-timing level ( "04. Self-timing Level" (page 4-32))
Reference To reset the span adjustment to the factory default state, select "Default"
in "30. Calibration function setting".
"30. Calibration Function Setting" (page 4-64)
4-68 GT-M-E
4-6 Calculation Mode
The calculation mode is a part of the function setting mode and is available only when
using one or more expansion units. Select the applications beforehand to easily
display values that require calculation, such as the degree of flatness or average.
"4-2 Function Setting Mode" (page 4-3)
* The available calculation methods vary depending on the number of expansion units added.
Note When the calculation function is set, if the number of
expansion units is changed the next time the power is turned
on, the error display as shown to the right appears. Number of expansion units
Reset the error by following the procedure below. error
• Turn off the power, and change the number of units
back to the previous number of units, then turn on the power again.
• Reset the calculation function from the main unit.
GT-M-E 4-69
4-6 Calculation Mode
Selecting an application
MODE
OFF
4 Maximum
Minimum
Setting Various Functions
Average
Reference
difference MODE
Press and hold for 2 sec.
Twist
Warpage
Thickness
4-70 GT-M-E
4-6 Calculation Mode
• On the main screen, press and hold the [MODE] button for at least two
seconds. Enter function setting mode, then select "00. CAL".
"Entering function setting mode" (page 4-4)
[MODE] button
4
Use one of the following methods to exit calculation mode and display the main screen.
• After the calculation mode settings are complete, press and hold the [MODE]
button for at least two seconds.
• After the calculation mode settings are complete, press the right arrow key
several times until the setting completion display appears, then press the
right arrow key once more.
GT-M-E 4-71
4-6 Calculation Mode
W Functions that can be set in function setting mode when the calculation function is used
The following restrictions are applied in the function setting mode when the
calculation function is used.
"4-2 Function Setting Mode" (page 4-3)
4 Sensor amplifier
Main unit
Functions that can be set in the function setting mode
"02. Response time", "00. Calculation function"
Expansion unit All functions except "02. Response time" *
Setting Various Functions
Reference All the setting modes other than the function setting can be set.
4-72 GT-M-E
4-6 Calculation Mode
GT-M-E 4-73
4-6 Calculation Mode
Sensor head When the R.V. value = -FFFF When the R.V. value = FFFF
GT-H10/H10L/A10 10.150
-0.050
GT-H22/H22L/A22 22.150
4 * The upper limit value/lower limit value varies depending on the settings (increase/
decrease direction, multiplier, calibration function).
Setting Various Functions
4-74 GT-M-E
4-6 Calculation Mode
If "C0. Calculation function OFF" is selected, the detection values are displayed for
each amplifier without using a calculation function, and each result is displayed
according to the settings.
W Amplifier display
Main screens that can be displayed are as follows:
Sensor
amplifier
P.V. value
display
Calculated
value
R.V. value
display
HIGH
setting
LOW
setting
Preset value
setting 4
display display display display
Main unit X - X X X X
W Output
Outputs are as follows:
Sensor amplifier Output
Main unit Result is output based on the settings for the main unit.
Expansion unit Result is output based on the settings for each expansion unit.
GT-M-E 4-75
4-6 Calculation Mode
C1. Maximum
When "C1. Maximum" is selected, the maximum value detected from the main unit
and expansion units is used as the calculated value.
A
B
W Amplifier display
Main screens that can be displayed are as follows:
Calculated HIGH LOW Preset value
Sensor P.V. value R.V. value
value setting setting setting
amplifier display display
display display display display
Main unit X*1 X*2 X*3 X X X
Expansion unit - - X*3 - - X
*1 Displays the largest detection value (criterion value) from the main unit and expansion
units according to the hold settings.
Note
If any of the detected values in the sensor amplifiers shows "FFFF", the
calculated value display of the main unit will be "FFFF".
"Values used in calculation" (page 4-74)
*2 The R.V. value of the maximum value detected from the main unit and
expansion units is displayed (without holding the value even when the hold
function is used).
*3 The detected value by the sensor head connected to each amplifier is
displayed as the R.V. value (without holding the value even if the hold function
is used).
W Output
Outputs are as follows:
Sensor amplifier Output
The maximum value detected with the main unit and expansion units
Main unit
is taken as the value and the result is output.
Expansion unit The same result as that of the main unit is output.
4-76 GT-M-E
4-6 Calculation Mode
C2. Minimum
When "C2. Minimum" is selected, the minimum value detected from the main unit and
expansion units used as the value.
A
B
*1 Displays the smallest detection value (criterion value) from the main unit and
expansion units according to the hold settings.
Note
If any of the detected values in the sensor amplifiers shows
"- FFFF", the calculated value display of the main unit will be
"- FFFF".
"Values used in calculation" (page 4-74)
*2 The R.V. value of the minimum value detected from the main unit and
expansion units is displayed (without holding the value even when the hold
function is used).
*3 The detected value by the sensor head connected to each amplifier is
displayed as the R.V. value (without holding the value even if the hold function
is used).
W Output
Outputs are as follows:
Sensor amplifier Output
The minimum value detected from the main unit and expansion units is
Main unit
taken as a value, and the result is output.
Expansion unit The same result as that of the main unit is output.
GT-M-E 4-77
4-6 Calculation Mode
When "C3. Degree of flatness" is selected, the maximum and minimum values
detected with the main unit and expansion units are extracted and the difference
between them (maximum value - minimum value) is used as a the value.
4 A
D
Setting Various Functions
W Amplifier display
Main screens that can be displayed are as follows:
P.V. value Calculated value R.V. value HIGH LOW Preset value
Sensor amplifier
display display display setting display setting display setting display
Main unit X*1 X*2 X*3 X X X
Expansion unit - - X*3 - - X
*1 Displays the result (criterion value) after calculating the maximum value minus
the minimum value from the main unit and expansion units according to the
hold settings.
Note If any of the detected values in the sensor amplifiers shows
"- FFFF" or "FFFF", the calculated value display of the main unit
will be "FFFF".
"Values used in calculation" (page 4-74)
*2 The R.V. value of the difference between the maximum value and minimum
value detected with the main unit and expansion units is displayed (without
holding the value even when the hold function is used).
*3 The detected value by the sensor head connected to each amplifier is displayed
as the R.V. value (without holding the value even if the hold function is used).
W Output
Outputs are as follows:
Sensor amplifier Output
The calculation result of the maximum value minus the minimum value
Main unit detected with the main unit and expansion units is taken as the value
and the result is output.
Expansion unit The same result as that of the main unit is output.
4-78 GT-M-E
4-6 Calculation Mode
C4. Average
When "C4. Average" is selected, the average detected value from the main unit and
expansion units is used as the value.
X2
X3
X4
X1
The formula for obtaining an average
value is shown below.
(X1 + X2 + X3 + ... Xn) / n
4
n: The total number of sensor amplifiers
that are connected.
*1 Performs calculations to average the detection values from the main unit and
expansion units, and then displays the calculation result (criterion value)
according to the hold settings.
Note
If any of the detected values in the sensor amplifiers shows "- FFFF",
the calculated value display of the main unit will be "- FFFF".
If any of the detected values in the sensor amplifiers shows "FFFF",
the calculated value display of the main unit will be "FFFF".
*2 Averages the values detected from the main unit and expansion units. The R.V.
value of the calculation result is displayed (without holding the value even
when the hold function is used).
*3 The detected value by the sensor head connected to each amplifier is displayed as the
R.V. value (without holding the value even if the hold function is used).
W Output
Outputs are as follows:
Sensor amplifier Output
Performs calculation to obtain an average value of those detected from
Main unit the main unit and expansion units. The resulting value of calculation is
taken as the value and the result is output.
Expansion unit The same result as that of the main unit is output.
GT-M-E 4-79
4-6 Calculation Mode
X3 (expansion unit 2)
X2 (expansion unit 1)
X1 (main unit)
The formula to obtain a reference
4 difference is shown below.
Expansion unit 1 = X2 - X1
Expansion unit 2 = X3 - X1
.
Setting Various Functions
.
.
W Amplifier display
Main screens that can be displayed are as follows:
Calculated HIGH LOW Preset value
Sensor P.V. value R.V. value
value setting setting setting
amplifier display display
display display display display
Main unit - - X*3 - - X
Expansion unit X*1 X*2 X*3 X X X
*2 The value displayed on the calculated value display is the same as that displayed
on the P.V. value display.
*3 The detected value by the sensor head connected to each amplifier is displayed
as the R.V. value (without holding the value even if the hold function is used).
4-80 GT-M-E
4-6 Calculation Mode
W Output
Outputs are as follows:
Sensor amplifier Output
When GO is output for all the expansion
GO output
units
The GO is not output for at least one of
Main unit HIGH output
the expansion units
Any of the sensor amplifiers is in the
HIGH output + LOW output
error status
Based on the respective detected values, results are individually output
Expansion unit
according to the setting contents.
GT-M-E 4-81
4-6 Calculation Mode
C6. Twist
If "C6. Twist" is selected, the degree of twist is calculated from the four detected
values and the calculation result is displayed as the value.
Note
"C6. Twist" can be selected only when three expansion units are
connected to a main unit (total of four-unit configuration).
X1 X2 X1 X2
W Amplifier display
Main screens that can be displayed are as follows:
P.V. value Calculated value R.V. value HIGH LOW Preset value
Sensor amplifier
display display display setting display setting display setting display
Main unit X*1 X*2 X*3 X X X
Expansion unit - - X*3 - - X
4-82 GT-M-E
4-6 Calculation Mode
GT-M-E 4-83
4-6 Calculation Mode
C7. Warpage
If "C7. Warpage" is selected, the degree of warpage is calculated from the three
detected values and the calculation result is displayed as the value.
Note
"C7. Warpage" can be set only when two expansion units are
connected to a main unit (total of three-unit configuration).
X2 Expansion unit
X2 Expansion unit
X3 Expansion unit
X1 Main unit
X1 X1
X3 X3
X2 X2
4-84 GT-M-E
4-6 Calculation Mode
W Amplifier display
Main screens that can be displayed are as follows:
Calculated HIGH LOW Preset value
Sensor P.V. value R.V. value
value setting setting setting
amplifier display display
display display display display
Main unit X*1 X*2 X*3 X X X
Expansion unit - - X*3 - - X
W Output
Outputs are as follows:
Sensor amplifier Output
Performs calculation to obtain a warpage value by using the values
Main unit detected from the main unit and expansion units. The resulting value
of calculation is taken as the value and the result is output.
Expansion unit The same result as that of the main unit is output.
GT-M-E 4-85
4-6 Calculation Mode
C8. Thickness
If "C8. Thickness" is selected, the thickness of the target is calculated from the
detected value obtained by placing the target between the two sensor heads. The
calculation result is displayed as the value.
Note
"C8. Thickness" can be set only when a single expansion unit is
connected to a main unit (total of two-unit configuration).
4 X1
Setting Various Functions
X2
W Amplifier display
Main screens that can be displayed are as follows:
Calculated HIGH LOW Preset value
Sensor P.V. value R.V. value
value setting setting setting
amplifier display display
display display display display
Main unit X*1 X*2 X*3 X X X
Expansion unit - - X*3 - - X
W Output
Outputs are as follows:
Sensor amplifier Output
Performs calculation to obtain a thickness by using detected values
Main unit from the main unit and expansion units. The resulting value of
calculation is taken as a criterion value and the result is output.
Expansion unit The same result as that of the main unit is output.
Main unit
Set the following preset values:
GT-M-E 4-87
4-6 Calculation Mode
MEMO
4
Setting Various Functions
4-88 GT-M-E
Specifications 5
GT-M-E 5-1
5-1 Specifications
Sensor Head
■ GT-H10/GT-H22/GT-H10L/GT-H22L
Item Specifications
5 Model
Detection range
GT-H10
10 mm
GT-H22
22 mm
GT-H10L
10 mm
GT-H22L
22 mm
Mechanical response (Ambient
Max. 10 Hz Max. 8 Hz Max. 7 Hz Max. 5 Hz
temperature 23°C )
Repeatability *1 3 µm 8 µm 3 µm 8 µm
When installed
1.0 N 1.5 N 0.3 N 0.35 N
down
Force *2 When installed
0.9 N 1.4 N 0.25 N 0.3 N
sideways
5 When installed up
Ambient
0.8 N 1.3 N 0.2 N 0.25 N
5-2 GT-M-E
5-1 Specifications
■ GT-A10/GT-A22
Item Specifications
Model GT-A10 GT-A22
Detection range 10 mm 22 mm
Repeatability *1 3 µm 8 µm
When installed
1.0 N 1.5 N
down
Force *2 When installed
0.9 N 1.4 N
sideways
When installed up 0.8 N 1.3 N
Ambient
Environmental 0°C to +55°C
temperature
resistance
Relative humidity 35 to 85% RH
Usable fluid Dry air
Usable pressure range 0.35 to 0.5 MPa 5
Withstanding pressure 1.0 MPa
No. of times for sliding 3 million times
Specifications
Protection rating*3 IP67
Main unit housing SUS430
Indicator Polyarylate
Dust boot NBR
Material
Contact SUS440C, SUS304
Cable PVC
Air cylinder Aluminum allow, polyacetal, brass-nickel plating
Weight *4 Approx. 150 g Approx. 180 g
GT-M-E 5-3
5-1 Specifications
Sensor Amplifier
Item Specifications
Mounting type*1 DIN rail mounting Panel mounting
Type Main/expansion Expansion Expansion
Main unit Main unit
unit*2 unit unit
NPN output GT-71A GT-72A GT-75A GT-76A
Model
PNP output GT-71AP GT-72AP GT-75AP GT-76AP
Power supply voltage*3 12 to 24 VDC, ripple (P-P), 10% max, Class 2
Display range -99.999 to 999.999
Display resolution 1 µm
Sampling rate 2000 times per second
Normal - Using 24 V: 1200 mW (50 mA) max.
5 Current consumption
Using 12 V: 1140 mW (95 mA) max.
Eco mode - Using 24 V: 840 mW (35 mA) max.
Using 12 V: 600 mW (50 mA) max.
Response time hSP (1.5 ms), 5 ms, 10 ms, 100 ms, 500 ms, 1 s, 5 s
Specifications
Control output NPN NPN open collector: 40 V, 50 mA max., residual voltage 1 V max.
(HIGH/GO/LOW) PNP PNP open collector: 30 V, 50 mA max., residual voltage 1 V max.
Non-voltage input (with contacts, without contacts), input time
Timing input
2 ms min.
Control input Preset input
Non-voltage input (with contacts, without contacts), input time
Bank input
20 ms min.
Reset input
Ambient
-10 to +55°C (No freezing)
temperature
Environmental Relative
35 to 85% RH (No condensation)
resistance humidity
10 to 55 Hz, compound amplitude 1.5 mm, 2 hours for each
Vibration
of XYZ axes
Main unit housing: polycarbonate, Key top: polyacetal,
Material
Front sheet: polycarbonate, Cable: PVC
Approx. 110 g
(Including the panel
Approx. 110 g
Weight mounting bracket, front
(Including the power cable)
protection cover, and power
cable)
*1 When using the DIN rail mounting type, be sure to mount the unit to a DIN rail
(mounted on a metal plate). When adding expansion units, make sure to use the
end unit (OP-26751).
*2 The maximum number of units is 10 (1 main unit and 9 expansion units). When
adding expansion units, set the output current for each unit to 20 mA or less.
*3 When adding expansion units, the voltage required is 24 VDC.
5-4 GT-M-E
5-1 Specifications
Status Table
Control Sensor
Amplifier Control input
output head
State Bank
Digital Criterion switching Criterion
Timing Preset indicator
display indicator * During key lock output
only
No. of
digits for
Normal the Yes
detection
result On *2
Over Criterion Criterion
Yes When all
result result
Detecting HIGH FFFF are OFF:
(P.V./R.V.) range
Over
Off 5
LOW -FFFF No
range
Specifications
Yes
All OFF
Criterion (All ON when
"------" Off No Off
standby N.C. is
selected)
Setting Criterion According According Criterion
During setting
display result to detection to detection result Flashes*3
HIGH/
Error HIGH/ Lights in
During error No LOW
display LOW on red
("Er.Chk" or output
"ErC" No*1 All OFF
When power is according to
(All ON when
Off Off detection) Off
turned on N.C. is
selected)
*1 "Er.Chk", "ErC", "Er.com", and "Er.CAL" are on ("ErH", "ErE", and "Er.Unit" are off)
*2 GO: green, HIGH/LOW: red
*3 GO: green, HIGH/LOW/All OFF (All ON when N.C. is selected): red
Reference When the power save function is set, the state changes to the following.
"21. Power Save Function" (Page 4-52)
Amplifier
State Sensor head
Digital display Criterion indicator
* Normal display or Normal display or
HALF Animation display normal operations normal operations
ALL Off Off Off
* Displays one segment at a time sequentially.
GT-M-E 5-5
5-1 Specifications
The following table shows the time from when the power is turned on until the sensor
outputs a signal.
Note that the times depend on the response times.
5 5000 ms Approx. 10 s
5-6 GT-M-E
5-2 I/O Diagram
I/O Diagram
■ GT-71A/GT-72A/GT-75A/GT-76A
Brown*1
12-24 VDC*2
protection circuit
*1 Main unit only (GT-71A, GT-75A). Sub units (GT-72A, GT-76A) do not have Brown/Blue
connections since they receive power directly from the Main unit.
5
*2 When connecting sub units, 24VDC MUST be supplied to the Brown wire.
*3 For details about external input, see "External Input Circuit Diagram" (Page 5-8).
Specifications
■ GT-71AP/GT-72AP/GT-75AP/GT-76AP
Brown*1
12-24 VDC*2
protection circuit
Blue*1 Load
0V
Pink, purple, pink/purple
Input circuit External input*3
*1 Main unit only (GT-71AP, GT-75AP). Sub units (GT-72AP, GT-76AP) do not have Brown/Blue
connections since they receive power directly from the Main unit.
*2 When connecting sub units, 24VDC MUST be supplied to the Brown wire.
*3 For details about external input, see "External Input Circuit Diagram" (Page 5-8).
GT-M-E 5-7
5-2 I/O Diagram
■ GT-71A/GT-72A/GT-75A/GT-76A
DC5V
Pink: Preset input
Purple: Bank A input, reset input switching
Pink/purple: Bank B input, timing input switching
Main circuit
Blue*1
■ GT-71AP/GT-72AP/GT-75AP/GT-76AP
12 to 24 VDC
Brown*1
Main circuit
5-8 GT-M-E
5-3 Timing Chart
Standard Detection
Hold released
Sample hold
P.V. value
Criterion output
5
Peak Hold, Bottom Hold, Peak-to-Peak
Specifications
Sampling period 1 Sampling period 2
P.V. value
Previous result Sampling period 1 result Sampling period 2 result
Criterion output
NG Hold
GT-M-E 5-9
5-3 Timing Chart
Preset Input
P.V. value
Criterion value/output before preset Criterion value/output after preset
Criterion output
Max. 20 ms
ON
Reset input OFF
Min. 20 ms Min. 20 ms
5 Bank Input
Specifications
P.V. value
Criterion output Bank 0 Bank 1 Bank 2 Bank 0 Bank 3
Max. 20 ms
ON
Bank A input OFF
Min. 20 ms
Bank B input ON
OFF
Min. 20 ms Min. 20 ms
Max. 8 ms
5-10 GT-M-E
5-3 Timing Chart
Reset Input
P.V.="-----"
P.V. value
Before resetting All are OFF After resetting
Criterion output
NG-H After the response time
Max. 20 ms Std
P-H has passed
b-H Set with timing
Reset input ON P-P
OFF
Min. 20 ms Min. 20 ms
5
Minimum Input Time
Specifications
Timing input
ON time 2ms
OFF time 2ms
ON OFF
ON time 20ms
OFF time 20ms
ON OFF
GT-M-E 5-11
5-4 Dimensions
Sensor Amplifier
■ GT-71A/GT-71AP
5
24
15
11.6
35°
x. 1
3 Ma
79
37
5
42
17.6
18
Specifications
■ GT-72A/GT-72AP
24
11.6
15
35°
x. 1
3 Ma
79
37
42
17.6
18
5-12 GT-M-E
5-4 Dimensions
■ GT-75A/GT-75AP/GT-76A/GT-76AP
44.7
16.3
12.5
51.8
6.8
5
1.6
2.3
12.4
Specifications
24
21
13.4
3.5
60
6.6 15 Unit: mm
GT-M-E 5-13
5-4 Dimensions
Sensor Head
■ GT-H10/H10L
30.2 to 41.7
26.4 to 37.9
3.8 14 48
22.4
6
(16.4)
φ2.5×2
(Twisted shield)
5
12
φ4
-0.006)
Specifications
φ8 h5(
+0
Unit: mm
Note
• If the end of the cable is bent repeatedly, the cable may break.
• GT-H10L has no dust boot.
■ GT-H22/H22L
40.3 to 64.8
36.5 to 61
3.8 19 72
22.4
6
(16.4)
φ2.5×2
(Twisted shield)
12
φ4
φ8 h5(-0.006 )
+0
Unit: mm
Note
• If the end of the cable is bent repeatedly, the cable may break.
• GT-H22L has no dust boot.
5-14 GT-M-E
5-4 Dimensions
■ GT-A10
30.6 to 41.7
26.8 to 37.9
3.8 14 48
22.4
φ8 h5 ( -0.006 )
(16.4)
+0
φ2.5×2
(Twisted shield)
12
5
6
φ4
12.1
18.15
Specifications
1.4
108.7
12
43.4
Unit: mm
GT-M-E 5-15
5-4 Dimensions
■ GT-A22
41.1 to 64.8
37.3 to 61
3.8 19 72
6
22.4
(16.4)
φ8 h5 ( -0.006 )
+0
φ2.5×2
(Twisted shield)
12
6
5
φ4
12.1
18.15
1.4
Specifications
136.5
12
55.7
Unit: mm
5-16 GT-M-E
5-4 Dimensions
45
(73)
X = 24 x
11.1
((No. of amplifier units)-1) + 21
X
Panel thickness: 1 to 6
Min. 63 mm
21(+0.4)
-0
45
Unit: mm
5
Specifications
Head Mounting Bracket A (OP-76874) Mounting Hole Diagram
027 )
+0. .005
( +0
G8
φ10
φ 0.025 A
0.5 to 1.0 *
5.5 to 11.3 *
A
Side to install nut
Unit: mm
*Processing accuracy : ±0.05
GT-M-E 5-17
5-4 Dimensions
MEMO
5
Specifications
5-18 GT-M-E
Appendix A
A
1 Troubleshooting....................................................... A-2
2 Error Messages....................................................... A-5
3 Replacing the Contact............................................. A-7
4 Replacing the Dust Boot ......................................... A-9
5 Index ..................................................................... A-13
GT-M-E A-1
1 Troubleshooting
A-2 GT-M-E
1 Troubleshooting
Criterion
output cannot
Is the unit properly wired? properly.
"5-2 I/O Diagram" (Page 5-7) A
Are the tolerances set Reset the tolerances
be output
correctly? "3-3 Setting the Tolerances" (Page 3-15)
properly.
Appendix
Is the hysteresis set Set it up correctly.
correctly? "14. Hysteresis" (Page 4-43)
The unit of Set it up correctly.
Is the displayed number of
display is not
digits set correctly? "13. Displayed Number of Digits" (Page 4-42)
0.001.
- Are expansion units
added? They are displayed
The 00.CAL Check the connections for the expansion units.
only on the main with an
parameters
expansion unit is added. "2-1 Mounting and Wiring the Sensor
do not
- Are the amplifiers Amplifier" (Page 2-2)
appear.
connected to each other
correctly?
Are the number of expansion
Calculations
units correct? Each of those
for twist,
calculations are available Connect the correct number of units.
warpage, and
only when there are 4, 3, or 2
thickness "4-6 Calculation Mode" (Page 4-69)
units.) Are the amplifiers
cannot be
connected to each other
set.
correctly?
When
calculating
the reference Are you attempting to set the During calculation, set the response time on the
distance, the response time on an main unit.
response expansion unit? "02. Response Time" (Page 4-29)
time cannot
be set.
GT-M-E A-3
1 Troubleshooting
cylinder head
is used, but
Adjust the air pressure to be between 0.35 and
the spindle Is the air pressure enough?
0.5 MPa.
does not
extend
completely.
A-4 GT-M-E
2 Error Messages
Appendix
The spindle sets off the alarm.
- Check whether the external timing
input is functioning properly.
(Press the [SET] button to restart.)
The delay time is too long. Change the
The timing level falls (or rises) in setting.(Press the [SET] button to restart.)
the set delay time. "06. User-specified Delay
Time" (Page 4-34)
- When using static hold for
internal timing, the value falls
below (rises above) the timing
level and the detection ends
Set the trigger level so that it does
before the detection becomes
not fall (or rise) the timing level
stable and the criterion value
until the measured value is
(P.V. value) is determined.
determined. Or increase the static
- When using the delay timer for
hold delay stability width so that
internal timing, the value falls
the measured value can be
below (rises above) the trigger
determined more easily.
level and the detection ends
"08. Static Hold Delay Stability
within the period since the value
Width" (Page 4-36)
rises above (falls below) the
timing level until the criterion
value (P.V. value) is determined
after the specified time.
After turning the power off, check
The amplifier units cannot whether the connections between
communicate with each other. the amplifiers are correct and
restart the power.
GT-M-E A-5
2 Error Messages
Reference
Even when "com.Loc" is
displayed, the external input can
be used for bank switching, reset
input, and preset input.
Check the calculation mode
setting again.
The calculation mode is used, so The function setting mode can be
the function setting mode cannot changed only when the
be changed. calculation mode setting on the
main unit is set to "C0.OFF" or
"C5.REF".
A-6 GT-M-E
3 Replacing the Contact
This section describes replacing the contact. Remove the sensor head before
performing this operation.
Spindle
Contact
Appendix
2 Use pliers or a similar tool to grasp the contact and remove it.
Pliers and similar tools cannot be used on the fluorocarbon contact (OP-
80228). Remove this contact with your fingers.
Note
• Cover the contact with a piece of cloth to cushion it while you grasp the
contact with the pliers. Pliers and similar tools cannot be used on the
fluorocarbon contact (OP-80228). Remove this contact with your fingers.
• Do not rotate the unit while performing these steps.
GT-M-E A-7
3 Replacing the Contact
3 As described in step 1, use the supplied key wrench to hold the spindle
securely.
Note
• For the roller contact or offset contact, secure the side of the
contact being used with a fixing nut. We also recommend
using a coating of adhesive to prevent wiggling.
• When using a roller contact, only apply force rotational in the
direction of movement and not along other axes.
• When installing the contact, never apply tightening torque
over 0.2 N•m. Use a tightening torque of 0.002 N•m or less for
the fluorocarbon contact (OP-80228) only.
A-8 GT-M-E
4 Replacing the Dust Boot
This section describes the replacement of the dust boot. Remove the sensor head
from the device before replacement.
Note
After replacing the dust boot, the protection rating IP67 cannot
be guaranteed.
1 Use the supplied key wrench to secure the spindle and pliers or similar
tool to remove the contact.
Pliers and similar tools cannot be used on the fluorocarbon contact (OP-
80228). Remove this contact with your fingers.
Spindle A
Appendix
Cover the contact with a cloth
Note
Cover the contact with a piece of cloth to cushion it while you
grasp the contact with the pliers. Pliers and similar tools
cannot be used on the fluorocarbon contact (OP-80228).
Remove this contact with your fingers.
Dust boot
GT-M-E A-9
4 Replacing the Dust Boot
A
Appendix
5 As shown in the drawing below, press on the end of the dust boot with
your fingers and pass the tip of the spindle through the dust boot.
A-10 GT-M-E
4 Replacing the Dust Boot
6 Thread the spindle completely through the end as shown in the drawing
below.
8 Appendix
As described in step 1, use the supplied key wrench to secure the spindle and
use pliers or similar tool to remove the contact.
GT-M-E A-11
4 Replacing the Dust Boot
9 Rotate the dust boot either up or down until the ridges on the dust boot
(which protrude slightly) line up completely straight.
A
If the dust boot is not straight, the spindle may apply
Appendix
A-12 GT-M-E
5 Index
This index provides a list of terms used in this document. It is arranged in alphabetic
order.
Appendix
Average ............................................ 4-79 Panel Cutout Dimensions ............ 5-17
Sensor Amplifier .......................... 5-12
Sensor Head ............................... 5-14
B Display Filter Function ...................... 4-61
Bank Function .................................. 3-23 Displayed Number of Digits ............. 4-42
Basic Setting Mode .......................... 4-37 Displaying the Detection Result ....... 3-21
C E
Calculation Function OFF ................. 4-75 Entering additional function setting
Calculation Mode ............................. 4-69 mode ................................................ 4-50
Calibration Function Setting ............. 4-64 Entering basic setting mode ............ 4-38
Calibration Setting Mode .................. 4-62 Entering calculation mode ............... 4-71
Checking the Package Contents ....... 1-7 Entering calibration setting mode .... 4-63
Sensor Amplifier ............................ 1-7 Entering function setting mode .......... 4-4
Sensor Head ................................. 1-8 Error Messages .................................. A-5
Communicating with External Device 1-6 Exiting additional function setting
Connecting the Sensor Head to mode ................................................ 4-50
the Amplifier Unit .............................. 2-15 Exiting basic setting mode ............... 4-38
DIN rail mounting type ................ 2-15 Exiting calculation mode .................. 4-71
Panel mounting type ................... 2-15 Exiting calibration setting mode ....... 4-63
Core Alarm ............................. 4-53, 4-54 Exiting function setting mode ............. 4-4
When "USEr" is selected .............. 4-55 External Input Circuit Diagram ........... 5-8
GT-M-E A-13
5 Index
A-14 GT-M-E
5 Index
Determining the workpiece height 1-2 Sensor Amplifier Wiring ...................... 2-8
Using the fixture for indirect DIN rail mounting type .................. 2-8
detection ....................................... 1-2 Panel mounting type ..................... 2-9
Setting the Tolerance Manually ........ 3-16
Setting the Tolerances ..................... 3-15
P Setting Up the Tolerances
P.V. value display screen ................... 3-3 Automatically .................................... 3-17
Parameters in Additional Function Setting Specifications ..................................... 5-2
Mode ................................................ 4-49 Sensor Amplifier ............................ 5-4
Parameters in Basic Setting Mode ... 4-37 Sensor Head ................................. 5-2
Parameters in Calibration Setting Status Table .................................. 5-5
Mode ................................................ 4-62 Standard detection ............................ 4-6
Power On Reset Time ......................... 5-6 When the external timing is used .. 4-6
Power Save Function ........................ 4-52 When the internal (self-) timing
Precautions When Using Calculation (delay timer) is used ..................... 4-8
Mode ................................................ 4-72 When the internal (self-) timing A
Functions that can be set in function (static hold) is used ..................... 4-12
setting mode when the calculation Static Hold Delay Stability ................ 4-35
Appendix
function is used ........................... 4-72 Static Hold Delay Stability Width ...... 4-36
Response time when the calculation Store Preset Value ............................ 4-51
function is used ........................... 4-72 Switching the display unit ................ 3-28
Preset Function ................................ 3-22
Preset value setting screen ................ 3-7
T
Thickness ......................................... 4-86
R Origin alignment when "C8. Thickness"
R.V. value display screen ................... 3-4 is selected ................................... 4-87
Reference Difference ....................... 4-80 Timing Chart ....................................... 5-9
Replacing the Contact ....................... A-7 Bank Input ................................... 5-10
Replacing the Dust Boot .................... A-9 Minimum Input Time .................... 5-11
Reset input ....................................... 4-46 NG Hold ........................................ 5-9
Reset input when using the hold Peak Hold, Bottom Hold,
function ........................................ 4-47 Peak-to-Peak ................................. 5-9
Reset input when using the hold Preset Input ................................. 5-10
function ............................................. 4-47 Reset Input .................................. 5-11
Response Time ................................ 4-29 Standard Detection ....................... 5-9
Timing chart in the NG hold detection
mode ................................................ 4-17
S Timing Type ..................................... 4-31
Self Timing Delay ............................. 4-33 Tolerance calibration ....................... 3-17
Self-timing Level ............................... 4-32 Troubleshooting ................................. A-2
GT-M-E A-15
5 Index
U
Useful Functions ............................... 3-22
User-specified Delay Time ............... 4-34
W
Warpage ........................................... 4-84
A
Appendix
A-16 GT-M-E
5 Index
MEMO
Appendix
GT-M-E A-17
5 Index
MEMO
A
Appendix
A-18 GT-M-E
5 Index
MEMO
Appendix
GT-M-E A-19
Revision History
Date of printing Version Revision contents
April 2006 Official release
E 1110-2
Copyright (c) 2007 KEYENCE CORPORATION. All rights reserved.
189005E 1041-5 96029E Printed in Japan