You are on page 1of 109

PRO 1000

SERVICE MANUAL
DINAMAP PRO 1000 SERVICE MANUAL
DINAMAP PRO 1000 Monitor

Service Manual
List of Effective Pages
Part No./Rev. Page No. Date of Latest Revision
2008072 All Original Oct. 2001)

U.S. Patent 5,170,795 U.S. Patent 4,349,034


U.S. Patent 5,052,397 U.S. Patent 4,360,029
U.S. Patent 4,754,761 U.S. Patent 4,501,280
U.S. Patent 4,638,810 U.S. Patent 4,546,775
U.S. Patent 4,543,962 U.S. Patent 5,518,000
U.S. Patent 5,704,362 Patents Pending
CAUTION: Federal (U.S.A.) law restricts this device to sale
by or on the order of a health care practitioner
The content of this document, including all figures and
drawings, is proprietary information of General Electric
Medical Systems Information Technologies, provided
solely for purposes of operation, maintenance or repair of
PRO 1000 Monitors. Dissemination for other purposes or
copying thereof without the prior written consent of General
Electric Medical Systems Information Technologies,
Tampa, Florida, is prohibited. Illustrations may show
design models; production units may incorporate changes.
 CRITIKON 2001 TAMPA, FL 33614
Printed in the U.S.A. All rights reserved.

United States
Critikon, L.L.C.
4502 Woodland Corporate Boulevard
Tampa, FL 33614

EU Representative
GE Medical Systems
Information Technologies GmbH
Munzinger Strasse,
Freiburg, Germany

ii
TABLE OF CONTENTS
SECTION 1 INTRODUCTION
1.1. Scope of Manual........................................................................................ 1-1
1.2. Manual Changes ....................................................................................... 1-2
1.3 Service Policy ............................................................................................ 1-2
1.3.1 Extended Warranties ................................................................................. 1-2
1.3.2 Assistance ................................................................................................. 1-2
1.3.3 Service ......................................................................................................1-3
1.3.4 Service Loaners......................................................................................... 1-4
1.3.5 Repair Parts .............................................................................................. 1-4
1.3.6 Replacement Accessories ......................................................................... 1-5
1.4 Product Description ................................................................................... 1-5
1.4.1 General Description................................................................................... 1-5
1.4.2 Storage Batteries....................................................................................... 1-6
Table 1-1 Specifications.......................................................................................... 1-7

SECTION 2. PRODUCT DESCRIPTION


2.1. Introduction ................................................................................................ 2-3
2.2. Product Configurations ..............................................................................2-3
2.3. Controls, Indicators, and Connectors......................................................... 2-3
2.3.1. PRO Monitor Rear Panel Connections ...................................................... 2-4
2.3.2. Front Panel Controls and Indicators........................................................... 2-5
2.4. Host Port Connector (rear panel) ............................................................... 2-7
2.4.1. Pin Assignments ........................................................................................ 2-7
2.5. Compatible Parts ....................................................................................... 2-8
2.6. Specifications............................................................................................. 2-9
2.6.1. Power Requirements ................................................................................. 2-9
2.6.2. Environmental ............................................................................................ 2-9
2.6.3. Mechanical.............................................................................................. 2-10
2.6.4. NIBP ....................................................................................................... 2-10
2.6.5. Temperature ........................................................................................... 2-10
2.6.6. SpO2....................................................................................................... 2-11
2.6.7. ECG ........................................................................................................ 2-12

SECTION 3. PRINCIPLES OF OPERATION


3.1. Introduction ............................................................................................... 3-3
3.2. Overall Principle Of Operation .................................................................. 3-3
3.2.1. Nellcor SPO2............................................................................................. 3-3
3.2.2. Cuff Blood Pressure (BP) and Pulse ......................................................... 3-3
3.2.3. Alaris Oral and Rectal Thermometry ......................................................... 3-4
3.2.4. ECG with Heart Rate and Respiration ....................................................... 3-4

iii
3.2.5. Host Communication Ports ........................................................................ 3-4
3.3. Functional Description .............................................................................. 3-5
3.3.1. PSU PWA .................................................................................................. 3-5
3.3.2. Mains Converter Module ........................................................................... 3-6
3.3.3. Main Board ................................................................................................ 3-6
3.3.4. Keyboard PWA .......................................................................................... 3-7
3.3.5. ECG PWA ................................................................................................. 3-8
3.3.6. Pneumatic Control ..................................................................................... 3-8
3.3.7. LCD Assembly........................................................................................... 3-9
3.3.8. Printer (Optional) .................................................................................... 3-10

List of Figures
3-1 General System Diagram .............................................................................. 3-12

SECTION 4. GENERAL MAINTENANCE


4.1. Introduction................................................................................................ 4-3
4.2. Configuring the PRO 1000 Monitor for the First Time................................ 4-3
4.2.1 Unpacking and Preparation for Installation................................................ 4-3
4.2.2 Set the Date and the Clock........................................................................ 4-5
4.2.3 Parameter Level Functional Testing .......................................................... 4-6
4.3. Periodic Maintenance ................................................................................ 4-7
4.3.1. As Required............................................................................................... 4-7
4.3.1.1 Integrity of Cuffs and Hoses .................................................................. 4-7
4.3.1.2 External DC Supply and Battery ............................................................ 4-7
4.3.1.3 Cleaning of Accessories ........................................................................ 4-7
4.3.1.4 Long Term Storage................................................................................ 4-8
4.3.2 Annual Procedures .................................................................................... 4-8
4.4. Care of Storage Batteries ..........................................................................4-9
4.4.1. Procedures for First Use............................................................................ 4-9
4.4.2 Battery Charging........................................................................................ 4-9
4.5 Safety Resistance Testing...................................................................... 4-12
4.6. Alarm Code Interpretation ...................................................................... 4-14
4.6.1. System Failures...................................................................................... 4-14
4.6.2. Hardware Errors ..................................................................................... 4-15
4.6.3. Parameter Failures ................................................................................. 4-15
4.6.3.1 ECG/RESP/TEMP Errors ................................................................... 4-15
4.6.3.2 NIBP Messages.................................................................................. 4-15
4.6.3.3 Temperature Messages...................................................................... 4-16
4.6.3.4 SpO2 Messages ................................................................................. 4-16
4.7. Service Mode Operation......................................................................... 4-16
4.7.1 SpO2 Tests ............................................................................................ 4-19
4.7.2 NIBP Tests ............................................................................................. 4-20
4.7.2.1 Leak Test............................................................................................ 4-21
4.7.2.2 NIBP Calibration Check ...................................................................... 4-23

iv
4.7.2.3 Pressure Recalibration ....................................................................... 4-24
4.7.2.4 Overpressure Test.............................................................................. 4-25
4.7.3 EKG Tests .............................................................................................. 4-27
4.7.4 Temp Tests ............................................................................................ 4-28
4.7.5 Recorder Tests ....................................................................................... 4-30
4.7.6 Battery Tests .......................................................................................... 4-31
4.7.7 Test Failsafe Logic ................................................................................. 4-32
4.7.8 Keypad LED Test ................................................................................... 4-33
4.7.9 Keypad Key Test .................................................................................... 4-33
4.7.10 Sound Test ............................................................................................. 4-33
4.7.11 Turn off the System ................................................................................ 4-33
4.8 Service Mode Exit................................................................................... 4-33

Chapter 4 Appendices
Test Record ......................................................................................... Appendix A
Monitor Configuration Log ................................................................... Appendix B

SECTION 5 ASSEMBLY DRAWINGS & ELECTRICAL SCHEMATICS


Assembly Drawings (Monitor Assembly & Disassembly)
Front Case 1 ......................................................................................................5-1/2
Front Case 2 ......................................................................................................5-3/4
Rear Case 1.......................................................................................................5-5/6
Rear Case 2.......................................................................................................5-7/8

Electrical Schematics
ECG Board – 315589........................................................................ 5-9 through 5-18
Main Board – 315592...................................................................... 5-19 through 5-42
Power Supply Board – 315593 ....................................................... 5-43 through 5-52
Keyboard ........................................................................................................ 5-53/54
Probe Warmer ................................................................................................ 5-55/56

v
SECTION 1 INTRODUCTION
Contents
1.1. Scope of Manual .............................................................................................. 1-1
1.2. Manual Changes..............................................................................................1-2
1.3 Service Policy ................................................................................................... 1-2
1.3.1 Extended Warranties................................................................................... 1-2
1.3.2 Assistance................................................................................................... 1-2
1.3.3 Service ........................................................................................................1-3
1.3.4 Service Loaners .......................................................................................... 1-4
1.3.5 Repair Parts ................................................................................................ 1-4
1.3.6 Replacement Accessories........................................................................... 1-5
1.4 Product Description........................................................................................... 1-5
1.4.1 General Description..................................................................................... 1-5
1.4.2 Storage Batteries......................................................................................... 1-6
Table 1-1 Specifications.......................................................................................... 1-7
This page intentionally left blank.
SECTION 1. INTRODUCTION

1.1
SCOPE OF MANUAL

This Service Manual provides service and parts repair information about the
DINAMAP PRO 1000 Monitor. This manual is intended for use by trained service
technicians who are familiar with electromechanical devices and digital and
analog circuit techniques.

WARNING
To reduce the risk of electric shock, do not remove
cover or back of any component. Refer servicing to
qualified service personnel.
CAUTION

Only qualified service-technicians should perform repairs to this


equipment.

Voltages dangerous to life exist in this unit. Take care when servicing
power supply and display assembly.

For information about operating the Monitor in a clinical environment, refer to the
separate Operation Manual.

This Service Manual consists of the following four sections:

Section 1 describes this volume and tells you how to use it. Information is
also provided about the physical and functional characteristics of the Monitor,
and how to get assistance in the event the unit fails to function properly.

Section 2 provides a general overview of the PRO 1000 including user


controls, external connections, and product/ parameter specifications.

Section 3 presents principles of operation for the Monitor, including an overall


system description and principles of operation at the component level.

Section 4 provides information about periodic and corrective maintenance of


the Monitor. Procedures include module performance tests and calibration
procedures. Information is provided to facilitate isolating faults to the
subassembly level.
Section 5 provides component information about the Monitor, including
disassembly and reassembly procedures, parts lists, and assembly drawings,
and electrical schematics.

1-1
1.2 MANUAL CHANGES

If, in the normal use of this manual, you notice errors, omissions, incorrect data,
or if you can suggest comments that may help improve this manual, please
complete the Publications Change Request form in the back of this manual.
Submit the form to:

General Electric Medical Systems Information Technologies


Technical Publications
4502 Woodland Corporate Boulevard
Tampa, Florida 33614

Changes to the Service Manual, either in response to user input or to reflect


continuing product improvements, are accomplished through reissue.

Changes occurring between reissues are addressed through Change


Information Sheets and replacement pages. If a Change Information Sheet does
not accompany your manual, the manual is correct as printed.

1.3
SERVICE POLICY

The warranty for this product is enclosed with the product in the shipper carton.
All repairs on products under warranty must be performed or approved by
Product Service personnel. Unauthorized repairs will void the warranty. Only
qualified electronics service personnel should repair products not covered by
warranty.

1.3.1
Extended Warranties

Extended warranties may be purchased on most products. Contact your Sales


Representative for details and pricing.

1.3.2
Assistance

If the product fails to function properly, or if assistance, service or spare parts


are required, contact Customer Support. Before contacting Customer Support,
it is helpful to attempt to duplicate the problem and to check all accessories to
ensure that they are not the cause of the problem. If you are unable to resolve
the problem after checking these items, contact General Electric Medical
Systems Information Technologies. Prior to calling, please be prepared to
provide:

1-2
• product name and model number
• a complete description of the problem

If the repair parts or service are necessary, you will also be asked to provide

• the product serial number


• the facility's complete name and address
• a purchase order number if the product is to need of repair or when you
order spare parts
• the facility's account number, if possible
• the 6-digit part number for spare or replacement parts

1.3.3
Service

If your product requires warranty, extended warranty or non-warranty repair


service, call Customer Support and a representative will assist you. Estimates
for non-warranty repairs are provided at no charge; however, the product must
be sent to the General Electric Medical Systems Service Center in order to
provide you with an estimate.

To facilitate prompt service in cases where the product has external chassis or
case damage, please advise the Customer Support representative when you
call.

The Customer Support representative will record all necessary information and
will provide you with a Return Merchandise Authorization Number (RMA). Prior
to returning any product for repair, you must have a RMA number. Contact
technical support at 1-877-274-8456
Monday through Friday, 8:00 a.m. to 7:00 p.m. EST, excluding holidays.

Packing Instructions
Follow these recommended packing instructions.

• Remove all hoses, cables, sensors, and power cords from the monitor
before packing.
• Pack only the accessories you are requested to return; place them in a
separate bag and insert the bag and the product inside the shipping carton.
• Use the original shipping carton and packing materials, if available.

If the original shipping carton is not available

• Place the product in a plastic bag and tie or tape the bag to prevent
loose particles or materials from entering openings such as hose ports.

1-3
• Use a sturdy corrugated container to ship the product; tape securely to
seal the container for shipping.
• Pack with 4 to 6 in. of padding on all sides of the product.

Insurance
Insurance is at the customer's discretion. The shipper must initiate claims for
damage to the product.

1.3.4
Service Loaners

A loaner unit is provided at no charge during the service life of the product
when we perform the repair service. Within 48 hours of your request, a loaner
will be shipped to your facility.

• General Electric Medical Systems will pay shipping charges for a loaner
sent to the customer for product repairs under the warranty.
• Shipping charges for a loaner sent to the customer for product repairs not
under warranty will be billed to the customer.
• The customer will pay shipping charges to return a loaner.

All loaners provided to customers must be returned within the specified time
stated on the loaner agreement or a rental fee will be incurred.

1.3.5
Repair Parts

Repair parts can be ordered from General Electric Medical Systems:

Via phone 1-877-274-8456, or

Via FAX 1-813-887-2430

Exchange replacement assemblies such as Circuit Board Assemblies also are


available; ask the Customer Support representative for details.

Please allow one working day for confirmation of your order. All orders must
include the following information.
• Facility's complete name, address, and phone number
• FAX number
• Your purchase order number
• Your account number

1.3.6
Replacement Accessories

1-4
Replacements such as hoses, sensors, etc. must be purchased from General
Electric Medical Systems at 1-877-274-8456. Please have the 4-digit or 6-digit
Reorder/Product Code of the item you wish to order, your purchase order and
account number available.

1.4
PRODUCT DESCRIPTION

The Monitor and storage batteries are described below. Refer to Table 1-1 for
specifications.

1.4.1
General Description

The DINAMAP PRO 1000 is designed for patient monitoring in acute care
settings such as critical care, emergency room, radiology, labor and delivery,
and operating room. It allows the clinician to view, record, and recall clinical
data derived from each parameter. This data includes heart rate, respiration
rate, oxygen saturation, noninvasive blood pressure, and temperature. Alarm
limit conditions are also detected.

The recorder provides numeric and waveform printouts of monitored data. Up


to 2 waveforms can be traced simultaneously. Each monitor can monitor one
patient at the bedside.

Patient sensor connections are made at the side of the unit, and network and
device connectors are at the rear.

Indicators for external DC operation (from AC mains), battery operation, and


battery charging are at the front of the unit.

At the time of publication, the available functioning parameters included the


following:

• NIBP
• Nellcor™ Pulse oximetry (SpO2)
• 3-lead ECG, with respirations
• 2-channel thermal recorder
• Alaris™ Oral and Rectal thermometry

The PRO 1000 Monitor series uses a TFT active-matrix-color liquid display.

The 10.4” diagonal display area contains 640 x 480 pixels and can display
262,144 colors simultaneously.

1-5
The LCD has the following specific characteristics. These are neither defects
nor malfunctions:

• The ambient temperature may affect the display condition of the LCD.
• The LCD uses replaceable cold cathode tubes for backlighting. Optical
characteristics, like luminance or uniformity will change during time.
• Uneven brightness and/or small spots may be noticed depending on
different display patterns.

Other DINAMAP PRO 1000 features include:

• The ability to uses industry standard accessories


• Remote alarm capability
• An intuitive graphical user interface, with a simple Select Knob that moves
the user through menus in a logical, and easy to understood format
• Five single-function keys for quick access to Alarm Silence, Record,
Freeze, NIBP Start/Stop, and STAT NIBP

1.4.2
Storage Batteries

The PRO 1000 Monitor operates from AC mains power, an external DC power
supply, or from the internal Nickel Metal Hydride storage battery. When
external DC power becomes available, the system rapidly switches from
battery power to external power.

1-6
Table 1-1. Specifications

Mechanical
Monitor 14.8 in (H) x 8.7 in (D) x 13.8 in (W)
37.0 cm (H) x 21.8 cm (D) x 34.4 cm (W)
Weight Less than 12 lb (9.5 kg)

Environmental*
Operating Temperature +41º F to +104º F (+5° C to +40° C)
Storage Temperature -40º F to +158º F (-20º C to +60º C)
Operating Humidity 5% to 95%, noncondensing
Storage Humidity 5% to 95%, noncondensing
Operating Atmospheric Pressure 700 hPa to 1060 hPa
Storage Atmospheric Pressure 500 hPa to 1060 hPa

Electrical
Power Supply
The PRO 1000 Monitor can be operated from AC power, external DC power, or the
rechargeable internal battery.
AC Input Voltage 120 - 240
AC Input Frequency 50 - 60 Hz
AC Input Power 60 - 120 Volt Amperes
AC Power Cable Detachable, 16-gauge, 10 ft (3 meters) long
DC Input Voltage 18-24 V (supplied from a source conforming to IEC 601-1)
DC Input Power 60 Watts (supplied from a source conforming to IEC 601-1)
Internal Battery 12 Volts, nickel-metal-hydride (NiMH)
Battery Life 120 minutes (± 10 minutes) using fully charged internal
battery, under specified load **
Charge time, The PRO Monitor typically charges the battery to within
internal charger 90% capacity within 3 hours.
Fuse (Battery) 10A 250V slow-blow

* The Monitor may not meet Performance Specifications (ANSI/AAMI SP10) if it


is stored or used out of environmental specification ranges.
** Monitor shall be capable of operating on battery power for 2 hours minimum
(NIBP @ 5 min., ECG/Resp. SpO2, temp, dual channel recording once every
20 minutes.

1-7
SECTION 2. PRODUCT DESCRIPTION

CONTENTS
2.1. Introduction ................................................................................................ 2-3
2.2. Product Configurations .............................................................................. 2-3
2.3. Controls, Indicators, and Connectors......................................................... 2-3
2.3.1. PRO Monitor Rear Panel Connections ...................................................... 2-4
2.3.2. Front Panel Controls and Indicators........................................................... 2-5
2.4. Host Port Connector (rear panel) ............................................................... 2-7
2.4.1. Pin Assignments ........................................................................................ 2-7
2.5. Compatible Parts ....................................................................................... 2-8
2.6. Specifications............................................................................................. 2-9
2.6.1. Power Requirements ................................................................................. 2-9
2.6.2. Environmental ............................................................................................ 2-9
2.6.3. Mechanical............................................................................................... 2-10
2.6.4. NIBP ........................................................................................................ 2-10
2.6.5. Temperature ............................................................................................ 2-10
2.6.6. SpO2 ........................................................................................................ 2-11
2.6.7. ECG .........................................................................................................2-12

2-1
This page intentionally left blank.

2-2
SECTION 2. PRODUCT DESCRIPTION

2.1. INTRODUCTION
DINAMAP PRO Monitors provide non-invasive
determination of systolic blood pressure, diastolic
blood pressure, mean arterial pressure, pulse rate, 3-
lead ECG, temperature, and oxygen saturation.
These portable AC and DC operated monitors are
primarily intended for use in hospital acute care
settings such as outpatient surgery, accident and
emergency, labor and delivery, GI/endoscopy, and
medical/surgical units.

2.2. PRODUCT
CONFIGURATIONS
Each PRO Monitor is supplied with an accessory
pack. The contents of the pack vary according to
model. Unpack the items carefully, and check them
against the checklists enclosed within the accessory
boxes. If an accessory is missing or if an item is in a
nonworking condition, contact Critikon immediately.
It is recommended that all the packaging be retained,
in case the PRO Monitor must be returned for service
in the future.

2.3. CONTROLS,
INDICATORS, AND
CONNECTORS
Descriptions of the items shown are listed on the
pages that follow. For symbol definitions, refer to
paragraph: 2.3.2 of this section.

2-3
2.3.1. PRO Monitor Rear Panel Connections

1. Serial Number/ Manufacturer labeling


2. Earth Ground (safety test compatible)
3. Battery fuse (10A 250V)
4. Mains input (Used to connect to AC power supply)
5. External DC Input: 18-24 VDC only.
6. Main speaker opening.
7. Socket to secure removable rear cover (see 10)
8. DB9 connection used for Host Communication.
9. DB15 used for Host Communication/ remote alarm.
10. Removable rear protective cover.

2-4
2.3.2. Front Panel Controls and Indicators

GO/STOP – Starts and stops any determination of noninvasive blood pressure.

AUTO-BP/STAT – Dual-function hardkey. Starts and stops auto BP determinations


by a single-press and gives you access to change the NIBP cycle time. Starts and
stops stat determinations with a double-press (5 minutes of continuous NIBP
cycles.)

OFF/ON – Turns Monitor off and on.

Silence – Temporarily silences alarms; acknowledges alarming crisis conditions.

Standby – Enters and exits standby mode.

Main – Closes the menu system and takes you back to the main screen.

Trend – Enters and exits trends (view patient trends data.) This hardkey can be
configured through config mode two ways: to view mini trends or to view full trends.
Freeze – Captures up to 16.8 seconds of waveforms on the screen (seconds vary
depending on the chosen sweep speed.)
Record – Prints with a single-press for a snapshot (timed recording) and a double-
press for a continuous recording of the chosen waveforms.

2-5
Optional Components
Note: Interconnected equipment must be installed by a qualified service person.
Symbols
CE Mark

External Communications Port Connector

Attention, consult accompanying documents

Type CF applied part

Battery in use

Canadian Standards Association

Storage temperature

External AC or DC power indicator


External DC power input
External AC power input

Keep away from heat

This way up

Keep dry

Fragile, handle with care

SN Serial number

REF Catalog number

Predictive temperature

Functional earth terminal (ground lug)

Serial Port 1
Serial Port 2

Ethernet Connector

2-6
2.4. HOST PORT CONNECTORS
(BENEATH REAR PANEL)

All host port signals are NON-ISOLATED and should be connected to equipment
conforming to IEC 601-1-1 ONLY. Where isolation of data communication is
required, the Critikon isolated level converter should be used. If external alarm
control is required, Critikon part number 487208 (Isolated Remote Alarm Cable
Assembly) should ALWAYS be used. Please refer to the Information Sheet included
with the isolated remote alarm cable for operational details.
Note: When using remote alarm, the PRO Monitor should be considered the
primary alarm source. The secondary alarm is used for secondary purposes
only.

2.4.1. DB15/ DB9 Connector Pin Assignments

Pin Function Pin Function


1 Ground 1 Ground
2 TX2_Inverted TTL Data 2 TX1 Inverted TTL Data
3 RX2_Inverted TTL Data 3 RX1
4 AUX5V (600mA max.) 4 TX2
5 AUX12V (250mA max.) 5 RX2
6 Serial Level Control (High=TTL Low=-RS-232) 6 +5V (600mA Max)
7 Ground 7 +12V (400mA Max)
8 Remote Alarm (open collector, 75mA Max Sink) 8 No Connection
9 No Connection 9 No Connection
10 No Connection
11 TX2_RS232
12 Port Enable Control <low=port 2> (when in use, DB9 4 & 5 disabled)
13 RX2_RS232
14 No connection
15 No connection

2-7
2.5. COMPATIBLE PARTS
The following parts are available from Customer Service.
Description of Compatible Part Code
SOFT-CUF, Cuff, Infant 2500
SOFT-CUF, Cuff, Child 2501
SOFT-CUF, Cuff, Small Adult 2502
SOFT-CUF, Cuff, Adult 2503
SOFT-CUF, Cuff, Large Adult 2504
SOFT-CUF, Cuff, Thigh 2505
SOFT-CUF, Cuff, Neonatal type 1 2521
SOFT-CUF, Cuff, Neonatal type 2 2422
SOFT-CUF, Cuff, Neonatal type 3 2523
SOFT-CUF, Cuff, Neonatal type 4 2524
SOFT-CUF, Cuff, Neonatal type 5 2525
DURA-CUF Cuff, Infant 2783
DURA-CUF Cuff, Child 2781
DURA-CUF Cuff, Small Adult 2779
DURA-CUF Cuff, Adult 2774
DURA-CUF Cuff, Large Adult 2791
DURA-CUF Cuff, Thigh 2796
DURA-CUF Cuff, Assortment cuff pack 2699
DURA-CUF Cuff, Child pack 2697
CLASSIC-CUF , Cuff, Infant 2618
CLASSIC-CUF, Cuff, Child 2613
CLASSIC-CUF, Cuff, Small Adult 2608
CLASSIC-CUF, Cuff, Adult 2603
CLASSIC-CUF, Cuff, Large Adult 2643
CLASSIC-CUF, Cuff, Thigh 2648
CLASSIC-CUF, Cuff, Neonatal type 1 2638
CLASSIC-CUF, Cuff, Neonatal type 2 2633
CLASSIC-CUF, Cuff, Neonatal type 3 2628
CLASSIC-CUF, Cuff, Neonatal type 4 2623
CLASSIC-CUF, Cuff, Neonatal type 5 2619
12 Foot (approx. 3.7 meters) Long Adult / Pediatric Hose 107365
24 Foot (approx. 7.3 meters) Long Adult / Pediatric Hose 107366
12 Foot (approx. 3.7 meters) Long Neonatal Hose 107368
12 Foot (approx. 3.7 meters) Long A/P Hose Quick Disconn. 107368
IVAC** Oral Temperature Probe 088012
IVAC** Rectal Temperature Probe 088013
IVAC** Temperature Probe Covers 088015
DINAMAP PRO Monitor Operation Manual 776995*
DINAMAP PRO Monitor Service Manual 777358*
Accessory Pole/Basket/Base 3215
Printer Paper (Box of 10) 089100
Power Cable 316579
NELLCOR*** SpO2 Extension Cable SCP10*
NELLCOR Finger Sensor DS100A
NIBP Calibration Kit 320246
* PRO Monitor unique parts
** IVAC is a trademark of ALARIS Medical Systems
*** NELLCOR is a trademark of Mallinckrodt, Inc.

2-8
2.6. SPECIFICATIONS

This product conforms with the essential requirements of the


Medical Device Directive. Accessories without the CE Mark are not
guaranteed to meet the Essential requirements of the Medical
0086 Device Directive.
The PRO Monitor is protected against vertically falling drops of
water and conforms to the IEC 529 standard at level of IPX1. No
IPX1
harmful effects will come of vertically falling drops of water making
contact with the monitor.

2.6.1. Power Requirements


MAINS Protection against electrical shock - Class 1

AC INPUT VOLTAGE 115 / 230 VAC, 50 / 60 Hz (nominal),


90 ~ 253 VAC, 47 ~ 63 Hz (range)
ALTERNATE SOURCES Protection against electrical shock – Class 1

DC INPUT VOLTAGE 24 VDC (nominal), 12-30 VDC from supplied power converter

EXTERNAL DC FUSE Internal, auto-resetting.

BATTERY 12 volt, 2.3 amp-hours. Protected by auto-resetting fuse.


Minimum operation time: 2 hours (5 minute auto cycle with adult
cuff at 25°C (77°F) with power save mode enabled) from full
charge. Time for full recharge: 1 hr 50 min from full discharge when
the Monitor is switched off and 8 hrs when Monitor is switched on.

2.6.2. Environmental
OPERATING TEMPERATURE + 5° C to + 40° C (+ 41° F to + 104° F)
OPERATING ATMOSPHERIC
700 to 1060 hectoPascal
PRESSURE RANGE
STORAGE TEMPERATURE – 20° C to + 50° C (– 4° F to + 122° F)
STORAGE / TRANSPORTATION
500 to 1060 hectoPascal
ATMOSPHERIC PRESSURE
HUMIDITY RANGE 0 % to 95 % non-condensing
Complies with IEC Publication 601-1-2 (April 1993) Medical
Electrical Equipment, Electromagnetic Compatibility
RADIO FREQUENCY
Requirements and Tests, and CISPR 11 (Group 1, Class A)
for radiated and conducted emissions.
The Monitor is protected against vertically falling drops of
water and conforms with the IEC 529 standard at level of
INGRESS OF LIQUIDS
IPX1. No harmful effects will come of vertically falling drops
of water making contact with the Monitor.

2-9
2.6.3. Mechanical
Height 9.8 in. (25.0 cm)
DIMENSIONS Width 9.8 in. (24.8 cm)
Depth 6.9 in. (17.5 cm)
WEIGHT including battery 7.8 lb (3.5 kg)
MOUNTINGS Self-supporting on rubber feet or pole mountable
PORTABILITY Carried by recessed handle or pole mounted
CLASSIFICATION Mode of Operation: Continuous Degree of Protection against
INFORMATION harmful ingress of water: Drip-proof IPX1

2.6.4. NIBP
Adult 0 mmHg to 290 mmHg
CUFF PRESSURE RANGE
Neonate 0 mmHg to 145 mmHg
DEFAULT TARGET: CUFF Adult 150 ± 15 mmHg
INFLATION Neonate 110 ± 15 mmHg
100 to 250 mmHg
Adult
TARGET CUFF INFLATION 5 mmHg increments
ADJUSTMENT RANGE 100 to 140 mmHg
Neonate
5 mmHg increments

BLOOD PRESSURE Adult 120 seconds maximum


DETERMINATION TIME Neonate 85 seconds maximum

Adult 30 – 200 BPM ±3%


PULSE RATE RANGE
Neonate 30 – 220 BPM ±3%
Adult 300 – 330 mmHg
OVERPRESSURE CUT-OFF
Neonate 150 – 165 mmHg
BLOOD PRESSURE Systolic MAP Diastolic
MEASUREMENT RANGES mmHg mmHg mmHg
Adult 30 - 290 20 – 260 10 - 220
Neonate 30 - 140 20 – 125 10 - 110
Meets AAMI/ANSI standard SP-10
BLOOD PRESSURE ACCURACY AAMI/ANSI standard: ± 5 mmHg mean error
Intra-arterial method: ± 8 mmHg standard deviation
PULSE RATE ACCURACY ± 3.5 percent

2.6.5. Temperature
SCALES Celsius Fahrenheit
Max 42.2 °Celsius 108.0° Fahrenheit
RANGE
Min 31.6°Celsius 88.9° Fahrenheit
± 0.2oF (when tested in a calibrated liquid
o
MONITOR MODE ACCURACY ± 0.1 C bath; meets ASTM E1112, Table 1, in
range specified)
PREDICTIVE MODE ACCURACY ± 0.6oC ± 1.0oF
DETERMINATION TIME Less than 60 seconds

2-10
2.6.6. SpO2
SpO2 RANGE AND adult/neonate: 70% to 100% ± 3.5 digits
ACCURACY adult/neonate: 0% to 69% ± (unspecified)
PULSE RATE RANGE AND
30 BPM - 250 BPM ± 3 BPM
ACCURACY
SATURATION PITCH Pitch changes with saturation
INDICATOR Volume selectable from 0 (off) to 9
WAVEFORMS Pulse plethysmograph waveform on LCD gain compensated
SENSOR CONNECT /
The monitor detects the attachment or disconnection of a sensor
DISCONNECT FROM
from the patient within 15 seconds
PATIENT
SENSOR CONNECT /
The monitor detects the attachment or disconnection of a sensor
DISCONNECT FROM
from the Monitor within 5 seconds
MONITOR
The monitor detects a pulse or enters a no signal state within 15
PULSE DETECTION
seconds of being attached to the patient
The monitor detects loss of pulse from patient and enters a no
LOSS OF PULSE
signal state within 10 seconds

NELLCOR SENSORS
ADULT ACCURACY (70% - 100%) ACCURACY
OXICLIQ-P pediatric sensor 2.5 digits
OXICLIQ-I infant sensor 2.5 digits
OXICLIQ-N neonatal/adult sensor 2.5 digits
OXICLIQ-A adult sensor 2.5 digits
OXIBAND pediatric/infant sensor 3.0 digits
OXIBAND adult/neonatal sensor 3.0 digits
DURA-Y ear clip 3.5 digits
REFLECTANCE sensor 3.5 digits
DURASENSOR adult 3.5 digits
PEDI-CHECK pediatric spot-check clip 3.5 digits
OXISENSOR II D-20 pediatric sensor 2.0 digits
OXISENSOR R-15 adult nasal sensor 3.5 digits
OXISENSOR II D-25 adult sensor 2.0 digits
OXISENSOR II N-25 neonatal/adult sensor 2.0 digits
OXISENSOR II I-20 infant sensor 2.0 digits
OXISENSOR II D-25L adult sensor, long cable 2.0 digits
When sensors are used on neonatal subjects as recommended,
Neonatal Accuracy the specified accuracy range is increased by ± 1 digit to account
NOTE: Refer to NELLCOR for the theoretical effect on oximeter measurements of fetal
sensor specifications hemoglobin in neonatal blood, e.g., N-25 accuracy on neonates
is ± 3, rather than ± 2.

2-11
2.6.7 ECG
Leads Available 3-lead configuration:
I, II, III, MCL1
QRS amplitude 0.2 to 5.0 mV
QRS duration range 15 to 200ms (does not reject 10 ms, 1mV QRS)
Heart rate accuracy 10 to 300 (adult) / 10 to 350 (neonate) beats/min ±3 beats/min or
3% of reading, whichever is greater
Heart rate resolution 1 beat/min
Bandwidth:
Display/Recorder 0.5 to 40 Hz
0.05 to 40 Hz
0.05 to 100 Hz
Standardizing voltage 1 mV marker
Common mode rejection 1 mV RTI or 10 mm peak-to-peak maximum displayed noise
allowed with 20 Vrms, 50-60 Hz input
Input impedance: > 2.5 MΩ @ 10 Hz
60 Hz tolerance Up to 10 mV (with artifact detector off)
Up to 300µV (at 1 mV QRS and artifact detector on)
Pacemaker detection/rejection
Pacer amplitude ± 2 mV to ±700 mV
Pacer width 0.1 ms to 2 ms
With under or overshoot of 2 mV, 70 ms duration
Pacer amplitude ± 2 mV to ± 700 mV
Pacer width 0.1 ms to 2 ms
Tall T wave rejection: 100%
Lead off sensing current: <0.1 µA DC signal leads
< 1 µA DC driven lead
Time to alarm: High heart rate < 10 s per AAMI EC13 – 1992
Low heart rate < 10 s per AAMI EC13 – 1992
Cardiac standstill < 10 s per AAMI EC13 – 1992
Tachycardia waveforms < 10 s per AAMI EC13 - 1992

2-12
SECTION 3. PRINCIPLES OF OPERATION

CONTENTS
3.1. Introduction............................................................................................... 3-3
3.2. Overall Principles Of Operation ................................................................ 3-3
3.2.1. Nellcor SPO2 ............................................................................................ 3-3
3.2.2. Cuff Blood Pressure (BP) and Pulse ......................................................... 3-3
3.2.3. Alaris Oral and Rectal Thermometry......................................................... 3-4
3.2.4. ECG with Heart Rate and Respriation....................................................... 3-4
3.2.5. Host Communication Ports ....................................................................... 3-4
3.3. Functional Description .............................................................................. 3-5
3.3.1. PSU PWA ................................................................................................. 3-5
3.3.2. Mains Converter Module ........................................................................... 3-6
3.3.3. Main Board................................................................................................ 3-6
3.3.4. Keyboard PWA ......................................................................................... 3-7
3.3.5. ECG PWA ................................................................................................. 3-8
3.3.6. Pneumatic Control .................................................................................... 3-8
3.3.7. LCD Assembly .......................................................................................... 3-9
3.3.8. Printer (Optional)..................................................................................... 3-10

LIST OF FIGURES
3-1 General System Diagram .......................................................................... 3-11/12

3-1
This page intentionally left blank.

3-2
SECTION 3. PRINCIPLES OF OPERATION

3.1 INTRODUCTION

This section provides overall theory of operation and functional description of the
DINAMAP PRO 1000 (hereinafter referred to as PRO monitor). The PRO monitor
has Blood Pressure (BP), Pulse, Temperature, SPO2, and ECG monitoring
capability. The printer module is optional.

3.2 OVERALL PRINCIPLES OF OPERATION

The following paragraphs provide a general system interface relationship. The


general block diagram is located in Figure 3-1.
The PRO monitor is a portable unit that receives input power from an external AC
source, external DC source, or internal rechargeable battery.
When the ON/OFF button is pressed, the Main Board is brought out of a sleep
mode and turns on the power regulators. The power regulators provide conditioned
power from one of the input power sources: AC Mains, External DC, or the internal
battery. The regulated power is routed to the Printed Wiring Assemblies (PWAs) via
the cable harnesses. Once the PRO monitor is energized, a self-test is performed.
The self-test automatically tests the main functions of the PRO monitor. Failure of
the self-test will set the PRO monitor into a fail-safe mode with an audio alarm.
Under normal operating condition, the PRO monitor is ready to monitor the patient
vital signs using five external attachments: two temperature probes (rectal and
oral), SPO2 sensor, ECG leads, and cuff. Interface with a central station or other
device is accomplished through the 9-pin host communication port or 15-pin
wireless communication port on the back of the PRO monitor.

3.2.1 Nellcor SPO2


When the SPO2 sensor is attached to the SPO2 connector and patient, it senses
the heart rate and oxygen saturation. These analog signals are routed to the SPO2
PWA. The analog signals are analyzed on the SPO2 PWA. The results are digitized
and sent to the Main Board via opto couplers. The couplers provide for patient
isolation as well as serial data interface. The Main Board routes the data to the
appropriate displays and/or printer.
A reset signal to the SPO2 PWA is also provided for power up sequencing.

3.2.2 Cuff Blood Pressure (BP) and Pulse


When the cuff and hose are attached to the PRO monitor and Non-Invasive Blood
Pressure (NIBP) determination is initiated, the pump inflates the cuff. Pressure
transducers PT1 and PT2 monitor pressure information. The pneumatic manifold

3-3
has two valves, which are used to deflate the cuff. Valve control is through the Main
Board. Determinations are made for the systolic BP, diastolic BP, pulse rate, and
Mean Arterial Pressure (MAP). The results are displayed on the front panel Liquid
Crystal Display (LCD).
If an over-inflation condition occurs, it is detected by PT2, resulting in assertion of
OVERPRESSURE. The OVERPRESSURE signal is routed to the PVM to release
the air pressure. The Main Board also generates an alarm condition with the
speaker sounding and a message in the LCD.

3.2.3 Alaris Oral and Rectal Thermometry


The PRO monitor has two temperature channels, one each for oral and rectal
determinations. When a TEMPERATURE probe is attached to the temperature
connector and patient, TEMP input is routed to the Main Board. This input
represents the temperature to be measured. The Main Board converts the TEMP
signal to a digital signal. During the conversion, the Main Board determines the
patient temperature using a predictive or monitor mode algorithm depending on
user setup. The patient temperature is distributed as a digital signal to the LCD
display or printer.
The PRO monitor has a probe check feature to help prevent erroneous temperature
readings stemming from using the wrong probe for determinations (i.e. oral probe
for rectal determination and vice versa.)

3.2.4 ECG with Heart Rate and Respiration


The ECG parameter provides electrocardiographic waveform in a 3-electrode
configuration. The 3-electrode configuration derives waveforms for leads I, II, or III.
It includes a waveform cascade feature and can display one waveform as the
primary lead.
Breath rate is calculated by measuring the thoracic impedance between two
electrodes. As the patient breathes, the movement of the chest changes the
measured impedance to produce the respiration rate.
3.2.5 Host Communication Ports
There are two Host Comm Ports provided on the back panel of the PRO monitor.
The DB9 connector provides +5V(600mA Max), +12V(400mA Max), and two
channels of TTL compatible communications. The DB15 connector provides
+5V(600mA Max), +12V(250mA Max), Remote Alarm Signal, and a TLL/RS-232
selectable communication channel.
Note: When the DB15 port is enabled, channel 2 of the DB9 port is disabled.
The Host Comm Ports are used to interface the PRO monitor with other electronic
devices (a central nurse’s station or remote alarm device.) Signals can be sent to
the PRO monitor to initiate blood pressure determinations and other functions.
Patient data can also be retrieved through this port. The DB9 connector should be
used with Critikon Adapters ONLY. The host port signals on the DB15 connector
are NON-ISOLATED and should be connected to equipment conforming to IEC

3-4
601-1-1 ONLY. Where isolation of data communication is required, the Critikon
isolated level converter should be used. If external alarm control is required,
Critikon part number 487208 (Isolated Remote Alarm Cable Assembly should
ALWAYS be used. Please refer to the Information Sheet included with the isolated
remote alarm cable for operational details.
Note: When using remote alarm, the PRO monitor should be considered the
primary alarm source. The secondary alarm is used for secondary purposes only.

3.3 FUNCTIONAL DESCRIPTION

The following paragraphs provide the functional interface relationship. The PRO
monitor contains a number of electrical & electro-mechanical assemblies. These
assemblies are:

• Power Supply Unit (PSU) PWA


• PSU Module
• Main Board
• Keyboard PWA
• ECG PWA
• Pneumatic control device
• Liquid Crystal Display (LCD) Assembly
• Printer PWA w/printer (optional)

3.3.1 PSU PWA


The PSU supplies regulated DC power to PRO monitor. The PSU PWA is designed
to operate from the output of the AC MAINS PSU module (+24VDC), EXTERNAL
DC (+24VDC to +28VDC) source, or from an internal NiMH rechargeable battery
(+12VDC). The PSU will automatically select the power source based on the
following priority:
• Valid EXTERNAL DC input = +16VDC (If greater than or equal to output of
Mains Converter)
• Valid AC MAINS input
• Valid NiMH battery
The PSU PWA converts the selected power source into the following main voltages:
• VRAW1 (14.4VDC)
• VRAW2 (14.4VDC)
• VBAT
The +12V printer supply voltage is down converted from VRAW1 and maintained by
a boost regulator to +12V when VRAW1 falls under 12V. ANA+ is regulated to
+14.4VDC from VRAW2 by a MAX668 step up controller. AUX +12V is down

3-5
converted from ANA+ using an LM340 regulator. ANA- is down converted from
VRAW2 to –14.4VDC using a LM2594 step down regulator.
VBAT is the battery voltage protected by a 500mA auto-reset fuse. It is also used to
power the failsafe alarm circuits on the Main Board.
The PSU PWA contains firmware that reports the charge status of the battery to the
secondary processor on the main board. The secondary processor will charge the
battery at the fastest rate allowable while keeping the PRO monitor power
consumption under 60W.
The fan control circuitry is located on the PSU PWA and is powered by VRAW1.
The circuitry is thermally controlled and will start the fan when the thermistor (TH1)
reaches approximately 50°C.
The host communication port control circuitry selects whether channel 2 is routed to
the Comms connector (DB-9) or the wireless connector (DB-15). If channel 2 is
routed to the wireless connector, it can be configured for RS-232 or inverted TTL
signals. Channel 1 is only available on the Comms connector as inverted TTL.

3.3.2 Mains Converter Module


The Mains converter module is an AC Mains to DC converter. The module receives
AC power from the mains input connector. When AC INPUT is applied to the
module, the module AC/DC Converter changes the AC INPUT supply via rectifier
circuit to a high voltage DC. The DC power is then routed through a high frequency
switching converter and regulated to 24 VDC. This supply is connected to the PSU
PWA for further regulation. The PSU only selects this source when DC Output is
greater than +16VDC AND greater than the EXTERNAL DC input voltage.

3.3.3 Main Board


The Main Board is configured with Flash ROM, EEPROM, RAM, 16-bit ADC,
Primary Processor, Secondary Processor, SPO2 Module, and Temperature
Module. The Primary Processor operates from a 4.9152 MHz crystal stepped up to
49.152 MHz. The Primary Processor services and controls the RAM, Flash ROM,
EEPROM, the physiological interface modular devices and display backlighting.
The Secondary Processor monitors the power supply circuit, controls the power-
on/off sequences, performs watchdog task on itself and the Primary Processor,
monitors signals within the NIBP module, and monitors board temperature. The
Secondary Processor monitors the power supply circuit and controls the battery
back up enable when no external sources are present and will shut down the unit
when the battery is exhausted. It will enable the battery charging circuit based on
the battery charge status, unit power consumption, and the availability of an
external power source.
The SPO2 processor monitors pulse oximetry signal and passes the processed
signals to the SPO2 interface chip. The interface chip takes the signals and derives
the oxygen saturation and heart rate data and converts them into serial data. The
serial data from the SPO2 interface chip are send across an isolation barrier
(optocouplers) and passed to the Primary Processor via a dual-channel UART.

3-6
The Temperature Module is a dual channel system. Two IVAC probes and the
calibration resistor chain are connected into a multiplexer that is controlled by a
single chip micro controller. The micro controller will then switch the output of the
multiplexer between the calibration chain and the two probes. The multiplexer
output is buffered, level shifted and amplified before being connected to a
MAX1241 12 bit serial DAC, which is read by the microcontroller. The
microcontroller will compute the resistance for each probe (and associated leads)
and transmit both resistance values to the Primary Processor in a 34-bit data
stream. A switch controls each probe; the Primary Processor will interpret the status
of the switches, and select the appropriate part of the data stream from the
temperature circuit.
The Random Access Memory (RAM) is comprised of a SRAM chip and two SDRAM
chips. The 512 Kbytes of battery-backed SRAM is provided to store trend data and
to provide space for working algorithms and is accessed on bits D[0:15] of the data
bus. The two 64 Mbit SDRAM chips are set up to form a 32 bit data bus on bits
D[31:0] that is used for running the program and working memory. This gives
16Mbytes of memory with an access time of less that 20ns. The program is loaded
(including the boot code) from the 16 bit FLASH Read Only Memory (ROM). The
Electronically Erasable Programmable Read Only Memory (EEPROM) is an 8 bit
chip that is used to store the calibration and other “setting” variables that have to be
maintained in the event of a complete power failure.
If a hardware or software error causes a malfunction, its watchdog will provide an
internal and external RESET(L) signal. The FAILSAFE controller will cause the
FAILSAFE(L) signal to go low. This signal passes to the Secondary Processor,
which will disable the Primary Processor’s power supplies, thus turning it off.
FAILSAFE(L) also passes to the PAL (NIBP control logic), which will dump the cuff
pressure. The system is left in a safe state but is remains on to enable the
Secondary Alarm to stay active. The Primary Processor monitors the activity of the
secondary via its handshaking communications. If the Secondary fails, the Primary
can assert the FAILSAFE line by overriding the FAILSAFE controller. The
Secondary Alarm is a hardwired alarm that will sound in the event of a FAILSAFE
condition. Pressing the OFF-key can immediately reset this alarm although it will
eventually time-out after about 10 minutes.

3.3.4 ECG PWA


The ECG PWA accepts signals from a 3-electrode cable for processing. The 3-
electrode cable provides a single lead configuration with Lead I, Lead II, or Lead III
available. The cables specified by Critikon are shielded and provide 1k-Ohm series
(safety) resistors internal to the cable that are part of the current limiting defibrillator
protection circuitry. Gas surge arrestors on the PWA provide lead-to lead
defibrillator protection. In addition, a passive R/C network located on this PWA
provides the first stage of high frequency filtering for EMC and ESU interference
rejection. Two leads are selected for ECG measurement by a multiplexer (LS0, LS1
signal controlled) and passed to a differential amplifier. A second multiplexor
selects the third electrode (the one not sent to the differential amplifier) and drives

3-7
the signal with an amplified and inverted version of the common mode voltage of
the two measuring electrodes. This feedback action cancels most of the common
mode signal applied to the differential amplifier. The output signal from the
differential amplifier is then routed to bandpass filter and pacemaker detection
circuit.
The ECG PWA uses the pacemaker detection circuit to prevent pacemaker signals
from interfering with heart rate measurements. The ECG signals are sent through a
bandpass filter designed to pass pacemaker pulses in preference to ECG signals.
The filter output is applied to a comparator that asserts an output signal when the
input signal exceeds its positive or negative threshold. This output signal is used by
the controller to blank the ECG signal channel and alert the host to the presence of
a pacemaker pulse. The filtered ECG signal is routed to the A/D converter for
transfer to the Main Board.
The respiration circuit uses the ECG electrodes to measure respiration rate. This is
achieved by applying an excitation current (61.5 kHz, well outside the bandwidth of
normal ECG signals) generated by a square wave switch onto two selected
electrodes. The measured voltage drop is filtered, the baseline component
removed, and amplified. The analog voltage representing the impedance (resp
value) is routed to A/D converter for transfer to the Main Board.
The ECG PWA provides isolated power to its circuitry using an isolation
transformer. A transformer driver drives the transformer primary at a frequency of
about 350 kHz. The voltage of the transformer secondary is full-wave rectified using
two Schottky barrier diodes. The isolated voltage is filtered by capacitors and
regulated by a +5V regulator. Isolated ground is obtained from the center tap of the
transformer. The data transferred from the A/D converter to the host is isolated
using optocouplers.

3.3.5 Keyboard PWA


The Keyboard PWA provides access to the basic functions of the PRO monitor. The
buttons that control each function are integrated with its status LED to form a touch
pad front panel. LEDs indicate the status of those functions by illuminating green
when active and yellow when inactive. The exception is the “SILENCE” button,
which is red when active. The function LEDs are driven by latches on the Main
Board. The battery LED is continuously yellow when the unit is running on battery
and flashes yellow when the battery is charging. The AC LED is green when an
external power source is present. The keyboard is connected to the Main Board via
a 36 way board-to-board connector.

3.3.6 Pneumatic Control


The pneumatic functional block includes the control signal decode logic, the valve
driver circuitry, the pump driver circuitry; pump current measurement circuit, and a
safety interlock circuit.
There are two transducers on board, PT1 and PT2. PT1 is used for main readings
while PT2 confirms readings and is used to derive overpressure signals. The

3-8
following signals are multiplexed into a 16-bit SAR A/D converter via an 8:1 channel
DG408:
• PT1A - the output of the measurement pressure sensor after
amplification and filtering by means of a passive 1KHz low pass anti-
alias filter.
• PT1B - the output of the measurement pressure sensor after
amplification and filtering by means of a passive 20Hz low pass anti-
alias filter.
• PT2 - the output of the confirmation/over-pressure sensor after
amplification.
• TH REF - the voltage that the amplified PT2 has to attain before the
safety circuit cuts in
• PT1 REF - the reference voltage that provides an offset voltage for the
PT1 amplifier
• PT2 REF - the reference voltage that provides an offset voltage for the
PT2 amplifier
• PUMPC - the pump current
• VBAT- 1/11 ratio of VBAT voltage

The 16-bit value out of the ADC is available on the data bus at D[15:0].
Control signals for the board are derived via four different sources: direct control
from outputs of the processor, controls signal derived from processor address write
commands (which are stored in an addressable latch), signals derived from the
watchdog timer, and signals generated by the overpressure functional block. The
four valve control signals and the pump control signal are derived from the write
address and stored in an addressable latch. Latch values are cleared by application
of system RESET generated by the processor. Each latch signal is individually
gated in a programmable logic device (16V8) with the fail safe input signal
(watchdog timer) and the overpressure latch output to ensure pressure is removed
from the patient cuff should either overpressure or processor failure condition occur.
A cross-coupled latch for overpressure is included in the programmable logic
device. It is set by the occurrence of an overpressure condition existing for a period
greater than 500 milliseconds. When this condition occurs, Filter_OVP-0 transitions
low setting the internal latch. The latch output state is indicated by the
Latched_OVP signal. The latch can only be cleared by the PNEURESET input.

3.3.7 LCD Assembly


The PRO monitor uses TFT (thin film transistor) active matrix color liquid display.
The 10.4” diagonal display contains 640 x 480 pixels and can display 262,144
colors simultaneously. The display is backlit by cold-cathode fluorescent lamps.
Note: the backlight power inverter is mounted separately from the other PCB’s.
The LCD is driven from the Primary Processor via buffers (HCT244) on a dedicated
LCD driver port:

3-9
Signal Name
Clk Clock
Vsync Vertical Sync
Hsync Horizontal Sync
R[0:3] Red bits (0:3)
G[0:3] Green bits (0:3)
B[0:3] Blue bits (0:3)

The display module has a 31-way control signal connector and a 4-way backlight
driving connector.

3.3.8 Printer (Optional)


The PRO monitor uses a thermal graphics printer. The printer requires a 5V supply
for its logic circuitry and 12V (nominal) for the motor. The power and data lines are
connected to the Main Board by a 40-way cable. The data lines are connected to
the SCC3 port on the Primary Processor.
The printer has a built-in sensor to monitor the printer paper level. When the printer
is out of paper, it sends a PAPER OUT signal to the Secondary Processor.

3-10
SECTION 4. GENERAL MAINTENANCE
Contents
4.1. Introduction ......................................................................................................4-3
4.2. Configuring the PRO 1000 Monitor for the First Time ...................................... 4-3
4.2.1 Unpacking and Preparation for Installation.................................................. 4-3
4.2.2 Set the Date and the Clock ......................................................................... 4-5
4.2.3 Parameter Level Functional Testing............................................................ 4-6
4.3. Periodic Maintenance ...................................................................................... 4-7
4.3.1. As Required ............................................................................................... 4-7
4.3.1.1 Integrity of Cuffs and Hoses .................................................................. 4-7
4.3.1.2 External DC Supply and Battery ............................................................ 4-7
4.3.1.3 Cleaning of Accessories ........................................................................ 4-7
4.3.1.4 Long Term Storage................................................................................ 4-8
4.3.2 Annual Procedures...................................................................................... 4-8
4.4. Care of Storage Batteries .............................................................................. 4-9
4.4.1. Procedures for First Use ............................................................................ 4-9
4.4.2 Battery Charging ......................................................................................... 4-9
4.5 Safety Resistance Testing ............................................................................. 4-12
4.6. Alarm Code Interpretation .......................................................................... 4-14
4.6.1. System Failures....................................................................................... 4-14
4.6.2. Hardware Errors ...................................................................................... 4-15
4.6.3. Parameter Failures.................................................................................. 4-15
4.6.3.1 ECG/RESP Errors .............................................................................. 4-15
4.6.3.2 NIBP Messages.................................................................................. 4-15
4.6.3.3 Temperature Messages...................................................................... 4-16
4.6.3.4 SpO2 Messages ................................................................................. 4-16
4.7. Service Mode Operation............................................................................. 4-16
4.7.1 SpO2 Tests .............................................................................................. 4-19
4.7.2 NIBP Tests ............................................................................................... 4-20
4.7.2.1 Leak Test............................................................................................ 4-21
4.7.2.2 NIBP Calibration Check ...................................................................... 4-23
4.7.2.3 Pressure Recalibration ....................................................................... 4-24
4.7.2.4 Overpressure Test.............................................................................. 4-25
4.7.3 ECG Tests................................................................................................ 4-27
4.7.4 Temp Tests .............................................................................................. 4-28
4.7.5 Recorder Tests......................................................................................... 4-30
4.7.6 Battery Tests ............................................................................................ 4-31
4.7.7 Test Failsafe Logic ................................................................................... 4-32
4.7.8 Keypad LED Test ..................................................................................... 4-33
4.7.9 Keypad Key Test ...................................................................................... 4-33
4.7.10 Sound Test ............................................................................................. 4-33
4.7.11 Turn off the System ................................................................................ 4-33
4.8 Service Mode Exit .......................................................................................... 4-33
Appendix A - Test Record
Appendix B – Monitor Configuration Log

4-1
This page intentionally left blank.

4-2
SECTION 4. GENERAL MAINTENANCE

4.1 INTRODUCTION

This section contains general Monitor service procedures, including alarm code
interpretation, service mode operation, and periodic maintenance and battery care.
Refer to Section 5 for disassembly and reassembly procedures and related
component service information.

4.2 Configuring the PRO 1000 Monitor for the First Time

4.2.1 Unpacking and Preparation for Installation

1. Unpack and identify the contents of all shipping materials.


2. Remove the PRO 1000 monitor.
3. Unpack the AC cord but do not plug the monitor in at this time.
4. Turn the monitor for access to the Host Comms Cover.
5. Use a Phillips-head screwdriver to remove the single screw that secures the
Host Comms cover.
6. The Battery fuse and the Fuse Holder are not connected at time of shipment.
Locate and remove the fuse and fuse holder from the protective plastic bag.
7. Identify the Battery Fuse holder located within the Host Comms well, near the
lower left side.
8. Insert the Battery Fuse into the Battery Fuse holder
9. Press the Battery Fuse Holder into the Battery Fuse mount using thumb
pressure until it is securely snapped in place.
10. Replace the Host Comms cover; refasten the Phillips screw. Tighten using
hand-tools only.
11. Plug the AC cord into the AC Mains input at the back of the monitor.
12. Plug the AC cord into a Hospital Grounded AC receptacle. A green LED will
illuminate on the front of the monitor indicating that an AC source is available.
Prior to usage it is necessary to charge the monitor for 12 hours. This
charge calibrates the battery charging circuitry with the charge status of
the battery.

4-3
4-4
4.2.2 Set the Date and the Clock

The DINAMAP® PRO 1000 monitor uses a rotor knob to navigate through the
menu systems. Rotating the rotor moves the arrow cursor, and pressing the rotor
makes the selection.

1. Power on the PRO monitor using the OFF/ON key


2. Choose no when the monitor prompts to admit a new patient
3. Press or turn the rotor to access the main menu.
4. Turn the rotor and press to select other system settings option.
5. Turn the rotor and press to select go to config mode option.
6. Turn the rotor and press yes to continue to Configuration Mode.
7. The PRO Monitor will now prompt for the Configuration Mode password.

8. Turn the rotor and press to select after each numeral of the password is
selected.
9. The password to enter the configuration mode is 2-5-0-8.
10. Following the password entry, turn the rotor and press to select done.
11. In the process of entering the Configuration Mode, the PRO 1000 monitor will
reset. Successful entry into the Configuration Mode can be noted by the
words CONFIGURATION MODE in red capital letters at the top-center of the
screen.
12. Turn the rotor and press to select other system settings.
13. Turn the rotor and press to select config settings…
The Options available within this menu are:
• Select Date Format – Press the rotor to choose the order Month, Day, and
Year appears
• Select Time Format – Press the rotor to choose either a 12-hour clock or a
24-hour clock.
• Adjust Date & Time – Press the Select Knob to enter the correct Date and
Time. After setting the date and the time, make sure to press set new time
and date to save.
• Language – Press the rotor to choose a different language.
• Display Units – Choose yes to display all units of measure.
• Display Limits – Choose yes to display all alarm units.
• Reset all to Factory – CAUTION: resets all user-set options to the factory
defaults. Yes enables.
• Send all defaults – Choose this option sends all defaults to another
connected PRO 1000 monitor through the Host Communications ports.

4-5
4.2.3 Parameter Level Functional Testing

After the initial configuration is complete, perform functional testing of each of the
parameters. Using the accessories supplied with the PRO Monitor, initialize the
monitor in such a way that only one parameter is functioning at a time.

• Perform a blood pressure by connecting the


supplied hose and cuff together, then attaching
to the left side of the PRO Monitor. Press the
Start Key on the front to begin the NIBP cycle.

• Connect the supplied temperature probes to the


corresponding connections (see right). A
predictive temperature will begin once one
probe is removed from its’ holster. Replace the
probe after completion of the Temp cycle.

• The SpO2 sensor is an assembly consisting of


two parts: the DS-100A, and the extender cable
SCP-10. Connect the cables prior to attaching
to the monitor. A SpO2 reading will be displayed
within moments of attaching the sensor to either
a Nellcor simulator or to your finger.

• Connect the ECG lead connector to the ECG trunk cable prior to connecting
to the monitor. The simplest way to function test the ECG circuits is through
the usage of an ECG simulator.

1. Set your simulator to normal heart rate.


2. Set ECG amplitude to 1.5mV, BPM to 80.
3. Set respirations to 20 RPM, delta ohms to 1.0.
4. Verify that the ECG waveform is displayed.
5. Remove and reattach leads I, II, and III sequentially, and verify that
LEAD OFF is displayed.
6. From the ECG menu, select turn parameter off.

4-6
4.3 PERIODIC MAINTENANCE

4.3.1 As Required

Perform the following maintenance procedures as required.

4.3.1.1 Integrity of Hoses and Cuffs

When the pneumatic integrity of any NIBP cuff and hose is in doubt, replace the
cuff and hose, and discard the questionable accessories.

4.3.1.2 Cleaning of Monitor

CAUTION: Do not clean Monitor with isopropyl alcohol or other


solvents.

Wipe the exterior of the Monitor with a cloth slightly dampened with mild
detergent or normal hospital bactericides. Use dishwashing detergents such as
IVORY and JOY (registered trademarks of Procter & Gamble Corp.), or
PALMOLIVE (registered trademark of Colgate-Palmolive Corp.)

Do not immerse unit.

4.3.1.3 Cleaning of Accessories

Clean the adult cuffs supplied for use with the monitor by hand washing in
warm, soapy water. However, take care to avoid entry of water into the cuff and
hoses at any time. If water enters the cuff, dry the cuff by passing air through it.

The neonatal cuffs are for single patient use - discard if they become soiled.

Clean cuffs and hoses with a cloth slightly dampened with mild detergent.

Do not immerse hoses.

Do not immerse cuffs without prior application of cuff hose caps.

Clean SpO2 sensor surface before and after each patient use. Clean SpO2
sensor with a cloth slightly dampened with a mild detergent. Wipe SpO2 sensor
to ensure all detergent residue has been removed.

Follow manufacturer's instructions for cleaning ECG lead wires and cable.
Compatible cleaning and disinfecting solutions are:

4-7
Dishwashing detergents such as IVORY and JOY (registered trademarks of
Procter & Gamble Corp.), or PALMOLIVE (registered trademark of Colgate-
Palmolive Corp.)

Chlorine bleach disinfectant, 5.25%, 0.75 cup per gallon of water

CAUTION: Do not apply isopropyl alcohol to the Monitor - some parts


can become marred and cracked.

Isopropyl alcohol (for accessories only)

Cidex Formula 7 (registered trademark of Johnson & Johnson Medical


Products, Inc.) or pHisoHex (registered trademark of Winthrop-Breon
Laboratories)

Quaternary-based germicidal detergents like VESTAL INSURANCE (registered


trademark of the Vestal Corp.), HI-TOR PLUS (registered trademark of the
Huntington Corp.), or VIREX (registered trademark of S.C. Johnson & Son
Corp.)

For the above, follow manufacturers' recommendations for dilution rate and
use. These recommendations are not an endorsement of the manufacturers or
of the effectiveness of these materials for cleaning or disinfecting.

4.3.1.4 Long-Term Storage

If it becomes necessary to store the Monitor for an extended period of time,


remove all attached accessories. Attach the original packing inserts, and place
the monitor into the original shipping container.

Generally, long-term storage of a nickel-metal hydride battery in either a


charged or discharged condition has no permanent effect on capacity. Capacity
loss due to self-discharge is reversible, and nickel-metal hydride batteries can
recover to full capacity by proper recharging. For example, cycling through
repeated charge/discharge cycles can restore a full capacity of a nickel-metal
hydride battery that was stored at room temperature for up to one year.

Long-term storage at high temperatures can lead to deterioration of seals and


separators and should be avoided.

4.3.2 Annual Procedures

Perform the test procedures described in paragraph 4.7 every twelve months, or
whenever the accuracy of any reading is in doubt.

4-8
Note: An internal, 3.6V NiMh battery acts as an alarm backup and maintains the
nonvolatile RAM memory when the Monitor is off or away from AC mains.
A system alarm message will be generated if backup battery replacement
is required.

4.4 CARE OF STORAGE BATTERIES

The Monitor uses one nickel-metal-hydride (NiMH) storage battery. The battery can
be charged at any time without reducing the charging capacity.

4.4.1 Procedures For First Use

Follow these procedures to condition a new NiMH battery and optimize its
performance:

The internal battery will automatically charge when the AC power supply is in use.
When the battery is charged for the first time, the charger may indicate
prematurely that charging is incomplete. This is normal and can happen with all
rechargeable batteries when first charged.

4.4.2 Battery Charging

The Monitor charges the NiMH battery whenever the AC power supply is in use.
The Monitor automatically senses if the battery needs recharging. Battery
charging will continue as long as the Monitor is connected to the AC power
supply, even when the Monitor is turned off.

• Batteries should be charged before first use or after prolonged periods of


storage.

• The battery should be charged before use, as a charged battery loses


some charge when left in storage.

• The battery should be charged at room temperature (59° F - 86° F; 16° C -


30° C).

• It is normal for the battery to become warm during charging or after use.

• Batteries can be charged or topped-off at any time. It is not necessary to


wait until they are fully discharged.

• If the monitor is idle for extended periods, it should be fully charged once a
month to ensure optimum performance.

4-9
Table 4-1. Battery Alarms
Alarm Type Indication Probable Cause
BATT WRONG TYPE -
REMOVE Message appears in
Unapproved battery engaged
or INTERNAL BATT - alarm message field
WRONG TYPE -REMOVE
BATT CHECKING
Message appears in Noncommunicating battery
or INTERNAL BATT -
alarm message field engaged
CHECKING
Internal battery loses voltage or
INTERNAL BATT FAIL - Message appears in
communication or is not accepting
REPLACE NOW alarm message field
proper charge
Message appears in
< 00:30 BATTERY alarm message field 30 minutes remaining in battery life
and in SelectBox
Message appears in
< 00:10 BATTERY 10 minutes remaining in battery life
alarm message field
< 1 minute remaining in battery life.
Message appears in
SHUTTING DOWN Monitor may shut down anytime
alarm message field
after 45-60 seconds
Message appears in Upon loss of AC power, the
AC FAIL - < 00:30 alarm message field internal battery is engaged with
BATTERY and procedural less than 30 minutes of life (but
alarm sounds more than 10 minutes) remaining
CHECK COOLING FAN Message appears in Monitor's internal temperature is
the alarm message too high. Cooling fan may be
field blocked or inoperative
Message appears in Upon loss of AC power, the
AC FAIL - < 00:10 alarm message field internal battery is engaged with
BATTERY and crisis alarm less than 10 minutes of life
sounds remaining

4-10
Table 4-2. Battery Troubleshooting
Trouble Probable Cause Remedy
Battery inoperative Battery not fully Charge and discharge battery up
or does not last very charged. to three times for optimum
long. performance.
Battery in long-term Remove and reinstall battery so
storage or nonuse. connector is properly seated.
Battery charged for When charging battery for first
only a short period Improper procedure for time, charger may indicate
of time before charging battery for first prematurely that charging is
indicating full time use. completed. Discharge battery and
charge. repeat charging procedure.
Charge at basic room temperature
of 59° F (16° C) to 86° F (30° C).
Slowly bring battery to basic room
Charging battery in
Battery will not temperature before recharging.
unusually cold or hot
charge. Batteries cannot be fully charged
temperatures.
unless internal temperatures
between 57° F (15° C) and 109° F
(40° C).

4-11
4.5 SAFETY RESISTANCE TESTING

Using a safety analyzer (Dynatech Nevada Model 235A or equivalent),


check the ground resistance of the PRO 1000 Monitor; refer to the Rear
View graphic for locations of test points.

Earth-To-Secondary Continuity
Verify that the resistance between the AC Mains ground pole and the External
DC connector ground is less than or equal to 1Ω.

AC Mains Leakage – Normal Polarity


For the following tests, 260 VAC is applied at the Monitor’s AC Mains input in
normal polarity.

No Fault
Verify that the leakage from the Line pole to the Ground pole is less than or
equal to 500 µA.

Open Ground
Disconnect the Monitor’s ground lead from earth ground (for the duration of this
test only) and verify that the leakage from the Line pole to the Ground pole is
less than or equal to 500 µA.

Open Neutral
Open the Monitor’s neutral lead (for this test only) and verify that the leakage
from the Line pole to the Neutral pole is less than or equal to 500 µA.

AC Mains Leakage – Reverse Polarity


For the following tests, 260 VAC is applied at the Monitor’s AC Mains input in
reverse polarity (inputs to Line pole and Neutral pole reversed).

No Fault
Verify that the leakage from the Line pole to the Ground pole is less than or
equal to 500 µA.

Open Ground
Disconnect the Monitor’s ground lead from earth ground (for the duration of this
test only) and verify that the leakage from the Line pole to the Ground pole is
less than or equal to 500 µA.

Open Neutral
Open the Monitor’s Neutral lead (for the duration of this test only) and
verify that the leakage from the Line pole to the Neutral pole is less than
or equal to 500 µA.

4-12
Rear View of Monitor with safety connection exposed

4-13
4.6 ALARM CODE INTERPRETATION

Refer to Table 4-1 for information about procedural alarms that involve battery
operation. If any other alarms appear that are not listed in the paragraphs that follow,
record the error message and report the failure to Customer Support. Refer to the
Operation Manual for information about patient alarms and general procedural
alarms.

4.6.1 System Failures

When a system failure is encountered, the error code is displayed on the screen
for five seconds and the system enters failsafe mode. The error code is logged in
the history log.

General system error codes are listed below. If any other SY or similar code
appears, report it to Customer Support.

Error Code Explanation Possible Cause(s)


Power fail signal Circuit that drives POWERFAIL* signal is defective
SY-16
true time is too long
Software detected 1. Go to service mode to observe current values
SY-19 power supply out of for power supplies
limits 2. Digital to analog converter is defective
Checksum of code 1. Defective flash memory chip.
SY-20 in flash memory is 2. Error during programming of flash memory.
not valid
Real time clock 1. RTC chip running too slow or not at all.
SY-43
running too slow 2. System clock running too fast.
1. Noise getting into RTC chip crystal input.
Real time clock
SY-44 2. Defective or wrong crystal on clock chip.
funning too fast
3. System clock running too slow.

4-14
4.6.2 Hardware Errors

These error codes, which are common to all parameters, indicate some internal
self-check test of the hardware has failed, and service is required.

Error Code Description


8193 HW, Time base failure
8202 HW, Power supply, System
8222 HW, RAM test failure
8232 HW, ROM checksum failure
8242 HW, Isolation interface comm failure
8252 HW, Secondary processor not
compatible

4.6.3 Parameter Failures

4.6.3.1 ECG/ TEMP Errors

Fail Code Description


101 ECG board data rate error.
201 ECG board cmd queue overrun

4.6.3.2 NIBP Messages

Fail Code Description


110 Overpressure circuit failure
112 Overpressure watchdog error
120 FPT test failure
130 EEProm read failure
131 EEProm write failure
140 Transducer initialization failure
141 Calibration of a transducer channel's zero failed
142 Calibration of a transducer channel's span failed
150 Auto zero failure
151 Auto Zero. Verify failed.
160 PT1 reference failure
161 PT2 reference failure
162 OVC reference failure
170 Pump current failure
171 Pump current value out of range
180 Excessive leakage

4-15
Fail Code Description
190 Commands out of sequence
200 OVP setpoint not found
210 Pump stuck on during idle
220 Valve in illegal state
221 Pressure too high for too long

4.6.3.3 Temperature Messages

Fail Code Description


114 Temp data line out of sync with clock
115 Temp date frames out of sync

4.6.3.4 SpO2 Messages

In operate mode, the Fail Code is reported as described in section 4.6.3. In


service mode, the Service Mode Code (hex) is displayed on the screen as a
Parameter Fatal Error (hex).

Fail Code Description


125 Too many reset requests
126 Nellcor has posted a “serious” FE error
127 Nellcor FE requests power cycle
128 Nellcor over-current error
129 FE data OK – processing stalled

4.7 SERVICE MODE OPERATION

The Monitor service mode exercises the built-in diagnostic features of the Monitor
and the installed parameters. Access the service mode from a cold start by
proceeding as follows:

1. Momentarily press the on/off button at the front of the Monitor. Observe that a
beep sounds and that the power up screen displays.

2. Press the Rotor to answer no to the admit patient prompt, rotate the Rotor, and
observe the main menu is displayed on the left side of the screen.

3. Rotate the Rotor to select other system settings.

4-16
Select go to service mode. Rotate and press the knob again to answer yes at the
prompt. Note that the menu changes to what is shown below.

4. Observe that a row of numbers is displayed at the bottom of the screen. Turn
the rotor and press to select after each numeral is selected.

5. The password to enter service mode is: 2213. Following completion of the
password entry, turn the rotor and press to select DONE.

6. In the process of entering the Service Mode, the PRO 1000 Monitor will reset.
Successful Entry into Service Mode can be noted by the Service Menu title
display on the upper left side of the display.

NOTE: The service mode can also be entered directly from a cold start by
pressing and holding the following three keys until full power-up: OFF/ON, AUTO-
BP, ands GO/STOP. To make any changes to the Service Menu, the password
will still have to be entered: press the rotor to enter service password.

7. At this point the Service Mode main screen should be present in the main
display, as shown below. The service menu service parameters area displays a
list that corresponds to the number and type of parameters that have been
detected by the Monitor. If the service mode was entered directly (as described
in the NOTE above), enter service password appears above the service
parameters on the service menu. If recalibration of any component is required,
the password MUST be entered (as described in steps 5 and 6) before any
changes to calibration can be made.

4-17
Main Service Menu

For each parameter, there are one or more service screens that display operating
values and tests that are applicable to the parameter type. Refer to the following
paragraphs for information about each parameter. At the conclusion of the tests,
select go to service menu at the top of the screen to return to the Service Menu
main screen.

4-18
4.7.1 SpO2 Tests

1. Disconnect all sensor cables from the SpO2 Parameter, and ensure that the
SpO2 parameter is listed within the main Service menu.

2. From the Service Menu, Turn and press the rotor to select the SpO2 service
parameter. The SpO2 service menu should appear, with the SpO2 STATUS
displaying SENSOR Disconnected and the SPO2 MESSAGES display
empty. Example shown below.

SpO2 Service Menu

3. Insert the Nellcor SRC-2 pocket tester into the Parameter front panel SpO2
sensor socket (use the extender cable SCP-10), and press until fully seated in
the socket.

4-19
4. After a few seconds, the two red LEDs should light on the pocket tester. The
SpO2 service menu should soon display a saturation of 80±1 and also
displaying the heart rate as set on the SRC-2 pocket tester. Cycle through the
3 RATE settings and verify that the monitor responds accordingly.

5. Remove the SRC-2 pocket tester from the extender cable and verify that the
monitor reports a SENSOR Disconnected message in the STATUS box.

4.7.2 NIBP Tests

NIBP Service Menu

4-20
Perform the following tests to determine that the NIBP parameter is functioning
normally.

4.7.2.1 Leak Test

1. Using the calibration kit (part number 320-246), an adult cuff and air hose,
and a manometer, set up the equipment as shown in Figure 3-1. Connect
the hose to the NIBP Parameter. Make sure all fittings are tight, and that
the valve on the manual inflation bulb is fully closed.

Figure 4-1. NIBP Test Setup

2. From the Service Menu, Turn and press the rotor to select the NIBP service
parameter.

3. Turn and press the rotor to select pneumatic reset.

4. Turn and press the rotor to select valve close.

4-21
5. Turn and press the rotor to select start leak test. Observe that the Leak
Test Status message on the menu indicates Busy.

6. Observe that the pump begins inflating the system to 200 ~ 210 mmHg, at
which point the pump operation will cease. The Monitor will begin to
calculate system pressure loss rate.

7. After about 60 seconds, the pressure is released, and the menu should
display Leak Test Status Passed, and the Leak Test Results indication
should be a value less than 6. Service Error: None should continue to
display.

8. If the menu displays Leak Test Failed, continue to step 9.

9. Using the calibration kit (part number 320-246), an adult cuff and air hose,
and a manometer, set up the equipment as shown in Figure 4-2.

Figure 4-2 Leak Test Setup

10. Close the pressure release valve on the manometer inflation bulb and
slowly increase the pressure to 200-mmHg ±1 mmHg.

4-22
11. Verify the pressure indicated on the manometer remains within 5 mmHg of
200 mmHg for 60 seconds. If not, either the cuff or hose or both may be
defective. If the cuff and hose pass this test, repeat steps 1 through 7 to try
to isolate the leak. Repeat the leak test for all cuff and hose combinations
to be used with the Monitor.

4.7.2.2 NIBP Calibration Check

1. Using the calibration kit (part number 320-246), an adult cuff and air hose,
and a manometer, set up the equipment as shown in Figure 4-1. Connect
the hose to the NIBP Parameter. Make sure all fittings are tight, and that
the inflation bulb valve is closed tightly.

2. From the Service Menu, Turn and press the rotor to select the NIBP service
parameter.

3. Turn and press the rotor to select


pneumatic reset.

4. Turn and press the rotor to select


valve close.

5. Observe that both PT1 Pressure


and PT2 Pressure equal initial
values of zero mmHg (0 mmHg).

6. Connect the pneumatic hose to


the Monitor’s NIBP port.

7. Fold the adult cuff so the index line is aligned with the inner range mark on
the inside of the cuff. Make sure all fittings are tight, and that the valve on
the inflation bulb is closed tightly. If there is doubt about the integrity of the
system, perform the leak test (paragraph 4.7.2.1) before continuing.

8. Close the pressure release valve on the manometer inflation bulb and
manually pump up the pressure until the manometer indicates
approximately 220 mmHg.

4-23
9. Allow the pressure to stabilize for at least a minute. Then open the
pressure release valve on the manometer inflation bulb and carefully bleed
off pressure until the manometer indicates 200 mmHg.

10. Observe that the values of PT1 Pressure and PT2 Pressure on the menu
indicate within 1 mmHg of the pressure shown on the manometer.

11. Verify the system linearity by repeating steps 8 & 9 using manometer
readings of 250 mmHg, 150 mmHg, and 50 mmHg. Observe that the PT1
and PT2 Pressures are within 3 mmHg of manometer readings for each of
these pressure indications.

4.7.2.3 Pressure Recalibration

1. Always enter Service Mode with the password, as described in paragraph


4.7, before attempting to recalibrate equipment.

2. Using the calibration kit (part number 320-246), an adult cuff and air hose,
and a manometer, set up the equipment as shown in Figure 4-2. Do not
connect the pneumatic hose to the NIBP port yet.

3. From the Service Menu, Turn


and press the rotor to select the
NIBP service parameter.

4. Turn and press the rotor to


select pneumatic reset.

5. Turn and press the rotor to


select valve close.

7. Observe that both PT1


Pressure and PT2 Pressure
display initial values of 0 on the
menu.

8. Turn and press the rotor to select cal press zero. Observe that the
message Inflate System to 200 mmHg Then Hit ‘Cal Press 200’ is
displayed on menu.

9. Connect hose to NIBP Parameter.

4-24
10. Fold the adult cuff so the index line is aligned with the inner range mark on
the inside of the cuff. Make sure all fittings are tight, and that valve on
inflation bulb is closed tightly. If there is doubt about the integrity of the
system, perform the leak test (paragraph 4.7.2.1) before continuing.

11. Close the pressure release valve on the manometer inflation bulb and
manually pump up the pressure until the manometer indicates
approximately 220 mmHg.

12. Allow the pressure to stabilize for at least a minute. Then open the
pressure release valve on the manometer inflation bulb and carefully bleed
off pressure until the manometer indicates a little more than 200 mmHg.

13. Turn and press the rotor to select cal press 200, but do not press the knob
at first.

14. When the manometer indicates exactly 200 mmHg, press the Rotor.
Observe that system pressure is released, and the message: !!!!! CAL
INFO NOT SAVED!!!!! is displayed on menu.

15. Turn and press the rotor to select save cal info. If the system is operating
normally, the menu displays Service Error: None, and the calibration
setting is saved.

16. Repeat the calibration check procedure (paragraph 4.7.2.2) to confirm the
calibration setting.

4.7.2.4 Overpressure Tests

1. Using the calibration kit (part number 320-246), an adult cuff and air hose,
and a manometer, set up the equipment as shown in Figure 4-1. Connect
the hose to the NIBP Parameter. Make sure all fittings are tight, and that
valve on inflation bulb is closed tightly.

2. From the Service Menu, Turn and press the rotor to select the NIBP service
parameter.

3. Turn and press the rotor to select pneumatic reset.

4-25
4. Turn and press the rotor to select valve close.

5. Observe that the menu displays Overpressure Selected Adult. If not, turn
and press the rotor to select adult ovp select.

6. Turn and press the rotor to select inflate on. The pump should begin to
inflate the system.

7. Watch the pressure indication increase on the manometer, and observe


that the pump is shut down and the pressure is released when the
manometer indicates in the range of 300 to 330 mmHg. Observe that the
menu displays Service Error: None.

8. Turn and press the rotor to select pneumatic reset.

9. Turn and press the rotor to select valve close.

10. Turn and press the rotor to select neo ovp select. Observe that the menu
displays Overpressure Selected Neo.

11. Turn and press the rotor to select inflate on. The pump should begin to
inflate the system.

12. Watch the pressure indication increase on the manometer, and observe
that the pump is shut down and the pressure is released when the
manometer indicates in the range of 150 to 165 mmHg. Observe that the
menu displays Service Error: None.

13. If the overpressure test results in an “out of tolerance” condition, contact


Critikon Technical Support at 877-274-8456 for assistance.

4-26
4.7.3 ECG Tests

Connect the ECG leads to the ECG trunk cable prior to connection to the
monitor. The simplest way to function test the ECG circuitry is through the usage
of an ECG simulator with the monitor in normal monitoring mode.

1. Set your simulator to normal heart rate.

2. Set the simulator’s ECG amplitude to 1.5 mV, BPM to 80.

3. Set respirations to 20 RPM, delta ohms to 1.0.

4. Remove and reattach leads I, II, and III sequentially, and verify that LEAD OFF
is displayed in the main window.

4-27
4.7.4 TEMP Tests

1. Disconnect the all sensor cables from the Alaris temperature connections.

2. From the Service Menu, Turn and press the rotor to select the TEMP service
parameter. The TEMP service menu should appear as shown below.

Temperature Service Menu

4-28
3. Connect both the Oral and Rectal Temp probes to their respective
connections, reference the graphic below.

4. Perform the following test to verify the integrity of the temperature parameter:

• Verify top probe IN/OUT


• Verify top probe warmer ON. Observe that the temperature reading rises
to 105.0º F.
• Verify top probe warmer OFF. Observe that the temperature reading
drops slowly.
• Verify bottom probe IN/OUT
• Verify bottom probe warmer ON. Observe that the temperature reading
rises to 105.0º F
• Verify bottom probe warmer OFF. Observe the temperature reading drops
slowly.

4-29
4.7.5 RECORDER tests (if fitted)

1. Ensure that paper has been loaded into the


Recorder Parameter, and you are presently in the
Service Mode.

2. From the Service Menu, Turn and press the rotor to


select the RECORDER test option. Turn and press
the rotor to choose the 3 waveforms option. Verify
that all printouts are of even tone and all pixels are
present.

3. Select the 6.25mm/S option.

4. Allow for the paper to spool out a 12 inch printed


section.

5. Select Vertical Text test. Verify that the printed text


is legible and evenly spaced.

6. Select Horizontal Text test. Verify that the printed


text is legible and evenly spaced.

Horizontal Text Sample 6.25 mm/sec – 1 waveform chosen


print test

4-30
4.7.6 Battery Tests

From within the Service Menu, battery status information is displayed on the right
upper 1/4th of the display.

Battery/ Power Supply menu

1. Battery Health: the Monitor’s software approximates the true status of the
battery’s health. The value indicated is displayed as both a number (in
percentage) and as an icon on the bottom-left area of the display.

2. External Supply available: True indicates a source other than the internal
battery is providing power for the monitor, and a source to charge the
internal battery.

3. External Supply Sufficient to Charge: If the voltage from the external supply
is greater than that of the internal battery, the monitor will display the results
as TRUE. False will result if either the voltage is equal to or lower than the
power available from the internal battery.

4. Charge Type: Fast or Slow.

5. Battery Failed: Any result other than FALSE, indicates that the internal
battery has suffered a failure and should be investigated.

4-31
6. Charger Supply Enabled: Should always be TRUE as the monitor
consistently attempts to keep the battery at its’ fullest capacity. A FALSE
indicates the battery may be faulty or not installed, or the charge circuit may
have failed. Also, if no external source of power is available, the monitor will
register a FALSE result.

4.7.7 Test Failsafe Logic

1. From the Service Menu, turn the rotor to select test fail-safe logic. A dialogue
box will appear:
2.
CAUTION! This will cause the system to freeze for approx.
2 seconds then enter fail-safe mode. Continue?

Answer yes.

2. After two seconds, the system will freeze and an alarm will sound. Recycle
the system power using the on/off button. To return to the Service Mode,
repeat the procedures as described in section 4.7.

4-32
4.7.8 Keypad LED Test

1. From the Service Menu, rotate and press the rotor to select keypad LED test.
Observe that each of the keys on the PRO 1000 monitor-face, illuminate one
key at a time. With the exception of the ON/OFF key, observe whether any of
the keys fail to illuminate.

2. After all keys have been tested, press the rotor again to stop the test.

4.7.9 Keypad KEY Test

Verify that the keypad LEDs are illuminated (except Main, Trend, and Standby)

4.7.10 Sound Test

Verify that the Monitor produces tones of various pitches when this option is
selected.

4.7.11 Turn off system

Selection of this menu item will bring up a dialogue window requesting you to
confirm your decision:

CAUTION! This will turn the system off. Are you sure you want to do this?

Selecting yes will power off the monitor; no will return you to the Service Menu.

4.8 SERVICE MODE EXIT

To exit the service mode and power off the Monitor, locate and press the key
marked ON/OFF at the front of the Monitor.

4-33
This page intentionally left blank.

4-34
TEST RECORD (Appendix A) Model# Serial# .
Step Description Min Max Actual Pass Fail N/A
SpO2 Tests
4.7.1 Verify Pleth waveform
4.7.1 SpO2 reading at 100% Saturation 96 100
4.7.1 SpO2 reading at 90% Saturation 86 94
Internal Tests (Perform in Service Mode)
4.7.2.1 Leakage Test
4.7.2.2 UUT Pressure - 50 mmHg 46 54
4.7.2.2 UUT Pressure - 150 mmHg 145 155
4.7.2.2 UUT Pressure - 250 mmHg 244 255
4.7.2.4 Verify adult overpressure occurs between 300~350 mmHg
4.7.2.4 Verify neo overpressure occurs between 150~165 mmHg
4.7.4 Verify top probe IN/OUT
4.7.4 Verify top probe warmer ON
4.7.4 Verify top probe warmer OFF
4.7.4 Verify bottom probe IN/OUT
4.7.4 Verify bottom probe warmer ON
4.7.4 Verify bottom probe warmer OFF
4.7.5 Recorder Test
4.7.8 Keypad Test
4.7.10 Speaker Test
ECG Tests (Perform in Monitor Mode)
4.7.3 Verify Waveform
4.7.3 Verify BPM (@ 80 BPM) 76 84
4.7.3 Verify Paced Signal
4.7.3 Verify Lead-Off
RESP Tests (Perform in Monitor Mode)
Verify Waveform
Verify RESP (@ 20 RPM) 17 23
Verify RESP (@ 5 RPM) 2 8
Verify alarm sounds & displays below 6 RPM
BPM Tests (Perform in Monitor Mode)
Verify low rate alarm at 45 BPM
Heart Rate reading at 50 BPM (SpO2) 46 54
Heart Rate reading at 120 BPM (SpO2) 116 124
Heart Rate reading at 80 BPM (SpO2) 76 84
NIBP Tests (Function test in Monitor Mode)
Initial cuff inflation (Adult cuff) 161 195
Heart Rate reading @ 80 BPM (NIBP) 76 84
Inflate/ Deflate cycle time <120 seconds
Initial cuff inflation (Neonatal cuff) 94 151

4-35
Appendix A (continued)
Temperature Test (service or monitor mode - requires Alaris Temp Simulator)
Measured top Temp in ºF (98.6º nominal) 98.4 98.8
Measured top Temp in ºF (80.2º nominal) 79.9 80.5
Measured top Temp in ºF (107.8º nominal) 107.5 108.1
Measured bottom Temp in ºF (107.8º nominal) 107.5 108.1
Measured bottom Temp in ºF (80.2º nominal) 79.9 80.5
Measured bottom Temp in ºF (98.6º nominal) 98.4 98.8
Front Panel Key Tests (Perform in Service Mode)
4.7.9 Verify appropriate responses to keypresses
Battery System Test (Perform in Monitor Mode)
Verify AC Mains indicator
Remove AC, Verify uninterrupted battery operation
Verify Battery LED is lit
Safety Testing
External DC to GND Resistance (m ) 0 1000
SpO2 leakage (µA) 0 150
Temp Leakage (µA) 0 50 d
ECG Leakage (µA) 0 50
Normal no-fault leakage (µA) 0 500
Normal open-ground leakage (µA) 0 500
Normal open-neutral leakage (µA) 0 500
Reverse no-fault leakage (µA) 0 500
Reverse open-ground leakage (µA) 0 500
Reverse open-neutral leakage (µA) 0 500

Tested by: Date:

Signature: Facility:

4-36
MONITOR CONFIGURATION LOG - Appendix B
DINAMAP PRO 1000 Monitor Series
Note: Please refer to the PRO 1000 Pre-Service and Calibration Procedures for instructions.
Date: City:

Hospital:

Serial Numbers:

How to Enter Configuration Mode


1. Choose other system settings from the Main Menu.
2. Choose go to config mode. The message This will initiate the sequence for entering Configuration Mode. Do you want to do this? appears.
3. Choose Yes to enter configuration mode.
4. The message Please enter the Config Mode password appears. Enter the password. FACTORY SET CONFIG PASSWORD: 2508
5. Choose Done.
6. The system will restart in configuration mode. Press the SelectKnob to access the Configuration Menu.

How to Configure Default Tables


1. Choose admit patient from the Configuration Menu.
2. Choose Choose patient settings. Select the table you wish to configure (default 1 through default 6).
3. A popup window appears: All unsaved changes to the current default will be lost! Are you sure you want to do this? Choose Yes.
4. Choose Patient type and select either Adult, pediatric, or neonate.
5. Change all other available settings as desired.
6. To save your changes for the selected table go to other system settings, choose save default changes.
7. A popup window appears: Enter the name for this default. Rename or accept the default table and choose DONE. Your data will be saved.
8. Repeat steps 1 through 7 for configuring the remaining five default tables.

How to Exit Configuration Mode


1. Choose other system settings from the Configuration Menu.
2. Choose exit config mode.
3. A popup window appears: This will exit configuration mode. All unsaved changes will be lost. Are you sure you want to do this? Choose Yes.
4. The system will automatically restart in patient monitoring mode.

Warning: All monitoring will cease when entering configuration mode. Do not enter this mode if actively monitoring a patient.
Default Table Name Factory Default Adult Pediatric Neonate Adult2 Pediatric2 Neonate2
Adjust Alarms
Adjust alarm volume (0 to 5) 4
Choose autoset % 20%
HR/Pulse auto-set
Config settings
Alarm volume low range 1
Alarm silence time (in min) 2
Admit Patient
Choose Patient settings DEFAULT 1
Patient type adult
enter bed number
enter unit number
View Patient Trends
Choose graphs to print 0 chosen
Display as numbers
View vitals every NIBP
Mini trends…
View trends on main screen? yes
Display as numbers
View vitals every NIBP
Config Settings
Save previous patient data? yes
Trend key default mini trends
Setup HR/Pulse
Select source auto
Adjust QRS volume 0
Adjust limits auto-set
hi 150
lo 50
Advanced settings…
Limit alarms priority warning
Change color based on source? yes
Select HR/Pulse's color light green
(A=adult, P=pediatric, N=neonate. If not specified the factory default setting is the same for adult, pediatric, and neonate)
Setup ECG
Lead selection Lead II
Waveform size 1X
Pacer off? PACE OFF
Arrhythmia detection yes
Advanced settings…
Cardiac sweep speed 25.0 mm/s
Cascade ECG? no
Display filter 0.5 to 40 Hz
other alarm priorities
VTACH crisis
lead fail procedural
Replace electrodes procedural
Artifact message
Select EKG's color light green
Config settings…
Fixed ECG sweep speed? no
Setup NIBP
Auto BP Manual
setup custom series
1st BP Series q 10 mins
repeat x3
2nd BP Series q 30 mins
repeat x3
3rd BP Series q 60 mins
repeat x3
4th BP Series q 120 mins
repeat x3
Adjust limits auto-set
systolic hi A=200, P=150, N=100
lo A=80, P=70, N=40
diastolic hi A=120, P=90, N=60
lo A=30, P=30, N=20
mean hi A=140, P=100, N=70
lo A=40, P=40, N=30
Advanced settings…
Initial target pressure auto
(A=adult, P=pediatric, N=neonate. If not specified the factory default setting is the same for adult, pediatric, and neonate)
Setup NIBP (continued)
Limit alarms priority warning
other alarm priorities
No determination procedural
Overpressure procedural
Pump timeout procedural
Total timeout procedural
Level timeout procedural
Select NIBP's color purple
Setup SpO2
View Waveform? yes
Adjust limits hi 100 lo 90
Advanced settings…
View signal strength bar? yes
View SpO2 PR? no
Spot check enable yes
Cardiac sweep speed 25.0 mm/s
Limit alarms priority warning
other alarm priorities
Lost pulse procedural
Sensor disconnected procedural
Replace cable procedural
Select SpO2's color white
Setup RESP
Lead to analyze A-LII, P-LII, N - L1
View waveform? yes
Waveform size 1X
Adjust limits hi A=30, P=60, N=100
lo A=6, P=10, N=15
Advanced settings...
Resp sweep speed (mm/s) A/P=12.5, N=6.25
Cardiogenic filter auto
Limit alarms priority warning
other alarm priorities
Resp approaching warning
Lead Fail procedural
Saturation procedural
(A=adult, P=pediatric, N=neonate. If not specified the factory default setting is the same for adult, pediatric, and neonate)
Setup RESP (continued)
Artifact procedural
Select RESP's color procedural
Config settings…
Turn on RESP with ECG? no
Setup TEMP
Unit of Measure °F
Choose mode predictive
Advanced Settings…
other alarm priorities
Disconnected procedural
Two probes out procedural
Timed out procedural
Probes same type procedural
Check probes procedural
Select TEMP's color yellow
Config settings…
Allow °C units only? no
Setup RECORDER
print on alarm no
vitals summary on printout no
Auto printout of vitals summary OFF
setup continuous
Waveforms to record 1 chosen
setup timed
Chart speed 25.0 mm/s
Length of strip (in seconds) 8
Record key printout at bedside
Config settings…
setup continuous
Delayed memory (in seconds) 8
Length of strip (in seconds) 20
setup timed
Delayed memory (in seconds) 4
(A=adult, P=pediatric, N=neonate. If not specified the factory default setting is the same for adult, pediatric, and neonate)
Other System Settings
Always display battery icon? no
Advanced Settings…
Select color format full color
Adjust keyclick volume 2
Adjust system volume 8
Config settings…
Select date format mm/dd/yy
Select time format military
Language english
Display units? yes
Display limits? yes
Config HostComm
Unit address ""
IP address 2.0.0.0
Waveforms to send 2 chosen
Remote access OFF
Serial 1 setup
Startup mode ASCII cmd
Baud rate 9600
Serial 2 setup
Startup mode ASCII cmd
Baud rate 9600
(A=adult, P=pediatric, N=neonate. If not specified the factory default setting is the same for adult, pediatric, and neonate)

Notes
Columns that are grayed out are non-configurable.
SECTION 5 ASSEMBLY DRAWINGS & ELECTRICAL SCHEMATICS

Assembly Drawings (Monitor Assembly & Disassembly)


Front Case 1 ..................................................................................................... 5-1/2
Front Case 2 ..................................................................................................... 5-3/4
Rear Case 1 ...................................................................................................... 5-5/6
Rear Case 2 ...................................................................................................... 5-7/8

Electrical Schematics
ECG Board – 315589........................................................................ 5-9 through 5-18
Main Board – 315592..................................................................... 5-19 through 5-42
Power Supply Board – 315593....................................................... 5-43 through 5-52
Keyboard......................................................................................................... 5-53/54
Probe Warmer................................................................................................. 5-55/56
Front Case - 705110

Rotor Knob- 733152

3
4
1 2

Pneumatic hose connects to the


bottom terminal of NIBP socket

1 - Cable Assy, Defib, to Main Board - 316701


Foot, Polyurethane - 732175 2 - Cable Assy, ECG to Main Board - 316680
Fascia, Key Pad, (english) - 701502 3 - PVC Tubing
4 - Pneumatic Tubing (NIBP) - 740185 Assembly Drawing
5 - Nellcor SpO2 Connector - 608530 Front Case (1 of 2)
6 - Alaris Temperature Connectors - 608531
page 5-1/2
7 - Temp Probe Housing - 705108
Main Board - 315592
ECG Backplane
Front Case - 705110

Protective Lens - 705120

Rotor Knob - 733152 Display, TFT Color - 614235

Rotor Assy - 705112

Keyboard, PWA - 315594


Support Post - 735339

Main Board - 315592

ECG Cable
ECG Backplane
Assembly - 316680

Bumper, Polyurethane - 732175

Keyboard, PWA - 315594

Pneumatic Tubing (NIBP) - 740185

Assembly Drawing -
Front Case (2 of 2)
page 5-3/4
Note: The Circuit Boards depicted within this drawing are shown as unpopulated.

Rear Case - 705116

AC Cable Assembly - 316698

PSU Board - 315593

Pneumatic
Assembly - 320744

PSU Module -
320757
External DC Cable
Assembly - 316694 Vibration Dampers

Speaker Assembly - 320747

NiMH Battery - 633177


(replaced as a whole)

For NiMh battery removal/replacement:


The battery retention screws are recessed
between the cells on the rear case.
Fan, DC - 680452 Assembly Drawing
Rear Case (1 of 2)
page 5-5/6
DB9 Serial Connector

Rear Case - 705116

DB15 Host
Comms Connector

Velcro - 736249

Note 1

AC Cable
Assembly - 316698

Bed Hook - 710123

External DC Cable
Assembly - 316694
Bumper - 732175
Bumper - 732175
Battery Fuse/Holder

Bumper - 732175

Antenna Blanking Plugs - 705119

- Indicates a screw removal point.


Note 1: This particular screw is longer than all of the other screws.
It is critical to return this screw to the same slot; placement
Host Comms Cover - 705117
elsewhere can potentially cause damage to the front of the
Monitor, requiring case replacement.
Phillips Screw, #2 - 705132 Assembly Drawing
Rear Case (2 of 2)
Host Comms Cover removed
page 5-7/8
P5AI

R71
E_LA
R29 R30 R31 100K
100M 100M 100M R72
E_RA
100K
R73
E_LL
P5AI 100K

C27 C4 C5 C6
100N 100N 100N
PL1 100N
OPA4340
1 R51 R77 3 4 AG VREF
1 AG AG AG
2 499R 332K
3 1 2
4
CG5-90L 11 R47
CP2 C33 U20 P5AI
5
2 LA
220P C18 200K P5AI
6 100P
7
P5AI MAX399
AG AG
AG 4 14 R59
IN0A +VS
AG
OPA4340
5 10K0 OPA4340
IN1A R34
R53 R78 6 8 10 4
12 4 IN2A OUTA
AG 7 9 8
499R 332K 14 IN3A OUTB 33K2 9
ECG1
1
CG5-90L 13 1
13 A0 R33 11
11 IN0B 16
CP1 C34 12 A1 U21
2
RA C19 U20 11
IN1B
EN
2
33K2
220P 100P IN2B
10 15
P5AI IN3B GND
3 AG
AG AG -VS
AG
AG R46
OPA4340 U13
R52 R79 10 4 R44 R45 200K
8 200K 200K
1 499R 332K 9 AG
CG5-90L
11
CP3 C35
2
LL C20 U20
220P 100P
AG
AG
AG
AG

Maintain 0.060 in. clearance to ground plane in this area. P5AI

P5AI
MAX399 VREF
4 14 R60 OPA4340
IN0A +VS
5 10K0 4 5
IN1A R69
6 8 7
IN2A OUTA
7 9 6
IN3A OUTB 4K99 11
1
13
IN0B
A0
16
U20
12 A1 R35
IN1B 2
11 EN 33K2
IN2B AG
10 15
IN3B GND
3 R74
-VS

U14 301K
R10 C17
AG R38 17K8 4N70
1K00

LS1

LS0

Schematic - ECG Board


P/N 315589 (1 of 5)
page 5-9/10
P5AI
C24
C23 VREF P5AI 1N00 C66
R20 R37 P5AI VREF P5AI
25K5 1N00 61K9 P5AI
OPA4340 100N P5AI R96
OPA4340 3 4 AG C52 34K0 C25
5 4 1 MAX323 8
10N0 1N00
OPA4340
7 R84 R57 2 + OPA4340 12 4
R19 R21 6 11 2 C NO 1 R58 R97 R99 10 4 MAX7401 14
ECG1 12K4 41K2 R16 R98 OUT_ECG
25K5 100K
11 U22 7 IN
1K43 33K2 7K87
8 2
IN OUT
5 13
U22 - 9
20K5 69K8 11
C37 U3 4
C47 C51 11 8 4 U22
CLK +V
C36 10N0 AG
22N0 10N0 U22
10N0 VREF 1 7
AG AG C85 COM SHDN C38
1N00 6 3 10N0
OS GND AG
AG
R94 U29
AG 10K0 AG
AG C48 C49
100N 100N
AG
R22
DAC_ECG R95
100K BLANK_ECG-0
VREF 10K0 AG
AG
C83 C26
1N00 1U00

AG AG

CLK_ECG_LPF

P5AI P5AI
P5AI
P5AI P5AI
C72 R7 C71 R92
VREF P5AI
C21 20K0 R62 100K
VREF P5AI 100P 100N VREF
C46 R32 100N LM393 AG 10K0 OPA4340
AG
R75 330P 143K OPA4340 5 8 12 4
19K1 OPA4340 3 4 7 14
OUT_PD-0
5 4 1 6 13
7 R8 R43 2 4 11
C9 R12 R11 6
20K0 7K87
11 U17 U21
2K67 17K8 11 U21
100N U21
AG C50 R93
C41 47UF 100K AG
C40 AG 10N0 AG R70 16V
10N0 4K99
P5AI
AG AG
AG
AG
LM393
3 8
1
2
4
U17

C53 R6 AG
10N0 20K0

AG AG

Schematic - ECG Board


P/N 315589 (2 of 5)
page 5-11/12
P5AI P5AI
C75
VREF
100N
U31 11 OPA4350 AG
R104 3 CC + NC 4 R80 5 4
T1
7 R49 R50 C10 1
1:1 8
C3 R56
100R C77 1 CO NO 16
16K2 6
RA
100N 11
1K65 1K65 100N 3N30 499R
15 IN 13 C22 U23 VREF P5AI P5AI
GND
- 100P C16 C44
14 33P0 R17
MAX383 AG 10P0
OPA4350
750R AQW214S
AG 3 4
C1 R54
1 8
1
R39 4 5 2
A1 S1A
7
LA
AG 2
RLS1-0 K1 S1B 3N30 499R
CLK_RESP_MOD 11
100R 3 6
AG A2 S2A
U23 RLS2-0
4
K2 S2B
5
C2 R55
R109 LL
U30 499R
4K99 3N30
AG Place close to T1

C80 R106 C79 R107 R108 C78


Maintain 0.060 in. clearance to ground plane in this area. 220P 1M00 220P 1M00 1M00 220P Make GND connections short and fat.

AG AG AG AG AG AG

Maintain 0.060 in. clearance to ground plane in this area.

R40 CR1
BL_RESP
1K00 MMBD914
P5AI

R48 C8
OPA4350
200K 33N0
12 4
14
Place close to U23 C81 13
33P0 C29
11 P5AI
U23 AG AG
P5AI 1U00
AG
AG C39
R82
2K67 VREF P5AI
100N
OPA4340 AG
C82 10 4 OPA4340
U31 11 8 R36 R26 12 4
100P 6 CC + NC 5 R63 9 14
11
5K90 150K 13
OUT_RESP
R81 20K0 R105
8 CO NO 9 U24 11
2K67 20K0 U24
10 IN 13 R76
GND AG
C30 C76
- 191K 10N0
VREF P5AI 14 1U00
MAX383 R28 AG
R85 499K
OPA4350
33K2 AG Route CLOCK signals away from these parts and connections.
10 4
8
C54
9 AG
10N0 R41
11
10K0
U23 VREF
P5AI
AG
OPA4340
R100 5 4
DAC_RESP
7
CLK_RESP_DEMOD 34K0 6
11
U24
C84 C7 R101 AG
1N00 47UF 10K0
16V

AG AG

Schematic - ECG Board


P/N 315589 (3 of 5)
page 5-13/14
P5AI

1
VIN
U36 2
VOUT
MAX6002 1 9
GND OUT_ECG P1.0/ADC0/T2 DAC0 DAC_ECG
2 10
3 OUT_RESP P1.1/ADC1/T2EX DAC1 DAC_RESP
3
BL_RESP P1.2/ADC2
E_RA
4
P1.3/ADC3 P0.0/AD0
43 P5DI
11 44
AG E_LA P1.4/ADC4 P0.1/AD1
12 45
E_LL P1.5/ADC5/~SS P0.2/AD2
13 46
P5DI 14
P1.6/ADC6 P0.3/AD3
49
P1.7/ADC7 P0.4/AD4
P0.5/AD5
50 R86 R87 R88 R89 R90 R91
51 25K5 25K5 25K5 25K5 25K5 25K5
3 C11 P0.6/AD6
AG 7 52
VCC CREF P0.7/AD7
P5AI 8 28
P5AI P5AI
U37 2
100N VREF P2.0/A08/A16
29
LS0
RESET
AG P2.1/A09/A17 LS1
30
MAX810 P2.2/A10/A18
31
BLANK_ECG-0 MAX323 8 OPA4340
GND P2.3/A11/A19 + 3 4
36
1 P2.4/A12/A20 6 C NO 5 1
15 37
RESET P2.5/A13/A21 3 IN 2
38
P2.6/A14/A22 - AG 11
AG 39
X1 P2.7/A15/A23
16
U3 4 U24
P3.0/RXD D_FROM_HOST
32 17
XTAL1 P3.1/TXD D_TO_HOST AG
33 18
XTAL2 P3.2/~INT0 OUT_PD-0
C63 11.0592MHZ C64 P3.3/~INT1/MISO
19
AG
33P0 33P0 22
P3.4/T0
23
P3.5/T1/~CONVST
5 24
AG AG AVDD P3.6/~WR RLS1-0 Spare Parts
20 25
P5DI DVDD1 P3.7/~RD RLS2-0
34 26
DVDD2 SCLOCK
48 27
DVDD3 SDATA/MOSI
P5DI
6 ADUC812
C56 C45 C31 C28 AGND
10N0 10N0 10N0 21 40 J2
DGND1 ~EA R102
10N0 35 41 FLASH-0 1
DGND2 ~PSEN
47
DGND3
U1 ALE
42
1K00 2
AG AG AG
AG
AG P/D_NUM=MMT-102-01-T-SH

P5DI
Keep this trace short. AG
C58
P5DI
10N0 P5DI
P5DI 74HC393 P5DI 74HC390
74HC161 AG 14 11
VCC Q1 74HC00 P5DI
16
VCC QA
13 C57
3 14
P0 Q0
13 10
EN_ECG40 1
+
14
3 10N0 74HC390
4 P1 Q1 13 CLOCK-0 Q2 15 16
5
U33 12
2
- 74HC00 CLOCKA-0 QB
11
VCC QA
3
CLK_ECG_LPF
P2 Q2 7
6 11 12 9 12 14
P3 Q3 RESET Q3 U35 13
+ 11 12
CLOCKB-0 QC
10 AG 1
CLOCKA-0 QB
5
2 CLOCK -
RCO 15 P5DI 7 U34 8 7
1 RESET-0 GND Q4 AG 14 9 4 6
U35 RESET QD CLOCKB-0 QC
9 LOAD-0 VCC 16 P5DI 8 U25 2 7
AG 7 ENABLE_P C55 AG
AG GND RESET QD
10 ENABLE_T GND 8 10N0 74HC00
EN_ECG100 14
8
GND
U25
9
+ 8 AG
10
AG P5DI -
7
U35 P5DI AG
P5DI
C59 AG C60
74HC00
14 4
10N0 74HC393 10N0
6 +
-
5 14
VCC Q1
3 74HC164
7 AG AG
14 VCC QA 3
U35 1 4 4
CLOCK-0 Q2
1
U28 QB
5
CLK_RESP_DEMOD
AG A1 QC
2 5 2 6
RESET Q3 A2 QD
8 CLOCK QE 10
7
GND U34 Q4
6 9 RESET-0 QF 11
EN_RESPCLK QG 12
7 GND QH 13 CLK_RESP_MOD
AG

AG

Schematic - ECG Board


P/N 315589 (4 of 5)
page 5-15/16
Isolation Barrier

+5V

P_RAW
P5AI
C12
100N T2
1 1:1.6 6
CR3
4 5 1
4U70 2
IN OUT
DG RB160L-40
8
SD
SENSE 6 L2
U18
C15 C32 TAP
3
+ C65 C13
LP2986IMX-5.0
6
100N 100UF 10U0 100N
3
VCC
1
10V 7
ERR
FB 2
FS D1
U8 2 5 1
GND

5
NC
AG
MAX253CSA AG AG AG
4 8
SD D2
GND1 GND2
2 7 AG
AG
R68 CR2
3 4 P5DI
15K0
DG RB160L-40
1
4U70 2
DG L1
C62
100N
P5DI
SK1
AG
1
2
3
+5V
4
5 C42
10N0
DG
R9
8 7 3K83
VCC ENAB AG
Maintain 0.050 in. clearance 2 ANODE U7
to ground plane in this area. HCPL-261N VO 6
D_FROM_HOST

R2 3
CATHODE
604R GND
E 2 R3 5

B MMBT3906LT1 604R
1
Q1
C 3 AG
R67
15K0

DG DG P_RAW

R4 R5
+5V 1K43 1K43

C43
10N0

DG CR4
7 8 3V0
ENAB VCC
U6 ANODE 2

6 VO
HCPL-261N
AG

3
NOTE: Host must provide pull-up on output data line. CATHODE
D_TO_HOST
GND
5

DG

Schematic - ECG Board


P/N 315589 (5 of 5)
page 5-17/18
A21
A20

A17
A16
A19
A27

A13
A14
A15

A12
A9
A11
+3V3D823 +3V3D823 +3V3D823 +3V3D823

D3
D2
D14
D15

D11
D9
D10
D17

D1
D4
D27
D23

D0
D8
D13
D12
R254 0R0

R202 0R0

R204 0R0

R206 0R0

R207 0R0

R209 0R0

R211 0R0

R213 0R0

R214 0R0

R359 0R0
R244 0R0

R245 0R0

R247 0R0

R250 0R0

R252 0R0
VRAM +3V3D823
C223 C224 C225
R57 100nF 100nF 100nF

R253 0R0

R255 0R0

R203 0R0

R205 0R0

R208 0R0

R210 0R0

R212 0R0
0R0

R243 0R0

R246 0R0

R248 0R0

R249 0R0

R251 0R0

R379 0R0
R28 +3V3D823
NF R215

M10
M11
M12
M13
M14
M15
M16
K10
K11
K12
K13
K14
K15
K16

L10
L11
L12
L13
L14
L15
L16
J10
J11
J12
J13
J14
J15
J16

M1
M2
M3
M4
M5
M6
M7
M8
M9
K1
K2
K3
K4
K5
K6
K7
K8
K9

L1
L2
L3
L4
L5
L6
L7
L8
L9
J1
J2
J3
J4
J5
J6
J7
J8
J9
10K R216

D3
D2
D14
D15
VDDH15
GND19
GND20
GND21
GND22
GND23
GND24
VDDH16
NC5
A21
A20
VDDL2
D11
D9
D10
D17
VDDH17
GND25
GND26
GND27
GND28
GND29
GND30
VDDH
A17
A16
A19
A27
D1
D4
D27
D23
VDDH18
GND31
GND32
GND33
GND34
GND35
GND36
VDDH19
A10
A13
A14
A15
D0
D8
D13
D12
VDDH20
VDDH21
VDDH22
VDDH23
VDDH24
VDDH25
VDDH26
VDDH27
A6
A12
A9
A11
+3V3D823 R217 10K
A1 VSSSYN A[31:6]
A2 IRQ0 N1 10K
VSSSYN1 N2
IRQ1 R390
A3
A4
KAPWR
IRQ7 N3 LCD_CLK D[31:0]
XTALOUT N4 49R9
A5 SHIFT/CLK
R25 XTALIN N5 LCD_G3
A6 LD4
EXTCLK N6 CHB
10M R155 10K A7 PC06
VDDL N7 LCD_B1
5% SWG A8 LCD_C
+3V3D823 IP_B0
PC08 N8 TEMP_O +3V3D
SWE A9
XL5 R95 IP_B3
NC6 N9 PL12
A10 FREEZE
SMRXD2 N10 RX1
4.915MHz 1K0 R136 10K A11 BB N11 HOST_TX2 1 2
C171 R292 2K21 A12 SMTXD1 SWG SRESET
C170 TA N12 3 4
22pF R138 2K21 A13 TDO/DSDO TCK_DSCK
22pF GPL_B5 N13 5 6
CS1[L] R137 10K A14 NC7 SWD
CS1 N14 LCD_R1 7 8
CS3[L] A15 LCD_A RESET[L] DSD1_TD1
+3V3D823 CS3
A07 N15 9 10
SRESET A16 NC1 R378 A8
L7 B1 A08 N16
VDDSYN P1 0R0 R391
47uH B2 NC8 VSYNC
XFC P2
+3V3D823 RESET[L] B3 PORESET
VSYNC 49R9 R392
C115 C3 C102 B4 HSYNC P3 HSYNC
10uF 100nF 10nF SRESET P4 LCD_R2 49R9
16V R256 10K HRST[L] B5 LD2
HRESET P5 LCD_B4
B6 LD6
MODCK2 P6 CHA
R139 1K0 B7 PC[5]
SPKROUT P7 PNEU_RESET
LCD_BACKLIGHT[L] R313 10K B8 PB[18]
DSCK P8 SOUND_TIMER1
SWF B9 TOUT1
IP_B4 P9 TEMP_R
B10 PC[10]
BURST P10
R140 10K B11 NC9
BUSREQ P11 RESET[L]
R145 4K75 B12 TRST
BURSTINH P12 I2CSCL
R141 4K75 B13 I2CSCL
GPL_B4 P13 FS_ARMED[H]
CS2[L] R368 10K B14 PC[13]
CS7[L] B15
CS2 U12 SPIMOSI P14 SPI_MOSI
NIBPCS[L] B16
CS7 MPC823 SPICLK P15 SPI_CLK
CS4 P16 LATCHED_OVC_3V
C1 PA[15]
NC2 R1 LCDOE[L]
C2 PD6
IRQ6 R2 LCD_R4
R142 10K C3 LD0
IRQ3 R3 LCD_G2
R146 10K C4 LD5
+3V3D823 WAIT_B R4 LCD_B2
+3V3D R143 10K C5 RSTCONF
LD8
PB16 R5 SPO2_RESET R154
R59 R144 10K C6 NC3 R6 SOUND_TIMER2 LCD_BRIGHTNESS
SWB C7 TOUT2 1K0
0R0 IP_B6 R7 LCD_G1
SWD C8 LCD_B
C195 C1 IP_B1 R8 PRINTER_CTS C2
SWC C9 PC09
47uF 100nF IP_B5 R9 ADC_BUSY[L] 100uF
16V C10 PB22 10V
BUSGRANT R10 TIMECHK
R293 2K21 C11 PC11
TR_ERR_ACK R11
GPL5 R58 2K21 C12 TDI/DSDI
GPL_A5 R12
CYCLE_PWR C13 TMS
R/W R13 PRINTER_TX
R386 CS6[L] R369 10K C14 TXD2
CLKO CS6 R14 PRINTER_RX
GPL2 C15 RXD2
49R9 GPL_A2 R15
GPL1 C16 PA14
CAS/OE R16 LATCHED_OVP_3V
D1 PC15
+3V3D823 CLKOUT
LD1 T1 LCD_R3
+3V3D823 D30 D2 D30
D3 LD3 T2 LCD_G4 PL8
IRQ5 T3 LCD_B3
R148 10K D4 LD7
R357 IRQ4
PC04 T4 PRINTER_PRES[L] +3V3D823 TCK_DSCK 1
R261 R350 10K R149 10K D5 TEXP T5 ROTOR TMS 2
D6 PC07
10K 10K MODCK1 T6 UNFIL_0VP[L] DSD1_TD1 3
R349 R352 SWH R150 10K D7 IP_B2
PA05
10K 10K T7 DSD0_TD0 4
U72 SWA D8 IP_B7
VDDL4
LV244 CLK1 T8 WARM_T 5
R416 2 A1 D4
D9 IRQ2 T9 TX1 D19
Y1 18 R151 10K D10 SMTXD2 RESET[L] 3 1
0R0 R413 4 A2 D5 TRANS_STRT T10 FAIL_PRI[L]
Y2 16 R152 10K D11 PB23
6 A3 Y3 14 D7 GPL_A4
SMRXD1 T11 HOST_RX2 BAS16
R417 NF 8 A4 D8
CS0[L] R370 10K D12 CS0 T12
Y4 12 CS5[L] D13 TCK/DSCK
0R0 R418 G CS5 T13 REM_AL_REQ[H]
GPL3 D14 PC12
1 GPL_A3 T14 I2CSDA
0R0 GND;10 BS2[L] D15 WE2
I2CSDA R108
+3V3D823;20 SPIMISO T15 TEMP_DATA[L] 1K0
+3V3D823 BS0[L] D16 WE0
PC14 T16 WARM_B
+3V3D823
C218
VDDH14

VDDH13

VDDH12

VDDH11

VDDH10

GPL_A0
BS_AB3

BS_AB1
ET5 ET6 ET7

VDDH9

VDDH8
VDDH7
VDDH6
VDDH5
VDDH4
VDDH3
VDDH2
VDDH1
GND18
GND17
GND16
GND15
GND14
GND13

GND12
GND11
GND10

VDDL1
GND9
GND8
GND7

GND6
GND5
GND4
GND3
GND2
GND1
TSIZ0

TSIZ1
NC4
D22
D18
D16

D25
D20
D19

D28
D24
D21
D06

D26
D31
D29
D07
A29
A23
A25
A24

A28
A30
A22
A18

A31

A26
D5

R356 R354 100nF


H16
H15
H14
H13
H12
H11
H10

F16
F15
F14
F13
F12
F11
F10
H9
H8
H7
H6
H5
H4
H3
H2
H1

F9
F8
F7
F6
F5
F4
F3
F2
F1
G16
G15
G14
G13
G12
G11
G10

E16
E15
E14
E13
E12
E11
E10
G9
G8
G7
G6
G5
G4
G3
G2
G1

E9
E8
E7
E6
E5
E4
E3
E2
E1
10K 10K
R355 R353
10K 10K U72
R371 10K
R223 0R0

R225 0R0

R231 0R0

R233 0R0

R235 0R0

R237 0R0

R239 0R0

R241 0R0
LV244
BS3[L]
R219 0R0

R221 0R0

R227 0R0

R229 0R0

R414 11 A1 D9
Y1 9

R153 10K
NF R415 13 A2 Y2 7 D10 C229 C228 C227
15 A3 D11 100nF 100nF 100nF C226
Y3 5
R222 0R0

R224 0R0

R232 0R0

R234 0R0

R236 0R0

R238 0R0

R240 0R0
R419 NF +3V3D823 +3V3D823 +3V3D823 100nF
0R0

0R0

17 A4 Y4 3 D12
R218 0R0

R220 0R0

R226 0R0

R228 0R0

+3V3D823
0R0 R420 G
Schematic - Main Board
BS1[L]

GPL0
19
0R0
R230

R242
D22
D18
D16

D25
D20
D19

D28
D24
D21

D26
D31
D29
D5

D6

D7
P/N 315592 (1 of 12)
A29
A23
A25
A24

A28
A30
A22
A18

A31

A26

Page 5-19/20
+3V3OP
C242
A[31:6]
U46 10uF
16V
+3V3D MAX809R 3
VCC

RESET 2
U39 C7
KM416S4030C 1 27
100nF GND
VDD1 VDD2
1
A29 23 A0 VDDQ1 3 +3V3D
A28 24 A1 VDDQ2 9 C8
A27 25 A2 VDDQ3 43 +3V3OP
A26 26 49 U42 100nF
A3 VDDQ4 AM29LV800B
A25 29 A4 37
A24 30 2 D31 VCC
A5 DO A30 25
A23 31 4 D30 A0
A6 D1 A29 24
A22 32 5 D29 A1
A7 D2 A28 23 A2 DO 29 D15 R372
A21 33 A8 D3 7 D28 10K
A27 22 A3 D1 31 D14
A20 34 A9 D4 8 D27
A26 21 A4 D2 33 D13
GPL0 22 A10 D5 10 D26
A25 20 A5 D3 35 D12
A18 35 A11 D6 11 D25
A24 19 38 D11
R366
VRAM 13 D24 A6 D4 EXT_POWER_LED
D7 A23 18 40 D10 3 1K2
C4 A9 21 A12/BA0 D8 42 D23 A7 D5 TH1 5%
A22 8 A8 D6 42 D9 G25C
A8 20 A13/BA1 44 D22
100nF D9 A21 7 A9 D7 44 D8 DC_INPUT
0 5% 2 Q21
D10 45 D21 DTC143ZKA
GPL2 18 RAS A20 6 A10 D8 30 D7 1
D11 47 D20 R423
U13 GPL3 17 CAS
48 D19
A19 5 A11 D9 32 D6 VBAT R367
KM6164000B 11 33 16 WE D12 A18 4 34 D5 49K9 BATT_LED
GPL1 50 D18 A12 D10 3 1K2
VCC1 VCC2 D13 A17 3 A13 D11 36 D4 R257 C247 +3V3OP 5%
BS3[L] 15 LDQM D14 51 D17 10K 100nF
A30 5 A16 2 39 D3
A0
BS2[L] 39 UDQM D15 53 D16 A14 D12 C12 2 Q22
A29 4 A1 A15 1 A15 D13 41 D2 DTC143ZKA
1
A28 3 A2 DO 7 D15 CS2[L] 19 CS A14 48 A16 D14 43 D1 100nF +3V3OP
VSSQ1 6 +3V3OP
A27 2 A3 D1 8 D14 R387 38 CLK 12
A13 17 A17 D15 45 D0
A26 1 9 D13 CLKO VSSQ2 A12 16 +12VA
A4 D2 49R9 37 CKE VSSQ3 46 A18 R365
A25 44 10 D12 A11 9 10K
A5 D3
GPL5 VSSQ4 52 A19 R96 U49
A24 43 A6 D4 13 D11 +3V3D 47 BYTE RY/BY 15 1K0 PIC16LC72 20
A23 42 A7 D5 14 D10 VSS1 VSS2 R258 1 VPP/MCLR[L] VDD
CS1[L] 26 CE 12K1
A22 27 A8 D6 15 D9 2 RA0/AN0 RB7 28 BATT_OFF[L]
28 54 GPL1 28 OE RESET 12 RESET[L]
A21 26 A9 D7 16 D8 3 RA1/AN1 RB6 27 CHARGER_ON[H]
BS0[L] 11 WE
A20 25 A10 D8 29 D7 ANALOG_V 4 RA2/AN2 RB5 26 CHARGE_STATUS
A19 24 30 D6 VSS1 VSS2 5 25
A11 D9 RA3/AN3/VREF RB4 R397
A18 23 A12 D10 31 D5 27 46 R4 C248 6 RA4/TOCKI RB3 24 FAILSAFE[L]
A17 22 32 D4 15K 100nF 7 23 10K
A13 D11 +3V3D RA5/AN4 RB2
A16 21 35 D3 22
A14 D12 C202 RB1
A15 20 A15 D13 36 D2 C5 +3V3D 9 OSC1/CLKIN RB0/INT 21
A14 19 A16 D14 37 D1 C6 -12VA 15pF 10 OSC2/CLKOUT
A13 18 38 D0 100nF U70 11 18
A17 D15
100nF XL6 R60 RC0/T1OSO RC7 FAIL_SEC[L]
AT27BV010 32 4.0MHz 0R0 12 RC1/T1OSI RC6 17 TIMECHK
SRAM_CS3[L] 6 CS
GPL1 41 OE U40 1 27 VCC C203 13 RC2/CCP1 RC5 16 CYCLE_PWR
KM416S4030C A31 12 A0 14 RC3/SCK/SCL RC4/SDI/SDA 15 I2CSDA
VDD1 VDD2 A30 11
40 UB A1 15pF VSS1 VSS2
A29 23 A0 VDDQ1 3 A29 10 A2 8 19
39 LB
A28 24 A1 VDDQ2 9 A28 9 A3 OO 13 D7
17 WE
VSS1 VSS2 A27 25 A2 VDDQ3 43 A27 8 A4 O1 14 D6
12 34 A26 26 49 A26 7 15 D5
A3 VDDQ4 A5 O2
A25 29 A25 6 17 D4 +5+3V3EN[H]
A4 A6 O3
A24 30 A5 DO 2 D15 A24 5 A7 O4 18 D3
A23 31 4 D14 A23 27 19 D2 ARM_FS[L]
U20 A6 D1 A8 O5
1 A22 32 A7 D2 5 D13 A22 26 A9 O6 20 D1 +3V3D VRAM
3
2 A21 33 A8 D3 7 D12 A21 23 A10 O7 21 D0 C10 C11
AC08 +3V3D;14 A20 34 A9 D4 8 D11 A20 25 A11
U20 GND;7
22 10 D10 A19 4 100nF 100nF
4 GPL0 A10 D5 A12 +3V3D
6 A18 35 A11 D6 11 D9 A18 28 A13
5
D7 13 D8 A17 29 A14 U15 U36
AC08 A9 21 A12/BA0 42 D7 A16 3 R358 8 PCF8582C C162 XL4 8 PCF8563T +3V3OP
U20 D8 A15
10K VDD 12PF 32.768kHz
BS0[L] 9 A8 20 A13/BA1 D9 44 D6 A15 2 A16 1 A0 TEST 7 1 OSCI VDD CLKOUT 7
8
BS1[L] 10 D10 45 D5 +3V3OP 2 A1 SCL 6 2 OSCO SCL 6
18 RAS 22 CE 31
AC08 GPL2
D11 47 D4 CS0[L] PGM 3 A2 SDA 5 3 INT SDA 5 R422
GPL3 17 CAS GPL1 24 OE VSS VSS 20K
D12 48 D3
GPL1 16 WE 4 4
50 D2
D13
+3V3D 1 VPP R175 ADDR= ADDR=
15 LDQM 51 D1 GND 100K 1010000R/W 1010001R/W
BS1[L] D14 3
BS0[L] 39 UDQM D15 53 D0 16
CS2[L] 19 CS 2 Q30
VSSQ1 6 DTC143ZKA
R388 38 CLK 12 I2CSCL 1
CLKO VSSQ2
49R9 37 CKE VSSQ3 46
GPL5 VSSQ4 52
1 U18
VSS1 VSS2 2 12
9 U18 POWER[H]
28 54 13
10 8 3 U18
HC10 SIGNAL=+3V3OP;14
11 4 6
POWER[L] HC10 5 +3V3OP
D[31:0] HC10
ROTOR[L] C9
100nF

Schematic - Main Board


P/N 315592 (2 of 12)
Page 5-21/22
ANA+
+12VA
VBAT U54 U56
U57 +12VA R91 C106 +12VA LM78L12 LM78L05
5 AD822 C140 C17
18K 2u2 100nF
7 470nF 25V 8 VIN 1 8 VIN 1
R407 VOUT VOUT
6 C27 R6 C144 C231 1M0 C161 C117
R80 330nF G1 G2 G3 G4 C208 C26 G1 G2 G3 G4
100nF 3K3 47nF 100nF 47uF 100nF
R274 21K U57 U75 2 3 6 7 35V 2 3 6 7 10uF
0.1% AD822 AD822 16V R2
3 8 3 8 10R
30K 1 1 +12VA R408 C25
U75
R156 R18 2 5 AD822 2 C15 100K C116
10K 750R
0.1%
4
7
4
CR12 100nF
6 100nF 10uF
3 16V
U55 1
DG408 13 U58 35 36
+12VA 4 S1
V+
D 8 5 AD822
R13 BAT54S

AVDD1

AVDD2
PT1 C29 TH_REF 5 7 1 26 D15
MPX2050 S2 AIN+ D0
PT2 6 S3 6 510R 2 AIN- D1 25 D14
VS 3 100nF PT2_REF 7 C101 3 24 D13
S4 VREF D2
U47 1nF U29
VOUT+ 2 12 S5 4 REFCOMP D3 23 D12
VOUT- 4 R93
3
8
+ +VS7 AD620 PT1 11 S6
LTC1604 D4 22 D11
R61
6 PT1_REF 10 S7 33 SHDN D5 21 D10
GND 1 C14 549R 1 C174 C18
100nF 0.1% REF 9 S8 47uF 100nF 32 CS D6 20 D9 0R0
2
- -VS
4
5 +5VD
C28
16V CONV_ST[L] 31 CONV D7 19 D8
+5VD 1 A0 RD_RES[L] 30 RD D8 18 D7
C31 16 A1 100nF 27 BUSY D9 17 D6
15 A2 +5VD 16 D5
100nF C13 U61 9 DVDD D10
100nF 12 C118 D11 15 D4
-12VA R389 2 EN A30 13
11 10 DGND
14 D3
U24 D12
+12VA HCT174 10K 14 DGND HCT32 +3V3D 10uF 29 OVDD 13 D2
PT2 16V C119 D13
C33 D15 3 1D 1Q 2
V- U61 28 OGND 12 D1
MPX2050 GND;8 3 9 D14
D14 4 2D +5VD;16 2Q 5 8 10uF D15 11 D0
VS 3 100nF D13 C16 A29 10 16V

AGND1
AGND2
AGND3
AGND4
6 3D 3Q 7 HCT32

VSS
VOUT+ 2 U48 11 4D 4Q 10
3
+ +VS 7 AD620 100nF U61 5 6 7 8 34
VOUT- 4 R94 8
13 5D 5Q 12 1
3
6 14 6D 6Q 15
-12VA A24 2 U30 U33 C24
GND 1 C32 549R 1 CLK GND;7
LM79L12 LM79L05
100nF 0.1% REF CLR HCT32 +5VD;14 C206 C207
2
- -VS
4
5
9 1
U61
47uF
35V 2 VIN1
47uF
35V 2 VIN1
100nF
4 3 VIN2 3 VIN2 C120
6 VOUT 1 VOUT 1
C34 5 6 VIN3 6 VIN3
HCT32 7 7 10uF
+12VA 100nF ANA_SEL ANA- C160 VIN4
G3 C23 VIN4
G3
16V
C35 330nF 100nF ADC_BUSY[L]
-12VA 5 5
R81 100nF +5VD U62
21K C211 1 -12VA
0.1% U58 3
AD822 A25 2
3 8 100nF GND;7 BS0[L]
HCT32 +5VD;14

R19 2
1
D[31:0]
750R 4
0.1% A[31:6]
U51
HCT259
A30
R384 GND;8 Q0 4 12V_ON[H]
PUMPC U62 13 D +5VD;16 Q1 5 DEFLATE
4
1K0 A26 5
6 PNEU_CS[L] 14 G Q2 6 DUMP[L]
C232 HCT32 Q3 7 PUMP_ON
100nF A29 1 S0 Q4 9 DEFIBEN[L]
A28 2 S1 Q5 10 ECG_EN[L]
A27 3 S2 Q6 11 ADC_NAP[L]
+5VD Q7 12 ADULT[L]
C30 CLR

100nF 15
TH_REF +5VD RESET[L]
NIBPCS[L]
R1
10R UNFIL_0VP5V[L]
+12VA R70
C21
ANA+ U17 C19 UNFIL_0VP[L]
+5VD 15K
REF-01 5 100nF
100nF R304 +5VD R105
TRIM U59 D5 R26 27K4
R14 AD822 R98 1 3
2 VIN VOUT 6 3 8 R73 U59 1M0
AD822 U53 1K0 10M R158
21K 1 5
R421 BAS16 5%
GND 0.1% 2 10K 7 3 LM393 U53 10K
C20 4 8 R305 R157
100nF 4 0.1% 6 1K0 1 5 LM393
R16 R275
CW R24
2 4 330K 10K 7
FLT_0VP[L]
10K 9K09 30K C130 6
VR2 0.1% 1uF
90K9 C22 16V
ADULT OP Adjust 0.1%
R52 100nF +5VD
10K
0.1%
R15
NEO OP Adjust
VR1 R92
41K2 10K CW 68K1
Q7
+5VD 2N7002
3
R46
49K9
1

2
R306
20K
Schematic - Main Board
P/N 315592 (3 of 12)
Page 5-23/24
+5VD

ECG_EN[L] C212

+3V3D
10nF
+3V3D
C172
C36 R351
10K PL10
22pF
100nF
R377 U14 1
XL2
3.6864MHz 1M0 XR16C2850 44 2
VCC 3 ECG
C173
18
XTAL1 TXA
13 ECG_TX 4 CONNECTOR
22pF 19
XTAL2 RXA
11 ECG_RX 5
14
39 TXB
RESET[H] RESET 10
RXB

24 1
GPL1 -IOR -TXRDYA
20 34
U82 -IOW -RXRDYA
1 16 12
BS0[L] 3 -CSA -TXRDYB
2 17 23
+3V3D;14
-CSB -RXRDYB TO_SPO2
HCT32
33 35
U20 INTA OPA
12 32 15 FROM_SPO2
CS5[L] 11 INTB OPB
13 42
CS7[L] +3V3D;14 -CDA
AC08 A31 31
A0 -CDB
21
A30 30 41
A1 -DSRA
A29 29 25
A[31:6] A2 -DSRB U82
43 4
-RIA 6
26 5
-RIB +3V3D;14
D7 2
D0 HCT32
D6 3 36 U82
D1 -RTSA 9
D5 4 27 8
D2 -RTSB 10
D4 5
D3
-DTRA
37 HCT32
D3 6
D4 38 U82
D2 7 -DTRB 12
D5 11
D1 8 40 13
D6 -CTSA
D0 9 28 HCT32
D[31:0] D7
GND
-CTSB

22

Schematic - Main Board


P/N 315592 (4 of 12)
Page 5-25/26
PL14 +3V3D +5VD
+3V3D +3V3D
1
R373
2 R171 R170 10K
3 2K21 2K21
SW3
4
3 PRINTER_RESET
5
1 COM
6
CHA 2 CHA RESET[L]
2 Q23
3 CHB 1 DTC144EKA
CHB

ANA+
Do not populate SW4 +3V3OP +3V3D Do not populate SW3
SW4 U77
A B
1 VIN VOUT 3
A' B'
5N 78T12C
ROTOR R172 R169 R159 R160 R163 R161 R168 R167 R166 R165 R47 R48 R164 C125 COM 100nF 1
12V
3
10K 10K 10K 10K 10K 10K 10K 10K 10K 10K 49K9 49K9 10K 220uF 2 C245 BZX84C
25V
D21 D8
100nF BAS16
D9 3
C41 1
3 1
ROTOR
A SW1 B BAS16 3 3 PL6
A' B'
5N ROTOR[L] D6 D7 1
MOTOR+
BLUE Probe Switch (Upper) BAS16 BAS16
TEMP_O
2
MOTOR-
1 1 3
DEFLATE VALVE+
A SW2 B TEMP_R
4
DEFLATE VALVE-
A'
5N B'
5
DUMP VALVE +
RED Probe Switch (Lower) SWA
6
7
DUMP VALVE -
Q1
SI9956DY 5
SWB +5VD
Switches are open when +5VD C37 4
6
Probes are in the holders SWC
100nF 3
SWD
SK1 R310
20K 1 CLK VCC
20

1 POWER[L]
11 OE
2 PALCE16V8 Q1
U63 SI9956DY 7
3 SWE
17 15 DEF 8
4 UNFIL_0VP5V[L]
8
I/O5 I/O3
16 DUMP 2
5 SWF FLT_0VP[L]
12
I7 I/O4
9 OVERCURRENT
6 LATCHED_OVC
13
I/O0 I8
14 PUMP 1
7 SWG LATCHED_OVP
3
I/O1 I/O2
ANA+
8 12V_ON[H]
4
I2
9 SWH DEFLATE
5
I3
10 DUMP[L]
6
I4 R259 ANA+
11 PUMP_ON I5 100K
EXT_POWER_LED 7 1
12 PNEU_RESET
2
I6
13 FAILSAFE[L] I1 Q2 R307
BATT_LED SI9956DY 5 2 Q3
14
SWITCH & LED PCB
POWER_YELLOW DTA143XKA
+5VD U62 GND 6 20K 3
15 C42 9 CS6[L] R428 10 4 5

CONNECTOR
8
16
17
POWER_GREEN ABT374 HCT32
10 BS0[L] 1K0 Q8
SI9410DY
6
STANDBY_YELLOW 100nF
OE 1 D[31:0] 3 7
18 STANDBY_GREEN +5VD;20 8
19 TREND_YELLOW U65 CLK 11 Q2
SI9956DY 7
4
20 TREND_GREEN R109
2 1Q 1D 3
5 2Q 2D 4 D1 U62 8 2
21 STAT_YELLOW
R111 330R D2 11
12 2 R276
22 6 3Q 3D 7
13 30K
3
9 4Q 4D 8 D3
23 STAT_GREEN 121R R110 D4
HCT32 1
24 SILENCE_YELLOW
12 5Q 5D 13
+5VD
R115 330R 15 6Q 6D 14 D5
25 SILENCE_RED
D6
26 FREEZE_YELLOW 121R R114
16 7Q 7D 17
19 8Q 8D 18 +5VD
27 FREEZE_GREEN
R113 330R
C39 +5VD
28 NIPB_YELLOW 100nF
29 121R R127
30 NIPB_GREEN
+5VD
C43 ABT374 33K2
U35
R309
31 MAIN_YELLOW R260 20K
+5VD;20 OE 1
3 LM392 U35 R277
32 100nF
33
MAIN_GREEN
RECORD_YELLOW
R112 U66 CLK 11
D7 100K
8
1
LM392
7
5
3K09
2 1Q 1D 3
2 1%
34 RECORD_GREEN R123 330R 5 2Q 2D 4 D8 1uF
C136
4 6
35 121R R122
6 3Q 3D 7 D9
100nF R7
36 9 4Q 4D 8 D10 C38 0R1
R121 330R 12 5Q 5D 13 D11 R273
15 6Q 6D 14 D12
330R R120 200K
16 7Q 7D 17
R119 330R 19 8Q 8D 18 R308 C40
20K 100nF R162
121R 10K
+5VD
C44 ABT374
100nF
+5VD;20 OE 1
PUMPC
R118 U67 CLK 11
D13
2 1Q 1D 3
R117 330R 5 2Q 2D 4 D14
6 3Q 3D 7 D15
121R R116 D16
9 4Q 4D 8
R126 330R 12 5Q 5D 13 D17
15 6Q 6D 14 D18
121R R125 16 7Q 7D 17
R124
121R
330R 19 8Q 8D 18
Schematic - Main Board
DTA143XKA is a digital transistor with a 4K7 base resistor and a 10K base pull-down resistor
VDC_OP P/N 315592 (5 of 12)
DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor
Page 5-27/28
VDC_OP
U37 +3V3D VRAM
LP2951CD-3V3 C48
+3V3D +3V3D VRAM
8 VIN VOUT 1 +3V3OP

7 2 100nF
FB SENSE U50
6 3 C131 MAX793 R383 R262 R174
C46 TAP SDN 2 1
4 100uF 10K 1K0 10K
100nF 5 ERR GND 10V VCC OUT
D12 5
BATT-ON
1 3
16 12
BATT CEOUT SRAM_CS3[L]
11
BAS16 CEIN CS3[L]
10
WDI
C49 9
WDO LOWLINE
14
100nF 8
MR
7 13
PFO RESET RESET[H]
15
RESET RESET[L]
D20 D11
R74 1 3 1 3
4
PFI BATT-OK
3
VRAW1
GND
2K0
3 BAS16 BAS16 6

R385 B1
D15
BZX84C 3V6
2K0 5V6
3/V110H

U19
8
REM_AL_REQ[H] 10
9 REMOTE_ALARM[L]

HC02

U19
5
4 FAILSAFE[H]
6

HC02
R409
FS_ARMED[H] U19
R399
10K 2
1 FAILSAFE[L]
10K 3
+3V3D;14
HC02
R398

10K

U79 U79 U79


R403
1 2 3 4 9 8
ARM_FS[L]
10K
74HC04 74HC04 74HC04
PL7 C237
1uF
3V3_BAT
1
DC_INPUT
U80
2 16 VCC Q14 3
U76
3 U76 9 U76 2
12 8 1 R405 Q13
FAIL_SEC[L]
4
BATT_OFF[L]
11 10 3 FSA[H] 9 OSCOUT2 Q12 1
13 2
5 HC00 249K 15
POWER MONITOR 6
CHARGER_ON[H]
+5+3V3EN[H]
HC00 3V3_BAT;14 HC00 C236
10 OSCOUT1
Q10
Q9 13 V_FSA

CONNECTOR 7
12V_ON[H]
FAIL_PRI[L]
R406
2
U78 100nF Q8 14
8
VDC_OP
1 11 OSCIN Q7 6
1K0 3
9 U76 4
VBAT U79 4 HC02 Q6
10
CHARGE_STATUS R404 6 FSA[L] 12 RESET Q5 5 1
POWER[H] 5 6 5 WD1
CLEAR_FS[L]
8 GND Q4 7
100K HC00 2 DB-E698
74HC04
R400
C238 74HC4060 1K0
1uF

R435
10K
V_FSA 3
3V3_BAT
U78 U78
5 U78 11
D22 U81 4 8 13 2 Q27
6 10 12 DTC143ZKA
2 LP2951CD-3V3 9 1
HC02 3V3_BAT;14 HC02
3
8 1 U79 D24 HC02
VIN VOUT R437
1 3 1
7 FB SENSE 2 11 10 C239
3 10nF
6 TAP SDN 20K
BAT54C 74HC04 BAS16 R402 R401
C240 5 ERR GND 4
47uF C235 D23
35V 10uF 3 1 33K2 12K1
16V

BAS16

U79 3V3_BAT;14
13 12

DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor
74HC04 Schematic - Main Board
P/N 315592 (6 of 12)
Page 5-29/30
+5VIT +5VIT
SK3
R182 R178 L6
10K 10K BLM31B601S
1
FL7 2 Temp probe 2
8 1 L5 3
BLM31B601S 4
7 2 L9 5
BLM31B601S
6 3 6
3 5 4 L26
Z>200/100MHZ BLM31B601S
2 Q28
1 DTC143ZKA
3 VREF
0VITA L12
2 Q29 BLM31B601S
1 DTC143ZKA FL5
8 1

0VITA 7 2 C137 C138


1uF 1uF
6 3 L1 SK4
VREF BLM31B601S
5 4
1
R17
Z>200/100MHZ L10
BLM31B601S 2 Temp Probe 1
4K32 R133 R134 3
0.1% 22K1 22K1 4
+5VIT 0.1% 0.1% 0VIT
5
C54
0VITA 0VITA 6
100nF R128
U38 0VITA 2K74
HEF4051B 16 0.1% +5VIT
11 A0
VDD
Y0 13 +5VIT
10 14
A1 Y1 R318
9 A2 Y2 15 R79 +5VIT 3K09
6 12 16K9 R183 1%
E Y3 0.1% 10K C50
1
Y4 C204
0VIT 5 0VITA 100nF
Y5 U43
2 15pF PIC16C622A-20 0VIT
Y6
0VIT
14 R325
Y7 4 XL1 VDD 1K0
19.6608MHz 16 17
C205 R62 CLKIN RA0
Z 3 15 CLKOUT RA1 18
15pF 0R0 4 VPP RA2 1
VEE VSS
0VITA 0VIT 3 2
7 8
T0CKI RA3 R45
5K76
0VIT RB0 6
0VITA 0VIT RB1 7
RB2 8 0VIT
RB3 9
RB4 10
RB5 11
RB6 12
+5VIT RB7 13
+3V3D
+5VIT VSS
R50 +5VIT
C55 C56 5
47K R297
1% 100nF 100nF 0VIT U4 2K0
U60 0VITA R180 +5VIT
0VITA U2 0VITA 10K 4N35
5 AD822 U60
R68 AD822 MAX1241 1
7 3 8 1 AN C 5 TEMP_DATA[L]
R181
6 7K87 1 2 DO 6 R296 B 6
C129 2 4 10K 3 CS 7 2 CTH E 4
100nF 4 470R
C53 SCLK 8
100nF 1%
0VITA
0VITA R69 R49 5

7K87 68K1 0VITA


0VITA

ISOLATION BARRIER

+5VIT

R200 VREF
150R

U32 R197 C51 C187


TL431 2K7 100nF 47uF R316
1
1% 16V 4K99
8
0VITA 0VITA R63
2 3 6 7
R177 0R0
10K R179 C52

0VITA
10K 100nF 0VIT 0VITA Schematic - Main Board
0VITA
0VITA 0VITA
P/N 315592 (7 of 12)
Page 5-31/32
U52
XC62FP5002PR
D1 L3 2 VIN VOUT 3 +5VIT
RB160L-40 4u7
C108 VSS
220pF 1
C156 C157 R198 R199 C57
10uF 10uF 2K7 2K7 100nF
20V 20V 1% 1%
R9
39R
1%
T1
L2 CEPH205 0VIT
ANA+ 11
4u7 1
C201 C100 C141 C142 C98 C168 2
220uF 100nF 2u2 2u2 100nF 470pF
25V 50V 50V 12
7 D2
5 L4 FS1
R78 6 +5VSPO2
30R RB160L-40 4u7 0.3A
1% 8 C107
220pF
Q6 C158 C159 R295 R438
PHT6N06T 4
10uF 10uF 750R D18 750R
2 20V 20V 1SMB5919BT3
1 R8 5V6
39R Q37
1% 2N7002
3 3
1

R282 R55 SP
2R7 C123
5% 330pF 10K
0.1%
Q15
PHT6N06LT 4
2
1 C167
FAILSAFE[L]
U5 470pF
3
R286 CNY17F-3 C97 R184
1K0
5C AN 1 U16 100nF 10K
6B TL431 1
4E CTH 2
8

R54
R294 R430 2 3 6 7 10K
R300 750R 750R 0.1%
470R U69
1% UCC3809D-1
+3V3D
1 FB REF 8
SP +5VD
2 SS U11
VDD 7 R299 C233 HCPL-2601
R3 10K
3 RT1 OUT 6 100nF 8 VCC
AN 2
4 RT2
10R ISOLATION BARRIER 7 VE
R21
GND 5 C112 C99 C113
6 VO
R427 5 GND CTH 3 FROM_SPO2_ISO
1uF 100nF 1uF 470R
50V 50V
0R0 1%
R303 R320
33K2 15K C145 R319
10nF 15K R426 PL13
VRAW1
NF R192
1 10K
2 +5VD U10
Q32 3 PROBE HOLDER HCPL-2601
SI9956DY 7 4 CONNECTOR 8
U83 8 5 2 AN VCC
3 WARM_T 1
3
R431 2 6 VE 7
2 VO 6 TO_SPO2_ISO
HCT32 10K 1
3 CTH
5
C146 C92 C124 D16 GND
10nF 100pF 330pF BZX84C C234
5V1 100nF
1 SP
PG 1%
Q32 470R
SI9956DY 5 R20 +5VD SP

U83 6
WARM_B 4
6
R432 4 FROM_SPO2
5 R298
10K 3 470R
HCT32 3 U19 1% U6
11 4N35
TO_SPO2 13 2 Q26
12 DTC143ZKA
+3V3D;14 1 1 AN C 5 RESET_TO_SPO2_ISO
HC02
U83 U83 B 6
9 12 2 CTH E 4
8 11
10 13
GND;7 3
HCT32 HCT32 +5VD;14 SP

SPO2_RESET
2 Q12
1 DTC143ZKA

Schematic - Main Board


P/N 315592 (8 of 12)
Page 5-33/34
PL4

HOST
HOST_RX2
2
HOST_TX2
3

TX1
4
5
CONNECTOR
RX1
6
REMOTE_ALARM[L]

+3V3D
+3V3D823 C58

5V-3.3V CONVERTER 100nF


+5VD
+5VD

R439
10K U7
74LCX244 C246
2 18
LATCHED_OVP A1 Y1 LATCHED_OVP_3V
4 16
22UF C214
LATCHED_OVC A2 Y2 LATCHED_OVC_3V 16V
6 14
10nF
PRINTER_TX A3 Y3 C59 L23
8 12
A4 Y4
R393 BLM41P181S
G
LCD_CLK 100nF
1 GND;10 49R9 L24 PL3
+5VD
+3V3D;20 R394 BLM31B601S
HSYNC
U44 HCT244 1

49R9 2
A1 Y1
18 2
R395 L25
4 16 3
VSYNC A2 Y2 BLM31B601S
49R9 6
A3 Y3
14 4
R410 R411 R412 5

LCD CONNECTOR
8 12
U7 LCDOE[L] A4 Y4
330K 330K 330K G R337
6
74LCX244
1 GND;10 100R R338 7
11 9 +5VD;20
A1 Y1 PRINTER_RX 8
13
A2 Y2
7
PRINTER_CTS R339 100R
9
15 5 U44 HCT244
A3 Y3 PRINTER_PRES[L]
100R 10
17
A4 Y4
3
LCD_R4 11
A1 Y1
9 R340
11
G
13 7
LCD_R3 A2 Y2
100R 12
19
LCD_R2 15
A3 Y3
5 R341
13
17 3
LCD_R1 A4 Y4
G 100R R342 14
15
19 R343 100R
16
U45
HCT244 100R 17
R344
LCD_G1 2 18 18
A1 Y1

+5VD LCD_G2 4
A2 Y2
16 100R 19
R345
LCD_G3 6 14 20
A3 Y3
+12VP 8 12 100R R346 21
PL1 LCD_G4 A4 Y4
G
22
C213 1 R347 100R
10nF 1 GND;10 23
2 +5VD;20
U45 HCT244 100R 24
3 R348
LCD_B4 11 9 25
A1 Y1
4
LCD_B3 13
A2 Y2
7 100R 26
5
LCD_B2 15 5 ENAB 27
A3 Y3
6
7 PRINTER LCD_B1 17
A4
G
Y4
3 28
29
PRINTER_PRES 8
9
CONNECTOR 19 +5VD
C60
30
31
10
11 100nF
+5VD
C217 C216 12
FAILSAFE[H]
220uF 220uF R66
13
25V 25V
14 0R0
R425
15
16 NF
R424
17
18 NF
R65
19
20 0R0
21
22
23
PL9
PRINT_DATA_IN 24
PRINT_DATA_OUT 25
1 ANA+
26
2
27
3 +12VP PL11
PRINT_DIN_EN 28
29 POWER 4
5 VRAW1 1

PRINTER_RESET
30
31
CONNECTOR 6
7
2
3
LCD BACKLIGHT
32 ANA-

33
34
8
9
VRAW1
+5VD
C215
10nF
4
CONNECTOR
10 +3V3D
35
11
36 R433
37
20K
3
38
LCD_BRIGHTNESS
39
C221 C151 C148 C150 C197 C196 C147 2 Q31
40 10nF 10nF 10nF 10nF 47uF 47uF 10nF DTC143ZKA
+3V3D 1
16V 16V 3
1
2 Q35
2 Q36 DTC143ZKA
LCD_BACKLIGHT[L] DTA143XKA 1
3

R436
10K Schematic - Main Board
P/N 315592 (9 of 12)
Page 5-35/36
LIM
VDD VDD
L18
BLM31B601S
R56 R191
FL2 10K R37 10K
RCAL-RTN 3 2 L17 SP
0.1% 200R
BLM31B601S R327 U21
RCAL 4 1 7 TLC3704CD
R36
Z>200/100MHZ 100K 1 SNSRDET-L
C78 C79 R102 6 200R
100nF 100nF 2K0
R101 R72
SP SP
2K0 10K5
L16 VLED_SP C80 C91
BLM31B601S 100nF
C143 SP 22nF SP
SP
SP

220uF 44 42 41 58 57
15 10V
SK2 FL1 R281 LIMFLT LIM ENB+ RCAL+ RCAL-
LED+ 3 2 37
1 LED- L19 10R 38
CHL
RCALFB
56
8 4 1 BLM31B601S LED+ LED RCAL 53
2 RCAL-RTN 39 DRIVE RCALTP
R33
9 Z>200/100MHZ R280 LED- 52
SPO2 3 RCAL R278 40 RCALOUT RCALOUT
10 CLL
200R
4
11 33K2
SP 10R 64
CONNECTOR DET-ANODE R32
5 ALCSW
12 65 WINDOW ERROR 45 ERROR-L
6 INR-SHIELD ALCIN ALC OUT
13 C114 66
IN
200R
7 DET-CATHODE ALCOUT
14 L15 R288 2 51
FL6 BLM31B601S R23 39nF DIFFFB+ TESTFB

Z>200/100MHZ 249K 1 - ZERO/TEST


26 ZERO/TEST
16 L14 100R + I
-
N PGAOUT
7 R268
5 4 BLM31B601S DIFFIN+ +

6 3
VREF5
DIFFIN- PGA TEST SW 15K8 R269 C169
+ C139
L13 68 - FLTOUT 9
15K8
7 2 BLM31B601S SP
R22 R287 470pF
67 DIFF- - 820pF SP
8 1 FLTIN 8
100R 249K DEMOD + C75 R35
18
C45 C230 IN LOWSIGC LOWSIG-L
R130 R129
62 DARK LOWSIG
25
LOWSIG
100nF SP
R131 200R R132
1n5 1n5 13 RED IR
14
3K15V 3K15V
41K2 41K2 RDMGIN LEVEL SHIFTERS IDMGIN 41K2 41K2
R82 C181 C182 C180 C179 IDMGOUT
49K9 100nF 100nF 12 IN OUT OUT IN
15 100nF 100nF
CAP CAP

ISOLATION BARRIER
VREF25 VREF25
SP C166
IDMGC 17
60 DKAMPFB
IDMGOUT 16 22nF RDMGOUT
61
VDD
- RDMGOUT 11
R31 C188 C164
499K 10nF + DKAMPIN RDMGC 10
VREF25
REFERENCES 22nF R34
R326 DARK 59 6 VREF25 +2V5VREF
100K C165 63
DKAMPOUT VREF25
3 VREF5 VDDA 200R
DARK DKCAP VREF5
5 ISET
R265 R103 C93 C198
ISET
22nF 20K 100pF 4u7
C175 2K0
RESET_TO_SPO2_ISO 21 STIN
16V
23 CYCLE
34 SP C154
RESET-OUT STOUT
35 10nF 47uF
VDD C134 DOUBLER SATIN-L 16V
SP SP
46 36
C81 47
C1+ GATE
30 SP SP
100nF C133 1uF C1- MODE MODE
16V 48 31
R104 C2+ V10 DACLOAD DACLOAD-L
2K0 1uF 49 C2- DOUT DACDATA
32 DACDATA
SP
16V 50 33
DACCLK DACCLK
4 RESET-L
24 ASIC-RESET-L
43
R106 ENLIMRST-L

U9 28
49R9 OXICHIP2
ADCCLK
29
SP
22 GPTCLK
C200 C199 VPP 27 CLK10
4u7 4u7 CLK10MH

L20 VDD 16V 16V VCC DGND VCCA AGND


20 54 19 55 VDDA
BLM31B601S
+5VSPO2
SP SP VDD
C77 C76
VLED_SP C127 100nF 100nF
L21 220uF
C128 16V C95 C94
220uF 23R-25MHz
16V
SP
100pF 100pF
C177 C178 SP SP
47uF 47uF
16V 16V
SP
SP SP
RESET-OUT
VDDA
R279
10R
R67 C176 C153
47uF 10nF
0R0 16V

SP SP SP SP

Schematic - Main Board


P/N 315592 (11 of 12)
Page 5-39/40
ANALOG OUTPUT
C67
100nF
+5VD
C72 +5VD
100nF C73
U3
MAX504 13 100nF
VDD U64
2 AD8052A
SPI_MOSI DIN DOUT 6 2 8 R190 C68
RESET[L] 3 CLR RFB 14 1
R186 R187
SPI_CLK 4 SCLK VOUT 12 3 4 4K12 100nF
DEFIBEN[L] 5 CS 15K 15K C69 R382
BIPOFF 1
8 AGND REFIN 9 C74 10K
7 DGND 47nF 100nF
REFOUT 10 PL5
VSS U64
11 6 AD8052A
R188
C105
33uF
R376 5
7
1K0
1
2 ANALOG OUTPUT
16V 4K12 L27
BLM31B601S
3 CONNECTOR
R440
10K

U34
LM79L05 C155
2 10uF
VIN1 16V
3 VIN2
6 VOUT 1 -5VD
VIN3
7 VIN4
G3
5
ANA-

+5VD

R75
1R0
C122
10uF
16V
U68
+5VD +5VD
SOUNDER
LM386 PL2
R314 C61 R185 6 L28
3 1
C70 C126 BLM31B601S
SPEAKER
8
R322 1M0 1nF 49K9 5 1
L29
CONNECTOR
15K 100nF 2 7 220uF 2
+3V3D C243 C152 4 R283 16V BLM31B601S
1nF R100 2R7
3 U23 +3V3D 158K 5%
39P MAX518 U8 C71 C62
74HC4053 C63 C121 100nF
R284 Q13 100nF 10uF C64
2
7 VDD
R263 100nF 16V
49K9 1 DTC143ZKA OUT0 1 1 2Y1 VCC 16
R315 100nF
R264 200K 2 2Y0 2Z 15
R321 3 SCL OUT1 8 3 3Y1 1Z 14 1M0
4K99 200K 4 3Z 1Y1 13
4 SDA AD0 6 5 3Y0 1Y0 12
+5VD 6 E S1 11
I2CSCL 2 GND AD1 5 C65 C66
7 VEE S2 10
100nF 100nF 8 GND S3 9
R323
15K
+3V3D C244
39P 3
R429
R285 1K0
2 Q14
49K9 1 DTC143ZKA

R324 D14
4K99 2
3
1
I2CSDA SOUND_TIMER2
BAT54C
SOUND_TIMER1 Schematic - Main Board
P/N 315592 (10 of 12)
Page 5-37/38
R90 IRFB
49K9
C109 R89 C194
220pF
49K9 VDDA 10nF
R11 C90
SP 110K C193
100nF
U26 SP 10nF
R271 R291 2 TLC2262CD
IDMGOUT 8 R329 U26
158K 162K 1 6 TLC2262CD U21
C185 3 4 100K 7 11 TLC3704CD
100nF 5 13 IRSD
R328 10
SP 100K R336
SP 100K
+2V5VREF +2V5VREF

R87 REDFB
49K9
C111 R88 C192
220pF
R10 49K9 VDDA 10nF VDD
110K C89 C191
SP C85
U25 100nF 10nF
100nF
R272 R290 2 TLC2262CD SP

RDMGOUT 8 R334 U25 SP

158K 162K 1 6 TLC2262CD U21


3 100K 7 5 TLC3704CD
C184 4 3
100nF 5 2 REDSD
4
R333 12
SP 100K SP
R335
100K
SP
+2V5VREF

+1V63VREF
VDD
VDDA C82
C88 R51 U1
150K MAX4053A 100nF
100nF 0.5% 6
11 INH VCC 16 SP

SP
U27 10 A
B
2 TLC2262CD 9 C X 14 IRFB
8 R77 U27 12 15 REDFB
1 6 TLC2262CD 13 XO Y
R301 X1
4 DARKFB
3 4 100K 7 +3V81VREF 2 Y0
Z

53K6 0.1% 5
1 Y1
VEE 7
0.5% 5
3 Z0
Z1 VSS 8
R76 SP
R84
100K
0.1% +2V5VREF 49K9 R266 SP

20K
SP

SP
R86 DARKFB
49K9 C189
R85 C190 VDD
C110
220pF 49K9 10nF C83
R12 10nF
110K VDDA C87
SP 100nF U31
HC175

DARKSDFF
IRSDFF
100nF

REDSDFF
SP
U28 SP
4 1D 2
R270 R289 2 TLC2262CD 1Q
DARK 8 R330 U28 SP_DG;8 1Q 3
158K 162K 1 6 TLC2262CD U21 5 2D VDD;16 2Q 7
3 100K 7 9 TLC3704CD 6
C183 4
5 14 DARKSD
2Q
100nF 12 3D 3Q 10
8 3Q 11
SP
R331 SP
R332 13 4D 4Q 15
100K 100K 14
CLK 4Q
+2V5VREF CLR
SP VDD
9 1
R196 VDD VDD
SDCLK
10K
VDD R194 R267
R41 U22 10K 20K
DACCLK R83
200R R42 PIC16C66-20
DACDATA 49K9 21 2 IRSDFF
R39 200R 22
RB0 RA0
3 REDSDFF LOWSIG-L
DACLOAD-L RB1 RA1
200R 23 RB2 RA2 4 DARKSDFF R44
LIM 24 5 LOWSIG-L MODE
25
RB3 RA3
6
200R
ERROR-L RB4 RA4
R43
26 7 ASIC-RESET-L
R40 RB5 RA5

ZERO/TEST 27 RB6 200R R38


200R 28 RB7 MCLR 1 RESET-OUT
200R
11 9
SNSRDET-L RC0 OSC1 VDD VDD
12 10 CLK20
RC1 OSC2
VDD XL3
RCALOUT 13 RC2
C86
14 20 20MHz U41 R195
R193 15
RC3 VDD
HC74 VDD 10K 100nF
RC4
C104 C103 U41
10K 16 RC5 C84 R107 18pF 18pF 4 PRE VCC 14
5 HC74
SP

17 8 100nF 49R9 2 D
Q
10 14 R201
FROM_SPO2_ISO RC6 VSS1
C135 PRE VCC
9
3 Q CLK10
TO_SPO2_ISO 18 RC7 VSS2 19 1uF SP SP
CLK
Q 6 12 D
150R
16V 1 CLR GND 7 11 CLK
Q 8
13 CLR GND 7
SP SP

SP
SP

Schematic - Main Board


P/N 315592 (12 of 12)
Page 5-41/42
R17
+5V
10M
5%
R93 VDC_OP
127K
U12
5 LM339
3
2
BATTERY_SWITCH
4 12 (SHT. 2)
R30
47K5
DC_INPUT
(SHT. 4)
PL4
D13
1
MAINS AC/DC 2
PBYR1045
+12V_+32V
(SHT. 2)
MODULE 3 R4
4 110K C75 C76
C22 470uF 470uF
50V 50V
100nF
D11 C23
SM6T39CA 100nF U2
LM2931D2T
R75 C24
10K7 100nF 4 +VIN 5
OUT +12V_IN
2 SHDN 1 (SHT. 2)
ADJ
GND R33 C73
C25 1K0 47uF
PL3 100nF 3 16V

FS2 D12
EXTERNAL DC 1
2 5A0 PBYR1045
+12V TO +32V R64
9K09

D10
VDC_OP U12
7 LM339
1
6 BAV70 C74
U17 C51 47uF
R3 LT1339 1uF 16V
16K9 35V Q15
U12 1 SYNC VBOOST 20 PHP21N06T
9 LM339
14
C53 2 5VREF
1uF TG 19
8 16V
C10 3 CT
1nF 4 SL/ADJ TS 18
D2
R21 12VIN 17 RB160L-40 L2
U12 5 IAVG BG 16 79uH
11 LM339 0R0 6 SS Q16
13 PGND 15 PHP21N06T
10 C54 7 VC PHASE 14
8 SGND RUN/SHDN 13
330pF
SENSE+ 11
9 VFB
C11 R45 D3
10 VREF SENSE- 12
RB160L-40
DTC143ZKA is a digital transistor with a 4K7 base resistor and a 47K base pull-down resistor 1nF 10K
D12 , D13 , Q16 Require heatsink type F13-2-220D-2LS
Q15 Requires heatsink type F14-2-220-2L C14
100nF R87
R34 R74 100K
1K0 10K7
R24 C99
0R0
ET4 ET3 ET2 ET1 10nF R5
0R022
1W

VDCIN
(SHT. 2)

C6 C7
470uF 470uF
35V 35V

Schematic - PSU
P/N 315593 (1 of 5)
Page 5-43/44
+12V_+32V
(SHT. 1)

+12V_IN
(SHT. 1) U5

MIC2950-06BZ
C91 C92
220uF 220uF
50V 50V

1 3
+5V VOUT VIN VDC_OP

D16 (SHT. 1) (SHT. 4)

C52
GND
C28
100nF
R8 C89
2 100nF D21
16K9 1uF
C96 R103 R99 R102 BAV70
16V
100nF 100K 47K5 100K
U23
C84 BAV70 C90
LT1339
C86
1uF 47uF
1uF 35V 16V Q21
R44
16V
1 20 PHP21N06T
2K67 SYNC VBOOST
C30
Q2 Q3
1nF
2
5VREF IRF5305S IRF5305S
19
TG L1 D20
R111
VRAW1
3
CT
18 149uH 0R033
TS (SHT. 3,4)
4 1W PBYR1045
SL/ADJ
17
12VIN
D15 C12
5 16 R18
IAVG BG RB160L-40 470uF
35V 10K
C85
BATTERY_SWITCH
6
SS Q20
15 (SHT. 1)
PGND
PHP21N06T
1uF
7 14
16V VC PHASE

13
RUN/SHDN
8
SGND
11 C18 R94
SENSE+
100nF 3K32
9
VFB Q11
C31 R51
10 12 D14 DTC143ZKA
VREF SENSE-
RB160L-40 Q9
10K BATT_OFF[L] ZVP3306F
10nF
DTC143ZKA
Q22
(SHT. 4)

C78 C55
100nF
R104
100nF

R50 R101 100K VRAW

1K0 14K3 D19


R112
VDCIN
R43
C83 0R010
(SHT. 1)
0R0 VRAW PBYR1045
U16 1W
100pF
LT1620CS8
R91 C81
R14 R13
100K 4u7
C100 100R 100R
U11 35V
R105 6 5
VCC IN+
MAX472
100K
10nF
VRAW2
1 4
SENSE IN- 2 6
Q12 NC RG2 (SHT. 3)

DTC143ZKA
C97 R90 MTD2955V
C17
8 7
AVG PROG 3 7 100K Q6
RG1 VCC

100nF
Q17 C98 100nF
CHARGER_ON[H]
3 2
R121
DTC143ZKA GND IOUT 1 5
SHDN SIGN
(SHT. 4)
100nF FS3
1K0
5A0 R95
C15
4 8 R9
GND OUT
100nF
39K2
9K09

C87

1uF
Q10
16V
12V_ON[H]
DTC143ZKA
(SHT. 4)

R78
R1

53K6
1M0
FS5
Q1 VBAT
U22
FDV302P R56
LM339 11 0A5 (SHT. 4)
R97
C77
13 10K
R113 FS6 100nF
56K2 10

0R010
R47 1W BATTERY DISCONNECT

150K
C103
R19 R25
22UF
100R 100R
U13 50V

R69 MAX472
R96

13K
1M0 2 6
NC RG2

U22
R52 C26 PL7
LM339 7
3 7
R98 RG1 VCC
1 10K C27
100nF 1
R120 + BATTERY
33K2 6
1 5
R79 SHDN SIGN 2
- BATTERY
100nF
237K 1K0 3
R58 - THERMISTOR
4 8
GND OUT 4
+THERMISTOR (33K @ 25 C)
R88 10K
R71 5

20K
C88
1M0 C29
R100
R106 100nF
U22 U24 10K
100K
R55 1uF BATTERY CONNECTOR
LM339 9 PIC12C672-04SM 16V
1
R89
14 10K VDD

7 4
26K7 8 AN0 GP3
D29

TRICKLE_CHARGE[L] 3 6 BZX84C
GP4 AN1
5V1

2 5
GP5 AN2
R57
Q18
VSS
10K
DTC143ZKA
8

NiMH BATTERY CHARGER CHARGE_STATUS

(SHT. 4)

Q19

DTC143ZKA

Schematic - PSU
P/N 315593 (2 of 5)
Page 5-45/46
L4 D7
VRAW2 ANA+
(SHT. 2) 33uH RB051L-40 (SHT. 4)

ANA+ C36
U10
100nF

U8 Q7
R20
127K
LM340-12
AUX 12V
MAX668 9 10 PHT6N06LT ANA+
1 VIN 3 VOUT
VCC SHDN AUX_12V
1 LDO
R22 R119 (SHT. 5)
C67 EXT 8 C68 C38
49R9 49R9 GND
330uF 330uF 100nF
25V 2 FREQ 25V C47 2 C35 FS1
CS+6
220nF 100nF AUX_12V[F]
4 REF 0A5 (SHT. 5)
C56 R31 R65
1uF 5 FB
16V PGND 7 0R047 12K1
0.25W
GND
3
R28 C48 C80 C37 C69
237K 220nF 4n7 100nF 330uF
25V
C4
2n2
R15 R70
1K21 13K
U19
LM2594
ANA-
4
FB
D4 L7 C34
7 +VIN OUT 8
RB160L-40 100uH 100nF
ON /OFF GND
C9 C33 5 6
100uF 100nF
50V

D5 C13 D6
R80 RB160L-40 100uF RB160L-40
25V
47K5

R81
47K5
ANA-
(SHT. 4)

VRAW1
U21
D17 6 LM324A
L3 7
VRAW1 +12VP
(SHT. 2) 12uH PBYR745B (SHT. 4) 5

+12V PRINTER SUPPLY C40 NC=1


C32
100nF C70
100nF 330uF
R7 25V
R118 210K
U9 Q5
MAX668 9 10 49R9 PHT11N06LT
VCC SHDN R23
1 LDO EXT 8 R115
C63 C64 C65 C66 C71 C39 1K0 C101 R61 R109
330uF 330uF 330uF 49R9 330uF 330uF 100nF 100nF

FS8
0A5
25V 25V 25V 2 FREQ 25V 25V 53K6 150K
CS+6
R49 U21
4 REF LM324A 13
1K0
C57
1uF
16V
R67
237K 5 FB PGND 7
R6
0R022
1W
C3
2n2
R2
24K3
TH1
G25C
5%
U22
LM339
R62
20K
14
12 FAN
GND 0 5 3 R60 PL6
3 2 150K
C46 4 R12 R63 1
R117 220nF R110 R68 12
10K 100K 53K6 2
20K 24K3
10
U14 8
TL431 C82
1 9 LM324A D33 R107 C16
U21
8
R11 100K 100nF 100nF
2 3 6 7 100K U21 PHT6N06LT
2 LM324A Q8
4 R59
BAV70 1 C102
4u7
R48 R108 3 11 20K 35V
R66
20K 20K 20K

FAN CONTROL

Schematic - PSU
P/N 315593 (3 of 5)
Page 5-47/48
C5
PL2

2n2

1
ANA+

2
R72 R16
+5VD 3
+12VP
(SHT. 5)
324R 1K07 4

FS4 5
POWER
U6 U6 Requires heatsink type F14-2-220-2L AUX_5V[F] 6

LM2596
4
0A9 (SHT. 5)
7
CONNECTOR
ANA-
FB VRAW1
8
L5
VRAW1 1 2 9
+VIN OUT
(SHT. 2)
47uH 10

11

GND ON /OFF

3 5
R82 D9

47K5 RB051L-40
C8 C41
C79 C42
470uF 100nF
35V 470uF 100nF
25V

PL1

+5VD SUPPLY DC_INPUT


1

3
Q13
4
DTC143ZKA BATT_OFF[L]

5
CHARGER_ON[H]
+5+3V3EN[H]
6
POWER MONITOR
7
12V_ON[H]

8
CONNECTOR
VDC_OP

9
VBAT

10
CHARGE_STATUS

C72

10nF

R73 R10
+3V3D

(SHT. 5)
332R 562R

U3

LM2595
4

FB

L6
2 1
+VIN OUT

100uH

GND ON /OFF

3 5
R83 D8

47K5 RB051L-40
C1 C44
C2 C43
220uF 100nF
25V 220uF 100nF
25V

+3V3D SUPPLY

Q14

DTC143ZKA

Schematic - PSU
P/N 315593 (4 of 5)
Page 5-49/50
+3V3D
U7 U7 U7
C94

9 8 11 10 13 12

100nF
74HC14 74HC14 74HC14

U7
R76
TX1
1 2

GND;7 1K0
74HC14 +3V3D;14
D25

BZX84C
5V1

U7
R77
RX1
4 3

1K0
74HC14
R86
D26 SK2
47K5
BZX84C
5V1
1

3
COMMS
4
R40
PL5 5 CONNECTOR
1K0 (SHT. 4) +5VD 6
HOST_RX2
1
(SHT. 3) 7
D22 AUX_12V
HOST_TX2 R84
2
BZX84C 8
47K5
HOST 3
5V1
C49
9
TX1 10uF
4
U15 16V
CONNECTOR 5
RX1
HC240
REMOTE_ALARM[L] R41
6
2 18
A1 Y1

4
A2 Y2
16 1K0

6 14
A3 Y3
D23
8 12
A4 Y4 BZX84C
G 5V1
AUX_5V[F]

1 GND;10

+3V3D;20
+3V3D
C45
C50
10uF
100nF 16V

R38

1K0

AUX_12V[F]

U15

HC240

11 9
A1 Y1 R114
13 7
A2 Y2

15
A3 Y3
5 1K0

17 3
A4 Y4
SK1
G

LK1
19
1

1 1 2
U7 R39
2 2 3
6 5

1K0 3 3 4

74HC14 R85 D24 4 4 5

47K5 BZX84C
+3V3D 5 5 6
5V1

C58 6 6 GND2 7
WIRELESS
7 7 8

U20 1uF
8 8 9
CONNECTOR
16V
MAX3221E 15
9 9 10
C60
VCC
1uF 10 10 11
2 3
16V C1+ V+
12
4
C1- V-
16V C61 13
C59 5 7 +3V3D +3V3D
C2+ U1
1uF 1uF
1uF 2 14
6
16V C2- C62 16V
1
3 15
GND;7
+3V3D;14 HC02 R54
11 T1IN T1OUT 13 R53
10K
10K

9 R1OUT 8 R1IN
HC02 R35
ENABLE_AND_TTL
9
10 1K0
8
R36
1 10 U1
EN INVALID
+3V3D U1 1K0
11 ~ENABLE
ENABLE_AND_RS232
12 16 13
FORCEON FORCEOFF
12 ~RS232
GND

HC02
14

FS7
D27 D28

0A5
U1
5 BZX84C BZX84C
4 5V1 5V1
6 +3V3D
C95
HC02

100nF

Q4

PHT6N06LT

REMOTE_ALARM[L]
D1

SM6T15C

D32

BZX84C
5V1

HOST INTERFACE
Schematic - PSU
P/N 315593 (5 of 5)
Page 5-51/52
PL1

SW1
A B SWA 2

NIPB A' B' SWB 3


5N

SWC 4

SWD 5

POWER[L] 6

SW8
A B SWE 7

A' B' 8

STAT 5N

SWF 9

SWG 10

SWH 11

SW7
A B EXT_POWER_LED 12

STANDBY A' B' BATT_LED 13


5N
VDC_OP
POWER_YELLOW 14

15

POWER_GREEN 16

SW6
A B STANDBY_YELLOW 17

SILENCE A' B' EXT_PWR STANDBY_GREEN 18


5N
LED1
TREND_YELLOW 19
LSG53D

TREND_GREEN 20

STAT_GREEN 21

SW5
A B
LED17
BATT 22

LED2
POWER A' B'
LUY53C STAT_YELLOW 23
5N POWER LUY53C

SILENCE_YELLOW 24

SILENCE_RED 25

FREEZE_YELLOW 26
LED3

SW4
LSG53D
A B FREEZE_GREEN 27

TREND A' B' NIPB_GREEN 28


5N

29

NIPB_YELLOW 30

MAIN_YELLOW 31
LED19

SW3
LUY53C
A B MAIN_GREEN 32

STANDBY
MAIN A' B' RECORD_YELLOW 33
5N

RECORD_GREEN 34

35
LED9

LED18
LSG53D
36
LUY53C
SW9
A B TREND
VDC_OP

FREEZE A' B'

5N

LED8

LSG53D

SW2
A B

RECORD A' B'

5N
LED21

LSG53D

STAT

LED11

LED20
LUY53C

LUY53C

SILENCE

LED10

LMR53DG

LED14

LUY53C

FREEZE

LED4

LED13
LSG53D

LSG53D

NIPB

LED12

LUY53C

LED15

LUY53C

MAIN

LED5

LED16
LSG53D

LUY53C

RECORD

LED6

LSG53D

Schematic - Keyboard
P/N 315594 (1 of 1)
page 5-53/54
Not Populated

VCC 1
J1

R4 R5 R6 R7 R8 R9 R10 R11 R12 R13


2K43 2K43 2K43 1K00 2K43 2K43 1K00 2K43 2K43 1K00

R14 R15 R1 R3 R2 R16 R17 R18 R19


2K43 2K43 1K00 1K00 2K43 2K43 1K00 2K43 1K00

Not Populated

GND 1
J2

Schematic - Probe Warmer


P/N 315595 (1 of 1)
page 5-55/56

You might also like