Professional Documents
Culture Documents
CS-T240
BIO-CHEMISTRY ANALYZER
CS-T240 Operation Manual
Instruction:
Dear user, thanks for purchasing our CS-T240 Auto-Chemistry Analyzer.
Please read the user manual carefully in order to operate the instrument correctly. Incorrect operation may
effect the precision and accuracy of the test results, or endanger personal safety.
Please keep the user manual safely for your any time reference.
Note:
● Instrument should be operated by medical inspection specialist, physician, nurse or lab assistant whom are
specially trained.
● Instrument should be controlled by special software. Please install the software that is appointed by our
company. Installation of other software/hardware may interfere normal operation. Don’t operate other software
when instrument operating.
● Dust may accumulate on the surface of instrument after long time storage. Soft cloth or gauze can be used for
cleaning work, and a little detergent can be used if necessary. Please cut off the power supply before cleaning.
When instrument is not used, make sure shut the lid down.
●As to the use and storage method of the sample, reagent, Controls, Calibrator, please refer to the relevant
instructions.
●Sample, Controls, Calibrator and waster solution have the potential biochemical infectivity, the detergents are
corrosive that may hurt eyes, skin and mucosa. Operator should refer to the safety regulation for lab operation.
Protective measure should be taken to operator (Such as lab protective clothes and gloves).
●Avoid contact with eyes and skin, in case of skin contact, flush the area with water, rinse immediately with
plenty of water and seek medical advice.
● Operator should comply with the local regulation when draining and dealing with reagent, waste solution,
waste sample, consumable etc. Please dispose the waste solution and instrument consumable according to the
regulation of medical waste, infective waste and industrial waste.
Warning:
●. Instrument should be operated in a good ground condition, and an independent power supply is a must, the
input power should be conformed to instrument requirement.
●. Don’t pull the electrical wire with wet hand, or there is a risk of electrical shock.
●. Don't stamp, twist, drag the wire and cable, or it may cause a fire.
●. Please don’t open the back and side cover board before cutting the general power supply except DIRUI
special service staff.
●. If liquid occurs in instrument interior or there is an internal pipeline leakage, please immediately cut off the
general power supply, and contact DIRUI customer service dept.
● Please don’t touch sample probe, reagent probe and stirring rod, etc. when instrument operating, don’t put your
hand into the opening part, or it may cause body injury or instrument damage.
● Cut off the power supply before replace light source lamp. Don’t touch the lamp before it is cool to avoid
CS-T240 Operation Manual
burning.
● Periodic maintenance should be executed strictly according to the user manual. Or it may cause instrument
malfunction, and effect the accuracy and precision of test results.
● Make sure that the Auto-Chemistry Analyzer is operated according to the user manual, or the measuring result
is not a reliable one, and the damage on instrument may endanger human safety.
● Please don’t place combustible material around the instrument.
CS-T240 Operation Manual
Catalogue
Chapter 1 Brief Introduction................................................................................................................................. 7
1.1 Summary........................................................................................................................................................ 7
1.2 Main Technical Index .................................................................................................................................... 7
1.3 Composition of instrument ............................................................................................................................ 8
1.3.1 Front picture ............................................................................................................................................... 8
1.3.2 Back picture ................................................................................................................................................ 9
1.3.3 Leftside picture ........................................................................................................................................... 9
1.4 Configuration and function.......................................................................................................................... 10
1.4.1 Operating system ...................................................................................................................................... 10
1.4.2 Analytical system...................................................................................................................................... 10
1.4.2.1 Sample reagent disk ........................................................................................................... 10
1.4.2.2 Sampling mechanism ......................................................................................................... 11
1.4.2.3 Reaction disk ..................................................................................................................... 13
1.4.2.4 Incubation bath .................................................................................................................. 14
1.4.2.5 Stirring mechanism ............................................................................................................ 14
1.4.2.6 Reaction cuvette rinsing mechanism ................................................................................. 15
1.4.2.7 Reagent cooling system ..................................................................................................... 17
1.4.2.8 Optical system ................................................................................................................... 17
1.5 Instrument Symbol ...................................................................................................................................... 18
Chapter 2 Measuring Principle ........................................................................................................................... 19
2.1 Mechanism movement principle.................................................................................................................. 19
2.1.1 Operation flow .......................................................................................................................................... 19
2.1.1.1Rinsing unit......................................................................................................................... 19
2.1.1.2 Sample and reagent adding unit ......................................................................................... 19
2.1.1.3 Stirring unit ........................................................................................................................ 20
2.1.1.4 Reaction disk unit .............................................................................................................. 20
2.1.2 Metering Characteristics ........................................................................................................................... 22
2.2 Analytical mode ........................................................................................................................................... 22
2.2.1Assay mode variety ................................................................................................................................... 23
2.2.2 Calibration Method................................................................................................................................... 28
2.3 Check of measure ........................................................................................................................................ 42
2.3.1 Calibration check ...................................................................................................................................... 42
2.3.2 Absorbance limit....................................................................................................................................... 43
2.3.3 Linearity Abnormal Check ....................................................................................................................... 44
2.3.4 Prozone check........................................................................................................................................... 45
Chapter 3 Instrument Installation ....................................................................................................................... 47
3.1 Installation requirement ............................................................................................................................... 47
3.1.1 Space Requirement ................................................................................................................................... 47
3.1.2 Environment requirement ......................................................................................................................... 47
3.1.3 Power requirement.................................................................................................................................... 47
3.2 Open package .............................................................................................................................................. 47
3.2.1 Procedure .................................................................................................................................................. 47
3.2.2 Handling method ...................................................................................................................................... 48
3.3 Installation procedure .................................................................................................................................. 48
3.3.1. software installation................................................................................................................................. 48
3.3.2 Peripherial device connection................................................................................................................... 51
3.3.2.1 Connection of pure water inlet pipeline............................................................................. 51
3.3.2.2 Connection of waste liquid outlet pipeline ........................................................................ 51
3.3.2.3 Connection of computer .................................................................................................... 51
3.3.2.4 Printer installation.............................................................................................................. 51
3.3.3 System login ............................................................................................................................................. 51
Chapter 4 Accessory Device............................................................................................................................... 56
4.1 Barcode reader ............................................................................................................................................. 56
4.1.1 Scan range of barcode reader.................................................................................................................... 56
4.1.2 Sample container requirement .................................................................................................................. 56
CS-T240 Operation Manual
4.1.3 Barcode using requirement ....................................................................................................................... 56
4.1.4 Stick requirement of sample barcode ....................................................................................................... 57
4.1.5 Reagent bottle requirement:................................................................................................................... 57
4.1.6 Reagent barcode using requirement.......................................................................................................... 57
4.1.7 Reagent barcode stick requirement........................................................................................................... 57
4.1.8 The rule of reagent barcode ...................................................................................................................... 57
4.1.9The using of sample reagent barcode reader ............................................................................................. 59
4.2 Purified water equipment............................................................................................................................. 59
Chapter 5 Software Operation ............................................................................................................................ 60
5.1 Software interface instruction...................................................................................................................... 60
5.1.1 Main interface composition ...................................................................................................................... 60
5.1.2 Keyboard function .................................................................................................................................... 62
5.1.3 Software function frame ........................................................................................................................... 62
5.2 Software Operation...................................................................................................................................... 64
5.2.1 Icon move ................................................................................................................................................. 64
5.2.2 Function key selection .............................................................................................................................. 64
5.2.3Open Form ................................................................................................................................................. 64
5.2.4The operation of list box and scroll bar ..................................................................................................... 65
5.2.5 Pull down menu operation ........................................................................................................................ 66
5.2.6Button box and check box ......................................................................................................................... 66
5.3 Instrument standard specification ................................................................................................................ 66
Chapter 6 Instrument Operation ......................................................................................................................... 68
6.1Operation overview ...................................................................................................................................... 68
6.2Detailed operation......................................................................................................................................... 69
6.2.1 Check before measurement....................................................................................................................... 69
6.2.2 Power on and software login .................................................................................................................... 69
6.2.3Check instrument status ............................................................................................................................. 69
6.2.3.1 Alarm check....................................................................................................................... 69
6.2.3.2 Light quantity check .......................................................................................................... 71
6.2.3.3 Cuvette blank check........................................................................................................... 72
6.2.3.4 Check the temperature of incubation bath ......................................................................... 73
6.2.4 Check analyze condition........................................................................................................................... 73
6.2.4.1 Check analysis conditions of colorimetric item ................................................................. 73
6.2.5 Reagent preparation .................................................................................................................................. 75
6.2.5.1 Reagent usage and important notice .................................................................................. 75
6.2.5.2 Reagent manual registration .............................................................................................. 76
6.2.5.3 Barcode scanning ( automatic registration) ....................................................................... 77
6.2.5.4 Reagent Horizontal ............................................................................................................ 78
6.2.5.5 Delete reagent information ................................................................................................ 79
6.2.6.2 QC item registration .......................................................................................................... 81
6.2.7 Sample registration and testing(Sample registration).......................................................................... 81
6.2.7.1 Single sample registration.................................................................................................. 81
6.2.7.2 Registration of batch routine sample ................................................................................. 83
6.2.7.3 Edit the patient info ........................................................................................................... 84
6.2.7.4 Data application ................................................................................................................. 86
6.2.7.5 Modification and deletion of sample information ............................................................. 86
6.2.8 Test preparation ........................................................................................................................................ 87
6.2.8.2 Test..................................................................................................................................... 87
6.2.9 Testing process.......................................................................................................................................... 87
6.2.9.1 System monitor .................................................................................................................. 87
6.2.9.2 Emergence stop.................................................................................................................. 90
6.2.9.3 Sample addition ................................................................................................................. 90
6.2.10 Test result checkup (Result data) ............................................................................................................ 91
6.2.10.1 Daily result ...................................................................................................................... 91
6.2.10.2 Check results within three days ....................................................................................... 97
6.2.11 Sample recheck ..................................................................................................................................... 100
6.2.12 Analyze complete ................................................................................................................................. 103
CS-T240 Operation Manual
Ambient condition:
──Ambient temperature: 15℃~32℃,suitable temperature:18℃~25℃;
──Relative Humidity: 40%~85%;
Relative humidity: 40%~85% Appearance dimension:
Chemistry Analyzer dimension: 998×752×517mm(length×width×height);
With cabinet: 998×752×1142mm(length×width×height);
Output power: 650VA Weight: About 120Kg
②
③
⑥
④ ⑤
①
②
③
④ ⑤
① Cooling Indicator ②Analysis Indicator ③Power Switch ④ Electrical outlet ⑤
Analysis Switch
Operating system is composed of mainframe, 17 inch CRT display monitor, keyboard, mouse and printer.
Mainframe: Windows XP system
Special applied software and database.
Computer configuration: CPU basic frequency ≥2.8GHz , hard disk≥160 G ,
Memory≥1G,with RS-232 serial port、internet port and USB interface
with RS-232 serial interface, website interface and USB interface.
Display monitor: Display all kinds of form, curve and test data of CS-T240 software.
Keyboard : Operation control and data input.
Mouse: Carry out software operation
Printer : Print out test data and chart.
1.4.2 Analytical system
Analytical system is composed of sample reagent disk, sample reagent pipetting mechanism, reagent disk,
stirring mechanism, cooling system, rinsing mechanism, optical system etc.
! Warning:
△
Do not touch sample reagent cover when the instrument is running, or it may cause body injury or instrument
damage.
① ② ③ ④ ⑤ ⑥ ⑦
①Sample reagent disk cover ② Disk cover lock knob ③sample tube ④sample reagent
disk handle ⑤ Disk cover detection switch ⑥Inner reagent bottle ⑦Outer reagent bottle
⑧Disk Lock Buckle ⑨Disk-oriented pin
Figure 1-4 Sample reagent disk
CS-T240 Operation Manual
(1) Function
Sample reagent disk is used for sample and reagent bottle placing. Place the containers ( standard cup, micro
cup , test tube) which contain calibrator, sample, control on the sample position, and then place the reagent,
CS-anti-bacterial phosphorus-free detergent on the reagent positon, the disk will send them to the sampling
position in the sampling mechanism.
Cooling system provide cooling condition for sample reagent disk to facilitate low-temperature reagent storage.
Refrigerated warehouse with a bar code reader window, and can scan the barcode of outer reagent and sample.
(2) Specifications
The reagent and sample share one disk, totally 66 positions. User-defined proportion of reagent position and
sample position.(The maximum reagent position is 42,the minimum is 6), no. 45 positon should be
CS-anti-bacterial phosphorus-free detergent
Reagent bottle volume:20ml、70ml、100ml。
Sample cup:standard cup, micro-cup, test tube.
(3) Movement
At Power on: it turns counterclockwise to move No.1 position to the pipetting mechanism sucking positon.
At analysis: At the beginning of analysis, sample disk makes the same movement as ―power on‖. During
analysis, sample disk turns to the direction allowing a quicker access.
At resetting: Make the same movement as at ―power on‖.
(4) Dismounting
The two locking buckle is used for two fix the plate. In dismounting, release the lock buckle fisrst, be sure
to set the position port matching with the guide pin.
Be sure to secure the cooling unit lid on the inner track, The outer track can be demounted without
removing the inner track.
Note: The instrument will issue alarm when the cover is opened under the condition of standby or testing.
Under standby, the instrument will carry out reagent horizontal scan.
(5) Action check
Single-click ― maintenance‖ key, select ― mechanism operation checkup‖, input the check times,
single-click ― Execute ‖ button. If abnormality exits,instrument will issued alarm.
! Warning
△
● Make sure that the sample reagent disk cover is well covered when the instrument is running.
CS-T240 Operation Manual
① ② ③ ④
(1) Function
Assmilates a specified amount of sample from sample container and a specified amount of reagent from
reagent container, and put them into the reaction cuvette.The pipetting probe is also a liquid level sensor.
Calculate the left reagent volume through the decrease distance of the probe. The left reagent volume will
be displayed in “reagent information‖form.
(2) Specification
Sample setting volume: 3~50ul, set in 0.1ul stepping.
Reagent setting volume: 10~450ul, set in 1ul stepping.
In sample pre-dilution, the specified amount of purified water from the inner wall of pipetting probe will be
added into reaction cuvette.The diluent volume is 10~450ul.
Important Notice: residual reagent volume、remaining tests is calculated upon(Setting amount + residual).
(3) Movement
At power on: The sample probe comes over above the reaction cuvette, and then returns above the
sample probe rinse trough.
At analysis: The probe moves follow the sequence of sample cup, reagent bottle, reaction cuvette,probe
rinsing bath.
At resetting: Makes the same movement as at power on.
(4) Automatic rinsing
After reagent pipetting, assimilate CS-anti-bacterial phosphorus-free detergent from the 45th position of
sample reagent disk. And pipet them into the reaction cuvette, and then return to pipetting probe washing
tank to wash the inner and outer wall. Adding detergent for 3 times, totally 1.05ml.
(5) Operation check
Single-click the ― System Maintenance‖ key, select ― mechanism operation checkup‖, and input the check
times. Click ―Execute ―. If abnormality exists, instrument will issue alarm.
CS-T240 Operation Manual
! Warning:
△
● Please don’t touch the lid of the reaction disk when running, or it may cause body injury and instrument
damage.
① ② ③ ④ ⑤ ⑥
① Reaction cuvette rinsing unit ② Reaction disk ③Reaction disk fixed knob
④Cup holder fixing screw ⑤Guide pin and guide hole ⑥Reaction cup component handle
Fix the reaction cuvette to the rotating reaction disk with screw, the reaction liquid reacts at 37 ℃ reaction
tank and conduct absorbance measurement in the rotation.
(2) Specifications
Reaction cuvette No.:20/unit×6 unit,totally 120 reaction cuvettes.
Light path :6mm
Reaction cuvette material:optical plastic
(3) Movement
Usually counter clockwise rotation.
At power on: Rotate, stop at the starting position. No. 1 reaction cuvette is under the first cleaning nozzle.
At analysis: Initial operation is the same as at power on. And then add two reaction cuvettes after one
circle (122 reaction cuvettes). Repeat this action process. It takes about 18 seconds to rotate a
circle
At resetting: Make sure same as at power on.
(4) Reaction cuvette cleaning
Place a anti-bacterial phosphate-free detergent bottle at 45 position of sample reagent disk. Open the reagent
bottle cover and conduct "reaction cuvette cleansing" in the "system maintenance" form, all of the reaction
cuvette can be cleaned. However, due to automatic cleaning by using CS-alkaline detergent in CS-alkaline
detergent box of the working analyzer, everyday maintenance do not needed.
(5) Operation check
Single-click the ― Maintenance‖ key, select ― mechanism operation checkup‖, and input the check times.
Click ― Execute―. If abnormality exists, instrument will issue alarm.
(6)Mounting/ Dismounting
Reaction disk: First remove the reaction cuvette cleaning unit of the reaction disk(top), then screw the
central knob of the reaction cuvette,the reaction disk can be lifted. In the installation, matching the guide
hole with the guide pin of the reaction disk seat, and then tighten the fixed knob.
Reaction cup: remove the screw of the reaction cuvette, grasp the handle of reaction cuvette component
upward, the reaction cuvette can be removed from the reaction disk.
Note: Place the removed the reaction cuvette in pure water to save. In addition, if the analyzer has been
shutdown for at least 3 days, reaction cuvette need to be removed, and placed in pure water.
! Warning:
△
● Keep the cleanness of purified water in incubation bath, or it may effect the test precision.
● When instrument startup or rinsing incubation bath, make sure there is enough CS-anti-bacterial
phosphor-free detergent at No.45 position.
(1) Function
Keep the reaction solution in the reaction cuvette at a constant temperature.
(2) Operation
At power on: Automatic exchanges the constant temperature water once, the CS-anti-bacterial phosphor-free
detergent in position No.45 of both reagent disks is added in incubation bath.
At analysis: Incubation bath water is circulating. Instrument may automatically supply water when water
shortage comes in operation process.
Exchange water: In ―maintenance‖ window, select ―rinsing incubation bath‖, and then the constant
temperature water may exchange, and then add 2.7ml CS anti-bacterial phosphor-free detergent in
incubation bath water.
Note: After running for 24 hours, instrument may require ―incubation bath water exchange‖, please carry
out ―Rinsing incubation bath‖.
CS-T240 Operation Manual
! Warning:
△
● Please don’t touch stirring mechanism when operate, or it may cause body injury or instrument damage.
② ③
① mixer ② mixer rinsing bath ③mixer arm
(1) Function
Stirring the reaction solution in each reaction cuvette.
(2) Operation
At power on: Move to the side of reaction cuvette and then stops above the rinsing bath, move to the side
of reaction cuvette again, and then stops above the rinsing bath.
At analysis:The mechanism descends, rotates, risees and stops between two locations: reaction cuvette and
stirring rod rinsing bath.
Stirring is carried out after each addition of reagent.
(3) Automatic rinsing
Automatic rinsing of mixer: when mixer descends into mixer rinsing trough, mechanism may automatically
rotates and washes the mixer with purified water.
Sampling finishing: mixer is stirring in reaction cuvette in which detergent is added, thus rinse the mixer.
(4) Operation check
Single-click the ―maintenance‖ key, select ―mechanism operation check‖, and input the check times. Click
―Execute‖. If abnormality exists, instrument will issue alarm.
● Avoid directly contact with body, or it may cause infection. Please adopt protective measure. In case of skin
contact, flush the area with water, rinse immediately with plenty of water and seek medical advice.
(1) Function
Eliminates the reaction solution, rinse the reaction cuvette,injects and eliminates purified water which
used for test cell blank
(2) Rinsing composition of nozzle
Reaction Disk
Figure 1-8 Rinsing nozzles arrange
CS-T240 Operation Manual
(3) Operation
Power on:First descend by about 5mm and then rise .
Analysis :According to the direction of figure1-8―Rinsing nozzles arrange‖ to conduct reaction cuvette
cleaning and bottle blank measurement.
(4) Operation check
Single-click the ―maintenance‖ key, select ―mechanism operation check‖, and input the check times. Click
―Execute‖. If abnormality exists, instrument will issue alarm.
(5)Mounting/ Dismounting
Unscrew the screw counter-clockwise, lift the cleaning unit; in installation, matching the cleaning unit with
seat pin, and tighten the screws.
1.4.2.7 Reagent cooling system
(1) Function
When the reaction disk rotates, the absorbance of purified water or reaction solution is measured in each
reaction cuvette. As figure1-9 shows.
(2) Specifications
Carry out photometry with dual-wavelength or single-wavelength at wavelengths: 340 nm,380 nm,405
nm,450 nm,480 nm,505 nm,546 nm,570 nm,600 nm,660 nm,700 nm,750nm.
Wavelength accuracy: ±2nm
Measuring range: 0 -3.3 Abs
Spectral bandwidth: FHW 8 to 10nm
Detector: Silicon photodiode
Symbol Meaning
AC symbol
Storage at
Batch code
Use by
Serial number
Measurement Control
Date of Manufacture
Manufacture by
Grounding terminal
Figure 1-1
CS-T240 Operation Manual
Chapter 2 Measuring Principle
The measuring principle is composed of mechanism movement principle and analyzing assay.
Reaction cuvette
rinsing unit
Reaction disk
Reagent resetting
sample
adding
position
Reference position
No.
Figure 2-1
2.1.1 Operation flow
2.1.1.1Rinsing unit
Rinse from the first reaction cuvette (Rinse each position twice), the reaction disk will pause after rotating
38 reaction cuvette position, and then pause after two reaction cuvette position, and then pause again after two
reaction cuvette position, then stop after 80 reaction cuvette position. In the cleaning process, due to reaction
cuvette go through metering section, so the cell blank can be tested. The cell blank value can be used as the
benchmark value of absorbance (Abs). The detergent in the reaction cuvette will be sucked by detergent nozzle .
2.1.1.2 Sample and reagent adding unit
Sample and reagent adding share one disk ,one set of injector, the order is reagent and then sample. When
the reagent control panel get adding sample command, the sample reagent disk will move to the corresponding
reagent position, the probe will move to the up direction of the reagent. And the then, the probe conduct rinsing,
the sample reagent disk will move to the corresponding sample position, the probe move to above the sample. At
this point, the probe will wait above the sample cup. When the reaction disk pause after two reaction cuvette
CS-T240 Operation Manual
position, the probe will move to the reaction cuvette position to add sample, and the move to the rinsing position.
In addition, in reagent 1adding, reagent and sample is added at the same time. In reagent 2 adding, only
reagent is added.
Probe unit will move to the reagent or sample position upon receipt of an sample adding command, the
probe will move to the top of reagent or sample. With the liquid level sensor, the probe tip will stop after enter
into the sample. And then the probe will move to above the reaction cuvette, to discharge reagent or sample.
Then the probe will conduct rinsing.
38 、 2 、 2 、 80
poisition , reaction
cuvette 3 begin to
rinse
Original position
Reaction cuvette 1
begin to rinse
⑤ Sampling:
Sampling is conducted after10 times’ rinsing. After position resetting, reaction cuvette 1 is at the position of
71,after 10 times’ rinsing, reaction cuvette 1is at the position of sampling position. As show in figure 2-5
Figture 2-4
Therefore, the sequence of sampling is the same as reaction cuvette sequence
Odd No.:1 → 3 → 5 → 7 → 9 → …… →117 → 119
Even No.:2 → 4 → 6 → 8 → 10 → …… → 118 → 120
The first 10 times rinsing need 5×18s=90s,and then sampling after 18s. The sampling time is after 2
positions. In the process of continuous sampling,the sample reagent probe need to finish sucking, sampling,
rinsing in 18 seconds. The individual dilution steps should be added. When the ISE functioned added, a sampling
process of sucking and ISE position is needed.
21
CS-T240 Operation Manual
Figure 2-5
Therefore,reagent 1sampling and mix is donein one rotate(18s).
Reagent 2 sampling and mix:
Reatgent 2 sampling is after the beginning of test. Reaction disk begin to sample at the 25 th circle, at 38+
2+2+80 reaction cuvette.The mix of reagent 2is after reagent 2 adding.
1 I
A = lg( )= lg( 0 )= ε b c
T It
CS-T240 Operation Manual
t :time(minute)b
2-point L– m–0–0 B1 + B 2 + B 3 ( AM + AM −1 ) − k( AL + AL −1 ) etween
Assay 1≤ L<m ≤ 49 photometry point
3 2
L,m
AM + AM −1 AL + AL −1
2-point L– m–0-0 B1 + B 2 + B 3 −
rate assay 1≤ L<M ≤ 49 2 2
3
t
L– m–0-0
Rate A B1 + B 2 + B 3 △A(M-L)
1 ≤ L<m ≤ 49
Assay
L +2<m 3
23
CS-T240 Operation Manual
Explanation of symbols:
S :Sample volume
Rj,Ri a: No. of reagents without correction
b: No. of reagents with correction
Note 1: The 21 th Photometric point won’t be stirred after adding reagent 2. Stirred when the reaction disk
pauses after rotates one circle plus 2 pitches plus 80 more pitches.
Note 2: liquid in the reaction cuvette should be more than or equal to 150 ul, less than or equal to 550ul.
Note 3: Do input 0 if the photometric point is not used.
(1) 1-point Assay
Endpoint assay in which absorbance is measured at a designated photometric point (specific time point when
reaction reach balance) after addition of sample and reagent. Figure 2-6 explains the 1-point assay.
Measurement is made twice at different measurement points (The two point are neither measured initial nor
endpoint) to determine the change in absorbance per minute in order to calculate sample concentration. For
check of reaction limit level, refer to Figure 2-8:
C X = {K × ( AX − B ) + C1 }× IFA + IFB
CS-T240 Operation Manual
C X = {K × ( A− B ) + C1 }× IFA + IFB
Cx: Concentration of standby sample
AX: Absorbance or change in absorbance per minute
IFA and IFB: Instrument constants, representing slope and intercept.
(d) Applicable assay
1-point assay, 2-point rate assay, 2-point assay, rate A assay
(3) Multi-point linearity
Blank (or standard 1) and standard samples (standard 2 and standards 6) are measured and linear working
curve. Figure 2-12 explains the linear method.
Absorbance
Concentration
C X = (C + C1 ) × IFA + IFB
K
AX = B +
1 + aC)
1 ⎧ K − ( A X − B) ⎫
C= ×⎨ ⎬
a ⎩ AX − B ⎭
Cx: Concentration of standby sample
C1: Blank concentration.
AX: Absorbance of sample or its change per minute.
K: Constants in approximation formula. The more Cx approaches ∞, AX approaches B
When K<0, AX≤B+K or K>0,When AX≥B+K,C=C1
IFA,IFB Instrument constants, representing slope and intercept.
(d) Calculation of SD value
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 3
(N=3~6,j=1or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
(e) Applicable assay
1-point assay, 2-point rate assay, 2-point assay, Rate A assay.
(5) Logit-log4P (Non-linearity method 2)
It is applied to a working curve in which the absorbance converges as the concentration increases. Figure
2-104explains the non-linearity method.
1 ⎧ K − ( AX − B) ⎫
C = b× ×⎨ ⎬
a ⎩ AX − B ⎭
Cx: Concentration of standby sample
C1: Blank concentration.
AX: Absorbance of sample or its change per minute.
K: Constants in approximation formula. The more Cx approaches ∞, AX approaches B
When K<0, AX≤B+K or when K>0,AX≥B+K C1=0
IFA,IFB Instrument constants, representing slope and intercept.
(d) Calculation of SD value
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=4~6,j=1or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
(e) Applicable assay
1-point assay, 2-point rate assay, 2-point assay, Rate A assay.
(6) Logit-log5P (Non-linear method 3)
There is no distinct difference between the working curves prepared by the non-linear method 2 and 3.
However, in some cases, the non-linear method 3 allows more accurate approximation this method has one
more calculation parameter than the non-linear method 2. Figure 2-15 explains the non-linear method 3.
CS-T240 Operation Manual
Absorbance
Concentration
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=5~6,j=1 or 2)
CS-T240 Operation Manual
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Au is 12 at maximum
(e) Applicable assay
1-point assay, 2-point rate assay, 2-point assay, Rate A assay.
(7) Exponential function method (Non-linear method)
Unlike non-linear methods 1,2 and 3.Exponetial function method prepares a working curve in which the
absorbance disperses as the concentration increases. Figure 2-16 explains the exponential function method.
Absorbance
Concentration
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 5
(N=5~6,j=1 or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
(e) Applicable assay
1-point Assay, 2-point Rate assay, 2-point Assay, Rate A assay.
(8) Spline function method (Non-linear method)
In this method, line is connected in each section so as to form a curve as a whole. Since each section is
smoothed including the error in measured value, more accurate approximation is possible than the polygonal
line approximation. Figure 2-17 explains this method.
Absorbance
Concentration
∑∑ (A )
N 2
, 2
IJ − A1
i =1 j =1
SD =
2N − 4
(N=5~6,j=1 or 2)
(Aij-Ai’): Difference between approximate absorbance Ai’ and measured value Aij or A12. Each standard
sample is measured in duplicate so the number as measurement points Aij is 12 at maximum
(e) Applicable assay
1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A Assay.
Absorbance
Concentration
S1 ABS is the average nalue of two measure value (absorbance or it’s change)
C 2 − C1
K=
A2 − B
B: Absorbance or it’s change rate (1)
A2: Absorbance or it’s change rate (2)
C1:calibrator (1) concentration (input value)
C2: calibrator (2) concentration (input value)
Same as calibrator(2)~calibrator(6),calculate K2、K3、K4、K5。
(c) Calculation of concentration.
C X = {K N × ( AX − AN ) + CN }× IFA + IFB
(e) Applicable assay
1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A assay.
(10)Isozyme Method
In a sample in which 2 different isozymes coexist, a reagent containing inhibitor may fail to completely
suppress the activity of either isozyme alone. In this case, isozyme activity is determined from total activity
and activity residual rate. Each working curve for total activity and isozyme activity are prepared by using 2
channels. If the activity of a specific isozyme of the coexistent two can be suppressed completely by using
monoclonal antibody, etc. this calibration method is unnecessary. As figure 2-19, 2-20 shows:
Figure 2-2
S1 to S4 are calibrator code numbers of calibrator 1 to calibrator 4 respectively. Enter the same calibrator
code number in both channels for each of calibrator 1,3,4. place the calibrator of isozyme M at position
Calibrator 3 and that of isozyme N at position Calibrator (4). It is unnecessary to enter the concentrations of
calibrators 3,4 of isozyme P,Q, and it is unnecessary to enter the concentrations and positions of calibrator 2
of isozyme Q. The isozyme Q item name must be specified when that of isozyme P is set, or the parameter
can not be saved and alarm occurs, and instrument stops testing. However, the item name of isozyme P is not
needed to be specified at isozyme Q side.
Note: The item name of isozyme Q should be specified while setting that of isozyme P.
(c) Calculating K value
C 2 − C1
A −B
K= 2
B: Absorbance of blank (standard 1) or its change per minute
A2: Absorbance of calibrator F (standard 2) or its change per minute
C1: Concentration of blank (calibrator 1 )
CS-T240 Operation Manual
α=
{K × ( A , -B )+ C }× IFA + IFB
3 1
β=
{ -B )+ C } × IFA + IFB
K ×(A , 4 1
{K × ( A -B ) + C } × IFA + IFB
4 1
A3, -B,
α=
A3-B
A4, -B,
β=
A4-B
(g) Calculation of isozyme M activity CM
CM’=α×CM+β×CN
CM , − α × CM
C M = C F- C N = C F-
β
β×CM=β×CF- CM’+α×CM
(α-β)×CM= CM’-β×CF
CM , − β × CF
CM =
(α − β )
(h) Applicable assay
1-Point Assay, 2-Point Rate Assay, 2-Point Assay, Rate A Assay.
Note: Suggest operator use 6 or 5 calibrators in Logit-log5P, exponential,spline non-linear calibration.
2.2.3 Calibration types
According to the number of calibration, there are four types of calibration. Blank calibration: Only reagent
blank (calibrator 1) is calibrated. Only one calibrator other than the reagent blank is calibrated. Reagent
blank and a single calibrator are calibrated. All standard solutions specified on the Chemistry Parameters
screen are calibrated. These are selectively usable so as to meet your analytical purpose. Each calibration is
explained below.
CS-T240 Operation Manual
Only reagent blank (calibrator 1) is calibrated. Table 2-3 lists the calculation method for each calibration type.
(a) S1ABS calculation.
In calibration, a warning –level alarm is issued if the measured absorbance of blank is not within the input
range of standard 1 absorbance. In this case, the result of measurement and alarm (S1ABS) are printed out.
To avoid check, enter-―-3.3~3.3‖.
(2) Discrete check
A warning level alarm is issued if the difference of the two times measured absorbance value is larger than
the set value. In calibration, each Calibrator (include reagent blank: Calibrator 1) is tested twice.
To avoid the check, enter‚3.3‛.
Discrete check is performed by the below formula:
If the difference of standard absorbance (average between 2 times measure) from Max. concentration
standard absorbance (sensitivity) exceeds the permissible absorbance sensitivity value (Sensitivity Limit), a
warning- level alarm is issued. In this case, alarm mark is printed out together with the result of
measurement.
The working curve of the alarmed analytical item will be renewed, and the Kvalue won’t be renewed. To
avoid the check, enter-―0‖.
Check the permissible sensitivity by the below formula:
Acalibratior(N)-Acalibration (1) > sensitivity value (ABS)
(4) K factor check
If the fluctuation in factor K value between previous calibration and current calibration is 20% ,a warning
level alarm is issued. The working curve and K factor will be renew and testing can be carried out. Make
sure check the reason of alarm.
Check the K factor by the below formula:
K this − Klast
×100%
(K last + Kthis ) / 2 ≤20%
In calibration, a warning –level alarm is issued if the difference between the calculated absorbance and
tested absorbance has exceeded the drift rate set value. The working curve and K factor will be renew and
testing can be carried out. Make sure to check the reason of alarm. To avoid the check, enter―3.3‖.
2.3.2 Absorbance limit
As to the Rate A assay and 2 point tate assay, correct data won’t be obtained when concentration or activity
exceeds the quantitative span. Thus, set the upper limit value and lower limit value of the absorbance, print
the alarm sign. Input the calibration value on the screen. To avoid the check, enter-0 (decrease) or 3.3
(increase).
When 4 or more than 4 tested absorbance value is not accord with the set value of reaction limit absorbance,
alarm is issued as figure 2-21shows:
CS-T240 Operation Manual
Time
Figure 2-21
Figure 2-22
If not selected, even if input value in textbox, linearity check is not carried out.
(1) When number of measurement points (N) more than 9 (N>9)
Linearity is checked by dividing the difference in absorbance change between the first and last 6
measurement points by the average absorbance change for all. If the value thus obtained is beyond the limit
linearity value, alarm is printed out together with the result of measurement as figure 2-23 shows:
ΔAf − ΔAb
ΔA
×100>Limit linearity value(%)
CS-T240 Operation Manual
Photometry point
Time
ΔAf ,− ΔAb,
ΔA, ×100>Limit linearity value(%)
Photometry point
Time
Figure 2-24 Linearity check(4≤N≤8)
No linearity check is needed if the following happens.
1) The No. of photometric points under absorbance of reaction limits is less than 3.(N<3)
2) ΔA/ΔA,≤0.006 or∣ΔAf-ΔAb∣≤0.006
2.3.4 Prozone check
In immunoreaction, working curve descends if antigen concentration is abnormally high beyond the suitable
range (prozone area). This is called prozone or zone phenomenon.
This instrument can check whether concentration is in the absorbance decreasing range (post zone). For
prozone check, the following 2 methods are available: antigen readdition method in 1-point assay with
prozone check and reaction rate method in 2-point assay. To Avoid check, input ―-3.3 lower limit‖ in
―checkup value‖ function box in ―analyze parameter‖ menu.
Antigen supplement method:
Take 1-point essay for example, measure reagent 1, and take it’s value as reference value. Replace reagent
with serum diluent, which contain antigen, add 20ul. Compare the prozone limit value with absorbance
difference (before add reagent 2 and after add reagent 2).
CS-T240 Operation Manual
Input method:
Prozone check value (PC value): 【 】
Upper /lower limit: 【】
Analytical method: 【】
Photometric point: 【Q1】 【Q2】 【0】 【0】
1≤Q1≤Q2≤49
Aq 2 + A( q 2 − 1) Aq1 + A( q1 − 1)
PC = −k×
2 2
K=total liquid volume when test q1/total liquid volume when test q2
When 1≤Q1≤Q2≤16 or 17≤Q1≤Q2≤110, K=1.
AQ1,AQ2:absorbance of photometric point:Q1,Q2
Absorbance
Time
To make sure normal operation, install or initialize the CS-T240 only by authorized staffs from DIRUI
Company.
3.1 Installation requirement
There are two types of CS-T240 auto-chemistry analyzer, one is sold with a cabinet, the other without a
cabinet.The first one should be placed on the cabinet,the second one should be placed on smooth cabinet or
table, and then put it on the ground.But it should not be put directly on the ground. Before installation,
operator should check the space, power and environment requirement.
3.1.1 Space Requirement
To make sure the space of maintenance, please follow the instruction as below:
● Space between left (right) side of analyzer and the wall should ≥50cm
● Space between rear panel of analyzer and the wall should ≥50cm
● Space in front of analyzer should≥100cm
● Make sure there is enough space for waste device and purified water equipment.
3.1.2 Environment requirement
Operate or store the analyzer according to the following requirement:
● Working environment: 1 5 ℃ ~32 ℃
● Relative humidity: 4 0 % ~85 %
● Atmospheric pressure: 7 6 k P a ~106kP a
● Environment should with no dust, no mechanical vibration, no noise source and power interference
● Do not put the analyzer in the vicinity of brush motor, flicker fluorescent tube and other constant on-off
electrical equipment.
● Avoid direct sunlight, do not put the analyzer in front of heat source and wind source.
● The maximum sound 1m distance around the instrument is limited at 40dB when it is working.
3.1.3 Power requirement
Note: The unqualified environment may cause test value inaccuracy , analyzer damage and it is also harmful to
human body.
● Check the packing and appearance of the analyzer, if there is a damage, please contact DIRUI company
or local distributor.
Figure 3-1
CS-T240 Operation Manual
Figure 3-2
In Figure 3-2, click "Next" to pop up the selection form of software installation folder, and the software default
installation directory (full installation) is "C: \ Program Files \ DIRUI\CS-T240 automatic biochemical analyzer
application software \ ", as figure 3-3 shows below:
Figure 3-3
In figure 3-3,click ―change(C)…‖function key,change the install catalog of the software,and the
click―next‖function key,installation of the starting software,as shown in figure 3-4:
CS-T240 Operation Manual
Figure 3-4
Click ―install‖ button in figure 3-4, initialize the software as figure 3-5, 3-6 shows:
Figure 3-5
CS-T240 Operation Manual
Figure 3-6
Click ―finish‖ button in figure 3-6 to complete the installation process.After installation, the menu ―CS-T240
auto-chemistry analyzer software‖ can be shown on the screen .
3.3.2 Peripherial device connection
3.3.2.1 Connection of pure water inlet pipeline
Connect outlet of pure wate machine and pure water inlet of the analyzer.( ③ in figure1-2)
3.3.2.2 Connection of waste liquid outlet pipeline
Connect one end of the concentrated waste pipeline taken with the analyzer with the concentrated waste liquid
outlet interface ④ in figure 1-2, and place the other end into the waste liquid collector.
Then, click the icon on computer screen or click ―start-up‖, then find the software CS-T240 in
―program‖ and click it, after that, enter ―system logging‖ window as the figure3-7 and 3-8 show.
CS-T240 Operation Manual
Figure 3-7
Figure 3-8
Input user name, password, click ―login‖ or ―enter‖ to get into the main menu of software, as figure 3-9 shows:
( Initial user name: 001, password: 001)
Note:If the inputted user name or password is wrong, wrong login will be displayed on the screen. If the
inputted user name or password is incorrect for continuous three times , the system will exit
automatically.
CS-T240 Operation Manual
Figure 3-9
After successfully login, the software show as offline state, browse menu, check alarm information, user
logout function can be used at this state.
a) Connection: in figure 3-9 interface, click ― ”, connecting will show, after success of access, the
status bar will display ―standby……‖. At this time, all operation and tests can be carried out.
After logging, the instrument partially connected or not connected the cable, click the ― ‖, the
screen will show as figure 3-10 shows.
Figure 3-10
If this shows on the screen, connect after get the power supply connected.
b) Exit system:
CS-T240 Operation Manual
In the window as figure 3-9 shows, click ― ‖, enter exit confirrm form figure 3-11 shows:
Figure 3-11
Click ―ok‖ in figure 3-11 to exit software.
Only exit system at off-line state. If analyzer is on-line, click ―off-line‖ button to exit system.
c)Logout user:under figure 3-9 interface,click shortcut key―Logout user‖ the screen will display as figure
3-8,the user can switch.
Note:
● In order to prevent data from being damaged or revised by other people, exiting software when doctor takes
a rest is strongly suggested. Periodically backup database in order to avoid data lose.
●. Input initial user name and password when first login, select ―user information‖ in ―management‖ menu, set
user name, password and access authority for next time login.
The analyzer will be in sleep status after 20 mins of power on.(waiting for the stability of power and
temperature).
3.3.4 Uninstallation of software
Method 1:If the uninstallation of applied software of auto-chemistry analyzer of CS-T240 is needed, please
enter ―addition or cancel program‖ in setting board, click ―delete‖ button, window ―addition or cancel
program‖ will pop up. Then window like figure 3-12 will show:
Figure 3-12
In the figure 3-12, click ―Yes‖ to complete the uninstallation.
Method 2: Click ―start‖ form, find ―Di Rui CS-T240 Auto-chemistry Analyzer‖, click‖uninstall CS-T240
Auto-chemistry Analyzer‖ can also uninstall that software. As show in figure 3-13:
CS-T240 Operation Manual
Figure 3-13
Figure 4-1
Figure 4-1
CS-T240 Operation Manual
(a) No cockle, no contamination one the label, no line deformity when stick barcode , otherwise the barcode
reader cannot correctly read the barcode.
(b) Stick barcode on correct place:
In order to obtain correct barcode, 15mm-20mm between cuvette bottom and barcode lower line is
required. Make sure barcode is on the outer side of sample disk when place on the test tube rack., as figure
4-2 show:
Figure 4-2
Note: The ―‖, ‘’,()are not allowed for the barcode, or it cannot be identified.
When the CODE39 is not capitalized, please add ―+‖ before the corresponding capitalized letters of the ID
code displayed on the screen and the report sheet printed after scanning.
Specifications:70ml、20ml、100ml
4.1.6 Reagent barcode using requirement
(a) Barcode type: code 128(17)
(b) Borcode label size:its width should be within 12mm~15mm,the length should be no more than
40mm(Figure 4-1)。
(c) Blankness between start and finish should be within 3mm when cutting barcode as figure 4-1 show.
4.1.7 Reagent barcode stick requirement
(a) Stick the barcode with no cockle, make sure there is no deformity on barcode line. Contamination is not
allowed on lable, or barcode cannot be read correctly.
(b) Stick the barcode on the correct place.
Blankness between bottle bottom and barcode should be within 15mm-25mm, thus barcode could be
read correctly.
4.1.8 The rule of reagent barcode
Barcode can be made by user according to their needs with the rule listed in the table 4-2:
CS-T240 Operation Manual
Numeric or alphanumeric
15~17 Parity bit
(automatically generated)
Figure 4-2
Reagent barcode information can only be read by barcode reader, the information will coupling with
chemistry parameter which stored in instrument, this process is called reagent registry information. Reagent
information registration could check reagent position on reagent disk.
The read information could be showed in ―reagent information‖ menu as ―disc No.‖, ―position‖, ―reagent
name‖, ―reagent type‖.
CS-T240 Operation Manual
Figure 5-1
Description:
: Represent display status: stand-by, testing, emergence stop, sampling stop, maintenance
operation, sleeping mode.
: Communication monitor mark. When communication is under normal status, the icon turns blue,
when communication abnormal, the icon turns black. Once the icon color turns from blue to black, that
indicate communication failure.
: Alarm icon. This icon occurs in status bar when alarm is issued. Click the icon, start alarm checkup,
solve the problem according to the remedy.
Figure 5-2
(c) Working space: According to the function selected by the user, corresponding function interface will show up,
Single click system setting in keypad area, as shown in figure 5-3
CS-T240 Operation Manual
Figure 5-3
(d) Hint bar: instruct user how to use software, hint the input range, input method, operation error, as figure
5-4 show:
Figure 5-4
(e) Shortcut key space: for convenience use, as figure 5-5 shows.Some commonly used function key is set in
shortcut keypad area.
Figure 5-5
Figure 5-6
Pop-up window as a model window, such as to turn off, click "Confirm" button.
5.1.2 Keyboard function
Test result delete result, result query, review, preview, audit, print,
batch print, batch print, reaction curve
Reagent Info Manual registry, remove reagent info, barcode scan, reagent level
System setup Report format Report Info, print sequence, print format setup
Figure 5-7
CS-T240 Operation Manual
Figure 5-8
Modeless form: other menu can be opened when modeless menu open like form ―check th result‖As figure 5-9
show:
CS-T240 Operation Manual
Figure 5-9
Figure 5-10
Figure 5-11
Check box: more than two functions can be chosen at the same time, for instance ―Calibration Register‖
interface,multiple calibration items can be selected at the same time, as figure 5-12 shows:
Figure 5-12
±2nm
precision
e
Reagent position 、 The reagent and sample share one disk, totally 66 positons, User
sample position defined proportion of reagent position and sample position. (The
maximum reagent position is 42, the minimum is 6.)
Sample type Serum,plasma, urine, cerebrospinal fluid, ascitic fluid
Reagent+Sample System
150~550ul
volume
Light source 20W/12V long life quartz halogen lamp
Absorbance range 0~3.3ABS
Quality control QC interval, Montly QC
Automatic rinsing Rinsing cuvette automatically.
Stirring mechanism Singleness stirring after reagent adding
port Standard RS-232
Data Printer stylus printer,support user-defined mode
system
Connecting LIS/HIS
Connect with LIS/HIS system
system
Weight About 120Kg
Integrated
Equipment size 998 mm×752 mm×517mm(length×width×height)
system Power consumption
650VA
(VA)
Table 5-1
Note: (1) It may affect the accuracy of test results when the total volume of reaction solution is150 ul.
(2) According to different test condition,sometimes instrument processing capability is lower than 200 test/hour.
Test condition Processing capability lower degree(estimated)
Table 5-2
CS-T240 Operation Manual
6.1Operation overview
Table 6.1 shows the operation flow. For detailed operation, please refer to 6.2.
Consult
Operation step Form / key Operation
index
1. Check before operation Check before turning on power 6.2.1
2. Connect water unit and CS-T240 Logging in Open water faucet, turn on the power of purified
Auto-chemistry Analyzer power water unit and CS-T240, input operator’s ID No.
6.2.2
Log Software and password.
3. Check instrument status Please refer to chapter 13 Alarm and Handling of
1) Check alarm Alarm information Instrument Failure
2) Check light quantity of photometer System maintenance Execute light quantity checkup, check if test value
3) Check cell blank is within regular range
6.2.3
4)Check temperature of incubation bath. System maintenance Execute cuvette blank check, check if test value is
Status bar witin regular range
Check if temperature of incubation bath is within
37.0℃±0.1℃
4. Check analytical conditions Add chemistry item
1) Item adding System setup Check chemistry parameter
2) Chemistry parameter input and System setup Check calibration curve and K factor 6.2.4
confirm Calibration info
3) Check K value
5. Reagent preparation (reagent info)1) Check reagent remaining volume and remaining
1)Check reagent residuel volume Reagent info test times. Place the reagent at the corresponding
6.2.5
2) Preparation for photometry item and position.
preparation for ISE item Reagent info
6. Setup of calibration and control item Calibration info Check item name of calibration analysis
6.2.6
Quality control Check item name of QC
7. Sample registration and test Sample registration Conducting single or batch sample registration,
single emergency sample registration, patient 6.2.7
information editing, modification and deletion.
8. Test Place sample on the sample+reagent disk
Sample Test Start test Implement "Start test" to carry out analysis
(1) sample preparation
(2) Send test instruct Place control on the sample+reagent disk
QC Test Implement "QC" ->‖QC test‖to carry out analysis
(1)Control preparation QC test 6.2.8
(2) Send test instruct
Calibration Test
Place calibrator on the sample+reagent disk
(1)Calibrator preparation Calibration test
Implement "Calibration infomation"
(2) Send test instruct
->‖calibration test‖to carry out analysis
9. Testing Process System moniter Moniter the instrument when testing.
(1)System moniter Sampling Carry out sample edit when testing and click start
(2) Sampling stop/continue stop/continue analyze. 6.2.9
(3) Emergence stop Emergence stop
(4) Sample superaddition. Sample registry
10. Check test result (result data) Result To search, amend, delete test result, check, print
6.2.10
sample reaction curve
11. Recheck sample test Test result Check the recheck item condition. In ―recheck‖
form, click ―start recheck‖, and send the recheck 6.2.11
instruction.
12 Completion of analysis Test result Check, confirm and printout the test result. 6.2.12
1) Recheck result System management Periodical backup database, one week is
2) Database backup suggested.
3) System dormancy Instrument could automatic start at specified time
4) Turn off instrument after setting
5) Preparation of next operation Cutting instruments, computers, pure water
supply device power for next time use.
6.2Detailed operation
! Warning :
△
CS serial detergent is corrosive liquid. In case of skin contact, flush the area with water, rinse immediately
with plenty of water and seek medical advice.
6.2.2 Power on and software login
(a)Turn on the power of purified water, and open the water faucet.The purified water device should use one
power supply alone.
(b)Turn on the power supply of CS-T240.The main power switch lies in the left lower side of instrument.
Turn on the switch when there are reagents in sample reagent disk, therefore the cooling system may under
normal working condition. The power of analyzed part lies in the left lower side of instrument.