Intelligent Illumination Chamber

GTOP series

User Manual

Please read this manual carefully before use


Dear user:
Thank you for choosing our products to use , Please save the user manual
properly. As the product is continuously upgraded and perfected, please be
aware that if the manual is not completely consistent with the product.

The following symbols appear in specification or product, it means that:

1
CONTENTS

I. Introduction......................................................................................................................................3
II. Technical indicators........................................................................................................................ 3
III. Instrument structure.......................................................................................................................3
IV. Operation and use.......................................................................................................................... 4
V. Precautions and troubleshooting.....................................................................................................9
VI. Warranty information.................................................................................................................. 12
VII. List of accessories......................................................................................................................12
VIII. Appendix...................................................................................................................................13

2
I. Introduction

GTOP Intelligent Illumination Incubator is a biological incubation experimentation instrument,


which adopts microcomputer automatic control system, programmable control mode, and
intermitting cooling.
Through setting the multiple parameters, it is able to mimic a natural environment. It has
illumination, temperature –constant, day and night shift automatically functioning. It is able to
meet various biological incubation tests.
Our incubator is able to apply to various areas such as seed germinate & breeding, plant group
foster and planting, animalcule incubation, breeding of insects and small animals, anti-oxidation
experiment of medicines and environmental experiments of articles; and it is especially
applicable for the study of Bioengineering, Medical study, agricultural, Aquatic, Animal
husbandry etc.

II. Technical indicators

● Temperature range: 0~50℃


●Temperature fluctuation: ±0.5℃
●Temperature uniformity: ±1℃
●Cooling power: 190W-450W(different models have different power)
●Compressor dynamic delay protection time: about 3 minutes
●Working mode: continuous operation
●Environmental conditions: temperature 0~35°C, humidity ≤85%RH
●Power supply: 220V, 50HZ

III. Instrument structure

1. Chamber structure
The chamber is composed of a top cover, a working chamber, a cooling and heating room and a
bottom chamber. The top cover is equipped with a computer controller strong electroplax, and the
top cover panel is equipped with a computer controller weak board, which includes a liquid crystal
display, an indicator light and an operation button. In addition, the top is also equipped with a
power switch, network interface, and fuse. There are LED panels on the side of the cabinet, and a
temperature sensor is installed in the cabinet. The cooling and heating room is located below the
working room, and the left and right (or front and rear) are respectively connected to the entrance
and exit of the working chamber of the studio. The indoor is equipped with an evaporator, an
anti-condensation blunt shunt, an electric heating tube and two circulating fans to work. The room
is heated or cooled, and the thermistor is equipped with a thermistor for overheat protection and a

3
bimetal temperature protection switch. The bottom chamber is equipped with a compressor, a
condenser, a cooling fan, an instrument incoming junction chamber and a fuse.

2. The panel structure is as follows:

3. Chamber description:

IV. Operation and use

1. Key Description:

(1) Move the menu bar during menu operation;


(2) Check the icon you want to work with;
(3) Adjust parameters when parameter setting.

(1) When parameter setting or menu operation, select the parameter to be modified;
(2) Check the icon you want to work with.

4
2. Specific operation
Turn on the power, close the power switch, The startup screen displays 10 seconds to
automatically jump to [Function]

Detailed instructions for each function in the menu are given below:
(1) Continue run
Just press the [OK] button during startup or running, then enter the screen as shown below or wait

5
for 10 seconds (as shown below)

In the [Continue] interface, press [OK] to jump to the [In-Progress] (as shown below)

Display the real-time situation that is currently running. The [TimeLeft] circle flashes to indicate
the countdown operation; [curPd] indicates the running time period; [TotPd] is how many working
hours are set; [Temp.],[Lux] indicates the real-time temperature and illumination of the current

operation. Press the key to observe the ambient and protection temperatures under
current operation and the current time.

(2) Setup process

In the [Function] interface, you can use the key to move the arrow. The arrow stays
that you can press [OK] to enter, press the [OK] button to enter the [Proc Set] interface, and then
you can enter the temperature, Lux, and work hour. Time and multiple time periods can be set.

6
In this interface, you can press the number keys to set the required parameters one by one. The
flashing position of the cursor is the current digital input. The parameters that can be changed
include temperature and running time. After setting the required parameters, press [OK] button.
Save the set parameters (as shown below)
Note: The maximum time cannot exceed 999 hours. If [-] appears when the temperature is set,
please change the [+] number when you re-enter any number from 1-9. Because this controller can
set 99 time slots.

(3) Shift time

The function of shifting time is to modify the set working time of the process. For the
modification of the current running time, it is necessary to change from [Pd. S/W] to [Process
switching] (as shown in the figure below).

(4) Process switching

7
The function of the process switching is to switch from the current working period process to the
other working time process, and if you want to switch to the required process, you need to press
[OK] from the [Proc S/W] to enter the process switching interface. (As shown below)

In this interface, you can change the current running time period by pressing the button,

cut to the current time period you need, press [OK] to switch to the new time period, and return to
the running interface to view the information parameters of the time period.
(5) Parameter settings

Since the equipment is rigorously tested before leaving the factory, it is not recommended for the
user to enter the settings. If the environment is not good when used, the outside temperature is too
low or too high, which will cause the display value to be inaccurate with the actual temperature
inside the chamber. If it exceeds the index range, the after-sales service personnel can guide the
operation, the system time and the factory reset can be With this setting, once the factory settings
are restored, the system parameters will be restored to the factory parameters. Press [OK] to
confirm the save modification information and change the initial password to [888888].
(6) About us
In order to better apply these products, please read this manual carefully. If you have any problems
during the use, you can contact our technical support staff through this interface.

8
V. Precautions and troubleshooting

1. Precautions
(1) The temperature deviation exceeds ±2°C alarm;
(2) When the touch switch is short circuited, it will alarm;
(3) When the temperature sensor is off or short circuited, it will stop alarm and stop heating and
cooling;
(4) The working temperature of the working temperature is above 60°C to stop heating;
(5) The temperature of the heating chamber is higher than 65°C, the protection relay is closed, and
the heating power is cut off;
(6) The chamber suddenly does not work. Please check whether the fuse tube (after the chamber)
is burnt out and check the power supply.
(7) The load-bearing shelf in the chamber has limited load. It is recommended that the weight
control of each layer is best within 10Kg;
(8) After the chamber has landed, if the ground is not flat, it should be flattened;
(9) In order to ensure personal safety, the ground wire (E) of the power socket should be
connected to the earth;
(10) When the chamber is in normal operation, the placement of the load in the chamber should
not affect the air circulation to ensure uniform temperature and humidity inside the chamber;
(11) The chamber should be placed in a cool, dry, well ventilated place away from heat and
sunlight. Place it smoothly to prevent vibration from generating noise;
(12) The surface and inner wall of the chamber should be wiped frequently to keep it clean;
(13) If the chamber is not used for a long time, unplug the power cord to prevent the chamber
from injuring people, and open the chamber door to keep the chamber dry. It is necessary to run it
for 1 to 2 days according to the conditions of use to avoid aging damage of some devices. ;
(14) When the temperature is below 10°C, the ambient temperature should not be too high, and the
temperature difference between the inside and outside of the chamber should not exceed 20°C to
ensure the Function Continue Proc Set Pd. S/W Proc S/W ParaSet About us
8 normal operation of the light chamber;
(15) If the chamber fails, it should be repaired by a professional or a manufacturer;
(16) The sterilization lamp is an ultraviolet lamp. When using the sterilization lamp, the operator
should remove the seed in the chamber, otherwise the seed will be damaged;
(17) If the accessories are common parts of the market, the warranty and performance
requirements are generally in accordance with the general requirements of the market and are not
included in the product warranty;
(18) All pictures are factual presentations, the products are subject to the actual product;
(19) The final interpretation right belongs to our company.

9
2.Easy troubleshooting

Fault
cause of issue Approach cause of issue Approach
phenomenon
1 Circulating fan ring Replace the fan
Reasonably place internal things to
Circulating air duct blocked
2 avoid

The test method is unreasonable:


1. The test should be carried
1. Turn on the air conditioner;
out at an ambient temperature
2. Empty chamber test, the
of about 25 degrees. The
location of the thermometer is
ambient temperature above too
arranged according to the
high or too low will have an
requirements of the national
effect on the internal uniformity;
standard;
2. When the uniformity is tested
3. After the temperature of the
Bad temperature according to the national
cabinet reaches the set
uniformity standard, the internal should be
temperature, it should be
empty, and the 9-point or
stabilized for another 1-2 hours
3 13-point test method is adopted;
before testing.
3. The cabinet does not fully
4. According to the specified time
enter the constant temperature
interval, after 6 consecutive tests,
state
each thermometer is averaged and
4. When the uniformity is
then compared;
obtained according to the
5. Correct each thermometer
average value of multiple tests, it
before the test. After the test is
cannot be judged by a certain
finished, remove the correction
measurement;
value and then calculate.
5. The error between the
thermometer itself should be
removed
When the temperature is lower
than the target temperature,
Heating lamp and heating is required. When the
cooling lamp are humidity is higher than the target
No need to deal with
4 on at the same humidity, the humidity needs to
time be removed by cooling, so it is
normal to have heating and
cooling in a short time.

The temperature The main temperature is Call the after-sales technician for
5 of the instrument reversed with the protection guidance
is set within the temperature sensor plug-in,

10
Temperature causing the temperature to drop
control range and directly.
immediately
drops to the
lowest
temperature.
Check and replace the fan (can call
Damage to the internal fan of
6 the technician for guidance)
the cabinet
The cooling light
Check and replace the condenser
is
fan
7 on, but it does Condenser fan damage
(can call the technician for
not drop the
guidance)
temperature.
The compressor is not working
(check if the compressor is
8 Call after sales
vibrating)

The instrument
does not control
the temperature,
does not heat, set
a certain
temperature
9 Damaged motherboard Replace the motherboard
control range, the
temperature does
not go, the
indicator light
does not light.

Unable to set
The keypad or keypad is
10 temperature Replace the keypad and cable
damaged.

When the temperature sensor


is detected to be normal and Repair or replace the high
Power on alarm
11 the humidity sensor is normal, temperature sensor.
the high temperature sensor is
judged to be damaged or broken.
Compressed compressor Adjusting the fastening
12 The boot sound is
transport compressor
too loud
13 Fan damaged Replace the fan
Dry the button board, if not,
The button is not
The button board is bad due to replace the button board and
14 working, long
moisture follow the
light
moisture-proof treatment

11
In order to prevent the cabinet
For example, if from being sent to an area with
there are more a high altitude, the glass is
Immediately after receiving the
copper tubes on broken due to the air pressure
goods, follow the label on the
the sides of the problem, and a diversion hole
15 chamber to plug the copper tube
glass on some is intentionally added so that
with the supplied sealant.
sides of the the air pressure before the two
cabinet or holes in layers of glass can be consistent
the glass. with the outside.

VI. Warranty information

1. warranty period
For our company's products, due to the failure of its materials or product quality, the company
provides a free warranty within one year from the date of shipment. During the warranty period,
any faulty product may be warranted or replaced by the company, but in any case, the company's
liability is limited to the original price range at the time of purchase.
This warranty does not cover accessories or options for products supplied by the company.
2. Exceptions
The product is not covered by the warranty if it is caused by the following conditions:
(1) Repair or change without permission;
(2) Misuse, negligence or accident.

VII. List of accessories

Name Quantity

Chamber 1 set
Fuse 2
Manual, warranty card 1
Water tray 1

12
VIII. Appendix

Temperatue
Dimensions
Model Volume control Illuminance
(width*depth*height)
range

GTOP-158Y 150 L 525×540×1580 mm 0~50℃ 0~8000LX

GTOP-268Y 268 L 666×630×1755 mm 0~50℃ 0~8000LX

GTOP-310Y 310 L 666×630×1895 mm 0~50℃ 0~8000LX

GTOP-400Y 400 L 744×658×1760 mm 0~50℃ 0~8000LX

GTOP-500Y 500 L 825×705×2030 mm 0~50℃ 0~8000LX

GTOP-1000Y 1000 L 1370×660×1990 mm 0~50℃ 0~5500LX

13