BOUCLE 4-20mA
3 programmable “user settings” per range
The GIV is a portable generator, used to generate standard signals (4-20 mA / 0-20 mA loop or 0-10
VDC) for field-testing and calibration of measuring systems (transmitter, digital panel meter,
controller, relays, valves and so on).
The GIV is able to generate these signals and can also be used as a 2-wire transmitter: in that case,
it makes the current varies and enables test continuity and reception of 4-20 mA signals.
The GIV is programmable with a user-friendly design. Fixed set points: 0%, 25%, 50%, 75% and
100% makes easier zero and span adjustments and linearity tests on current loops. Pushbuttons on
keyboard can provide a continuous adjustment of the signals and extend the range up to 110% to
check threshold or loop simulation behaviour.
A memory setting of 3 values makes this system quicker and easier to use.
The GIV can also measure a current up to 23.5 mA DC, providing a full-scale diagnostic feature.
The GIV is a simple, comprehensive system, ideal for test and calibration of measuring loops.
The equipment bears the CE mark, in accordance with directive 89 336/CEE – electromagnetic
compatibility. This system meets the following standards: EN50081-2 et EN50082-2.
Complementary tests have been carried out as per other standards. Please call our technical
team for further information.
Maximal load available at 20 mA : 600 Ω, (560 Ω at Minimal load resistor 650 ? Umax = 11 VDC
22 mA)
Imax : 22 mA - Uo = 13,5 VDC
GIV
GIV
4-20mA CURRENT LOOP
0/10 VDC
STROM KREIS 4-20mA
+ -
- Gen +
load
- +
A flashing “BATTERY” (PILE) indicates that batteries are
load loop supply +
low and must be shortly replaced
ENVIRONMENTAL
ACCURACY Operating temperature : 0 to +50°C
-4 Storage temperature : -20 to +85°C
Output and display : 5.10
Display resolution : 0.01 mA or 0.01 VDC
-4
Thermal drift = 1.10 /°C
GIV GIV
- Gen +
XT
- +
Umax 13.5 V
- +
load loop supply
A puissance 3
GIV
On
Keys for Fixed values 0% M1 START/STOP
Off
M2 25%
Choice of the signal type :
50% M3 I
U 4-20mA, 0-20 mA or 0-10 VDC
BAT 75%
% Choice of display :
100% MES.
% or physical units
Key"FCT" to enable access to
Memories, battery test Fct.
and Measurement features mA
0 Instant 0 mA or 0VDC
V
Increasing or decreasing
the signal
- Mes +
+0/10V- - XT + - Gen +
BOUCLE 4-20mA
I
U
Choice of the signal generated by the GIV: 0-20 mA, 4-20 mA or 0-10 VDC by pushing the key then on
I
U
keys or , then confirm the choice by pushing
TEST FUNCTION
I
Fct. U
Pushing then , will start a test of the loop load. When you work in current loop, this function will
generate a ramp from 0 to 20.5 mA on a period of 5 seconds to detect the possible overload of the loop.
When you work in voltage loop, this function will generate a ramp from 0 to 10.5 VDC on a period of
5 seconds to detect the possible overload of the loop
Current mode
- Select the 0-20 mA range or connect to a 100-Ohm shunt
- Press the 0-mA key
- Use the or keys to obtain an output of 0.10 mA (+/-1 count).
- Adjust the zero current potentiometer to obtain an output of 0.100 mA (+/- 1 count)
- Press the ON/OFF switch to turn the GIV off
- Press the % key and the ON/OFF key at the same time: the “AJI” message is displayed
- Use the or keys to obtain an output of 20.000 mA (+/-1 count).
- Press the % key to validate the setting
Check the linearity
Fct. %
- Push on key then on key, then the message
‘’CONTR’’ is displayed zero current