Forward:

Firstly , thanks for the buyers those who have bought our machines . This manual is to help clients to use
this machine better. Hope our product and service could make you satisfied.

Attention:
This product must be strictly used in accordance with the method of this manual, if not, you have to take
the responsiblity for eventual consequences. company does not assume any responsibility. This manual
contains proprietary information protected by patents issued. Copyright, without the prior written consent
of our company, any one is prohibited copying and translating any part of this manual.

Manufacturer's responsibility:
The company are responsible for instrument safety and performance and reliability under the following
circumstances : the assembly operations, re-adjust, improvement and maintenance are recognized by the
Company personnel, related electrical equipment are in line with national standards, and cosmetic
instrument is operated in accordance with instructions.
Warning:
This is a cosmetic device.
If beauty institutions that are using this machine can not achieve a satisfactory operation of the program, it will
result in abnormal equipment failure, and may endanger personal safety.
Today, the Company could provide a circuit diagram with pay, also a list of instructions proofing methods and other
information to help the user by a suitable qualified technical personnel to repair those parts of the instrument
companies classified by user-serviceable.

Guarantee:
Manufacturing processes and raw materials: We provide a year' warranty from the date of shipment while machine
is in regular use and maintenance of the state.

All the examination of our company is not perfectly complete.

Free service:
Not including the shipment cost and other fees .
We are not responsible for the below items: Default use ,replace the machine part without our permission or repair
machine by a non-authorized of our company people ,these conditions caused by direct, indirect and ultimately
damage and delay.
Safety, reliability and operating conditions.
If the following conditions occur, we will not be responsible for the machine safety , reliability and
operating conditions .
Take apart the machine component,stretch ,and re-debugging . Repairing machine by a non-authorized of
our company .
Return Method:
If really need to return back the machine , please follow the below process:
Picking machine right : Please contact our after-sale department,please inform the machine serial
number ,which is written in the machine ourside package .If the machine serial number is not clear , can
not return the machine . Please write the machine type,product serial number ,and explain the return
reasons .
Shipment cost : The return machine should be shipped back by our company ,the shipment cost should
be borne by the buyer.

Content Summary:
This user manual contains all the general information of the machine.This is the starting point for new
users to start using liposunic,Because it is a general introduction to the system,the manual explains how
it works, introduces the installation and various methods of operation of the liposunic machine. you can
be found the necessary information through catalogue.
Contraindications:
HIFU treatments are not recommended for people who have:
1.have heart pacemaker or other similar electronic devices in body.
2.pregnant
3.Take some medicine such as Accutane,Roaccutane and Oratane in the past nine months.
4.there is spun gold on facial.
5.Heart disease,diabetes and hyperthyreosis patient.
6.have wound after operation.
7.have severe skin disease

Before treatment:
Prepare medical things:
Tweezers, cotton swabs, shave wool implement, alcohol, cooling gel, small brush,glasses,Eye mask.
Notice:mustn't wear Metal accessories in the treatment,it is very dangerous.

Precautions before surgery facilities:


1. Preoperative keep skin, hair parts clean, not makeup the day of surgery, before the make-up to try to remove the
traces.
2, before surgery to fully communicate with your physician, early and late each time, ten grams of collagen repair
ACMETEA, taking 10-20 days.
3, dental implants, orthodontics patients before surgery facilities to inform the process will be painful treatments, the
patient can not be applied to thyroid surgery.
During treatment:
Energy must be adjusted from low to high;(important)
(Deep color of skin need low ennergy and light color need higher energy)
The handle must be touched on skin and then push switch;
(If there is any problem in treatment,must close switch.If start,repeat before operation,other wise,it will hurt skin.)

The area forbidden for treatment:


1.Near 1.5cm-2.0cm of hairline
2.Temples(Eye corner of parallel lines),
3.Inside of orbital bone(warning! it is eyeball)
4.Apple muscle,
3.1cm area beside of mouth
4.1cm beside of ear
5.Philtrum
6.Adam's apple
7.Also need forbidden do the treatment on bone.
UPPER NECK
1.Treat the upper neck, making multiple passes and moving between
columns as necessary, based on the patient’s response.
-Deliver the first line of energy in each column as close to the jaw bone
(mandible) as possible, without hitting it. Align the transducer with the
mandible, then slide it back and forth so the mandible “flashes” acros
the screen. Deliver energy when the mandible is not in the treatment
window.
-Treating directly over the mandible or facial nerves may increase
discomfort and the likelihood of potential side effects. *
CHEEK
1. Treat the cheek region, making multiple passes and moving between columns as necessary, based on the
patient's response.
-Do not treat directly into the cheek bone or into the mandible.

-Ensure that the columns of treatment touch each other without overlapping, to ensure proper density.

-Treating directly over the mandible or facial nerves may increase discomfort and the likelihood of potential side
effects.

Refer to the Instructions For Use (IFU) for complete instructions on operating the Ulthera System

LATERAL ORBIT
1.Align the transducer with the outside corner of the eye (closest to the ear),positioning the handpiece so the cord
is towards the patient’s eye.
2.Deliver an equal number of lines above and below the marked line at the outside corner of the eye.
-When treating, the end of the transducer closest to the ear will move more than the opposite side, creating a
fanning pattern.
-Reposition the transducer after delivering each line, instead
BROW:
1. Place the treat line of the transducer on the brow region, positioning the handpiece so the cord is towards the
patient’s eye. Deliver the determined number of lines spread evenly in the region.-Shorten the treatment line if the
frontal bone is visible above the green line, if the shape of the frontal bone prevents good coupling or if the patient’s
sensory response is significant.

INFRAORBITAL
1.Align the transducer on the mark below the eye, while gently pulling the infraorbital skin in a downward direction
(towards the feet).
2.Deliver the first row of energy at the mark and then deliver each additional row below the last row delivered. Never
tilt the transducer towards the orbit.

GENERAL NOTES:

-Do not treat over the eye, in a location or with a technique, where the ultrasound energy may reach the eye. *

- Do not deliver energy if bone is visible at or above the


treatment line.
UPPER NECK:
Treat the upper neck, making multiple passes and moving between columns as necessary, based on
the patient’s response.
- Deliver the first line of energy in each column as close to the jaw bone (mandible) as possible,
without hitting it. Align the transducer with the mandible, then slide it back and forth so the mandible
“flashes” across the screen. Deliver energy when the mandible is not in the treatment window.
-Do not deliver energy below the line above the thyroid cartilage.
-Treating directly over the mandible or facial nerves may
increase discomfort and the likelihood of potential side effects.
CHEEK:
Treat the cheek region, making multiple passes and moving between columns as necessary, based on
the patient’s response
-Do not treat directly into the cheek bone or into the mandible.

-Ensure that the columns of treatment touch each other without overlapping, to ensure proper density.

-Treating directly over the mandible or facial nerves may


increase discomfort and the likelihood of potential side effects.Refer to the Instructions For Use (IFU)
for complete instructions on operating the Ulthera System

LATERAL ORBIT:

1.Align the transducer with the outside corner of the eye (closest to the ear),positioning the handpiece so
the cord is towards the patient’s eye.
2.Deliver an equal number of lines above and below the marked line at the outside corner of the eye.

-When treating, the end of the transducer closest to the ear will move more than the opposite side,
creating a fanning pattern.

-Reposition the transducer after delivering each line, instead


of just pivoting on a point, to prevent retreating the same tissue.
BROW:
1.Place the treat line of the transducer on the brow region, positioning the handpiece so the cord is towards the
patient’s eye. Deliver the determined number of lines spread evenly in the region.

-Shorten the treatment line if the frontal bone is visible above the green line, if the shape of the frontal bone
prevents good coupling or if the patient’s sensory response is significant.

INFRAORBITAL:
1.Align the transducer on the mark below the eye, while gently pulling the infraorbital skin in a downward direction
(towards the feet).
Deliver the first row of energy at the mark and then deliver each additional row below the last row delivered. Never
tilt the transducer towards the orbit.

GENERAL NOTES:
-Do not treat over the eye, in a location or with a technique, where the ultrasound energy may reach the eye. *

-Do not deliver energy if bone is visible at or above the treatment line.

-Frequently verify proper coupling between the transducer and the skin (see REFERENCE section).
Operation---Machine screen Interface
1)Operation interface :
When you start the machine ,you can choose HIFU mode and Liposonic
mode:
Function Description Interface Terms Description
1 Screen brightness
adjustment
Mainly adjust the Display screen brightness

2 Tools Back stage adjust the voice and other functions

3 Times/Areas Adjust the time / setting or adjust the treatment areas

4 Energy density Energy adjustment range(1.5-5.0J)

5 Start Beginning working

6 Pause Stop working

7 Treatment Area Treatment area is referred to the treatment times

8 4-4.5mm depth Choose the treatment depth

9 7-4.5mm depth Choose the treatment depth

10 7-3.0mm depth Choose the treatment depth

11 10-1.5mm depth Choose the treatment depth

12 Case Record Patient ID,Choosing Tips,Times record and so on

13 Scanner depth Skin subcutaneous tissue,Ultrasound imaging display


Item Description

Switch button Imaging start, Stop Treatment/Start Treatment mode


Work button Indicator lamp is on when machine is working
Lock head of tip position Treatment tip is locked in the Cartridge

Treatments tips Treatment tips are using in the patient’s different


parts
Soft rubber cable sheath Connecting the handle to control section
Our Hifu machine provide transducer DS-1.5,DS-3.0,DS-4.5, DS-8.0,DS-13mm,
DS-1.5/3.0/4.5 Transducers have 5000 stable shots.DS-8.0&DS-13mm transducers
have 9000 shots The shot times of transducer is 10000 shot times
(there are 14 points on one line at this situation)
1.prompt treatment area in Figure 1
2.To facilitate memory and operating system with a single count function shown in
Figure 2

When treating an area, such as when the maximum number is 40, when the count
reaches 40, the system stops working, press the RESET button to reset after 0, re-
count and work.
This machine Treatment Tips has Temperature Detection , Output Control
Testing ,Alarm Prompt ,and Power supply work protection function .When the
treatment tips temperature achieve the designed temperature ,S1 and S2 indicator
light will flash at the same time,Buzzer sounded, the machine will stop
working ;When finding the energy output is out of control, S1 and S2 will blink and
stop working at the same time .Power supply has overcurrent protection and
temperature control protection .

1.The operation person must be specially trained , should know well of the Human
Face Structure ,and is proficient in the equipment operation .
2.Dividing the treatment area as well as avoiding the nerve sensitive areas.
The pre-treatment Guide of Ulthera program established in clinical trials.
The menu images shows guidelines details, as they appeared on the screen.
Two pre-programmed treatment guide:
Note: Operation is according to the below flow chart. Note: After finishing the treatment , please restore
defaults engergy settings from the same and different operations screen guide.

3.According to the characteristics of the treatment of people, the reasonable choice for the treatment of head
focus depth, energy, and point spacing Length.General rule is focused on small spacing-energy small;The focused
on big spacinghigh energy.Due to the everyone's different characteristics, it is suggested that increased from low
energy up.

4.When treating, you need coated cooling gel in the treatment area, and the treatment handle should vertical
contact the skin completely (Prohibit tilt with skin, vacant, also a high one low is not allowed.)
5.The times of treatment area can not exceed the number of standard interface
prompts.Please refer to Picture 5.

6.After the treatment is completed,


the facial hair is forbidden esides, please remember not eating hot and cold food. Meanwhile, you must
avoid stimulating the facial nerve system.
After self-testing, machine will run to handpiece interface automatically.
Screen will show the cartridge information. For example: cartridge focus depth is 0.8cm ,

cartridge connection status, cartridge Remaining sites:

then you can press the “Select moulding”,you can enter the following menu:
Screen will show the cartridge output style, For example: Full Style is 576 piece

Output points,when side fat less than 25mm, Use the edge style for shooting ,around the

navel area it will need to use the special pattern

Screen displays the position of the current focus point: ,which indicates that the

current site is in the positon of 75 point,energy of the focus point is adjustable, ,

press and you will see the adjustment button ,that you can adjust the

energy here,and the adjustment range isfrom 30 to 75.At the same time, the interface will also shows

the focusing depth and the cartridge sites remaining:


1 The production of products should meet the product standards, such as there is not in line with my
company responsible for the replacement, return.
2 Our company shall bear the corresponding liability for damage caused by personal or property damage
due to the defects in the product manufacturing, but we shall not be liable for any of the following
circumstances:
1). Buyers, the operator does not operate in accordance with the instructions to use, not according to the
safety precautions, warning of the use of the note
2). Product, personal and property damage in violation of the provisions of the product transportation,
installation, use, maintenance, storage, etc..

3 My company to implement a lifetime of products and technical services, in the quality assurance period of
manufacturing quality problems, will be free for customer maintenance.

The control handle or ultrasonic probe must be purchased from the company when it is required to be
replaced.
1. storage:Storage device should be placed in the original packaging box, clean and well ventilated indoor,
packing box should be high pad, environmental temperature should be to - 10 degrees to 40 degrees,
relative humidity is not more than 93%, are not allowed to have harmful gases, flammable, explosive and
corrosive gases.

2.Transportation: under package conditions, the instrument suits for road, rail, air and sea transport,
handling and transportation process should prevent severe
vibration and shock, and shall not be exposed to moisture, and can not be flammable, corrosive substances
mixed, mixed transport, about the details requirements,please according to order contract.

1.For this instrument, the warranty is 1 year (note: except artificial damage).
2.Parts warranty is three months (Note: except artificial damage), consumables (Probe 300 more than 300
shots) do not warranty.
3.Beyond the warranty of instruments, if want to continue to use, need engineers to detect approval and
subsequent , and need regular instrument performance testing by engineers.
4.Over the period of warranty of scrap instrument and accessories should be handled according to local
regulations.