Professional Documents
Culture Documents
Measurement by Velar
Measurement by Velar
Search... Search
Download presentation
Buttons:
Cancel Download
1 / 217
Similar presentations
By Velar H. Elias Measurement.
Published by Cordelia Whitehead Modi ed about 1 year ago
Embed
Download presentation
3
Download presentation
2 Introduction
CHAPTER ONE
We think you have liked this presentation. If you wish to
download it, please recommend it to your friends in any social
3 system. Share buttons are a little bit lower. Thank you!
INTRODUCTION
Measurement: measurement means, to monitor a process or a
Buttons:
operation and using an instrument, express the parameter, quantity or
a variable in terms of meaningful numbers.
Measurement of a given parameter or quantity is the act or result of a
quantitative comparison between a prede ned standard and an
3
unknown quantity to be measured.
5 Evolution of Instruments II
Electrical:
it is faster than mechanical, indicating the output are rapid than
mechanical methods. But it depends on the mechanical movement of
the meters. The response is between 0.5 to 24 seconds.
Electronic:
it is more reliable then other system. It uses semiconductor devices
and weak signal can also be detected.
7 Performance Characteristics
/
Static Characteristics: the set of criteria de ned for the instruments,
which are used to measure the quantities which are slowly varying with
time or mostly constant, i.e., do not vary with time is called static
characteristics.
Download presentation
Dynamic Characteristics: when the quantity under measurement
changes rapidly with time, it is necessary to study the dynamic
relations existing between input and output which is expressed as
We think you have liked this presentation. If you wish to
di erential equations.
download it, please recommend it to your friends in any social
system. Share buttons are a little bit lower. Thank you!
8 CALIBRATIONButtons:
calibration is the process of making an adjustment or making a scale
so that the reading on an instrument agree with the accepted and
certi ed standard.
Calibration is a comparison between a known measurement (the 3
standard) and the measurement using your instrument. Typically, the
accuracy of the standard should be ten times the accuracy of the
measuring device being tested. However, accuracy ratio of 3:1 is
Cancel Download
acceptable by most standards organizations.
In practice, calibration also includes repair of the device if it is out of
calibration.
9 CALIBRATION
Calibration of your measuring instruments has two objectives:
It checks the accuracy of the instrument
And it determines the traceability of the measurement.
Calibration improves the accuracy of the measuring device. Accurate
measuring devices improve product quality.
When should you calibrate your measuring device?
According to recommendation of the manufacturer.
After any mechanical or electrical shock.
Periodically (Daily, monthly, before use)
10 De nitions
Instrument: a device for determining the value or magnitude of a
quantity or variable.
Accuracy: closeness with which an instrument reading approaches the
true value of the variable being measured.
Precision: a measure of the reproducibility of the measurement i.e.,
given a xed value of a variable, precision is a measure of the degree
to which successive measurements di er from one another.
11 De nitions
Sensitivity: the ratio of output signal or response of the instrument to a
change of input or measured variable.
Resolution: the smallest change in measured value to which the
instrument will response.
Error: deviation from the true value of the measured variable. Several
techniques may be used to minimize the e ect of error.
/
the algebraic di erence between the indicated value and the true value
of the quantity to be measured is called an ERROR
Download presentation
12 Precision and accuracy comparison
Accuracy refers to the degree of closeness or conformity to the true
value at the quantity under measurement. Precision refers to the
We think you have liked this presentation. If you wish to
degree of agreement within a group of measurements or instruments.
download it, please recommend it to your friends in any social
In critical work, good practice dictates that the observer make an
system. Share buttons are a little bit lower. Thank you!
independent set of measurements, using di erent instruments or
di erent measurement techniques, not subject to the same systematic
Buttons:
errors. He must also make sure that the instruments function properly
and are calibrated against a known standard, and that no outside
in uence a ects the accuracy of his measurements.
3
Cancel Download
14 Sources of Errors Faulty design of instrument
Insu cient knowledge of quantity and design conditions
Improper maintenance of the instrument.
Sudden change in the parameter to the measured.
Unskilled operator
E ects of environmental conditions.
15 Errors
A study of errors is a rst step in nding way to reduce them.
Errors types:
Gross errors: Largely human errors, among them misreading of
instruments incorrect adjustment and improper application of
instruments, and computational mistakes.
Systematic errors: Short coming of the instruments, such as defective
or wrong parts, and e ective of environment on the equipment or the
user. (Instrument Error)
Random errors: these errors are due to unknown causes and occur
even when all systematic errors have been accounted for.
/
18 Systematic-Instrumental Errors
De ned as shortcoming of the instrument, and can be avoided by:
Download presentation
Selecting a suitable instrument for the particular measurement
application
Applying correction factors after determining the amount of
We think you have liked this presentation. If you wish to
instrument error
download it, please recommend it to your friends in any social
Calibrating the instrument against the standard
system. Share buttons are a little bit lower. Thank you!
Buttons:
19 Systematic-Environmental errors
Due to external conditions a ection the measurement, environmental
errors are due to conditions external to the measuring device,
including conditions in the area surrounding the instrument, such 3as
the e ect of change in temperature, humidity, barometric pressure, or
of the magnetic or electrostatic elds
Systematic errors can also be subdivided into static or dynamicCancel
errors. Download
20 Systematic-Environmental errors
To eliminate the error:
Proper correction factors given by the manufacturer.
Make arrangements to keep surrounding constant .
Sealing the components to avoid dust, humidity
Providing magnetic or electrostatic shields
22 Random errors
In well designed experiments, few random errors usually occur, but
they become important in high accuracy work.
The only way to o set these errors is by increasing the number of
readings and using statistical means to obtain the best approximation
of the true value of the quantity under measurement
They cannot be corrected by any method.
24 UNITS
It is necessary to specify type and magnitude for the reading. Where
unit represents the type of the physical quantity and reading on the
instrument represents its magnitude.
Di erent system of units are
M.K.S
/
C.G.S
S.I (System International units)
Download presentation
25 System International Units (SI)
Quantity
Units
We think you have liked this presentation. If you wish to
Symbol
download it, please recommend it to your friends in any social
Length
system. Share buttons are a little bit lower. Thank you!
Meter
M Buttons:
Mass
Kilogram Electromagnetic
Kg Induction
Time 3
Second
S
Electrical current
Cancel Download
Ampere
A
Thermal temperature
Kelvin
K
Luminous intensity
Candela
cd
Amount of substance
Mole
mol
35 International Standards
These standards are maintained at the (IBS) international bureau of
weights and measures and are periodically evaluated and checked by
absolute measurements.
These standards are not available for ordinary users for calibration.
For accuracy they are replaced by absolute units which are more
accurate than international standards.
36 Primary Standards
They are maintained at national standard laboratories in di erent
countries (NBS).
/
These standards represents fundamental units as well as electrical and
mechanical derived calibrated by absolute measurements at each
national laboratories.
Used for calibration and veri cation of secondary standards.
Download presentation
37 Secondary Standards
We think you have liked this presentation. If you wish to
Since primary standards are not available for outside users, various
download it, please recommend it to your friends in any social
industries need some reference.
system. Share buttons are a little bit lower. Thank you!
They are used by measurement and calibration laboratories and are
maintained by the particular
Buttons: industry to which they belong.
Each industry has its own standards
38 Working Standards 3
These are the basic tools of a measurement laboratory.
Use to check and calibration for accuracy and performance or to
perform comparison measurements in industrial applications. Cancel Download
Ex: resistor industry maintains a standard resistor for checking the
values of manufactured resistors
39 IEEE Standards
Institute of Electrical and Electronic Engineering standards are checking
of secondary standards.
Used for testing and evaluating of electronic systems and components.
The most important standards is (IEEE 488) digital interface for
programmable instrumentation for test and other equipment.
IEEE classi cation (word le)
Cancel Download
/
48 Active and Passive Instruments
1. Active Instruments-
Download presentation
The change in petrol level moves a potentiometer arm, and the output
signal consists of a proportion of the external voltage source applied
across the two ends of the potentiometer.
We think you have liked this presentation. If you wish to
The energy in the output signal comes from the external power source:
download it, please recommend it to your friends in any social
the primary transducer oat system is merely modulating the value of
system. Share buttons are a little bit lower. Thank you!
the voltage from this external power source.
Buttons:
53 Measuring Instruments
Classi ed measuring instruments in to two groups:
Absolute Instruments (Standard instruments)
Secondary Instruments
54 Absolute Instruments
/
These instruments gives the magnitude of quantity under
measurement in terms of physical constants of the instrument
e.g. Tangent Galvanometer. These instruments do not require
comparison with any other standard instrument
Download presentation
These instruments give the value of the electrical quantity in terms of
absolute quantities (or some constants) of the instruments and their
de ections.
We think you have liked this presentation. If you wish to
In this type of instruments no calibration or comparison with other
download it, please recommend it to your friends in any social
instruments is necessary.
system. Share buttons are a little bit lower. Thank you!
Buttons:
55 Absolute Instruments
• They are generally not used in laboratories and are seldom used in
practice by electricians and engineers. They are mostly used as means
of standard measurements and are maintained lay national 3
laboratories and similar institutions.
• Some of the examples of absolute instruments are:
* Tangent galvanometer * Raleigh current balance * Absolute
Cancel Download
electrometer
57 Absolute instruments
Tangent galvanometer is an early measuring instrument for small
electric currents. It consists of a coil of insulated copper wire wound on
a circular non-magnetic frame. Its working is based on the principle of
the tangent law of magnetism
An electrometer is an electrical instrument for measuring electric
charge or electrical potential di erence. The absolute electrometer was
rst proposed by Lord Kelvin
Raleigh current balance (or ampere balance), a device used to
reproduce the unit of electric current, the ampere. Made of
nonmagnetic materials.
58 Secondary Instruments
There are direct reading instruments.
These instruments are calibrated by comparison with an absolute
instrument or another secondary instrument, they are used in general
for all laboratory purposes.
Therefore secondary instruments are most commonly used.
They are direct reading instruments. The quantity to be measured by
these instruments can be determined from the de ection of the
instruments.
Some of the very widely used secondary instruments are: ammeters,
voltmeter, wattmeter, energy meter (watt-hour meter), ampere-hour
meters etc.
59 Secondary Instruments
60 Secondary Instruments
Secondary instrument may be grouped on the basis of various e ects
of electric current as follows: /
Magnetic e ect (usually for Ammeters and Voltmeters)
Heating e ect (for Ammeters and Voltmeters)
Electromagnetic e ect (ammeter, voltmeter, wattmeter, watt-hour
meter)
Download presentation
Electrostatic e ect (for Voltmeters only)
Chemical e ect (DC Ampere-Hour meter
62 Secondary instruments
Buttons:
Secondary instruments can be classi ed due to the method of taking
reading from the instruments as follow:
Indicating Instruments
Recording instruments 3
Integrating instruments
Cancel Download
63 1. Indicating Instruments
Indicating instruments are those which indicate the instantaneous
value of the electrical quantity being measured at the time at which it is
being measured.
Their indications are given by pointers moving over calibrated dials.
The moving system is subjected to the following three torque forces:
De ecting (Torque) force
Controlling (Torque) force
Damping (Torque) force
/
66 1. B. Controlling Torque Function of controlling torque:
It balance the de ecting torque,presentation
and insures that the magnitude of the
Download
de ection is always same for a particular value of the quantity to be
measured.
It brings back the moving system
We think you (pointer)
have likedtothis
its zero position. If you wish to
presentation.
Tc=Td (Equilibrium) download
at measured (position)
it, please value.
recommend it to your friends in any social
Controlling torque obtained be:
system. Share buttons are a little bit lower. Thank you!
Controlling by Spring.
Controlling by Gravity.
Buttons:
/
71 1. B. I. Controlling Torque by Spring
The control torque Download
is provided by two hair springs, coiled in opposite
presentation
directions acting one against the other.
Equilibrium force of two springs equal to zero, therefore the pointer
remains at zero position.
We think you have liked this presentation. If you wish to
Springs are made of such material which
download it, please recommend it to your friends in any social
Are non-magnetic
system. Share buttons are a little bit lower. Thank you!
Are not subject to fatigue
Have low speci c resistance
Buttons:
Have low temperature-resistance coe cient
Cancel Download
79 1. C. Damping torque
The degree of damping should be adjusted to a value which is
su cient to enable the pointer to rise quickly to de ected position
without overshooting .
80 1. C. Damping torque
If the instrument is under damping, the moving system oscillates a lot
before it nally settles down to its steady value.
In case of critically damped instruments, the pointer reaches its nal
steady position rapidly and smoothly (such an instrument is also called
dead-beat instrument).
An over damped instrument produces damping torque more than the
required value, as such the pointer moves slowly to its nal steady
value.
There are three types of damping torque
/
83 1. C. II. Fluid Friction Damping system
Download presentation
84 1. C. II. Fluid Friction Damping system
In this case the disc or vane attached to the spindle moves in a
damping oil. We think you have liked this presentation. If you wish to
The oil used must ful ll the following
download requirements:
it, please recommend it to your friends in any social
It should not evaporate quickly
system. Share buttons are a little bit lower. Thank you!
It should not have any corrosive action upon metals.
Its viscosity should not change with temperature.
Buttons:
It should be good insulator.
In the rst system, a disc is immersed in the oil. The friction drag
developed during the motion of the disc attached to the moving
system always opposes the motion. 3
In second system, vans are used.
Cancel Download
85 1. C. III. Eddy Current Friction Damping system
88 2. Recording Instruments
these instruments record continuously the variations of an electrical
quantity or physical quantities such as ow, pressure, temperature as a
function of time.
instruments like recording devices, X-Y plotter, and oscilloscope.
89 3. Integrating Instruments
These instruments measure the total amount of quantity of electricity
(Ampere-Hour) or the total energy (Watt-Hour) supplied to a circuit
over speci ed period.
These type of energy meters is used both for AC and DC.
97 PMMC construction
(a) Stationary part or magnet system: In the present time we use
magnets of high eld intensities, high coercive force instead of using U
shaped permanent magnet having soft iron pole pieces.
(b) Moving coil: The moving coil can freely moves between the two
permanent magnets as shown in the gure given below. The coil is
wound with many turns of copper wire and is placed on rectangular
/
aluminum which is pivoted on jeweled bearings.
98 PMMC construction
Download presentation
(c) Control system: The spring generally acts as control system for
PMMC instruments. The spring also serves another important function
by providing the path to lead electric current in and out of the coil.
We think you have liked this presentation. If you wish to
(d) Damping system: The damping force hence torque is provided by
download it, please recommend it to your friends in any social
movement of aluminum former in the magnetic eld created by the
system. Share buttons are a little bit lower. Thank you!
permanent magnets.
(e) Meter: Meter of these instruments consists of light weight pointer to
Buttons:
have free movement and scale which is linear or uniform and varies
with angle.
3
99 Error in PMMC instruments
(a) Errors due to permanent magnets: Due to temperature e ects and
aging of the magnets the two magnet may lose their magnetismCancel to Download
some extent. The magnets are generally aged by the heat and vibration
treatment.
(b) Error may appear in PMMC Instrument due to the aging of the
spring. However the error caused by the aging of the spring and the
errors caused due to permanent magnet are opposite to each other,
hence both the errors are compensated with each other.
(c) Change in the resistance of the moving coil with the temperature:
Generally the temperature coe cients of the value of coe cient of
copper wire in moving coil is very low. Due to lower value of
temperature coe cient the temperature rises at faster rate and hence
the resistance increases. Due to this signi cant amount of error is
caused.
/
102 Moving Iron Instruments
Moving iron type instruments
Downloadare of mainly two types. Attraction type
presentation
and repulsion type instrument.
Whenever a piece of iron is placed nearer to a magnet it would be
attracted by the magnet.
We think you have liked this presentation. If you wish to
The force of this attraction
downloaddepends
it, pleaseupon the strength
recommend of magnetic
it to your friends in any social
eld. system. Share buttons are a little bit lower. Thank you!
If the magnet is electromagnet then the magnetic eld strength can
easily be increased Buttons:
or decreased by increasing or decreasing electric
current through its coil.
Accordingly the attraction force acting on the piece of iron would also
be increased and decreased. Depending upon this attraction the
moving iron instrument was developed. 3
114 EX1/
We think you have liked this presentation. If you wish to
A 250-volt moving download
iron voltmeter takes
it, please a current
recommend it toofyour
0.05A when
friends in any social
connected to a 250system.
volt DCShare
supply. The coil has an inductance of 1
buttons are a little bit lower. Thank you!
Henry. Determine the reading on the meter when connected to a 250
Buttons:
volt, 100 Hz AC supply.
Ex2/
The change of the inductance for a moving-iron ammeter is
2μH/degree. The control spring constant is (5*10^-7 N.m/degree). The
maximum de ection of the pointer is 100 ͦ , what is the current 3
corresponding to the maximum de ection
Cancel Download
115 Suspension Galvanometer Ballistic Galvanometer
116 Introduction
Early measurements of Direct Current (DC) required a suspension
galvanometer.
This instrument was the former of the moving coil instrument, basic to
most DC indicating movements currently used.
A coil of ne wire is suspended in a magnetic eld produced by a
permanent magnet.
The coil will rotate in the magnetic eld when it carries an electric
current.
117 Introduction
the coil de ection is a measure of the magnitude of the current carried
by the coil.
The coil continue to de ect until its electromagnetic torque balances
the mechanical counter torque of the suspension.
118 Introduction
The suspension galvanometer is still used certain high sensitivity
laboratory measurements when the accuracy of the instrument is not
objectionable and portability is not required.
Galvanometer are used to indicate or measure small current in bridge
circuits, potentiometers and other measuring equipment.
/
There are three possible cases, depending on the quantity under the
root:
Case I:
Case II:
Download presentation
Case III:
Roots real and unequal
Roots conjugate-complex
We think you have liked this presentation. If you wish to
Roots real and equal
download it, please recommend it to your friends in any social
system. Share buttons are a little bit lower. Thank you!
129 Ex/
A PMMC instrument has a coil of dimension (10mm*8mm). The ux
density in the air gap is 0.15Wb/m² . If the coil wound for 100 turns,
carrying a current of 5mA . Then calculate the de ection torque .
Calculating the de ection if the spring constant is (0.2*10^-6
N.m/degree)
Download presentation
132 Ballistic Galvanometer
In fact, the moment of inertia of the coil is made so large that whole of
the charge passes through
We think the
yougalvanometer
have liked thisbefore its coil Ifhas
presentation. youhad
wish to
time to move su ciently.
download it, please recommend it to your friends in any social
In that case, the rst swingShare
system. of the coil isare
buttons proportional to theThank
a little bit lower. charge
you!
passing through the galvanometer.
After this swing hasButtons:
been observed, the oscillation coil may be rapidly
brought to rest by using eddy-current damping.
i.e. the coil moves after the charge to be measured has passed through
it. Obviously, during the movement of the coil, there is no current
owing through it. 3
143 Di nitions
Magnetomotive force: it is that force which drives or tends to drive the
ux through a magnetic circuit of number of conductors N and current
I passing through them (mmf=NI)
Similar to the way that electromotive force (EMF) drives a current of
electrical charge in electrical circuits, magnetomotive force (MMF)
'drives' magnetic ux through magnetic circuits.
144 EMF
/
152 A) Measuring of Flux / Flux density
Φ = ux linking withDownload
search coil presentation
R = resistance of the ballistic galvanometer circuit
N = number of turns in the search coil
t = time taken to reverse the you
We think ux have liked this presentation. If you wish to
average emf induced in the search
download coil recommend it to your friends in any social
it, please
And the average current through
system. Sharethe ballistic
buttons aregalvanometer
a little bit lower. Thank you!
Buttons:
153 Measurement of value of magnetizing force (H)
The value of H inside a specimen can be calculated from the formula,
given by:
i.e., ampere .turns / meter
3
N = number of turns in the specimen
I = current owing through the specimen
l = mean circumference of the ring (m) Cancel Download
/
158 B-1) Method of Reversal The circuit 11.8
For determination Download
of B-H curve, a ring shaped specimen whose
presentation
dimensions are know is used
A layer of thin tape is put on the ring and a search coil insulated by wax
is wounded over theWe tape. Another
think you have layer of this
liked tapspresentation.
is put over the search
If you wish to
coil and the magnetizing winding
download is uniformly
it, please wounditover
recommend this friends
to your tape in any social
The circuit 11.8 system. Share buttons are a little bit lower. Thank you!
The B-H curve may be plotted from the measured values of B
corresponding to the various value of H.
Buttons:
/
162 B-2) Step-by-step method
Download presentation
163 C) Determination of Hysteresis loop
Similar to determination of B-H curve, there are two methods for
determination of hysteresis
We thinkloop
you for a magnetic
have liked this material specimen:
presentation. If you wish to
Method of reversalsdownload it, please recommend it to your friends in any social
Step-by-step methodsystem. Share buttons are a little bit lower. Thank you!
Buttons:
164 C-1) Method of reversals:
R1, R2, and R3 are the variable resistances for adjusting the resistances
in the ballistic galvanometer and magnetizing coil circuits, R4 is a
3
variable shunting resistance, which can be connected across the
magnetizing coil by means of switch S2 thus reducing the magnetizing
current from its maximum value down to any desired value depending
upon the value of R4. Cancel Download
/
167 Determination of leakage factor in Dynamo-electric
machinery
Download presentation
In dynamo-electric machine, the leakage factor is de ned as:
Where “useful ux” is the ux in the ARMATURE (i.e., ux crossing the
air gap) We think you have liked this presentation. If you wish to
And “total ux” in the pole bodies
download which
it, please in turn is equal
recommend to the
it to your (useful
friends in any social
ux + leakage ux) existing on the pole body at its root.
system. Share buttons are a little bit lower. Thank you!
Thus in order to measure the leakage factor, we have to measure the
ux in the pole bodies and ux in the armature.
Buttons:
For measurement of leakage factor the (Flux meter) is used,
Ballistic galvanometer is unsuitable due to high inductance of the eld
system of the machine.
3
/
173 Lloyd- sher square for measuring iron loss
Iron losses
Download presentation
When ferromagnetic materials are subjected to an alternating eld,
power loss due to hysteresis e ects and eddy currents occurs.
Hysteresis loss: ThisWeloss depends
think you haveupon thethis
liked frequency, and maximum
presentation. If you wish to
ux density of the magnetic
downloadeld to which
it, please the specimen
recommend it toisyour
subjected.
friends in any social
Eddy current loss: system.
this loss depends
Share buttonsupon
are athe
littlewaveshape of the
bit lower. Thank you!
variation of ux with time, frequency and maximum ux density of the
magnetic eld to which the specimen is subjected.
Buttons:
Cancel Download
179 Lloyd- sher square for measuring iron loss
B--Wattmeter method-Lloyd- sher square
Each strip (usually 0.25mm long and 50 to 60 mm wide) is
perpendicular to the plane of the square. The strips are built up into
four stacks which are placed inside four similar magnetizing coils of
large cross-sectional area. These four coils are connected in series to
form the primary winding.
Each magnetizing coil has two similar single layer coils underneath it;
these are called secondary coils. These secondary coils are connected
in series in group of four, one from each core, to form two separate
secondary windings.
The magnetic circuit is completed by bringing the four stacks together
in the form of a square and joining them at the corners. The corner
joints are made by a set of standard right angled corner pieces.
There is an overlapping of corner pieces and strips at the corners due
to which cross-section of iron is doubled at the corners; therefore, a
correction must be applied for this. Also the measured losses has to be
corrected for loss in the corner pieces.
3
191 DC Ammeters (Shunt resistor)
The basic movement of a DC ammeter is a PMMC galvanometer, since
the coil winding of a basic movement is small and light, it can carry only
very small currents. Cancel Download
When a large currents are to be measured, it is necessary to bypass
the major part of the current through a resistance, called a shunt.
/
197 Shunt type Ohmmeter
Buttons:
Cancel Download
201 Shunt type Ohmmeter
4) H.M Calculate the external resistance that causing half-scale
de ection of pointer (shunt type Ohmmeter).
211 Electrodynamometer
Electrodynamometer is the one of the most important AC movement,
used in accurate AC voltmeters and ammeters, not only at the
powerline frequency but also in the lower audio-frequency range.
/
With some slight modi cation, can be used as a wattmeter, a
VARmeter, a power-factor meter, or a frequency meter.
May also serve as a transfer instrument, because it can be calibrated
on DC and then used directly on AC.
Download presentation
The d’Arsonval movement uses a permanent magnet to provide the
magnetic eld in which the movable coil rotates.
The electrodynamometer uses the current under measurement to
We think
produce the necessary eld you
ux. have liked this presentation. If you wish to
download it, please recommend it to your friends in any social
system. Share buttons are a little bit lower. Thank you!
212 Electrodynamometer
Buttons:
A xed coil, split into two equal halves, providing the magnetic eld in
which the movable coil rotates.
The two halves are connected in series with moving coil and are fed by
current under measurement. 3
Like d’Arsonval movement, its rotation is controlled by a springs.
And damping is provided by aluminum air vanes, moving in sector-
shaped chambers.
Cancel Download
215 Electrodynamometer
The scale of the electrodynamometer is usually calibrated in terms of
the square root of the average current square:
Therefore, the meter reads the rms or e ective value of the AC.
The transfer properties of the electrodynamometer becomes apparent
when we compare the e ective value of alternative current and direct
current in terms of their heating e ect or transfer of power.
216 Electrodynamometer
The average rate of producing heat by an AC of i amperes during one
cycle in resistance R: is
/
The average rate of producing heat by a DC of I amperes in a
resistance R: is
Download presentation
217 Thermo-instruments
Cancel Download
Search... Search