Manual Fluke Serie 80
Manual Fluke Serie 80
80 Series V
Multimeters
Users Manual
May 2004
2004 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies.
Lifetime Limited Warranty
Each Fluke 20, 70, 80, 170 and 180 Series DMM will be free from defects in material and workmanship for its lifetime. As used herein,
“lifetime” is defined as seven years after Fluke discontinues manufacturing the product, but the warranty period shall be at least ten years from
the date of purchase. This warranty does not cover fuses, disposable batteries, damage from neglect, misuse, contamination, alteration,
accident or abnormal conditions of operation or handling, including failures caused by use outside of the product’s specifications, or normal
wear and tear of mechanical components. This warranty covers the original purchaser only and is not transferable.
For ten years from the date of purchase, this warranty also covers the LCD. Thereafter, for the lifetime of the DMM, Fluke will replace the
LCD for a fee based on then current component acquisition costs.
To establish original ownership and prove date of purchase, please complete and return the registration card accompanying the product, or
register your product on http://www.fluke.com. Fluke will, at its option, repair at no charge, replace or refund the purchase price of a
defective product purchased through a Fluke authorized sales outlet and at the applicable international price. Fluke reserves the right to
charge for importation costs of repair/replacement parts if the product purchased in one country is sent for repair elsewhere.
If the product is defective, contact your nearest Fluke authorized service center to obtain return authorization information, then send the
product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk
for damage in transit. Fluke will pay return transportation for product repaired or replaced in-warranty. Before making any non-warranty
repair, Fluke will estimate cost and obtain authorization, then invoice you for repair and return transportation.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL
DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. AUTHORIZED RESELLERS ARE
NOT AUTHORIZED TO EXTEND ANY DIFFERENT WARRANTY ON FLUKE’S BEHALF. Since some states do not allow the exclusion or
limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of
this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the
validity or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett WA 5602 B.D. Eindhoven
2/02 98206-9090 The Netherlands
Table of Contents
Title Page
Introduction.................................................................................................................... 1
Contacting Fluke............................................................................................................ 1
Safety Information ......................................................................................................... 2
The Meter's Features..................................................................................................... 6
Power-Up Options .................................................................................................... 13
Automatic Power-Off................................................................................................. 13
Input Alert Feature ................................................................................................. 13
Making Measurements .................................................................................................. 13
Measuring AC and DC Voltage................................................................................. 13
Zero Input Behavior of True RMS Meters (87).......................................................... 15
Low Pass Filter (87).................................................................................................. 15
Measuring Temperature (87) .................................................................................... 16
Testing for Continuity................................................................................................ 16
Measuring Resistance .............................................................................................. 18
Using Conductance for High Resistance or Leakage Tests ..................................... 20
Measuring Capacitance ............................................................................................ 21
Testing Diodes.......................................................................................................... 22
i
80 Series V
Users Manual
ii
List of Tables
iii
80 Series V
Users Manual
iv
List of Figures
v
80 Series V
Users Manual
vi
Introduction
1
80 Series V
Users Manual
2
Safety Information
• Inspect the test leads for damaged • When making electrical connections,
insulation or exposed metal. Check the connect the common test lead before
test leads for continuity. Replace connecting the live test lead; when
damaged test leads before you use the disconnecting, disconnect the live test
Meter. lead before disconnecting the common
• Do not apply more than the rated test lead.
voltage, as marked on the Meter, • Do not use the Meter if it operates
between the terminals or between any abnormally. Protection may be impaired.
terminal and earth ground. When in doubt, have the Meter serviced.
• Never operate the Meter with the cover • Do not operate the Meter around
removed or the case open. explosive gas, vapor, or dust.
• Use caution when working with voltages • Use only a single 9 V battery, properly
above 30 V ac rms, 42 V ac peak, or 60 V installed in the Meter case, to power the
dc. These voltages pose a shock hazard. Meter.
• Use only the replacement fuses • When servicing the Meter, use only
specified by the manual. specified replacement parts.
• Use the proper terminals, function, and • When using probes, keep fingers
range for measurements. behind the finger guards on the
• Avoid working alone. probes.
• When measuring current, turn off circuit • Do not use the Low Pass Filter option to
power before connecting the Meter in the verify the presence of hazardous
circuit. Remember to place the Meter in voltages. Voltages greater than what is
series with the circuit. indicated may be present. First, make a
voltage measurement without the filter to
detect the possible presence of
hazardous voltage. Then select the filter
function.
3
80 Series V
Users Manual
4
Safety Information
5
80 Series V
Users Manual
Table 2. Inputs
Terminal Description
A Input for 0 A to 10.00 A current (20 A overload for 30 seconds maximum), current frequency, and duty cycle
measurements.
mA µA Input for 0 µA to 400 mA current measurements (600 mA for 18 hrs.) and current frequency.
I Input for voltage, continuity, resistance, diode, capacitance, frequency, temperature (87), and duty cycle
measurements.
6
The Meter's Features
Any Position When the Meter is turned on, the Meter model number briefly appears on the display.
J AC voltage measurement
Press A for low pass filter (K) (87 only).
L DC voltage measurement
e Resistance measurement
Press A for capacitance measurement.
O Diode test
7
80 Series V
Users Manual
Table 4. Pushbuttons
Switch
Button Function
Positon
A N Selects capacitance
(Yellow) Selects temperature (87 only)
M
J Selects ac low pass filter function (87 only)
Power-up Disables automatic power-off feature (Meter normally powers off in 30 minutes).
The Meter reads “PoFF” until A is released.
B Any switch Starts recording of minimum and maximum values. Steps the display through MAX, MIN, AVG
position (average), and present readings. Cancels MIN MAX (hold for 1 second)
Power-up Enables the Meter’s calibration mode and prompts for a password.
The Meter reads “CAL“ and enters calibration mode. See 80 Series V Service Information.
C Any switch Switches between the ranges available for the selected function. To return to autoranging, hold
position the button down for 1 second.
Power-up Enables the Meter’s smoothing feature. The Meter reads “5___” until C is released.
8
The Meter's Features
Switch
Button Function
Positon
D Any switch AutoHOLD (formerly TouchHold) captures the present reading on the display. When a new,
position stable reading is detected, the Meter beeps and displays the new reading.
MIN MAX Stops and starts recording without erasing recorded values.
recording
Frequency Stops and starts the frequency counter.
counter
Power-up Turns on all LCD segments. Releasing D, briefly displays the Meter’s software version.
H Any switch Turns the backlight on, makes it brighter, and turns it off.
position
For Model 87, hold H down for one second to enter the HiRes digit mode. The “HiRes” icon
appears on the display. To return to the 3-1/2 digit mode, hold H down for one second.
HiRes=19,999
9
80 Series V
Users Manual
10
The Meter's Features
G RS AutoHOLD is active.
aom1_af.eps
Figure 1. Display Features (Model 87) H S Display Hold is active.
11
80 Series V
Users Manual
Hertz, Kilohertz
-- 0L Overload condition is detected.
Hz, kHz
12
Making Measurements
W Caution
Power-Up Options
Placing the probes across (in parallel with) a
Holding a button down while turning the Meter on
powered circuit when a lead is plugged into a
activates a power-up option. Table 4 includes the power-
current terminal can damage the circuit you
up options.
are testing and blow the Meter's fuse. This
Automatic Power-Off can happen because the resistance through
the Meter's current terminals is very low, so
The Meter automatically turns off if you do not turn the the Meter acts like a short circuit.
rotary switch or press a button for 30 minutes. If MIN MAX
Recording is enabled, the Meter will not power off. To Making Measurements
disable automatic power-off, refer to Table 4.
The following sections describe how to take
Input Alert Feature measurements with the Meter.
If a test lead is plugged into the mA/µA or A terminal, but Measuring AC and DC Voltage
the rotary switch is not set to the correct current position,
the beeper warns you by making a chirping sound and the Model 87 features true rms readings, which are accurate
display flashes “LEAd”. This warning is intended to stop for distorted sine waves and other waveforms (with no dc
you from attempting to measure voltage, continuity, offset) such as square waves, triangle waves, and
resistance, capacitance, or diode values when the leads staircase waves.
are plugged into a current terminal. The Meter's voltage ranges are 600.0 mV, 6.000 V,
60.00 V, 600.0 V, and 1000 V. To select the 600.0 mV dc
range, turn the rotary switch to mV.
13
80 Series V
Users Manual
REL
AutoHOLD
Hz %
mV
mA
V A
V
voltage, measure the ac voltage first. Note the ac voltage OFF
higher than the ac range. This procedure improves the 10A MAX
FUSED
400mA
FUSED
mV MIN MAX
REL Hz %
4½ DIGITS
1 Second
mV
mA
V A
V LO
V A
OFF
A mA A COM V
+
400mA
10A MAX FUSED
FUSED
aom2f.eps
Figure 2. Measuring AC and DC Voltage
14
Making Measurements
15
80 Series V
Users Manual
Display ranges are –200.0 °C to +1090.0 °C and Press Eto turn the continuity beeper on or off.
–328.0 °F to 1994.0 °F. Readings outside of these ranges
The continuity function detects intermittent opens and
show OL on the Meter display. When there is no
shorts lasting as little as 1ms. A brief short causes the
thermocouple connected, the display also reads OL.
Meter to emit a short beep.
To measure temperature, do the following:
16
Making Measurements
Activates
87 V TRUE RMS MULTIMETER 87 V TRUE RMS MULTIMETER
continuity
beeper
˚C/˚F ˚C/˚F
MIN MAX RANGE AutoHOLD MIN MAX RANGE AutoHOLD
4½ DIGITS 4½ DIGITS
1 Second 1 Second
mV mV
mA mA
V A V A
LO
V A ON LO
V A OFF
(closed) (open)
OFF OFF
A mA A COM V A mA A COM V
400mA 400mA
10A MAX FUSED 10A MAX FUSED
FUSED FUSED
aom4f.eps
Figure 4. Testing for Continuity
17
80 Series V
Users Manual
Measuring Resistance The following are some tips for measuring resistance:
18
Making Measurements
Circuit Power
87 V TRUE RMS MULTIMETER
OFF
1 2
3
Disconnect
˚C/˚F
MIN MAX RANGE AutoHOLD
4½ DIGITS
1 Second
mV
1 2
mA
V A
3
LO
V A
A mA A COM V
400mA
10A MAX FUSED
FUSED
Disconnect
aom6f.eps
Figure 5. Measuring Resistance
19
80 Series V
Users Manual
Using Conductance for High Resistance or To measure conductance, set up the Meter as shown for
Leakage Tests measuring resistance (Figure 5); then press Cuntil
the nS indicator appears on the display.
Conductance, the inverse of resistance, is the ability of a
circuit to pass current. High values of conductance The following are some tips for measuring conductance:
correspond to low values of resistance. • High-resistance readings are susceptible to
The Meter's 60 nS range measures conductance in electrical noise. To smooth out most noisy readings,
nanosiemens (1 nS = 0.000000001 Siemens). Because enter the MIN MAX recording mode; then step to the
such small amounts of conductance correspond to average (AVG) reading.
extremely high resistance, the nS range lets you • There is normally a residual conductance reading
determine the resistance of components up to 100,000 with the test leads open. To ensure accurate
MΩ, 1/1 nS = 1,000 MΩ. readings, use the relative (REL) mode to subtract
the residual value.
20
Making Measurements
Measuring Capacitance
87 V TRUE RMS MULTIMETER
WCaution
To avoid possible damage to the Meter or to Select
the equipment under test, disconnect circuit Capacitance MIN MAX
˚C/˚F
RANGE AutoHOLD
The Meter's capacitance ranges are 10.00 nF, 100.0 nF, A mA A COM V
1.000 µF, 10.00 µF, 100.0 µF, and 9999 µF. 400mA
10A MAX FUSED
FUSED
+
aom10f.eps
Figure 6. Measuring Capacitance
21
80 Series V
Users Manual
Testing Diodes To test a diode out of a circuit, set up the Meter as shown
WCaution in Figure 7. For forward-bias readings on any
semiconductor component, place the red test lead on the
To avoid possible damage to the Meter or to component's positive terminal and place the black lead
the equipment under test, disconnect circuit on the component's negative terminal.
power and discharge all high-voltage
capacitors before testing diodes. In a circuit, a good diode should still produce a forward-
bias reading of 0.5 V to 0.8 V; however, the reverse-bias
Use the diode test to check diodes, transistors, silicon reading can vary depending on the resistance of other
controlled rectifiers (SCRs), and other semiconductor pathways between the probe tips.
devices. This function tests a semiconductor junction by
sending a current through the junction, then measuring A short beep sounds if the diode is good (< .85 V). A
the junction's voltage drop. A good silicon junction drops continuous beep sounds if the reading is ≤ .100 V. This
between 0.5 V and 0.8 V. reading would indicate a short circuit. The display shows
“OL” if the diode is open.
22
Making Measurements
Reverse Bias
87 V TRUE RMS MULTIMETER
Forward Bias 87 V TRUE RMS MULTIMETER
Typical +
˚C/˚F
Reading + ˚C/˚F
MIN MAX RANGE AutoHOLD
MIN MAX RANGE AutoHOLD
Peak MIN MAX
Peak MIN MAX
REL Hz %
REL Hz %
mV mV
mA mA
A V A
V
A LO
V A
LO
V
A mA A COM V A mA A COM V
400mA
400mA FUSED
FUSED 10A MAX
10A MAX FUSED
FUSED
MIN MAX
˚C/˚F
RANGE AutoHOLD
or
REL Hz % Peak MIN MAX
REL Hz %
mV
mV
mA
V A mA
V A
LO
V A
LO
V A
OFF
OFF
A mA A COM V
A mA A COM V
400mA
10A MAX FUSED
FUSED 400mA
10A MAX FUSED
FUSED
aom9f.eps
Figure 7. Testing a Diode
23
80 Series V
Users Manual
To avoid possible electric shock or personal To measure current, refer to Figure 8 and proceed as
injury, never attempt an in-circuit current follows:
measurement where the open-circuit
potential to earth is greater than 1000 V. You 1. Turn off power to the circuit. Discharge all high-
may damage the Meter or be injured if the voltage capacitors.
fuse blows during such a measurement.
2. Insert the black lead into the COM terminal. For
WCaution currents between 6 mA and 400 mA, insert the red
lead into the mA/µA terminal. For currents above
To avoid possible damage to the Meter or to 400 mA, insert the red lead into the A terminal.
the equipment under test:
Note
• Check the Meter's fuses before
measuring current. To avoid blowing the Meter's 400 mA fuse, use
the mA/µA terminal only if you are sure the
• Use the proper terminals, function, and
current is less than 400 mA continuously or less
range for all measurements.
than 600 mA for 18 hours or less.
• Never place the probes across (in
parallel with) any circuit or component
when the leads are plugged into the
current terminals.
24
Making Measurements
˚C/˚F
AutoHOLD mA
3
MIN MAX RANGE
V
mA
A 5
LO
V A
400mA
10A MAX FUSED
FUSED
aom7f.eps
Figure 8. Measuring Current
25
80 Series V
Users Manual
3. If you are using the A terminal, set the rotary switch The following are some tips for measuring current:
to mA/A. If you are using the mA/µA terminal, set the
rotary switch to µA for currents below 6000 µA • If the current reading is 0 and you are sure the Meter
(6 mA), or mA/A for currents above 6000 µA. is set up correctly, test the Meter's fuses as described
under "Testing the Fuses".
4. To measure dc current, press A.
• A current Meter drops a small voltage across itself,
5. Break the circuit path to be tested. Touch the black which might affect circuit operation. You can calculate
probe to the more negative side of the break; touch this burden voltage using the values listed in the
the red probe to the more positive side of the break. specifications in Table 14.
Reversing the leads will produce a negative reading,
but will not damage the Meter.
26
Making Measurements
Measuring Frequency The following are some tips for measuring frequency:
The Meter measures the frequency of a voltage or current • If a reading shows as 0 Hz or is unstable, the input
signal by counting the number of times the signal crosses signal may be below or near the trigger level. You
a threshold level each second. can usually correct these problems by selecting a
lower range, which increases the sensitivity of the
Table 6 summarizes the trigger levels and applications for Meter. In the L function, the lower ranges also have
measuring frequency using the various ranges of the lower trigger levels.
Meter's voltage and current functions.
• If a reading seems to be a multiple of what you
To measure frequency, connect the Meter to the signal expect, the input signal may be distorted. Distortion
source; then press G. Pressing Eswitches the can cause multiple triggerings of the frequency
trigger slope between + and -, as indicated by the symbol counter. Selecting a higher voltage range might solve
at the left side of the display (refer to Figure 9 under this problem by decreasing the sensitivity of the
"Measuring Duty Cycle"). Pressing Dstops and starts Meter. You can also try selecting a dc range, which
the counter. raises the trigger level. In general, the lowest
The Meter autoranges to one of five frequency ranges: frequency displayed is the correct one.
199.99 Hz, 1999.9 Hz, 19.999 kHz, 199.99 kHz, and
greater than 200 kHz. For frequencies below 10 Hz, the
display is updated at the frequency of the input.
Below 0.5 Hz, the display may be unstable.
27
80 Series V
Users Manual
Approximate
Function Range Typical Application
Trigger Level
L 1000 V 100 V
R e E Gd Frequency counter characteristics are not available or specified for these functions.
µAF 600 µA, 6000 µA 30 µA , 300 µA Refer to the measurement tips given before this table.
AF 6 A, 10 A .30 A, 3.0 A
28
Making Measurements
Measuring Duty Cycle frequency function, you can change the slope for the
Meter's counter by pressing E.
Duty cycle (or duty factor) is the percentage of time a
signal is above or below a trigger level during one cycle For 5 V logic signals, use the 6 V dc range. For 12 V
(Figure 9). The duty cycle mode is optimized for switching signals in automobiles, use the 60 V dc range.
measuring the on or off time of logic and switching For sine waves, use the lowest range that does not result
signals. Systems such as electronic fuel injection in multiple triggering. (Normally, a distortion-free signal
systems and switching power supplies are controlled by can be up to ten times the amplitude of the selected
pulses of varying width, which can be checked by voltage range.)
measuring duty cycle.
If a duty cycle reading is unstable, press MIN MAX; then
To measure duty cycle, set up the Meter to measure scroll to the AVG (average) display.
frequency; then press Hz a second time. As with the
+Slope -Slope
Trigger Point Trigger Point
30% Above 70% Below
+Slope -Slope
100%
iy3f.eps
Figure 9. Components of Duty Cycle Measurements
29
80 Series V
Users Manual
Pulse Width = % Duty Cycle ÷ 100 In the 60 V range, for example, the major divisions on the
(in seconds) Frequency scale represent 0, 15, 30, 45, and 60 V. An input of -30 V
lights the negative sign and the segments up to the
middle of the scale.
30
HiRes Mode (Model 87)
Zoom Mode (Power Up Option Only) For peak adjustments, set the Meter to the desired
function, connect the leads to the circuit under test; then
To use the Rel Zoom Bar Graph: press F. The display reads zero. As you adjust for a
1. Hold down F while turning the Meter on. positive or negative peak, the bar graph length increases
The display reads “2rEL”. to the right or left of zero. If an overange symbol lights
(< >), press F twice to set a new reference; then
2. Select the relative mode by pressing F continue with the adjustment.
again.
HiRes Mode (Model 87)
3. The center of the bar graph now represents zero
and the sensitivity of the bar graph increases by On a Model 87 Meter, pressing H for one second causes
a factor of 10. Measured values more negative the Meter to enter the high-resolution (HiRes), 4-1/2 digit
than the stored reference activate segments to mode. Readings are displayed at 10 times the normal
the left of center; values more positive activate resolution with a maximum display of 19,999 counts. The
segments to the right of center. HiRes mode works in all modes except capacitance,
frequency counter functions, temperature, and the 250 µs
Uses for the Zoom Mode (peak) MIN MAX modes.
The relative mode, combined with the increased To return to the 3-1/2 digit mode, press H again for one
sensitivity of the bar graph's zoom mode, helps you make second.
fast and accurate zero and peak adjustments.
31
80 Series V
Users Manual
MIN MAX Recording Mode since the start of recording (overloads are discarded).
The average reading is useful for smoothing out unstable
The MIN MAX mode records minimum and maximum inputs, calculating power consumption, or estimating the
input values. When the inputs go below the recorded percentage of time a circuit is active.
minimum value or above the recorded maximum value,
the Meter beeps and records the new value. This mode Min Max records the signal extremes lasting longer than
can be used to capture intermittent readings, record 100 ms.
maximum readings while you are away or record
readings while you are operating the equipment under Peak records the signal extremes lasting longer than 250
test and cannot watch the Meter. MIN MAX mode can µs.
also calculate an average of all readings taken since the
MIN MAX mode was activated. To use MIN MAX mode,
Smooth Feature (Power Up Option Only)
refer to the functions in Table 7. When the input signal changes rapidly, “smoothing”
provides a steadier reading on the display.
Response time is the length of time an input must stay at
a new value to be recorded. A shorter response time To use the smooth feature:
captures shorter events, but with decreased accuracy.
Changing the response time erases all recorded 1. Hold down C while turning the Meter on. The
readings. Model 83 has 100 millisecond response time; display will read “5___” until C is released.
Model 87 has 100 millisecond, and 250 µs (peak)
response times. The 250 µs response time is indicated
2. The smooth icon (g) will appear on the left side
by "p" on the display.
of the display to let you know that smoothing is
active.
The 100 millisecond response time is best for recording
power supply surges, inrush currents, and finding
intermittent failures.
32
Smooth Feature (Power Up Option Only)
Enter MIN MAX recording mode. The Meter is locked in the range displayed before you
entered MIN MAX mode. (Select the desired measurement function and range before
B
entering MIN MAX.) The Meter beeps each time a new minimum or maximum value is
recorded.
B
Step through maximum (MAX), minimum (MIN), average (AVG) and present values.
(while in MIN MAX mode)
Model 87 only: Select 100 ms or 250 µs response time. (The 250 µs response time is
E
indicated by pon the display.) Stored values are erased. The present and AVG
PEAK MIN MAX
(average) values are not available when 250 µs is selected.
D Stop recording without erasing stored values. Press again to resume recording.
B
Exit MIN MAX mode. Stored values are erased. The Meter stays in the selected range.
(hold for 1 second)
33
80 Series V
Users Manual
34
Maintenance
35
80 Series V
Users Manual
REL
AutoHOLD
Hz %
6F22, or 006P).
mV
mA
XWWarning
V A
LO
V A
OFF
Touch top half A mA A COM V To avoid false readings, which could lead to
of input contacts 10A MAX
FUSED
400mA
FUSED
possible electric shock or personal injury,
replace the battery as soon as the battery
indicator (b) appears. If the display shows
“bAtt” the Meter will not function until the
battery is replaced.
87 V TRUE RMS MULTIMETER
4½ DIGITS
mV
mA
1.005 kΩ LO
V
V A
A
leads from the terminals.
Replace fuse: OL OFF
36
Service and Parts
Replacing the Fuses 5. Remove the fuse by gently prying one end loose,
then sliding the fuse out of its bracket.
Referring to Figure 11, examine or replace the Meter's
fuses as follows: 6. Install ONLY specified replacement fuses with the
amperage, voltage, and speed ratings shown in
1. Turn the rotary switch to OFF and remove the test Table 8.
leads from the terminals.
7. Verify that the rotary switch and the circuit board
2. Remove the battery door by using a standard-blade switch are in the OFF position.
screwdriver to turn the battery door screws one-
quarter turn counterclockwise. 8. Replace the case top, ensuring that the gasket is
properly seated and case snaps together above the
3. Remove the three Phillips-head screws from the LCD (item A).
case bottom and turn the case over.
9. Reinstall the three screws and the battery door.
4. Gently lift the input terminal-end of the top case to Secure the door by turning the screws one-quarter
separate the two halves of the case. turn clockwise.
Service and Parts
If the Meter fails, check the battery and fuses. Review
this manual to verify proper use of the Meter.
37
80 Series V
Users Manual
F1
F2
aom12f.eps
Figure 11. Battery and Fuse Replacement
38
Service and Parts
39
80 Series V
Users Manual
40
Service and Parts
TL75
MP8
Test Lead Set
MP5 MP31
MP66
U5
J1-2
CR6
S2 MP2
MP9
Alligator Clips
MP13 MP41
J3 MP64
F2 MP81 MP71
MP27-30
W F1
MP4
MP14 Holster
MP15
H5-9 (5)
Tilt Stand
MP6
MP10-11
TM1
H2-4 (3)
BT1
MP22 TM3 TM2
MP390-391
aom015c.eps
Figure 12. Replaceable Parts
41
80 Series V
Users Manual
Table 9. Accessories
Item Description
AC72 Alligator Clips for use with TL75 test lead set
42
Specifications
Specifications
Maximum Voltage between any Terminal and Earth Ground: 1000 V rms
W Fuse Protection for mA or µA inputs: 44/100 A, 1000 V FAST Fuse
W Fuse Protection for A input: 11 A, 1000 V FAST Fuse
Display: Digital: 6000 counts updates 4/sec; (Model 87 also has 19,999 counts in high-resolution mode). Analog: 32 segments, updates
40/sec. Frequency: 19,999 counts, updates 3/sec at > 10 Hz.
Temperature: Operating: -20 °C to +55 °C; Storage: -40 °C to +60 °C
Altitude: Operating: 2000 m; Storage: 10,000 m
Temperature Coefficient: 0.05 x (specified accuracy)/ °C (< 18 °C or > 28 °C)
Electromagnetic Compatibility: In an RF field of 3 V/m total accuracy = specified accuracy.
Relative Humidity: 0 % to 90 % (0 °C to 35 °C); 0 % to 7 0% (35 °C to 55 °C)
Battery Type: 9 V zinc, NEDA 1604 or 6F22 or 006P
Battery Life: 400 hrs typical with alkaline (with backlight off)
Vibration: Per MIL-PRF-28800 for a Class 2 instrument
Shock: 1 Meter drop per IEC 61010-1:2001
Size (HxWxL): 1.25 in x 3.41 in x 7.35 in (3.1 cm x 8.6 cm x 18.6 cm)
Size with Holster and Flex-Stand: 2.06 in x 3.86 in x 7.93 in (5.2 cm x 9.8 cm x 20.1 cm)
Weight: 12.5 oz (355 g)
Weight with Holster and Flex-Stand: 22.0 oz (624 g)
Safety: Complies with ANSI/ISA S82.01-2004, CSA 22.2 No. 1010.1:2004 to 1000 V Overvoltage Category III, IEC 664 to 600 V
Overvoltage Category IV. UL listed to UL3111-1. Licensed by TÜV to EN61010-1.
43
80 Series V
Users Manual
Detailed Specifications
For all detailed specifications:
Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18° C to 28° C, with relative humidity up to 90
%, for a period of one year after calibration. For Model 87 in the 4 ½-digit mode, multiply the number of least significant
digits (counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range. Model 87 is true rms
responding. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 %
full scale) typical, for a crest factor up to 3.
44
Specifications
45
80 Series V
Users Manual
Accuracy
Function Range Resolution
Model 83 Model 87
46
Specifications
47
80 Series V
Users Manual
48
Specifications
G 3.000 V 0.001 V ± (2 % + 1)
49
80 Series V
Users Manual
50
Specifications
e 1000 V rms < 1.3 V dc < 450 mV dc < 1.3 V dc 1 mA 100 µA 10 µA 1µA 1 µA 0.5 µA
1. 106 V Hz max
51
80 Series V
Users Manual
120 ms to 80 % Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of range
(ac functions)
250 µs (peak) Specified accuracy ± 100 counts for changes > 250 µs in duration
(Model 87 only)1 (add ± 100 counts for readings over 6000 counts)
(add ± 100 counts for readings in Low Pass mode)
52