You are on page 1of 12

Evolve your care

Philips MX16EVO CT specifications


Table of contents

1 Introduction 3
1.1 Imaging 2.0 3

2 Gantry 4

3 X-ray 4
3.1 Tube 4
3.2 Generator 4

4 Patient couch 5

5 Operator console 5

6 Host computer system 6

7 System performance 7
7.1 EVOEye algorithm 7
7.2 MAR artifact reduction 7
7.3 Scan modes 8
7.4 Axial scanning 8
7.5 Spiral scanning 8
7.6 Other specifications 8

8 Applications 9
8.1 Standard applications 9
8.2 Optional applications 9

9 Site planning 10
9.1 Dimensions and weights 10
9.2 Power supply requirements 10
9.3 Environmental requirements 10

10 Other configurations 11

2
1. Introduction

The Philips MX16EVO CT scanner is evolving the way that


clinicians are able to care for patients by offering them Imaging 2.0
exceptional image quality with low dose, high patient
throughput, and attractive total cost of ownership.
Exclusive technologies make it possible for the MX16EVO Imaging 2.0 from Philips is an innovative approach
to deliver performance and productivity that are at the to clinical integration and collaboration, patient
top of its class, allowing institutions to excel at routine focus, and improved economic value to help
applications and also move into neuro and CTA imaging
with high patient throughput. The system features
radiologists worldwide find success with new ways
a compact design, as well as the most capable and of practicing medicine.
reliable tube in its class. MX16EVO is a truly outstanding
workhorse that is easy for hospitals to operate and own.

3
2. Gantry

Feature Specification
Slip ring Low voltage slip ring; capacitive data transfer
Aperture 700 mm ± 10 mm
Scan localizer Sagittal and transverse localizer, can perform PA and lateral scan
(laser positioning lamp)
Scan collimation 16 x 1.5 mm, 12 x 1.5 mm, 10 x 1.5 mm, 16 x 0.75 mm, 12 x 0.75 mm,
10 x 0.75 mm, 8 x 0.75 mm, 2 x 0.75 mm, 4 x 0.75 mm
Data acquisition rate 1160, 2320 views/rotation
Scan speed – s/360° 0.5 s (optional), 0.6 s, 0.75 s, 1.0 s, 1.5 s, 2 s
Maximum SFOV 500 mm
Detector type GOS solid
Number of detectors 672 x 24
Detector arrangement 24 rows
Detector channels per row 672
Data transmit Digital

3. X-ray

The MX16EVO is designed for long tube life with no tube cooling delay.

3.1 Tube
Feature Specification
Tube type 5 MHU
Focal spots 0.5 x 1.3 mm (small); 1.0 x 1.3 mm (large)
Anode rotation speed 6300 rpm
ISO center to focal spot distance 570 mm
Anode heat capacity 5.0 MHU
Anode continuous heat dissipation 6 kW
Maximum anode heat dissipation rate 9.6 kW
Cooling mode Oil cooled

3.2 Generator
Feature Specification
Maximum kW output 50 kW
Range of mA output 30 mA – 420 mA
KV switches 90 KV, 120 KV, and 140 KV (average tolerance ≤10% )

4
4. Patient couch

Feature Specification
Range of vertical movement 430 mm to 970 mm
Maximum range of horizontal movement 1580 mm
Horizontal movement speed 0.375 – 100 mm/sec, tolerance ≤10%
Horizontal movement accuracy ≤ 0.25 mm
Vertical movement speed 9 mm – 15 mm/second
Maximum scanning range 1500 mm with head holder
Maximum load capacity 200 kg (440 lbs)

5. Operator console

Feature Specification
Recon FOV 50-500 mm
Image matrix 512 x 512, 768 x 768, 1024 x 1024
Maximum display matrix 1024 x 1024
Image recon time Up to 20 images/second
Maximum number of images displayed 16
Image enlargement mode Bilinear interpolation
CT number range -1024 ~ +3072
Grayscale range 256 (24-bit color)

5
6. Host computer system

Feature Specification
CPU frequency ≥ 2.4 GHz, multi core (advanced workstation)
Graphic processor 1 GPU (specific CPU on graphics card)
Memory ≥ 16 GB
Hard disk capability ≥ 1 TB
Maximum image storage capacity Storage capacity ≥ 800,000 images, 512 x 512
(512 KB/image)
DVD – RW driver 7500 images of 512 matrix per 4 GB disk
User interface Multi-language
Monitor 19’ LCD
Monitor resolution matrix 1280 x 1024 (256 gray level, 24-bit color space)
DICOM 3.0 configuration Standard configurations:
• DICOM print/store
• DICOM send
Optional configurations:
• DICOM MPPS/DICOM modality worklist (HIS/RIS)
AutoVoice Standard
AutoFilm Standard

6
7. System performance

Designed for productivity and performance, the MX16EVO offers the widest detector coverage
in its class, high pitch at high quality with ASTR (Adaptive Spiral Tracking Reconstruction Algorithm),
and fast reconstruction speeds of up to 20 ips, so you can scan more patients each day.

7.1 EVOEye algorithm


The exclusive EVOEye algorithm allows
for excellent images with reduced noise.

The 1024 matrix enables high resolution scanning,


while EVOEye improves low-contrast resolution.

7.2 MAR artifact reduction


Exclusive Metal Artifact Reduction (MAR) technology data, enhancing visualization of surrounding anatomy
helps isolate the effects of metal objects in the image for increased diagnostic confidence.

Without MAR With MAR

7
7.3 Scan modes
The MX16EVO allows for surview scan, axial scan,
and spiral scan modes.

7.4 Axial scanning


Feature Specification
FOV 500 mm
Scan time (s/revolution) 0.5 s (optional), 0.6 s, 0.75 s, 1s, 1.5 s, 2 s,

7.5 Spiral scanning


Feature Specification
Scan time (s/revolution) 0.5 s (optional), 0.6 s, 0.75 s, 1s, 1.5 s, 2 s,
Maximum scan time 100 s
Maximum continuous revolutions 200 rev
Maximum continuous scan length 1500 mm
Range of pitch 0.5 to 1.5
Recon matrix 512 x 512, 768 x 768, 1024 x 1024
Maximum display matrix 1024 x 1024

7.6 Other specifications


Feature Specification
Dynamic scan rate 1920 slices/min
Spatial resolution 15 lp/cm @ 0% MTF; CTDIw (head phantom) should be
no more than 40 mGy
Low-contrast resolution 2 mm @ 0.3%; CTDIw (head phantom) should be no
more than 40 mGy
Image noise ≤ 0.35%; CTDIw (head phantom) should be no more
than 40 mGy system phantom
Uniformity of CT number Water CT number is +4HU
Standard CT number range -1024 ~ +3072 HU
CTDIw dose • Head: 16 mGy/100 mAs
• Body: 8.5 mGy/100 mAs
(120 KV, 16 x 1.5 mm scan collimation)
Recon speed Up to 20 images per second (ips)

8
8. Applications

The MX16EVO provides the capabilities to allow institutions to excel at routine applications
and also move into neuro and CTA imaging and increase patient throughput.

8.1 Standard applications 8.2 Optional applications


3-D SSD VA (Vessel analysis)
MIP/MinIP CT perfusion package for stroke
MPR Dental planning
Curved MPR Bolus tracking
AutoFilm CT reporting
AutoVoice SAS
CT angiography CCT (single)
Volume rendering Barcode reader
DoseRight pediatric protocols MPPS
Virtual endoscopy Worklist

9
9. Site planning

Siting is simple
This compact system has similar siting requirements to
most existing 1-, 2-, and 4-slice scanner rooms, and is
designed for fast, simplified, cost-effective installation.

9.1 Dimensions and weights


Feature Specification
Gantry 2210 mm (L) x 890 mm (W) x 1890 mm (H)
Gantry weight ≤ 1900 kg
Gantry package 2370 mm (L) x 1030 mm (W) x 2250 mm (H)
Couch 2420 mm (L) x 575 mm (W) x 1055 mm (H)
Couch weight ≤ 430 kg
Couch package 2570 mm (L) x 970 mm (W) x 1230 mm (H)
Console table 1400 mm (L) x 800 mm (W) x 760 mm (H)

9.2 Power supply requirements


Feature Specification
Power capacity 80 kVA
Input voltage • 380/400 VAC 3-phase 5-line
• 3-phase 4-line (export is equipped with isolate
transformer); power supply from below options:
200/208/220/230/240/380/400/415/440/460/480 VAC
Voltage variation Tolerance ≤ ± 10% (supply power)
Drop with loading ≤ 5% of the rating
3-phase unbalance ≤ 5%
Frequency 50 Hz/60 Hz ± 1 Hz
Grounding resistance Independent grounding resistance ≤ 4Ω;
common grounding resistance ≤ 1Ω

9.3 Environmental requirements


Feature Specification
Minimum area of scanning room 18 m2
Minimum area of operator room 4.8 m2
Recommended room size 20 m2
Minimum height of ceiling 2300 mm
Temperature of scanning room • Scan room 18°C ~ 24°C
• Control room 18°C ~ 2 8°C
(Gradient less than 5°C/hour)
Humidity of scanning room • Scan room 30% ~ 60%
• Control room 20% ~ 80% (non-condensing)
Atmospheric pressure 70 kPa ~ 106 kPa
Temperature of transportation and storage -20°C ~ +55°C; shock is less than 10 G with package
Humidity of transportation and storage • 10% ~ 90% non-condensing
• Atmospheric pressure ~ 50 kPa ~ 106 kPa
Running noise ≤ 70 dB(A) (1 meter distance)

10
10. Other configurations

Feature Specification
Laser camera DICOM 3.0 Interface (not provided by Philips)
High-pressure injector Supports (not provided by Philips):
• MEDRAD Envision
• Medtron Injektron 82 CT
• Nemoto Auto Enhance A-60 injector
• SHENZHEN DONGDA NSJ-200C
• MEDRAD Vistron
• Tyco CT-9000
• Covidien: Mallinckrodt Optivantage
• MEDRAD: Medrad Stellent D(SX)
• Other injector models on request
Power conditioner Optional configuration (for domestic systems,
with isolation transformer)
Isolation transformer Optional configuration
UPS for console Optional configuration

11
Philips Healthcare
is part of Royal Philips

How to reach us
www.philips.com/healthcare
healthcare@philips.com

Asia
+49 7031 463 2254

Europe, Middle East, Africa


+49 7031 463 2254

Latin America
+55 11 2125 0744

North America
+1 425 487 7000
800 285 5585 (toll free, US only)

Please visit www.philips.com/MX16EVO

© 2013 Koninklijke Philips N.V.


All rights are reserved.

Philips Healthcare reserves the right to make changes in specifications and/or to discontinue any product at any time
without notice or obligation and will not be liable for any consequences resulting from the use of this publication.

Printed in The Netherlands.


4522 962 88838 * DEC 2013