TECHNICAL DATA
Technical Data
5w (7z) - digital fast beam densitometer
Generator
X-ray Continuous
High Frequency Monoblock
Generator
Manufacturer IMD Generators
Cooling system Immersion in oil + cooling fans
Maximum/Nominal High
110kV/90kV
Voltage
Maximum/Nominal Tube
2.4mA/2mA
Current
X-Ray Tube
Type Tungsten Fixed Anode
Localization Under the patient
Anode angle 12°
Anode-Cathode direction Horizontal
X-ray Beam Pencil Beam
Focal Point Dimensions 0.5 mm x 0.5 mm
X-ray Spectrum peaks 43 keV & 70 keV (Filtering: Samarium 200 µm + Aluminium 2 mm)
Anode Capacity 40 kJ (54kHU)
Tube Housing Thermal
500 kJ (675kHU)
Capacity
Source Collimator
Material Lead
height 5mm
Diameter 1 mm & 2 mm
Collimator-patient
50 mm
distance
Tube-Patient distance 160 mm
Shutter 5 mm Lead
tda_5w(7z)_m_02
2
Technical Data
5w (7z) - digital fast beam densitometer
Detector Collimator
Material Brass
Height 38 mm
Diameter 30 mm
Detector
Quantity 1 (Scintillator & Photomultiplier)
Type of Detection High Performance Photomultiplier Tube
Material Cerium Bromide (CeBr3)
Localization Above the patient
Scanner
Scanning Method Rectilinear scan
Acquisition Method
Method Dual Energy X-ray Absorptiometry (DXA)
Type Digital Fast Beam Technology
Acquisition Windows
Scan Window Size Adjustable to patient’s morphology
tda_5w(7z)_m_02 3
Technical Data
5w (7z) - digital fast beam densitometer
Software - Diagnostic
Intuitive solution for workflow improvement and optimized patient care
tda_5w(7z)_m_02
4
Technical Data
5w (7z) - digital fast beam densitometer
tda_5w(7z)_m_02 5
Technical Data
5w (7z) - digital fast beam densitometer
Tools
FRAX :
Fracture risk tool from The University of Sheffield, to evaluate the 10 years probability of
osteoporotic fractures. Fracture risk information using Hip Structural Analysis (HSA).
DVO recommendations (Certified by DachVerband Osteologie).
Reference :
Calculation of standardized BMD (comparison to NHANES III normative data of the femur).
Reference curve : Displays BMD according to the age for the examined regions and supplies
T-score and Z-score values as diagnosis values. Body Composition according to NHANES
(2009).
Reference population (reference normality curve): Caucasian, Asian, NHANES III, African,
Turkish, Hispanic, Japanese and Korean.
Personalized multiple reference populations (normality curves editor).
Advanced Morphometric tools (distance, angle and area) :
Automated Hip Structural Analysis (HAS): Hip Axis Length (HAL), Femoral Neck Axis Length
(FNAL), Intertrochanter to Femoral Head Center Distance (IH) and Femoral axis versus Neck
axis Angle (FNA)
Personalized ROI & Image tools :
Manual analysis of BMD in any site, BMC, Area.
Contrast, Brightness and zoom for Image display. Density display in colour scale.
Exclusion of non-bone region from calculation with advanced masking (metal, calcification,
etc...)
Multi-user :
Different profiles (operator, physician...etc) can be configured with security access.
Several workstations can be linked to an equipment
Exportation & Importation :
Patient’s data follow-up: Data base importation from other devices + previous data input.
Customizable automatic/semi-automatic database archiving
Print : Multi-report for comparative purposes with personalization
Customizable reports (header, footer, predefined forms, letters, etc.).
Detailed colour print out of reports (bone + reference curve + analysis report + operator
comments + patient and physician letters + follow-up), configurable by the physician.
Email or fax reports sending
Automatic and customizable letter (patient and physician)
Patient follow-up graphs, patient’s pictures & parameters evolution
Multi-language software: Chinese simplified, Czech, German, English, Spanish, French, Italian,
Polish, Russian, Taiwan & Turkish (other languages can be translated)
Customizable user interface (colours, trend, results, print-out, etc.)
Help menu
Telemaintenance and fault detection software (network connection required)
tda_5w(7z)_m_02
6
Technical Data
5w (7z) - digital fast beam densitometer
Examination
Automatic & Manual selection of the Region Of Interest (ROI)
Low Dose adapted to patient Morphology
Radiation Dose
(Effective/DAP/Input Dose displayed in the report)
Staff Dose at 1m Spine (90s) : 0.0264 µSv
(Annual Staff Dose 0.678 mSv*) Proximal Femur (90s) : 0.0301 µSv
*6 patients/h, 1 spine & 1 femur, 2000h/year Whole Body (300s) : 0.0009 µSv
Normal mode Paediatric mode
Dose to patient Femur < 0.28 µSv (0,028 mrem) -
Using ICRP60 weighting factors in Dual Femur < 0.6 µSv (0.06 mrem) -
AP Spine
Normal mode
Using ICRP103 weight factors in < 2 µSv (0.2 mrem) 1.1149 µSv
Paediatric mode Forearm < 0.2 µSv (0.001 mrem) -
Whole Body < 0.02 µSv (0.002 mrem) 0.0323 µSv
Normal mode Paediatric mode
Femur Pixel Size Scan time DVA Pixel Size Scan time
Fast 3x1.5 mm² 60 s Fast 3x1.5 mm² 3 min
Normal 2x1 mm² 90 s Normal 2x1 mm² 5 min
Precision 1x1 mm² 3 min Precision 1x1 mm² 8 min
AP Spine Pixel Size Scan time Forearm Pixel Size Scan time
Fast 3x1.5 mm² 60 s Fast 3x1.5 mm² 60 s
Normal 2x1 mm² 90 s Normal 2x0.8 mm² 90 s
Precision 1x1 mm² 3 min Precision 1x0.8 mm² 3 min
Lateral Pixel Size Scan time Whole Pixel Size Scan time*
Spine Body
Fast 3x1.5 mm² 60 s Fast 20x10 mm² 4 min
Normal 2x1 mm² 90 s Normal 12x6 mm² 7 min
Precision 1x1 mm² 3 min Precision 8x4 mm² 10 min
*depending on BMI
tda_5w(7z)_m_02 7
Technical Data
5w (7z) - digital fast beam densitometer
Method of Examination
Preset exam modes: Exam parameters adjusted automatically based on patient’s
morphology (Thin, Healthy and Overweight)
Customized options: Scan speed, scan area and resolution (up to 0.50 mm)
Automatic and manual selection of the Region Of Interest (ROI) = Image acquisition with
Intelligent Scan
IntelliScan: Smart reduction of the scan window and of the examination time
Patient Positioning
Laser Light
PC controlled easy scan repositioning
Accessories
Cube, Mattress, DVA Cushion, Spine phantom, Calibration phantom
PC data cable, Power cable, USB Adapter
Double foot support, Foot support strap
tda_5w(7z)_m_02
8
Technical Data
5w (7z) - digital fast beam densitometer
Computer configuration
Minimal Hardware Configuration
Windows 7, 64 bits, or more recent
Operating System
(Windows 10, 64 bits, recommended)
Processor Intel Core i5 or better
RAM 8 GB
Hard Disc 300 GB (About 60.000 standard exams)
Archiving External Hard-Drive or Network Share
Monitor 15” or 17” SVGA or larger (1024x768 minimum)
Printer Color: Any printer compatible with windows 7 or higher
2 LAN port for communication & DICOM
Connectivity LAN for DICOM can be supplied by USB to LAN converter +
1 LAN for 3D-DXA (Option Box 3D)
Connectivity
(Medix90 Conformance Statement available on request)
tda_5w(7z)_m_02 9
Technical Data
5w (7z) - digital fast beam densitometer
Environmental Data
& Electrical Specifications
Environmental Data
Operating Temperature 10 - 30°C (50-86°F)
Operating Humidity 30% - 75% (without condensation)
Operating Pressure 700 - 1060 hPa
Storage Temperature 0 - 40°C (32-104°F)
Storage Humidity 10% - 80% (without condensation)
Storage Pressure 700 – 1060 hPa
Radioprotection No external shielding required
Electrical Specifications
Frequency 50 - 60 Hz
Power Consumption 560W
Recommended UPS
(Medix90 without PC &
Min 700 W, 1000 VA
Printer)
tda_5w(7z)_m_02
10
Technical Data
5w (7z) - digital fast beam densitometer
Physical Description
Characteristics
Room Layout
tda_5w(7z)_m_02 11
Technical Data
5w (7z) - digital fast beam densitometer
Attached Documents
Dual Femur
DVA Spine
tda_5w(7z)_m_02
12
Technical Data
5w (7z) - digital fast beam densitometer
AP DVA Spine
In ROW Scan
tda_5w(7z)_m_02 13
Technical Data
5w (7z) - digital fast beam densitometer
Triple-Exam
SCM DES SALYENSReport 1/3
Left Femur - Total Hip 0.866 28.90 33.39 -1.3 (-19%) -1.0 (-14%)
3 - Right Forearm - Total 0.564 11.33 20.09 -0.1 ( -1%) 0.9 ( 8%)
2D Fan Beam DEXA Printing date/hour 06-14-2018 14:44:42 High Energy: 70 keV, Low Energy: 43 keV Ver.V4.1.9.0 demo 06-14-2018 / H0 0 - SN: 900 OT 100
Normality Curve : Caucasian F RightFemur from DMS normality curves, 2003/2004.
* Effective and input doses are measured for rachis and femur
in normal mode with normal patient (18<BMI<25). dosimetry (February 2009).
Patient : WHOLE BODY Xxxx12 Height : 179 cm Weight : 74 kg Prescriber : Self ref Patient : WHOLE BODY Xxxx12 Height : 179 cm Weight : 74 kg Operator : Default
Patient's ID : Ethnic : Caucasian Sex : Male Scan date: 09-03-2014 12:52:36 Patient's ID : Ethnic : Caucasian Sex : Male Prescriber : Self ref
Birth Date : 05-24-1976 Current Age : 42 Years Analysis date : 09-03-2014 12:52:36 Birth Date : 05-24-1976 Current Age : 42 Years Physician : Rachael
*Effective/DAP/Input dose: 0.1 µSv/3.28 mGy.cm²/0.27 µGy BMI: 23.10 kg/m² Health [18.5 - 24.9]
Composition indices
Total
ROI Tissues Tissues Tissues Area Fat Lean BMC BMC Area Total mass
(%Fat) (g) (cm²) (g) (g) (g) (cm²) (kg) Total
Left Arm 2.4 4662 1737.0 113 4549 351.8 293.4 5.0 Age Fat Change/previous Change/ref. Change/ref.
Right Arm 3.2 4409 1452.5 140 4269 345.3 287.0 4.8 Scan date
(years) (g) (%) (%) (%/year)
Left Ribs 5.1 12748 797.7 651 12097 361.3 121.4 13.1 09-03-2014 38.3 4130 -45.2 -74.8 -229.5
Right Ribs 5.0 3519 322.8 175 3343 149.4 61.0 3.7 07-09-2014 38.1 7536 -53.9 -53.9 -312.7
T Spine 5.3 2284 148.6 120 2164 115.4 77.2 2.4 05-07-2014 38.0 16362 0.0 reference reference
L Spine 2.9 565 67.8 16 549 20.2 6.4 0.6
Pelvis 5.1 7471 429.4 378 7094 257.6 108.3 7.7
Left Leg 3.8 13923 2806.9 535 13388 886.7 510.0 14.8
Right Leg 4.0 14861 2806.4 600 14261 952.8 505.2 15.8
SubTotal 4.2 64441 10569.2 2728 61713 3440.4 1969.9 67.9
Head 20.0 3913 1480.2 783 3131 458.6 242.2 4.4 Body Mass Index (BMI) = 23.10
Total 5.1 68354 12049.3 3511 64843 3899.0 2212.2 72.3 13 18.5 25 30 (BMI) 40
Android 5.3 5198 317.0 277 4921 107.7 29.1 5.3 Underweight Normal Overweight Obesity
Gynoid 6.3 9302 740.1 586 8715 383.4 135.8 9.7 42 59 80 96 (kg) 128
2D Fan Beam DEXA Printing date/hour 06-14-2018 14:49:29 High Energy: 70 keV, Low Energy: 43 keV Ver.V4.1.9.0 demo 06-14-2018 / H0 0 - SN: 900 OT 100
Body composition references values from NHANES, 2009,BMR according to Harris-Benedict 2D Fan Beam DEXA Printing date/hour 06-14-2018 14:46:59 High Energy: 70 keV, Low Energy: 43 keV Ver.V4.1.9.0 demo 06-14-2018 / H0 0 - SN: 900 OT 100
Body composition references values from NHANES, 2009,BMR according to Harris-Benedict
tda_5w(7z)_m_02
14
Technical Data
5w (7z) - digital fast beam densitometer
HSA
tda_5w(7z)_m_02 15
Technical Data
5w (7z) - digital fast beam densitometer
3D-DXA
tda_5w(7z)_m_02
16
Technical Data
5w (7z) - digital fast beam densitometer
tda_5w(7z)_m_02 17
Photos non-contractual | TDA_5W(7Z)_M_02
Medix90 is a Class IIb medical device manufactured by Medilink. The conformity assessment has been carried out by SGS, n°0120
and bears the CE marking. It is intended for bone densitometry evaluation. Read the operating instructions carefully. This product is
certified as medical devices in the European Union according to Directive 93/42/EEC on medical devices by SGS CE0120, exclusively
for the indication of bone densitometry evaluation. Other non-medical uses attributed to this device does not fall within the scope
of CE certification, and users should be aware that the performance and/or safety of the product has not been evaluated by SGS
for these purposes. 3D-DXA: Class I medical device, manufactured by DMS. Photos are non-contractual.