ECLOS
Daily QC Phantom
INSTRUCTION MANUAL
Before using this system, be sure to thoroughly read this manual and make yourself
familiar with this system.
After reading this manual, keep it in a legible place adjacent to the system.
Q1E-BW1311-1
QS4 45462
This manual explains the Daily QC Phantom for quality control of the Hitachi Whole-body
X-ray CT System.
This symbol indicates possible danger that could result in light or moderate
injury, partial equipment damage, or computer data loss.
Revision history:
First edition: April 2007
QS4 45462
Preface
This instruction manual describes the Daily QC Phantom for quality control of the Hitachi
Whole-body X-ray CT System. Before using the Daily QC Phantom, read this instruction
manual thoroughly to understand functions of the Daily QC Phantom for correct and safe use.
The Daily Check ROI function is used for evaluation of CT images of the Daily QC Phantom;
therefore, keep the relevant instruction manual on hand for reference.
Q1E-BW1311 i
QS4 45462
ii Q1E-BW1311
QS4 45462
Contents
Preface ......................................................................................................................................................i
1. General ........................................................................... 1
4. Specification.................................................................. 17
Appendix ........................................................................... 19
A.1 Entering the CT number ..................................................................................................................19
INDEX ............................................................................... 25
Q1E-BW1311 iii
QS4 45462
iv Q1E-BW1311
QS4 45462
1 General
1. General
The Daily QC Phantom can be used for quality control of the Whole-body X-ray CT System.
The CT number can be checked using the CT image of the Daily QC Phantom.
Q1E-BW1311 1
QS4 45462
1 General
2 Q1E-BW1311
QS4 45462
2 Composition of the Daily QC Phantom
CT number label
Positioning mark
(3 locations)
Bearing
Acrylic Polyethylene
Teflon Air
Water
Q1E-BW1311 3
QS4 45462
2 Composition of the Daily QC Phantom
Reservoir
4 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
Inlet
When it is hard to pour distilled water into the reservoir due to air inside, remove
the air vent screw for easy filling.
Air bubbles in the Daily QC Phantom can be removed relatively easily if adding
the surface active agent (e.g. Fuji WELL, Kodak PHOTO-FLOW of
concentration 0.1%, etc).
Q1E-BW1311 5
QS4 45462
3 Using the Daily QC Phantom
Mounting the phantom with water leak left as it is can cause a failure of the CT
system.
Tightening the cap or air vent screw too much can cause damage of the Daily QC
Phantom. Tighten them to the extent that there is no looseness of the screw and
no water leak.
Setting and saving the scan protocol and scan parameters should be carried out only
when the Daily QC Phantom is scanned for the first time.
6 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
Q1E-BW1311 7
QS4 45462
3 Using the Daily QC Phantom
Reservoir
Air bubble
Water
When the reservoir is filled with air and air returns to the inside of the phantom when
the Daily QC Phantom is stood vertically, distilled water is short. Replenish the
phantom with distilled water.
8 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
Before using the Daily QC Phantom, be sure to check that there is no water leak. If
using it with water leak left as it is, a failure of the CT system can result.
Mount the Daily QC Phantom with the reservoir upward. If the phantom is set with
the reservoir downward, air enters the phantom and a correct image may not be
acquired.
4 Turning on the light localizer and moving the tabletop until the light
localizer strikes the Daily QC Phantom
Q1E-BW1311 9
QS4 45462
3 Using the Daily QC Phantom
Light
localizer
Positioning
mark
Light localizer
Turn the Daily QC Phantom to align the positioning mark and light localizer on
one side.
Light
localizer
Positioning
mark
10 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
Check the positioning mark on the other side, and turn the Daily QC Phantom by a
half of deviation.
Light localizer
Turn by a half of
deviation
Positioning mark
When the positioning mark and light localizer are apart largely, adjust the height of
the tabletop.
When the positioning marks at the left and right sides do not align with the light
localizer, the angle of rotating direction is not correct. Turn the Daily QC Phantom
again to adjust the angle of rotating direction.
When the position cannot be adjusted even if fully turning the height adjusting dial,
adjust the height of the tabletop and use the height adjusting dial to adjust it again.
The positioning mark should align with the light localizer within 2mm. If a
deviation is too big, a correct slice image may not be acquired.
Q1E-BW1311 11
QS4 45462
3 Using the Daily QC Phantom
When the scan protocol and scan parameters have not been set and saved, refer to
"3.2.1 Setting a scan protocol and scan parameters".
12 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
Q1E-BW1311 13
QS4 45462
3 Using the Daily QC Phantom
Fit ROI's to the center of each peg as far as possible. It fitting it to the peripheral of
the peg, accurate CT number measurements are not available due to the edge
enhanced effect.
14 Q1E-BW1311
QS4 45462
3 Using the Daily QC Phantom
When the CT number is not written on the CT number label, the CT number to be a
reference needs to be measured and registered. For details, refer to "Appendix A.1
Entering the CT number".
Q1E-BW1311 15
QS4 45462
3 Using the Daily QC Phantom
16 Q1E-BW1311
QS4 45462
4 Specification
4. Specification
Name Daily QC Phantom
Outside
210 × 232 × 90 (width × height × depth [mm])
dimensions
Weight (before feed water) 1.0 kg
(after feed water) 2.1 kg
CT number Water 0 (depending on definition of the CT number)
Teflon Written on the phantom (+930 ±60 [ref.])
Acrylic Written on the phantom (+125 ±10 [ref.])
Polyethylene Written on the phantom (-55 ±15 [ref.])
Air -1,000 (depending on definition of the CT number)
Q1E-BW1311 17
QS4 45462
4 Specification
18 Q1E-BW1311
QS4 45462
Appendix
Appendix
A.1 Entering the CT number
When the Daily QC Phantom is used for the first time, the CT number of each peg needs to be
entered in the CT number label.
Measure the CT number to be a reference in the following procedure after completing
installation and various adjustments, and enter the CT number in the CT number label stuck
near each peg of the Daily QC Phantom.
When entering the CT number in the CT number label, use a oil felt pen so that the
values entered do not vanish.
Q1E-BW1311 19
QS4 45462
Appendix
1 Mounting
Turn the knob counterclockwise and loosen the fixing plate L under the condition
shown below.
knob
Lift up the phantom mounting bracket and insert the phantom mounting bracket in
the direction until the end bracket touches the tabletop end as illustrated
below.
Tabletop
End brancket
20 Q1E-BW1311
QS4 45462
Appendix
Adjust the fixing plate R to the right edge of tabletop and clip the tabletop by
sliding the fixing plate L in the direction.
Turn the knob to fix the phantom mounting bracket.
If the bracket is not fixed properly, the bracket may possibly fall down during the
calibration resulting in damage of the scanner cover.
Knob
Fixing plate L
Fixing plate
Tabletop
Fixing plate R
Q1E-BW1311 21
QS4 45462
Appendix
DU nameplate
DU adjusting dial
LR adjusting dial U
L
R
D
LR nameplate
22 Q1E-BW1311
QS4 45462
Appendix
Knob
Keep the phantom leveled by tilting the phantom by a deflection component of the
table top as illustrated below.
Horizontal direction
Tabletop
Angle
adjustment
Q1E-BW1311 23
QS4 45462
Appendix
Knob
Fixing plate L
Tabletop
Fixing plate L
Before storing the phantom mounting bracket, ground the insertion part of the
phantom as illustrated below.
24 Q1E-BW1311
QS4 45462
INDEX
INDEX
Mounting and positioning ............................................7
A
Appendix.................................................................... 19
P
Phantom Mounting Bracket........................................20
C Preparatory Work for the Daily QC Phantom...............5
Composition of the Daily QC Phantom ....................... 3
S
E Scan............................................................................12
Entering the CT number............................................. 19 Scanning the Daily QC Phantom..................................6
Setting a scan protocol and scan parameters ................6
G Specification...............................................................17
General......................................................................... 1
U
M Using the Daily QC Phantom.......................................5
Measuring the CT number ......................................... 13
Q1E-BW1311 25
QS4 45462