2-1 Correspondence between the main detector channels and the DAS channels
Main detector Backplane DAS channel Converter-16 Main detector Backplane DAS channel Converter-16
channel PWB PWB slot channel PWB PWB slot REF
connector connector
1 to 32ch CNN1 1 to 24ch 1 633 to 656ch CNN25 577 to 600ch 25 (REF ch)
33 to 80ch CNN2 25 to 48ch 2 657 to 680ch CNN26 601 to 624ch 26
81 to 128ch CNN3 49 to 72ch 3 681 to 704ch CNN27 625 to 648ch 27
129 to 152ch CNN4 73 to 96ch 4 705 to 728ch CNN28 649 to 672ch 28
153 to 176ch CNN5 97 to 120ch 5 729 to 752ch CNN29 673 to 696ch 29
177 to 200ch CNN6 121 to 144ch 6 753 to 800ch CNN30 697 to 720ch 30
201 to 224ch CNN7 145 to 168ch 7 801 to 848ch CNN31 721 to 744ch 31
225 to 248ch CNN8 169 to 192ch 8 849 to 896ch CNN32 745 to 768ch 32 (1ch side)
249 to 272ch CNN9 193 to 216ch 9
273 to 296ch CNN10 217 to 240ch 10
297 to 320ch CNN11 241 to 264ch 11
321 to 341ch CNN12 265 to 288ch 12
345 to 368ch CNN13 289 to 312ch 13
Main detector
369 to 392ch CNN14 313 to 336ch 14
393 to 416ch CNN15 337 to 360ch 15
417 to 440ch CNN16 361 to 384ch 16
441 to 464ch CNN17 385 to 408ch 17 Converter-16 Backplane
465 to 488ch CNN18 409 to 432ch 18 (The channel bundling
DAS
489 to 512ch CNN19 433 to 456ch 19 of the main detector
array output is
513 to 536ch CNN20 457 to 480ch 20
performed according to
537 to 560ch CNN21 481 to 504ch 21 the connection pattern
Front view of the gantry
561 to 584ch CNN22 505 to 528ch 22 on the backplane.)
585 to 608ch CNN23 529 to 552ch 23
609 to 632ch CNN24 553 to 576ch 24
For this system, because channels at both ends (1-channel
Note: For the 16-slice detector, the correspondence table (shown above) for 16 rows is side: for 128 channels, 896-channel side: for 144
provided. The REF detector is connected to channel 1 (row 1) of the DAS. channels) of the main detector are bundled, the detector
channel number does not match the DAS channel number.
No. 2D201-065E*D
19
Table 4.2-2 Correspondence between the DCA channels and the DAS Converter-16 PWB slot numbers
Note:
(Example) If the channel 319 data is abnormal on the DCA, there may be a circuit failure of the 23rd channel on the 11th Converter-16 PWB.
No. 2D201-065E*D
20