You are on page 1of 72

TD850

E LECT RO SURGICAL UNIT

Service manual

TE698883-10
Read these Instructions before use
Preliminary
Information
Keep these ‘Instructions’ in a safe convenient place for future reference. Read in
conjunction with the relevant Publications detailed in the preliminary information section.
Technical data

Eschmann After Sales


Safety notes
Eschmann is staffed and equipped to provide advice and assistance during normal office
Introduction hours. To avoid delays when making enquiries, please quote the Model and Serial Number of
your Electrosurgical Unit which is shown on the Serial Number plate, the location of which is
Description
shown below. Please ensure you include all alpha and numeric digits of the Serial Number.

Maintenance

Illustrated parts list

Cleaning, disinfection
and care

For further information visit www.eschmann.co.uk

All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :

UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 875711.
Service manual

Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.

Patents and Trade marks

The ESCHMANN name and logo are trade marks of Eschmann Holdings Limited.
“Eschmann Equipment” is a trading name of Eschmann Holdings Limited.
“TD850” is a trade mark of Eschmann Holdings Limited.

Patents : US5540682, GB2276551, AU673883, ZA94/2173, US5480399,


EP617925, IEE69100, GB2261379, NZ236021, GB2237997,
EP429204,AU633156, US5246439 and other Patents Pending.

Copyright © 2013 Eschmann Holdings Limited

All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.

The CE marking affixed to the product certifies that it complies with the
European Medical Devices Directive 93/42/EEC and related legislation.

TE698883-10 July 2013


TD850 ELECTROSURGICAL UNIT

READ THESE INSTRUCTIONS BEFORE USE


Keep these Instructions in a safe convenient place for future reference. Read in conjunction
with the relevant Publications detailed in the preliminary information section.

CONTENTS
Page Page
1. PRELIMINARY INFORMATION .. 4 Component replacement .. .. .. .. 48
Access .. .. .. .. .. .. .. 48
2. TECHNICAL DATA .. .. .. .. 5
Removing/Refitting-Top Cover .. .. 48
General .. .. .. .. .. .. .. 5
Removing/Refitting-Front Panel .. 48
Dimensions .. .. .. .. .. .. 5
Removing/Refitting-Socket moulding 48
Electrical data .. .. .. .. .. .. 5
Removing/Refitting-Back panel .. 49
Internal fuses .. .. .. .. .. .. 5
Removing/Refitting-Back panel board 49
Monopolar outputs .. .. .. .. .. 5
Removing/Refitting-Sliding doors .. 49
Bipolar outputs .. .. .. .. .. .. 6
Removing/Refitting-Instruction cards 50
Audible indicators .. .. .. .. .. 6
Removing & Refitting PCBs .. .. .. 52
Visual indicators .. .. .. .. .. 6
General .. .. .. .. .. .. 52
Safety .. .. .. .. .. .. .. 6
Boards fitted in card frame .. .. .. 52
Symbols explaination .. .. .. .. 7
System checks .. .. .. .. .. .. 54
3. SAFETY NOTES .. .. .. .. .. 8 Eschmann Test sets .. .. .. .. 54
4. INTRODUCTION .. .. .. .. .. 9 Performance Checks .. .. .. .. 54
General .. .. .. .. .. .. .. 9 Output Power tests .. .. .. .. 54
Monopolar mode of operation .. .. .. 9 Alarm Function tests .. .. .. .. 55
Bipolar mode of operation .. .. .. 9 Footswitch Control Checks .. .. 56
Parameter change mode .. .. .. .. 10 7. ILLUSTRATED PARTS LIST .. .. 64
Accessories .. .. .. .. .. .. 10 Main assembly .. .. .. .. .. .. 64
5. DESCRIPTION .. .. .. .. .. 12 Rear panel assembly .. .. .. .. .. 68
Electrical .. .. .. .. .. .. .. 12 8. CLEANING DISINFECTION & CARE 70
Printed circuit boards .. .. .. .. 12
ILLUSTRATIONS
100V PSU Board .. .. .. .. 12
Fig.
Back Panel Board .. .. .. .. 12
1 TD850 electrosurgical unit .. .. .. 4
User interface Board .. .. .. .. 12
2 Front and rear views of TD850 unit .. 11
Front Panel Controller Board 2 .. .. 12
3 General arrangement of TD850 unit .. 14
Front Panel Controller Board 1 .. .. 13
4 Location of boards in card frame .. .. 15
Display Board .. .. .. .. .. 13
5 Power Output Vs Output Control
Card Frame Mother Board .. .. .. 13
Setting, Monopolar and Bipolar .. .. 17
12V Regulator Board .. .. .. .. 13
6 Power Output Vs Load Resistance
Monopolar Regulator Boards 1 & 2 .. 13
Monopolar and Bipolar .. .. .. .. 29
Bipolar Regulator Board .. .. .. 13
7 Power Output Vs Load Resistance
Pattern Generator Board .. .. .. 13
Monopolar Only .. .. .. .. .. 36
Monopolar/Bipolar output Boards .. 16
8 Power Output Vs Load Resistance
Monopolar Output Logic Boards 1&2 16
Monopolar Only, With and Without
Patient Alarm Board .. .. .. .. 16
Video Synchronisation .. .. .. .. 41
Tone Generator Board .. .. .. .. 16
9 Rear Panel (connector board) .. .. .. 49
Power output graphs .. .. .. .. .. 16
10 Sicket moulding sliding doors .. .. 50
6. MAINTENANCE .. .. .. .. .. 46 11 Removing/refitting card frame PCBs .. 51
General .. .. .. .. .. .. 46 12 Output waveform diagrams .. .. .. 57
Fault finding .. .. .. .. .. 46 13 ETS-3 test set .. .. .. .. .. .. 62
Systems check .. .. .. .. .. .. 46 14 TD800 Series test set .. .. .. .. 63
Alarms .. .. .. .. .. .. .. 46 15 TD850 Main assembly .. .. .. .. 64
Alarm conditions .. .. .. .. .. 46 16 Rear panel assembly .. .. .. .. 68
Alarm remedies .. .. .. .. .. 47 17 TD850 Block diagram .. .. .. .. 71

TE698883-10 Page 3 of 71
Part 1 TD850 ELECTROSURGICAL UNIT

1. PRELIMINARY INFORMATION
1.1 This Service Manual should be referred to for details of the TD850 Electrosurgical Unit,
REF 83-255-05 and 83-255-13, (serial numbers 85BB8E0000 and 85AB8E000 respectively or above).

1.2 Instruction and Service Manuals should be readily accessible for reference prior to and when
operating, cleaning and servicing the TD850 Electrosurgical Unit. All manuals are available from
Eschmann Equipment, see inside front cover for address details.

Related Technical Publications:-


Service Manual - TD850 - E-SM40 (issue ‘b’ or above)
Service Manual Addendum - TD850 - E-SM41 (issue ‘b’ or above)

Fig. 1 TD850 Electrosurgical Unit

Page 4 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 2

2. TECHNICAL DATA
GENERAL MONOPOLAR OUTPUTS
Equipment .. High power electrosurgical Effective carrier frequency 490kHz nominal
unit with monopolar and Power control - Variable amplitude with fixed
bipolar outputs duty cycle. Spray/soft spray coag-
Type .. .. Portable ulation have variable amplitude and
duty cycle control.
DIMENSIONS
Width .. .. .. .. .. 43.5 cm Monopolar (Normal) Peak
Height .. .. .. .. .. 20.0 cm Open
Peak Load Circuit
Length .. .. .. .. .. 47.4 cm
Power (ohms) Voltage
Weight .. .. .. .. .. 18.1 kg
Normal cut 160W 700 850V
ELECTRICAL DATA
Blend 130W 750 1100V
200-240V
Power Supply .. 200-240V a.c., 50-60Hz Specialist cut 210W 750 1100V
Current (max.) .. .. .. .. 3.1A
Fuse rating .. .. .. .. 250V, F4.0A Pinpoint coagulation 90W 550 1250V
Fuse type .. .. 2 x 20 mm HRC to IEC 127
(BSEN60127) located on back panel. Spray coagulation 80W 300 2200V
100-120V
Power Supply .. 100-120V a.c., 50-60Hz
Current (max.) .. .. .. .. 6.4A
Monopolar (parameter change) Peak
Fuse rating .. .. .. .. 250V, F8.0A Open
Fuse type .. .. .. 2 x 20 mm HRC Peak Load Circuit
located on back panel Power (ohms) Voltage

INTERNAL FUSES Endoscopy cut 100W 500 600V

PCB Fuse Eschmann Endoscopy blend 65W 450 900V


Type Part No.
Regulator Board Arthroscopy cut 100W 700 800V

Mono 1 (F1) FF/5A/250V 713056 MIS pinpoint 45W 450 850V


(F2) FF/8A/250V 713697
Mono 2 (F1) FF/5A/250V 713056 Soft coagulation 25W 250 2200V
(F2) FF/8A/250V 713697
Bipolar (F1) FF/5A/250V 713056
(F2) FF/8A/250V 713697
Low Volts Reg
(F1, F2) FF/2A/250V 713699
(F3) 500mA A/S 715595
Patient Alarm
Board (qty.1) F/125mA/125V 713849
Front Panel Control
(No.2.PCB,qty.1) FF/2A/250V 713681
Oscillator PCB
(1 fuse/oscillator)F/125mA/125V 713849

TE698883-10 Page 5 of 71
Part 2 TD850 ELECTROSURGICAL UNIT

2. TECHNICAL DATA
BIPOLAR OUTPUTS VISUAL INDICATORS
Effective carrier frequency 490kHz nominal Three separate displays, one each for User 1 and
Power control - Bipolar cut has variable User 2 Monopolar output users and one for Bipolar
amplitude control. Bipolar coagula- output. Nominal available power output for each
tion has variable duty cycle control. function is displayed numerically in the
appropriate display area. Numerals are displayed
in ‘green’ for normal usage and ‘amber’ when
Bipolar (normal) Peak
Open ‘parameter change’ selected.
Peak Load Circuit IMPORTANT NOTE
Power (ohms) Voltage
The displayed value gives an indication of the
Cut (high) 40W 400 400V power in Watts (± 20%), but only at the given
load impedance and at ambient temperature.
Micro coagulation 8W 80 160V The graphs of output power against impedance
(see Part 3) must be referred to for a full
Macro coagulation 50W 80 300V understanding of the particular mode
characteristics.
Power output ‘on’ indicators (green (normal) and
Bipolar (parameter change) Peak amber (parameter change)) are situated above each
Open function touchbutton. When any function is
Peak Load Circuit selected, its indicator will start to flash, and
Power (ohms) Voltage continue flashing while the output power is being
set. The indicator will become continuous (and
Cut (low) 20W 350 330V
further power adjustment disabled) approximately
10 seconds after adjustment is completed, or if
the output (at the level now set) is activated by
AUDIBLE INDICATORS either a footswitch or fingerswitch.
Cut or Blend running tone*
– High .. .. .. 1000Hz (approx.) SAFETY
Coagulation running tone* Designed to IEC601-1 1988 Part 1, Safety of
– Medium .. .. .. 800Hz (approx.) medical electrical equipment, IEC601-2-2: 1991
Specification for high frequency surgical
Bipolar running tone*
equipment.
– Cut .. .. .. .. 900Hz (approx.)
General classification, Class 1, Type CF
Bipolar running tone*
(Defibrillator proof).
– Coagulation .. .. 700Hz (approx.)
If the CE mark is affixed to the product, it indicates
Alarm – High/low, two-note at maximum volume
compliance with Council Directive 93/42/EEC of
Touch button ‘bleep’* 14 June 1993 concerning medical devices.
– indicates when a function button is pressed
Leakage (risk) current - always less than 10 micro
*Adjustable volume amps to earth (ground) from all patient circuits at
nominal mains voltage as required by IEC601 for
the unit in normal condition.

Page 6 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 2

2. TECHNICAL DATA
DRIP PROOF DUTY CYCLE

The symbol (drip proof) together with IPX 1 The duty cycle rating of 10 seconds ‘on’, 30 seconds
indicates that the equipment is protected against ‘off’, indicates that the equipment can be used at full
dripping water. output power in any mode for 10 seconds with a 30
second rest period. The unit can remain connected
to the mains electrical supply in the ‘standby’ or
ELECTRODE ISOLATION ‘outputs selected but not enabled’ conditions
The plate electrode circuit of this equipment is continuously. This duty cycle well exceeds that
isolated from earth and is protected by voltage encountered in normal use. The ‘on’ period at lower
monitoring, earth contact and patient attached power levels can be extended.
monitoring circuits. The bipolar output is also
fully isolated. NON-IONIZING RADIATION

The symbol on the front panel denotes that This symbol warns the user of the possibility
of Non-ionizing Radiation being caused by this
the neutral electrode (plate) is isolated from earth
equipment.
at high frequency.
VOLUME CONTROL
CLASS
Class 1 denotes that the equipment must be The symbol above item 40 (Fig. 2) indicates
earthed via the protective conductor in the 3-core the running tone volume control which has a
mains cable, connected to a plug with protective minimum sound level of 40 dBa.
earthing contact.
The symbol above item 40 (Fig. 2)
SAFETY CATEGORY indicates increasing running tone volume.

The symbol denotes that the equipment is SUPPLY


in the category type CF, i.e. that it is manufactured
to a safety standard commensurate with The symbol on the rear panel specification
international regulations for medical electrical plate indicates that the equipment is for use on
equipment. It also provides a degree of protection alternating current only.
higher than that for type BF equipment against
electric shock, particularly regarding leakage
currents. The symbol also denotes that the
equipment will not be damaged by defibrillator
discharge and that the plate electrode need not be
removed from the patient if a defibrillator is used.

TE698883-10 Page 7 of 71
Part 3 TD850 ELECTROSURGICAL UNIT

3. SAFETY NOTES
Attention to the following points will prolong the life and efficiency of your
electrosurgical unit and will help to avoid the risk of accidents, or damage:

DO: DO NOT:
♦ Read the instructions carefully before using ♦ Do not use uninsulated forceps.
the electrosurgical unit. Keep this instruction ♦ Do not use other than Eschmann accessories.
manual close-to-hand. ♦ Do not place monitoring electrodes close to
♦ Switch off and disconnect from the mains the operating site. When high frequency
electrical supply prior to cleaning the surgical equipment and physiological
equipment. monitoring equipment are used
♦ Use Eschmann accessories, in particular simultaneously on the same patient, any
cables, which should preferably be no more monitoring electrodes should be placed as far
than 3 metres in length. as possible from the surgical electrodes.
♦ Use bipolar techniques in preference to Needle monitoring electrodes are not
monopolar, whenever possible. For surgical recommended. In any case, monitoring
procedures on parts of the body having a systems incorporating high frequency current
relatively small cross-sectional area, the use limiting devices are recommended.
of bipolar techniques may be desirable to ♦ Do not allow active cables to drape across the
avoid unwanted coagulation. patient's body or to contact the cables of other
♦ Seek advice of Cardiology Department before equipment.
using the equipment on patients with ♦ Do not use flammable anaesthetics or
implanted pacemakers. Monitor such patients flammable solvents.
carefully. ♦ Do not allow the patient's body to touch
♦ Use a quiver to hold active accessories when conductive objects. The patient should not
not in use. come into contact with metal parts which are
♦ Use only insulated forceps. earthed or which have an appreciable
♦ Use 4 mm or 8 mm plug type active cables in capacitance to earth, e.g. operation table,
preference to hook type cables. supports, etc. The use of antistatic sheeting is
recommended for this purpose.
♦ Set power output controls to minimum setting
before use. ♦ Do not use hook type active cables, if possible.
♦ Use minimum power setting to achieve the ♦ Do not use hook type adapters for active
desired effect. cables.
♦ Grasp the connector(s) not the cable when ♦ Do not reuse disposable plate electrodes.
disconnecting cables from the equipment. ♦ Do not rely solely on surgical gloves to provide
♦ Check all cables and accessories routinely insulation.
before use. ♦ Do not insert a 4mm active plug into either of
♦ Use monitoring systems incorporating high the fingerswitch sockets (items 28 and 32,
frequency current limiting devices. Fig. 2). 4mm active plugs MUST ONLY be
inserted into the red active sockets (items 29
♦ Contact the hospital electronics engineer or and 33, Fig. 2).
Eschmann After Sales Service Department if
the equipment fails to function after checking.
♦ Ensure the equipment is serviced regularly.
Contact Eschmann After Sales Service
Department for details.

Page 8 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 4

4. INTRODUCTION
GENERAL 4.8 The Eschmann TD850 has been designed
4.1 The TD850 electrosurgical unit provides for ease of use and includes a quick set-up mode
two outlets for monopolar cutting and coagulation, that allows the user to rapidly access pre-selected
together with a separate outlet for bipolar cutting power settings in all major modes. The feature is
and coagulation. The unit incorporates particularly useful in trauma surgery, where rapid
touchbutton controls, digital power displays and equipment set-up is vital. A sterilisable remote
alarm code indicators. control enables either monopolar user to increase
or reduce power levels at the touch of a button.
4.2 The TD850 electrosurgery generator
provides a range of monopolar and bipolar cutting 4.9 Self diagnostics continuously monitor
and coagulating outputs to cover the needs of all internal circuits and if a fault is detected an alarm
surgical specialists including the latest Minimal will sound, the display will change colour and an
Access techniques. It incorporates important new ‘error’ code will appear. Diagnostic cards located
safety systems and user convenience is enhanced in the base of the unit enable the operator to
by rapid setup, remote control and self diagnosis identify the source of the problem and take
features. corrective action.

4.3 The TD850 has been designed to meet the MONOPOLAR MODE OF OPERATION
needs where two surgeons operate at the same 4.10 The monopolar mode of operation offers a
time, a requirement in cardiovascular, trauma and choice of fingerswitch or footswitch controls for
some general surgery procedures. It has two actuation of Cut or Coagulation outputs.
independent monopolar generators enabling it to Fingerswitch control is carried out using the
supply two surgeons simultaneously. Eschmann two-button fingerswitch; footswitch
4.4 The TD850’s two monopolar generators can control is carried out using an active electrode
be used simultaneously in all five major modes: handle in conjunction with one of a range of
pure cut, specialist cut, blend, pinpoint and spray footswitches. A plate electrode is required to be
coagulation. In addition to these essential macro attached to the patient in the monopolar mode.
modes, the TD850 also offers a range of specialist 4.11 In the monopolar mode, two users can use
Parameter Change (PC) modes. the unit simultaneously.
4.5 The Specialist Cut mode is designed for wet 4.12 To help reduce the possibility of accidental
field surgery as typically encountered in Urology, burns (if the patient’s body should contact a
while the Blend output is preferred in MAS and conductive material, see section 3), the unit is
Large Loop Excision of the Transformation Zone fitted with a leakage limiting circuit in the
(LLETZ). monopolar section. This circuit detects the
4.6 In addition to macro (50 Watts max.) and conditions that may lead to an increased possibility
micro (8 Watts max.) coagulation outputs, the of accidental burns from leakage and limits the
bipolar generator offers bipolar cutting, useful in leakage current accordingly. The HF leakage
a wide variety of surgical applications. limiter does not generate an alarm.

4.7 Innovative safety features include a split BIPOLAR MODE OF OPERATION


plate monitoring system called PAM (Plate 4.13 Bipolar coagulation is an efficient method
Attachment Monitor), as well as the standard of effecting haemostasis and for closure of vessels
Eschmann Patient Voltage Monitor, Patient Earth such as Fallopian tubes. Bipolar cutting or
Monitor, and Plate Continuity Monitor. A high coagulation are extremely safe to use as the
frequency leakage control circuit is included in current path is between the electrodes of the
the monopolar mode. bipolar instruments which are in contact with
An Excess Power Monitor (EPM) is also included. tissue. A plate electrode is not required, thus

TE698883-10 Page 9 of 71
Part 4 TD850 ELECTROSURGICAL UNIT

4. INTRODUCTION
reducing the risk of patient burns. In the bipolar 1. Bipolar output display zone
coagulation mode, a choice of two power modes 2. Activity selected indicator (green)
3. Parameter change selected indicator (amber)
is available; micro (8 Watts output) or macro 4. Bipolar Cut touchbutton
(50 Watts output). Selection of each power mode 5. Bipolar Macro Coag. touchbutton (50W output)
is by panel touchbuttons in the bipolar zone. 6. Bipolar Micro Coag. touchbutton (8W output)
7. Monopolar User 1 display zone
PARAMETER CHANGE MODE 8. Pure (normal) Cut touchbutton (User 1)
9. Blend Cut touchbutton (User 1)
4.14 The parameter change mode is available for 10. Specialist Cut touchbutton (User 1)
both the monopolar and bipolar modes of 11. Pinpoint Coagulation touchbutton (User 1)
operation. In this mode, power levels, as an 12. Spray Coagulation touchbutton (User 1)
alternative to the normal power levels, are 13. Monopolar User 2 display zone
14. Pure (normal) Cut touchbutton (User 2)
available for carrying out specialist procedures 15. Blend Cut touchbutton (User 2)
such as arthroscopy or endoscopy. 16. Specialist Cut touchbutton (User 2)
17. Pinpoint Coagulation touchbutton (User 2)
ACCESSORIES 18. Spray Coagulation touchbutton (User 2)
4.15 The TD850 Electrosurgical Unit is designed 19. User 2 select touchbutton
20. Remote control indicator
for use with the Eschmann range of active, plate 21. Standby touchbutton and indicator
and bipolar cables. For information on the full 22. Pre-set power level touchbutton
range of TD850 Electrosurgical Unit accessories, 23. Power level setting control
please contact Eschmann Equipment Sales 24. ‘S’ (Scope) connector socket
25. Plate electrode connector socket
Department. 26. 8 mm active electrode socket (User 2)
27. Sliding door (User 2)
28. Fingerswitch socket, (together with 29, User 2)
29. 4 mm active electrode socket (User 2)
30. 8 mm active electrode socket (User 1)
31. Sliding door (User 1)
32. Fingerswitch sockets (together with 33, User 1)
33. 4 mm active electrode socket (User 1)
34. Bipolar output socket.
35. Bipolar fingerswitching socket
36. Mains power on/off switch
37. Mains power input plug
38. Mains fuse (see Tech. Data for full details)
39. Mains fuse (see Tech. Data for full details)
40. Running tone volume control
41. Potential equalising terminal
42. User 2 monopolar (Cut or Coag.) footswitch socket
43. User 2 monopolar (Cut or Coag.) footswitch socket
44. Remote control accessory socket
45. User 1 monopolar (Cut or Coag.) footswitch socket
46. User 1 monopolar (Cut or Coag.) footswitch socket
47. Bipolar Coagulation footswitch socket
48. Bipolar Cut footswitch socket
49. Bipolar pneumatic footswitch socket (Coag. only)
50. Video synchronisation connectors control panel
51. Video in socket (only applicable to monopolar
User 1)
52. Auxiliary socket
53. Horizontal adjustment, (spray coagulation only)
54. Horizontal adjustment (any mode except spray
coag. (only applicable to monopolar User 1)
55. Vertical adjustment (all modes)
56. Serial plate
Key to Fig. 2

Page 10 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 4

4. INTRODUCTION

Fig. 2 Front and rear view of TD850 unit

TE698883-10 Page 11 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION
ELECTRICAL PRINTED CIRCUIT BOARDS
5.1 The Electrosurgical Unit is of modular 5.4 The following descriptions explain the main
design incorporating the following printed circuit functions of each board fitted in the unit. The
boards. locations of the boards are shown on Figs 3 and 4.
The following boards are fitted to the chassis, the A system block diagram is shown in Fig 16.
front panel moulding, or to the outside of the 100V PSU Board (Fig. 3, item 11)
cardframe: 5.5 The 100V PSU board is an off-line switcher,
♦ Front Panel Display Board converting the mains electricity supply at either a
♦ Front Panel Controller 1 Board nominal 230V or 110V to 100V d.c. It is connected
♦ Front Panel Controller 2 Board to the Card Frame Mother Board by a loom, and
♦ Front Switch Panel Board all power for the unit is derived from it.
♦ 100V Power Supply Board Back Panel Board (Fig. 3, item 12)
♦ User Interface Board 5.6 All the footswitch input sockets, the remote
♦ Back Panel (connector) Board control input sockets and the volume control, are
mounted on this board. A ribbon cable links it to
The following boards are fitted in the card frame: the User Interface Board.
♦ 12V Regulator Board
User Interface Board (Fig. 3, item 13)
♦ Monopolar Regulator 1 (HV) Board
♦ Monopolar Regulator 2 (HV) Board 5.7 The User Interface Board provides
individual electrical isolation of all fingerswitches,
♦ Bipolar Regulator (HV) Board
footswitches, and the remote control accessory,
♦ Tone Generator Board and passes the status of each of these to the Front
♦ Patient Alarm Board Panel Controller Board 2 via a ribbon cable. A
♦ Pattern Generator Board dedicated isolation/detection module is provided
♦ Monopolar Output (combined assemblies) for each of these switch functions (a total of 19).
Boards An on-board regulator converts the 12V supply
to the 5V required by these modules.
♦ Monopolar 1 Logic Board
♦ Bipolar Output Board Front Panel Controller Board 2 (Fig. 3, item 4)
♦ Monopolar 2 Logic Board 5.8 The 5V supply required for the Front Panel
Controller Board is derived from the 12V supply
5.2 Printed circuit boards which are not fitted
on the Mother Board by an on-board switch-mode
in the card frame are connected to the system by
DC-DC convertor. The microcontroller located
cable looms, ribbon connectors, or single wires
on this board sends instructions and information
with appropriate terminations (see Fig. 5). The
to other boards in response to command signals
printed circuit boards fitted in the card frame plug
from:
into connectors fitted in a mother board in the
(a) Fingerswitches, footswitches or the remote
centre of the card frame. The mother board is
control accessory.
connected to the 100V power supply board, and
to other system boards and components by cable (b) Front panel buttons (attached to the front
looms. panel moulding).
(c) Alarm signals.
5.3 The boards are locked into the card frame
by side plates and a lever fitted to the board, which (d) Front panel control knob and the two sliding
engages a stirrup fitted to the card frame. When safety covers.
released, the lever is used to partly pull the board 5.9 Group (a) instructions are received from the
from the card frame (see Fig. 11). User Interface Board via a ribbon cable.

Page 12 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION
5.10 Group (b) instructions are received from the (b) Determines the colour (red/green/amber) of
Front Panel Switch Board, connected via another each of the numerical display digits.
ribbon cable to a small daughter board clipped on
to Controller Board 2. Display Board (Fig. 3, item 8)

5.11 Group (c) comprises alarm signals sent by 5.15 The Display Board carries the running
other boards mounted in the card frame via the lamps, the Mode indicators, and the power level
card frame Mother Board, and linked to Controller indicators. The seven segment displays and the
Board 2 by another ribbon cable, which also Mode indicator LEDs are driven by circuitry on
carries RS485 data communication and brings the Display Board from information received from
power (12V) for the Controller Boards. Controller Board 1. However, the running lamps
and the red/green/amber colour information for
5.12 A ribbon cable linking Controller Board 2 the seven segment displays comes from Controller
with Controller Board 1 brings information about Board 2.
control knob setting and the position of the two
sliding safety covers; the sensors for these items Card Frame Mother Board (Fig. 4, item 13)
are mounted on Controller Board 1. 5.16 All the card frame boards are plugged into
the Card Frame Mother Board which carries both
5.13 Outgoing instructions and information from
power and control signals between them.
Controller Board 1 include:
12V Regulator Board (Fig. 4, item 8)
(a) Numerical data for the front panel setpoint
indicators. 5.17 The 12V Regulator Board is a switch-mode
(b) An RS485 data link for communication with convertor which is used to generate +12V from
other boards as follows: the main 100V supply module. In addition, a low
power -12V supply is also generated.
Bipolar
Monopolar Regulator Boards 1 and 2 (Fig. 4,
♦ Tone Generator Board
items 1 and 7)
♦ Pattern Generator Board
♦ Bipolar Regulator Board 5.18 The Monopolar Regulator Boards 1 and 2
generate a voltage in the range 0 - 85V from the
♦ Bipolar Output Board
100V supply. This sets the power output level to
Monopolar User 1 be generated by the Output Board pair (Fig. 4,
♦ Tone Generator Board items 4 and 11). The actual voltage level is
♦ Pattern Generator Board determined by the RS485 microlink.
♦ Monopolar Regulator Board 1 Bipolar Regulator Board (Fig. 4, item 2)
♦ Monopolar Output Logic Board 1
5.19 The Bipolar Regulator Board is similar in
Monopolar User 2 function to the Monopolar Regulator Boards, but
♦ Tone Generator Board sets the level of Bipolar power output.
♦ Pattern Generator Board Pattern Generator Board (Fig. 4, item 3)
♦ Monopolar Regulator Board 2
5.20 The Pattern Generator Board includes three
♦ Monopolar Output Logic Board 2 dedicated ASICs which produce signals to drive
Front Panel Controller Board 1 (Fig. 3, item 5) the various output boards, which in turn produce
the required output wave form. The timing of
5.14 The Front Panel Controller Board 1
these signals is determined by the microcontroller
performs two functions relevant to the front panel
(Front Panel Controller board 2).
display:
(a) Controls which running lamps are
illuminated.

TE698883-10 Page 13 of 71
Part 5
TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

1 Chassis back panel


2 Chassis base
3 Handle moulding
4 Front panel controller board 2
5 Front panel controller board 1
6 Socket moulding
7 Front panel moulding
(complete with switch board)
8 Display board
9 Chassis vertical bulkhead
10 Cardframe Front and back panel
11 100V PSU board details are shown on Fig. 2
12 Back panel (connector) board
Circuit board locations in the
13 User interface board
card frame are shown on Fig. 4
14 Side plates
15 Cable protection card

Fig. 3 General arrangement of TD850 Electrosurgical Unit

Page 14 of 71 TE698883-10
Part 5
TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

1 Monopolar Regulator Board 1 8 12V Regulator Board


2 Bipolar Regulator Board 9 Bipolar Output Board
3 Pattern Generator Board 10 Monopolar Logic Board 2
4 Monopolar Output Module (User 1) 11 Monopolar Output Module (User 2)
5 Monopolar Output Logic Board 1 12 Tone Generator Board
6 Patient Alarm Board 13 MotherBoard
7 Monopolar Regulator Board 2

Fig. 4 Location of boards in the card frame

TE698883-10 Page 15 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION
Monopolar Output Boards (Fig. 4, items 4 & Patient Alarm Board (Fig. 4, item 6)
11) & Bipolar Output Board (Fig. 4, item 9) 5.23 All plate related alarm conditions are
5.21 These boards are the high frequency power detected by the Patient Alarm Board. The Patient
generators for monopolar 1 and 2 and bipolar Earth Monitor (PEM) alarm looks to see if there
outputs respectively. They each receive dc power is a path with a resistance of less than 220ohms
from their respective Regulator Boards, and a between the plate and earth. It does this each time
signal (to drive the push-pull output transistors) (and just before) a monopolar output is activated,
from the Pattern Generator Board. Each and every 20 seconds in the absence of use.
monopolar output also has a resonant tank circuit Further, similar, circuits look for Plate Continuity
which is used to generate the high voltage spray Monitor (PCM) and PAM alarms, and yet another
coagulation output waveform. One of the signals monitors the absolute value of voltage on the plate/
from the Pattern Generator controls this function. patient and, if excessive, initiates a Patient Voltage
The output at high voltage is derived from the Monitor (PVM) alarm.
output transformer, and the output waveform is
Tone Generator Board (Fig. 4, item 12)
monitored to measure both the load current and
voltage (or just current in the case of bipolar). 5.24 The Tone Generator Board generates
These measurements are used to control the output running, alarm and button bleep tones. The alarm
power according to a “load curve” as programmed tones are produced at full volume, whereas the
into the microcontroller. This output power is button bleep and running tones are adjustable by
monitored to ensure that it remains within safe the rear panel mounted volume control; the three
operating limits of the front panel settings. connections to the volume control run through the
Mother Board to Front Panel Controller Board 2,
Monopolar Output Logic Boards 1 and 2 down to and across the User Interface Board, and
(Fig. 4, items 5 and 10) then to the Back Panel Board.
5.22 The Monopolar Output Logic Boards 1 and
Power Output Graphs (Figs 5 to 8)
2 monitor their respective output loads and control
output power accordingly, so as to follow the 5.25 The power output of the unit has been
“power curve” for the mode in use that is optimised to provide effective cutting and
programmed into the microcontroller. coagulation over the wide range of tissue
Measurements of both load current and load impedance that will be encountered. The low
voltage are provided by the respective Output impedance region of the power output graphs
Board via a loom, and a voltage level derived from (below normal operating range) is controlled by a
these adjusts the voltage supplied to the Output power limiting circuit which protects the
Board by the Regulator Board. This function is equipment and the accessories connected to the
referred to as “power control”. If a low impedance unit. All measurements against which the power
load is presented to the output, a more direct output graphs were plotted were made at
control of output current must then be employed. approximately 25°C.
Thus, the load current is separately monitored and,
if excessive, the logic board will interrupt the
applied power (by progressively disabling the
Pattern Generator ASIC) so that an average safe
value is achieved. RF leakage is also monitored
using a sense toroid situated near the output
sockets and controlled within safe operating
limits.

Page 16 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

200

180

160

140
POWER OUTPUT (WATTS)

120

100

80

60

40

20

0
0 20 40 60 80 100 120 140 160
OUTPUT CONTROL SETTING

(a) Pure Cut at 500 ohms, with Video Synchronisation

180

160

140
POWER OUTPUT (WATTS)

120

100

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160
OUTPUT CONTROL SETTING

(b) Pure Cut at 700 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 1 of 12)

TE698883-10 Page 17 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

120

100
POWER OUTPUT (WATTS)

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100
OUTPUT CONTROL SETTING

(c) Parameter Change Pure Cut at 350 ohms, with Video Synchronisation

120

100
POWER OUTPUT (WATTS)

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100
OUTPUT CONTROL SETTING

(d) Parameter Change Pure Cut at 500 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 2 of 12)

Page 18 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

160

140

120
POWER OUTPUT (WATTS)

100

80

60

40

20

0
0 20 40 60 80 100 120 140
OUTPUT CONTROL SETTING

(e) Blend at 650 ohms, with Video Synchronisation

140

120

100
POWER OUTPUT (WATTS)

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100 110 120 130
OUTPUT CONTROL SETTING

(f) Blend at 700 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 3 of 12)

TE698883-10 Page 19 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

80

70

60
POWER OUTPUT (WATTS)

50

40

30

20

10

0
0 10 20 30 40 50 60 70
OUTPUT CONTROL SETTING

(g) Parameter Change Blend at 350 ohms, with Video Synchronisation

70

60

50
POWER OUTPUT (WATTS)

40

30

20

10

0
0 5 10 15 20 25 30 35 40 45 50 55 60 65
OUTPUT CONTROL SETTING

(h) Parameter Change Blend at 450 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 4 of 12)

Page 20 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

300

250
POWER OUTPUT (WATTS)

200

150

100

50

0
0 50 100 150 200 250
OUTPUT CONTROL SETTING

(i) Specialist Cut at 650 ohms, with Video Synchronisation

250

200
POWER OUTPUT (WATTS)

150

100

50

0
100

110

120

130

140

150

160

170

180

190

200

210
10

20

30

40

50

60

70

80

90
0

OUTPUT CONTROL SETTING

(j) Specialist Cut at 750 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 5 of 12)

TE698883-10 Page 21 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

120

100
POWER OUTPUT (WATTS)

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90 100
OUTPUT CONTROL SETTING

(k) Parameter Change Specialist Cut at 650 ohms, with Video Synchronisation

100

90

80

70
POWER OUTPUT (WATTS)

60

50

40

30

20

10

0
0 10 20 30 40 50 60 70 80 90 100
OUTPUT CONTROL SETTING

(l) Parameter Change Specialist Cut at 700 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 6 of 12)

Page 22 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

120

100
POWER OUTPUT (WATTS)

80

60

40

20

0
0 10 20 30 40 50 60 70 80 90
OUTPUT CONTROL SETTING

(m) Pinpoint Coagulation at 400 ohms, with Video Synchronisation

100

90

80

70
POWER OUTPUT (WATTS)

60

50

40

30

20

10

0
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90
OUTPUT CONTROL SETTING

(n) Pinpoint Coagulation at 550 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 7 of 12)

TE698883-10 Page 23 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

50

45

40

35
POWER OUTPUT (WATTS)

30

25

20

15

10

0
0 5 10 15 20 25 30 35 40 45
OUTPUT CONTROL SETTING

(o) Parameter Change Pinpoint Coagulation at 350 ohms, with Video Synchronisation

50

45

40

35
POWER OUTPUT (WATTS)

30

25

20

15

10

0
0 5 10 15 20 25 30 35 40 45
OUTPUT CONTROL SETTING

(p) Parameter Change Pinpoint Coagulation at 450 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 8 of 12)

Page 24 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

70

60

50
POWER OUTPUT (WATTS)

40

30

20

10

0
0 10 20 30 40 50 60 70 80
OUTPUT CONTROL SETTING

(q) Spray Coagulation at 300 ohms, with Video Synchronisation

80

70

60
POWER OUTPUT (WATTS)

50

40

30

20

10

0
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80
OUTPUT CONTROL SETTING

(r) Spray Coagulation at 300 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 9 of 12)

TE698883-10 Page 25 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

25

20
POWER OUTPUT (WATTS)

15

10

0
0 5 10 15 20 25
OUTPUT CONTROL SETTING

(s) Parameter Change Spray Coagulation at 250 ohms, with Video Synchronisation

25

20
POWER OUTPUT (WATTS)

15

10

0
115
10

11

12

13

14

16

17

18

19

20

21

22

23

24

25
0

OUTPUT CONTROL SETTING

(t) Parameter Change Spray Coagulation at 300 ohms, no Video Synchronisation

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 10 of 12)

Page 26 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

45

40

35
POWER OUTPUT (WATTS)

30

25

20

15

10

0
0 5 10 15 20 25 30 35 40
OUTPUT CONTROL SETTING

(u) Normal Bipolar Cut at 400 ohms

25

20
POWER OUTPUT (WATTS)

15

10

0
0 2 4 6 8 10 12 14 16 18 20
OUTPUT CONTROL SETTING

(v) Parameter Change Bipolar Cut at 400 ohms

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 11 of 12)

TE698883-10 Page 27 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

6
POWER OUTPUT (WATTS)

0
0 1 2 3 4 5 6 7 8
OUTPUT CONTROL SETTING

(w) Micro Bipolar Coagulation at 80 ohms

50

45

40

35
POWER OUTPUT (WATTS)

30

25

20

15

10

0
0 3 6 9 12 15 18 21 24 27 30 33 36 39 42 45 48 51
OUTPUT CONTROL SETTING

(x) Macro Bipolar Coagulation at 80 ohms

Fig. 5 Power Output Vs Output Control Setting, Monopolar and Bipolar (Sheet 12 of 12)

Page 28 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

180

160

140
POWER OUTPUT (WATTS)

120

100

100%
80

60

40 50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(a) Pure Cut Normal, no Video Synchronisation

120

100

80
POWER (WATTS)

60

100%
40

50%
20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(b) Parameter Change Pure Cut, no Video Synchronisation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 1 of 7)

TE698883-10 Page 29 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

140

120

100
POWER OUTPUT (WATTS)

80 100%

60

40
50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(c) Blend Normal, no Video Synchronisation

70

60

50
POWER (WATTS)

40

30
100%

20

50%

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(d) Parameter Change Blend, no Video Synchronisation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 2 of 7)

Page 30 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

250

200
POWER (WATTS)

150 100%

100

50%
50

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(e) Specialist Cut Normal, no Video Synchronisation

90

80

70
100%

60
POWER (WATTS)

50

40

30 50%

20

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(f) Parameter Change Specialist Cut, no Video Synchronisation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 3 of 7)

TE698883-10 Page 31 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

100

90

80

70
POWER (WATTS)

60

50

100%
40

30

20 50%

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(g) Pinpoint Coagulation Normal, no Video Synchronisation

50

45

40

35
POWER (WATTS)

30

25

20
100%

15

10
50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(h) Parameter Change Pinpoint Coagulation, no Video Synchronisation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 4 of 7)

Page 32 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

80

70

60

100%
50
POWER (WATTS)

40

50%
30

20

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(i) Spray Coagulation Normal, no Video Synchronisation

25

20

100%
POWER (WATTS)

15

10

50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(j) Parameter Change Spray Coagulation, no Video Synchronisation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 5 of 7)

TE698883-10 Page 33 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

45

40

35

30 100%
POWER (WATTS)

25

20
50%

15

10

0
0 100 200 300 400 500 600 700 800 900 1000
LOAD (OHMS)

(k) Bipolar Cut Normal

20

18
100%

16

14
POWER (WATTS)

12

10
50%

0
0 100 200 300 400 500 600 700 800 900 1000
LOAD (OHMS)

(l) Parameter Change Bipolar Cut

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 6 of 7)

Page 34 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

50

45

40

35
POWER (WATTS)

30

25

20

15 100%

10
50%

0
0 100 200 300 400 500 600 700 800 900 1000
LOAD (OHMS)

(m) Bipolar Macro Coagulation

6
POWER (WATTS)

100%
2

50%
1

0
0 100 200 300 400 500 600 700 800 900 1000
LOAD (OHMS)

(n) Bipolar Micro Coagulation

Fig. 6 Power Output Vs Load Resistance, Monopolar and Bipolar (Sheet 7 of 7)

TE698883-10 Page 35 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

200

180

160

140
POWER (WATTS)

120

100

80 100%

60

40 50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(a) Pure Cut Normal, with Video Synchronisation

120

100

80
POWER (WATTS)

60

40
100%

20
50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(b) Parameter Change Pure Cut, with Video Synchronisation

Fig. 7 Power Output Vs Load Resistance, Monopolar Only (Sheet 1 of 5)

Page 36 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

160

140

120

100
POWER (WATTS)

80
100%

60

40
50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(c) Blend Normal, with Video Synchronisation

70

60

50
POWER (WATTS)

40

30
100%

20

50%
10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(d) Parameter Change Blend, with Video Synchronisation

Fig. 7 Power Output Vs Load Resistance, Monopolar Only (Sheet 2 of 5)

TE698883-10 Page 37 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

300

250

200
POWER (WATTS)

150

100%

100

50%
50

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000

LOAD (OHMS)

(e) Specialist Cut Normal, with Video Synchronisation

120

100

80
POWER (WATTS)

60 100%

40

50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(f) Parameter Change Specialist Cut, with Video Synchronisation

Fig. 7 Power Output Vs Load Resistance, Monopolar Only (Sheet 3 of 5)

Page 38 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

100

90

80

70
POWER (WATTS)

60

50

40
100%

30

20
50%

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(g) Pinpoint Coagulation Normal, with Video Synchronisation

50

45

40

35
POWER (WATTS)

30

25

20
100%
15

10
50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(h) Parameter Change Pinpoint Coagulation, with Video Synchronisation

Fig. 7 Power Output Vs Load Resistance, Monopolar Only (Sheet 4 of 5)

TE698883-10 Page 39 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

60

50

100%

40
POWER (WATTS)

30 50%

20

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(i) Spray Coagulation Normal, with Video Synchronisation

25

20

100%

15
POWER (WATTS)

10

50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(j) Parameter Change Spray Coagulation, with Video Synchronisation

Fig. 7 Power Output Vs Load Resistance, Monopolar Only (Sheet 5 of 5)

Page 40 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

200

Video
180

160

140
POWER (WATTS)

120

100
Video
80
100%

60

40 50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(a) Pure Cut Normal

120

Video
100

80
POWER (WATTS)

60

100%
40
Video

20 50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(h) Parameter Change Pure Cut

Fig. 8 Power Output Vs Load Resistance, Monopolar Only, With and Without
Video synchronisation (Sheet 1 of 5)
TE698883-10 Page 41 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

160

140 Video

120

100
POWER (WATTS)

80 100%

Video
60

40
50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(c) Blend Normal

70

60

Video

50
POWER (WATTS)

40

30 100%

20
Video
50%

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(d) Parameter Change Blend

Fig. 8 Power Output Vs Load Resistance, Monopolar Only, With and Without
Video synchronisation (Sheet 2 of 5)
Page 42 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

300

Video
250

200
POWER (WATTS)

150

Video 100%

100

50%

50

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(e) Specialist Cut Normal

120

Video
100

80
POWER (WATTS)

100%
60
Video

40 50%

20

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(f) Parameter Change Specialist Cut

Fig. 8 Power Output Vs Load Resistance, Monopolar Only, With and Without
Video synchronisation (Sheet 3 of 5)
TE698883-10 Page 43 of 71
Part 5 TD850 ELECTROSURGICAL UNIT

5. DESCRIPTION

100

90 Video

80

70
POWER (WATTS)

60

50 Video
100%
40

30

20 50%

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(g) Pinpoint Coagulation Normal

50

45

40

Video
35
POWER (WATTS)

30

25

100%
20

15

Video
10 50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(h) Parameter Change Pinpoint Coagulation

Fig. 8 Power Output Vs Load Resistance, Monopolar Only, With and Without
Video synchronisation (Sheet 4 of 5)
Page 44 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 5

5. DESCRIPTION

80

70

60

Video
50
POWER (WATTS)

100%

40

30
50%

20

10

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(i) Spray Coagulation Normal

25

Video

20

100%
POWER (WATTS)

15

10

50%

0
0 200 400 600 800 1000 1200 1400 1600 1800 2000
LOAD (OHMS)

(j) Parameter Change Spray Coagulation

Fig. 8 Power Output Vs Load Resistance, Monopolar Only, With and Without
Video synchronisation (Sheet 5 of 5)

TE698883-10 Page 45 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
MAINTENANCE CODE CONDITION
General
01 Multiple activation
6.1 Ensure that electrosurgical safety checks and 02 Patient voltage monitor (PVM)
routine maintenance are carried out at regular 03 Patient earth monitor (PEM)
intervals and only by Eschmann trained personnel 04 Plate continuity monitor (PCM)
otherwise the warranty for the equipment could 05 Plate attachment monitor (PAM)
be infringed. 06 Over temperature.
(NOTE: applicable to early units only).
07 Power output without enable or in
WARNING
standby
Potentially dangerous voltages are exposed
08 Disallowed footswitch combination
when the unit covers are removed. Switch
(both users intending to use foot-
'off' and disconnect the unit from the 'mains' switches)
supply before doing any maintenance. (NOTE: applicable to early units only).
80 A foot or fingerswitch has been
CAUTION operated while the unit is in standby,
This unit contains static sensitive devices, use or may be faulty.
appropriate handling procedures when 87/88/89 Monopolar user 1, or User 2, or
handling printed circuit boards and Bipolar output is faulty.
electronic components. Notes:
(i) No plate alarms should operate with bipolar
mode only selected.
Fault Finding (ii) There are additional alarm codes which
6.2 If the unit malfunctions, check all cables and indicate an internal fault. If an alarm code
accessories by ‘substitution’ before doing further other than those listed in this manual occur,
checks. contact the Eschmann After Sales Service
Department.

SYSTEMS CHECK Alarm Conditions

6.3 A systems check should be carried out 6.6 The user alarms operate as follows:
before the unit is issued for use. This will provide (01) Multiple Activation Alarm
a complete check of the correct functioning of the Operates if more than one fingerswitch or
unit. footswitch is pressed at the same time in a
6.4 Routine checking of the unit should be given user section.
limited to external checks of the unit, which (02) Patient Voltage Monitor (PVM)
should also include all accessories and cables. Is designed to prevent dangerous
electrosurgical voltages appearing on the
patient’s body. Such voltages could occur
ALARMS because of insulation faults in the active
6.5 The equipment incorporates alarm circuits, circuit and could cause burns at points of
each of which is given a code number that is contact between the patient and conductive
displayed in flashing red numerals while all other objects. This alarm will only operate when
displays are blank. Activation of an alarm is monopolar output power is ‘on’ and the
indicated by a two-tone audio signal. The alarm fingerswitch or footswitch is operated, and
codes and their interpretation are as follows: will not be enabled in standby use. This
monitor will not detect a poor contact between
the return electrode and the patient.
Page 46 of 71 TE698883-8
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE
(03) Patient Earth Monitor (PEM) (87/88/89) Excess Power Monitor (EPM)
Current safety regulations relating to patients In accordance with council directive
on an operation table require that, for protection 93/42/EEC (Medical Devices Directive) the
against electric shock, there should be no unit monitors its output power in monopolar
direct earth connection to the patient’s body. or bipolar modes, compares this with the
This circuit detects if such a condition exists, displayed power and alarms, disabling output,
it also reduces the chances of secondary contact if the delivered power exceeds the displayed
electrosurgical burns. The circuit detects paths value by more than a certain margin.
with a resistance of less than 220ohms to earth
Alarm Remedies
every 20 seconds (approx) the plate electrode
is attached to the patient and the unit is switched 6.7 Some remedies for alarms are as follows:
‘on’. If such an earth path is present, the alarm 01 Alarm: (i) Footswitch may be jammed under
will operate until the path has been removed. table, trolley etc.
WARNING (ii) Another footswitch/fingerswitch
The patient earth monitor (PEM) will not may be faulty and constantly on.
detect a low impedance path between the 02 Alarm: (i) Check for insulation fault(s) in
patient’s body and earth when the active circuit.
electrosurgical output is activated. (ii) Check plate electrode for correct
(04) Plate Continuity Monitor (PCM) application to patient.
Ensures that the unit cannot be used with an Note: This alarm resets when foot or
electrically defective plate electrode or cable. It fingerswitch is released.
does not monitor the contact between the plate 03 Alarm: (i) Check for accidental direct con-
and the patient’s body. It is reset automatically tact between patient and earthed
when a satisfactory plate and cable are connected. object, e.g. drip stand or conduc-
(05) Plate Attachment Monitor (PAM) tive part of operation table.
Applies only when a split plate with (ii) If applicable, disconnect suspect
appropriate lead is used. Ensures that the ECG electrodes from monitor.
unit cannot be used in a monopolar mode if a 04 Alarm: (i) Check plate electrode is connected
split plate electrode is not attached to the to unit.
patient’s body correctly. (ii) Check plate electrode and cable for
(06) Over Temperature Alarm condition. If faulty, substitute new
Activated if unit has been run far in excess of plate and/or cable. This procedure
the recommended duty cycle. should normally be carried out before
(NOTE: applicable to early units only). the plate is attached to the patient.
(07) Power Output Without Enable or in 05 Alarm: (i) Check that plate electrode is fully
Standby in contact with patient.
WARNING (ii) Check that no obstructions have
Output electrodes may be energized been introduced between plate
if alarm 07 is activated. surface and patient.
(iii) Check plate electrode and
Activated by an internal fault.
cable for condition. If faulty, sub-
(08) Disallowed Footswitch Combination stitute new plate and/or cable.
(User 1 and User 2) This procedure should normally
At least one user must be using a fingerswitch be carried out before the plate is
in Dual User Mode.(NOTE: applicable to attached to the patient.
early units only).
TE698883-10 Page 47 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
06 Alarm: (i) Allow machine to cool down for Refitting the Top Cover
several minutes. 6.10 Refitting of the top cover is the reverse of
(ii) If a cooling interval does not re- the removal procedure. However, when fitting
set alarm, unit may have an inter- the top cover ensure the following points are
nal fault and the Eschmann After noted:
Sales Service Department should
be contacted. (1) Avoid chaffing or trapping cables.
07 Alarm: (i) Turn machine off at mains. Turn (2) Slide top cover forward to front panel
on again and set up as required. moulding before tightening screws.
(ii) If procedure above does not cure (3) Ensure top cover enters slots around handle
fault, contact Eschmann After moulding.
Sales Service Department. (4) Ensure top cover lugs (one at centre front and
08 Alarm: (i) Ensure that one user has a two at rear) are correctly seated.
fingerswitching accessory con- Removing the Front Panel Moulding
nected.
80 Alarm (i) Check for electrically defective 6.11 Remove front panel moulding (Fig 15, 6)
footswitch. as follows:
87/88/89 (i) Call an Eschmann Engineer (1) Remove top cover as described in section 6.9.
COMPONENT REPLACEMENT (2) Remove power setting knob (Fig 15, 17) by
pulling it carefully from its shaft.
Access
(3) Remove six screws (Fig 15, 74) and withdraw
6.8 Accessing the unit for maintenance will the front panel moulding. Disconnect the
involve removal of the top cover: ribbon cables connecting switch board
WARNING (Fig 15, 36) fitted in the front panel moulding
Potentially dangerous voltages are exposed board, to Controller Board 2 daughter board
when the unit covers are removed. Switch (Fig 15, 22) fitted to the chassis vertical
'off' and disconnect the unit from the 'mains' bulkhead.
supply before doing any maintenance. Refitting the Front Panel Moulding
CAUTION 6.12 Refitting of the front panel moulding is the
This unit contains static sensitive devices, use reverse of the removal procedure.
appropriate handling procedures when
handling printed circuit boards and Removing the Socket Moulding
electronic components. 6.13 Remove socket moulding (Fig 15, 7) as follows:
Removing the Top Cover (1) Remove top cover as described in section 6.9.
6.9 Remove top cover (Fig 15, 3) as follows: (2) Remove the card frame assembly (Fig 15, 19)
(1) Turn the unit upside-down on a clean and and RFI shield (Fig 15, 34) as described in
protected surface. section 6.35.
(2) Remove four screws (Fig 15, 81) securing the (3) Remove front panel moulding (Fig 15, 6) as
top cover to the chassis and loosen three screws described in section 6.11.
(Fig 15, 73) securing back panel to chassis. (4) Remove three screw covers (Fig 15, 1) and
(3) Turn the unit the right-way-up, ensuring that remove three screws (Fig 15, 75) securing
the top cover does not fall off. socket moulding (Fig 15, 7) to the chassis and
pull the socket moulding from the chassis.
(4) Carefully lift the cover, disconnect the earth
wire connecting the cover to the chassis, and (5) Disconnect each front panel socket from the
remove the cover. user interface board (Fig 15, 33), noting, and

Page 48 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE
where necessary recording, the connections complete with connector core inserts (Fig 9, 3)
for re-assembly. out of chassis back panel.
Refitting the Socket Moulding (3) Disconnect airswitch loom (Fig 16, 24) and
remove socket wire loom (Fig 16, 22).
6.14 Refitting of the socket moulding is the
reverse of the removal procedure. When the
socket moulding is fitted, check that when
operated, sliding doors (Fig 15, 9) pass through
photocells on Front Panel Controller Board 1
(Fig 15, 21) without fouling them. If necessary,
loosen the socket moulding, and re-adjust its
position until the sliding doors operate correctly,
and then re-tighten the socket moulding.
Note: When fitting the socket moulding ensure
that the cables are not trapped.
Removing the Back Panel Fig. 9 Rear panel (Connector board)
6.15 Remove back panel (Fig 15, 2) as follows:
CAUTION
(1) Remove cover as described in section 6.9.
The connector cores are specifically
2) Turn the unit onto its side onto a protected positioned with respect to the connector
and clean surface. bodies on the back panel. If further
(3) Remove three screws (Fig 15, 73) securing maintenance is required ensure that:
the back panel to the chassis base, and ease ♦ The connector body orientation index
the panel clear of the chassis until the ribbon (Fig 9, 5) is correctly located in notch
connector, remote control socket (Fig 16, 23) (Fig 9, 6) in back panel (Fig 9, 4).
and air switch (Fig 16, 10) can be dis-
♦ Connector core (Fig 9, 3) is correctly soldered
connected. Access is now available to all
and positioned on the back panel connector
components on the back panel.
board, to ensure that the insert orientation
(4) Disconnect mains loom from power supply matches the insert index in the connector
unit (PSU) (Fig. 3, 11). body.
(5) Disconnect video looms from Pattern
Refitting the Back Panel (Connector) Board
Generator Board.
(6) Disconnect earth cables from chassis base 6.18 Refitting of the back panel (connector)
earth stud. board is the reverse of the removal procedure (see
CAUTION in section 6.8 ).
Refitting the Back Panel
Removing the Socket Moulding Sliding Doors
6.16 Refitting of the back panel is the reverse of
the removal procedure. 6.19 Remove the socket moulding sliding doors
(Fig 10, 7) as follows:
Removing the Back Panel (Connector) Board
(1) Remove the top cover as described in
6.17 Remove the back panel (connector) board section 6.9.
(Fig 16, 3) as follows: (2) Remove the front panel moulding as described
(1) Remove back panel as described in in section 6.11.
section 6.15. (3) Remove ‘R’ shaped spring clip (Fig 10, 1)
(2) Unscrew socket core insert retaining ring from sliding door rod (Fig 10, 2).
(Fig 9, 1) on each footswitch connector, and (4) Carefully withdraw sliding door rod
ease back panel (connector) board (Fig 9, 2),

TE698883-10 Page 49 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
(Fig 10, 2) from the socket moulding, and moulding and re-adjust its position as
remove, in the following order (see Fig 10): required.
Washer (3) - Spring (4) - Washer (5) Removing the Instruction Cards
Doorstop (6) - Sliding door (7) - Doorstop (8)
Washer (9) - Spring (10) - Washer (11) 6.21 Remove the instruction cards (Fig 15 items
14, 15 and 16) as follows:-
(5) If it is necessary to remove both doors,
continue withdrawing sliding door rod (2) (1) Turn the unit upside-down on a clean and
and repeat step (4). protected surface.
(2) Remove the ten screws (Fig 15, 84) securing
Refitting the Socket Moulding Sliding Doors
the instruction card plate (Fig 15, 4) to the
6.20 Refitting of the socket moulding sliding base of the unit.
doors is the reverse of the removal procedure, but (3) Lift away the three instruction cards (Fig 15
note that: items 14, 15 and 16) complete with their
(1) The doors are correctly located around the mouldings (Fig 15, 13).
bottom of the socket moulding.
Refitting the Instruction Cards
(2) The doors operate freely and return to the
centre position when released. 6.22 Refitting of the instruction cards is the
reverse of the removal procedure, but ensure that
( 3) The sliding doors pass through the photocells
when re-fitting the instruction card plate
on Front Panel Controller Board 1 without
(Fig 16, 4), all of the cards are free to slide and
fouling them. If necessary, loosen the socket
are not trapped when tightening the fixings.

Fig. 10 Socket moulding sliding doors

Page 50 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE

Fig. 11 Removing/Refitting card frame printed circuit boards

TE698883-10 Page 51 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
REMOVING AND REFITTING PRINTED (2) Remove the front panel moulding as described
CIRCUIT BOARDS in section 6.11.
General (3) Disconnect ribbon cable.
6.23 Printed circuit boards are fitted in the unit (4) Remove six screws (Fig 15, 73) securing
either in the card frame assembly, or are secured Display Board (Fig 15, 20) to the chassis and
to the chassis or front panel moulding by nuts, remove the Display Board.
washers, and where necessary spacer pillars. The Refitting The Display Board
user interface board is attached to the chassis
6.28 Refitting of the Display Board (Fig 15, 20)
baseplate by plastic clips and cannot be removed
is the reverse of the removal procedure, but ensure
until the card frame has been taken out.
studs are located centrally in the slots.
6.24 Removal of boards fitted to the chassis or
Removing Controller Board 1
to the front panel moulding is self-evident.
However, care must be taken to ensure that all 6.29 Remove Controller Board 1 (Fig 15, 21) as
connecting wiring is removed, and where follows:
necessary the connections are recorded for correct (1) Remove top cover as described in section 6.9.
re-assembly. (2) Remove the front panel moulding as described
Boards Fitted In The Card Frame in section 6.11 and socket moulding as
described in section 6.13.
6.25 Boards fitted in the card frame plug into
connectors on the central mother board and are (3) Disconnect ribbon cable connectors.
held in position by retaining plates. Some boards (4) Remove four nuts and nylon washers and
in the card frame are interconnected by flyleads. three screws securing Controller Board 1 to
Ensure that all interconnecting leads are the spacer pillars (Fig 15, 77) on Controller
disconnected before removing the boards from the Board 2, and remove Controller Board 1.
card frame. Note also, that the Monopolar Output Refitting Controller Board 1
Module, the Pattern Generator board and Bipolar
Output board each comprise multiple boards 6.30 Refitting of Controller Board 1 is the reverse
secured together and are each removed as a unit. of the removal procedure.

6.26 The boards are locked into the cardframe Removing Controller Board 2
by a lever fitted to the board which engages a 6.31 Remove Controller Board 2 (Fig 15, 22) as
stirrup fitted to the card frame. When released, follows:
the lever is used to pull the board from the card (1) Remove top cover as described in section 6.9.
frame. When replacing the board, it is pushed (2) Remove the front panel moulding as described
into the card frame until the foot of the lever can in section 6.11 and socket moulding as
engage the stirrup, the lever is then raised which described in section 6.13.
pushes the board into the motherboard connector
and also locks the board in the card frame (Fig 11). (3) Remove Controller Board 1 as described in
section 6.29.
Note: Coding pegs are installed on the connectors (4) Remove four spacer pillars (Fig 15, 77) and
of both the board and mother board to ensure that three spacer pillars (Fig 15, 66) securing
the correct PCB is installed in a particular position. Controller Board 2 to spacer pillars
Removing The Display Board (Fig 15, 76) and remove Controller Board 2.
Note position of spacer pillars when removing
6.27 Remove Display Board (Fig 15, 20) as
them.
follows:
(1) Remove top cover as described in section 6.9.

Page 52 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE
Refitting Controller Board 2 Removing the 100V Power Supply Board
6.32 Refitting of the Controller Board 2 is the 6.37 Remove 100V Power Supply Board
reverse of the removal procedure. (Fig 15, 65) as follows:
Removing the Switch Panel Board (1) Remove card frame assembly (Fig 15, 19) as
described in section 6.35.
6.33 Remove Switch Panel Board (Fig 15, 36)
as follows: (2) Disconnect mains supply connector noting
orientation for correct reassembly.
(1) Remove top cover as described in section 6.9.
(3) Disconnect the earth wire connecting the
(2) Remove the front panel moulding as described mains connector to the card frame assembly.
in section 6.11.
(4) Disconnect the cable connecting 100V Power
(3) Remove eight nuts (Fig 15, 69) securing the Supply Board (Fig 15, 65) to the card frame
Switch Panel Board to the front panel assembly mother board.
moulding and remove.
(5) Remove four screws (Fig 15, 82) washers
(4) Disconnect the ribbon cable. (Fig 15, 79) and spacer pillars (Fig 15, 83)
Refitting the Switch Panel Board securing the Power Supply Board to the rear
6.34 Refitting of the Switch Panel Board is the panel of the card frame, and remove the 100V
reverse of the panel removal procedure, but ensure Power Supply Board.
that the nuts (Fig 15, 69) are fully tightened. Refitting the 100V Power Supply Board
Removing the Card Frame Assembly 6.38 Refitting of the 100V Power Supply Board
6.35 Remove the card frame assembly (Fig 15, 65) is the reverse of the removal
(Fig 15, 19) complete with 100V Power Supply procedure.
Board (Fig 15, 65) as follows: Removing the User Interface Board
(1) Remove top cover as described in section 6.9. 6.39 Remove User Interface Board (Fig 15, 33)
(2) Raise protective plastic covers (Fig 15, 91). as follows:
(3) Disconnect and remove all card frame boards (1) Remove card frame assembly (Fig 15, 19) as
as described in section 6.25. described in section 6.35 to expose User
(4) Disconnect the mains connector on 100V Interface Board (Fig 15, 33).
Power Supply Board (Fig 15, 65) and (2) Remove RFI shield (Fig 15, 34).
disconnect the mains connector earth wire (3) Disconnect all connections from the User
from card frame (Fig 15, 19). Interface Board to the following boards noting
(5) Ensure all other 100V Power Supply Board connector orientation for correct reassembly:
connections are disconnected. ♦ Rear panel (connector) board (Fig 16, 3)
(6) Remove four screws (Fig 15, 80) securing the ♦ Socket moulding (Fig 15, 7)
card frame to the chassis, and remove the ♦ Front Panel Control Board 2 (Fig 15, 22)
card frame. (4) Release the User Interface Board from the six
Refitting the Card Frame Assembly nylon clips (Fig 15, 71) securing it to the base
of the chassis, and remove the user interface
6.36 Refitting of the card frame assembly is the board.
reverse of the removal procedure. Replace
protective plastic covers (Fig 15, 91) on
completion.

TE698883-10 Page 53 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
Refitting the User Interface Board of the electrosurgical unit under test at impedances
6.40 Refitting of the User Interface Board is the between 100 ohms and 1200 ohms in steps of 100
reverse of the removal procedure, also ensure that: ohms. The ETS-3 test set must be used for all tests
as it has the current measuring devices installed and
(1) The six board securing clips are not damaged. is able to dissipate most power. For monopolar tests,
(2) The insulating sheet (Fig 15, 70) fitted the plate test lead must always be connected to the
between the base of the chassis and the User ETS-3 test set.
Interface Board is not damaged.
6.45 The TD800 Series test set has a group of 4 x
(3) When refitted, the User Interface Board is
200 ohm resistors which can be configured
securely held in position.
individually in series or in parallel. Each resistor
SYSTEM CHECKS is rated at 100W, but this is with convection
Eschmann ETS-3 and TD800 Series Test Sets cooling at an ambient temperature of 25 deg. C,
which is not achievable in the TD800 Series test
6.41 The Eschmann ETS-3 and TD800 Series set. For safety, no more than 50W per resistor for
test sets (see Fig 13 and 14) should be used for 10 seconds in 40 seconds should be dissipated,
routine output power checking. They can also be i.e. 25% duty cycle. This dissipation is adequate
used for checking alarm trip point positions and to obtain the test results.
operation of the footswitch functions of the
electrosurgical unit under test. 6.46 For routine checks, refer to the technical
data for the peak power and appropriate test
Note: The Eschmann ETS-3 Eletrosurgical Test Set impedance for each mode. Be aware that the peak
and TD800 Series Test Set have been discontinued. power indicated on the meter of the ETS-3 test
Although the following refer to the ETS3, any ESU set is only applicable to a 400 ohm load. For
power analyser may be used provided that it meets calculation of power at other impedances, take
the load requirements as detailed. To test alarm current reading and apply the equation:
settings, a suitable alternative is the Eschmann Alarm
Test Set, Tool 1880 (part number 111715). W (Watts) = I x I x R
Where I is the current in Amps and R is the load
Performance Checks resistance in ohms.
6.42 The following performance checks should Note: All the power measurements are subject
be carried-out every six months. Any errors found to a ±20% error.
during the checks indicate a fault in the unit and 6.47 Examples of the set ups required for various
will require the attention of an Eschmann trained test loads are as follows:
engineer.
 700 ohms Load
Test Equipment To achieve this, use the 400 ohm load in the
6.43 The following test equipment will be ETS-3 test set with the 1 Amp meter. The
required to carry out the performance checks: remaining 300 ohm load can be obtained by
♦ An oscilloscope with a time base resolution using a series 200 ohm resistor in the TD800
of at least 10 nanoseconds. Series test set together with a further two 200
♦ A current probe, or inductive pick-up loop. ohm resistors in parallel, i.e. Active from the
electrosurgical unit to the COM socket on the
♦ The Eschmann ETS-3 test set.
TD800 Series test set, with a link between the
♦ The Eschmann TD800 Series test set.
200R and the 600R sockets, and the 400R
Note: Instructions for using the ETS-3 test set
socket linked to the Active input socket (6)
are contained in the test set lid.
on the ETS-3 test set. Take reading from the
Output Power Tests 1 Amp meter and apply the equation given in
6.44 Using the ETS-3 and TD800 Series test sets it section 6.46.
is possible to check the output power of any mode
Page 54 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE
 800 ohms Load Bipolar test configuration on the ETS-3 test set
Apply an 800 ohm load by using the 400 ohm gives a load of 100 ohms, which is very close to
load in the ETS-3 test set together with the 1A the peak point for Micro and Macro. The Bipolar
meter and a further 400 ohm load via the TD800 Cut output peaks at 350 ohms and may be tested
Series test set (Active from electrosurgical unit with either 300 ohm or 400 ohm loads.
under test to the COM socket on the TD800
Series test set and link the 400R socket to the Alarm Function Tests
Active input socket (6) on the ETS-3 test set). 6.48 To test the alarm functions and verify that the
The power reading (after taking current reading correct power is being delivered from the
and applying the equation given in section 6.46) electrosurgical unit under test proceed as follows:
should be compared with the appropriate mode Notes:
graph.
(1) When carrying out the alarm function tests select
 400 ohms Load Specialist Cut setting 2 on Monopolar User 1
Use ETS-3 test set for the 400 ohms load as display on the electrosurgical unit under test.
detailed in the test set instructions. (2) All DVM measurements must be made with the
 500 ohms Load jack plug removed from the TD800 Series test
Use the ETS-3 set to 400 ohms with 100 ohms set.
on TD800 Series test set, i.e. connect  Plate Continuity Monitor (PCM) Alarm (04)
electrosurgical unit Active to the COM socket (1) Connect jack plug of TD800 Series test
on the TD800 Series test set. Also link the set lead between ‘PLATE’ socket on test
COM socket to the 200R socket and the 200R set and plate socket (53) on electrosurgi-
socket should be linked to the Active input cal unit.
socket (6) on the ETS-3 test set. Readings (2) On TD800 Series test set, set ‘ALARM
obtained should be compared with the SELECT’ switch to ‘PCM’ position.
appropriate mode graph. (3) Adjust ‘PCM’ potentiometer until the
 300 ohms Load PCM alarm on the electrosurgical unit
A test load of 300 ohms can be obtained by sounds.
using a 200 ohm load from the ETS-3 test set (4) Disconnect jack plug from ‘PLATE’ socket
and 100 ohm load from the TD800 Series test and, using DVM, measure resistance
set, i.e. connect the electrosurgical unit Active across ‘1’ (Red) and ‘2’ (Yellow) sockets
to the COM socket on the TD800 Series test on TD800 Series test set. The resistance
set. Also link the COM socket to the 400R must be between 300 and 500 ohms.
socket and the 200R socket should be linked  Plate Attachment Monitor (PAM) Alarm (05)
to the Active input socket (6) on the ETS-3 (1) Connect jack plug of TD800 Series test
test set. Link sockets 5 and 3 on the ETS-3 test set lead between ‘PLATE’ socket on test
set and connect socket 2 to the plate test lead set and plate socket (53) on electrosurgi-
of the electrosurgical unit. This configuration cal unit.
uses the more sensitive 1 Amp meter for the (2) On TD800 Series test set, set ‘ALARM
current measurement. SELECT’ switch to ‘PAM’ position.
 Parameter Change Modes (3) Adjust ‘PAM’ potentiometer until the PAM
Testing of the parameter change modes is alarm on the electrosurgical unit sounds.
carried out in a similar manner to the (4) Disconnect jack plug from ‘PLATE’ socket
monopolar tests. and using DVM, measure resistance across
‘2’ (Yellow) and ‘3’ (Black) sockets on
 Bipolar Coagulation Modes
TD800 Series test set. The resistance must
Testing of the Bipolar Coagulation modes is be between 90 and 130 ohms.
carried out using the ETS-3 test set only. The
TE698883-10 Page 55 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE
 Patient Earth Monitor (PEM) Alarm (03) ♦ Monopolar Cut and Coagulation, User 2 left and
(1) Connect jack plug of TD800 Series test right.
set lead between ‘PLATE’ socket on test
6.50 NOTE:- For safety reasons, this test must
set and plate socket (53) on electrosurgi-
be carried out at minimum power settings. To
cal unit. Connect earth cable between
check the footswitch operation of the
‘EARTH’ socket on test set and electro-
electrosurgical unit, connect the monopolar plug
surgical unit earth.
on the TD800 Series test set into each monopolar
(2) On TD800 Series test set, set ‘ALARM
footswitch socket in turn and enable the
SELECT’ switch to ‘PEM’ position.
appropriate output by pressing the appropriate
(3) Adjust ‘PEM’ potentiometer for short cir-
‘MONOPOLAR CUT or COAG’ buttons on the
cuit conditions and wait until PEM alarm
TD800 Series test set. Repeat the procedure to
sounds. With alarm on, adjust ‘PEM’
test the bipolar footswitch sockets, but using the
potentiometer until alarm stops.
‘BIPOLAR CUT and COAG’ buttons and the
(4) Disconnect jack plug from ‘PLATE’
bipolar plug.
socket and earth cable from ‘EARTH’
socket and, using DVM, measure 6.51 To check the remote control function of the
resistance between ‘3’ (Black) socket electrosurgical unit, proceed as follows:
and ‘EARTH’ green socket on TD800  Remote User 1 and 2.
Series test set. (The resistance must be (1) Connect remote control plug of TD800 Series
between 220 and 260 ohms). test set to electrosurgical unit remote control
 Patient Voltage Monitor (PVM) Alarm (02) socket.
(1) Connect active diathermy current to (2) Set up a low monopolar output power level
‘ACTIVE’ socket on TD800 Series test on the electrosurgical unit and depress a
set. Connect earth cable between footswitch or fingerswitch button to activate
‘EARTH’ socket on test set and elec- the unit (the remote control only functions
trosurgical unit earth. after the unit has been activated once). Check
(2) Adjust Specialist Cut to give 0.33A (dis- that the output power of the selected mode
play power setting 38 ± 10). With power increases or decreases when the appropriate
applied, press ‘PVM’ test button to ob- ‘REMOTE’ button on the TD800 Series test
tain alarm. set is operated.
Note: Do not press ‘PVM’ test button Note: The ‘USER’ selected button and
until a stable reading is displayed on meter. the ‘+ (UP)’ or ‘— (DOWN)’ buttons on the
TD800 Series test set must be operated
(3) Adjust Specialist Cut to give 0.27A (dis- simultaneously.
play power setting 34 ± 10). With power
applied, press ‘PVM’ test button. Alarm 6.52 The output waveforms may be sampled
should not be activated. using an oscilloscope and current probe or
Note: The trip point for the PVM alarm is inductive pickup loop on the plate connector lead
approximately 0.30A. to the power load test set. Adjust the time base
and gain of the oscilloscope to obtain waveforms
Footswitch Control Checks similar to those shown on the following pages.
6.49 The footswitch sockets on the Variation in test equipment will mean that you may
electrosurgical unit under test can be checked for not achieve exactly these waveforms. They are
the following: intended as a guide only.
♦ Bipolar Cut and Coagulation.
♦ Monopolar Cut and Coagulation, User 1 left and
right.

Page 56 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE

(a) Monopolar Pure Cut (160W) - 400 ohm load

(b) Monopolar Specialist Cut (210W) - 400 ohm load

Fig. 12 Output wave form diagrams (Sheet 1)

TE698883-10 Page 57 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE

(c) Monopolar Blend (130W) - 400 ohm load

(d) Monopolar Pinpoint Coagulation (90W) - 400 ohm load

Fig. 12 Output wave form diagrams (Sheet 2)

Page 58 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE

(e) Monopolar Spray Coagulation (80W) - 400 ohm load

(f) Bipolar Coagulation Normal full power (50W) - 100 ohm load

Fig. 12 Output wave form diagrams (Sheet 3)

TE698883-10 Page 59 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE

(g) Bipolar Coagulation Normal full power (50W) - 100 ohm load

(h) Bipolar Coagulation Normal half power (24W) - 100 ohm load

Fig. 12 Output wave form diagrams (Sheet 4)

Page 60 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE

(i) Bipolar Cut Normal full power (40W) - 200 ohm load

Fig. 12 Output wave form diagrams (Sheet 5)

TE698883-10 Page 61 of 71
Part 6 TD850 ELECTROSURGICAL UNIT

6. MAINTENANCE

Fig. 13 ETS-3 Test Set (REF 83-099-65)

Page 62 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 6

6. MAINTENANCE

Key to leads:- A - Bipolar footswitch B - Monopolar footswitch


C - Remote D - Plate
E - Earth

Fig. 14 TD800 Series Test Set (REF 83-099-73)

TE698883-10 Page 63 of 71
Part 7 TD850 ELECTROSURGICAL UNIT

7. ILLUSTRATED PARTS LIST

NOTE: The front panel seal (item 18) was 713751,


this is no longer available. The replacement seal
731116 is now supplied in a kit (731118) of 3
strips with revised fitting instructions which
should be followed when changing the seal.

Fig. 15 TD850 Main assembly

Page 64 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 7

7. ILLUSTRATED PARTS LIST

Fig. 15 Part No. Description Qty. per


Item No. Unit

1 713803 Chassis 1
2 714120 Back panel assy. (see Fig 17) 1
3 714102 Top cover 1
4 714016 Instruction card plate 1
5* 714106 Serial label 1
6 713960 Front panel moulding 1
7 713801 Socket moulding 1
7a 715582 Socket moulding assembly (with sockets) 1
8 713802 Handle moulding 2
9 713804 Sliding door moulding
10* 714188 Mains lead
11 713911 Adhesive screw covers 3
12* 713797 Feet 8
13 713808 Instruction card moulding 3
14 713928 Instruction card 1 (Basic set-up 1) 1
15 713935 Instruction card 2 (Basic set-up 2) 1
16 713936 Instruction card 3 (Alarms and codes) 1
17 713807 Knob 1
18 731118 Front panel seal kit (see note in Fig. 15) 1
19 713709 Card frame assy.(inc. mother board) 1
20 713829 Front panel display board 1
21 713838 Front panel control board 1 1
22 714396 Front panel control board 2 1
(including daughter boards)
23 715594 Regulator board 12V 1
24 713720 Regulator board monopolar 1 1
25 713889 Regulator board monopolar 2 1
26 713890 Regulator board bipolar 1
27 713738 Tone generator board 1
28 713827 Patient alarm board 1
29 715576 Pattern generator board 1
(including daughter board)
30 715575 Monopolar output board 2
(assy. of 2 boards)
31 714606 Monopolar 1 output logic board 1
32 714392 Bipolar output board 1
(including daughter board)
33 713836 User interface board 1
(including daughter boards)
33a 713840 User interface daughter board 19
34 714112 RFI shield 1
35 714647 Monopolar 2 logic board 1
36 714006 Switch panel board 1
37* 713906 Back panel wiring loom 1
38 713969 Knob clamp spring 1
39* 713964 Cable assy. (100V PSU to 1
card frame mother board)
40* 713951 Output loom part 1 (monopolar 1 1
output socket to user interface board)
* Item not illustrated Continued:

TE698883-10 Page 65 of 71
Part 7 TD850 ELECTROSURGICAL UNIT

7. ILLUSTRATED PARTS LIST

Fig. 15 Part No. Description Qty. per


Item No. Unit

41* 713952 Output loom part 2 1


(bipolar output socket to user
interface board)
42* 715578 Output loom part 3 1
(monopolar 2 output socket to
user interface board)
43* 713954 Output loom part 4 1
(bipolar board to user
interface board)
44* 715579 Output loom part 5 1
(plate and active board to
alarm and user interface boards)
45* 713993 Cable assy. 1
(plate socket to alarm board)
46* 713994 Cable 1 (output boards, logic 2
to spray and TX)
47* 713995 Cable 2 (Output boards, logic 2
to push/pull
48 713923 Socket, white, 4mm 2
49 713924 Socket, red, 4mm 2
50 714399 Socket, black, 4mm 2
51 714399 Socket, black, 4mm 2
52 713947 Socket, white, 2mm 1
53 713948 Plate socket, 3-way, 6.35mm 1
54 713946 ‘S’lead socket, 3mm, 2W 1
55 713898 Bovie socket, 8mm 2
56 713900 Bovie socket spring 2
57 713899 Bovie socket spacer 2
58 713901 Bovie socket nut 4
59 711917 Plate socket nut 1
60 713885 Sliding door stops 4
61 713886 Sliding door rod 1
62 713927 Sliding door spring 4
63 713929 Sliding door ‘R’clip 2
64 710700 Washer, plain, M4 20
65 713745 100V PSU 1
66 714131 Spacer (attaching controller 3
board 2 to controller board 1)
67 714114 PCB retaining plate 2
68 709991 Screw, M2.5 x 10 10
(attaching retaining plates)
69 713979 Nut, half, M4 8
(attaching switch panel board)
70 695150 Elephant hide (user interface board 1
/chassis base)
71 695577 Nylon clips (attaching user 6
interface board to chassis base)
* Item not illustrated Continued:

Page 66 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 7

7. ILLUSTRATED PARTS LIST

Fig. 15 Part No. Description Qty. per


Item No. Unit

72 713959 Spacer, M4 x 16 (display board 6


attachment)
73 710171 Screw, pan head, M4 x 10 (attaching 9
display board to chassis panel and
back panel to chassis)
74 710580 Screw, socket butt, M4 x 10 (attaching 6
front panel moulding to chassis)
75 710156 Screw, socket head, M6 x 65 3
(attaching front panel socket
moulding to chassis)
76 713931 Spacer, male/female, M4 x 10 4
(attaching control boards 1 and 2
to chassis)
77 713932 Spacer, male/female, M4 x 25 7
(attaching control boards 1 and 2
together)
78 710839 Nut, full, M4 (attaching control 11
board 2 to chassis and for earth studs)
79 307231 Washer, nylon, M4 (attaching 100V 18
PSU, control boards 1 and 2 and
display board)
80 710219 Screw, socket cap, M4 x 20 4
(attaching card frame to chassis)
81 713933 Screw, socket head, M4 x 45 4
(attaching top cover to chassis)
82 710213 Screw, socket head, M4 x 20 4
(attaching 100V PSU to card frame)
83 730246 Insulated spacer, 12mm 4
(attaching 100V PSU to card frame)
84 710170 Screw, pan head M4 x 5 (attaching 4
instruction card baseplate to chassis)
85 713971 Screw, shoulder, M6x12 (attaching 4
handle to chassis)
86* 697395 Sleeve, 1 in. (large) 6
87* 714005 Plastic cap 2
88* 714007 Earth lead (card frame to chassis) 1
89* 714013 Earth lead (back panel to chassis) 1
90 714012 Earth lead (top cover to chassis) 1
91 714218 Cable protection card 2
92* 695777 Tie-wrap (attaching cable 6
protection cards)
93 710731 Washer, shakeproof 3
94* 714156 Top cover label (loose) 1
95* 714170 Front panel label (loose) 1
96* 714186 Back panel label (loose) 1
97* 714189 Instruction manual 1

* Item not illustrated Continued:

TE698883-10 Page 67 of 71
Part 7 TD850 ELECTROSURGICAL UNIT

7. ILLUSTRATED PARTS LIST

Fig. 16 Rear panel assembly

Page 68 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT Part 7

7. ILLUSTRATED PARTS LIST

Fig. 16 Part No. Description Qty. per


Item No. Unit

1* 714120 Back panel assy. RF


(see Fig 4.1, item 2)
2 714017 • Back panel 1
3 713842 • Back panel board 1
4 718668 • Potential equalisation point 1
5 718669 • Colour coded washer 1
6 713963 • Washer, shakeproof, M6 1
7 710372 • Nut, M6 2
8 710702 • Washer, plain, M6 11
9 714013 • Earth lead (rear panel to chassis) 1
10 696480 • Airswitch 1
11 709891 • Screw, csk, M3 x 5 2
12 714082 • Potentiometer, 100k 2
13 714111 • Potentiometer, 250k 1
14 714079 • Potentiometer knob, grey 3
15 714080 • Knob cap, grey 3
16 714081 • Nut cover, grey 3
17 714019 • Grommet (for volume control) 1
18* 713906 • Back panel wiring loom 1
19 713902 • • Mains block 1
20* 714088 • Video synch cable 1 1
21* 714089 • Video synch cable 2 1
22* 714002 • Remote socket wire loom 1
23 713914 • • Remote socket 5-way 1
24* 713958 • Airswitch wire loom 1
25* 606312 • Tie-wrap base (for item 18) 2
26* 695776 • Tie-wrap, small (for item 18) 2
27 714097 • BNC connector, insulated 2

* Item not illustrated

TE698883-10 Page 69 of 71
Part 8 TD850 ELECTROSURGICAL UNIT

8. CLEANING DISINFECTION AND CARE


CLEANING AND DISINFECTION
CAUTION
The electrosurgical unit MUST be
disconnected from the mains power supply
prior to cleaning.
8.1 The cabinet should be cleaned with a
sponge or cloth dampened in a mild detergent or
soap solution and wiped dry.
(1) Do not use excessive amounts of fluid which
could enter the cabinet and damage internal
components.
(2) The cabinet front panel can be wiped, as it is
a continuous membrane. Pressure on the
touchbuttons, with the power supply off, will
not cause mode changes.
8.2 For disinfection, Eschmann Equipment
recommend primarily the use of alcohol based
disinfectants, e.g. 70% isopropanol/water or 70%
methyl alcohol/water.

CARE
8.2 Ensure that electrosurgical safety checks and
routine maintenance are carried out at regular
intervals and only by Eschmann trained personnel
otherwise the warranty for the equipment could
be infringed.

Page 70 of 71 TE698883-10
TD850 ELECTROSURGICAL UNIT

Fig. 17 TD850 System block diagram

TE698883-10 Page 71 of 71

You might also like