General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 3 di 37
Oggetto
INDEX
1. GENERAL NOTES ........................................................................................................................5
1.1. Typographic conventions .......................................................................................6
1.1 Symbols of advice positioned on the equipment: ...................................................6
2. INTRODUCTION ...........................................................................................................................7
3. GENERAL INFORMATION ...........................................................................................................8
3.1. The range ...............................................................................................................8
4. POWER SUPPLY..........................................................................................................................8
5. ENVIRONMENTAL CHARACTERISTICS AND FUNCTIONING LIMITS.....................................8
6. RISKS LINKED WITH THE WASTE ELIMINATION .....................................................................9
7. THE PERFORMANCES ..............................................................................................................10
7.1. Image Intensifiers .................................................................................................10
7.2. Dynamic Recursive Filter (DRF)...........................................................................10
7.3. Pulsed fluoroscopy ...............................................................................................10
7.4. Images memorization ...........................................................................................10
9 TV CHAIN CHARACTERISTICS ........................................................................................................11
10 ANALOG TV MONITOR CHARACTERISTICS ................................................................................12
10.1 Dose Control Device ............................................................................................13
10.2 Monitor Support -optional ....................................................................................13
1.2 Dolly .....................................................................................................................13
11 ANALOG TV CHAIN O-I HR 9”, 12”HR CALIBRATION PROCEDURE .....................................14
11.1 Image centering and sizing adjustments ..............................................................14
11.2 Image Direction Adjustment .................................................................................15
11.3 Camera Focus Adjustment...................................................................................16
11.4 Iris /Dose Calibration/ABC calibration ..................................................................17
11.5 Output video signal adjustment ............................................................................18
11.6 Antivignetting........................................................................................................19
11.7 Black level adjustment..........................................................................................20
12 MAINTENANCE ..........................................................................................................................21
12.1 GMM X-ray dose calibration.................................................................................21
12.2 X-ray automatic dose control................................................................................21
13 GENERATOR CONFIGURATION ..............................................................................................22
14. CAMERA INTERFACE................................................................................................................27
15. INTERFACE WITH THE TABLE OPERA T ................................................................................28
INTERFACCIA N33 CON GENERATORE INDICO...............................................................................29
CONNESSIONE CONTROLLO ABS E SYNC ......................................................................................30
CONNESSIONE N33 CON INTENSIFICATORE THALES ...................................................................31
TOSHIBA TUBES ADAPTER 9” 12” ......................................................................................................31
CONNESSIONE N33 CON TELECOMANDATO ..................................................................................32
16. NICAL INTERFACE.....................................................................................................................33
18. TOPOGRAPHIC SCHEME .........................................................................................................35
19. SPARE PARTS ...........................................................................................................................36
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 4 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 5 di 37
Oggetto
1. GENERAL NOTES
WARNING!
X-RAY EQUIPMENTS CAN BE DANGEROUS TO THE HEALTH OF BOTH PATIENTS AND OPERATORS
UNLESS PROPER SAFETY MEASURES ARE STRICTLY OBSERVED.
Although this equipment was designed and manufactured according to the most up-to-date
safety Standards, the X-ray beam always represents a danger, should not the operator be
properly qualified and trained. Excessive exposures to X-rays cause damages to the human
body: therefore, all the necessary precautions must be taken to prevent unauthorized or
unskilled personnel from operating this equipment, thus jeopardizing themselves and other
people.
Note: This equipment is in compliance with the requirements established by the 89/336/CEE
Directive applying to the Electromagnetic Compatibility.
In order to comply with such requirements, special technical measures were adopted (such as
the application of anti-noise filters and/or ferrites on some circuits) and precise earth
connections were defined. The metallic covers and their clamps are realized and checked by
means of fixing devices or special conductive material-gaskets ensuring both the immunity to
external sources of disturbance and the limitation of possible internal disturbances generated
by the electrical circuits.
Therefore, it is recommended that the technicians executing possible maintenance or
fixing interventions, should not modify the path of the connections and carefully
reconnect all earth connectors of the gaskets (should these be removed during the
maintenance or fault research operations) and restore the protection gaskets with the
utmost attention in the same conditions and with the same fixing devices originally
foreseen.
GENERAL MEDICAL MERATE S.p.A. shall decline all responsibility with regard to
the good functioning of the equipment should the installation and/or the maintenance of
if be executed by unauthorized personnel.
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 6 di 37
Oggetto
Important information
Example:
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 7 di 37
Oggetto
2. INTRODUCTION
The Analog Tv chain Opera I 9” HR -12” HR identify a “radiological system” consisting in:
a high-resolution camera, Monitor and I. I. The Analog Tv chain aim is to acquire images
through the X ray digital fluoroscopy.
1) OPERA I 9 HR :
¾ Image intensifier 9”
¾ TV chain 1024x1024 (Last Image Hold) with dynamic recursive filter
and with the opportunity to memorize one image at the release of the foot
switch fluoroscopy
¾ Possibility to memorize 99 images in cine loop
¾ Possibility to acquire images in pulsed fluoroscopy mode 1-3-6-12 F/sec
¾ B&W Monitor LCD 19”
Optional: Additional Monitor and monitor dolly
2) OPERA I 12HR:
¾ Image intensifier 12”/13”
¾ TV chain 1024x1024 (Last Image Hold) with dynamic recursive filter
and with the opportunity to memorize one image at the release of the foot
switch fluoroscopy
¾ Possibility to memorize 99 images in cine loop
¾ Possibility to acquire images in pulsed fluoroscopy mode 1-3-6-12 F/sec
¾ B&W Monitor, LCD 19”
Optional: Additional Monitor and monitor dolly
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 8 di 37
Oggetto
3. GENERAL INFORMATION
19 "LCD"
O-I 12” 12 "/13“ 1024x1024 DRF
4. POWER SUPPLY
Power supply:
Available power supply: 24Vdc ± 20%
Electrical input: 20W
Insulation class:
First class
With applicable parts: no one
For monitor power supply make reference to monitor manual
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 9 di 37
Oggetto
The following table gives all information applying to the material composing the equipment,
as well as the relevant quantity and characteristics, so that at the end of its functioning life, it
will be possible to provide for the disposal of the product to be carried out with minimized
risks, according to the specifications established by the provisions of the law about waste
elimination.
In any case, at the end of the functioning life of this equipment, upon the Customer's
request, GMM will be glad to supply a proper assistance for the elimination and/or
scrapping of the product.
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 10 di 37
Oggetto
7. THE PERFORMANCES
For the 12"/13”version, a particular attention was paid to the magnetic shielding in order to
contain the typical S distortion
7.3.Pulsed fluoroscopy
The chain system can aquire images in fluoroscopy mode with variable speed of 1-3-6-12
frame for a second
7.4.Images memorization
It is possbile to to memorie until 99 images and review them in in cineloop mode with
variable speed of 12-6-3-1 frame for a second.
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 11 di 37
Oggetto
9 TV CHAIN CHARACTERISTICS
Operating conditions :
Working temperature range: -10°C to +45°C
Store temperature range: -40°C to +55°C
Humidity: Operating to 95% relative humidity (non-
condensing)
Mechanical specifications :
Control unit: dimension (WxDxH)/ Weight : 235x192x65mm 1500gr
Camera head: dimension (φ x H)/Weight : (Standard) 127x108mm 1400gr
(with Rotation) 125x95mm 1400gr
Approvals :
Safety: EN 60601-1
EMC: EN 60601-1-2
Others: CE-label according 93/42 CEE directive for
medical devices.
Options :
360° Continuous rotating camera head (with slip
ring)
Neutral density filter
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 12 di 37
Oggetto
Model
MOLVL 1964C/MED
Specifications
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 13 di 37
Oggetto
1.2 Dolly
The dolly is the most common and simple sytem in order to support the TV monitor . It has
cartwheels and accessories shelves
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 14 di 37
Oggetto
ALLEN SCREW
ADJUSTABLE
DISTANCE RING
CCD
ADAPTER
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 15 di 37
Oggetto
Fig 3.1 Allen screw at the I. I. center field Fig. 3.2 MonitorImage adjustment direction
¾ Loosen the Allen screw positioned on the head camera side and turn it around until
you will reach the right position in the monitor (Fig. 3.2 Monitor image adjustment
direction)
¾ Fasten the Allen screw.
¾ In case you will not reach the right image direction, please move the jumpers
positioned on the interface board as follows:
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 16 di 37
Oggetto
¾ Place on the patient tabletop a “MIRA FUNK TEST or CONVERGIN LINE PAIR”
¾ Loosen the head camera cover
¾ Find on the printed circuit board the focus hole place (Fig. 4CCD head camera focus
adjustment)
Insert the screwdriver inside the focus hole place and turn it around to the right or to the left
side until you will obtain the maximum resolution
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 17 di 37
Oggetto
¾ Cover the X-ray output collimator area with a copper sheet of 1millimetre thick.
¾ Make sure that the grid is in
¾ Select 70 kV on the X ray generator and adjust the mA to obtain 35 µR/s (±10%).
¾ Loosen the camera head cover.
¾ Find the hole for Iris adjustment (Fig.5 Iris CCD camera) on the printed circuit
¾ While performing fluoroscopy, insert the screwdriver into the regulation diaphragms
hole and set the Iris screw in order to obtain the two LEDS (red and green) turned off
on N30 PWR20 board.
¾ Check if J8 and J7 on N30 PWR20 board, are both positioned in 2-3 position
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 18 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 19 di 37
Oggetto
11.6 Antivignetting
Fig. 7 Antivegnitting
Fig. 8 Vignetting
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 20 di 37
Oggetto
N30 PWR20
NDR_ADV0
1 1
J8
2 2 1 2
3 3 J7
J5 J6 1 2
J4
R R R84 R79 R
101 100 106
¾ Check if the Auto control black level function is inserted (J6 pin position 1-2
(Fig.9 Trimmers adjustments)
¾ Adjust the black level to 30mV by moving theTrimmer R101, but without starting the
X- ray function
¾ Place the PHILIPS phantom on the patient tabletop and start the X- ray function
¾ Adjust the Trimmer R100 to reach the black level to 50 mV
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 21 di 37
Oggetto
12 MAINTENANCE
Every 6/12 months it is advise to control if the x-ray beams dose is in compliance with the
requirements established by the country low directive where the Tv chain equipment is
installed.
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 22 di 37
Oggetto
13 GENERATOR CONFIGURATION
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 23 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 24 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 25 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 26 di 37
Oggetto
Fig. 13 JW22
Warning!
Check if the internal monitor switch is positioned onto ON, make reference
to the following figure
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 27 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 28 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 29 di 37
Oggetto
J2
J4C
N33 J2 to TDI J2
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 30 di 37
Oggetto
0V24
ABS
SYNC
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 31 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 32 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 33 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 34 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 35 di 37
Oggetto
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 36 di 37
Oggetto
MODEL DESCRIPTION
Ver .9
General Medical MERATE SpA 06/Nov/07
Via Partigiani, 25 File Opera I 9HR 12HR
24068 Seriate (BG) Italy Tipo Service Manual
e-mail: info@gmmspa.com pag. 37 di 37
Oggetto