CR Mammography Solution
based on CR 30-Xm
CR Mammography Solution
Service Manual
► Document History
► Referenced Documents
Document Title
For a list of referenced documents see section 3.
Edition 1, Revision 0
07-2012 Printed in Germany Document Node ID: 36784739
eq_standard_e_template_v09
Agfa Company Confidential Copyright © 2012 Agfa HealthCare N.V.
DD+DIS159.12E CR Mammography Solution
► Manufacturer
Agfa HealthCare N.V.
Published by
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright 2012 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
Agfa and the Agfa rhombus are trademarks of Agfa-Gevaert N.V., Belgium,
or its affiliates.
WARNING:
Improper operation or service activities may cause damage or injuries.
INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see Agfa HealthCare Library > General Info > Agfa HealthCare > Publications >
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.
IMPORTANT:
The installation and service of the product(s) described herein is to be performed
by qualified personnel who are employed by Agfa HealthCare or one of its affiliates
or who are otherwise authorized by Agfa HealthCare or one of its affiliates to provide
such services.
NOTE:
To verify the latest version of a chapter of the Service Manual refer to the “Checklist for
Completeness” in the Agfa HealthCare Library.
LIST OF CONTENTS
2 SAFETY DIRECTIONS.............................................................................................................8
3 PREREQUISITES.....................................................................................................................9
4 CR MAMMOGRAPHY SOLUTION SET-UP...........................................................................11
5 PREPARATION ......................................................................................................................12
5.1 Checklist for the Sales Representative...................................................................................12
6 INSTALLATION ......................................................................................................................21
6.1 Installation Checklist ...............................................................................................................21
7 CONFIGURATION..................................................................................................................23
7.1 Configuring the Central Monitoring System (optional)............................................................23
1 Solution Overview
The monitor of the CR Mammography workstation has not been approved for final
interpretation of examinations. Images shown on this monitor are for validation of
patient positioning purposes only.
Constancy Quality Control (QC) ensures that the performance level of an existing
CR Mammography Solution is still equivalent to the level defined during solution setup.
This is the responsibility of the customer.
An X-ray modality alignment is mandatory to use the CR Mammography Solution.
CAUTION:
Incorrect X-ray Modality setup can cause serious performance problems.
The solution components and the X-ray Modality have to be configured before
Mammography examinations will be performed.
The solution components have to be configured by an Agfa CR Mammography
specialist (FSE/CAS), whereas the X-ray Modality needs to be configured by the
modality service engineer, in presence of the Agfa CR Mammography specialist
(FSE/CAS).
FSE - Field Service Engineer
CAS - Clinical Application Specialist
1.1 Components
The components released for the CR Mammography Solution are listed in the
CR/DR Interoperability Matrix, see section 3.
The NX workstation and the CR 30-Xm Digitizer always belong together meaning that
each CR Digitizer has to be connected to one dedicated NX workstation.
The identification of the CR Mammography cassettes are done via Fast ID workflow,
this means no ID tablet is required: When the cassette is inserted into the CR Digitizer
the user enters the patient data at the NX workstation and it is transferred directly to
the digitizer.
IMPORTANT:
The deviations to the released configuration of the CR Mammography Solution have to
be reported in the Final Acceptance Certificate, which must be signed by the customer.
2 Safety Directions
CAUTION:
Incorrect X-ray Modality setup can cause serious performance problems.
The solution components and the X-ray Modality have to be configured before
Mammography examinations will be performed.
The solution components have to be configured by an Agfa CR Mammography
specialist (FSE/CAS), whereas the X-ray Modality needs to be configured by the
modality service engineer, in presence of the Agfa CR Mammography specialist
(FSE/CAS).
FSE - Field Service Engineer
CAS - Clinical Application Specialist
WARNING:
X-ray dose could be too high for the patient.
Do not use uncalibrated dosimeters.
Only order the Agfa recommended dosimeters.
WARNING:
X-ray dose could be too high for the patient.
It is mandatory to remove older versions of PMMA blocks you bought locally and
to use the ones delivered by Agfa, which are internally tested and approved.
This ensures that the measurement tolerances of the alignment procedure are fulfilled.
3 Prerequisites
REQUIRED TOOLS
USB memory stick (checked to be virus-free), for Configuration purposes
Cleaning Material:
Please refer to the CR 30-X/CR 30-Xm Plates and Cassettes User Manual to
order the appropriate cleaning material for the CR MM3.0T Mammo Cassettes.
See in the table below.
DOCUMENTS:
Document Hyperlink
a) Project Management Guide
for the CR Mammography Solution based Intranet Link / Extranet Link
on CR 30-Xm
b) CR Mammography Solution User
Intranet Link / Extranet Link
Manual(s),
Intranet Link / Extranet Link
document number UM 2344/2444
c) CR 30-X/CR 30-Xm Plates and Cassettes
Intranet Link / Extranet Link
User Manual, document number UM 2387
d) CR/DR Interoperability Matrix Intranet Link / Extranet Link
e) CR 30-Xm Digitizer Service Manual Intranet Link / Extranet Link
f) NX Service Manual
Agfa HealthCare Library
version 2.0.8000/NX 3.0.8000 or higher,
Computed Radiography
- Chapter 4 - Installation and Configuration
CR Workstation Software*
- Chapter 5 - Monitor Options
g) Electronic Licensing Management System -
Intranet Link / Extranet Link
Service Documentation
h) Connectivity Release Document for XRDI Agfa HealthCare Library General
Software Info Connectivity & Application*
i) Marketing Newsletter on Mammo XRG
Intranet Link*
connection with NX Mammo
j) Application Note - Correct Use of Musica
Mammo Image Processing for NX 2008 and Intranet Link / Extranet Link
higher
* Only applicable for Agfa Employees.
For external partners: Contact your local Agfa representative.
Document Hyperlink
k) Musica MCE (Micro Calcification
Enhancement) Image Processing Software Intranet Link / Extranet Link
document
l) CR Mammography ATP Service Manual
Intranet Link / Extranet Link
DD+DIS010.12E
m) CR Mammography ATP Tool – Version 5.2
Intranet Link / Extranet Link
or higher
n) Tools for the CR Mammography Solution
Intranet Link / Extranet Link
document
o) Application Note - Basic Manual Quality
Control Procedure Intranet Link / Extranet Link
DD+DIS 007.12E
p) Interface Status Report (ISR) Intranet Link*
Agfa Healthcare Library General
Info Connectivity & Application
q) Connectivity Release Documents
To [Manufacturer] Connectivity
Release Document *
* Only applicable for Agfa Employees.
For external partners: Contact your local Agfa representative.
5 Preparation
Services
(including Professional Services):
Licenses:
End
X-ray Modality
o Vendor:
o Type:
o Software Version:
o X-ray exposure parameters:
o Acceptance test report including the X-ray modality alignment
results (if available):
o Connectivity Set foreseen or available*:
Film/screen solution
o Film, vendor and type:
o Screen, vendor & type:
o Chemistry, vendor and type:
List of data printed in the data field of the film (if applicable):
Professional services
Check that the following tasks must be ordered as professional Services:
Additional acceptance testing (e.g. EUREF tests) beyond
Agfa Acceptance Test procedure
Constancy Quality Control (QC) to be performed by Agfa
Additionally optimization of Mammography image processing in the
workstation (from NX 2008 on)
Sketch of environment
Make a sketch of the involved rooms, e.g. X-ray room, reporting room.
2 remarks have to be taken in account:
Existing equipment involved in the CR Mammography Solution and
the location of the new equipment should be marked.
CR Mammography workstation shall be in the X-ray room in case
there will be connectivity with the X-ray modality (workflow
improvement).
Radiologists:
o Name:
o Name:
Remarks
End
CR equipment
In case of a pure Mammography environment (only Mammo X-ray
modality available), ensure that the Al-Filter is available.
Define and plan the necessary soft- and hardware upgrades for the
components of the CR Mammography Solution.
Ensure that the license configuration files will be generated and
available for the on site visit.
Professional Service
Inform sales representative about potential options to sell professional
service based on the experience of the site.
End
Professional Service
Check for customer specific agreements.
Remarks
Explain the customer representative that with his/her signature on the Final
Acceptance Certificate he/she accepts the present conditions of the CR
Mammography Solution.
End
6 Installation
Following checklist provides the necessary steps for installing the CR Mammography
Solution based on CR 30-Xm.
These steps have to be performed by the CR Mammography Specialist (FSE)
CR Workstation
Check the software version: NX 2.0.8600/NX 3.0.8600 or higher
Remarks
End
7 Configuration
IMPORTANT:
It is the responsibility of each country/region to optimize the default configuration
according to the local requirements and to re-use this configuration to avoid
unnecessary repetition of work in the field.
As a starting point, the Agfa default configuration has to be used.
Required time:
Approximately 60 minutes (without Central Monitoring System)
All properties of the NX configuration are taken by the CMS, except the User Interface.
This has to be configured separately on the CMS.
Therefore, the GenRad properties can not be used accidentally for CR Mammography.
For the NX 2.0.8600 / NX 3.0.8600 and higher (Service Updates and Service Packs
inclusive), the configuration of the Operating System is described in the NX Service
Manual, Chapter 4 - Installation and Configuration.
See referenced documents in section 3.
For NX 2.0.8600 / NX 3.0.8600 and higher (Service Updates and Service Packs
inclusive), the Setup of a Default Configuration for NX Application is described in the
NX Service Manual, Chapter 4, - Installation and Configuration.
See referenced documents in section 3.
The Start-up Exam Tree for the CR Mammography application is installed by default in
the Configuration Tool. If not, proceed as described below:
NOTE:
In case there is a connection to an IMPAX softcopy station in the CR Mammography
Solution, keep in mind that any adaptation of the site configuration like hospital name
and department name has to be synchronized with IMPAX.
(1) Go to:
General General Settings Configure Workstation
Settings
NOTE:
The IP address of the CR Mammography workstation is automatically transferred from
Windows level to the NX Workstation Settings.
(2) Fill in and select the data of the local installation, such as: Character Set
(3) Mark the checkbox <Enable Fast ID> and fill in the appropriate AE Title.
(3) Enter:
IP Address
AE Title
Default Router
Subnet Mask
NOTE:
The Emergency Procedure settings have no impact on the Mammography application.
These are required settings for a faultless digitizer operation.
IMPORTANT:
If no printer is available on site, a dummy printer has to be configured to display the
images in the proper size in the Editing screen (refer to section 7.2.6.3).
NOTE:
The following procedure covers the default settings for the Agfa printers released
for CR Mammography.
For other vendors refer to the respective Connectivity Release Documents
as listed in the Interface Status Report (ISR). See referenced documents in section 3.
(4) If a Mammography printer is selected, the settings are already entered by default.
Make sure the AE title is correct and enter the missing IP Address as
shown in Figure 9.
NOTE:
It is supported to print two images on the 11 x 14 format.
Be aware that due to the True Size printing a certain part (opposite
side of the chest wall side) of the image is not printed and therefore
lost. Especially on the 24 x 30 images the diagnostic area can be
affected. For such films, configure the Text box with the mention
“Not for diagnostic”.
It is the responsibility of the customer to decide if the printed image
can be used for diagnose or not.
(5) Measure the view box Luminance (View Box Illumination) as follows:
Switch on the view box and wait about 5 minutes.
Remove any film from the view box.
Collimate for a Mammography study (four films).
Measure the luminance (View Box Illumination) in the center of this
area and in each corner 10 cm from the edges.
Calculate the median of the measured values.
(6) Fill in the View Box Illumination as follows:
NOTE:
The following procedure covers the default settings for the softcopy stations released
for CR Mammography. For other vendors refer to the respective Connectivity Release
Documents on Agfa HealthCare Library, see section 3.
NOTE:
In case of non Agfa softcopy devices, select Generic as Type and configure the
device according the Connectivity documents.
NOTE:
In case of sending images to CAD systems, use MG for Processing or DX for
Processing as output types.
NOTE:
The feature Background darkening is implemented in the NX workstation. The role
of this feature is to enhance the breast details by darkening the background of the
image. The background darkening can be transmitted to the softcopy station.
To do so, following configurations are needed:
Either the new attribute Pixel Padding Range Limit is supported in the
softcopy station.
Or the option GSPS supported is checked on the NX (see Figure 14).
In both cases, the Output Type must be 15bit PVAL VOIT LUT.
NOTE:
For NX 2.0.8200 / NX 3.0.8200 and higher, the feature Musica MCE is implemented in
the NX workstation. The role of this feature is to enhance potential micro calcifications
on mammography images. The enhancement is only visible on a softcopy station.
To get the enhanced images on the softcopy station, check the box Accepts Musica
Micro Calcification Enhanced images (see Figure 14).
For details about the Musica MCE, refer to the Musica MCE (Micro Calcification
Enhancement) Image Processing Software document, see section 3.
NOTE:
It is possible to store more than one export destination with different export settings.
When exporting images, it is easily possible to switch between the different configured
destinations in the NX application Main Menu without having to change it in the
Configuration Tool.
Export functionality is only available with NX 3.0 Features License.
Before starting the configuration of the X-ray device in the NX workstation, following
steps have to be performed:
NOTE:
If the license X-ray Mammo Modality connection is not available, the XRDI Software
can not be installed.
(5) If it is not there, install the XRDI software using the NX Starterkit DVD 1 as
explained below. Alternatively, download the XRDI software from the Agfa
HealthCare Library.
IMPORTANT:
In case the XRDI software has to be upgraded:
A new Application License file (ALF) is needed. Refer to CR Licensing - Service
Documentation, see referenced documents in section 3.
The status of the license X-ray Mammo Modality connection has to be checked.
When performing the upgrade, the old XRDI software version is removed and the
new one is installed. This procedure disables the X-ray Mammo Modality
connection license and the connection status in the examination pane does not
appear.
IMPORTANT:
If the UPS will not be used anymore, remove the UPS software (on Windows level)
before connecting the X-ray device to the COM port. Otherwise all COM ports are
blocked.
The default Examination Groups, Exposures and Exposure Groups have to be adapted
to the local requirements.
The image processing parameters can be adapted.
(1) Modify the Exposures in the Examination Group MAMMO according to the needs
of the customer, but only under certain conditions:
All clinical Mammography studies must be processed with the
CR Mammography image processing.
This means they must make
use of the CR
Mammography image
processing packages.
For more details, refer to section 7.2.5.6, Configure Musica Mammo Image
Processing
Do not use any entry under the Examination Group System Diagnosis
to build a new Examination Group or Exposure for Mammography. Those
are not linked with the dedicated CR Mammography image processing.
In case of a connection with IMPAX softcopy station ensure that the
Examination Group contains a string which will identify the study as a
Mammography study. By default MAMMO should be shown.
The following table shows the mapping of the NX Examination Tree DICOM
parameters with the IMPAX DICOM fields.
For non - Agfa PACS, check if View code and View modifier code
are used by the PACS for the hanging order. If so, synchronize them with
the PACS.
ACTION:
Go to Exam or Exam Tree Configuration to get an overview of the available
Examination Groups and Exposure Types.
The following list shows values for the parameters which shall not be changed for all
Clinical Mammography exposures:
Chapter Details
Package: Mammography*
Textbox position **
* 5 packages are available, refer to section 7.2.5.6.
** The choice of the textbox position is described below.
The following list shows values for the parameters which can be changed for all
Clinical Mammography exposures.
Mandatory exposures:
Name View Code View Modifier Image Implant
Code Laterality present
L-CC cranio-caudal None Left No
R-CC cranio-caudal None Right No
L-MLO medio-lateral oblique None Left No
R-MLO medio-lateral oblique None Right No
L-ML medio-lateral None Left No
R-ML medio-lateral None Right No
Optional exposures:
Name View Code View Modifier Image Implant
Code Laterality present
L-LM latero- medial None Left No
R-LM latero- medial None Right No
L-LMO latero- medial None Left No
oblique
R-LMO latero- medial None Right No
oblique
L-XCC exaggerated None Left No
cranio-caudal
R-XCC exaggerated None Right No
cranio-caudal
L-CV cranio-caudal Cleavage Left No
R-CV cranio-caudal Cleavage Right No
L-Implant Displ. cranio-caudal Implant Left Yes
or displaced
Implant Displ. L-CC
R-Implant Displ. cranio-caudal Implant Right Yes
or displaced
Implant Displ. R-CC
Implant L-CC cranio-caudal None Left Yes
Implant R-CC cranio-caudal None Right Yes
Implant L-ML medio-lateral None Left Yes
oblique
Implant R-MLO medio-lateral None Right Yes
oblique
The following two exposure groups need to be defined with the corresponding
exposure types:
(5) Configure the second exposure group according to the table above.
NOTE.
Some of the following configurations are only applicable with appropriate Features
licenses.
Described configurations are only recommendations; wishes of the customer must be
taken in account in any case.
For the UI of the Worklist pane in the NX workstation, no specific configuration task
with regard to CR Mammography application is necessary.
For details about the UI configuration of the Worklist pane, refer to NX Service
Manual, Chapter 4 – Installation and Configuration.
See referenced documents in section 3.
(3) Go to:
User Interface Examination Settings Configure Tools
in Examination
(4) In Tools & Annotations, define which button will be displayed and on which
position in the UI:
NOTE:
Only in case the X-ray parameters are entered manually:
Configure the Target/filter combination.
Ask the X-ray modality engineer for the available options.
NOTE:
The mammo modifier codes are divided in:
View modifier codes, and
Reason for imaging
Their label (Text) can be customized, but their content (Meaning) is fixed.
NOTE:
This feature is only available in case the licenses package NX Quality Assurance is
installed and the license NX_Repeat_Reject has been activated.
NOTE:
If no printer is available on site, a dummy printer has to be configured to display the
images in the proper size in the Editing pane (refer also to section 7.2.4.5).
Image (1) Go to:
Processing User Interface Editing Settings Configure Tools &
Tool Annotations
(2) In Tools & Annotations, look if the options are set according to table below.
Selection Palette
Delete
Select
Revert to Original
Annotations Palette
Left Marker Predefined Polygon Rectangle PVI Measure an
Text angle
Right Marker Arrow Freehand Rectangle EI Measure a
distance
Custom Markers Rectangle Perpendicular Calibrate a line Measure a leg
difference
Free text Circle Line Calibrate a Measure
circle scoliosis
Annotation Color Calibrate ERMF
Transformation Palette
Rotate Clockwise Show /Hide
Square Marker
Rotate Counter
Clockwise
Flip
View Palette
Zoom In Reset Zoom Enable/
Disable
Full Screen
Zoom Out Enable/
Disable Shutters
Zoom to ROI Enable/
Disable Split View
In case of connectivity to the X-ray modality (dedicated license) or in case the X-ray
exposure parameters will be entered manually, the default contents of the Film text box
have to be modified.
(1) Go to: Devices Printer Sheet/Image Text Box
(2) Create a default Mammo Text Box as recommended below:
Text box Height is 15 mm.
Text box Width is 160 mm.
Text box is composed of 3 lines to be divided with the proportions and filled
in with the parameters shown in the table below.
IMPORTANT:
The monitor calibration for a CR Mammography workstation can only be performed
on BARCO monitors.
7.3 Digitizer
8 Customer Training