Professional Documents
Culture Documents
VIEWED/PRINTED:
Calibration Only
Spectris Solaris
©
MEDRAD, Inc. 2007
Prior to use, you are responsible to confirm this copy's revision is the latest released.
This copy of 205254 Rev A is valid as of May-30-2007 at 11:06:00.
TABLE OF CONTENTS
05/31/2007 at 16:47:36.
1. SCOPE .................................................................. 3
VIEWED/PRINTED:
2. AUDIENCE ............................................................ 3
3. EQUIPMENT.......................................................... 3
4. DOCUMENTATION ............................................... 4
5. CALIBRATION ....................................................... 4
6. PRESSURE CHECK.............................................. 5
1. SCOPE
2. AUDIENCE
VIEWED/PRINTED:
3. EQUIPMENT
3.2 Leakage Test Box (Fixture No. 1-0-01), Bender Leakage Tester
or NeTech LGK 601 Electrical Leakage Analyzer – Equipment
Number 11153
4. DOCUMENTATION
5. CALIBRATION
5.3 Press Flow Rate. With a pencil point, touch all four corners of
the number 2 and verify that the key activates. Recalibrate
Touch Screen, if needed.
5.8 Verify that the arm lights stop flashing when the plungers are
05/31/2007 at 16:47:36.
moving.
VIEWED/PRINTED:
5.9 Verify that A and B syringe injects 20 ml and that the plungers
are now on 40 ml (A syringe) and 80 ml (B syringe) marks on
the syringe.
5.10 Go to the History Screen and open the latest file. Verify that the
Total Contrast = 20 ml, Total Flush = 20 ml and pressure limit
indication is NO.
6. PRESSURE CHECK
6.1 Fill both syringes with water and install both onto head. Install
the pressure gauge assembly onto syringe A and remove any
air bubbles.
6.3 Cycle CRU power and at safety screen activate the Swift
program.
6.5 Set DVM to DC millivolts, set pressure module to PSI and rotate
the “zero” adjustment until the DVM reads 0 mVDC. Close the
flow control valve.
6.6 Press “Start” on the PC screen. (Do not press the Enable key)
Using the A Forward Load key pressurize the system to the
following pressures on the digital gauge and press the Check for
Air key after achieving each pressure +/- 1 psi. Refer to the
Swift Op Manual, if needed.
• 75 PSI acceptable PC reading
6.7 Using the Load key, reverse plunger A to remove the pressure
VIEWED/PRINTED:
in the syringe.
6.8 Move the pressure assembly to the B syringe and redo Steps
12.1 to 12.7 substituting the B syringe for the A syringe.
6.9 Verify that the PC readout is within the acceptable ranges for all
pressure transducers for each reading.
6.10 Retract both pistons and remove the pressure gauge assembly.
Remove and discard the syringe.
7.1 Perform this step only if the Bender Box is required. Test
leakage according to 97057-00-SV-01.
7.2.1 Connect the CRU power cord to the leakage box. Inspect
power cord condition. Plug the leakage box into the power
source
7.2.2 Leakage box set-up: Set DMM to the lowest millivolt AC range
and connect to the points labeled Meter. With a test lead,
jumper Ground to Probe on the leakage box.
7.2.5 Set Line to Reverse. Leakage must be less then 100 mVAC.
Record value.
05/31/2007 at 16:47:36.
VIEWED/PRINTED:
7.2.7 Disconnect the test lead from the probe jack on the leakage box
and touch any metal part on the head with test lead, ensure the
leakage is less than 10 mVAC. Record the leakage value.
7.2.8 Remove power. Remove the leakage box power cord from the
outlet. Reconnect the CRU power cord and secure.
7.3.1 Inspect power cord condition. Connect the power cord to the
leakage box. Plug the leakage box into the appropriate power
source.
7.3.3 Plug unit into the NeTech Electrical Safety Analyzer and the
NeTech Electrical Safety Analyzer into the AC supply.
7.3.6 Set Outlet to NORMAL. Apply power. Press the Ground switch
to OPEN. Leakage must be less than 100 mVAC for voltages
less than 130VAC (300mVAC with AC voltage greater than
131VAC). Record the leakage value.
7.3.10 Connect a test lead to the Case Ground jack on the NeTech
Electrical Safety Analyzer. Touch any metal part on the head
with test lead. Ensure the leakage is less than 10 mVAC.
Record the leakage value.
7.3.11 Remove the leakage test box from unit. Reconnect unit to AC
supply.
05/31/2007 at 16:47:36.
VIEWED/PRINTED:
8. FINAL PREPARATION
8.1 Clean the area around the injector and sink of all spilled water.
Ensure syringes used during the PM are removed from the
injector and disposed properly. Discard all disposable materials
used while performing this PM. Place all linens used during
calibration in the soiled linen receptacle.
8.3 Record the software versions from the safety screen on the
PMP checklist.
8.7 Ensure that all covers are in place and all hardware is present
and secure.
This copy of 205254 Rev A is valid as of May-30-2007 at 11:06:00.