Professional Documents
Culture Documents
Revision 007
Service Manual
for Affinity Series Mammography Systems
Table of Contents
P/N 9-500-0255 i
Service Manual
ii P/N 9-500-0255
Service Manual
iv P/N 9-500-0255
Service Manual
List of Figures
List of Figures
P/N 9-500-0255 v
Service Manual
List of Figures
vi P/N 9-500-0255
Service Manual
List of Tables
List of Tables
Table B-16 Host Microprocessor Board (PN 1-003-0493) LED Indicators .................................... B-8
Table B-17 Host Microprocessor Board—C-arm Switch Multiple Function Indicators ................. B-8
Table B-18 HV Control Filament Board (PN 1-003-0485) LED Indicators .................................... B-8
Table B-19 Image Receptor Sensor Combinations ....................................................................... B-9
Table B-20 Affinity Series Fuses ................................................................................................ B-10
Table B-21 Microprocessor Board (PN 1-003-0493) Jumper Settings ......................................... B-12
Table B-22 Communications Interface Board (PN 1-003-0434) Jumper Settings ........................ B-12
Table B-23 C-Arm Safety Board (PN 1-003-0440) Jumper Settings ............................................. B-12
Table B-24 HV Control Filament Board (PN 1-003-0485) Jumper Settings ................................. B-13
Preface
Chapter 1: General Information: This chapter describes the general system, X-ray, electrical,
and mechanical safety precautions, and compliance information.
Chapter 4: Calibrations and Performance Tests: This chapter describes system calibration,
adjustment, and performance and compliance test procedures.
Chapter 7: Parts List: This chapter provides tabular listings of the replacement parts.
Appendix B: Technical References: This appendix identifies test points, LED indications,
fuses, jumpers, and accessory identification sensor combinations in the Affinity Series
System.
P/N 9-500-0255 ix
Service Manual
Preface
Warnings, Cautions, and Notes
WARNING! This warning alerts you to procedures that you must follow
precisely to avoid causing potentially serious or fatal injury to
yourself or others.
Warning: This warning alerts you to procedures that you must follow
precisely to avoid injury to yourself or others.
Caution: Cautions point out procedures that you must follow precisely to avoid
damage to equipment, loss of data, or corruption of files in software
applications.
Note… Notes indicate important information that must be followed to ensure the
proper operation of the system.
x P/N 9-500-0255
Service Manual
Preface
Acronym and Abbreviation List
Acronym Definition
21 CFR FDA Code of Federal Regulations, Title 21
A Amps
ACR MAP American College of Radiology Mammography Accreditation Program
ACR/CDC American College of Radiology/Center for Disease Control
AEC Automatic Exposure Control
BR-12 Breast Equivalent Material
CF Compression Force
cm centimeter
DMM Digital Multimeter
EMI Electromagnetic Interference
EMO Switch Emergency OFF Switch
ESD Electro-static Discharge
F.A.S.T. Fully Automatic Self-Adjusting Tilt Paddle
HV High Voltage
HTC High Transmission Cellular Grid
HVL Beam Quality Half-Value Layer
I/O Input/Output
IR Image Receptor
IRSD Image Receptor Support Device
lb Pound(s)
LVPS Low Voltage Power Supply
mA milli-Amps
Mag Magnification Mode
max maximum
mGy milli-Gray
min minimum
mm millimeters
Mo Molybdenum
MQSA Mammography Quality Standards Act
mrad milli-Radiation Absorbed Dose
N Newtons
OD Optical Density
PCB Printed Circuit Board
PMMA Polymethyl Methacrylate
PN Part Number
PWM Pulse-Width-Modulated
Rh or Rho Rhodium
SID Source to Image Distance
V Volts
VAC Volts, Alternating Current
VDC Volts, Direct Current
VTA Vertical Travel Assembly
P/N 9-500-0255 xi
Service Manual
Preface
Acronym and Abbreviation List
1.0 Introduction
The Affinity is a dedicated mammography system that features automatic and manual exposure
modes, a compact Control Panel, and optional film labeling capabilities.
Table 1-1: Required Tools and Equipment for Installation and Maintenance
• Standard Tool Set
• Scale/Force Gauge—1 - 50 lb (min)
• Tool, Field Service Belt Adjustment (PN 9-061-0107)
• Digital Multimeter (DMM)
• Oscilloscope—100 MHz (min) 2 channel
• High-Voltage Adapter Cable (PN 2-425-3015)
• High-Voltage Divider—Ratio of 10,000:1 or 100,000:1 with output impedance of 1 kW
• mAs Meter—dual range
• Densitometer
• Light Meter—UDT Instruments - Model 351
• Radiation Meter with probe (Victoreen Model 600 or equivalent)
• ACR Mammographic Phantom (ACR MAP) (Nuclear Associates #156 or equivalent)
• Aluminum Aperture for Light meter with 1.0 mm aperture
• Aluminum Filter Pack—Type 1100 (99.9% pure)
• Image Receptor Phantom Tool—Cassette Simulator (PN 9-060-0403)
• Lead Shield/Blocker—Approximately 4.0 x 4.0 cm
• Light Field Alignment Bars (PN 9-060-0173)
• Light/X-ray Field Template (PN 3-405-8010)
• Loctite, 242 Blue (PN 2-580-0506)
• PMMA Acrylic (or BR-12) - 0.5 cm thick (1 ea)
• PMMA Acrylic (or BR-12) - 1.0 cm thick (1 ea)
• PMMA Acrylic (or BR-12) - 2.0 cm thick (4 ea)
• Resolution Line Pair Phantom
2.0 Safety
This section describes electrical, mechanical, and radiation safety, as well as precautions
concerning static electricity and magnetic media storage. The equipment complies with IEC
60601 (General, Collateral and applicable Particular Standards), UL 2601 and CSA
22.2601.1.
Note… The Affinity does not contain any mercury or latex in its construction.
WARNING! LETHAL voltages are present within the system. Use extreme
caution to avoid contacting, directly or indirectly (through
tools), any connector pins, terminals, or test points. Remove
all jewelry before working on the system, and avoid wearing
loose fitting clothing.
Warning: Never perform service alone. Only service this equipment in the
company of someone who is capable of rendering aid should an
accident occur.
WARNING! To ensure the isolation for the system, attach only Hologic
accessories or options to the system.
1.5m
2.5 Interlocks
In addition to the Emergency Off Switches, the Affinity Series has several safety interlocks,
including:
• C-arm Movement Interlock - This interlock immediately stops vertical travel and
disables all C-arm vertical movements (except C-arm UP) if the C-arm encounters an
obstruction during travel (until the obstruction is cleared). C-arm movement is also
inhibited when compression force exceeds 50 N (12 lb).
• Automatic Compression Release Interlock - This interlock disables the compression
release function, including automatic compression release (if enabled) when a
Localization Paddle is installed.
• Early Release Interlock - This interlock immediately terminates the exposure if the X-ray
Button is released prior to the end of the exposure.
• Cassette Sense Interlock - This interlock prevents X-ray exposure if the system does not
detect a cassette in the Bucky, or if the cassette is not ejected after an exposure.
• Aperture Interlock - This function detects the installed aperture and image receptor and
prevents X-ray exposure if the combination is not compatible.
• Mirror/Filter Interlock - This interlock prevents X-ray exposure when the Light Field
Mirror and/or the Filter is not positioned correctly.
Note… The Hologic Bucky devices (18 x 24 cm and 24 x 30 cm) are certified
components that do not have to be reported on form FDA 2579 (Ref BRH:
Doc MA3499).
Item Manufacturer
X-ray Tube Varian (IM-113R)
Beam Limiting Device Hologic
High-Voltage Generator Hologic
X-Ray Control Hologic
Image Receptor Support Device Hologic
Bucky 18 x 24 cm Hologic
Bucky 24 x 30 cm Hologic
Note… If there is a discrepancy between the contents and the packing list or sales
order, contact Hologic immediately. If it is necessary to re-pack any items
for future installation, use the original packaging materials.
4. Inspect all items for damage, then safely store them near the exam site.
Caution: To prevent damage, do not store the Operator Radiation Shield flat. Avoid
damaging the Operator Radiation Shield by impact or scratching.
Warning: Never pull on the C-arm as the system is top heavy. To prevent
injury to personnel and/or damage to equipment, take care when
uncrating the system.
5. While still in the loading area, carefully move the pallet into an upright position.
6. Remove the top collar securing the system to the pallet.
7. Using the installed dolly, carefully roll the system off the pallet.
8. Transport the system, on the installed dolly, from the loading area to the exam area for
installation.
9. Remove the 4 bolts securing the system to the dolly.
10. Carefully maneuver the system off the back end of the dolly, slide the dolly out and
position the system in the area of installation.
11. Return the dolly to Hologic according to the shipping instructions posted on the dolly.
Where:
R = Ground Impedance
VNL = Voltage at no load
VFL = Voltage at full load (32 kV, 100 mA)
IMAX.L = Current at full load
Ground Impedance must be no more than:
After determining the input voltage range, verify the system’s isolation transformer is
correctly set (reconfigure as required).
The following describes how to access the input power assembly chassis and reconfigure the
isolation transformer taps to match the previously measured source voltage.
1. Turn off the Circuit Breaker on the rear of the Gantry (down position).
2. Remove the Gantry Upper Rear panel as described in Chapter 6, Section 2.1.2, Upper
Rear Gantry Panel, p. 6-4. Set the cover aside.
3. Remove the Rear Connector panel and shroud as described in Chapter 6, Section 2.1.3,
Rear Connector Panel, p. 6-5. Pull the panel slightly away from the system, then let it
hang by the attached wiring harness.
4. To gain access to the input power assembly chassis, remove the Lower Rear panel as
described in Chapter 6, Section 2.1.4, Lower Rear Panel/Power Distribution Assembly, p.
6-6, Steps 1 through 3. The isolation transformer is mounted on the lower shelf of the
input power assembly chassis.
5. Configure the input wiring to tap the transformer so that it matches the previously
measured source voltage.
STOP
Connections are made inside the removable power supply distribution assembly,
(Figure 2-4 #1), located on the input power assembly chassis:
1. Locate the removable power distribution assembly. For removal instructions, refer to
Chapter 6, Section 2.1.3, Rear Connector Panel, p. 6-5.
2. Connect the remote X-ray on and Power on cables to the terminal board as shown
below.
3. Complete the remote cables and light installation following local guidelines.
4. Reinstall the Gantry lower rear, upper rear, and rear connector panels. Do not reinstall
the rear connector panel shroud at this time.
Warning: For those systems with the Control Panel mounted to the Gantry,
the Operator Radiation Shield is required. Install the Operator
Radiation Shield after completing the initial startup procedures.
Figure 2-7: Rear Connector Panel (Shroud removed) and Lower Rear Panel Connections
Note… Do not reinstall left and right side panels at this time.
Note… The X-ray Footswitch is delivered already mounted on the X-ray Footswitch
mounting bracket.
1. Secure the X-ray Footswitch mounting bracket to the Gantry base plate using
two screws (supplied) on the right side of the Gantry.
Note… When the Gantry right side cover is installed, the X-ray Footswitch is
positioned below the Control Panel).
2. Connect the X-ray Footswitch cable to the X-ray connector (Figure 2-7, #5).
a. Align the key in the footswitch connector to the keyhole in the receptacle.
b. Push the connector straight in until it stops.
3. Locate DIP switch (S2) at the lower right side of the Host Microprocessor board
(Gantry right-side panel removed). Enable the X-ray Footswitch by setting S2
switch 3 to OFF.
2. Position the footswitch in the desired location on the floor below the C-arm.
Note… The film labeling device (AutoFilm I.D. or Printer) must be installed so that it
is situated at least 1.5 meters (approximately 5 feet) from the patient.
AutoFilm I.D.
Refer to the AutoFilm I.D. User’s Guide for installation and operating procedures.
Rapid ID
Refer to the Rapid ID User’s Guide for installation and operating procedures.
Label Printer
The Label Printer prints a self-adhesive label of the exposure information. When
connected and on-line, the printer immediately prints a label after each exposure.
1. Insert the one end of the printer cable into the receptacle on the rear of the
Printer.
2. Connect the other end of the printer cable to the printer receptacle (Figure 2-7,
#2) on the rear of the Gantry.
3. Plug the printer power cord into the power receptacle on the rear of the printer,
then plug the opposite end into the printer power outlet (Figure 2-7, #4) on the
rear of the Gantry.
Notes… Always perform the pre-startup Inspections before turning on the system.
Always correct discrepancies before turning the system on and using it
clinically.
1. If a film labeling device is installed, turn it on.
2. Turn on the Circuit Breaker on the rear of the Gantry (up position). The back light of the
display indicates that power is applied.
3. The Control Panel displays a message indicating that internal tests are being performed.
4. Ensure that the Control Panel (Figure 3-1, #9) displays the Run Mode screen after startup.
5. If the AutoFilm I.D. or Rapid ID is installed, ensure that the Patient Information Screen is
displayed. Refer to the respective User’s Guide.
Figure 3-2: Location of the Emergency Off Switches and the Circuit Breaker
2.3.2 C-arm
Item # Button
Tests
(Figure 3-3) Description
1 Filter Indicator • Confirm that the Mo filter light is on. (The system always
(if equipped) comes up with Moly filter chosen.)
• Use the Control Panel to change the filter to Rh. Move the
highlight with the down arrow key until Filter is selected,
then press the right arrow key to change to the Rh Filter.
• Ensure that the Rh light is now on.
• Change the filter back to Moly.
2 Light Field Lamp • Press and release the Light Field Lamp Button.
• Verify the Light Field Lamp illuminates and then turns off
after approximately 30 sec.
3 Compression Down • Verify the Light Field Lamp illuminates, and the
Button compression device moves down while the button is held.
• Verify the compression thickness readout changes
accordingly.
• Ensure the compression device stops when the button is
released and the compression brake is active (handwheel-
up has resistance).
4 C-arm Down Button • Verify the C-arm moves down while the button is held and
it stops when the button is released.
5 C-arm Rotation • Verify the C-arm can be rotated while the button is held,
Release Button and the C-arm brake prevents rotation when the button is
released.
• Verify that when the C-arm rotates, the rotation readouts
on the Gantry change accordingly. Verify that the C-arm
rotation angle display shows zero degrees of rotation
when the C-arm is in the “CC” position.
6 C-arm Up Button • Verify the C-arm moves up while the button is held and it
stops when the button is released.
1 2 3
Item # Button
Tests
(Figure 3-4) Description
1 C-arm Up Key • Verify the C-arm moves up while the button is held
and stops upon release.
2 C-arm Down Key • Verify the C-arm moves down while the button is
held and stops upon release.
3 C-arm Rotation • Verify the C-arm can be rotated while the button is
Release Key held and the brake prevents rotation when the
button is released.
Footswitch Controls • Push the C-arm up footswitch and verify the C-arm
moves up and stops when the footswitch is
released.
• Push the C-arm down footswitch and verify the C-
arm goes down and stops.
• Verify that the Compression Up and Down
footswitches function correctly.
The user uses the User Defaults Screen to set the defaults for:
• Compression Release
• Compression Force
• Cassette Sense
• Film ID
• Force Units
• Date
• Time
The choices made in the User Defaults Screen are displayed after startup.
The user uses the Exposure Defaults Screen to set system defaults for exposure techniques.
The choices made in the Exposure Default Screen display in the Run Mode Exposure settings
after startup.
Press Reset + Right Arrow at the same time to select the User Default Screen from the Run
Mode Screen.
To select the Exposure Default Screen from the Run Mode Screen, press Reset + Right Arrow
twice.
To go back to the Run Mode Screen from the Exposure Default Screen, press Reset + Left
Arrow twice; from the User Default Screen, press Reset + Left Arrow once.
Note… Access to the Calibration Mode (the Debug Screen and the four Calibration
Screens) requires the service engineer to configure a DIP switch on the Host
Microprocessor Board (Chapter 4, Section 2.1,Entering Calibration Mode,
p. 4-2).
The user uses the Calibration Screen #1 (below) to make changes to system calibration
values such as Filament Current.
The user uses the Calibration Screen #2 to set up and calibrate the Automatic Exposure
Control system, including the AEC Gain, AEC Offset, and AEC Density. When Cal Screen #2
is selected, the system stays in Auto-Time Mode.
The user uses the calibration Screen #3 (below) to calibrate electro-mechanical systems and
perform the backup timer test. When setting CF CAL and SPECIAL commands, pressing Reset
+ Comp Rel sends the selected command.
The user uses the Calibration Screen #4 to set the X-ray tube type, perform AEC Gain
Calibration, set the wait time between exposures when in service mode, and set the Auto
Filter Threshold.
Note… When the Cassette Sense is set to ON, the system remains in Standby until a
new cassette is installed.
Note… At kVs above 32, maximum mAs is restricted to less than 500 mAs for large
focal spot and 150 mAs for small focal spot.
• mAs—Large FS, 3 to 500 mAs; Small FS, 3 to 150 mAs; manual exposures require
that the user set a fixed mAs for exposure timing.
• Density—+5 to Normal to -5 (not available in Manual or Mag Manual modes)
• Screen Film—Film 1, Film 2, Film 3
• Focus—Large or Small Focal Spot
• Auto-kV Window—When the exposure mode is set to Auto-kV or Auto-Filter the
system determines the exposure factors necessary to make films at the optimum
Optical Density (OD), with ideal contrast. To do this, the user must set a timing
“window” that determines when to terminate the exposure. The timing window
choices for Large FS are 125, 165, or 200 mAs and the choices for Small FS are 38,
50, or 60 mAs. The Auto timing window (based on breast thickness) should be
selected by those users who are required to meet the 2 mGy (200 mrad) maximum
average glandular dose requirement.
• Default Exposure Mode—Mode that system is set to at startup.
• Default Mag Mode—Displays default mode when a magnification device is
installed.
• Filter—Upon start up, Filter is always set to Molybdenum; user can change this
setting to Rhodium.
1.0 Introduction
This chapter contains the Affinity Series calibration, adjustment, and performance and
compliance test procedures.
Caution: X-Ray exposures made in rapid succession may exceed the thermal
capacity of the system. During periods of numerous X-ray exposures, such
as during service calibration or physicist accreditation, please wait a
minimum of 40 seconds between exposures to allow adequate cooling of
the X-ray tube and high voltage electronics.
Note… When servicing the Affinity, if a procedure instructs you to remove the
Gantry right-side cover or the tubehead covers, do not reinstall these covers
until all required procedures (for example, calibrations, adjustments, tests,
remove and replace, etc.) are completed.
The fields for each calibration screen are described in Table 4-2.
Caution: Always return the system to the user mode (S2 switch 1 = OFF) after
performing calibrations to prevent inadvertent changes to system
calibrations.
5. Using a coaxial cable (#1) and a BNC to banana plug splitter (#4), connect a Digital
Multimeter (DMM, #2) to the anode terminal (#3) of the HV divider (#6). Position the
meter where it can be read from behind the protective Operator Radiation Shield.
6. Connect the HV divider’s ground wire (#8) to an unpainted metal surface on Gantry.
7. Connect one end of the HV Adapter cable (#10) into the voltage divider receptacle
(#11), and the opposite end into the HV test well (#12) on the HV Generator Assembly.
8. Turn the system on.
Note… Skip Step 9 when performing the Tube Voltage Calibration as part of the
initial installation procedures.
9. Perform the kV Control Offset Calibration Procedures, Section 9.0, p. 4-64
10. From the Run Mode, set the system for a Manual mode exposure at 30 kV, 150 mAs.
Monitor the voltmeter from behind the Operator Radiation Shield (if equipped), then
make the exposure.
11. The voltage reading on the voltmeter for 30 kV should be 3.0 VDC (±0.06 VDC using
10,000:1 voltage divider), or 30 VDC (±0.6 VDC using 1,000:1 voltage divider).
12. If the voltage reading is not within the specification above, adjust R41 on the HV Control
Filament Board, which is just below the HV Inverter Board on the HV Generator
Assembly. See Figure 4-7 and Figure 4-8. Using a screwdriver, turn R41 clockwise to
increase kV, or counter-clockwise to decrease kV.
13. Repeat Steps 9, 10, and 11 until the voltage reading at 30 kV is +2% of the VDC
specified in Step 11.
14. Set the system for a Manual mode exposure at 39 kV, 80 mAs, Rh filter. Monitor the
voltmeter from behind the Operator Radiation Shield, then make the exposure.
15. The voltage reading on the voltmeter for 39 kV should be 3.9 VDC (±0.078 VDC using
10,000:1 voltage divider), or 39 VDC (±0.78 VDC using 1,000:1 voltage divider).
16. If the voltage reading is not within the specification above, readjust R41 on the HV
Control Filament Board.
17. Repeat the exposure and test the voltage reading. Repeat Steps 14, 15, and 16 until the
voltage reading at 39 kV is within the ±2% specification.
18. Test the voltage at 20 kV to verify that it remained within the ±2% specification. If
necessary continue to make adjustments at 20 kV, then at 39 kV until calibration at both
levels is achieved.
19. Select intermediate kV levels and make an exposure at each setting. Confirm that the
voltage readings on the meter are within 2% of the kV setting. If they are not, make small
adjustments at 20 kV and 39 kV to “dial” in the intermediate kV levels while maintaining
the ±2% at all kVs.
20. For initial setup and calibration, skip this step. If all service procedures have been
completed, return the system to the user mode (S2 switch 1 = OFF), and reinstall all
previously removed covers and panels.
Note… This calibration must be performed using Manual Exposure mode only.
1. Turn the system off.
2. Remove the Gantry’s right side panel (to access the Host Microprocessor Board).
3. Connect an oscilloscope to the mA Sense Test Point on the Host Microprocessor Board.
4. On the Host Microprocessor Board, set the system for the service mode (S2 switch 1 =
ON).
5. Turn the system on.
6. After the Run Mode screen displays, highlight the Filter field, and select the Mo filter.
7. From the Run Mode screen, access CAL SCREEN 1 by pressing Reset + Right Arrow four
times.
8. Highlight the Auto Cal field, then press a Change key until the field reads: ON.
9. Highlight and change the appropriate fields to set the system for a Manual mode
exposure at 20 kV, 40 mAs, Large Focal Spot.
10. Make an exposure. The system automatically adjusts the starting filament values. Wait
60 seconds, and make a second exposure. Wait 60 seconds, and make a third exposure.
After each exposure, examine the waveform on the oscilloscope for undershoot or
overshoot (Figure 4-9). Verify that the waveform appears stable and is at the correct
amplitude for the set kV as shown in Table 4-3).
Notes… To fully calibrate the system, you must make 3 exposures (at 40 second
intervals between shots).
The mA Target (“T”) value and the mA Start (“ST”) value displayed at the
bottom of CAL SCREEN 1. The difference between the start value and the
target value should be less than 10.
11. Repeat steps 11 and 12 for each kV station using the Large focal spot and Mo filter, then
perform this procedure for each kV station using the Small focal spot, Mo filter, and 12
mAs. Record the values for each kV station on Table 4-3.
Note… During the next step, the system does not take X-rays and attempts a 7.0 sec
exposure.
13. Make an exposure.
14. When the exposure terminates, the message area along the lower part of the screen
indicates the exposure duration (when the timer terminated the exposure), and the status
of the test.
15. Verify that the test “PASSED”. If the indication is “FAILED”, immediately take corrective
action to repair the system. Do not use the system clinically.
16. The Timer Test field automatically resets to OFF.
17. Remove all test equipment, do not remove jumpers JP5 and JP6.
18. Return the system to the user mode (S2 switch 1 = OFF) and reinstall previously removed
panels if this is not the initial installation
Note… This procedure must be performed for each film type used with the system.
Before performing the AEC Calibration procedure, be sure you have
completed the Tube Voltage Calibration and the Filament Calibration in
Section 4.0.
AEC system performance is specified at the kVs that are normally used clinically. Table 4-5 is
a guideline for typical clinical values.
Table 4-5: Typical Clinical Values for Breast Composition and Thickness
2.0 cm thickness imaging should not be used above 26 kV; 6.0 cm thickness should not be
used below 26 kV.
Notes… When calibrating the AEC system, always use the same Bucky device and
the same cassette to ensure consistent results.
For optimum calibration results use the AEC Worksheets provided at the
end of this procedure (Table 4-6 through Table 4-12). Record the AEC
performance values on the worksheets when the AEC system is calibrated.
Note… The replacement resistor value typically falls within the 1.0 kΩ to 1.7 kΩ
range.
8. Replace R1 on the AEC Detect board with a 1/4 Watt, metal film, 1% resistor with the
value determined in Step 7 above.
9. Reverse Steps 2 through 6 to install the AEC Detector Assembly.
10. Perform the procedure: Section 5.1, AEC Detector Gain Calibration.
Note… For large focal spot calibration with an HTC Bucky, the system must be in 100
mA mode. Ensure the HTC Threshold is low enough to remain in this mode.
2. Set system to CAL SCREEN 3, set the HTC C-THK threshold for 1.0 cm.
3. Set system to CAL SCREEN 2.
4. Install the 18 x 24 cm Bucky onto the IRSD and insert a loaded cassette. Always use the
same cassette and Bucky for all calibration steps.
5. Set the system for an Auto-Time exposure at 25 kV, Large focus, Mo filter.
6. Set the Film field to the screen-film combination loaded in the cassette.
7. Place 3.0 cm of BR-12 or PMMA acrylic (BEM) on the Bucky, centered laterally, and
extending approximately 1.0 cm over the chest wall edge.
8. Place the AEC detector in the position closest to the chest wall.
9. Make an exposure, develop the film, then measure the OD.
Note… Always measure the OD with the lower edge of the film flush with the front
edge of the Densitometer and centered side-to-side (approximately 5.0 cm
into film plane).
10. Make Gain adjustments to calibrate the OD to the value requested by the customer.
11. Increase the Gain value if the density is too high. To maintain mAs requirements (40 - 80
mAs), it may be necessary to increase or decrease the amount of BR-12 or PMMA used.
12. Make another exposure, develop the film, then measure the OD.
13. Repeat Steps 10—12 as necessary, to obtain the mean OD requirement (±0.15 OD).
14. Replace the 3.0 cm BEM with the 6.0 cm BEM.
15. Using the same techniques as in Steps 10—12, make an exposure, develop the film, then
measure the OD.
16. Make Offset adjustments to calibrate the OD to the value requested by the customer.
17. Increase the Offset value if the density is too high. To maintain mAs requirements (240 - 350
mAs), it may be necessary to increase or decrease the amount of BR-12 or PMMA used.
18. Make another exposure, develop the film, then measure the OD.
19. Repeat Steps 16—18 as necessary, to obtain the mean OD requirement (±0.15 OD).
20. Replace the 6.0 cm BEM with the 3.0 cm BEM, make an exposure, develop the film,
then measure the OD.
21. Ensure the OD reading is within ±0.15 OD of the exposure at 6.0 cm.
22. If necessary, repeat the entire procedure to bring both the 3.0 cm and the 6.0 cm
exposures within the specified requirements.
23. Change the Gain, Offset and M-Density (if required) values for each of the remaining
large focal kV stations to the Gain, Offset, and M-Density values obtained in this
procedure.
Note… When calibrating with high-speed contrast film, each kV station needs
verification and input of the Gain, Offset, and AEC Density.
5. If the customer uses more than one film type, repeat this procedure for each one.
Note… For large focal spot calibration with an HTC Bucky, the system must be in
100 mA mode. Ensure the HTC Threshold is low enough to remain in this
mode.
2. Set system to CAL SCREEN 3, set the HTC C-THK threshold for 1.0 cm.
3. Set system to CAL SCREEN 2.
4. Install the 18 x 24 cm Bucky onto the IRSD and insert a loaded cassette. Always
use the same cassette and Bucky for all calibration steps.
5. Set the system for an Auto-Time exposure at 28 kV, Large focal spot, Rh filter. Set
the Film field to the screen-film combination loaded in the cassette.
6. Place 3.5 cm of BR-12 or PMMA (BEM) on the Bucky, centered laterally, and
extending approximately 1.0 cm over the chest wall edge.
7. Place the AEC detector in the position closest to the chest wall. Make an
exposure, develop the film, then measure the OD.
Note… Always measure the OD with the lower edge of the film flush with the front
edge of the Densitometer and centered side-to-side (approximately 5.0 cm
into film plane).
8. Make Gain adjustments to calibrate the OD to the value requested by the
customer. Decrease the Gain value if the density is too high. To maintain mAs
requirements (40 - 80 mAs), it may be necessary to increase or decrease the
amount of BR-12 or PMMA used.
9. Make another exposure, develop the film, then measure the OD.
10. Repeat Steps 8 and 9 as needed, to obtain the mean OD requirement (±0.15 OD).
5.5.2 Offset Adjustment
1. Replace the 3.5 cm absorber BEM the 7.0 cm BEM. Using the same techniques
as in Steps 8 and 9 of the previous procedure, make an exposure, develop the
film, then measure the OD.
2. Make Offset adjustments to calibrate the OD to the value requested by the
customer. Decrease the Offset value if the density is too high. To maintain mAs
requirements (240 - 350 mAs), it may be necessary to increase or decrease the
amount of BR-12 or PMMA used.
3. Make another exposure, develop the film, then measure the OD.
4. Repeat Steps 2 and 3 as necessary, to obtain the mean OD requirement (±0.15
OD).
5. Replace the 7.0 cm BEM with the 3.5 cm BEM, make an exposure, develop the
film, then measure the OD. Ensure the OD reading is within ±0.15 OD of the
exposure at 7.0 cm. If necessary, repeat the entire procedure to bring both the
3.5 cm and the 7.0 cm exposures within the specified requirements.
6. Change the Gain, Offset and M-Density (if required) values for each of the
remaining large focus kV stations to the Gain, Offset, and M-Density values
obtained in the Gain Adjustment and Offset Adjustment procedures.
Note… When calibrating with high-speed contrast film, each kV station needs
verification and input of the Gain, Offset, and AEC Density.
Note… It may be necessary to use BEM thicknesses greater than 6.0 cm to achieve
the desired mAs when calibrating AEC using the Rhodium filter at higher
kVs.
Note… Before starting this procedure, select the small focal spot, then set the Gain
and Offset values to the values used at 25 kV for the Large Focal spot.
1. Remove the Bucky device and install the Magnification Table or cassette holder
and Mag Platform. Always use the same cassette for all AEC calibrations.
2. On the Run Mode screen set the system for an exposure at 25 kV, Small focal
spot, Mo filter, Mag Auto-Time.
3. Place 2.0 cm of BR-12 or PMMA on the breast platform, centered laterally, and
extending 1.0 cm over the chest wall edge.
4. Place the AEC detector in position #1 (closest to the chest wall). Make an
exposure, develop the film, and measure the OD.
Note… Always measure the OD with the lower edge of the film flush with the front
edge of the Densitometer and centered side-to-side (approximately 5.0 cm
into film plane).
5. Make Gain adjustments as necessary to calibrate the OD target value. Increase
the Gain value if the density is too high. To maintain mAs requirements (6 - 14
mAs), it is necessary to increase or decrease the amount of BR-12 or PMMA
used. See Table 4-8 and Table 4-9.
6. Make another exposure, develop the film, then measure the OD.
7. Repeat Steps 5 and 6 as necessary, to obtain the mean OD requirement (±0.15
OD).
5.10.2 Offset Adjustment
1. Replace the BEM used for the Gain adjustments with the 5.0 cm BEM. Using the
same techniques as in Steps 5 and 6 of the previous procedure, make an
exposure, develop the film, then measure the OD.
2. Make Offset adjustments to calibrate the OD to the value requested by the
customer. Increase the Offset value if the density is too high. To maintain mAs
requirements (65 - 130 mAs), it may be necessary to increase or decrease the
amount of BEM used.
3. Make another exposure, develop the film, then measure the OD.
4. Repeat Steps 2 and 3 as necessary, to obtain the target OD requirement (±0.15
OD).
5. Replace the 5.0 cm BEM with the 2.0 cm BEM, make an exposure, develop the
film, then measure the OD. Ensure the OD reading is within ±0.15 OD of the
exposure at 5.0 cm. See Table 4-8 and Table 4-9.
6. If necessary, repeat the entire procedure to bring both the Gain and the Offset
exposures within the specified requirements.
7. Change the Gain, Offset, and M-Density (if required) values for each of the
remaining small focal spot kV stations to the Gain, Offset, and M-Density values
obtained in this procedure.
Note… Before starting this procedure, on CAL SCREEN 2, select the small focus,
then set the Gain and Offset values to the values used at 26 kV for the Large
Focal spot.
1. Set the system for an exposure at 28 kV, Small focal spot, Rh filter, Mag Auto-
Time.
2. Place 2.5 cm of BR-12 or PMMA on the Magnification Table or cassette holder,
centered laterally, and extending 1.0 cm over the chest wall edge.
3. Place the AEC detector in the position closest to the chest wall. Make an
exposure, develop the film, then measure the OD.
Note… Always measure the OD with the lower edge of the film flush with the front
edge of the Densitometer and centered side-to-side (approximately 5.0 cm
into film plane).
4. Make Gain adjustments to calibrate the OD to the value requested by the
customer. Decrease the Gain value if the density is too high. To maintain mAs
requirements (6 - 14 mAs), it may be necessary to increase or decrease the
amount of BR-12 or PMMA (BEM) used.
5. Make another exposure, develop the film, then measure the OD.
6. Repeat Steps 4 and 5 as necessary, to obtain the target OD requirement (±0.15
OD).
HTC FACTOR
Reduced
kV Thickness mAs OD HTC Factor
mA
26 30
18 x 24 cm Bucky 24 x 30 cm Bucky
Thickness mAs OD Thickness mAs OD
4.0 cm 4.0 cm
Final Large Bucky Factor _________
Note… For this adjustment the system must be in 100 mA mode. Ensure that the
Compression Thickness Threshold is 1.0 cm at the start of this procedure.
1. Set the system to CAL SCREEN 3, and set the HTC C-THK threshold for 1.0 cm.
2. Install the 18 x 24 cm HTC Bucky onto the IRSD and insert a loaded cassette. Always use
the same cassette and Bucky for all calibration steps.
3. Set the system for an Auto-Time exposure at 25 kV, Large focus, Mo filter. Set the Film
field to the screen-film combination loaded in the cassette.
4. Place the amount of BR-12 or PMMA acrylic that was used in the final 25 kV, low
attenuation exposure from Table 4-6 on the Bucky, centered laterally and extending 1.0
cm over the chest wall edge. Use motorized compression to lower the compression
paddle onto the BEM material.
5. Insert a loaded cassette into the Bucky, then make a 25 kV Auto-Time exposure. Record
the mAs value (as point “A”) on the semi log worksheet (Table 4-12).
6. Release compression and place the amount of BR-12 or PMMA acrylic that was used in
the final 25 kV, high attenuation exposure from Table 4-6 on the Bucky. Reapply
compression, then make a second exposure. Record this mAs value (as point “B”) on the
semi log worksheet.
7. Draw a line between points “A” and “B” on the semi log. Follow the left side of the graph
and locate the 80 mAs point. Draw a straight line and intercept the line between the 2
points.
8. Draw a vertical line from that point to the bottom of the graph and note the compression
thickness value indicated. This is the compression thickness threshold value.
9. On CAL SCREEN 3, highlight HTC C-THK and change the displayed value to the value
calculated on the graph.
500
400
300
200
100
90
80
70
60
50
40
30
20
mAs↑ 10
Thickness→ 0 1cm 2cm 3cm 4cm 5cm 6cm
10. Place a 0.2 mm aluminum filter (99.9% pure aluminum) on the surface of the
compression paddle, directly beneath the X-ray tube port (Figure 4-11).
11. Using the same technique factors as in Step 8, make another exposure and record the
milliroentgen reading on the semi-log.
12. Repeat Steps 10 and 11 with additional 0.1 mm sheets of aluminum between the X-ray
tube port and the radiation probe.
13. Record the milliroentgen reading on the semi-log after each exposure until the reading is
less than one-half the original step #9 exposure reading (without aluminum).
14. Draw a line on the semi-log through the plotted mR readings. It may be necessary to draw
the line to achieve a “best fit” path (in cases where the plotted readings are not linear).
15. Determine the exact half-value of the step #9 mR reading and record this on the semi-log.
16. Draw a horizontal line, starting from this plot, across the semi-log until it intersects with
the line plotted in Step 14.
17. Draw a vertical line through the intersection to the bottom of the semi-log. This is the
half-value layer.
18. Verify that the half-value layer is greater than 0.33 mm AL, and less than 0.42 mm AL
(Molybdenum filter).
19. Select the Rhodium (Rh) filter, then repeat Steps 9 through 16 to obtain the half-value
layer for the Rhodium filter. Verify that the half-value layer is greater 0.33 mm AL, and
less than 0.49 mm AL (Rhodium filter).
If the test fails, take corrective action to repair the system.
11. Continue to randomly change the technique settings for kV and mAs, then return to 25
kV and 60 mAs and Make an exposure.
12. Record the milliroentgen reading on the worksheet until ten individual readings have
been recorded.
13. Compute and record the average milliroentgen reading.
14. Subtract each actual reading from the average, and square each difference. Add the
squares and divide the sum by nine. Then take the square root of the result.
15. Divide the number calculated in Step 13 by the average milliroentgen reading to obtain
the coefficient of variation. The coefficient of variation must be less than 0.05.
16. Leave the configuration on the image receptor support the same. Change the technique
settings to 25 kV and 20 mAs.
17. Make an exposure and record the milliroentgen reading and the mAs value on the
screen on the Linearity Worksheet Figure 4-14, p. 4-43.
18. Repeat Step 16, changing the mAs to 60, 100, 150, 200, 250, 300, 400, and 500.
Record milliroentgens and mAs at each setting.
19. Divide each milliroentgen reading by its corresponding mAs value.
20. For each pair of successive tests, calculate the difference between each corresponding
Step 19 result.
21. For each pair of successive tests, calculate the sum of each Step 19 result.
22. Divide each Step 20 difference value by each Step 21 sum value. If the result for any pair
exceeds 0.10, the test is considered failed.
23. Repeat Steps 16 through 22 using the Small Focal Spot.
24. If the system fails any part of the above test, first redo all calculations, then repeat the
tests. Failure to meet these performance standards indicates the need to verify the X-ray
tube’s kV, mA, and mR outputs.
Note… Use the same Bucky device, cassette, and film type for all exposures.
1. With the C-arm to 0°, on the Run Mode screen, set the system for an Auto-Time, 25 kV
exposure, Large spot, Normal density. Turn off “AutoCompression Release.”
2. Select the Film Type being used.
3. Place a loaded cassette in the Bucky device, then place 2.0 cm of BR-12 (or PMMA
acrylic) on the Bucky device. Hold the BEM in place with the Compression Device.
4. Make an exposure. Verify a low mAs reading is obtained (nominally between 12 and 40
mAs, dependent on attenuation).
5. If the mAs reading is above or below the specified range, add BEM to increase mAs, or
remove BEM to decrease mAs.
6. Reload the Bucky (using the same cassette and film type), and make an exposure.
7. Verify that the mAs falls within the specified range. If it does, remove the cassette and
develop the film.
8. Verify the absence of grid lines (SRL 2000 Bucky) or grid patterns (HTC Bucky) in the
developed film.
9. Place 5.0 cm of BR-12 (or PMMA acrylic) on the Bucky and hold it in place with the
paddle.
10. Reload the same cassette with the same film type, then load it into the same Bucky.
11. Make an exposure using the technique factors from Step 1.
12. Verify a high mAs reading is obtained (nominally between 200 and 400 mAs, dependent
on attenuation).
13. If the mAs reading is above or below the specified range, add attenuation to increase
mAs, or remove attenuation to decrease mAs.
14. Reload the Bucky (using the same cassette and film type), and make an exposure.
15. Verify that the mAs falls within the specified range. If it does, remove the cassette and
develop the film.
16. Verify the absence of grid lines (SRL 2000 Bucky) or grid patterns (HTC Bucky) in the
developed film.
17. Repeat Steps 2 through 15, but with the C-arm rotated to -90° (grid travel in the direction
of gravitational pull).
18. Verify the absence of grid lines (SRL 2000 Bucky) or grid patterns (HTC Bucky) on the
developed film for all exposures.
19. Repeat Steps 2 through 15, but with the C-arm rotated to +90° (grid travel against the
direction of gravitational pull).
20. Verify the absence of grid lines (SRL 2000 Bucky) or grid patterns (HTC Bucky) on the
developed film for all exposures.
21. Repeat the entire procedure for all Bucky devices.
D
A
H
B
5c
F
SCATTER
PROBE
9. Repeat this entire procedure for positions “B,” “C,” and “D” in Figure 4-15.
10. Install the 24 x 30 cm Bucky (if used). Repeat this test for positions “E”, “F”, “G” and “H”
(Figure 4-15) for the 24 x 30 cm Bucky.
11. If there are any readings greater than 0.10 mR, take corrective action to repair the
system.
BS
NEWTONS/L
CM
Note… Any reading above 100.0 mR/Hour is unacceptable. Take corrective action
to repair the system.
Always position the X-ray Beam Alignment Template accurately on the image receptor
support by performing the following:.
1. Place the screen (facing up) onto the IRSD and center it.
2. Move the screen forward (away from the C-arm) approximately 1.5". Lay the template on
top of the screen.
3. Butt the rear edge of the template against the C-arm to place the image receptor tray
edge line directly above the edge of the IRSD.
4. Center the template left-to-right until the cut in edges (at the rear of the template) are set
flush with the sides of the image receptor support.
Since it is necessary to make repetitive exposures, the system can be set so that there is only
a fifteen-second Standby interval between exposures. On the Host Microprocessor Board, set
DIP Switch S2 switch 1 to On. Remember to set this switch to Off at the end of this
procedure.
The recommended method for making exposures and determining required adjustments
during this procedure:
1. Make one exposure per edge (for example, concentrate only on evaluating one edge
each time an exposure is made).
2. Use short exposure durations (one to two seconds).
3. Remember the direction of movement for that edge, and the approximate distance
required for the movement each time an exposure is made.
Keep in mind, attempting to remember the directions and amounts for more than one edge at
a time requires longer exposures, and usually, the same number of exposures as would be
necessary when doing one edge at a time.
Note… Not all apertures are standard equipment on the Affinity. Perform the
sections applicable to the apertures packaged with the system.
4. Adjust the Affinity apertures (shown in Figure 4-19) in the order shown below:
• Section 7.4, Auto-Aperture Alignment, p. 4-52
• Section 7.5, Magnification Full Field (18 x 24 cm) Aperture Alignment (SM FS), p. 4-54
• Section 7.6, 10 cm Coned-down Contact Aperture Alignment (LG FS), p. 4-55
• Section 7.7, 7.5 cm Spot Contact Aperture Alignment (LG FS), p. 4-55
• Section 7.8, 7.5 cm Spot Magnification Aperture Alignment (SM FS), p. 4-55
4 5
Note… A blocker is anything that can be inserted into the aperture sensor (for
example, cotton swabs, wadded paper) that make the sensor think an
aperture is installed.
3. Make an exposure and note the location of the edges of the X-ray beam relative
to the 18 x 24 cm markings on the alignment template.
4. If the edges of the beam are not aligned properly, adjust the Auto-Aperture as
follows:
a. Slightly loosen the four screws (Figure 4-20, #1) securing the Auto-Aperture
adjustment blades just enough so that minimal effort is required to move the
blades.
b. Making exposures as necessary, adjusting the rear blade so that it is parallel
to the rear edge markings on the alignment template.
Caution: Never backdrive any axis (manually turning worm gear (Figure 4-20, #3)
on the Auto-Aperture motor) because it will destroy the gearhead. Always
use the motor control (SW1 on Tubehead Control Board) to move around
each axis.
c. If the back edge field adjustment is off, reposition by relocating the Auto-
Aperture sensor to match the new location of the rear edge of the aperture.
d. Cycle the IN/OUT position of the Auto-Aperture by using SW1 on the
Tubehead Control Board, make exposures and verify position.
e. Making exposures as necessary, adjust the right blade to align within the
markings on the alignment template.
f. Tighten the two screws securing the blade.
g. Making exposures as necessary, adjust the left blade to align within the
markings on the alignment template.
h. Tighten all screws (including the sensor screw).
i. Re-verify the beam position to ensure that the blades did not move when
tightening the hardware.
7.5 Magnification Full Field (18 x 24 cm) Aperture Alignment (SM FS)
a. Longitudinal Position: alignment of front and rear edges. Loosen the screws
(Figure 4-21, #1) and rotate the lamp clockwise to move the Light Field area
forward (towards the patient); rotate it counterclockwise to move the Light
Field area toward the C-arm.
b. Lateral Position: alignment of left and right edges - Loosen the screws (Figure
4-21, #2) and move the lamp toward the left to move the light field to the
right; move the lamp toward the right to move the light field to the left.
c. Overall Size: overall area of pattern (combined length and width) - loosen
the screws securing the bracket to the C-arm frame (Figure 4-21, #3) and
move the entire bracket forward (toward patient) to expand the Light Field
area; move it back to contract the Light Field area.
Note… Inability to accurately align the Light Field with the X-ray image area may
indicate aperture plate mis-alignment.
4. Repeat this Illuminance test for positions “B,” “C,” and “D” in Figure 4-23.
5. If the difference in any quadrant is less than 160 lux, the Light Field Illuminance Test is
considered failed.
6. If necessary, remove the lower tubehead cover, then loosen the two screws that secure
the lamp socket to the lamp cradle (Figure 4-24).
7. Increase the light field illuminance by adjusting the lamp position within the cradle.
Rotate the lamp and socket at the same time until the lamp filament is parallel with the
plane of the mirror. Tighten the screws, reinstall the tubehead cover, then repeat the test.
8. Verify that the length-to-width ratio of the light field is not altered.
9. If the test still fails, replace the lamp and repeat the entire test. When complete, perform
the light field alignment as described in Section 7.10.
Note… Do not overly force the chain to achieve the desired deflection. Chain
movement should be effortless.
13. Press a compression down control to move the compression device down approximately
5.0 cm, then manually move it down an additional 2.0 cm. Repeat the chain tension
verification in Step 10.
14. Continue moving the compression device downward, alternating between motorized
and manual movement, until the entire compression range is verified.
15. If any chain tension adjustments are required, repeat the tension test from the beginning.
When complete, reinstall the covers removed previously.
16. Install a Bucky device, a compression paddle and 5.0 cm of BEM.
17. Use motorized compression to lower the compression paddle until it just touches the
acrylic. Use manual compression to apply 133.5 N (30 lb) of force onto the acrylic.
18. Press Reset + Comp. Release at the same time to send the compression thickness
calibration data to the microprocessor.
19. Raise the compression paddle, remove the acrylic from the receptor support, then return
the system to the user mode (S2 switch 1 = Off).
8.2 Compression Thickness Potentiometer Adjustment
1. Ensure the timing belt is mounted on lower bearing block assembly (see Figure 6-37).
2. Bring the compression device to the bottom of its travel.
3. Rotate the potentiometer in a counterclockwise direction (when viewed from the right
side of the C-Arm assembly) until it bottoms out, then make one full turn of the pot in a
clockwise direction.
4. Manually lower the
compression device until the
potentiometer is aligned with
the access hole in the left C-
arm side rail.
5. Insert a flat blade screwdriver
through the access hole in the
left C-arm side rail. Rotate the
potentiometer in a clockwise
direction until it bottoms out,
then make one half turn of the
potentiometer in a
counterclockwise direction.
6. Tighten clamp screw (Figure 4-
26, #1) on potentiometer shaft
at this time. Ensure clamp screw
is aligned with timing belt.
7. Perform the Compression
Thickness Calibration Figure 4-26: Compression Thickness Potentiometer
Procedures as described in
Section 3.2. Legend for Figure 4-26
1.Clamp Screw
2.Sprocket
3.Hex Nut
4.Potentiometer Shaft
5.Mounting Bracket
Chapter 5: Maintenance—General
1.0 Introduction
This chapter provides general information to aid the Service Engineer in maintaining the
Affinity. The information covered in this chapter is as follows:
• Preventive Maintenance (Section 2.0)
• Troubleshooting (Section 3.0, p.5-3)
Frequency PM Refer to
D; AR Pre-Startup Inspection Chapter 3, Section 2.1, Pre-
Startup Inspections, p. 3-2
D; AR Post Startup Tests Chapter 3, Section 2.3, Post-
Startup Tests, p. 3-3
S C-arm Safety Function Test Chapter 4, Section 5.19, C-arm
Safety Function Test, p. 4-35
AR C-arm Safety Microprocessor Chapter 4, Section 8.4, C-arm
Board Calibration Safety Microprocessor Board
Calibration/Vertical Drive Force
Load Cell Setting, p. 4-63
S Tube Voltage Calibration Chapter 4, Section 4.1, Tube
Voltage Calibration, p. 4-9
S Filament Current Calibration Chapter 4, Section 4.3, Filament
Current (mA) Calibration, p. 4-12
S Automatic Exposure Control Chapter 4, Section 5.0, Automatic
System Calibration Exposure Control System
Calibration, p. 4-19
S Compression System Calibration Chapter 4, Section 3.0,
Compression System Calibration
Procedures, p. 4-7
S X-ray System Performance Tests Chapter 4, Section 4.0, Exposure
System Calibration, p. 4-9
S System Level Performance Tests Chapter 4, Section 6.0,
Performance Tests, p. 4-36
A X-ray Field Alignment Test Chapter 4, Section 7.0, X-ray and
Light Field Test and Alignment
Procedures, p. 4-48
A Light Field Alignment Test Chapter 4, Section 7.10, Light
Field Alignment Test and
Adjustment, p. 4-56
3.0 Troubleshooting
This section provides an overview of system theory of operation and the system Alert Codes.
The theory of operation and Alert Codes can be used in conjunction with the Test Point and
LED Indicators Reference Tables found in Appendix B and the Affinity Schematic Package for
troubleshooting the Affinity.
Power Control
The Power Control PCB controls the following system functions:
• Collimator lamp power
• Compression motor drive
• Compression motor brake
• Rotation Lock release
• VTA actuator drive
All input signals come from the Host Microprocessor PCB.
Auto-Filter
This mode provides full automatic operation of all technique factors. Upon exposure
initiation, the AEC signal is sampled to determine if the kV needs to be incremented
upward to provide an exposure within the 200 mAs window, or the 200 mrad
window. When determined by the AutoFILTER algorithm, the exposure is
momentarily interrupted and the Rhodium filter is moved into the beam in place of
the Molybdenum filter. After the kV or kV/filter has been selected by the algorithm,
the Auto-Time function completes the exposure. The final kV, post-mAs, and Filter
are displayed upon termination of the exposure.
1. Connect the laptop computer to the Affinity System using the Null-Modem Cable as
shown in Figure 5-1.
Note… Care must be taken to avoid selection of INIT CAL DATA option in the
Special command field. When selected and initiated, this field resets all
calibration data.
2. Press Reset + Comp Rel to access the Alert Codes Log.
3. The title of the log, # of records, and CMD line appear on the display screen,
along with the first page of alert codes. Use scroll keys to view the log.
4. Select (using the change keys) the DOWNLOAD command.
5. Press Reset + Comp Rel to initiate data transfer.
6. Select (using the change keys) the CLEAR command.
7. Press Reset + Comp Rel to clear the log.
8. Verify the log is empty.
9. From the computer, the log can now be printed or appended.
Note… Care must be taken to avoid selection of INIT CAL DATA option in the
Special command field. When selected and initiated, this field resets all
calibration data.
2. Set up Hyperterm to display downloading data and select Capture Text from the
Transfer Menu. Type in an appropriate path and set the file name to any VALID
FILE NAME.TXT, include system identification (for example, DOWN1.TXT).
3. Press Reset + Comp Rel to automatically download the calibration data to the
laptop.
4. When complete, the message area on the display screen shows DOWNLOAD
COMPLETE. On the HyperTerm, select Transfer - Capture Data - Stop. Verify
that the data downloaded correctly by examining the text file established in Step
2.
5. From the computer, the log can now be printed or appended.
1.0 Introduction
This chapter details the remove and replace procedures for the Affinity. Where necessary,
references are given for performing required alignment and/or adjustment procedures.
• Section 2.0 provides the remove and replace procedures for the components within the
Gantry.
• Section 3.0 provides the remove and replace procedures for the components within the
C-arm Assembly.
Only authorized, trained service engineers can service this mammography system. The
system is designed for module-level repair. When a defective module (assembly) is
identified, the service engineer replaces it with a new one.
Note… When servicing the Affinity, if a procedure instructs you to remove any
covers or panels, do not reinstall the covers until all required procedures
(for example, calibrations, adjustments, tests, remove and replace, etc.) are
completed.
6
2
1
7 12
2
13
3
8
9 14
4
10
5
11
STOP
Caution: In step 6, be careful that you do not disconnect any connectors when
pulling the cover away from the system.
6. Pull the cover slightly away from the system, then set it upright on the floor.
Note that a wiring harness connected to the Angle Display Board and the
Emergency Off Switch is still attached to the Gantry. Proceed to Step 8 if the side
panel is to be replaced.
7. Do not perform steps 8 and 9 if the system is to be turned on at any point while
servicing the system. The system does not function if the cables/side panels are
disconnected.
8. Turn the cover away from the system to gain access to the wiring harness.
Follow the harness to the connector on the Gantry frame.
9. Disconnect the wiring harness, then set the cover in a secure location.
WARNING! Always turn the system off before performing any removal or
replacement procedure.
Note… The HV Generator is assembled and calibrated at the factory and is certified
as a complied device. It must be replaced as an assembly.
The type of mounting plate used with the HV Generator Assembly determines the
remove and replace procedures used. Affinity Mammography Systems with a Serial
Number between xxxmmyy0001 and xxxmmyy0018 (xxx represents the system
series designator; mmyy represents the month and year of manufacture) require
removal of the HV Inverter Board and the HV Control Filament Board prior to
removing the HV Generator Assembly Mounting plate. See Figure 6-7. For those
systems with a Serial Number of xxxmmyy0019 or above, the HV Generator
Assembly can be removed as a single item. See Figure 6-8. The system Serial
Number is located on a nameplate on the rear of the Gantry. Refer to Chapter 1,
Figure 1-4, Label 1.
Affinity Serial Numbers 0001 through 0018
Remove the HV Generator Assembly (see Figure 6-7)
1. Remove power to system and access the High Voltage Generator Assembly by
removing the left side Gantry cover as described in Section 2.1.1.
2. On the top of the HV Multiplier, disconnect the HV cable from the HV cable
receptacle.
3. On the HV Inverter board, disconnect GJ1. Wait approximately 6 minutes
before proceeding (see WARNING below).
4. On the HV Inverter board, disconnect GJ2, GJ3, and GJ5. Remove any installed
cables ties.
5. Remove the Ribbon Cable between the HV Inverter board and the HV Control
Filament board. Set aside for re-installation later.
Note… The replacement board must be calibrated before the system is released to
the user.
7. Plug the 2 previously removed power wires onto their respective lugs on the
replacement microswitch.
8. Install the new switch, but do not tighten the mounting screws.
9. Turn the system on.
10. With the C-arm still at the vertical (0°), verify that the trigger arm roller is in the
cam detent.
11. Adjust the switch height so that the switch actuator arm just closes the switch,
then tighten the mounting hardware.
12. Verify that the C-arm Rotation Angle Displays read 0°.
13. Rotate the C-arm and verify that the switch opens when the trigger arm roller is
out of the cam detent.
14. Rotate the C-arm to the following detent positions and verify correct rotation
angle display readings: ±45°, ±90°, ±135°, +180°.
10. Loosen the screw that fastens the jam nut bracket (Figure 6-14, #5) to the jam
nut (Figure 6-14, #6), then remove the screw that fastens the bracket to the
standoff (Figure 6-14, #7) on the lower portion of the VTA isolation mount
(Figure 6-14, #8).
11. Thread the jam nut down 3 to 4 inches.
12. Remove the two screws that secure the hub nut (Figure 6-14, #9) to the VTA
isolation mount, then thread the hub nut down to the jam nut.
13. Remove the actuator assembly from the unit by pulling the motor forward and
sliding the lead screw down, then out of the VTA isolation mount.
14. To install the replacement actuator assembly, tilt the actuator lead screw toward
the system, then slide the lead screw up and through the top isolation mount.
15. Position the actuator motor into the bottom isolation mount, then align the holes
in the mount with the hole in the base of the motor assembly.
16. Insert the clevis pin through the holes, then reinstall the spring clip to secure the
clevis pin.
17. Thread the hub nut along the lead screw until its top surface is approximately
5-5/8" from where the lead screw enters the motor assembly.
18. Plug the 4-pin connector into the motor.
Use the following illustration and legend to identify parts on the Tubehead and on the Beam
Limiting Assembly (see Figure 6-16).
Mo
Rh
C
M
Use the following illustration and legend to identify parts on the Compression Device (see
Figure 6-17).
Use the following illustration and legend to identify parts on the Image Receptor Support
Device (see Figure 6-18).
Note… For exposure other than magnification studies, always ensure that the Face
Shield is attached. Remember to verify its condition prior to use each day.
The Face Shield does not offer protection from exposure through the acrylic.
Mo
Rh
C
M
10. Install the high voltage lead and the filament connector to their receptacles on
the rear of the X-ray tube.
11. Reset the Exposure Counter. See Section 3.2.2.1.
12. Perform the following exposure system setup procedures:
• Section 4.1, Tube Voltage Calibration
• Section 4.3, Filament Current (mA) Calibration
• Section 5.3, Initial Calibration—Large Focal Spot (Mo)
13. When the exposure control system is calibrated, perform the following
compliance tests (Chapter 4):
• Section 6.1, Half-Value Layer
• Section 6.2, Reproducibility and Linearity in Manual Mode
• Section 6.3, Reproducibility in Auto-Time Mode
• Section 6.4, Reproducibility in Auto-kV Mode
• Section 6.5, Reproducibility in Auto-Filter Mode
• Section 7.0, X-ray and Light Field Test and Alignment Procedures
• Section 5.1, AEC Detector Gain Calibration
3.2.2.1 Exposure Counter Reset
The Affinity Series keeps track of the number of exposures made. Whenever the
X-ray tube is replaced, it is necessary to clear - or reset - the exposure counter to
accurately tabulate the number of exposures made with the new X-ray tube.
Note… This procedure is only necessary after installing a replacement X-ray tube.
1. If the system is not in service mode, refer to Chapter 4, Section 2.1, Entering
Calibration Mode, p. 4-2 and set the system for the service mode (S2 switch
1 = ON).
2. Access the Debug Screen (Chapter 3, Figure 3-9) and record the number of
exposures taken since last reset (data line EXP:).
3. Access Calibration Screen #3 (Figure 4-3).
4. Highlight the Special field. Press a Change key until the field reads: Clear
Exp Cnt.
5. Press Reset + Comp. Release at the same time to reset the exposure counter.
6. Return the system to the user mode (S2 switch 1 = OFF).
Note… If it is only necessary to replace the firmware chip (U9), skip to Step 6 now.
2. Disconnect the wiring harness connectors on the Tubehead Control Board
(Figure 6-26). Note their locations.
S1 S2 S3 S4
3. Remove the five screws that secure the Tubehead Control Board to the standoffs
on the tubehead chassis.
4. Pull the board out of the unit.
5. Reverse procedures to replace the board. Skip to Step 7.
6. Replace the Firmware Chip (U9) as described in Section 2.5.1.3, Firmware
Chip, then go to Step 7.
7. Verify the board operation as follows (refer to Table 6-1and Figure 6-26):
a. Turn the system on.
b. Press S1 once on Tubehead Control Board, verify the Auto-Aperture moves
out of the field.
c. Press S1 a second time, verify the Auto-Aperture moves back in.
d. Press S2, verify the lamp and mirror assemblies move out of field.
e. Press S4, verify the lamp and mirror assemblies move back in.
f. Press S3, verify the filter select mechanism changes position.
g. Press S4, verify the entire collimator assembly cycles in and out of field.
Switch Function
S1 Auto-Aperture Select
S2 Light Field Lamp Assy/Motorized Mirror Assy Select
S3 Filter Select
S4 Collimator Assy Select
Optical Sensor 1 2 3 4
Tubehead Control Board LED D3 D2 D5 D4
Note… The top and bottom Compression Device covers are fastened to the
compression bellows. Use care not to damage the bellows when removing
these covers.
1. Remove the mounting hardware that fastens the Compression Device covers
(top and bottom) to the Compression Device. (See Figure 6-17.)
2. Carefully lift the covers off the Compression Device to access the following
components:
• Compression Clutch
• Compression Clutch brake
• Compression Drive
• Compression Brake
• Compression Tray Sensor Board
• Compression Force and Thickness Boards
3. Disconnect the wiring between the covers and the Compression Device,
including 1 ground wire for each cover.
4. Reverse procedures to reinstall the Compression Device Covers.
Note… The cam-lock set screw is recessed further in the compression motor
sprocket than the key-way set screw, which is flush when tight.
4. Turn the handwheel again to continue rotating the motor sprocket to access the
key-way set screws. Loosen this set screw 1/2-turn.
5. Remove the 2 screws (Figure 6-32, #1) that hold the compression motor and the
motor mounting plate to the C-arm rail. Remove the motor and mounting plate
from the C-arm.
6. Note the polarity of the wires on the compression motor, then disconnect them.
Pull the motor out from the mounting plate, until the motor shaft clears the
plate, then pull the motor out from the C-arm.
7. Transfer the mounting plate to the new motor, then reverse this procedure to
install the replacement motor assembly.
8. Turn the system on and verify that the motor works by cycling the motor in each
direction using the compression controls.
9. When complete, perform the Compression Chain Tension Adjustment as
described in Chapter 4, Section 8.1.
14. Perform the following before releasing the system to the user:
• Chapter 4, Section 8.1, Compression Chain Tension Adjustment, p. 4-61
• Chapter 4, Section 3.1, Compression Force Calibration, p. 4-7
• Chapter 4, Section 3.2, Compression Thickness Calibration, p. 4-8
7. Gently thread the timing belt down and out of the pulleys. Inspect the timing
belt, and if it is damaged, replace it as follows:
a. Remove the IRSD covers.
b. Loosen the two screws securing the belt in the lower clamp block. They are
located through the rear of the IRSD and under the bellows assembly.
Remove the belt and discard it.
c. Insert the new belt through the lower clamp on the Tubehead Support Plate.
Tighten screws enough to secure belt in place.
d. Thread the timing belt through the pulleys, ensuring the potentiometer does
not move and the timing belt does not twist.
e. Insert the timing belt through the upper clamp block (see Figure 6-36).
Apply Blue Loctite 242 and tighten screws.
f. Loosen screws securing timing belt in lower clamp block. While holding on
to the bottom of the timing belt, raise the compression device to the top of
its travel.
g. Affix the Field Service Belt Adjustment Tool (PN 9-061-0107) to the bottom
of the timing belt and loosen the screws securing the belt in the bottom
clamp. Apply Blue Loctite 242 and tighten screws.
8. Locate the 3 colored wires soldered to potentiometer. Note the colored wire for
each post and desolder the wires. See Figure 6-38.
9. Remove the screws from potentiometer bracket and pull it from the compression
device.
10. Loosen the clamp screw (1) on the end of the potentiometer shaft (4). Remove
the clamp and the sprocket (2) together.
11. Remove the hex nut (3) that secures the potentiometer to the mounting bracket
(5). Slide the potentiometer off the bracket.
12. To replace the Compression Thickness Potentiometer, reverse Steps 8 through
11 above. Do not tighten clamp screw at this time.
13. Perform the Compression Thickness Potentiometer Adjustment as described in
Chapter 4, Section 8.2, Compression Thickness Potentiometer Adjustment, p. 4-
62.
14. Reinstall the IRSD covers, the tubehead covers, the compression device
assembly upper and lower covers, the C-arm bottom cover and left side cover.
15. Reinstall the compression accessory. Restore power to system.
1. Turn the system off. Remove the IRSD covers as described in Section 3.4.1. To
replace the board, go to Step 3.
2. If you are only replacing the software (Figure 6-42, #1), see Figure 6-43 and
remove and replace the chip. Reinstall the IRSD covers and go to Step 6.
3. Disconnect all cables and ground wires to the AEC Position Compression
Display Microprocessor Board (Figure 6-42).
4. Remove the four nuts securing the board to the IRSD bottom plate.
5. To install the replacement AEC Position Compression Display Microprocessor
Board, reverse these steps.
6. When complete, perform the procedures listed below before releasing the
system to the user:
• Chapter 4, Section 5.0, Automatic Exposure Control System Calibration, p.
4-19
• Chapter 4, Section 6.10, IRSD Leakage Test, p. 4-46
1.0 Instructions
This Chapter of the manual provides tabular listings of the serviceable components for the
Affinity. Part numbers and descriptions are provided for component identification.
Table 7-1: Gantry Cover, Control Panel and Internal Components—Parts List
Appendix A—Specifications
2.2 Compression
Manual Compression Force Limited to a maximum 300 N, +0/-88 N (67.4 lb, +0/-20
lb) between -150° through 0° to +150° of C-arm
rotation. It is not less than 150 N (33.7 lb) at all other
angles.
Motorized Compression Force User Selectable for either Full or Pre Compression
modes in User Defaults. 66 N to 178 N (15 lb to 40 lb),
user-selected within this range. Factory default of 132 N
(30 lb).
Full Compression User adjustable with a nominal range of 111 N
minimum to 178 N maximum (25 lbs minimum to 40
lbs maximum).
Pre Compression User adjustable with a nominal range of 66 N minimum
to 132 N maximum (15lbs minimum to 30 lbs
maximum).
Motorized Compression Activation Motorized Up/Down controls on both sides of the C-
arm and on the dual function footswitch.
Manual Release Manual Up/Down handwheel controls on both sides of
the Compression Device.
Mode kV mAs
Auto-Time 20 to 39 kV 5 to 500 mAs
Auto-kV 25 to 30 kV 5 to 500 mAs
Auto-Filter 25 to 32 kV 5 to 500 mAs
X-ray Collimation The system detects the attached image receptor and full
field collimation is provided by an auto-aperture
mechanism with additional fixed apertures for
magnification and reduced field operation. An aperture
detection system ensures that the correct aperture is
used with the installed image receptor. Exposures are
inhibited when improper combinations are detected.
The collimator has a beam-limiting device that allows
the entire chest wall edge of the X-ray field to extend
5 mm, +2.5 mm, beyond the edge of the Image
Receptor and provides a means to ensure the X-ray field
does not extend beyond any non-chest wall edge of the
Image Receptor of more than 2% of the SID.
Light Field to X-ray Field Congruency The light field that passes through the X-ray beam
limitation device is aligned with the X-ray field so that
the total of any misalignment of the edges of the light
field and the X-ray field along either the length or the
width of the visually defined field at the plain of the
breast support surface does not exceed 2% of the SID.
Light Field Lamp The lamp illuminates for 30 seconds ±5 seconds upon
pressing a Light Field Lamp Button on the C-arm, or by
pressing a Compression Down Button. The lamp
extinguishes automatically when X-ray exposure is
initiated. A shatter shield is provided.
Light Field Illuminance The light localizer has a minimum illuminance of 160
lux and meets all of the requirements of 21CFR, section
1020.31. The lamp voltage is the least amount required
to assure 160 lux at low line. A shatter shield surrounds
the lamp. Lamp is service adjustable to provide
alignment of the light field to the X-ray field.
Light Field Mirror The light field mirror is motor driven out of the X-ray
field within 1.0 sec of X-ray initiation signal. Position
sensors provide for X-ray interlock. The lamp position is
adjustable between large focal spot and small focal spot
to provide proper light field to X-ray field congruency
required for a bi-angle target X-ray tube.
4.0 Accessories
4.1 Compression Paddles
Compression Paddles The compression paddles are transparent and the full
field paddles are marked with the location of the AEC
sensor positions. These markings are not detectable on
film when imaged with 1 cm of acrylic attenuator at 22
kV to an Optical Density of 1.2 OD. The paddle
attenuation is less than 20% (reduction of mR/mAs) at
28 kV. The full field paddles are adjustable to provide
focal spot, compression device and image receptor
alignment requirements.
The compression paddles including the frame and
compression device maintain a parallel compression
and do not deflect by more than 1 cm difference from
any surface providing compression under maximum
motorized compression force (except for FAST paddles).
Compression Paddle Sizes Available paddles:
18 x 24 cm Full Field Contact
24 x 30 cm Full Field Contact
7.5 cm Spot Contact/Magnification
10 cm Contact/Magnification
10 cm Open (Rectangular) Localization
10 cm Perforated Localization
15 cm Contact/Magnification
15 cm Open (Rectangular) Localization
15 cm Perforated Localization
18 x 24 cm Full Field FAST Paddles
24 x 30 cm Full Field FAST Paddles
Ultrasound Paddle
Small Breast Paddle
The chest wall edge of the compression paddle does not
extend beyond the chest wall edge of the image
receptor by more than 1% of the SID when tested with
the compression paddle placed above the breast
support surface at a distance equivalent to standard
breast thickness. The shadow of the vertical edge of the
compression paddle is not visible on the image.
4.2 Apertures
There are five apertures for all combinations of paddles and receptors.
Auto-Aperture
7.5 cm Spot Contact
7.5 cm Spot Magnification
10 cm Coned
Magnification Full Field
4.3 Magnification
Magnification Factor 1.8 ±0.1 X (2.25 cm above the surface) for platform and
table devices
Material of breast support surface
Platform: Polycarbonate
Table: Carbon Fiber
Size of breast support surface 12 x 16.5 cm minimum to provide full 18 x 24 cm full
field magnification at 1.8X.
1.0 Introduction
This Appendix provides tables used for identifying test points, LED indications, fuse identification and
location, jumpers, and accessory identification sensor combinations (image receptors) in the Affinity
Series System.
Caution: Be aware of the damage static electricity can inflict on electronic components. Always wear a
grounded electrostatic discharge strap when handling static sensitive components.
Use the following test point tables as an aid for troubleshooting the Affinity.
• Table B-1: Filament Protect Board (PN 1-003-0289) Test Points, pg. B-1
• Table B-2: Power Control Board (PN 1-003-0445) Test Points, pg. B-2
• Table B-3: C-arm Safety Board (PN 1-003-0440) Test Points, pg. B-2
• Table B-4: Display Microprocessor Board (PN 1-003-0428) Test Points, pg. B-3
• Table B-5: Tubehead Control Board (PN 1-003-0403) Test Points, pg. B-3
• Table B-6: AEC Position Compression Display Microprocessor (PN 1-003-0471) Test Points, pg. B-3
• Table B-7: AEC Position Compression Display (PN 1-003-0472) Test Points, pg. B-4
• Table B-8: Auxiliary Power Distribution Board (PN 1-003-0489) Test Points, pg. B-4
• Table B-9: Communications Interface Board (PN 1-003-0434) Test Points, pg. B-4
• Table B-10: Host Microprocessor Board (PN 1-003-0493), pg. B-5
• Table B-11: HV Control Filament Board (PN 1-003-0485) Test Points, pg. B-6
Table B-6: AEC Position Compression Display Microprocessor (PN 1-003-0471) Test Points
Table B-7: AEC Position Compression Display (PN 1-003-0472) Test Points
Table B-8: Auxiliary Power Distribution Board (PN 1-003-0489) Test Points
LED “ON”
LED LED “ON” Indication LED
Indication
D1 +5 V D6 ACC0
D2 Actuator Fault D7 ACC1
D3 Load Cell Fault (load cell at limit) D8 ACC2
D4 Clock signal (flickers to indicate board is working) D9 VCC
D5 CAL Mode Initiated D10 -5 V
LED Color LED “ON” Indication LED Color LED “ON” Indication
D4 RED GRID D40 GRN R+ 14 V
D29 RED HV Interlock D41 GRN HV Enable
D30 RED INV Interlock D43 RED Rotor Over Current
D32 RED ARC D57 RED Tube Over Current
D33 RED INV Over Current D67 GRN Filament Enable
D35 RED Tube Over Voltage D68 GRN SM Focal Spot
D37 GRN -15 V D70 GRN LG Focal Spot
D38 GRN +5 V D71 RED Over Current
D39 GRN +15 V D75 RED Over Voltage
5.0 Fuses
Figure B-1, page B-11 (on the following page) shows the location of the fuses that are on the Affinity
Series Gantry circuit boards. One additional fuse is located on the Relay Protection Board (part of the
Input Power Assembly Chassis). Refer to Table B-20 for rating and part number information on the fuses.
6.0 Jumpers
The following tables show the Jumpers on the Affinity boards with the default (factory setting)
configuration. The jumpers should not be changed from the configuration shown unless directed to do so
during a test, or calibration procedure. Any time the jumper setting is changed, it must be returned to the
default position unless the change was made to establish new operating parameters.
• Table B-21: Microprocessor Board (PN 1-003-0493) Jumper Settings
• Table B-22: Communications Interface Board (PN 1-003-0434) Jumper Settings
• Table B-23: C-Arm Safety Board (PN 1-003-0440) Jumper Settings
• Table B-24: HV Control Filament Board (PN 1-003-0485) Jumper Settings, pg. B-13
JP1
JP2
Index
P/N 9-500-0255 i
Service Manual
Index
Connecting Film Labeling Devices 2-12 C-arm Safety Microprocessor Board 6-21
Connecting Power Cable 2-5 Communications Interface Board 6-21
Control Panel 3-7 Host Microprocessor Board 6-18
Screens 3-8 Low Voltage Power Supply 6-21
Control Panel Installation 2-8
remote location 2-8 H
to Gantry 2-8 Half-Value Layer Test 4-36
Control Panel, description 1-8 High Voltage Generator Assembly 6-2
Controls and Indicators 3-1 High Voltage Generator Calibration Setup 4-10
Host Microprocessor Board 6-2
D
download/upload data 5-12 I
Downloading Alert Codes Log 5-13 Image Receptor Detect Board, Sensor Combinations
Downloading Calibration Data 5-13 B-9
Dual Function Footswitches, Installing 2-11 Indicators and Controls 3-1
Indicators, LED B-7
E Initial Startup Procedures 2-13
Electrical Safety 1-3 Input Power Configuration 2-4
Exposure Control 5-8 Installing
Exposure System, Calibrating 4-9 Control Panel 2-8
Exposure Technique Defaults to Gantry 2-8
Setting 3-15 Control Panel, remote location 2-8
External Indicators 1-5 Dual Function Footswitches 2-11
Operator Radiation Shield 2-14
F System 2-4
Filament Current (mA), Calibrating 4-12 System Accessories 2-10
Filament Protection Board 6-35 X-ray Footswitch 2-10
Filament Waveforms Pattern 4-13 Interlocks 1-5
Film Labeling Devices, Connecting 2-12 International Symbols P-x
Filter Indicator 3-5 Introduction to Affinity 1-1
Final Operational Setup 2-14 IRSD
Footswitch Components 6-31
Controls 3-6 Leakage Test 4-46
Functional Tests 3-1 Isolation Integrity 1-4
Fuses B-10
J
G Jumpers B-12
Gain Adjustment
Initial Calibration, Large Focal Spot K
Mo 4-21 kV Control Offset Calibration 4-64
Rh 4-22
Initial Calibration, Small Focal Spot L
Mo 4-26 Label Location 1-11
Rh 4-27 Leakage Test, IRSD 4-46
Gantry and C-arm, description 1-6 LED Indicators B-7
Gantry Components Light Field Alignment Test and Adjustment 4-56
Auxiliary Power Distribution Board 6-17 Large Focal Spot 4-56
Firmware Chip 6-20 Small Focal Spot 4-57
HV Generator Assembly Light Field Edge Contrast Test 4-58
Removal 6-16 Light Field Illuminance Test 4-59
Power Control Board 6-17 Light Field Lamp 3-5
Right Location of Labels 1-11
Calibration Storage Battery 6-19 Low Voltage Power Supply 6-2
ii P/N 9-500-0255
Service Manual
Index
iv P/N 9-500-0255