Model ECLOS
Tokyo, Japan
CONTENTS
Page
1. X-ray tube assembly ···························································································· 1
2. Accuracy of radiation output ················································································· 1
3. Loading factors in CT system ················································································ 1
4. Radiation quality ································································································ 2
5. Computed Tomography Dose Index 100 ·································································· 2
6. Sensitivity profile and dose profile ········································································· 6
7. Stray radiation ··································································································· 9
8. Emergency stop ································································································ 10
E1E-BW1020-1
QS4 45456 i
QS4 45456
1. X-ray tube assembly
For information of X-ray tube assemblies, refer to the following instruction manuals
attached with X-ray tube assembly.
“GS3070 Rotating Anode X-ray tube”
“GS4570 Rotating Anode X-ray tube”
“Information for Users”
“Information for Assemblers”
“Information for Assembly and installation”
Loading factor for leakage radiation 130 kV, 50 mA, large focus
E1E-BW1020-1
QS4 45456 1
4. Radiation quality
4.1 First half-value layer in CT(IEC60601-2-44 cl. 29.201.5)
100 kV: 3.0 mm Al minimum
120 kV: 3.8 mm Al minimum
130 kV: 4.2 mm Al minimum
+50 mm
D( z )
CTDI 100 = ∫
−50 mm
N ×T
dz
Where
CTDI100 is Computed Tomography Dose Index;
D(z) is the dose profile along a line Z perpendicular to the tomographic plane, where dose is
measured in air kerma;
N is the number of tomographic sections produced in a single rotation of the radiation source;
T is the nominal tomographic section thickness.
E1E-BW1020-1
QS4 45456 2
5.3 Dose at typical CT condition of operation
The location of CTDI100(center) value is along the axis of rotation and the center of the
dosimetry phantom.
The CTDI100(peripheral) value is Average of the four value of CTDI100 located 10mm
interior to the surface at positions 0°, 90°, 180°, 270° along a line parallel to the axis of
rotation in the dosimetry phantom. The location of the maximum CTDI100 in
peripheral is 12 o’clock position in front of the gantry.
E1E-BW1020-1
QS4 45456 3
5.4 Normalized CTDI100
These below values are normalized CTDI100 by the typical dose in item 5.3.
E1E-BW1020-1
QS4 45456 4
Normalized CTDI100(peripheral) value as a scan parameter varies from the typical CT
condition of operation.
When a scan parameter varies from the typical conditions, you can estimate the dose
by multiplying the typical dose in item 5.3 by the Normalized CTDI100 listed above.
E1E-BW1020-1
QS4 45456 5
6. Sensitivity Profile and Dose Profile (IEC60601-2-44 cl. 29.1.103)
The sensitivity profile is the relative response of a system for computed tomography as a
function of positioning along a line perpendicular to the tomographic plane at the location
corresponding to the center location of the dosimetry phantom. The dose profile is the
graphical representation of the dose as a function of position along a line perpendicular to
the tomographic plane is measured in the center location of the head-dosimetry phantom
and the body-dosimetry phantom.
Here are Profile graphs. These sensitivity profiles are obtained from additional image of the
multi slice images at one scan. For example, using the image addition function, 4 multi
images of 5mm slice thickness become one image of 20mm slice thickness. These sensitivity
profiles include deviation of CT numbers. These dose profiles include variability inherent in
the measurement with TLD (thermoluminescence dosimeter) chips.
1 1
0.8 0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
1 1
0.8
0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
E1E-BW1020-1
QS4 45456 6
2.5mmx4 Sensitivity Profile 10mm Dose Profiles
1 1
0.8 0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
1 1
0.8 0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
E1E-BW1020-1
QS4 45456 7
0.625mmx4 Sensitivity Profile 2.5mm Dose Profiles
1 1
0.8 0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
1 1
0.8 0.8
Relative value
Relative value
0.6
0.6
0.4
0.4
0.2
0.2
0
0 -15 -10 -5 0 5 10 15
-15 -10 -5 0 5 10 15 Distance from scan center (mm)
Distance from scan center (mm)
Head Dose Profile
Sensitivity Profile Body Dose Profile
E1E-BW1020-1
QS4 45456 8
7. Stray radiation (IEC60601-2-44 cl. 29.208)
The stray radiation is measured in the horizontal plane(140cm hight). The body-dosimetry
phantom is positioned in the center of rotation of the CT scanner and centered on the
tomographic plane.
Loading factors: 130 kV, 200 mA , 2 s scan time, 5 mm thickness x 4 images
The unit of measurement is air kerma per mAs applied to the X-ray tube during normal use.
Here is the distribution of the stray radiation.
26
41
11
56
71
86
Scanner
Gantry
11
26
86
71
56
41
Patient Table
E1E-BW1020-1
QS4 45456 9
8. Emergency stop (IEC60601-2-44 cl. 22.4.101)
The stopping distance of the patient support table movement: within 10 mm
The stopping angle of the gantry tilting: within 0.5 degree
E1E-BW1020-1
QS4 45456 10