Professional Documents
Culture Documents
Noninvasive Glucose Monitoring
Noninvasive Glucose Monitoring
AlMughtaribeen University
College of Engineering
Biomedical Engineering Department
Presented By:
Mustafa Khalid Mustafa Elzubair
Noha Mohammed Talha Ahmed
ReemAwis Othman Awis
Supervisor:
Ola Salah
ACKNOWLEDGMENT
We are truly blessed, and utterly thankful for all the opportunities we have been
given in life, all these of which are the reason we are where we are today.
All the praise first and foremost goes to Allah, for guiding our path, and leading
our way.
Our parents who were always there for us with their time, money and all sorts
of support.
Our teachers and tutors throughout our years of education, and to anyone who ever
taught us a word.
A very deep and sincere acknowledgement to Dr. Ola Salah- our project
supervisor, who was always there for our group mates and we with her time,
guidance and support. We could have never asked for better guidance and
supervision.
To Ms. Eng. Fathia Garma and Mr. Mohammed Eltahir who they have generously
given their time and expertise to better our work. We thank them for their
contribution, the great co-ordination and their good-natured support.
To our friends and colleagues, for their never ending encouragement throughout
the years.
Thank you all.
I
DEDICATION
To caffeine and sugar, our companions through many a long nights of studying,
hard working, reading and writing.
II
TABLE OF CONTENTS
Title Page No
ACKNOWLEDGMENT………..…………………..………………….I
DEDICATION…………………….…………………………….…….II
TABLE OF CONTENTS……………………………………….……III
LIST OF FIGURES……….……………………………………….….V
LIST OF TABLES……………..……………………………………VII
NOMENCLATURE…….………………………………….………VIII
ABSTRACT………..…...……………...……………………………..IX
CHAPTER 1: INTRODUCTION……………………………….…….1
1.1.THESIS INTRODUCTION………………………………………………………………...….1
1.2.THESIS OBJECTIVES…………………………………………………………………………..1
1.3.PROJECT LAYOUT……………………………………………………………………..………1
III
4.3.2 Implementation…………………………………….………………………26
CHAPTER 5 : RESULTS AND DISCUSSION…………………….28
5.1. RESULTS………………………………………………………….28
5.2. DISCUSSION……………………………………….……………..32
CHAPTER 6 : CONCLUSIONS AND RECOMMENDATIONS…35
6.1. CONCLUSION……………………………………………………..35
6.2. RECOMMENDATIONS………………………………………..….35
REFERENCES………………………………………………………36
APPENDIX I………………………………………………………...37
IV
LIST OF FIGURES
Figures Page No
Figure (3.1) Visualization of the photo-acoustic effect. 10
Figure (3.2) Absorption spectra of tissue. 11
V
Figure (5.1) Relationship between voltage and glucose concentration for 29
10 persons by using photoacoustic technique.
Figure (5.2) Relationship between voltage and glucose concentration for 30
10 persons by using electrical technique.
Figure (5.3) Clarke's Error Grid Analysis for photoacoustic technique. 32
VI
LIST OF TABLES
VII
NOMENCLATURE
IR INFRARED
NBM NONINVASIVE HEMOGLOBIN MONITORING
NIR NEAR INFRARED
PRESS PRESSURE
OCT COHERENCE TOMOGRAPHY
PAS PHOTOACOUSTIC SPECTROSCOPY
TOF TIME-OF-FLIGHT
MC MONTE-CARLO
FWHM FULL WIDTH OF HALF MAXIMUM
PA PHOTO ACOUSTIC
EGA CLARKE’S ERROR GRID.
PZT LEAD ZIRCONATE TITANATE.
USART UNIVERSAL ASYNCHRONOUS RECEIVER TRANSMITTER
SPT™ SERIAL PERIPHERAL INTERFACE.
I²C™ 2-WIRE INTER-INTEGRATED CIRCUIT
EEPROM ELECTRICALLY ERASABLE PROGRAMMABLE READ ONLY
MEMORY.
LCD LIQUID CRYSTAL DISPLAY.
PIC PICK.
ECG ELECTROCARDIOGRAPHY.
Mg\dl MILLIGRAMS PER DECILITER
mv MILLIVOLT
VIII
ABSTRACT
This project demonstrated the feasibility study and design issues to measure
blood glucose concentration noninvasively in human subjects based on PA and
electrical techniques. PS used acoustic sensor and pulsed laser source (λ=650nm)
which was found to be the suitable range for detecting glucose signal. The mean
value of sensor output for non-diabetic individuals has been determined and
proportional relation between glucose concentration and sensor output voltage has
been derived. The electrical technique used low current source and ECG electrodes
to measure the impedance of human skin then compares it to invasive glucoses
reads.
IX
اﻟﻤﺴﺘﺨﻠﺺ
ﻓﻲ ھﺬا اﻟﻤﺸﺮوع ﺗﻤﺖ دراﺳﺔ ﺟﺪوى وﺗﺼﻤﯿﻢ ﺟﮭﺎز ﻟﻘﯿﺎس وﻣﺮاﻗﺒﺔ ﺗﺮﻛﯿﺰ اﻟﺠﻠﻜﻮز ﻓﻲ اﻟﺪم دون وﺧﺰ وأﻟﻢ
ﺑﺈﺳﺘﺨﺪام ﺗﻘﻨﯿﺔ اﻟﻠﯿﺰر اﻟﺴﻤﻌﻲ واﻟﺘﻘﻨﯿﺔ اﻟﻜﮭﺮﺑﺎﺋﯿﺔ.
ﺗﻘﻨﯿﺔ اﻟﻠﯿﺰر اﻟﺴﻤﻌﻲ ﺗﺴﺘﺨﺪم أﺟﮭﺰة اﻻﺳﺘﺸﻌﺎر اﻟﺼﻮﺗﯿﺔ وﻣﺼﺪر ﻟﯿﺰر ﺑﻨﻄﺎق ﻣﻌﯿﻦ ﻟﻨﺒﺾ اﻟﻠﯿﺰر ﻟﻠﻜﺸﻒ ﻋﻦ
إﺷﺎرة اﻟﺠﻠﻮﻛﻮز.
اﻣﺎ اﻟﺘﻘﻨﯿﺔ اﻟﻜﮭﺮﺑﺎﺋﯿﺔ اﻟﺘﻲ ﺗﺴﺘﺨﺪم ﻣﺼﺪر ﺗﯿﺎر ﻣﻨﺨﻔﺾ واﻗﻄﺎب ﺟﮭﺎز ﺗﺨﻄﯿﻂ اﻟﻘﻠﺐ ﻟﻘﯿﺎس ﻣﻘﺎوﻣﺔ ﺟﻠﺪ
اﻹﻧﺴﺎن وﻣﻦ ﺛﻢ ﻗﯿﺎس ﺗﺮﻛﯿﺰ اﻟﺠﻠﻜﻮز ﻓﻲ اﻟﺪم.
ﺛﻢ ﻣﻘﺎرﻧﮫ ﻗﺮاءة اﻟﺠﮭﺎزاﻟﺬي ﺗﻢ ﺗﺼﻤﯿﻤﮫ ﺑﻘﺮاءات ﺟﮭﺎز ﻗﯿﺎس ﺗﺮﻛﯿﺰ اﻟﺠﻠﻜﻮز ﻓﻲ اﻟﺪم ﺑﺴﺤﺐ ﻋﯿﻨﺎت ﻣﻦ اﻟﺪم؛
وﻣﻦ ﺛﻢ ﺗﻘﯿﯿﻢ اﻟﺠﮭﺎز
X
CHAPTER ONE
INTRODUCTION
1.1 General Overview
Diabetes is a chronic condition , it can usually be controlled with lifestyle changes
and medication. The main goal of treatment is to keep blood sugar levels in the
normal or near-normal range. Checking patient blood sugar is one of the best ways
to know how well his diabetes treatment plan is working. Continuous glucose
monitors have also become popular, especially for people who use an insulin
pump.
If the continuous monitoring is invasive by drawing blood it will cause pain to the
patient so this project intends to use a noninvasive glucose monitoring method to
measure the blood glucose levels. Noninvasive glucose refers to the measurement
of blood glucose levels without drawing blood, puncturing the skin or causing pain
or trauma.
1.2 Objectives
1.2.1 General objective
Designing multi-techniques noninvasive blood glucose monitoring prototype to
determine the blood glucose levels using electrical and photoacoutstic techniques.
1
four. Chapter five shows the results obtained and discussion. Chapter six deals
with conclusion and future recommendations.
CHAPTER TWO
LITERATURE REVIEWS
1. Jason J. Burmeister and Mark A. Arnold studied that The chemical and physical
properties of the tongue make it an excellent site for noninvasive glucose
measurements with first-overtone near-IR spectroscopy. Transmission through the
tongue produces effective optical path lengths through water of .5 mm, which are
required for measuring clinically relevant concentrations of glucose. Furthermore,
tongue tissue contains little fat, which permits low noise in the spectral region
where the most reliable glucose-specific information resides. Comparing the
putative sites examined here, webbing tissue contains the largest percentage of
fatty tissue and, therefore, is clearly the worst measurement site. Differences in the
amount of fat are considerably less between the remaining measurement sites.
Overall, the order of fat content (from highest to lowest) is webbing .. lower lip ..
upper lip .. cheek . . tongue . [1]
2. Orna Amir, Daphna Weinstein, Silviu Zilberman, et.al. said that In vitro
analysis was used to validate the technology and algorithms. A clinical study was
performed to quantify the in vivo performance of the NBM device. A total of 23
patients with type 1 (n = 12) and type 2 (n = 11) diabetes were enrolled in the
clinical study and participated in 111 sessions. Accuracy was assessed by
comparing NBM data with paired self-monitoring of blood glucose meter readings
In vitro experiments showed a strong correlation between calculated and actual
glucose concentrations. The clinical trial produced a total of 1690 paired glucose
values (NBM vs reference). In the paired data set, the reference glucose range was
40–496 mg/dl. No systematic bias was found at any of the glucose levels examined
(70, 100, 150, and 200 mg/dl). The mean relative absolute difference was 17.2%,
and a Clarke error grid analysis showed that 95.5% of the measurements fall within
the clinically acceptable A and B regions (zone A, 69.7%; and zone B, 25.7%).
This study indicates the potential use of OrSense’s NBM device as a noninvasive
sensor for continuous blood glucose evaluation. The device was safe and well
tolerated. [2]
2
3. R. Marbach, TH. koschinsky, F. A. gris, and H. M. heise proved that A genuine
correlation between capillary blood glucose concentration and the NIR spectrum
has been proven, and a time lag of about 10 min of the glucose concentration in the
spectroscopically probed tissue volume has been established. The mean-square
prediction error (PRESS 1/2) was estimated between 45 and 55 mg/dL with the use
of the "leave-one-out" strategy of cross-validation. An analysis of error variance
showed this limitation to be due to the reduced measurement reproducibility (i.e.,
variations in lip position and contact pressure), whereas the signal-to-noise ratio
would have allowed for a PRESS V2 value of approximately 25 mg/dL for log(l/R)
signals. Measurement irreproducibility occurring for single patients is of the same
order of magnitude as those between different individuals. The results show that
NIR spectroscopy is feasible for the noninvasive measurement of blood glucose
and may be used for building a self-monitoring device usable by diabetic patients
at home. However, for such a device, sampling reproducibility is of paramount
importance and has to be further improved if the prediction performance is to
achieve clinical acceptance. A different approach using a fiberoptic accessory may
be worth realizing. [3]
6. Arnold, Mark A., Lingzhi Liu, and Jonathon T. Olesberg Noninvasive near-
infrared spectra are collected through a fiber-optic interface attached to a thin fold
of skin on the back of an anesthetized laboratory rat while glucose levels are varied
in a controlled manner.
Glucose-specific spectral information is present within noninvasive near-infrared
spectra collected from a rat model using a transmission geometry. Apparently
functional, yet incorrect, calibration models can be generated, and the propensity to
create such false PLS calibration models calls into question the validity of past
reports. An analysis of calibration vectors can provide valuable insight into the
chemical basis of selectivity for multivariate calibration models of complex
systems. [6]
4
7. Hassan Ali. Al Naam , Mohamed Osman Idrees , Abdalsalam Awad , Ola S.
Abdalsalam and Fragoon Mohamed designed a non-invasive blood glucose
measurement tool using pulsed Photo-Acoustic Spectroscopy. A system has been
designed and implemented, as a result it was found that there is a linear
mathematical relationship between acoustic signal produced across measuring site
and blood glucose concentration. The experimental results were analyzed using
Clarke Error Grid Analysis (EGA) with 90% of the results falling into the
clinically accurate zone A. the system overall error was (2.6±6.28 mg/dl). This
study presents a new effective, simple, portable and safe method that reduces
discomfort due to the measurement procedures. [7]
8. Katsuhiko Maruo, Mitsuhiro Tsurugi and Jakusei Chin assumed that the glucose
content in dermis tissue traces the variations in blood glucose. For dermis spectra
measurements, epidermis, especially stratum corneum, acts as an interference in
skin tissue. Thus, we have developed a method for the selective measurement of
dermis tissue spectra, enabling us to obtain better quality spectra for an accurate
blood glucose assay. The selective measurement of the dermis spectra realized by
using a newly developed fiber-optic probe that consists of source and detector
optical fibers separated by 0.65 mm on a skin surface. The light path in the skin
tissue for this geometry has been simulated by a Monte Carlo method. The
simulation results show that detected light mainly interrogates dermis tissue. As
the absorbance signal of glucose in human tissue is extremely small, the quality of
the measured spectra is critical for the reliable assay. The present method for blood
glucose assay has been applied to one Type 1 diabetic. The correlation coefficient
between the blood glucose content predicted by NIR spectra and those measured
by finger-prick was 0.928 and the standard error of prediction was 32.2 mg/dL.
These results demonstrate the potential of our methodology for noninvasive NIR
blood glucose assay.[8]
9. Yang, Won S., and Yoon O. Kim studied a method and apparatus for measuring
blood glucose concentration by irradiating blood vessels with electromagnetic
radiation, where the method and apparatus uses near-infrared radiation diffuse-
reflection laser spectroscopy. This invention uses electromagnetic radiation of a
wavelength that is transmitted through the skin to the measurement region, for
example, a blood vessel. Since skin is mostly composed of water (H2 O), which
absorbs infrared radiation in nearly the entire infrared spectral range, only radiation
from a certain, narrow portion of the infrared spectral range called the "water
transmission window" is transmitted through the skin. The present invention uses
electromagnetic radiation with a wavelength of 1.3 μm- 1.8 μm radiation from a
5
semiconductor diode laser. When electromagnetic radiation of these wavelengths
irradiates the skin, light is transmitted through the skin to the blood vessel where
the light interacts with the heterogeneous components of the blood. The light
transmitted to the blood is then diffusely reflected by the blood. The reflected light
will have been modulated by the characteristic vibrations of the molecules which
are major components of blood. The reflected light is detected and provided as a
digital signal to a one-chip microcomputer. The one-chip microcomputer calculates
a blood glucose concentration from the digital signal by reference to a calibration
curve stored in the memory of the one-chip microcomputer. The one-chip
microcomputer causes the calculated blood glucose concentration to be displayed
on a digital display. [9]
6
CHPTER THREE
THEORETICAL BACKGROUND
3.1 Diabetes
Glucose is a form of sugar produced when the body digests carbohydrates
(sugars and starches). Glucose is the body's major fuel for the energy it needs.
When insulin is absent or ineffective, the blood glucose (blood sugar) level
increases. High blood glucose levels can lead to both short and long-term
problems. The normal cycle of glucose and insulin
Diabetes is the condition in which the body does not properly process food for
use as energy. Most of the food we eat is turned into glucose, or sugar, for our
bodies to use for energy. The pancreas, an organ that lies near the stomach, makes
a hormone called insulin to help glucose get into the cells of our bodies. When you
have diabetes, your body either doesn't make enough insulin or can't use its own
insulin as well as it should. This causes sugars to build up in your blood. This is
why many people refer to diabetes as “sugar.” Diabetes can cause serious health
complications including heart disease, blindness, kidney failure, and lower-
extremity amputations. Whether it’s termed type 1 (previously known as “juvenile-
onset”), type 2 (“adult-onset”), or the gestational diabetes that is a complication of
pregnancy, the end result is the same glucose may be present in the blood in
dangerously low (“hypoglycemia”) or high (“hyperglycemia”) amounts, and
without a means of measuring glucose, treatment is a dangerous guessing game of
taking pills, injecting insulin, or deciding how much and what kind of food to eat?
People who think they might have diabetes must visit a physician for diagnosis.
7
At present, no cure is available for diabetes but if patients adhere strictly to a
proper diet, exercise, medication and make frequent measurements of blood
glucose, they are able to maintain their health, and indeed, lead relatively normal
lives. If simple, inexpensive, reliable tests were available, they could make those
measurements as well and as often as required. The measurement of blood glucose
by any technique is inherently complex because of the wide range of potentially
interfering components. For a non-invasive technique, not only are there many
analytes within human blood that could interfere with the measurement, but there
are also other problems such as the variability and inhomogeneity of human skin
and the constantly changing human physiology. The invasiveness of the testing
procedure for diabetes plays a contributing role to the fact that nearly one-third of
the population with diabetes goes undiagnosed. It is for this reason that a method
for noninvasively monitoring glucose levels is highly desirable for patients with
diabetes which would allow for more frequent and possibly continuous monitoring
without the pain that is associated with current commercial glucose monitors.
According to the World Health Organization . “As per facts about diabetes, 90 to
95% of the world diabetics have diabetes type 2.They get it around age 40. It is an
alarming fact that, of late, this global epidemic has started coming up in children
too.
More alarming is that 90% of them are also obese. What is dangerous about
diabetes type 2 is that more than 15% of those do not know that they have it. It
comes up silently without showing its symptoms.
According to the World Health Organization, 170 million people have diabetes
worldwide and this number may well double by the year 2030. If a family history
of diabetes exists; there is a more likely opportunity to develop diabetes. [10]
9
3.3.1 Photo-acoustic effect
The photo-acoustic effect is a conversion between light and acoustic waves due to
absorption and localized thermal excitation. When rapid pulses of light are incident
on a sample of matter, they can be absorbed and the resulting energy will then be
radiated as heat. This heat causes detectable sound waves due to pressure variation
in the Surrounding medium.
Figure (3.1) Visualization of the photo-acoustic effect First, pulsed light that is incident on a sample is
absorbed (a) and the constituent molecules become thermally excited (b). Periodic heat flow from the
sample to the surrounding gas causes pressure waves (c) that are in turn detected by an acoustic sensor
(d).
10
3.3.3 Selection of wavelength and transducer material
Absorption of light by tissue causes localized heating, which causes expansion and
subsequent contraction, which gives rise to a PA signal. As composition of blood is
quite
complex and has various constituents, wavelength needs to be chosen judiciously.
Figure (3.2) shows how optical properties of tissue vary over the wavelength with
respect to melanin, water and oxyhaemoglobin. It can be seen that absorption by
water, melanin, and oxy-haemoglobin is less in red and NIR region. This region
extending from 600nm to 1300nm is known as optical window of biological tissue.
Also, in this region the depth of penetration of light in tissue is several millimeters.
Further, it can be seen from Figure (3.3) that glucose has highest absorption at
1037nm, but absorption of water at this wavelength his also high. One more fact
that needs to be considered is that, in wavelengths ranging from 1000nm to
1200nm absorptions for oxyhaemoglobin and deoxy-haemoglobin are different,
whereas this difference is fairly small at around 900nm. This may lead to errors as
light absorption will change with the changes in amount of blood oxygenation.
Considering the above phenomena tissue’s optical window (wavelength around
900nm) is the most suitable for glucose detection. Although glucose has relatively
lower absorption n around this wavelength, but due to minimum attenuation of
optical signal by other constituents, desired depth of penetration can be achieved
with substantial energy available for absorption by the glucose. Excitation signal
has nanosecond pulse duration and hence microphone cannot be used due to its
limited bandwidth. PZT (Lead Zirconate Titanate), being a highly sensitive
material is selected for detection of PA signal. [7]
11
Figure (3.3): Optical absorption spectra for glucose showing absorption peaks.
12
3.4 Electrical Technique
Electrical Technique is based on the flow of a low electrical current through the
skin, between an anode and cathode positioned on the skin surface. An electric
potential is applied between the anode and cathode, thus causing the migration of
sodium and chloride ions from beneath the skin toward the cathode and anode,
respectively. In particular, it is sodium ion migration that mainly generates the
current.[12] This measurement is possible because neutral molecules, such as
glucose, are extracted through the epidermis surface during this convective flow.
This flow causes interstitial glucose to be transported, collecting at the cathode,
where a traditional glucose sensor is placed to measure glucose concentration
directly. The “GlucoWatch” device (Cygnus Inc., Redwood City, CA) is based on
this technology, and it was approved by the US Food and Drug Administration.
The device collects glucose molecules through the cathode disk and measures the
amount with a sensor that contains enzyme glucose oxidase. Blood glucose
concentration is predicted by comparing the premeasured blood glucose value with
the signal generated by glucose molecules collected at the cathode. However, this
product was withdrawn from the market due to poor accuracy, skin irritation, and
long procedural problems. [13]
The advantage of this technology is that the electrodes are easily applied to the
skin, by which a physiologically relevant fluid sample is collected, in that there is a
correlation between glucose concentration in the physiological fluid and glucose
concentration in blood.
While electrical technology has great potential, the only device ever marketed
using it had such serious practical drawbacks that it was withdrawn from the
market. First, the electrodes irritated the skin. Second, the electrodes needed to be
in place for at least 60 minutes, which exceeded the patience of many users. Third,
readings were inaccurate, especially when the subject was sweating. Fourth, it was
not able to detect rapid changes in blood glucose, due to its long “wakeup” time.
[14]
13
CHAPTER FOUR
Material and Methodology
The designed system was developed and simulated for the purpose of measuring
glucose concentration using Photo-Acoustic and Electrical techniques.
In first technique- photoacoustic technique the NIR signal will act as a heating
source to the tissues and cause it to expand and contract according to the heating
pulse rate, which will consequently produce a pressure wave. The acquisition
system will be responsible for detecting the pressure wave. The acquired signal
will be preprocessed and prepared to become a proper input for the
microcontroller. The microcontroller will calculate the equivalent glucose
concentration according to the amplitude of the detected pressure wave.
In the second technique- electrical the ECG electrodes is there, to detect the
reverse iontophoresis signal due to the interaction of the low current with human
tissue. The acquired signal will be preprocessed and prepared to become a proper
input for the microcontroller. The microcontroller will process the signal and
calculate the equivalent glucose concentration.
Finally glucose values will be matched into one appropriate value and the final
value will be displayed to the user.
PREPROCESSI
LCD PROCESSOR
NG
IR Source
650nm Laser sensor Module 6mm 5V 5mW Red Laser Dot Diode Copper Head
Module Details
1\ Laser sensor Module
2\ Operating voltage 5V
3\ Output wavelength 650 nm
4\ 3 pins module
-Pin1 = Vdc
-Pin2 = Ref
-Pin3 = GND
15
Figure (4.2): Laser Dot Diode.
Measurement champer
It is a part of the components of the circuit used and it's made from (Teflon). It
occurs to generate a sound wave as we enter the finger in the cuvette and shine of
Pulses of a 650nm near-IR laser light. It occurs absorption in human tissue is
excited at the wavelength absorbed by the molecules of glucose absorption will
cause the local heating. Increase in temperature causes a rapid thermal expansion,
which generates a pressure wave.
Characteristics
From the specific characteristics of Teflon is that it:
1. A green waxy substance.
2. Strong resistance to chemical acid.
3. Smooth surface.
4. It retains physical properties unchanged at temperatures ranging from 260 - to
240 Silesia.
5. Insulating material for electricity.
6. To dissolve in organic solvents.
7. A flexible material.
16
Figure (4.3): Measurement champer.
Transducer (microphone)
Audio transducer that convert an acoustic wave generated by pulsed laser which
interact with the tissue into electrical signal.
17
Figure (4.4): Microphone.
Operation amplifier
The 741 is the godfather of all operational amplifiers (amplifiers on a chip).
Although most up-to-date designs beat it for speed, low noise, etc. It still works
well as a general purpose device. One of its advantages is that it is compensated
(its frequency response is tailored) to ensure that under most circumstances it won't
produce unwanted spurious oscillations.
18
Figure (4.5): 741 Amplifier.
19
LCD
The LCD screen consist of two lines with 16 characters, each character consist of
(5×8 or 5×11 dot matrix). Display contrast depends on the power supply voltage
whether messages are displayed in one or two lines. For this reason, varying
voltage 0-5volt is applied on the pin marked as Vee. In this project Vee had been
set to zero.
4.2.1 Simulation
The validity of the design had to be checked. Simulation is a useful mean of
validating electronic circuits. A version of PROTEUS 7.7 had been used as a
simulation tool.
20
Figure (4.8): Whole photoacoustic circuit simulation.
4.2.2 Implementation
The circuit applied as in figure (4.9) shows the implemented prototype of the
design
21
Figure (4.9) : Photoacoustic circuit layout.
22
Block Diagram
CURRENT MASUREMEN
ELECTRODE
SOURCE T
PREPROCESSI
LCD PROCESSOR
NG
Current source
We design a voltage to current convertor circuit to give us low current range
(5.1mA).
23
Figure (4.11): Low current source.
24
Electrodes ( ECG Electrodes)
For electrical systems which use alternating current, the electrodes are the
connections from the circuitry to the object to be acted upon by the electric current
but are not designated anode or cathode because the direction of flow of the
electrons changes periodically, usually many times per second.
4.3.1 Simulation
The validity of the design had to be checked. Simulation is a useful mean of
validating electronic circuits. A version of PROTEUS 7.7 had been used as a
simulation tool.
25
Figure (4.13): Whole electrical circuit simulation.
4.3.2 Implementation
The circuit applied as in figure (4.14) shows the implemented prototype of the
design
26
Figure (4.14): Electrical circuit layout.
27
CHAPTER FIVE
RESULTS AND DISCUSSION
5.1 Results
To design the device firstly, we measure the output voltage from the transducer
and compare it with invasive blood glucose concentration.
Figures (5.1) and (5.2) show the relationship between voltage and glucose
concentration on 10 persons by using the two techniques.
Table (5.1): The results of test for persons by using photoacoustic technique.
28
Gulcose
200
44.478x + 0.8084y =
180
0.9804= ² R
160
140
120
1Series
100
( 1Linear (Series
80 ( 1Linear (Series
60
40
20
0
0 20 40 60 80 100 120 140 160
Figure (5.1): Relationship between voltage and glucose concentration for 10 persons by using
photoacoustic technique.
29
Table (5.2): The results of test for persons by using electrical technique.
30
Gulcose
200
22.921x - 28.246y =
180 0.9862= ² R
160
140
120
100 1Series
( 1Linear (Series
80
60
40
20
0
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8
Figure (5.2): Relationship between voltage and glucose concentration for 10 persons by using
electrical technique.
The circuits was designed and tested on 10 volunteers in the age range of (17-44);
Table (5.3) shows the final result for measuring glucose concentration by the
circuits we designed compared with non-invasive method.
Table (5.3): Final result for measuring glucose concentration by both Invasive and Noninvasive
methods:
Noninvasive by using Noninvasive by using Invasive (mg\dl)
electrical (mg\dl) photoacoustic (mg\dl)
61.817 69.37 67
71.42 72.078 71
84.41 75.56 77
87.8 78.153 82
90.34 82.58 89
95.71 96.123 98
109.83 126.42 117
120 127.924 122
143.73 147.51 143
188.92 178.22 188
31
5.2 Discussion
The Clarke error grid approach is used to assess the clinical significance of
differences between the glucose measurement technique under test and the venous
blood glucose reference measurements. The method uses a Cartesian diagram, in
which the values predicted by the technique under test are displayed on the y-axis,
whereas the values received from the reference method are displayed on the x-axis.
The diagonal represents the perfect agreement between the two, whereas the points
below and above the line indicate, respectively, overestimation and
underestimation of the actual values. Zone A (acceptable) represents the glucose
values that deviate from the reference values by ±20% or are in the hypoglycemic
range (<70 mg/dl), when the reference is also within the hypoglycemic range. The
values within this range are clinically exact and are thus characterized by correct
clinical treatment. Zone B (benign errors) is located above and below zone A; this
zone represents those values that deviate from the reference values, which are
incremented by 20. The values that fall within zones A and B are clinically
acceptable, whereas the values included in areas C-E are potentially dangerous,
and there is a possibility of making clinically significant mistakes. A total of 10
persons were tested during the trial period, producing 10 data pairs. Using Clarke
error grid analysis, 100% of the readings fell in the clinically acceptable zones A
and B, with 90% in the A zone, and 10% in B zone.
32
Clarke's Error Grid Analysis
400
E C B
350
300
Concentracion Predicha [mg/dl]
250
B
200
150 D
D
100
50
A C E
0
0 50 100 150 200 250 300 350 400
Reference Concentration [mg/dl]
33
Clarke's Error Grid Analysis
400
E C B
350
300
Concentracion Predicha [mg/dl]
250
B
200
150 D
D
100
50
A C E
0
0 50 100 150 200 250 300 350 400
Reference Concentration [mg/dl]
34
CHAPTER SIX
6.1 Conclusions
This study proposed a noninvasive measurement system for glucose concentration;
using the photoacoustic and electrical techniques. The equations relating glucose
concentration with the two techniques was derived from experiments. The system
design was simulated using PROTEUS 7.7 and Matlab 7.12 was used in the EGA
evaluation. The design has been implemented and results were analyzed.
6.2 Recommendations
1. Increasing the number of experimental results from 10 to few hundred
including hypoglycemic and hyperglycemic readings, to provide better
evaluation of the photoacoustic device.
2. Laser used should be 905nm wavelength.
35
References
[1] Burmeister, Jason J., and Mark A. Arnold. "Evaluation of measurement sites for noninvasive
blood glucose sensing with near-infrared transmission spectroscopy." clinical chemistry 45.9
(1999): 1621-1627.
[2] Amir, Orna, et al. "Continuous noninvasive glucose monitoring technology based on
“occlusion spectroscopy”." (2007): 463-469.
[3] Marbach, Ralf, et al. "Noninvasive blood glucose assay by near-infrared diffuse reflectance
spectroscopy of the human inner lip." Applied Spectroscopy 47.7 (1993): 875-881.
[4] Alarousu, Erkki, et al. "Noninvasive glucose sensing in scattering media using OCT, PAS,
and TOF techniques." Proceedings of SPIE. Vol. 5474. 2003.
[5] Alexeeva, Natalia V., and Mark A. Arnold. "Near-infrared microspectroscopic analysis of rat
skin tissue heterogeneity in relation to noninvasive glucose sensing." (2009): 219-232.
[6] Arnold, Mark A., Lingzhi Liu, and Jonathon T. Olesberg. "Selectivity assessment of
noninvasive glucose measurements based on analysis of multivariate calibration vectors."
(2007): 454-462.
[7] Al Naam, Hassan Ali, et al. "Non invasive blood glucose measurement based on Photo-
Acoustic Spectroscopy." Computing, Control, Networking, Electronics and Embedded Systems
Engineering (ICCNEEE), 2015 International Conference on. IEEE, 2015.
[8] Maruo, Katsuhiko, et al. "Noninvasive blood glucose assay using a newly developed near-
infrared system." IEEE Journal of selected topics in quantum electronics 9.2 (2003): 322-330.
[9] Yang, Won S., and Yoon O. Kim. "Non-invasive method and apparatus for measuring blood
glucose concentration." U.S. Patent No. 5,267,152. 30 Nov. 1993.
[10]Patel CKN & Tam AC (1981) Pulsed optoacoustic spectroscopy of condensed matter.
Reviews of Modern Physics 53(3): 517-550.
[11] Amos AF, McCarty DJ & Zimmett P (1997) The rising global burden of diabetes and its
complications: estimates and projections to the year 2010. Diabetic Medicine, Supplement 5: S1-
S5.
[12] Sieg A, Guy RH, Delgado-Charro MB. Noninvasive glucose monitoring by reverse
iontophoresis in vivo: application of the internal standard concept. Clin Chem. 2004;50(8):1383–
1390
[13] The Diabetes Control and Complications Trial Research Group. The effect of intensive
treatment of diabetes on the development and progression of long-term complications in insulin-
dependent diabetes-mellitus. N Engl J Med 1993;329:977–86.
[14] Recent advances in noninvasive glucose monitoring Chi-Fuk So1 Kup-Sze Choi1 Thomas
KS Wong2 Joanne WY Chung2,3 Appendices
36
APPENDIX
Clarke’s error grid function
37
% [1] A. Maran et al. "Continuous Subcutaneous Glucose Monitoring in Diabetic Patients"
% Diabetes Care, Volume 25, Number 2, February 2002
% [2] B.P. Kovatchev et al. "Evaluating the Accuracy of Continuous Glucose-Monitoring
Sensors"
% Diabetes Care, Volume 27, Number 8, August 2004
%
% Edgar Guevara Codina
% codina@REMOVETHIScactus.iico.uaslp.mx
% File Version 1.1
% November 18 2008
%
% Ver. 1.1 corrected upper B-C boundary, lower B-C boundary slope ok
%
% MATLAB ver. 6.5.0.180913a (R13)
% -------------------------------------------------------------------------
% Error checking
if nargin==0
error('clarke:Inputs','There are no inputs.')
end
if length(yp) ~= length(y)
error('clarke:Inputs','Vectors y and yp must be the same length.')
end
if (max(y) > 400) || (max(yp) > 400) || (min(y) < 0) || (min(yp) < 0)
error('clarke:Inputs','Vectors y and yp are not in the physiological range of glucose
(<400mg/dl).')
end
% ---------------------- Determine Screen Size ----------------------------
scrsz = get(0,'ScreenSize')-[0 -5 0 72];
% ---------------------- Determine data length ----------------------------
n = length(y);
% ---------------------- Plot Clarke's Error Grid -------------------------
figure
h = plot(y,yp,'ko','MarkerSize',4,'MarkerFaceColor','k','MarkerEdgeColor','k');
xlabel('Reference Concentration [mg/dl]');
ylabel ('Predicted Concentration [mg/dl]');
ylabel ('Concentracion Predicha [mg/dl]');
title('Clarke''s Error Grid Analysis');
set(gca,'XLim',[0 400]);
set(gca,'YLim',[0 400]);
axis square
hold on
plot([0 400],[0 400],'k:') % Theoretical 45º regression line
plot([0 175/3],[70 70],'k-')
plot([175/3 320],[70 400],'k-')
plot([70 70],[84 400],'k-')
38
plot([0 70],[180 180],'k-')
plot([70 290],[180 400],'k-') % Corrected upper B-C boundary
plot([70 70],[0 175/3],'k-')
plot([70 400],[175/3 320],'k-')
plot([180 180],[0 70],'k-')
plot([180 400],[70 70],'k-')
plot([240 240],[70 180],'k-')
plot([240 400],[180 180],'k-')
plot([130 180],[0 70],'k-') % Lower B-C boundary slope OK
text(30,20,'A','FontSize',12);
text(30,150,'D','FontSize',12);
text(30,380,'E','FontSize',12);
text(150,380,'C','FontSize',12);
text(160,20,'C','FontSize',12);
text(380,20,'E','FontSize',12);
text(380,120,'D','FontSize',12);
text(380,260,'B','FontSize',12);
text(280,380,'B','FontSize',12);
set(gcf, 'color', 'white'); % sets the color to white
%print(gcf,'-dmeta','Clarke_EGA'); % Saves plot as a Enhanced MetaFile
total = zeros(5,1); % Initializes output
percentage = zeros(5,1); % Initializes output
for i=1:n,
if (yp(i) <= 70 && y(i) <= 70) || (yp(i) <= 1.2*y(i) && yp(i) >= 0.8*y(i))
total(1) = total(1) + 1; % Zone A
else
if ( (y(i) >= 180) && (yp(i) <= 70) ) || ( (y(i) <= 70) && yp(i) >= 180 )
total(5) = total(5) + 1; % Zone E
else
if ((y(i) >= 70 && y(i) <= 290) && (yp(i) >= y(i) + 110) ) || ((y(i) >= 130 && y(i) <=
180)&& (yp(i) <= (7/5)*y(i) - 182))
total(3) = total(3) + 1; % Zone C
else
if ((y(i) >= 240) && ((yp(i) >= 70) && (yp(i) <= 180))) || (y(i) <= 175/3 && (yp(i) <=
180) && (yp(i) >= 70)) || ((y(i) >= 175/3 && y(i) <= 70) && (yp(i) >= (6/5)*y(i)))
total(4) = total(4) + 1;% Zone D
else
total(2) = total(2) + 1;% Zone B
end % End of 4th if
end % End of 3rd if
end % End of 2nd if
end % End of 1st if
end % End of for loop
percentage = (total./n)*100;
39