ED-CAL PLUS

HEMATOLOGY CALIBRATOR
CALIBRATOR
ASSAY VALUES AND EXPECTED RANGES
LOT PLUS1119

2019-12-05
Instrument Parameter Reference Value Deviation

DS-500, WBC ×109/L 8.96 ±0.20


DS-500i RBC ×1012/L 4.30 ±0.08

HGB g/L 136 ±2

MCV fL 90.9 ±2.0

PLT ×109/L 265 ±12

H30 WBC ×109/L 9.3 ±0.3

RBC ×1012/L 4.55 ±0.06


HGB g/L 140 ±1
MCV fL 85.3 ±0.3

PLT ×109/L 245 ±11

MPV fL 11.0 ±0.2

H50 WBC ×109/L 9.3 ±0.5


RBC ×1012/L 4. 26 ±0.10
HGB g/L 126 ±3
MCV fL 84.1 ±0.5
PLT ×109/L 277 ±10
MPV fL 9.6 ±0.2

【NOTE】
1. The calibrator should be stored in refrigerator (2~8℃). After opening, it will keep stable for 7 days when it is stored airtight at 2~8℃.

2. Before mixing and running the calibrator after taking it out from the refrigerator, please keep it at least 15 minutes until reaching room
temperature (15~30℃).
3. Calibrator must be well mixed before using. Please mix gently, to avoid cells rupture and/or generating bubbles.

4. After using, put the calibrator back into the refrigerator to prevent contamination and evaporation.

P/N: 01.54.458222
MPN:01.54.458222010